Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility . Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT , INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF
THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS
BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION,
LOSSES SUSTAINED AS THE RESULT OF INJURY T O ANY PERSON, OR DAMAGE TO PROPERTY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Le vel letter increases alphabetically as the manual under goes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially . When a new Re vision is created, all Addenda associated with the previous Re vision
of the manual are incorporated into the new Revision of the manual. Each ne w Revision includes
a revised copy of this print history page.
Revision A (Document Number 6517A-903-01)...............................................................July 2000
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous
voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar
with the safety precautions required to avoid possible injury. Read the operating information carefully
before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for
ensuring that the equipment is operated within its specifications and operating limits, and for ensuring
that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures
and proper use of the instrument. They must be protected from electric shock and contact with hazardous
live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the
manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be
performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of prod-
ucts. Only properly trained service personnel may perform installation and service procedures.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable con-
nector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present.
safety practice is to expect that hazardous voltage is present in any unknown cir cuit before measuring.
Users of this product must be protected from electric shock at all times. The responsible body must ensure that users are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product users in these circumstances must be
trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or
above 1000 volts,
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital multimeter measuring circuits (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation
Category II. All other instruments’ signal terminals are Installation Category I and must not be connected to mains.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with
impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting
sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any
capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching
cards, or making internal changes, such as installing or removing jumpers.
no conductive part of the circuit may be exposed.
A good
Do not touch any object that could provide a current path to the common side of the circuit under test or power
line (earth) ground. Alw ays make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions
or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications
and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth
ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in
the manual.
The symbol on an instrument sho ws that it can source or measure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to av oid personal contact
with these voltages.
The
WARNING heading in a manual explains dangers that might result in personal injury or death. Always
read the associated information very carefully before performing the indicated procedure.
The
CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the
power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses,
with applicable national safety approvals, may be used if the rating and type are the same. Other components
that are not safety related may be purchased from other suppliers as long as they are equivalent to the original
component. (Note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component,
call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument
only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a
computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
If you have any questions after re viewing this information, please contact your local K eithle y
representative, or call one of our Applications Engineers at
only). Worldwide phone numbers are listed in the front of this manual.
1-800-348-3735 (U.S. and Canada
The Model 6517A is a 6
surement capabilities:
•DC voltage measurements from 1µV to 210V
•DC current measurements from 10aA to 21mA
•Charge measurements from 10fC to 2.1µC
•Resistance measurements from 10
•Surface resistivity measurements
•Volume resistivity measurements
•External temperature measurements from -25°C to 150°C using the supplied Model
6517-TP thermocouple
•Relative humidity measurements (0 to 100%) using the optional Model 6517-RH probe
½-digit electrometer/high resistance system with the following mea-
Ω to 210T Ω
Front Panel Operation1-3
Some additional capabilities of the Model 6517A include:
•Built-in V-Source — The 100V range provides up to 100V at 10mA, while the 1000V
range provides up to 1000V at 1mA
•Data storage
•Single button zeroing (REL)
•Built-in math functions
•Filtering; averaging and median
•Built-in test sequences
•Remote operation using the IEEE-488 (GPIB) bus or the RS-232 interface
•Model 617 emulation using the DDC (device-dependent command) programming language mode
•Scan (measure) channels of an external scanner (e.g., Model 706, 7001 or 7002)
•Scan (measure) channels of an internal scanner card (e.g., Model 6521 or 6522) installed
in the option slot
The Model 6517A front and rear panel overvie ws (Figures 1 and 2) include important abbre-
viated information that should be reviewed before operating the instrument.
