READ OPERATION MANUAL BEFORE USING EQUIPMENT. FAMILIARIZE
YOURSELF WITH CONTROLS AND ACCESSORIES.
ALWAYS DISCONNECT EQUIPMENT FROM POWER SOURCE BEFORE
SERVICING OR PERFORMING ANY MAINTENANCE PROCEDURES.
WHEN OPERATING THIS EQUIPMENT, ALWAYS WEAR PROTECTIVE CLOTHING,
GLASSES AND OTHER ACCESSORIES AS SPECIFIED BY THE SAFETY
REGULATIONS OF YOUR LABORATORY OR INSTITUTION.
WE RECOMMEND THAT YOU DO NOT LEAVE EQUIPMENT UNATTENDED WHILE
IT IS IN OPERATION. DO NOT OPERATE EQUIPMENT WITH A DAMAGED
ELECTRICAL CORD OR ALLOW CORD TO COME INTO CONTACT WITH A HOT
SURFACE.
DO NOT MODIFY CONSTRUCTION AND/OR ASSEMBLY OF EQUIPMENT. KEEP
GUARDS IN PLACE. DO NOT INSERT FINGERS INTO EQUIPMENT WHEN IT IS
OPERATING. DO NOT REMOVE LABELS, DECALS OR OTHER INFORMATION
CONCERNING OPERATION OR SAFETY FROM THE UNIT.
MAKE SURE ALL VESSELS ARE SECURELY CLAMPED BEFORE TURNING ON
UNIT. TO ELIMINATE HAZARD OF ELECTRICAL SHOCK, MAKE SURE FLOOR
AROUND MACHINE IS DRY. IN THE EVENT OF ACCIDENTAL SPILLING OR
SPLASHING OF LIQUIDS, CLEAN UP AND/OR NEUTRALIZE THE SPILLED LIQUIDS
BEFORE CONTINUING.
DO NOT USE EQUIPMENT FOR OTHER THAN ITS INTENDED PURPOSE. USE
ONLY THE ACCESSORIES AND ATTACHMENTS THAT ARE SHIPPED WITH THE
EQUIPMENT OR ARE SPECIFIED FOR IT. SUBSTITUTING OTHER ATTACHMENTS
OR ACCESSORIES CAN PRODUCE HAZARDS OR MAKE THE UNIT INOPERATIVE.
DO NOT OPERATE SHAKER FOR EXTENDED PERIODS UNDER HIGH HUMIDITY
CONDITIONS. DO NOT OPERATE SHAKER IN AMBIENT TEMPERATURES
EXCEEDING 40ºC. DO NOT OPERATE EQUIPMENT IN AN EXPLOSIVE
ATMOSPHERE. DO NOT OPERATE SHAKER IN A CO2 ENRICHED ATMOSPHERE.
PERFORM REGULAR MAINTENANCE SERVICE AS SPECIFIED IN THIS MANUAL
AND KEEP UNIT IN GOOD REPAIR. DO NOT OPERATE WITH KNOWN DEFECTS.
SECTION 3
5
Page 6
DESCRIPTION
Lab-Line Models 4625, 4625JPN and 4625-1 Titer Plate Shakers are
sturdy, heavy- duty units designed to produce an accurate 0.120-inch orbital
shaking motion that is ideal for many kinds of laboratory work with titer plates.
The shakers are identical except that Model 4625 is designed for a 120V,
60 Hz power source, Model 4625JPN for a 100V 50/60 Hz power supply and
Model 4625-1 is designed for 240V, 50 Hz power.
Controls are clearly marked, self-explanatory and make the shaker easy to
use and operate.
Users can operate the Titer Plate Shaker at a range of speeds from 40 to
1100 rpm to produce shaking movement ranging from gentle rotation to vigorous
agitation to meet specific user requirements.
For increased versatility, a rocker switch on the control panel enables users
to select either a constant shaking motion or a timed cycle up to 5 minutes for
accurate, repetitive runs.
The ribbed platform provides a non-skid surface and retaining springs
which act to hold the titer plates securely in place. In addition, rubber feet at the
corners hold the shaker firmly on a counter surface and prevent the shaker from
“walking.”
motor features high torque, full ballbearing construction, and ceramic
permanent magnets. Brushes are user
replaceable.
SHAKING SPEED: Speed range is 40 to 1100 rpm.
TIMER:Range: Up to 5 minutes in 1 minute
increments or continuous run operation.
SHAKER MOTION: 0.120-inch orbit
SECTION 4
DIMENSIONS: 11" W x 11" D x 7" H (28 x 28 x 18 cm)
SHIPPING WEIGHT:25 lbs. (11 kg.)
