Keyence FS-NEO Series Training Manual

DIGI TA L F IBE ROP TI C SE NS OR
T R A INI NG GUID E
Basics: Ca libration, T ips, and Trouble shooting
DIGITAL FIBEROPTIC SENSOR TR AINING GUIDE: FS Neo Series Basics
Operation 1
The Basics of Sensitivity Adjustment
1 Two-point Calibration
point as the setting value.
OPERATING PROCEDURE
(1) Make sure that the display shows the setting value/current value.
(2) Press the [SET] button once with no workpiece present.
(3) Press the [SET] button once with a workpiece present.
Target
2 Fully Automatic Calibration
With this method, the FS -NEO Series sets the intermediate value between the maximum and minimum received light
intensity within a certain period of time.
Do you have trouble adjusting the sensitivity for applications where a workpiece that is narrower than the optical axis
diameter continuously passes through the light beam? In such a case, fully automatic calibration will enable more stable
detection.
OPERATING PROCEDURE
(1) While holding down the [SET] button for 3 seconds or longer, let the workpiece(s) pass through the beam. The
sensitivity is set based on the maximum and minimum light intensity received while the [SET] button is pressed down.
Work piece
3 seconds or longer
After the calibration is complete, the setting value is displayed in green on the digital monitor.
2
The display f lashes when t he setting is complete.
Current value (red)Set ting valu e (green)
DIGITAL FIBEROPTIC SENSOR TR AINING GUIDE: FS Neo Series Basics
Operation 2
The Basics of Sensitivity Adjustment
3 Maximum Sensitivity Setting
[For thrubeam models] When a workpiece is larger than the fiber strand diameter (lens diameter), the light is
completely blocked. This method is used to set the sensitivity when the workpiece is large
enough to completely interrupt the light.
[For reflective models] This method is used to set the maximum sensitivit y while still ignoring the background.
OPERATING PROCEDURE
(1) Make sure that the display shows the setting value/current value.
(2) [For thrubeam models] Press and hold the [S ET] button for 3 seconds or longer with a workpiece present between
the transmitter and receiver fiber units.
Target
Pres s and hol d for 3 seconds or long er
[For reflective models] Press and hold the [SET] button for 3 seconds or longer with no workpiece present.
Pres s and hol d for 3 seconds or long er
4 Fine Adjustment
Set the sensitivity manually.
OPERATING PROCEDURE
(1) Make sure that the display shows the setting value/current value.
(2) While monitoring the current value display, press the manual buttons to perform fine adjustment of the setting value.
Pres sing here will increment the se tting value.Set ting valu e
Pres sing here will decrement t he setti ng va lue.
3
DIGITAL FIBEROPTIC SENSOR TR AINING GUIDE: FS Neo Series Basics
Operation 3
The Basics of Sensitivity Adjustment
5 Positioning Calibration
With this method, the FS -NEO Series detects the workpiece exactly when its edge reaches the predetermined position.
OPERATING PROCEDURE
(1) Make sure that the display shows the setting value/current value.
(2) Briefly press the [SET] button with no workpiece present.
Work piece
(3) Place the part of the workpiece to be detected (edge). Then, press and hold the [SET] button for 3 seconds or longer.
Work piece
Edge
3 seconds or
longer
When set ting is c omp lete, the display f lashes and then sho ws th e settin g val ue.
4
DIGITAL FIBEROPTIC SENSOR TR AINING GUIDE: FS Neo Series Basics
Operation 4
Simplified Sensitivity Adjustment
6 Preset Function
One simple click of the button will make all the displays uniform and set the sensitivity at the same time.
OPERATING PROCEDURE
After installing the sensor, briefly press the [PRESET] button with no workpiece present when using a thrubeam model
(or with a workpiece present when using a reflective model).
Pres s the [PRESET] but ton once with no workpie ce present.
PST i ndicator lig hts i n green.
The received light intensity is adjusted to 100.0 and
the setting value is adjusted to 50.0.
The setting is complete with just one click.
The setting value is displayed a s “50.0”.
The current value is displayed as “100.0”.
7 Saturation Recovery Function
When setting a reflective model, this function automatically adjusts the light intensity from a level that is too intense,
which prevents light intensity differences from being detected, to a lower value that allows for proper, reliable detection.
OPERATING PROCEDURE
After installing the sensor, briefly press the [MODE] button and [SET] button at the same time.
Pres s the [MO DE] button an d [SE T] b utton at the same time.
Power mod e
HSP*, FINE, T URB O 2047 ± 350
SUP ER 4095 ± 500
ULTR A, M EGA 5000 ± 600
* HI GH SP EED
Set ting range of received
light intensity
This function automatically adjusts the light intensity to the proper level.
5