1-4Front Panel Operation
Figure 1
Model 6517A front
panel overview
1 ANNUNCIATORS
ERRQuestionable reading
REMIn remote
TALKAddressed to talk
LSTNAddressed to listen
SRQService Request
RELRelative reading displayed
FILTDigital filter enabled
MATHMath calculation enabled
AUTOAutoranging enabled
ARMTrigger armed; not in idle
*(asterisk) Reading being stored
Secondary Display Line
Primary Display Line
19
EDIT ERR REM TALK LSTN SRQ REAR REL FILT MATH4W AUTO ARM TRIG SMPL
1234567891012345678910
PREV
DISPLAY
NEXT
POWER
6517 ELECTROMETER/HIGH RESISTANCE METER
VI RQ
RELZ-CHK
FILTER MATH
INFO LOCAL
STORE RECALL
OPTIONVOLTAGE SOURCE
CARD OPER
TRIG SEQ
EXIT ENTERCONFIG MENU
RANGE
AUTO
RANGE
VOLTAGE
SOURCE
OPERATE
CAL
52673 84
2 FUNCTION KEYS Select measurement function; volts (V), amps (I), resistivity (R) or
coulombs (Q)
3 RANGE KEYS
▲Moves to higher range; increments digit
▼Moves to lower range; decrements digit
AUTOEnables/disables autorange
Front Panel Operation1-5
4 HANDLE (not shown) Pull out and rotate to desired position.
5 DISPLAY KEYS PREV/NEXT Scroll through multiple displays of a function
6 OPERATION KEYS
Z-CHKEnables/disables zero check. Enable zero check before changing functions.
RELEnables/disables relative reading
FILTERDisplays digital filter status for present function and toggles filter on/off
MATHDisplays math calculation and toggles math on/off if configured
TRIGTriggers unit
SEQPerforms selected test sequence
and Moves cursor among data entry digits, menu selections, and information displays
INFOShows context-sensitive information about the present display
LOCALCancels remote operation
STOREEnables data storage
RECALL Displays reading data (reading, number, time). Use PREV/NEXT DISPLAY for
maximum, minimum average, standard deviation.
CONFIG Configures functions and operations
MENUSaves/restores instrument conditions; sets up communications; performs calibra-
tion and self-tests; defines limits, digital output, and other miscellaneous opera-
tions (see Table 2)
EXITCancels selection, moves back within menu structure
ENTERHolds reading, enters selection, moves down within menu structure
7 VOLTAGE SOURCE KEYS
OPERToggles V-source between operate and standby
▲ and ▼ Adjusts V-source value
8 VOLTAGE SOURCE OPERATE Indicator on when in operate, off when in standby
9 OPTION CARD KEYUse to program and operate an installed option. Also used to scan
external scanner channels.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
1-6Front Panel Operation
Figure 2
Model 6517A rear
panel overview
23 5 49 810
1010V
PEAK
HI
SELECTED
LINE VOLTAGE
90-110V
105-125V
OUT
DIGITAL
TRIG LINK
I/O
!
INTERLOCK
TTL HI
TTL LO
BNC CONNECTIONS
INPUT
250V PEAK
1
PREAMP OUT
TEMP
TYPE K
250 PEAK
OPTION SLOT
17
!
Input High
Input Low
Chassis Ground
Unguarded (Guard off)
Input High
Guard
Chassis Ground
Guarded (Guard on)
LO
COMMON 2V ANALOG
OUTPUT
HUMIDITY
750V
PEAK
EXT TRIGINMTR COMP
RS232
V SOURCE
6715 1614 1312
EXTERNAL TRIGGER INPUT
Trigger Reading
>10µsec
180-220V
210-250V
115V
LINE RATING
50VA MAX
50-60HZ
AC ONLY
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
LINE FUSE
SLOWBLOW
1/2A 90-125V
1/4A 180-250V
METER COMPLETE OUTPUT
Reading
Complete
>10µsec
TTL HI
TTL LO
HI
LO (Chassis)
MADE
U.S.A.
IN
11
Front Panel Operation1-7
1 INPUT CONNECTOR (see Figure 2)
Unguarded ConfigurationDisable GUARD for amps, ohms, coulombs and unguarded
volts measurements.
Guarded ConfigurationEnable GUARD for guarded volts measurements.
2 TEMP TYPE K Connect the Model 6517-TP type K thermocouple for temperature mea-
surements.
3 PREAMP OUT Follows the signal amplitude applied to the INPUT terminal. With
GUARD on, PREAMP OUT is internally connected to the inner shell of the INPUT triax
connector to configure the input for guarded volts measurements. Referenced to COMMON. See the Model 6517A User’s Manual for more information.
