KERN ADB Operating Instruction

KERN & Sohn GmbH
Ziegelei 1
E-Mail: info@kern-sohn.com
Internet: www.kern-sohn.com
Operating instructions Analytical balance
KERN ADB
Version 1.0
11/2015
GB
D-72336 Balingen
Fax: +49-[0]7433-9933-149
ADB-BA-e-1510
GB
KERN ADB
Analytical balance
Version 1.0 11/2015
Operating instructi ons
Contents
1 Technical data ................................................................................................ 4
1.1 Dimensions ............................................................................................................................... 5
2 Appliance overview ....................................................................................... 6
2.1 Keyboard overview ................................................................................................................... 7
2.2 Overview of display .................................................................................................................. 8
3 Basic Information (Gen eral) .......................................................................... 9
3.1 Intended use ............................................................................................................................. 9
3.2 Improper Use ............................................................................................................................ 9
3.3 Warranty .................................................................................................................................... 9
3.4 Monitoring of Test Resources ............................................................................................... 10
4 Basic Safety Precauti o ns ............................................................................ 10
4.1 Pay attention to the instructions in the Operation Manual ................................................ 10
4.2 Personnel training .................................................................................................................. 10
5 Transport and storage ................................................................................. 10
5.1 Testing upon acceptance ...................................................................................................... 10
5.2 Packaging / return transport ................................................................................................. 11
6 Unpacking, Setup and Commissioning ..................................................... 13
6.1 Installation Site, Location of Use .......................................................................................... 13
6.2 Unpacking, checking and install ation .................................................................................. 13
6.3 Levelling .................................................................................................................................. 14
6.4 Electric power supply ............................................................................................................ 15
6.5 Initial Commissioning ............................................................................................................ 15
6.6 Connection of peripheral devices ......................................................................................... 15
7 Adjustment ................................................................................................... 16
7.1 Adjustment with recommended adjustment weight ........................................................... 16
7.1.1 Adjustment with 100 g-weight .................................................................................................. 17
8 Basic Operation ........................................................................................... 18
8.1 Start-up .................................................................................................................................... 18
8.2 Switch into stand-by mode .................................................................................................... 18
9 Applications ................................................................................................. 21
10 Menu ............................................................................................................. 24
10.1 Navigation in the menu .......................................................................................................... 24
10.2 Menu overview ........................................................................................................................ 26
10.3 Description of individual functions ...................................................................................... 27
10.3.1 Automatic zero point correction............................................................................................ 27
2 ADB-BA-e-1510
11 RS232C-interface ......................................................................................... 28
11.1 Technical data ......................................................................................................................... 28
11.2 Interface cable: ....................................................................................................................... 29
11.3 Interface parameters .............................................................................................................. 29
11.3.1 Baud rate .............................................................................................................................. 29
11.3.2 Output condition ................................................................................................................... 30
11.4 Sample protocols (KERN YKB-01N) ..................................................................................... 31
12 Servicing, maintenance, disposal .............................................................. 32
13 Instant help ................................................................................................... 33
ADB-BA-e-1510 3

1 Technical data

KERN ADB 200-4
Readability (d) 0.1 mg Weighing range (max) 210 g Reproducibility 0.2 mg Linearity ± 0.4 mg Recommended adjusting weight
not supplied (class)
200 g (E2)
Stabilization time 3 sec. Warm-up tim e 8 h Weighing Units Smallest part weight for piece
counting
Reference quantities at piece counting
g, oz, ct, lb
0.1 mg
5, 10, 20, 50,100
Weighing plate, stainless steel ø 90 mm Net weight (kg) 6 Permissible ambient condition +10° C to +30° C Humidity of air 20 ~ 85 % relative (not condensing) Power supply unit
Input voltage Balance
Input voltage
AC 110 -230 V, 50-60Hz
DC 12 V, 4 A
Interface RS232C
4 ADB-BA-e-1510

1.1 Dimensions

ADB-BA-e-1510 5

2 Appliance overview

Pos.
Designation
1 Display 2 Handle for operation of the side windshield doors 3 Foot screws 4 Bubble level 5 Windshield 6 Weighing pan
6 ADB-BA-e-1510

2.1 Keyboard overview

Button Function
Turn on/off Exit menu
Adjustment
Taring Zeroing Save setting
Switch-over weighing unit Selecting an application
How to save the reference Switch on/off background illumination of display
(keep pressed the buttons long time)
Change setting in the menu Print out the displayed value
ADB-BA-e-1510 7

2.2 Overview of display

Display Description
Display of stable values The balance is in stand-by mode
g Weighing unit “Gram”
ct lb
oz
Pcs
%
Weighing unit „Carat“ Weighing unit „Pound“ Weighing unit „Ounce“ Application parts counting Application percentage determination
8 ADB-BA-e-1510

3 Basic Information (General)

3.1 Intended use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic balance”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan. As soon as a stable weighing value is reached the weighing value can be read.

