KERN ABT series, ABT-BA-e-1312, ABT 100-5M, ABT 120-4M Operating Instruction

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating instruct ion Analytical balance
KERN ABT
Version 1.2
03/2013
GB
#
ABT-BA-e-1312
ABT-BA-e-1312
2
GB
KERN ABT
Version 1.2 03/2013
Operating instruction
Analytical balance
Table of Contents
1 Technical data ................................................................................................ 4
2 Declaration of conformity ............................................................................. 7
3 Keyboard and display overview ................................................................... 8
3.1 Keyboard overview .................................................................................................. 8
3.2 Overview of display ................................................................................................10
4 Basic Information (General) ........................................................................ 11
4.1 Proper use ...............................................................................................................11
4.2 Improper Use ..........................................................................................................11
4.3 Warranty ..................................................................................................................11
4.4 Monitoring of Test Resources ...............................................................................12
5 Basic Safety Precautions ............................................................................ 12
5.1 Pay attention to the instructions in the Oper at i on Manual ..................................12
5.2 Personnel training ..................................................................................................12
6 Transport and storage ................................................................................. 12
6.1 Testing upon acceptance .......................................................................................12
6.2 Packaging / return transport ..................................................................................13
7 Unpacking, Setup and Commissioning ..................................................... 15
7.1 Installation Site, Location of Use ..........................................................................15
7.2 Unpacking ...............................................................................................................15
7.2.1 Scope of delivery ...................................................................................................................... 16
7.2.2 Installation ................................................................................................................................. 17
7.3 Mains connection ...................................................................................................18
7.4 Connection of peripheral devices .........................................................................18
7.5 Initial Commissioning ............................................................................................19
7.5.1 Turning On the Power............................................................................................................... 19
8 Adjustment ................................................................................................... 20
8.1 Automatic adjustment via PSC ..............................................................................20
8.1.1 PCS function ON/OFF .............................................................................................................. 21
8.2 Automatic adjustment via Clock-CAL ...................................................................22
8.2.1 Setting the time for Clock-CAL ................................................................................................. 23
8.3 Adjustment via preset process ..............................................................................24
8.3.1 Selecting default adjustment method ....................................................................................... 24
8.3.2 Default setting: Adjustment with internal weight ....................................................................... 25
8.3.3 Default setting: Adjustment test with internal weight ................................................................ 25
8.3.4 Default setting: Adjustment test with external weight ............................................................... 26
8.4 Carrying out alternative adjustment methods ......................................................27
8.4.1 Adjustment with internal weight ................................................................................................ 27
8.4.2 Adjustment test with internal weight ......................................................................................... 28
8.4.3 Adjustment test with external weight ........................................................................................ 29
9 Verification ................................................................................................... 30
10 ISO/GLP log .................................................................................................. 31
10.1 Setting the log printout of your adjustment data .................................................31
10.2 Balance ID Number Setting ....................................................................................32
10.3 Setting the date printout ........................................................................................33
ABT-BA-e-1312
3
10.3.1 Printout Date and Time, without weight value ...................................................................... 33
11 Basic Operation ........................................................................................... 34
11.1 Weighing .................................................................................................................34
11.2 Taring ......................................................................................................................34
11.3 Changing the display ..............................................................................................35
11.4 Changing r eadability ..............................................................................................36
11.5 Underf l oor weighing ...............................................................................................37
12 The menu ...................................................................................................... 38
12.1 Changing settings ..................................................................................................38
12.2 How to change settings .........................................................................................38
12.3 Call up menu ...........................................................................................................39
12.4 Sel ect i ng t he m enu .................................................................................................40
12.5 M enu overview ........................................................................................................41
12.6 M enu Lock ...............................................................................................................44
12.7 Resett ing the menu ................................................................................................44
12.8 Settings control display .........................................................................................45
13 Description of individual functions ............................................................ 46
13.1 Stability filter ...........................................................................................................46
13.1.1 Standard mode ..................................................................................................................... 46
13.1.2 Anti-convection mode ........................................................................................................... 47
13.1.3 High-stability Mode ............................................................................................................... 47
13.1.4 Apportioning mode ............................................................................................................... 48
13.2 Standstill display ....................................................................................................49
13.3 Auto Zero .................................................................................................................50
13.4 Setting date .............................................................................................................51
13.
5 Setting time .............................................................................................................52
13.6 Capacity display .....................................................................................................53
14 Application Functions ................................................................................. 54
14.1 Part s count ing ........................................................................................................54
14.2 Percent det er m i nation ............................................................................................55
14.3 Density determination ............................................................................................56
14.3.1 Density determination of solids ............................................................................................ 56
14.3.2 Determining density of liquids .............................................................................................. 57
14.4 Auto Print ................................................................................................................59
14.5 Add up mode ...........................................................................................................60
14.6 Recipe m ode ...........................................................................................................62
15 Data output ................................................................................................... 64
15.1 RS 232C int erface ...................................................................................................64
15.2 Data For m at s ...........................................................................................................64
15.3 Remot e cont r ol instructions ..................................................................................65
15.4 Standar d Set t ings ...................................................................................................66
15.5 User Set t i ngs ..........................................................................................................67
15.5.1 Communication speed settings ............................................................................................ 68
15.5.2 Delimiter settings .................................................................................................................. 68
15.5.3 Parity settings ....................................................................................................................... 68
15.5.4 Stop bit settings .................................................................................................................... 69
15.5.5 Input-output data format settings ......................................................................................... 69
15.5.6 Handshake settings .............................................................................................................. 69
16 Service, maintenance, disposal .................................................................. 70
16.1 Cleaning ..................................................................................................................70
16.
2 Service, mai ntenance .............................................................................................70
16.3 Disposal...................................................................................................................70
17 Instant help ................................................................................................... 71
ABT-BA-e-1312
4
1 Technical data
KERN ABT 100-5M ABT 120-4M
Readability (d) 0.01 mg 0.1 mg
Weighing range (max) 101 g 120 g
Minimum load (Min) 1 mg 10 mg
Verification value (e) 1 mg 1 mg
Verification class I I
Reproducibility 0.05 mg 0.1 mg
Linearity ± 0.15 mg ± 0.2 mg
Stabilization time 10 sec 3 sec
Adjustment weight internal
Weighing units (calibrated appliances)
g, ct
Smallest part weight for piece counting
1 mg 1 mg
Reference quantities at piece counting
10, 20, 50,100
Weighing plate, stainless
steel
ø 80mm
Dimensions of the housing (B x D x H)
217 x 356 x 338
Dimensions Glass wind screen [mm]
Weighing room 168 x172 x223
Net weight (kg) 7
Permitted environmental condition
+10° C to +30° C
Humidity of air max. 80 % relative (not condensing)
AC adapter (Primary) AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply DC 12 V, 1 A
Pollution Degree 2
Overvoltage Category Category II
Altitude Up to 2000 m
Installation Site device may only be used indoors
ABT-BA-e-1312
5
KERN ABT 220-4M ABT 320-4M
Readability (d) 0.1 mg 0.1 mg
Weighing range (max) 220 g 320 g
Minimum load (Min) 10 mg 10 mg
Verification value (e) 1 mg 1 mg
Verification class I I
Reproducibility 0.1 mg 0.1 mg
Linearity ± 0.2 mg ± 0.2 mg
Stabilization time 3 sec
Adjustment weight internal
Weighing units (calibrated appliances)
g, ct
Smallest part weight for piece counting
1 mg 0,1 mg
Reference quantities at piece counting
10, 20, 50,100
Weighing plate, stainless
steel
ø 80mm
Dimensions of the housing (B x D x H)
217 x 356 x 338
Dimensions Glass wind screen [mm]
Weighing room 168 x172 x223
Net weight (kg) 7
Permitted environmental condition
+10° C to +30° C
Humidity of air max. 80 % relative (not condensing)
AC adapter (Primary) AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply DC 12 V, 1 A
Pollution Degree 2
Overvoltage Category Category II
Altitude Up to 2000 m
Installation Site device may only be used indoors
ABT-BA-e-1312
6
KERN ABT 120-5DM ABT 220-5DM
Readability (d) 0.01/0.1 mg 0.01/0.1 mg
Weighing range (max) 42 g/120 g 82 g/220 g
Minimum load (Min) 1 mg 1 mg
Verification value (e) 1 mg 1 mg
Verification class I I
Reproducibility ±0.02/ 0.1 mg ±0.05/ 0.1 mg
Linearity ± 0.05/0.2 mg ± 0.1/0.2 mg
Stabilization time 3sec./10 sec.