4 2V OUTPUT Binding post provides a scaled 0 to 2V output that is referenced to COM-
MON. Typically connected to a measuring device such as a chart recorder. See the Model
6517A User’s Manual for more information.
5 COMMON Binding post that is internally connected to INPUT low through a 0.6Ω resis-
tor.
6 CHASSIS GROUND Binding post that connects the chassis to ground through the power
line cord. COMMON can be connected to chassis ground by installing the ground link. For
floating measurements, make sure the ground link connection between COMMON and
Chassis Ground is open.
7 HUMIDITY Connect the optional Model 6517-RH probe for relative humidity measure-
ments.
8 V-SOURCE HI and LO Banana jacks for the voltage source.
9 LINE VOLTAGE SWITCH Two-position slide switch configures instrument for avail-
able line power voltage. A 90 to 125V line voltage requires the 115V setting, while a 180 to
250V line voltage requires the 230V setting.
10 POWER LINE INPUT Connect to power line (50 or 60Hz) grounded outlet using 3-wire
power cord.
11 LINE FUSE For the 115V line voltage setting, use a ½ A, 250V, slow blow, 5×20mm
fuse. For the 230V setting, use a ¼ A, 250V, slow blow, 5×20mm fuse.
12 IEEE-488 CONNECTOR Connects the instrument to the IEEE-488 (GPIB) bus. Use
shielded IEEE-488 cables.
13 TRIG LINK An 8-pin micro DIN connector for sending and receiving trigger pulses to
and from other instruments.
14 DIGITAL I/O An 8-pin micro DIN connector for the four TTL-compatible digital output
lines.
15 RS-232 DB-9 connector for the RS-232 interface. Use a standard RS-232 cable.
16 INTERLOCK Connects the safety interlock to the Model 8009 Resistivity Test Fixture
or the Model 8002A High Resistance Test Fixture using the appropriate cable.
17 OPTION SLOT An option card, such as the Model 6521 or 6522 scanner card, installs in
this slot.
1-8Front Panel Operation
Display
As shown in the Model 6517A front panel ov erview (Figure 1), the front panel has three lines
of display information: the primary display line, the secondary display line, and annunciators.
•Primary Display Line: The top line displays readings along with units. It can also display
measurement type, “hold”, math operating type, channel number, menu headings and
messages. Error and status messages are listed in Section 2 of the User’s Manual.
•Secondary Display Line: The bottom line displays the range, V-Source value, menu
items, messages and multiple displays (see Measurement Options section of this manual). Longer text strings are indicated by arrows on either end of the display line. Use the
cursor keys ( and ) to display the additional information.
Power-up
WARNING
W arm-up
Before turning on the Model 6517A, make sure it is connected to a grounded
power receptacle using the supplied power cord or the equivalent. Failure
to properly ground the unit creates a shock hazard that could result in injury or death.
Use the front panel POWER switch to turn power on and off.
During power-up, the Model 6517A performs self-tests on its memory elements. If a failure
occurs, the instrument momentarily displays an error message and turns on the ERR annunciator. If the instrument passes the self-tests, the fi rmware re vision lev el, and bus interf ace information are displayed.
For complete information on the power-up sequence, see Section 2 in the User’s Manual.
The instrument is ready for use as soon as the power-up sequence is completed. However, to
achieve rated accuracy, allow the instrument to warm up for one hour. If the instrument has been
subjected to extreme temperatures, allow additional time to allow internal temperatures to stabilize.
Bench defaults
The Model 6517A can save ten user setups in non-v olatile memory. You can select one of the
user setups as the power-on default or ha v e the instrument power up to either of the two factory
defaults (optimized for “BENCH” or “GPIB” operation).
