Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of
shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries,
diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will
be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs
will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of
any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or
problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE
BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHIZR KEITHLEY INSTRUMBNTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND
SOmWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF
SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL
AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
Warranty Information
ManualAddenda..
Safety Symbols and Terms
Specifications
Using this Instruction Manual
Unpacking and Inspection
Getting Started
Preparation for Use
Repacking for Shipment.
Accessories
...................
..
...........
.............
.......
..................
.....
........
.................
.............
.........
....................
SECTION 2-OPERATION
Introduction
Power Up Procedure
Power Up Self Test and Display Messages
Front Panel Familiarization
Controls
Display and Indicators
TiltBan
Front Panel Programs
IEEE-488 Address
Exponent Mode (Alpha or Numeric)
Calibration
Rear Panel Familiarization.
Connectors and Terminals.
V,RGUARDSwitch
LineFuse..
Basic Measurement Techniques.
Warm Up Period
Input Connections
Making Voltage Measurements.
Guarded Operation
Making Current Measurements.
Making Charge Measurements
Resistance Measurements
Using the Ohms Function As A Current Source
Low-Level Leakage Current Measurements
DiodeCharacterization
CapacitorLeakageMeasurements
CapacitanceMeasurement
Voltage Coefficients of High-Megohm Resistors
Model617RearPanel
Input Connector Configuration
Connections For Voltage Measurements
Meter Loading Considerations
Unguarded Circuit
GuardedCircuit
Guarded Input Connections
CurrentMeasurements
Voltage Burden Considerations.
Coulombs Connections
Resistance Measurement Connections
Voltage Source Connections
V/l Resistance Measurement Connections
Typical 2V Analog Output Connections
Typical Preamp Out Connections.
Electrometer Input Circuitry (AmpsMode)
Shielded Fixture Construction
“Transdiode” Logarithmic Current Configuration
Non-Decade Current Gains
Equivalent Input Impedance with Zero Check Enabled.
External Trigger Pulse Specifications
Meter Complete Pulse Specifications
Exlemal Triggering Example
Multiple Ground Points Create Ground Loop
Eliminating Ground Loops
Leakage Resistance Effects
Input Capacitance Effects.
Simplified Model of Source Resistance and Source Capacitance Effects
..........................................
..................................
...........................
.................................
...........................................
...............................................
.....................................
..........................................
..................................
.........................................
.............................
.....................................
..........................
...........................
................................
............... .........
....................................
...................
......................................
...............
..............................
..............................
.....................................
......................
......................................
.......................................
.......................................
^.
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
2-2
2-7
2-8
2-9
2-10
2-13
2-13
2-14
2-16
2-17
2-18
2-31
2-33
2-34
2-34
3-l IEEE Bus Configuration
3-2
3-3
3-4
3-5
3-6
3-7 617 Rear Panel IEEE Connector.
3-B Contact Assignments
3-9 General Data Format
3-10 SRQ Mask and Status Byte Format.
3-11 UO Status Word and Default Values.
3-12
3-13 U2 Status (Data Condition) Format
vi
IEEE Handshake Sequence.
Commands Groups
System Types
IEEE-488 Connector
IEEE-488 Connections
Typical Bus Times For Various Functions and Trigger Modes
......................................
......................................
............
.. ........... ....
..................................
.. .... .....
...
......... ... ..
........... ....
.......
.....
.........
..................
............. ......
..............
.......
.....
.....................
......
......
......
......
......
......
......
......
......
......
......
3-4
3-5
3-7
3-7
3-9
3-11
3-12
3-13
3-13
3-15
3-19
3-21
3-28
3-31
3-36
4-l
5-l
5-2
5-3
5-4
5-5
5-6
6-l
Diode Currents and Voltages.
Recommended Test Equipment for Performance Verification
Limits for A mps Verlflcatlon
Limits for Volts Verification
Limits for Ohms Verification (2kn-2OMa Ranges).
Limits for Ohms Verification (2COMn, 2Gn and 2OOGil Ranges)
Voltage Source Verification Limits
The following safety precautions should be observed before operating the Model 617
This instrument is intended for use by qualified personnel who recognize shock hazards and are familiar with the
safety precautions required to avoid possible injury. Read over the manual carefully before operating this instrument.
Exercise extreme caution when a shock hazard is present at the instrument’s input. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V rms or 42.4V peak
are present. A good safety practice is to expect that a hazardous voltage is present in any unknown circuit before
measuring.
Do not exceed 5oOV peak between input low and earth ground. Do not connect PREAh4P OUT, COM, OI
2V ANALOG OUTPUT to earth ground when floating input.