Adjustment weight internal
Weighing units (calibrated appliances)
g, ct
Smallest part weight for piece counting
1 mg
Reference quantities at piece counting
10, 20, 50,100
Weighing plate, stainless
steel
ø 80mm
Dimensions of the housing (B x D x H)
217 x 356 x 338
Dimensions Glass wind screen [mm]
Weighing room 168 x172 x223
Net weight (kg) 7
Permitted environmental condition
+10° C to +30° C
Humidity of air max. 80 % relative (not condensing)
AC adapter (Primary) AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply DC 12 V, 1 A
Pollution Degree 2
Overvoltage Category Category II
Altitude Up to 2000 m
Installation Site device may only be used indoors
ABT-BA-e-1312
7
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Declaration of conformity
EC-Konformitätserklärung EC-Declaration of -Conformity EC-
Déclaration de conformité EC-Declaración de Conformidad
EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
Konformitäts­erklärung
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
GB
Declaration of conformity
We hereby declare that the product to which this declaration refers conforms with the following standards.
CZ
Prohlášení o shode
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferi­sce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
RUS
Заявление о соответствии
Мы заявляем, что продукт, к которому относится данная декларация, соответствует перечисленным ниже нормам.
Electronic Balance: KERN ABT
2004/108/EC EN 61326-1:2006
2006/95/EC EN 61010-1: 2010
Signatur
Signature
Datum
Date
27.03.2013
Ort der Auss tellung
Place of issue
72336 Balingen
Albert Sauter KERN & Sohn GmbH
Geschäftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
ABT-BA-e-1312
8
3 Keyboard and display overview
3.1 Keyboard overview
In weighing mode:
Key
Designa-
tion
Pressed once and released
Keep pressed for about 3 seconds
[ON/OFF]
Switches between the opera­tion and standby modes.
Switches the key notification buzzer on/off.
[CAL]
Invokes adjustment or menu selection.
Invokes adjustment or menu selection.
[TARE] Tare or set weight display to zero
[UNIT] Changing the display
[PRINT]
Issue of weighing value to external appliances (printer,
PC)
Issue of date and time to external applianc­es.
[1d/10d]
ABT 120-4M ABT 220-4M
ABT 320-4M
For verified applications cutting off the last
digit is only possible for models with a
readout of 0,01 mg. For all other models the button 1/10d is inop­erable.
ABT 100-5M
ABT 120-5DM
ABT 220-5DM
ABT-BA-e-1312
9
In menu:
Key
Designa-
tion
Pressed once and released
Keep pressed for about 3 se-
conds
[ON/OFF]
Back to menu Return to weighing mode
[CAL]
Menu selection
[TARE]
Parameter selection Storing settings
[UNIT] Entering numeric values Increases the numeric value of flashing digit by 1.
[PRINT] Entering numeric values. Shifts flashing digit.
[1d/10d]
No effect.
ABT-BA-e-1312
10
3.2 Overview of display
Display
Designation
Description
Standstill display
Indicates that the weighed value is stable. Highlights current set-
ting during menu item selection
Weight symbol
Appears during adjustment. Flashes prior to start of automatic
adjustment. Appears during parameter selection for adjustment. Blinks to ad-
vise necessity of adjustment.
Number symbol
Indicates numeric value entry.
Menu symbol
Appears during menu selection. Always shown when the menu is
locked.
Asterisk
Indicates that the displayed numeric value is not a mass value.
Add-on symbol
Indicates set-up of Add-on mode.
Communication symbol
Is lit up during communication with external appliances via RS­232C cable. Shown when communication functions are ON.
Battery symbol
Indicates a low battery voltage when using the balance with the optional battery pack.
Auto Print symbol
Indicates set-up of the Auto Print function.
Standby mark
Appears during standby status of balance.
Inverse triangle sym-
bol
Illuminates as part of the solid specific gravity measurement dis-
play.
Capacity display
Units display
ABT-BA-e-1312
11
4 Basic Information (General)
4.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the mate­rial to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
4.2 Improper Use
Do not use balance for dynamic add-on weighing procedures, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measur­ing system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be dam­age by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weigh­ing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
4.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
ABT-BA-e-1312
12
4.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the bal­ance and, if applicable, the testing weight, must be checked regularly. The responsi­ble user must define a suitable interval as well as type and scope of this test. Infor­mation is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
5 Basic Safety Precautions
5.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
5.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
6 Transport and stora ge
6.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appli­ance itself when unpacking for possible visible damage.
ABT-BA-e-1312
13
6.2 Packaging / return transport
Keep all parts of the original packaging for a possibly re-
quired return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
ABT-BA-e-1312
14
ABT-BA-e-1312
15
7 Unpacking, Setup and Commissioning
7.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the i nstallation site:
Operate the device only indoors.
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.
7.2 Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
ABT-BA-e-1312
16
7.2.1 Scope of delivery Serial accessories:
Balance
Weighing plate
Carrier of weighing plate
Screening ring
Mains adapter
Adapter cable holder
Protective cover
Operating Manual
ABT-BA-e-1312
17
7.2.2 Installation
Attach the adapter cable holder. Peel
the protective sheet of adhesive of the adapter cable holder, and stick it on the back of the balance as shown in the figure.
Carrier of weighing plate, attach
weighing plate and circular screen in order.
Level balance with foot screws until
the air bubble of the water balance is in the prescribed circle.
When the key panel and the display
must be protected from dirt and wear,
place the cover over the key panel.
ABT-BA-e-1312
18
1
7.3 Mains connection
Power is supplied via the external mains power supply. The stated voltage value must be the same as the local voltage. Only use original KERN mains power sup­plies. Using other makes requires consent by KERN.
7.4 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance. Terminal for external devices:
1 RS -232 interface 2 Mains connector 3 Anti-theft safety device (for chains or other fastenings) 4 Connection earthing terminal
3
4
ABT-BA-e-1312
19
7.5 Initial Commissioning
A warming up time of 4 hours after switching on stabilizes the measuring values. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
7.5.1 Turning On the Power
1. Supply power to balance via mains
adapter. After the balance performs a self check, calibration will be automati­cally executed. During this process, the display will change as follows. “CHE 5”,
“CHE 4”...... “CHE 0”, “CHE 2”, “CHE 1”
, “CAL 2 – CAL 0” , “CAL-End”, “oFF”. This adjustment can be interrupted im­mediately after switching-on the power
supply by actuating the [ON/OFF]-key.
However, at least one adjustment must be carried out before the balance can
be used.
2. Actuate the [ON/OFF] key. After all dis-
plays appear, the gram (g) symbol ap-
pears.
3. Pressing the [ON/OFF] key again turns
on the standby symbol and puts the balance in standby (warm up) status. Current time (see ch. 13.5) will be dis­played.
ABT-BA-e-1312
20
8 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To re­ceive accurate measuring values it is also recommended to adjust the balance peri­odically in weighing operation. Observe stable environmental conditions. Stabilisation requires a warm-up period of 1 or 4 hours (for models d = 0.01 mg). Ensure that there are no objects on the weighing plate.