Since the basic measurement procedures in this manual assume the BENCH defaults, reset
the instrument from the SAVESETUP item of the MAIN MENU by performing the following
steps:
1.Press the MENU key to display the main menu:
MAIN MENU
SAVESETUP COMMUNICATION CAL
2.If the SAVESETUP option is not blinking, press the key until it is blinking and then
press ENTER to view the setup menu:
SETUP MENU
SAVE RESTORE POWERON RESET
3.Select the RESET option and press ENTER to view the reset menu:
RESET ORIGINAL DFLTS
BENCH GPIB
4.Select the BENCH option and press ENTER. The following message is displayed:
RESETTING INSTRUMENT
Enter to confirm; EXIT to abort
5.Press ENTER to confirm. The display will go to the volts function with auto-range disabled and zero check enabled.
TEST LIMITS STATUS-MSG GENERAL
Front Panel Operation1-9
Other BENCH default settings for each function include:
•Triggers — Continuous measurements.
•Measurement speed (integration time) — Normal, 1 power line cycle.
•Digital filter — Enabled, averaging, 10 readings.
•Median filter — Enabled, rank 1.
•Display resolution — 5
Note: See “Menus” in the Measurement Options section of this manual for details on navi-
gating the Main Menu.
Changing functions
T o avoid erratic operation, always enable zero check (“ZeroCheck” displayed) before selecting one of the other measurement functions (V, I, R or Q). The Z-CHK key controls zero check.
See “Zero Check” in Section 2 for more information on the zero check feature.
½ digits.
1-10Front Panel Operation
V oltage measurements
Assuming “BENCH” reset conditions, the basic procedure is as follows:
1.With zero check enabled (“ZeroCheck” displayed), select the v olts function by pressing
V. The Z-CHK key is used to enable/disable zero check.
2.Enable or disable guard as needed. See “Guarding” at the end of this procedure.
NOTEWith guard enabled (ON), the “Grd” message is displayed. However, if zero correct
is already enabled (“ZCor” displayed), then “Grd” will not be displayed.
3.To achieve optimum accuracy for low voltage measurements, zero correct the Model
6517A. To do this, use the RANGE
and press REL. With zero correct enabled, the REL indicator is on.
▼ key to select the lowest measurement range (2V)
NOTEWith zero correct enabled, the “ZCor” message is displayed. However, if guard is al-
WARNING
CAUTION
ready enabled (“Grd” displayed), then “ZCor” will not be displayed.
4.Select a measurement range or use autoranging.
A. To automatically select the most sensitive range, enable autorange. The AUTO key
enables and disables autorange. When enabled, the AUTO annunciator is on.
B. For manual ranging, use the RANGE
consistent with the expected voltage.
5.Connect the test cable(s) to the voltage to be measured. Figure 3A sho ws connections for
unguarded measurements, and Figure 3B shows connections for guarded measurements.
Hazardous voltage may be present on the inner shield of the triax cable
when GUARD is on. A safety shield connected to safety earth ground (as
shown in Figure 3B) should be used for voltage measurements at or above
30V.
Do not apply more than 250V peak (DC to 60Hz) to the input or instrument
damage may occur.
6.Press Z-CHK to disable zero check and take a reading from the display.
▲ and ▼ keys to select a measurement range
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS
CAUTION:FOR CONTINUED PROTECTION AG
CAUTION:FOR CONTINUED PROTECTION AG
WARNING:NO INTERNAL OPERATOR
WARNING:NO INTERNAL OPERATOR
Front Panel Operation1-11
Figure 3
Voltage measurements
A. Unguarded (GUARD off)
B. Guarded (GUARD on)
Measured
Voltage
Shield (Optional)
Safety
Earth
Ground
Measured
Voltage
Guard
6517
237-ALG-2
Red (HI)
Black
Green
LO
Safety Shield
Cable
!
237-ALG-2
Cable
INPUT
250V PEAK
6517
!
250V PEAK
INPUT
COMMON
COMMON
Red (HI)
+
Vs
-
Black (LO)
+
Vs
-
Guarding
measurements that use long input cables. In a high impedance circuit, guarding greatly reduces
leakage currents. When using long input cables, guarding cancels the effects of cable capacitance that can significantly slow down the measurement response time.
tential to the inner shell of the INPUT triax connector. With this configuration, the COMMON
banana jack is used for input low . Figure 3 sho ws both the guarded and unguarded configuration
of the INPUT connector.