UNIT’S ENVIRONMENTAL OPERATING CONDITIONS:
POLLUTION DEGREE:2
INSTALLATION CATEGORY:II
ALTITUDE:2000 Meters MSL (Mean Sea Level)
HUMIDITY:80% maximum, non-condensing
ELECTRICAL SUPPLY:120VAC or 240VAC
VOLTAGE TOLERANCE:±10% of normal rated line
TEMPERATURE: 15ºC to 40ºC
PRODUCT USAGE:This product is intended for use indoors only
SECTION 5
7
Page 8
√SHIPPING CARTON:
This should be inspected upon delivery. When received, carefully
examine for any shipping damage before unpacking. If damage is discovered,
the delivering carrier should both specify and sign for the damage on your copy
of the delivery receipt.
Open the carton carefully making certain that all parts are accounted for
before packaging materials are discarded—after unpacking, if damage is found,
promptly report it to the carrier and request a damage inspection promptly.
IMPORTANT: Failure to request an inspection of damage within a few
days after receipt of shipment absolves the carrier from any liability for damage:
you must call for a damage inspection promptly.
LEVELING:
Place the shaker on a level surface in a convenient location and near a
properly grounded electrical outlet that matches the requirements of the unit as
indicated on the nameplate. A minimum of ½-inch clearance is required on all
sides of the unit to provide unobstructed rotation.
ELECTRICAL REQUIREMENTS:
120 VAC models require a 120 VAC, 50/60 Hz power source. They are
supplied with a 3-wire line cord. It should be plugged into an outlet designed for
3-prong plugs. If an extension cord is used, it also should be the 3-wire grounded
type. For an outlet designed to accept 2-prong plugs (ungrounded), it is
required that a qualified electrician replace the outlet with a new grounded type.
240 VAC models require a 240 VAC, 50/60 Hz power source. Because of
the variety of plug configurations in use worldwide for 240 VAC power, the unit is
furnished with the plug removed. The user must install a plug to conform with
local code and configuration requirements.
100 VAC models require a 100 VAC, 50/60 Hz power source. Because of
the variety of plug configurations in use worldwide for 100 VAC power, the unit is
furnished with the plug removed. The user must install a plug to conform with
local code and configuration requirements.
If a plug must be installed, use only the 3-prong grounded type, rated for
the unit load requirements and matching the power outlet. Make sure the green
ground wire is secured to the plug ground terminal.
INSTALLATION
NOTE: LEAVE THE SHAKER DISCONNECTED WHEN NOT IN USE.
SECTION 6
8
Page 9
DANGER: DO NOT USE IN THE PRESENCE OF FLAMMABLE OR COMBUSTIBLE MATERIALS OR EXPLOSIVE GASES. DO NOT USE IN THE PRESENCE
OF PRESSURIZED OR SEALED CONTAINERS. FIRE OR EXPLOSION MAY
RESULT, CAUSING DEATH OR SEVERE INJURY.
NOTE: DEPRESS METAL TAB PRIOR TO ROTATING KNOB. THE TAB FIXES KNOB
SETTING AND PREVENTS INADVERTENT MOTION OF THE KNOB.
Located on the front panel are easily accessible controls, from left to right:
speed control, timer and selector switch.
The speed control is easy to operate: rotate knob clockwise to increase
shaking speed and counterclockwise to decrease the speed.
The timer permits setting a timed shaking period from approximately 5
seconds to 5 minutes. Each timed cycle is exactly repeatable until a new timed
period is selected.
The selector switch allows the operator to select either a "Timed" cycle
or a "Constant" run position. The center position is the off position.
CAUTION: BEFORE PROCEEDING MAKE CERTAIN THAT SELECTOR
SWITCH IS IN THE OFF POSITION AND SPEED CONTROL IS AT THE ZERO
POSITION.
*AS LISTED ON REPLACEMENT PARTS LIST
OPERATION: (Con’t)
LOADING THE PLATFORM:
9
Page 10
Insert a filled titer plate in the spring-tensioned holder by pushing the plate
against the springs until the titer plate is resting on the ribbed bottom of the
platform and is securely locked in position.
STARTING/STOPPING THE SHAKER:
•For continuous shaking action, push the selector switch to the
CONSTANT position.
•In the event a timed period of shaking is desired, push selector switch to
TIMED. Next, turn the black arrow on the timer dial to the setting
corresponding to desired timing period. Press center button on timer to
start the timing period. The shaker shuts off automatically at the end of
the timed cycle. During the cycle the white dot on the dial will rotate to
indicate elapsed time.
•During either the CONSTANT or TIMED shaking periods, the operator can
stop the shaker by simply pushing selector switch to the center or OFF
position.
SETTING THE SPEED CONTROL:
After starting the shaker, turn the speed control clockwise to the desired
level of shaking action.