8.1 Automatic adjustment via PSC
The default setting for balances of the series ABT ("PSC ON") prompts automatic adjustment via the PSC function. The moment a change in temperature is detected this function uses the internal tare weight to carry out fully-automatic adjustment with the help of a temperature sensor. If PSC is left ON, when there is a temperature change that would influence sensitivi­ty, span calibration executes automatically to maintain the sensitivity of the balance. Adjustment in weighing mode is carried out automatically under the following condi­tions:
(1) If a change in environmental temperature (0.5 °C) occurs;
(2) When about four hours has passed since the previous calibration.
(3) When the balance is switched from standby status to weighing mode
and condition (1) or (2) has been met.
(4) When balance was disconnected from the mains.
If one of the above conditions was met in weighing mode, the weight symbol flashes for about two minutes in order to notify the pending adjustment; afterwards the dis­play will show "PSC run". During operation, the display will automatically change and the motor sound of the weight loading system is heard. In order to ensure proper PSC operation, prevent vibrations and air flow. As soon as the display in grams re­appears after completing adjustment via PSC, the balance returns to weighing mode. The sensitivity before and after adjustment is slightly different. Also, no measure­ments can be made during adjustment. To prevent a start-up of adjustment during a
measuring process you have to actuate the [ON/OFF] key as soon as the weight
symbol starts flashing. This will interrupt automatic adjustment.
ABT-BA-e-1312
21
8.1.1 PCS function ON/OFF
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Actuate the [CAL] key repeatedly until the current "PSC:∗∗" set-
ting appears:
To change the setting, actuate the [TARE] key as soon as "PSC:∗∗" appears.
You can use the [CAL] key to select between the following set-
tings:
“PSC-on” Function activated “PSC-oF” Function deactivated
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
PSC and Clock-CAL may be turned on and off separately. The setting control display (see ch. 12.6) shows the weight symbol ( ), when PSC or Clock-CAL or both func-
tions are activated.
ABT-BA-e-1312
22
8.2 Automatic adjustment via Clock-CAL (only models ABT 100-5M, ABT 120-5DM, ABT 220-5DM)
With the help of its internal adjusting weight and integrated clock the balance can be set to carry out automatic adjustment at set times (up to three times daily). Clock­CAL is a very convenient function, when calibration reports are desired to be made for regular calibrations, or when wishing span calibrations during break times to avoid interruption of measurement work. The weight symbol blinks for about two minutes as notification of span calibration
before it begins. Automatic adjustment can be stopped by actuating the [ON/OFF]
key during this message.
Clock-CAL function ON/OFF:
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Actuate the [CAL] key repeatedly until the current "tCal:∗∗" setting
appears:
To change the setting, actuate the [TARE] key as soon as "tCal:∗∗" appears.
You can use the [CAL] key to select between the following set-
tings:
“tCAL-on” Function activated “tCAL-oF” Function deactivated
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
ABT-BA-e-1312
23
8.2.1 Setting the time for Clo ck-CAL
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
(for setting 1)
Press the [CAL] key repeatedly until “tCAL t*” appears:
Actuate the [TARE] key until "t*HH:MM" is displayed. The *-
position symbolises a number between 1 and 3 (3 set times for automatic adjustment). The currently set time is displayed in "HH:MM" (HH for hours, MM for minutes) format with the first digit on the left flashing. The time appears as “_ _:_ _” when no time is set. The symbol, and the # symbol appear to indicate it is in the numerical input status.
Pressing the [PRINT] key moves the blinking digit to the next digit to the right. Press the [UNIT] key to change the value of the blink­ing digit. When the [UNIT] key is pressed, the numerical of the
blinking digit increases by 1 at a time. The numerals progress in this order: 0→1→2→...→9→_→0... Set the hours within the range of 00 to 23 and the minutes within the range of 00 to 59. Complete
your setting by pressing the [TARE] key. The display returns to “t
CAL t*”.
In order to set another time, press the [CAL] key to move to the
next “t CAL t*” setting and set the time in the same way.
Actuate the [ON/OFF] key repeatedly. Return to menu/weighing
mode.
Clearing the Settings
The clock settings “tCAL t1” to “tCAL t3” may each be reset by
using procedure 3 to set the time setting to “_ _:_ _”.
PSC and Clock-CAL may be turned on and off separately. The setting control display (see ch. 12.6)
shows the weight symbol (
), when PSC or Clock-CAL or both functions are activated.
ABT-BA-e-1312
24
8.3 Adjustment via preset process
It is possible to start the preset adjustment method without having to access the menu. When in weighing mode, the preset adjustment method can be carried out by
simply actuating the [CAL] key, followed by [TARE].
8.3.1 Selecting default adjustment method
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Press the [TARE] key. The display shows “E CAL”
Press the [CAL] key repeatedly until the desired setting appears.
The standstill display ( ) appears when current adjustment is displayed.
Select amongst the following four types:
Not documented
Adjustment test with external weight (see ch. 8.3.4)
Adjustment test with internal weight (see ch. 8.3.2)
Adjustment test with internal weight (see ch. 8.3.3)
Confirm your setting by pressing the [TARE] key. The display
shows “CAL dEF”.
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
The selected adjustment can be carried out by simply actuating
the [CAL] key followed by [TARE].
ABT-BA-e-1312
25
8.3.2 Default setting: Adjustment with internal weight
Condition: Function „i tCAL“ activated (see ch. 8.3.1)
Press the [CAL] key The display shows “i-CAL” Press the [TARE] key. The display changes automatically in this
order: “CAL 2”, “CAL 1”, “CAL 0”, and “CAL End”. After success­ful adjustment the balance automatically returns to weighing mode. In case of an adjustment error (e.g. objects on the weighing plate) the display will show an error message, repeat adjustment.
8.3.3 Default setting: Adjustment test with internal weight
During adjustment tests the balance automatically compares the saved value of the adjustment weight with the actual value. This is only a check, i.e. no values are changed.
Condition: Function „i tESt“ activated (see ch. 8.3.1)
Press the [CAL] key The display shows “i-tESt”
Press the [TARE] key. The display changes automatically in this
order: “tESt 2”, “tESt 1”, “tESt 0”
After this, the difference to the previous adjustment will be dis­played for several seconds.
After "tESt End" was displayed, the balance will automatically return to weighing mode.
ABT-BA-e-1312
26
8.3.4 Default setting: Adjustment test with external weight
Condition: Function „E tESt“ activated (see ch. 8.3.1)
Press the [CAL] key The display shows “E-tESt”
Press the [TARE] key. Testing will be started and zero display is
flashing (Ensure that no items are present on the weighing plate).
(Example)
Wait until the weighing value for adjustment is shown and flash­es.
Place the required adjusting weight on the weighing plate.
Wait until the blinking zero display reappears. (this may take about 30 seconds.)
Remove the weight from the weighing plate.
After this, the difference to the previous adjustment will be dis­played for several seconds.
After "tESt End" was displayed, the balance will automatically return to weighing mode.
ABT-BA-e-1312
27
8.4 Carrying out alternative adjustment methods
Here you start adjustment by selecting a setting in the menu.
8.4.1 Adjustment with internal weight
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press the [TARE] key. "CAL" is displayed.
Press the [TARE] key. "E CAL" is displayed.
Repeatedly press the [CAL] key until "i CAL" appears.
Press the [TARE] key. The display changes automatically in this order: “CAL 2”, “CAL 1”, “CAL 0”, and “CAL End”.
After successful adjustment the balance automatically returns to weighing mode. In case of an adjustment error (e.g. objects on the weighing plate)
the display will show an error message, repeat adjustment.
ABT-BA-e-1312
28
8.4.2 Adjustment test with internal weight
During adjustment tests the balance automatically compares the saved value of the adjustment weight with the actual value. This is only a check, i.e. no values are changed.
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press the [TARE] key. "CAL" is displayed.
Press the [TARE] key. "E CAL" is displayed.