Guarding should be used for high impedance ( ≥10
9
Ω) voltage measurements and for voltage
When GUARD is enabled, the INPUT triax connector is reconfigured to apply the guard po-
Perform the following steps to enable or disable guard:
1.Press CONFIG and then V to display the volts configuration menu.
2.Use the and keys to place the cursor (blinking menu item) on GUARD and press
ENTER.
3.Place the cursor on the desired selection (OFF to disable or ON to enable) and press ENTER.
4.Use the EXIT key to back out of the menu.
CAUTION:
FOR CONTIN
CAUTION:
FOR CONTIN
WARNING:
NO IN
WARNING:
NO IN
1-12Front Panel Operation
Current measurements
Assuming “BENCH” reset conditions, the basic procedure is as follows:
1.With zero check enabled (“ZeroCheck” displayed), select the amps function by pressing
I. The Z-CHK key is used to enable/disable zero check.
2.To achieve optimum accuracy for low current measurements, zero correct the Model
6517A. To do this, use the RANGE
(20pA) and press REL. W ith zero correct enabled, the REL indicator is on, and the message “ZCor” is displayed.
3.Select a measurement range or use autoranging:
A. To automatically select the most sensitive range, enable autorange. The AUTO key
enables and disables autorange. When enabled, the AUTO annunciator is on.
B. For manual ranging, use the RANGE
consistent with the expected current.
4.Connect the instrument to the current to be measured as shown in Figure 4.
▼ key to select the lowest measurement range
▲ and ▼ keys to select a measurement range
CAUTION
5.Press Z-CHK to disable zero check and take a reading from the display.
NOTETo disable zero correct, enable zero check and then press REL.
Figure 4
Current measurements
Do not apply more than 250V peak (DC to 60Hz; 10 seconds per minute
maximum on mA ranges), or instrument damage may occur.
6517
237-ALG-2
Black (LO)
Cable
!
INPUT
250V PEAK
COMMON
Measured
Current
Shield (Recommended
below 1µA)
CAUTION: Maximum Input = 250V peak (DC to
60Hz) ; 10s per minute maximum on
mA ranges.
I
Red (HI)
s
Input low connected
to shield
Resistance measurements
NOTEAccurate high resistance measurements require the use of a low-leakage test fixture,
such as the Model 8002A.
The following procedure requires “BENCH” reset conditions. See “Bench defaults” to reset
the instrument.
1.With zero check enabled (“ZeroCheck” displayed), select the ohms function by pressing
R.
2.Select Auto V-Source ohms. In this mode, the V-Source is automatically set to a voltage
(nominally 40V or 400V) that optimizes measurement accuracy . See “ Auto V-Source” to
select Auto V -Source ohms.
Front Panel Operation1-13
NOTE“BENCH” reset conditions selects Manual V-Source ohms. In this mode, you must
manually set the V -Source le vel for the measurement. See “Voltage Source” in Section
2 for details on using the V-Source.
WARNINGSelecting Auto V-Source ohms may set the V-Source to 400V.
3.Select a measurement range or use autoranging:
A. To automatically select the most sensitive range, enable autorange. The AUTO key
enables and disables autorange. When enabled, the AUTO annunciator is on.
B. For manual ranging, use the RANGE ▲ and ▼ keys to select a measurement range
consistent with the expected resistance.
4.Connect the Model 6517A to the test fixture. Generic connections are shown in Figure
5A. Note that this connection scheme requires that ammeter LO be internally connected
to V-Source LO. See “Configuring V-Source” in Section 2 to control this internal LO-toLO connection. If using the Model 8002A test fixture, refer to Figure 5B.
NOTEFor optimum accuracy, leakage currents in the test fixture can be cancelled by per-
forming REL on the current component of the measurement. To cancel leakage current, perform the procedure in “Amps Rel”.
5.Disable zero check by pressing Z-CHK and press OPER to source voltage to the test fixture DUT.
6.Take a reading from the display.
7.Place the V-Source in standby when finished by again pressing OPER.
