Process ECN-2
Technical Data and
V2CS
August, 2004
Crossover
Information
©Eastman Kodak Company, 2004

Note: While the data presented are typical of production coatings, they do not represent standards which must be met by Kodak.
Varying storage, exposure, and processing conditions will affect results. The company reserves the right to change and improve
characteristics at any time.
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Processed Control Strip - 35mm . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Important Improvements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Control Strip Log Exposure Range
ECN-2 Control Strip Exposed on KODAK VISION 200T Color
Negative Film 5274 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
ECN-2 Control Strip Exposed on KODAK VISION2 200T Color
Negative Film 5217 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Effects of Mechanical and Chemical Variations
Developer Mechanical Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Effects of Time and Temperature Variations . . . . . . . . . . . . . . . . . . 8
Developer Chemical Factors
Effects of pH and CD-3 Variations . . . . . . . . . . . . . . . . . . . . . . . . . 9
Effects of NaBr and AF-2000 Variations . . . . . . . . . . . . . . . . . . . . 10
Effects of Na2SO3 and Under and Over Replenishment Variations. 11
Effects of Prebath Contamination . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Effects of CD-2 for CD-3 and Stop Bath Contamination . . . . . . . . 13
Effects of Bleach Contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Effects of F-34a Fixer and Final Rinse Contamination . . . . . . . . . . 15
Effects of KI Contamination and Water Dilution . . . . . . . . . . . . . . 16
Effects of Bacterial Na2S and Na2CO3 Contamination . . . . . . . . . . 17
Control-Strip Crossover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Control-Strip Crossover for Process ECN-2 . . . . . . . . . . . . . . . . . . 19
MORE INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
. . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 KODAK VISION2 Color Negative Control Strips, Technical Data and Crossover Information

KODAK VISION2 Color Negative Control Strips
Introduction
KODAK VISION2 Color Negative Control Strips (Figure 1)
are intended for the control of Process ECN-2 using the
procedures described in KODAK Publication No. H-24,
Manual for Processing KODAK Motion Picture Films,
Module 1, Process Control.
The first batch of KODAK VISION2 Color Negative
Control Strips is code 5011 for 35 mm, CAT 152 0469 and
code 5012 for 16 mm, CAT 152 0444.
Figure 1 Processed Control Strip - 35mm
ECN-2
9071
Steps
4
HD
Important Improvements
KODAK VISION2 Color Negative Control Strips
incorporate several important improvements into the control
strips for Process ECN-2.
• A VISION2 emulsion is being used for process control.
The new format is exposed on KODAK VISION2 200T
Color Negative Film 5217, while the former ECN control
strip was exposed on KODAK VISION 200T Color
Negative Film 5274.
KODAK VISION2 Color Negative Control Strips,
Process ECN-2, are packaged in 100-foot rolls containing at
least 120 exposures and a processed reference strip. The
exposures on the roll are spaced at 9.5-inch intervals. Each
exposure has 21 gray-scale steps at 0.20 log H increments
(2/3 camera stop).
The reference strip accompanying each roll was exposed
along with all other control strips; it was then processed
under specified well-controlled conditions. An instruction
sheet enclosed with the package contains process deviation
(correction) factors, if they are required, to determine your
laboratory process aim numbers. A four-digit code number
appearing on the carton, can, control strips, reference strips,
and instruction sheet, identifies each production batch of
strips.
8
MD
14
LD
21
D-min
F009_0146CC
KODAK VISION2 Color Negative Control Strips, Technical Data and Crossover Information 3

Control Strip Log Exposure Range
The graphs in Figure 2 and Figure 3 demonstrate the
differences between the former ECN control strip exposed
on KODAK VISION 200T Color Negative Film 5274, and
the new format exposed on KODAK VISION2 200T Color
Negative Film 5217.
Physically, the center-to-center step distance has been
maintained on the new control strips; no problems should be
encountered with existing automatic densitometers.
Once the transition to the new control strips is complete,
and the new control step reference values have been
determined, continue to use KODAK VISION2 Color
Negative Control Strips in the same manner as your previous
control strips.
Please contact your Kodak representative for additional
information on monitoring your process with control strips.
4 KODAK VISION2 Color Negative Control Strips, Technical Data and Crossover Information

Figure 2 ECN-2 Control Strip Exposed on KODAK VISION 200T Color Negative Film 5274
3.0
2.0
DENSITY
1.0
Exposure:
Process:
1/50 sec
ECN-2
Densitometry:
Status M
B
G
R
0.0
-4.0 -3.0 -2.0 -1.0 0.0 1.0 2.0
LOG EXPOSURE (lux-seconds)
KODAK VISION2 Color Negative Control Strips, Technical Data and Crossover Informat ion 5

Figure 3 ECN-2 Control Strip Exposed on KODAK VISION2 200T Color Negative Film 5217
3.0
Exposure:1/50 sec
Process: ECN-2
Densitometry: Status M
B
G
2.0
R
DENSITY
1.0
0.0
-4.0 -3.0 -2.0 -1.0 0.0 1.0 2.0
LOG EXPOSURE (lux-seconds)
6 KODAK VISION2 Color Negative Control Strips, Technical Data and Crossover Information