Repeatedly press the [CAL] key until "i tESt" appears.
Press the [TARE] key. The display changes automatically in this
order: “tESt 2”, “tESt 1”, “tESt 0”
After this, the difference to the previous adjustment will be dis­played for several seconds.
After "tESt End" was displayed, the balance will automatically return to weighing mode.
ABT-BA-e-1312
29
8.4.3 Adjustment test with external weight
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press the [TARE] key. "CAL" is displayed.
Press the [TARE] key. "E CAL" is displayed.
Repeatedly press the [CAL] key until "i tESt" appears.
Press the [TARE] key. Testing is started and zero display is
flashing. (Ensure that there are no objects on the weighing plate).
(Example)
Wait until the weighing value for adjustment is shown and flash­es.
Place the required adjusting weight on the weighing plate.
Wait until the blinking zero display reappears. (this may take about 30 seconds.)
Remove the weight from the weighing plate.
After this, the difference to the previous adjustment will be dis­played for several seconds.
After "tESt End" was displayed, the balance will automatically return to weighing mode.
ABT-BA-e-1312
30
9 Verification
General introduction: According to EU directive 90/384/EEC balances must be verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory. c) For official purposes. d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification instructions
An EU type approval exists for balances described in their technical data as verifyable. If a balance is used where obligation to verify exists as described above, it must verified and re-verified in regular intervals. Re-verification of a balance is carried out according to the respective national regula­tions. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed! After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
Position of the “official seals“:
Balances with obligation to verify must be taken out of operation if:
- The weighing result of the balance is outside the error limit. Therefore, in
regular intervals load balance with known test weight (ca. 1/3 of the max.
load) and compare with displayed value.
- The reverification deadline has been exceeded.
Seals
ABT-BA-e-1312
31
10 ISO/GLP log
Quality assurance systems require printouts of weighing results as well as of correct adjustment of the balance stating date and time and balance identification. The easi­est way is to have a printer connected.
10.1 Setting the log printout of your adjustment data
This function enables automatic log issue after each adjustment. These logs may be issued by using an optional printer. Fully automatic and periodical adjustments are carried out and logs created in combination with the Clock-CAL function (see ch.
8.2). First make adjustments as specified in chap. 15.4 under „KERN-YKB-01N“. Then set the log printout as follows:
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Repeatedly actuate the [CAL] key until "GLP:**" appears (The ∗∗
symbolise the current setting).
To change settings, actuate the [TARE] key when "GLP:**" ap-
pears.
You can use the [CAL] key to select between the following set-
tings:
“GLP-on” Function activated “GLP-oF” Function deactivated
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
Actuate the [ON/OFF] key repeatedly. Return to menu/weighing
mode.
ABT-BA-e-1312
32
10.2 Balance ID Number Setting
This setting is for the balance ID number that is output along with the adjustment re­port.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Repeatedly actuate the [CAL] key until "id:****" appears (The ∗∗
symbolise the current setting).
Press the [TARE] key. In the upper part of the display panel, the
symbol and the # symbol appear in order to indicate numer-
ical input status. The leftmost digit of ∗∗∗∗ blinks. The numeral of the blinking digit can be changed.
When the [UNIT] key is pressed, the numerical of the blinking digit
increases by 1 at a time. You can determine the value of the flash­ing digit, or shift the flashing digit by one position to the right, by
pressing the [PRINT] key. Confirm your setting by pressing the [TARE] key.
Actuate the [ON/OFF] key repeatedly. Return to menu/weighing
mode.
ABT-BA-e-1312
33
10.3 Setting the date printout
This setting determines whether or not the date and time on the balance’s built-in clock is printed out along with the log.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Repeatedly actuate the [CAL] key until "Prtdt:∗∗" appears (The ∗∗
symbolise the current setting).
To change settings, actuate the [TARE] key when "Prtdt:∗∗" ap-
pears.
You can use the [CAL] key to select between the following set-
tings:
“Prtdt-on” Date and time are printed. “Prtdt-oF” Date and time are not printed.
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
Actuate the [ON/OFF] key repeatedly. Return to menu/weighing
mode.
10.3.1 Printout Date and Time, without weight value
To print out only date and time, without weight value,
press [PRINT] button for approx. 3 sec.
ABT-BA-e-1312
34
11 Basic Operation
11.1 Weighing
Note: Stabilisation requires a warm-up period of 1 or 4 hours
(for models d = 0.01 mg).
Turn on balance by pressing the [ON/OFF] key. The balance will carry out a
self-test Your balance is ready to weigh as soon as the “0.0000 g" display ap-
pears.
Note: The [TARE] key can be used to set the balance to zero at any time.
Place goods to be weighed on balance. Wait until the standstill display ( )
appears, then read the weighing result.
11.2 Taring
The dead weight of any weighing container may be tared away by pressing a button, so that the following weighings show the net weight of the goods to be weighed.
Place empty tare container on the weighing plate. The total weight of the
container is displayed.
To start the taring process press the [TARE] key. The weight of the con-
tainer is now internally saved.
Place the goods to be weighed into the tare container.
Read the weight of the goods on the display.
Note:
The balance is able to only store one taring value at a time. When the balance is unloaded the saved taring value is displayed with negative sign. Remove all items from the weighing plate in order to delete the stored tare value and
subsequently press the [TARE] key.
The taring process can be repeated any number of times. The limit is reached when the whole weighing range is exhausted.
ABT-BA-e-1312
35
11.3 Changing the display
You can change the display to activated units, piece counting, percentage and densi-
ty determination mode by pressing the [UNIT] key several times.
The default setting provides the following options:
[g] [%] [Pcs] [ct] [g]
Different settings have to be activated in the menu:
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
(Example)
Repeatedly press the [CAL] key until "Unit.SEL" appears.
Press [TARE] key
To select settings, use the [CAL] key:
Current settings are indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key. The [TARE] key is also used to deactivate a unit or function if
the corresponding setting with standstill display is shown in the display.
Repeatedly press the [ON/OFF] key. This takes you back to the
menu/weighing mode.
ABT-BA-e-1312
36
11.4 Changing readability (only models ABT 100-5M, ABT 120-5DM, ABT 220-5DM)
Models ABT 120-5DM and ABT 220-5DM are set to “lower range“ with a readability of 0.01 mg as soon as they are connected to power supply and turned on. To switch
to the “upper range” with a minimum display of 0.1mg, press the [1d/10d] key.
If the lower max. weighing range (82g for ABT 220-5DM, 42 g for ABT 120-5DM) is exceeded during the weighing process, the balance changes automatically to the upper range. If you tare the balance in the upper range, this setting will be maintained throughout.
To return to lower range, you will have to actuate the [TARE] key after the stability symbol ( ) has appeared.
Lower range:
Upper range:
ABT-BA-e-1312
37
11.5 Underfloor weighing
Objects unsuitable for placing on the weighing pan due to size or shape may be weighed with the help of the flush-mounted platform. Proceed as follows:
Switch off balance.
Open the closing lid (1) on the bottom of your balance.
Place your balance over an opening.
Suspend the goods to be weighed from the hook and carry out the weighing.
Fig. 1: Underfloor weighing device
CAUTION
Always ensure that all suspended objects are stable enough to hold the desired goods to be weighed safel y (danger of breaki ng).
Never suspend loads that exceed the stated maximum load (max) (danger of breaking)
Always ensure that there are no persons, animals or objects that might be dam­aged underneath the load.
NOTE
After completing the underfloor weighing the opening on the bottom of the bala nce must always be closed (dust protection).
(1)
!
ABT-BA-e-1312
38
12 The menu
To adapt the behaviour of the balance to your requirements, go to the balance mode menu. Usually the default setting of the balance mode menu is such that you do not need to make any changes. If you encounter special conditions of use, go to the bal­ance mode menu in order to set your balance according to your individual require­ments.