3.2 Improper Use

Do not use balance for dynamic add-on weighing procedures, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. T he serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and dest r ucti on o f the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

3.3 Warranty

Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
ADB-BA-e-1510 9

3.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN's accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.

4 Basic Safety Precautions

4.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

4.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

5 T ransport and storage

5.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
10 ADB-BA-e-1510

5.2 Packaging / return transport

Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Pack display unit, weighing plate + accessories and power
supply unit separately.
Secure glass windshield against slipping (e.g. using an
adhesive strip).
Reattach possibly supplied transport securing devices.
Secure all parts agains t shi fting and damage as depicted.
ADB-BA-e-1510 11
12 ADB-BA-e-1510

6 Unpacking, Setup and Commissioning

6.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
On the installation site observe the following:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (co ndensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.

6.2 Unpacking, checking and installation

Open packaging and remove all parts careful l y. Verify that there has been no damage and that all packing items are present.
Scope of delivery / serial accessories
1. Balance
2. Display Unit
3. Weighing pan
4. Carrier weighing plate
5. Air chamber
6. Bottom sheet (round)
7. Mains adapter
8. Country-specific mains plug
9. Operating instructions
ADB-BA-e-1510 13
Prior to any installation and assembly works, the balance must be separated from the mains supply.
Remove the transport securing from the lower side of the balance.
Install the balance at the intended workplace.
The right place is decisive for the accuracy of the weighing results of high­resolution analytic balances (see chap. 6.1).
Put the following parts upon in the right order
Round bottom sheet
Air chamber
Carrier of weighing plate
Weighing pan
Connect the display unit with the balance.
As the terminal is not rigidly assembled to the balance, it can be freely placed in the vicinity of the balance but not further away than this is permitted by the length of the cable.

6.3 Levelling

Level balance with foot screws until the air bubble of the water balance is in the
prescribed circle.
Check levelling regularly.
14 ADB-BA-e-1510

6.4 Electric power supply

Do not connect the scales to the power grid unless the information on
consent by KERN.
As soon as the balance is supplied with energy,
Select a country-specific mains plug and plug it into the mains adapter. Check, whether the voltage acceptance on the scales is set correctly.
the instrument (sticker) matches the local mains voltage. Only use KERN original mains adapter. Using other makes requires
Important:
Prior to commissioning check the mains cable for damage. Make sure that the mains adapter will not be in touch with liquids. The mains plug must be accessible at any time.
Connect the mains adapter to the connecting socket on the backside of the balance and to the power mains.
The display unit lights up. the indicator [ ] is displayed.

6.5 Initial Commissioning

In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). During this warming up time the balance must be connected to the power supply (mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.

6.6 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
ADB-BA-e-1510 15

7 Adjustment

Carry out adjustment as near as possible to the balance’s maximum
Ensure that there are no objects on the weighing pan.
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation. Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization. Ensure that there are no objects on the weighing pan.
weight (required adjustment weight see chap. 1). Info about adjustm ent weights can be found on the Internet at: http://www.kern-sohn.com
Observe stable environmental conditions. Stabilisation requires a certain warm-up time.

7.1 Adjustment with recommended adjustment weight

We recommend to carry out all adjustments as close as possible to the maximum load of the balance.
Press the CAL button until „CAL 100“ will be displayed.
Press TARE key to select „CAL 200“
Put the required adjustment weight (200 g) carefully in the centre of the weighing plate and close the windshield doors.
16 ADB-BA-e-1510
Press the CAL key
After successful adjustment the balance automatically returns to weighing mode.
Take away adjustment weight.

7.1.1 Adjustment with 100 g-weight

Adjustment is also possible with a weight of 100 g, but is not optimal for technical measuring.
Press the CAL key until „CAL 100“ will be displayed.
Put the adjustment weight (100 g) carefully in the centre of the weighing plate and close the windshield doors.
Press the CAL key
After successful adjustment the balance automatically returns to weighing mode.
Take away adjustment weight.
ADB-BA-e-1510 17

8 Basic Operation

8.1 Start-up

As soon as the balance is supplied with energy, the indicator [ ] is displayed.
To switch on press the ON/OFF key.
The balance carries out a display test. As soon as the weight display appears, the balance is
ready for weighing.

8.2 Switch into stand-by mode

In stand-by mode the balance is ready for operation immediately after
switching-on without warm-up time.
To switch off the balance completely, separate it from the mains. The balance starts in the mode, in which it has been switched off.
Press ON/OFF key, the display disappears
The power symbol [ ] will be displayed.
18 ADB-BA-e-1510

8.3 Zeroing

In order to obtain optimal weighing results, reset to zero the balance before weighing.
Unload the balance. Press the TARE key.
Wait until the zero display appears.

