laerdal Reusable, Serres User Manual

www.laerdal.com
Laerdal Suction Unit
Reusable & Serres
User Guide
EN
LSU Reusable Cat.no 78 00 00
3
Contents
Intended Use 4
Important Information 5
Overview 7
LSU Reusable 7
LSU Serres 8
User Interface 9
Assembly 10
LSU Reusable Canister System 10
LSU Serres Suction Bag System 11
Use 12
Emptying LSU Reusable 14
Emptying LSU Serres 15
Cleaning 16
Reusable Parts 16
Serres 18
LSU Cabinet 18
Device Test 19
Service and Maintenance 24
Battery 25
Troubleshooting 28
Specications 29
Accessories and Parts 35
Warranty 37
4
Intended Use
The Laerdal Suction Unit (LSU) is a portable, electrically powered, medical suction equipment intended for field and transport use. It is intended for intermittent operation to remove secretions, blood or vomit from a patient´s airway to allow ventilation.
Higher vacuum levels are generally selected for oropharyngeal suctioning, and lower vacuum levels are usually selected for tracheal suctioning and the suctioning of children and infants.
5
Important Information
This User Guide covers two configurations of the LSU: Reusable and Serres Suction Bag System. Unless otherwise specified, the information in this User Guide applies to both configurations.
Read this User Guide carefully, and become thoroughly familiar with the operation and maintenance of the LSU before using it. Read all Cautions and Warnings before using the LSU.
Warnings and Cautions
A Warning states a condition, hazard, or unsafe practice that can result in serious personal injury or death. A Caution states a condition, hazard, or unsafe practice that can result in minor personal injury or damage to the product.
Warnings
• The LSU is not suitable for use in the presence of flammable liquids or gases; there can be a danger of explosion or fire.
• Not intended for use in MRI environments.
• Do not use the LSU under environmental conditions that are outside the ranges specified in Specifications section. This can endanger safety and adversely affect operation of the device.
• Do not block the Exhaust Outlet during use. This will lead to reduced flow and can also cause damage to the LSU.
• The LSU Reusable must not be used without the Aerosol Filter or the Float Ball.
• Disconnect the LSU from external power prior to cleaning. Use a minimum amount of liquid to prevent any electrical shock hazard.
• Do not immerse the LSU or allow it to stand in water or other liquids. This might damage the device, and cause electrical hazard.
Cautions
• Do not pump any cleaning solution or other liquids through the vacuum pump, i.e. through the Vacuum Connector. This can damage the LSU.
• Use only parts and accessories supplied by Laerdal Medical or one of our authorised distributors to ensure that the LSU operates satisfactorily.
• Overflow of suctioned material can damage the device. If overflow of liquid from the Canister into the pump is suspected, contact your local Laerdal Medical representative.
• The LSU should only be used by persons trained in the use of medical suction equipment.
6
Battery
Cautions
• Only use batteries approved by Laerdal Medical. Other batteries will have problems related to the battery status indicator of the LSU, the batter y operation time and safety.
• The LSU must be charged between each clinical use.
• To maintain satisfactory operation of the battery, placing the LSU on continuous charge immediately after use and when not in use is recommended.
• If it is not possible to place the LSU on continuous charge when not in use, make sure the battery is charged for a minimum of 4 hours at least once a month.
• The LSU must be placed on charge for a minimum of 4 hours to reach full battery capacity. Fast charging gives approximately 80% battery capacity after 3 hours (for a new battery). Repetitive 3 hour charging is not recommended.
• Fully charging the battery is recommended. Repetitive charging to a lower capacity level will reduce battery life.
• Always fully charge the battery before storage.
• Do not store the battery when it is discharged.
• Do not store the LSU with a discharged battery.
• Laerdal recommends charging a spare battery every 6 months when stored in room temperature at 25 °C (77 °F).
Important Information
7
Vacuum Inlet
Aerosol Filter
Patient Inlet
User Interface
Canister Holder
Release Arm
Water Bottle
Battery Door
Canister
Float Ball
Vacuum Seal
Patient Suction Tubing
Holder for Water Bottle/Patient Suction Tubing.
Gasket
Canister Lid
Quick Reference
Power Inlet (on the back)
Angled Connector
Suction Catheter Adapter
LSU Reusable
Overview
8
Overview
Vacuum Inlet
Serres Vacuum Connector
Canister
Battery Door
Release Arm
Canister Holder
Holder for Suction Tubing
Suction Tube
Serres Suction Bag with filter
Patient Inlet
User Interface
Quick Reference
LSU Serres
9
User Interface
*Pressure conversion chart
mmHg 80 120 200 350 500
kPa 10.6 16.0 26.6 46.6 66.5
mBar 107 160 267 467 667
Test Button
Run the Device Test program
Vacuum Indicator*
Actual vacuum level during operation. Each fully lit segment represents 50 mmHg. If a segment has a dim light, this represents 25 mmHg (e.g 125 mmHg is shown by 2 fully lit and 1 dimly lit segments).
Battery Status Indicator
Operating Knob
On/Off switch Vacuum selector
Failure Mode Indicator
Possible malfunction detected
External Power Indicator
AC or DC power connected
Power On Indicator
Continuously lit - LSU is
switched on.
Rapid flash (approx.
twice per second) ­Device Test in progress.
Slow flash (approx.
once per second) ­Automatic power save mode activated; Device Test interrupted or battery discharged.
Overview
1 2
3
4
5
6
10
Assembly
LSU Reusable Canister System
CLICK
CLICK
A
B
C
D
E
A
B
1 2
3 4
11
Assembly
LSU Serres Suction Bag System
Note
Turn the operating knob to 500+ mmHg. Block the Patient Inlet with your thumb and simultaneously push the lid down. The lid is properly installed when vacuum has reached 500 mmHg. Release the Patient Inlet and make sure that the bag is fully inflated.
CLICK
A
B
C
D
1
2
12
Before Use Checklist
• Check that no parts are missing and that all parts are clean.
• To operate the LSU from an external power source, connect to either external AC or DC power. To operate the LSU from the internal battery, check that the battery is installed.
• Run Device Test.
LSU Reusable:
• Check that the Patient Suction Tubing is securely connected to the Patient Inlet on the Canister Lid; and the Aerosol Filter is securely fastened in the LSU and the lid.
• Check that a suction catheter is attached to the patient suction tube or suction adapter. Do not use the suction tube or suction adapter without a suction catheter attached.
LSU Serres:
• Check that the Serres Vacuum Connector is properly connected to the LSU and to the canister.
• Check that the lid of the canister is properly sealed: block patient inlet and turn on the LSU. Vacuum will build up when all parts are assembled correctly.
• Attach appropriate suction catheter if necessary. (Not supplied by Laerdal Medical).
Use
1. Unwind the Suction Tubing. Set the Operating Knob to the required vacuum level.
The LSU will switch on and start to operate. The Power On Indicator lights up during operating.
2. When suction is complete, set the Operating Knob to “0” to turn off.