12.1 Changing settings
To change the settings of specific functions, select the related functions.
Functions are changed by taking the following three steps:
 Cal l up m enu  Set function  Confirm and store
When setting a function the ON/OFF, CAL and TARE keys have special functions.
12.2 How to change se ttings
CAL key = Select menu and pass through menu items from top to bot-
tom ().
TARE key = Select function. After selecting the function on the display by pressing the CAL key, the change is invoked by pressing the TARE key.
CAL key = Select one of the possible settings within the function. Pass-
ing through menu items from top to bottom.
TARE key = Store the setting momentarily appearing on the display by actuating the TARE key.
The standstill display shows the present setting for the function.
ON/OFF key = Exit function
Brief actuation of ON/OFF key:
Back to previous menu.
Long actuation of ON/OFF key:
Back to weighing mode.
ABT-BA-e-1312
39
12.3 Call up menu
Please have a go and try to change a function yourself. Change the "Auto-Zero" function to OFF and then again to ON.
Turn on balance by pressing the ON/OFF key. Actuate the CAL key until FUnC.SEL appears Press the TARE key once until CAL appears Press the CAL key once until trC :on appears To confirm, press the TARE key once.
"Auto-Zero” function is now selected. The standstill display shows the current setting.
Press the CAL key
thereby means: trC-oF Auto-Zero is switched off trC-on Auto-Zero is switched on Select trC-oF
Press the TARE key once
The colons appear when storing the changed trC: oF
Press ON/OFF key
To quit the menu keep the ON/OFF
key pressed longer than 2 sec.
Note:
If you wish to make several changes in the weighing mode menu you do not have to exit the menu after each change. You can make several changes one after the other and finally exit the menu
ABT-BA-e-1312
40
12.4 Selecting the menu
In weighing mode press the [CAL] key. First function "i-Cal" (ex.) appears. The various functions of the menu are passed through by pressing the [CAL]
key again.
Weighing mode
Preset adjustment method (see ch. 8.3)
Not documented
Setting control display (see ch. 12.6)
Standard mode
P0VR�1nG�
Apportioning mode
Anti-convection mode
High-stability Mode
Entry to second level
(application functions, menu group for individual settings)
Entry to third level (system settings menu group)
Entry to forth level (communication settings menu group)
Weighing mode
Settings for stability and response (see chapter 13.1)
ABT-BA-e-1312
41
12.5 Menu overview
* = default setting
ABT-BA-e-1312
42
ABT-BA-e-1312
43
ABT-BA-e-1312
44
12.6 Menu Lock
The menu setting operations can be locked so that the settings cannot be inadvert­ently changed. This is called Menu Lock. The menu lock is activated with the follow­ing procedure.
Turn on the power supply for the balance and wait until "oFF" ap­pears.
Press the [CAL] key until "LoCKEd" appears. Menu is disabled,
menu symbol appears. This is followed by a reappearance of "oFF".
If a menu item selection is attempted in locked status, the message “LoCKEd” ap­pears and the menu selection is not allowed. To deactivate menu lock, follow the sequence of operations below:
Disconnect the balance from the power and connect again after a while.
When "oFF" is displayed, press the [CAL] key until "rELASE" ap­pears. The menu lock is deactivated.
12.7 Resetting the menu
This will return all the settings to default. The reference value stored in previous use of piece counting or percentage conversion will also be cleared. The default settings are indicated with “#” on the Menu Map.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows “CAL dEF”.
Press the [CAL] key repeatedly until “rESET“ appears.
Press the [TARE] key. The display shows confirmation enquiry
"rESET?
Confirm by pressing the [TARE] key; balance automatically returns to weighing mode.
ABT-BA-e-1312
45
12.8 Settings control display
To receive a confirmation of the current settings, repeatedly actuate the [CAL] key in
weighing mode until "St. b1 t" (ex.) appears.
Stability and response setting
St: Standard mode Co: Anti-convection mode Hi: High-stability mode Po: Apportioning mode
Stability and response setting
b1: 1 count b5: 5 count b10: 10 count
Appears when zero
function is turned on.
The weight symbol is displayed when either automatic adjustment PSC or Clock-CAL or both functions are turned on.
Appears when the automatic adjustment has been switched on by PSC or Clock
-CAL.
Appears when the issue of the adjustment data is switched on.
ABT-BA-e-1312
46
13 Descri pt ion of individual functi ons
13.1 Stability filter
Generally, data processing for greater stability slows the response and processing for higher response reduces stability. Balances of the series ABT are designed to ensure both qualities, that is, fast reaction time and high stability. Most measurements can be carried out using default settings, that is, standard mode. In specific environmental conditions and for specific measuring purposes anti­convection, high stability or apportioning mode is available for you to use. The cur­rently set mode is shown on the setting control display (see ch. 12.6).
13.1.1 Standard mode
This is the default setting. Use this mode unless stability or response has to be in­creased or reduced.
Repeatedly press the [CAL] key until "Stnd".
To set standard mode, press the [TARE] key.
Setting of this mode can be confirmed only in settings check dis-
play (see ch. 12.6).
Setting control di
splay
(when selecting sta
ndard mode)
ABT-BA-e-1312
47
13.1.2 Anti-convection mode
If measurements have to be taken under unavoidable ambience conditions, (e. g. changing air condition systems) convection occurring within the weighing space may result in fluctuations in the display on the balance after the stability display has ap­peared. The small range (readability 0.01 mg) of series ABT is particularly suscepti­ble to this effect. Anti-convection mode adjusts the timing of appearance of the stability mark. Note that the stability mark will then appear at a late timing.
Repeatedly press the [CAL] key until "ConvECt" appears.
To set anti-convection mode, press the [TARE] key.
Setting of this mode can be confirmed only in settings check
display (see ch. 12.6).
13.1.3 High-stability Mode
The balances of series ABT are designed to react only minimally to the effects of vibrations and drafts. However, if it must be set up in a location with poor conditions, use this function to reduce the effects of vibration or air current even further. There is a slight increase in the reaction time of the balance but the display itself is stabilized.
Repeatedly press the [CAL] key until "ConvECt" appears.
To set high stability mode, press the [TARE] key.
Setting of this mode can be confirmed only in settings check
display (see ch. 12.6).
Setting control di
splay
(when selecting anti
-convection mode)
Setting control
display
(when selecting high
-stability mode)
ABT-BA-e-1312
48
13.1.4 Apportioning mode
Use this function if you wish to increase display speed, e.g. during apportioning. However, please note that the balance is very susceptible to ambience conditions. You can set the degree of sensitivity according to site (qui­et/unquiet).
P0VR�1nG�
Repeatedly press the [CAL] key until "PoUr inG" appears.
nQlnAL.Eno�
Press the [TARE] key; display shows current setting of sensitivi-
ty
Setting of this mode can be confirmed only in settings check
display (see ch. 11.6).
5TABL.Eno�
nQlnAL.Eno�
Un5TABL.Eno
You can use the [CAL] key to select between the following set-
tings:
“StAbl.Env” very quiet site/sensitive and fast “normL.Env” normal site/ medium setting “ UnStAbl.Env ” very unquiet site/insensitive but slow
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
For mode setting in setting control display (see ch.12.6):
PQ.S.B1T�
very quiet site/sensitive and fast
PQ.n.B1T�
normal site/ medium setting
PQ.U.B1T�
very unquiet site/insensitive but slow
ABT-BA-e-1312
49
13.2 Standstill display
If the display shows the stability display ( ) the balance is in a stable status. The condition for judging stability is user-selectable. There are three levels to select: 1 count, 5 counts and 10 counts. The default setting is 1 count.
The setting of the standstill display can be checked on the setting control display (see ch.12.6).
Setting the standstill display
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
(Example)
Repeatedly press the [CAL] key until the current "b And:**" set-
ting appears.