8.4 Simple weighing

Place the goods to be weighed and close the windshield.
Wait for stability display . Read weighing result.
Ov erload warning

8.5 Unity change

By repeated pressing of the MODE key the weighed value can be switched over to the available weig hing and application units.
Overloading exceeding the stated maximum load (max) of the device, minus a possibly existing tare load, must be strictly avoided. This could damage the instrument. Exceeding the maximum load is indicated by the display „Err“. Unload weighing system or reduce preload.
ADB-BA-e-1510 19

8.6 Weighing with tare

When the balance is unloaded the saved taring value is displayed with
The dead weight of any weighing container may be tared away by pressing a button, so that the following weighing procedures show the net weight of the goods to be weighed.
Set balance to zero
Deposit empty weighing container. The res ult is displayed.
negative sign.
To delete the stored tare value, remove load from weighing pan and press the TARE key.
Wait for stability display, then press the TARE key
The zero display appears.
Weigh the material, the net weight will be indicated.
20 ADB-BA-e-1510

9 Applications

9.1 Parts counting

Before the balance can count parts, it must know the average part weight (i.e. the reference). Proceed by putting on a certain number of the parts to be counted. The balance determines the total weight and divides it by the number of parts, the so­called reference quantity. Counting is then carried out on the basis of the calculated average piece weight.
1. Select the reference piece number
In weighing mode call up menu item „C2“, see chap.
10.1 Select the desired reference number of pieces using
the PRINT key, see chap. 10.2. Confirm setting by pressing the TARE key. Exit menu using ON/OFF key: Co nfirm query „SAVE“
by pressing the TARE key. The balance returns automatically into weighing mode
2. Call application
Press the MODE key repeatedly until „Pcs“ is displayed.
3. Set to zero/taring Press the TARE key to set the balance to zero or to
tare when using a weighing container.
ADB-BA-e-1510 21
4. Weigh-in reference parts
of the minimum weight of the piece (see chap. 1 „Technical
The reference weight will remain stored even after the weighing balance
Place as many counted pieces to add-up as req ui r ed by the set reference piece number.
Sample protocol (KERN YKB-01N)
100 pcs s
Press the SET key to save the reference whereupon the weighing scale automatically calculates an average weight per part. Remove reference weight. The balance is now in parts counting mode and counts all units on t he weighing plate.
5. Count the item s
Place load on pan and read the number o f pie c es.
6. Switch-over display between number of items and weight
Use the MODE key to switch-over the display to the available units, see chap. 8.5
7. Printing
The display value will be printed out by connecting an optional printer and pressing the PRINT key (factory setting).
Take care data“).
was turned off until the reference is reset.
22 ADB-BA-e-1510

9.2 Percent determination

Percentage calculation facilitates weight display in percent related to a reference weight equivalent to 100 %.
1. Call application
Press the MODE key several times until „%“ is displayed.
2. Set to zero/taring Press the TARE key to set the balance to zero or to
tare when using a weighing container.
3. Reference setting (100 % value)
Put a reference weight which corresponds to 100 %.
Sample protocol (KERN YKB-01N)
49.99 % s
Store reference by pressing the SET key.
Remove reference weight.
4. Percent determination
Place goods to be weighed on balance. The weight of the sample is displayed in percentage in terms of the reference w eig ht.
5. Switch-over the display between percentage and weight
Use the MODE key to switch-over the display to the available units, see chap. 8.5
6. Printing
The display value will be printed out by connecting an optional printer and pressing the PRINT key (factory setting).
The reference weight (100 %) will remain stored even after the weighing balance was turned off until the reference is reset.
ADB-BA-e-1510 23

10 Menu

menu items showing the current settings one by

10.1 Navigation in the menu

Access to menu
1.
In weighing mode print first the SET key, then
the PRINT key. The first menu item „C1“ showing the current setting will be displayed.
Select menu items
2.
Press the TARE key to select the individual
one.
Change settings
3.
Press the PRINT key to change the setting of a
selected menu item. Each time the PRINT key is pressed the next setting will be displayed.
24 ADB-BA-e-1510
Confirm setting with TARE key, the next menu
item will be displayed. Either carry out more settings or exit and save menu (see step 4 or 5)
How to save settings and to exit the menu
4.
Press the ON/OFF key, „SAVE“ will be
displayed.
Any changes carried out are stored by pressing
the TARE key. The balance returns automatically into weighing mode.
Cancel
5.
Press again the ON/OFF key, no changes will
be saved. The balance retur ns aut om ati c all y into weighing mode.
ADB-BA-e-1510 25
Output for stable and instable weighing values