WARNINGPlace the V -Source in standby before making or breaking connections to the
test fixture or DUT.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QU
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QU
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TY
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TY
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSO
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSO
1-14Front Panel Operation
Figure 5
Resistance measurements
A. Generic connections
B. Connections using 8002A
test fixture
Red
Measured
Resistance
R
X
Black
LO connected
to shield
Shield
8002A HIGH RESISTANCE TEST FIXTURE
7078-TRX-3 Triax Cable
Note: Set fixture mode switch
to picoammeter operation.
Warning: Connect of fixture
to safety earth ground using
safety ground wire (supplied
with 8002A test fixture).
6517
237-ALG-2
Cable
!
INPUT
250V PEAK
Note: V-Source low internally
connected to electrometer
low.
8002-ILC-3 Interlock Cable
!
250V PEAK
8607 Banana Plug Cables
LOHI
COMMON
PREAMP OUT
250V PEAK
COMMON
PREAMP OUT
250V PEAK
INPUT
V SOURCE
LOHI
V SOURCE
!
INTERLOCK
IN
TRIGG
LINK
OUT
Model 6517
Auto V -Source
Auto V-Source should be used to achieve optimum accuracy for ohms measurements. With
Auto V-Source selected, the V-Source is automatically set for either 40V or 400V. For the 2MΩ
through 200GΩ ranges, the V-Source is set to 40V. For the 2TΩ through 200TΩ ranges, a 400V
test voltage is used. With Manual V-Source selected, you can manually set the voltage level of
the V -Source.
Perform the following steps to select the V-Source mode for the ohms function:
1.Press CONFIG and then R to display the ohms configuration menu.
2.Use the and keys to place the cursor (blinking menu item) on VSOURCE and
press ENTER.
3.Place the cursor on AUTO (Auto V-Source ohms) or on MANUAL (Manual V-Source
ohms), and press ENTER.
4.Use the EXIT key to back out of the menu.
Amps Rel
the following steps:
NOTEThis procedure assumes that steps 1 through 4 of the Resistance Measurement proce-
Front Panel Operation1-15
Amps Rel is used to cancel leakage current in a test fixture for ohms measurements. Perform
dure have been performed.
1.With the V-Source in standby, remove the DUT from the test fixture.
2.Select the amps (I) function and disable zero check. Make sure REL is disabled.
3.Place the V-Source in operate by pressing OPER. The VOLTAGE SOURCE OPERATE
indicator is on when in operate.
4.Select the lowest possible measurement range to display the leakage current in the test
fixture.
5.Enable relative (REL annunciator on) by pressing REL. This zeroes the display, cancelling the leakage current reading.
6.Press OPER to place the V-Source in standby and enable zero check.
7.Perform the following steps to establish the amps Rel value for the ohms function.
A. Press CONFIG and then R to display the ohms configuration menu.
B. Use the and keys, place the cursor (blinking menu item) on AMPSREL and
press ENTER.
C. Place the cursor on YES and press ENTER to use the amps Rel value for the ohms
measurement.
D. Use the EXIT key to back out of the menu.
8.Re-install the DUT in the test fixture.
9.Select the ohms function (R) and proceed to step 5 of the Resistance Measurement procedure.
1-16Front Panel Operation
Resistivity measurements
NOTEThe following procedur e to measur e resistivity (surface or volume) r equir es the use of
the Keithley Model 8009 Resistivity Test Fixtur e. Refer to the Model 8009 Instruction
Manual for details on using the test fixture.
The following procedure requires “BENCH” reset conditions. See “Bench defaults” to reset
the instrument.
1.With zero check enabled (“ZeroCheck” displayed), select the ohms function by pressing
R.
2.Select Auto V-Source ohms. In this mode, the V-Source is automatically set to a voltage
(nominally 40V or 400V) that optimizes measurement accuracy. See “Auto V-Source”
located under “Resistance Measurements” to select Auto V-Source ohms.
NOTE“BENCH” reset conditions selects Manual V-Source ohms. In this mode, you must
manually set the V -Source le vel for the measurement. See “Voltage Source” for details
on using the V-Source.