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
"b-1" 1 count "b-1" 5 count "b-1" 10 count
The current setting is marked by the stability symbol ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
ABT-BA-e-1312
50
13.3 Auto Zero
This function is used to tare small variations in weight automatically. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensa­tion“ in the balance. (e.g. slow flow of liquids from a container placed on the balance, evaporating processes). When apportioning involves small variations of weight, it is advisable to switch off this function.
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
(Example)
Repeatedly press the [CAL] key until the current "trC:**" setting
appears.
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
“trC-on” Function activated “trC-oF” Function deactivated
The current setting is marked by the stability symbol ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
ABT-BA-e-1312
51
13.4 Setting date
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key.
Repeatedly press the [CAL] key until "d-MM.DD" appears (MM and DD provide two digits each, indicating month and day).
(15. march 2003)
Press the [TARE] key. The date currently set appears. In the up­per part of the display panel, the symbol and the # symbol
appear in order to indicate numerical input status. The leftmost digit blinks.
(2. nov. 2004)
When the [UNIT] key is pressed, the numerical of the blinking digit
increases by 1 at a time. You can determine the value of the flash­ing digit, or shift the flashing digit by one position to the right, by
pressing the [PRINT] key. Confirm your setting by pressing the [TARE] key.
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
ABT-BA-e-1312
52
13.5 Setting time
Balances of the series ABT are equipped with an integrated clock. Set the clock be­fore using the functions Clock-CAL (ch. 8.2) or GLP log (ch. 10). Note that the cur­rent time is displayed at the stand by status (ch. 7.5.1).
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key.
Repeatedly actuate the [CAL] key until "t-HH.MM" appears (HH and MM provide 2 digits each for hour and minute display).
Press the [TARE] key. The currently set time appears. In the up­per part of the display panel, the symbol and the # symbol
appear in order to indicate numerical input status. The leftmost
digit blinks. When the [UNIT] key is pressed, the numerical of the blinking digit
increases by 1 at a time. You can determine the value of the flash­ing digit, or shift the flashing digit by one position to the right, by
pressing the [PRINT] key. Confirm your setting by pressing the [TARE] key.
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
ABT-BA-e-1312
53
13.6 Capacity display
This function displays a bar graph representation of the load on the weighing plate. This may be used to prevent sudden appearance of “oL” (overload) during measurement.
It is possible to turn the display of capacity on or off.
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
Repeatedly press the [CAL] key until the current "AdiSP:**" set-
ting appears.
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
“Ad-on” Function activated “Ad-oF” Function deactivated
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode.
Display will appear when about one third of capacity is exhausted.
ABT-BA-e-1312
54
14 Application Functions
14.1 Parts counting
With parts counting you can either count parts into a container or remove parts from a container. To count a greater number of parts the average weight per part has to be determined with a small quantity (reference quantity). The larger the reference quantity, the higher the counting exactness. High reference must be selected for small parts or parts with considerably different sizes. The process has four steps:
Tare the weighing container
Determine the reference unit
Weigh in the reference unit
Count the items
Condition: Function "U- PSC" activated (ch. 11.3)
(When PCS is used for the first time)
Repeatedly press the [UNIT] until display is in percentage mode. The display symbol PCS pops up.
If you are using a weighing container use the [TARE] key for
taring.
To determine the reference piece number, place 10, 20, 50 or 100 counting parts on the balance.
Press the [CAL] key
Repeated pressing of the [CAL] key allows the change between
the following reference piece numbers "Ld 10", “Ld 20”,“Ld 50” and “Ld 100”.
Important: The larger the reference quantity, the more accurate the parts counting.
(When loading 20 pieces)
When standstill control ( ) is complete, confirm your reference
piece number by pressing the [TARE] key.
The display shows "SEt" for several seconds and the stored reference piece number is displayed.
ABT-BA-e-1312
55
Remove reference weight.
Now you can fill the items to be counted into the container. The respective quantity is shown in the display.
Repeated pressing of the [UNIT] key changes the display value, e. g.: [g] → [%] → [Pcs] → [ct]
Note:
Error message "Err 20" indicates that the weight for the reference piece number is too low.
14.2 Percent determination
Percent weighing allows to display weight in percent, in relation to a reference weight. The displayed weighing value is stored as a standard percent value (default setting: 100%).
Condition: Function "U- %" is activated (ch. 11.3)
(When PCS is used for the first time)
Repeatedly press the [UNIT] key until the display is in percent-
age determination mode. The display symbol % pops up.
If you are using a weighing container use the [TARE] key for
taring.
Put on reference weight (=100%) (minimum weight: Readability d x 100)
When standstill control ( ) is complete, press the [CAL] key.
The reference weight is saved.
Remove reference weight.
From now on the placed weight will be shown as %.
Repeated pressing of the [UNIT] key changes the display value, e. g.: [g] → [%] → [Pcs] → [ct]
ABT-BA-e-1312
56
14.3 Density determination
Density determination with the help of the flush-mounted platform facility is described in the following. Density determination becomes even easier when an optional set for density deter­mination is applied. For further information please refer to the operating instructions enclosed with the set for density determination.
1. Remove the below-weigh hook cover on the floor of the balance, after remov­ing the two fixing screws.
2. Hang the user-prepared hanging pan from the below-weigh hook and sink that hanging pan into the sample liquid in the tank.
14.3.1 Density determination of solids Condition: Function ",d" (density of solids) is activated. See Chapter 11.3.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key.
Repeatedly actuate the [CAL] key until "LSG SEt" appears
(Example)
Press the [TARE] key. The currently set density for the liquid to be
measured appears. In the upper part of the display panel, the
symbol and the # symbol appear in order to indicate numer-
ical input status. The leftmost digit blinks.
(Example)
Enter density for your liquid to be measured. When the [UNIT] key
is pressed, the numerical of the blinking digit increases by 1 at a time. You can determine the value of the flashing digit, or shift the
flashing digit by one position to the right, by pressing the [PRINT] key. Confirm your setting by pressing the [TARE] key.
Repeatedly press the [ON/OFF] key until the balance is in weigh-
ing mode.
ABT-BA-e-1312
57
Repeatedly press the [UNIT] key until the balance is in density
determination mode for solids ",d". Note that “g” also appears dur-
ing weight measurement in air. Press the [TARE] key. Place the item to be measured on the
weighing pan. When standstill control is complete, press the [CAL] key
Place the item to be measured on the immersed hanging weighing pan. The display is showing the density of the measured item. “dSP oL” may be displayed when nothing is on the weighing pan, which is normal.
14.3.2 Determining density of liquids Condition: Function "d" (density of liquids) is activated. See Chapter 11.3.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key.
Repeatedly actuate the [CAL] key until "Sv SEt" appears
(Example)
Press the [TARE] key. The currently set density for the body to be
immersed appears. In the upper part of the display panel, the
symbol and the # symbol appear in order to indicate numer-
ical input status. The leftmost digit blinks.
(Example)
Enter density for your body to be immersed. When the [UNIT] key
is pressed, the numerical of the blinking digit increases by 1 at a time. You can determine the value of the flashing digit, or shift the
flashing digit by one position to the right, by pressing the [PRINT] key. Confirm your setting by pressing the [TARE] key.
Repeatedly press the [ON/OFF] key until the balance is in weigh-
ing mode.
ABT-BA-e-1312
58
Repeatedly press the [UNIT] key until the balance is in density
determination mode for liquids "d". Note that “g” also appears dur-
ing weight measurement in air. Press the [TARE] key. Place the body to be immersed on the
weighing pan. When standstill control is complete, press the [CAL] key
Immerse the body to be immersed without bubbles developing. The display is showing the density of the sample. “dSP oL” may be displayed when nothing is on the weighing pan, which is nor­mal.
ABT-BA-e-1312
59
14.4 Auto Print
Using Auto Print allows measurement results to be automatically output via the RS-
232C interface without pressing the [PRINT] key individually with every measure-
ment. An issue will be made after standstill control ( ) is complete. The next meas­urement is done after removing the items from the weighing plate and the display is returning to a value within the ±3-fold of the zero range.