10.2 Menu overvi ew

Menu item Settings Description
Adjustment mode
Reference quantity
Automatic zero point correction
Baud rate
C1 - 0 Not documented C1 – 1* Always use this setting C2 – 0 10 C2 - 1 20 C2 - 2 50 C2 - 3 100 C2 - 4 1000 C3 - 0 C3 – 1 C3 - 2 C3 - 3 C3 - 4 C3 - 5 C3 - 6
Automatic zero tracking off Automatic zero point correction ON <1d> Automatic zero point correction ON <21d> Automatic zero point correction ON <3d> Automatic zero point correction ON <4d> Automatic zero point correction ON <5d> Not documented
C4 - 0 2400 C4 - 1 1200 C4 - 2 4800 C4 - 3 9600
Data output
C5 - 0 Automatic output of stable weighing values C5 - 1 Via remote control command (P) C5 - 2 Continuous data output of stable and instable
C5 – 3
Sound by pressing the button
Not documented
C6 – 0 switched on C6 - 1* switched off C7 - 0* Always use this setting C7 - 1 Not documented
weighing values (interval 3 s)
after pressing PRINT key
26 ADB-BA-e-1510

10.3 Description of individual functions

10.3.1 Automatic zero point correction

This function is used to tare small variations in weight automatically. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“. (e.g. slow flow of liquids from a container placed on the balance, evaporating processes). When apportioning involves small variations of weight, it is advisable to switch off this function.
Call up menu item „C3“, see chap. 10.1 Select the desired setting using the PRINT key, see
chap. 10.2.
Save / back to weighing mode, see chap. 10.1
ADB-BA-e-1510 27

11 RS232C-interface

For the connection of a peripheral device (printer, computer) the balance is as per series equipped with a RS232C-interface.
The following conditions must be met to provide successful communication between the weighing balance and the peripheral devices.
Connect balance using a suitable cable with the interface of the peripheral device. Faultless operation requires an adequate KERN interface cable.
Communic a ti on par a m eters (baud rate, bits and parity) of balance and peripheral device must match.
This data exchange is asynchronous using ASCII - Code.

11.1 Technical data

Connection 9 pin d-subminiature bushing
Baud rate 1200 / 2400 / 4800 / 9600 optional Parity 8 bits, no parity / 1 stop bit / 1 start bit
Pin 2: Receive data Pin 3: Transmit data Pin 5: Signal ground
28 ADB-BA-e-1510

11.2 Interface cable:

Balance 9-poles
PC 9-poles
Balance
Printer poles
2 2 3 3 5 5
9-poles
2 2 3 3 5 7

11.3 Interface parameters

11.3.1 Baud rate

In this menu point the data transfer is adapted to different RS232C-peripheral devices. The baud rate determines the speed of the data transfer via the serial interface. For a faultless data transfer, balance and peripheral device must be set to the same value.
Call up menu item „C3“, see chap. 10.1 Use the PRINT key to select the desired setting.
25-
Options:
C4 - 0 2400 baud C4 - 1 1200 Baud
ADB-BA-e-1510 29
C4 - 2 4800 baud C4 - 3 9600 baud
Save / back to weighing mode, see chap. 10.1

11.3.2 Output condition

The data transfer type is determined in this menu item
Call up menu item „C5“, see chap. 10.1 Use the PRINT key to select the desired setting. Options:
C5 - 0 Automatic edition of stable weighing values C5 - 1 Via remote control command (P) C5 - 2 Continuous data output of stable and instable
weighing values (interval 3 s)
C5 - 3 Output for stable and instable weighing values
after pressing PRINT key
Save / back to weighing mode, see chap. 10.1
30 ADB-BA-e-1510

11.4 Sample protocols (KERN YKB-01N)

+ 10.0000 g s Stable/positive weighing value
- 10.0000 g us Instable/negative weighing value
100 pcs s Parts counting (stable weighing value)
49.99 % s Percentage determination (stable weighing value)

11.5 Remote control instructions

Command Terminator
ASCII Hex <CR> <LF>
Function
O 4F 0D 0A T 54 0D 0A C 43
M 4D P 50
0D 0D
0D
0A
0A 0A
Function such as ON/OFF key Function such as TARE key
Function such as CAL key Function such as MODE key
Function such as PRINT key
ADB-BA-e-1510 31

12 Ser vi c ing, maintenance, disposal

Before any maintenance, cleaning and repair work disconnect the appliance from the oper ati ng v ol tag e.

12.1 Cleaning

Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device. Polish with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods must be removed immediately.

12.2 Servicing, maintenance

The appliance may only be opened by trained service technicians who are
authorized by KERN.
Before opening, disconnect from power supply.

12.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
32 ADB-BA-e-1510

13 Inst a nt hel p

In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The displayed weight does not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plug ged in/faulty).
Power supply interrupted.
The displayed weight is permanently changing
The weighing result is obviously incorrect
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
The balance is on an uneven surface.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
ADB-BA-e-1510 33