WARNINGSelecting Auto V-Source ohms may set the V-Source to 400V.
3.Select a measurement range or use autoranging:
A. To automatically select the most sensitive range, enable autorange. The AUTO key
enables and disables autorange. When enabled, the AUTO annunciator is on.
B. For manual ranging, use the RANGE ▲ and ▼ keys to select a measurement range
consistent with the expected resistance.
4.Connect the Model 6517A to the Model 8009 test fixture as shown in Figure 6, and set
the switch on the test fixture to the desired measurement type (Surface or Volume).
5.Select the resistivity measurement mode as follows:
A. Press the CONFIG and then R to display the ohms configuration menu.
B. Use the and keys to place the cursor (blinking menu item) on MEAS-TYPE
and press ENTER.
C. Place the cursor on RESISTIVITY and press ENTER.
NOTEThe Model 6517A senses the switch position (surface or volume) on the Model 8009
through the interloc k cable, and automatically configur es the instrument for that measurement type.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
Front Panel Operation1-17
D. Cursor position (SURFACE or V OLUME menu item blinking) indicates the select-
ed measurement type:
•SURFACE — If Surf ace is the selected measurement type, you do not have to
proceed any further into the menu structure. Use the EXIT key to e xit from the
menu.
•VOLUME — If Volume is the selected measurement type, proceed as follows
to specify the thickness of the sample:
a.With the cursor on VOLUME, press ENTER to display the parameters menu
for volume.
b.Place the cursor on THICKNESS and press ENTER. The present thickness pa-
rameter for the sample will be displayed.
c.Use the cursor keys ( and ) and the RANGE ▲ and ▼ keys to specify
the thickness (in millimeters) of the sample, and then press ENTER.
d.Use the EXIT key to back out of the menu structure.
6.Disable zero check by pressing Z-CHK.
7.Press OPER to source voltage to the DUT and, after an appropriate electrification time,
take a reading from the display. Typically, an electrification time of 60 seconds is used.
8.Place the V-Source in standby when finished by again pressing OPER.
WARNINGPlace the V -Source in standby before making or breaking connections to the
test fixture or DUT.
Figure 6
Resistivity measurements
MAX INPUT
LID
INTERLOCK
1100V
METER SOURCE
TRIAX
250 MAX
HI-LO
8607 Banana Plug Cable
8607 Banana Plug Cable
7078-TRX-3 Triax Cable
PREAMP OUT
!
250V PEAK
INPUT
250V PEAK
!
Warning: Connect of fixture
to safety earth ground using
safety ground wire.
6517-ILC-3 Interlock Cable
LOHI
COMMON
V SOURCE
INTERLOCK
!
OUT
LINE RATING
90-134VAC180-250VAC
50, 60, 400HZ
55VA MAX
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
LINE FUSE
SLOWBLOW
1/2A, 250V
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,R
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,R
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,S
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,S
1-18Front Panel Operation
Charge measurements
Assuming “BENCH” reset conditions, the basic procedure is as follows:
1.With zero check enabled (“ZeroCheck” displayed), select the coulombs function by
pressing Q.
2.Select a measurement range or use autoranging:
A. To automatically select the most sensitive range, enable autorange. The AUTO key
enables and disables autorange. When enabled, the AUTO annunciator is on.
B. For manual ranging, use the RANGE ▲ and ▼ keys to select a measurement range
consistent with the expected charge reading.
3.Connect the test cable (see Figure 7) to the input of the Model 6517A and, with the input
open, disable zero check. If needed, press REL to zero the display.
4.Connect the instrument to the charge to be measured as shown in Figure 7.
CAUTIONDo not apply more than 250V peak (DC to 60Hz) or instrument damage
may occur.
5.Take a reading from the display.
NOTEThe reading can be r eset to zero by enabling zero check. If you wish to have the r ead-
ing reset to zero when a specified charge level is reached, enable Auto Discharge as
explained in the following paragraph.
Figure 7
Charge measurements
Measured
Charge
Shield (Optional)
Red (HI)
Q
s
237-ALG-2
Cable
Black (LO)
Input low connected
to shield
6517
!