Important:
This mode can be applied to any unit. The initial load of the balance must be kept within the ± 5-fold of the ze-
ro range.
Weight of sample 10 x zero range (zero range = 0.25 e)
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
(Example)
Repeatedly press the [CAL] key until the current "AtPrt:**" set-
ting appears.
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
“ AtPrt-on” Function activated “ At Prt-oF” Function deactivated
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode. Whilst the auto print function is active the display keeps showing the [AP] symbol.
ABT-BA-e-1312
60
14.5 Add up mode
An optional number of single weighing processes are automatically added up to a total sum, e.g. all single weighing processes of a batch or if you have to carry out a great number of measurements of small samples. When the standstill control ( ) is complete the weighing value is automatically is­sued to the optional printer. The displayed value is added into the total adding memory. Afterwards automatic taring will take place. This is repeated newly every subsequent time a new sample is placed on the weighing pan. When the last single
weighing process is finished, press the [ON/OFF] key to receive the total sum ("TO-
TAL=").
Important:
This mode can be applied to any unit. The initial load of the balance must be kept within the ± 5-fold of the ze-
ro range.
Weight of sample 10 x zero range (zero range = 0.25 e) If during the use of multiple range balances the smallest readability digit
was turned off by pressing the [1d/10d] key, evaluation is based on the
latest add-on value displayed before.
Activate function:
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
(Example)
Repeatedly press the [CAL] key until the current "Addon:**" set-
ting appears.
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
“ Adon-on” Function activated “ Adon-oF” Function deactivated
The current setting is indicated by the standstill display ( ).
ABT-BA-e-1312
61
Confirm your selection by pressing the [TARE] key. Printout "----
---- ADDON MODE ---------" takes place.
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key: Back to weighing mode. Whilst the add-on function is active, the display keeps showing the add-on symbol [ ].
Add up:
Operation Display
When using a container, place this on the weighing plate and tare.
Prior to weighing the component press the [PRINT] key 1x.
Weigh first component.
5.0003
g
Wait until standstill display ( ) appears, the weighing value for the first component (CMP001) will be issued automatically to the optional printer. This is followed by automatic taring.
0.0000
g
Weigh second component
5.0017
g
Wait until standstill display ( ) appears, the weighing value for the second component (CMP002) will be issued automatically to the optional printer. This is followed by automatic taring.
0.0000
g
Weigh third component
5.0010
g
When the last single weighing process is finished, press the
[ON/OFF] key to receive the total sum ("TOTAL=").
15.0030
g
ABT-BA-e-1312
62
Printout:
---ADDON Mode---
CMP001=
5,0003 g
CMP002=
5,0017 g
CMP003=
5,0010 g
TOTAL=
15,0030 g
Note: Only one function at a time may be activated (on) when working with func-
tions auto print, add-on and recipe.
Whilst the function is active, automatic adjustment (PSC/Clock-CAL) does
not take place. A flashing symbol, however, indicates that adjustment is required.
14.6 Recipe mode
The formula function allows to add on various components of a mixture. The total weight of all components can be invoked to check (separate memory for the weight of tare vessel and recipe components).
Activate function "FormU: on“ (see chap. 14.5 Add up):
Whilst the recipe function is active, the display keeps showing the "FormU -on" sym-
bol [ M].
Operation Display
When using a container, place this on the weighing plate and tare.
0.0000 g
Weigh first component.
0.5361
g
When standstill control ( ) is complete, the weighing value of the first component (CMP001) is issued to the
optional printer by pressing the [PRINT] key. The dis-
played value is added into the total adding memory. Af­terwards automatic taring will take place.
0.5361
g
0.0000
g
Weigh second component
0.5422
g
When standstill control ( ) is complete, the weighing value of the second component (CMP002) is issued to
the optional printer by pressing the [PRINT] key. The
displayed value is added into the total adding memory. Afterwards automatic taring will take place.
0.54221
g
0.0000
g
ABT-BA-e-1312
63
Place further components
0.4488
g
When the last single weighing process is complete press
[ON/OFF] key to receive the total sum (Total).
1.5271
g
Remove weight from weighing plate. The balance is now ready for further measurements.
Printout:
---Formulation Mode---
CMP001=
0,5361 g
CMP002=
0,5422 g
CMP003=
0,4488 g
TOTAL=
1,5271 g
Note: Only one function at a time may be activated (on) when working with func-
tions auto print, add-on and recipe.
Whilst the function is active, automatic adjustment (PSC/Clock-CAL) does
not take place. A flashing symbol, however, indicates that adjustment is required.
ABT-BA-e-1312
64
15 Data output
15.1 RS 232C interface
Pin allocation of balance output plug:
PIN No.
Signal
2
TXD
3
RXD
6
DSR
7
SG
20
DTR
5
CTS
4
RTS
15.2 Data Formats
Note: space character and DL the delimiter code.
1. Input Data Format COMMAND CODE + DL (see 15.2.3)
2. Output data format In weighing mode:
(ex.) S -200.000gDL
stability
polarity
Polarity data
Positive: space character (() Negative: minus (-)
Stability information (when output includes stability information)
Stable : S Unstable : U
In “oL” or “-oL” display
(ex.) U -

oLDL
stability
polarity
Polarity data
Positive: space character () Negative: minus (-)
Stability data (when output includes stability information)
Stable : S Unstable : U
3. Data format
ASCII (JIS) code Baud rate, parity (and bit length), delimiter, stop bit, format, and handshake
differ by menu item selections.
ABT-BA-e-1312
65
15.3 Remote control instructions
Attention:
Inputting characters and command codes not shown here into the balance may not only alter the previous settings but may also impair proper measurement. If by mistake characters or commands not shown here are entered into the balance, immediately unplug the power supply cable and wait about ten seconds before plug­ging again.
Command Code
Function
Description
D01
Continuous output
The balance continuously outputs every 110ms .
D05
1 time output
Corresponds to [PRINT] key
D06
Auto Print
See 14.4
D07
1 time output with stability
information
The status of the stability mark is appended to the head
of the data with output.
S: when the standstill display appears
U: when standstill display does not appear
D08 1 time output at stability
After command input, the data are output at the first
appearance of the standstill display.
D09
Halt output
Auto Print and continuous output halted
Q ON/OFF switching
Swit
ches between standby status and measurement
status. T Taring
Corresponds to the [O/T] key
TS Taring after stability wait
After command input, taring is done at the first appear-
ance of the stability mark.
C18
Measurement span calibration
+ Measurements in add-up mode
See ch. 14.5 R Total reset
All application measurements terminated and reset
mg
mg unit
PERCENT
Percent determination
PCS
Parts counting
CT
ct unit
SDENCE
Density of solids
LDENCE
Density of liquids
%
100% setting
G
g, % switching
- g
g unit removal
- mg
mg unit removal
- PERCENT
Deleting percentage determina-
tion t
- PCS
Piece counting removal
- CT
ct unit removal
- SDENCE
Solid specific gravity removal
- LDENCE
Liquid specific gravity removal
ABT-BA-e-1312
66
C02
High-stability mode setting
C13
Anti-convection mode setting
C14
Standard mode setting
C05
Standstill display,
setting to 1 count
C06
Standstill display,
setting to 5 count
C15
Standstill display,
setting to 10 count
C07
Auto Zero on
C08
Auto zero off
C10
Autom. CAL on
C11
Autom. CAL off
C17 Display setting status
Measurement conditions set by menu selection in abbre-
viated form is output.