It is not necessary to return the speed control to the start or zero position
after completion of each shaking cycle. Once a satisfactory shaking speed is
found, the control may be left at that setting.
To stop the shaker, turn the switch to OFF.
If the shaker will not be used again soon, disconnect the plug from its
electrical outlet.
POWER MAINS PROTECTION:
If a fuse or circuit breaker blows, have a qualified person replace them
with a properly rated fuse or circuit breaker—these now follow here: The motor
lead circuit breaker is: 1 AMP, ETA #44-100-P10 (Lab-Line Part #330-158-00);
The speed control fuse is: 4 AMP, Littlefuse #313.004, Slo-Blo Type (Lab-Line
Part #330-371-00).
SECTION 7
MAINTENANCE
10
Page 11
NOTE: MAKE NO ATTEMPT TO SERVICE OR REPAIR A LAB-LINE PRODUCT UNDER
WARRANTY BEFORE CONSULTING YOUR LAB-LINE DEALER. AFTER THE WARRANTY
PERIOD, SUCH CONSULTATION IS STILL ADVISED, ESPECIALLY WHEN THE REPAIR
MAY BE TECHNICALLY SOPHISTICATED OR DIFFICULT.
IF ASSISTANCE IS NEEDED BEYOND WHAT THE DISTRIBUTOR CAN PROVIDE, PLEASE
CALL THE LAB-LINE CUSTOMER RELATIONS DEPARTMENT AT (319) 556-2241 OR
(800) 522-5463. NO MERCHANDISE, HOWEVER, SHOULD BE RETURNED DIRECTLY TO
LAB-LINE WITHOUT PRIOR APPROVAL FROM LAB-LINE.
CAUTION: DISCONNECT PLUG FROM ELECTRICAL OUTLET BEFORE
ATTEMPTING ANY MAINTENANCE OR REPAIR OF THIS UNIT.
MAINTENANCE-FREE DESIGN:
The shaking mechanism is equipped with permanently lubricated bearings
and the unit is basically maintenance-free. The only parts that require periodic
inspection are the silicone rubber tubing sections under the platform and the
motor brushes.
Silicone tubing sections should be inspected periodically. If any cracking
is evident, replace the part.
REPLACING PARTS:
WARNING: TO AVOID RISK OF ELECTRIC SHOCK, DISCONNECT
EQUIPMENT FROM POWER SOURCE BEFORE REMOVING THE BOTTOM
COVER OR PERFORMING ANY SERVICE.
SPEED CONTROL:
Replacement of the speed control requires removal of the bottom cover
casting. Turn unit over and remove 6 screws holding cover and lift off.
To replace speed control, remove the control knob (2 setscrews) and a
retaining nut to separate it from the panel. Unplug the terminal connections from
the old terminal unit and plug them into the new one. Reverse this sequence to
put the control unit back into position.
The new control unit must be calibrated in the following manner. Locate
the maximum and minimum speed potentiometers on the speed control. Lay the
unit on its side and elevate the housing so that the titer plate platform will clear
the table to permit an unobstructed area of shaking motion.
MAINTENANCE: (Con’t)
SPEED CONTROL: (Con’t)
Start the unit and turn the speed control knob fully clockwise (maximum
speed). Using a suitable tachometer and a small plastic screwdriver, adjust the
maximum speed potentiometer to 1100 rpm.
11
Page 12
Next, rotate the speed control fully counterclockwise (minimum speed).
Now adjust the "minimum" potentiometer to the point at which platform stops
moving.
If the procedure as outlined has been followed carefully, the new speed
control is now calibrated.
Replace bottom cover casting and insert and securely tighten the 6
screws that were previously removed.
SILICONE TUBING SECTIONS:
To replace the silicone tubing sections on the platform, remove the 4
screws holding the platform and lift off. Silicone rubber tubing is available from
most laboratory supply houses or from Lab-Line. Use ¼" ID x 7/16" OD and cut to
1-5/16" length.
Reassemble the unit carefully by following the reverse order from the
preceding.
CAUTION: DO NOT OPERATE SHAKER WITHOUT PLATFORM. TO DO SO
WILL PRODUCE A SERIOUS IMBALANCE AND CAUSE THE ENTIRE UNIT TO
WALK. IN ADDITION, EXPOSED ROTATING PARTS WILL BE A HAZARD TO
FINGERS.
MOTOR BRUSHES:
Motor brushes should be checked every three months for wear. Turn the
unit upside down and remove bottom casting to expose the motor. 2 brushes are
located under plastic caps on opposite sides of the lower part of the motor.
Unscrew the plastic caps and slide the brushes out. Replace brushes when they
are worn down to 3/16" in length. Always replace both brushes even if only one
brush is worn down.