INPUT
250V PEAK
COMMON
Auto discharge
When enabled, Auto Dischar ge is used to automatically reset the charge reading to zero when
the specified charge level is reached. After the inte grator resets, the charge measurement process
simply restarts at zero. Perform the following steps to enable or disable Auto Discharge.
1.Press the CONFIG and then Q to display the coulombs configuration menu.
2.Use the and keys to place the cursor (blinking menu item) on AUTO-DISCHARGE and press ENTER.
3.Perform step A to enable Auto Discharge or perform step B to disable it:
A. Enable Auto Dischar ge — Place the cursor on ON nad press ENTER to display the
B. Disable Auto Discharge — Place the cursor on OFF and press ENTER.
4.Use the EXIT key to exit from the menu.
Front Panel Operation1-19
present discharge level. Use the and keys, and the RANGE ▲ and ▼ keys
to set a different discharge level and press ENTER.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
1-20Front Panel Operation
External temperature and
relative humidity measurements
The Model 6517A can make external temperature measurements from -25°C to 150°C using
the Model 6517-TP type K thermocouple (which is a supplied accessory).
The Model 6517A can make relative humidity measurements (0 to 100%) using the Model
6517-RH humidity probe (which is available as an option).
Connections
The Models 6517-TP (thermocouple) and 6517-RH (humidity probe) connect to the rear pan-
el connectors labeled “TEMP TYPE K” and “HUMIDITY” as shown in Figure 8.
NOTEIt is recommended that the thermocouple junction of the Model 6517-TP be electri-
cally insulated from ground and voltage potentials. Erroneous reading will result if
the thermocouple comes in contact with ground or a voltage potential.
Figure 8
External temperature and
humidity
Model 6517-TP
Thermocouple
!
6517
INPUT
250V PEAK
PREAMP OUT
250V PEAK
COMMON
LO
V SOURCE
Extension Cable (P/N CA-129-1)
LINE RATING
90-134VAC
180-250VAC
50, 60, 400HZ
55VA MAX
Model 6517-RH
Humidity Probe
Measurement control
BENCH reset disables temperature and humidity measurements. These measurements can be
enabled from the front panel as follows:
1.Press MENU to display the main menu.
2.Use the and keys to place the cursor (blinking menu item) on GENERAL and
press ENTER to display the menu items of the general menu.
3.Place the cursor on A/D-CONTROLS and press ENTER to display the A/D controls
menu.
4.Place the cursor on DAT A-STAMP and press ENTER. The present control state for temperature and humidity will be displayed. OFF indicates that the measurement is disabled,
while ON indicates that the measurement is enabled.
5.To change the control state for temperature and/or humidity, place the cursor on the appropriate menu item and use the RANGE ▲ and ▼ keys to toggle the control state. W ith
the desired states for temperature and humidity displayed (ON or OFF), press ENTER.
6.Use the EXIT key to back out of the menu.
T emperature units
All temperature readings by the Model 6517A can be displayed in °C, °F or K. BENCH reset
selects °C measurement units. Perform the following steps to change units:
Front Panel Operation1-21
1.Press MENU to display the main menu.
2.Use the and keys to place the cursor (blinking menu item) on GENERAL and
press ENTER to display the menu items of the general menu.
3.Place the cursor on DISPLAY and press ENTER to display the SET READING DISPLAY menu.
4.Place the cursor on TEMP-UNITS and press ENTER. The blinking cursor indicates the
present temperature units selection.
5.T o change temperature units, place the cursor on the desired selection (°C, K or °F) and
press ENTER.
6.Use the EXIT key to back out of the menu.
Display reading options
T emperature and humidity readings are available as multiple (NEXT) displays. Using multiple displays, temperature and humidity readings can accompany each normal A/D measurement,
and they can be included as data storage elements for each reading stored in he buf fer . See “Multiple Displays” in Section 2 of this manual.
External temperature and humidity are selected as data store elements from the data store
configuration menu. See “Buffer (Data Store)” in Section 2.
1-22Front Panel Operation
Loading...
+ 80 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.