15.4 Standard Settings
Display
for menu
selection
Baud
Rate
Delimiter
Parity
(Bit
length)
Stop bit
Data
format
Hand­shake
Standard set-
ting 1
iF:F1 1200 C/R None (8) 1 dF1 Hardware
Standard set-
ting 2
iF:F2 1200 C/R None (8) 1 dF2 Hardware
Standard
setting 3
iF:F3 2400 C/R+L/F Even (7) 1 dF3 Hardware
Standard set-
ting 4
iF:F4 1200 C/R+L/F Odd (7) 1 dF4 Hardware
Operator
settings
(see 15.5)
iF:USEr
User
settings
User
settings
User
settings
User
settings
User
settings
User set-
tings
KERN–
YKB-01N
iF:USEr 1200 C/R None (8) 1 dF1 off
Selecting one of standard Settings:
Repeatedly press the [CAL] key until "intFACE" appears.
Press [TARE] key
Repeatedly press the [CAL] key until the desired standard setting
appears.
Confirm your selection by pressing the [TARE] key. Repeatedly press the [ON/OFF] key. The balance returns to toler-
ance weighing mode.
ABT-BA-e-1312
67
15.5 User Settings
The user setting allows individual setting for each item in communication settings.
Repeatedly press the [CAL] key until "intFACE" appears.
Press [TARE] key
Repeatedly press the [CAL] key until "iF:USEr" appears.
(Example)
Press [TARE] key
Use the [CAL] key to select between the following settings
(The **** reflect the current setting, two to four characters):
“io.b:****” Speed of communication “io.d:****” Delimiter “io.P:****” Parity “io.S:****” Stop bit “io.F:****” Data format “io.H:****” handshake
Confirm your selection by pressing the [TARE] key and set the
parameters you require as follows:
ABT-BA-e-1312
68
15.5.1 Communication speed settings
1. The display changes from “io.b:****” to “b-300”. Pressing the [CAL] key changes the
display. The standstill display ( ) marks the current setting.
Display during setting
b-300
b-600
b-1200
b-2400
b-4800
Setting specifics
300bps
Baud Rate
600bps
Baud Rate
1200bps
Baud Rate
2400bps
Baud Rate
4800bps
Baud Rate
Display during setting
b-9600
b-19.2K
b-38.4K
Setting specifics
9600bps Baud Rate
19.2Kbps Baud Rate
38.4Kbps Baud Rate
2. Confirm your desired setting by pressing the [TARE] key.
3. To return to "io.S:****", press the [ON/OFF] key.
15.5.2 Delimiter settings
1. The display changes from “io.d:****” to “d-Cr”. Pressing the [CAL] key changes the
display. The standstill display ( ) marks the current setting.
Display during setting
d-Cr
d-LF
d-CrLF
d-Cn
d-win 1
d-win -
Setting specifics
C/R L/F C/R + L/F Comma Not documented
2. Confirm your desired setting by pressing the [TARE] key.
3. To return to "io.d:****", press the [ON/OFF] key.
15.5.3 Parity settings
1. The display changes from “io.P:****” to “P-no”. Pressing the [CAL] key changes the
display. The standstill display ( ) marks the current setting.
Display during setting
P-no
P-odd
P-EvEn
Setting specifics
No parity
(eight bits)
Odd parity
(seven bits)
Even parity
(seven bits)
2. Confirm your desired setting by pressing the [TARE] key.
3. To return to "io.P:****", press the [ON/OFF] key.
ABT-BA-e-1312
69
15.5.4 Stop bit settings
1. The display changes from “io.S:****” to “S-S1”. Pressing the [CAL] key changes the
display. The standstill display ( ) marks the current setting.
Display during setting
S-S1
S-S2
Setting specifics
Stop bit, 1 bit Stop bit, 2bit
2. Confirm your desired setting by pressing the [TARE] key.
3. To return to "io.S:****", press the [ON/OFF] key.
15.5.5 Input-output data format settings
1. The display changes from "io.F:****" to " F-dF1". Pressing the [CAL] key changes
the display. The standstill display ( ) marks the current setting.
Display during setting
F-dF1
F-dF2
F-dF3
F-dF4
Setting specifics
Data format 1. Standard format
Data format 2. Not documented
Data format 3. Not documented
Data format 4. Not documented
2. Confirm your desired setting by pressing the [TARE] key.
3. To return to "io.F:****", press the [ON/OFF] key.
Note: When set to data format 2, the balance will always send a process result
against commands from the computer.
15.5.6 Handshake settings
1. The display changes from "io.H:****" to " H-oFF". Pressing the [CAL] key changes
the display. The standstill display ( ) marks the current setting.
Display during setting
H-oFF
H-Soft
H-HArd
H-tr
Setting specifics
No handshake
Software
handshake
Hardware
handshake
Timer
handshake
2. Confirm your desired setting by pressing the [TARE] key.
3. To return to "io.H:****", press the [ON/OFF] key.
ABT-BA-e-1312
70
16 Service, maintenance, disposal
16.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vac­uum cleaner.
Spilled weighing goods must be removed immediately.
16.2 Service, maintenance
The appliance may only be opened by trained service technicians who are author­ized by KERN. Before opening, disconnect from power supply.
16.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
ABT-BA-e-1312
71
17 Instant help
In case of an error in the program process, briefly turn off the balance and discon­nect from power supply. The weighing process must then be restarted from the be­ginning.
Error code table:
Error code display
Explication
Remedy
CAL E2
High zero point shift dur­ing adjustment
Remove items from the weighing pan.
CAL E3
Great deviation of meas­uring values in PCAL.
Use correct adjusting weight.
CAL E4
Great deviation of meas-
uring values during ad-
justment
CHE X (X is a numeral)
(when the display stops
here)
Internal malfunction
Please contact a service representative.
Err 0X (X is a numeral) Internal malfunction
Please contact a service representative.
Err 20
An improper value setting was attempted.
Enter the correct numbers or decimal points.
Err 24
Power voltage error
Check the power voltage .
Possible causes of errors:
When Symptom Possible causes Remedy
Before
meas­urement
Nothing appears in the
display.
The AC adapter is not securely
connected. The power switchboard of the room is turned off. The power
voltage is incorrect.
Check power supply and
connect correctly.
During
meas­urement
The display fluctuates.
The standstill display does not appear fast enough.
The measured results have poor repeatabil­ity.
“CAL d” appears fre-
quently.
Vibrations or draft.
Change the installation site.
Change stability and reac­tion settings or standstill display.
ABT-BA-e-1312
72
Attempting to measure volatile sub­stances
Measure with a lid on it.
The weighed item is electrically
charged.
Measure in a metal contain-
er. Measure with a metal
object larger than the items.
Sample temperature and tempera-
ture inside weighing space differ.
Measure at the same tem-
perature. Leave the item in the chamber before meas­urement. Change to high-
stability mode.
Draft in weighing space.
Leave the glass doors of the
weighing chamber open 1 to
2 cm wide when not in use.
Effects of electronic noise or strong electromagnetic waves
Move away from the noise source.
Internal trouble with the balance
Contact a service repre-
sentative.
“oL” or “-oL” is dis­played
The load on the weighing pan is too
large. The weighing pan is de-
tached.
Use within the weighing
capacity. Place the weighing
pan on correctly.
Automatic adjustment carried out frequently.
Severe temperature variations in the room or the instrument
Move to a location with less temperature fluctuation.
Display is faulty.
Adjustment is not carried out.
Carry out correct adjust­ment.
No taring to zero before weighing.
To reset the display before
weighing, actuate the
[TARE] key.
The desired weighing
unit cannot be called
by [UNIT] key.
Unit was not activated beforehand.
Set unit beforehand.
Cannot transmit or
receive data to or from
computer or device.
Communication settings are wrong.
Make the proper communi-
cation settings.
Error message ap­pears.
Refer to the error code ta­ble.
During
adjust-
ment
Error message ap-
pears.
Refer to the error code ta-
ble.
During
menu item se-
lection
The menu settings
cannot be changed.
The menu is locked
Remove the menu lock.
Loading...