NEED A PART? CALL THE LAB-LINE PARTS HOTLINE.
CALL: (319) 556-2241 or (800) 522-5463; FAX: (319) 589-0516.
LAB-LINE RESERVES THE RIGHT TO MAKE CHANGES TO SPECIFICATIONS WITHOUT PRIOR NOTICE.
13
Page 14
WARRANTY
Page 1 of 2
LAB-LINE INSTRUMENTS, INC. (“Lab-Line”) warrants that the product manufactured by
Lab-Line shall be free of defects in materials and workmanship for a period of time
defined on the following page from the first to occur of (i) the date the product is sold by
Lab-Line or (ii) the date the product is purchased by the original retail customer (the
“Commencement Date”). Except as expressly stated above,
LAB-LINE MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED, WITH
RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIMS ANY AND ALL
WARRANTIES, INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN,
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
An authorized representative of Lab-Line must perform all warranty inspections. In the
event of a defect covered by Lab-Line’s warranty, Lab-Line shall, as its sole obligation
and exclusive remedy, provide free replacement parts to remedy the defective product.
In addition, for products sold by Lab-Line within the continental United States or
14
Page 15
Canada, Lab-Line shall provide free labor to repair the products with the replacement
parts, but only for a period of ninety (90) days from the Commencement Date.
Lab-Line’s warranty provided hereunder shall be null and void and without further force
or effect if there is any (i) repair made to the product by a party other than Lab-Line or
its duly authorized service representative, (ii) misuse (including use inconsistent with
written operating instructions for the product), mishandling, contamination, overheating,
modification or alteration of the product by any customer or third party or (iii) use of
replacement parts that are obtained from a party who is not an authorized dealer of LabLine.
IN NO EVENT SHALL LAB-LINE BE LIABLE TO ANY PARTY FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR
ANY DAMAGES RESULTING FROM LOSS OF USE OR PROFITS, ANTICIPATED
OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USE
OR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON
CONTRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF STRICT
LIABILITY OR REGULATORY ACTION.
The name of your nearest authorized Lab-Line dealer may be obtained by calling 1800-522-5463.
DESIGNERS AND MANUFACTURERS
A SUBSIDIARY of Barnstead|Thermolyne
1999 North 15th Ave.
PHONE: (319) 556-2241 or (800) 522-5463; FAX: (319) 589-0516
, Melrose Park, IL 60160-1491 USA
WARRANTY
Page 2 of 2
12 MONTH PARTS WARRANTY:
• All Environmental Chambers
• Low Temperature B. O. D. Incubators
• Animal Study Chamber
• Controlled Environment Centers
• Biological Work Station
• Refrigerators, Freezers
• Chromatography Refrigerators (5 year parts warranty on compressor only)
• Large Capacity Refrigerators and Freezers (5 year parts warranty on compressor
only)
24 MONTH PARTS WARRANTY:
• Frame Clamps, Frame Sets, Lab Jacks
• Saybolt Viscosimeter
15
Page 16
• Timers, Samplers, Flasks
• Saf-T-Shield, Safety Tongs
• All Incubators & Ovens
• Dual Action Open Air Shaker
• Reciprocating Shakers (open air and water bath)
• Rockers and Rotators
• Low Cost Shakers
• Environ Blok Shaker
• Titer Plate Shaker
• Multi Wrist Shaker
• Water Baths (excluding Aquabaths), Ultrasonic Cleaners
• Slide Warmers
• Mixers, Stirrers, Hotplates
• Thermal Cyclers
• Blok Heaters
• Aquabaths, lifetime warranty on heaters
LIFETIME PARTS WARRANTY:
• All ORBITAL Shakers (not carrying a 24 month parts warranty) offer a lifetime
parts warranty on the drive mechanism and a 5 year warranty on all other parts
• Refrigerated Orbital Shakers carry a lifetime warranty on the drive mechanism,
1 year parts warranty on the compressor, and a 5 year warranty on all other
parts.
16
Page 17
FIRST IN INSTRUMENTS SERVICING SCIENCE, INDUSTRY,
RESEARCH AND EDUCATION
SINCE 1908.
17
Page 18
18
Page 19
ACCESSORY CHECKLIST
The following loose parts and accessories are packed with this unit.
Before discarding any packing materials, please be sure that nothing
has been overlooked.
MODEL NO.: 4625, 4625JPN, 4625-1, 4625-1CE, 4625-2
INSPECTOR __________
DATE __________
PACKED BY __________
CHECKEDITEMPART NUMBERQUANTITY
_________Operation Manual057-057-001
_________Warranty Card528-022-001
Voltage Warning Tag
_________(240 Volt) 528-009-001
19