Inspect the test leads for possible wear, cracks or breaks before each use. If any defects are found, replace with test
leads that have the same measure of safety as those supplied with the instrument.
For optimum safety do not touch the test leads or the instrument while power is applied to the circuit under test.
Turn the power off and discharge all capacitors, before connecting or disconnecting the instrument.
Do not touch any object which could provide a current path to the common side of the circuit under test or power
line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface,
capable of withstanding the voltage being measured.
Do not exceed the instrument’s maximum allowable input as defined in the specifications and operation section.
Safe operation and good measurement practice dictates use of an external resistor when necessary to limit input
currents to less than 30mA.
SECTION 1
GENERAL INFORMATION
1.1 INTRODUCTION
The Keithley Model 617 Programmable Electrometer is a
highly sensitive instrument designed to measure voltage, current, charge, and resistance. Two forms of resistance
measurements are included in the standard configuration: a
constant current method, and a constant voltage method that
uses a built in voltage source for greater sensitivity. The
measuring range of the Model 617 is between 1OpV and 200V
for voltage measurements, O.lfA and 2OmA in the current
mode, O.ln and 200GO (up to 1OlQ using the built in voltage
source), and lOfC and 20°C in the coulombs mode. The very
high input impedance and extremely low input offset current
allow accurate measurement in situations where many other
instruments would have detrimental effects on the circuit being measured. A 4% digit display and standard IEEE-488 interface give the user easy access to instrument data.
1.2 FEATURES
Some important Model 617 features include:
l 4% Digit Display-An easy to read front panel LED display
includes a 4% digit mantissa plus a two-digit alpha or
numeric exponent.
l Autoranging-Included for all functions and ranges.
l Digital Calibration-The instrument may be digitally
calibrated from the front panel or over the IEEE-488 bus.
l Zero Correct-A front panel zero correct control allows the
user to cancel any offsets.
l Baseline Suppression-One button suppression of a
baseline reading is available from the front panel or over the
IEEE-488 bus.
l One-shot Triggering-A front panel control for triggering
one-shot readings from the front panel is included.
l Isolated IOOV Voltage Source-A built in 1OOV scwrce is
isolated from the electrometer section. The voltage source is
programmable in 50mV steps.
l Selectable Guarding-A selectable driven cable guard is in-
cluded to optimize speed.
l Standard IEEE-488 Interface-The interface allows full bus
programmable operation of the Model 617.
l Analog Outputs-Both preamp and 2V full range analog
outputs are included on the rear panel.
l 100~Point Data Store-An internal buffer that can store
up to 100 readings is accessible from either the front panel
or over the IEEE-488 bus.
l Minimum and maximum data points can be stored and are
accessible from the front panel or over the IEEE-488 bus.
1.3 WARRANG INFORMATION
Warranty information for your Model 617 may be found inside the front cover of this manual. Should you need to use
the warranty, contact your Keithley representative or the factory for information on obtaining warranty service. Keithley
Instruments, Inc. maintains service facilities in the United
States, West Germany, Great Britain, France, the
Netherlands, Switzerland, and Austria. Information concerning the operation, application, or service of your instrument
may be obtained from the applications engineer at any of
these locations.
1.4 MANUAL ADDENDA
Information concerning improvements or changes to the instrument which occur after the printing of this manual will be
found on a” addendum sheet included with this manual.
Please be sure that you read this information before attempting to operate or service your instrument.
1.5 SAFETY SYMBOLS AND TERMS
The following safety symbols and terns are used in this
manual and found on the instrument:
The A
should refer to the operating instructions in this manual for
further details.
The WARNING heading as used in this manual explains
dangers that might result in personal injury or death. Always
read the associated information very carefully before performing the indicated procedure.
The CAUTION heading used in this manual explains hazards
that could damage the instrument. Such damage may in-
validate the warranty.
symbol on the instrument indicates that the user
l-l
1.6 SPECIFICATIONS
The following items aw included with every Model 617 shipment:
Detailed Model 617 specifications may be found immedi-
ately preceding the table of contents of the manual. Note
that accuracy specifications assume that the insinxnent has
been properly zero corrected, as discussed in Section 2.
1.7 USING THIS INSTRUCTION MANUAL
This manual contains all the information necessary for you to
operate and service your Model 617 Programmable Elec-
trometer. The manual is divided into the following sections:
l Section 1 contains general information about your instru-
ment including that necessary to unpack the instrument and
get it operating as quickly as possible.
l Section 2 contains detailed operating information on how to
use the front panel controls and programs, make connec-
tions, and basic measuring techniques for each of the
available measuring functions.
l Information necessary to connect the Model 617 to the
IEEE-488 bus and program operating modes and functions
from a controller is contained in Section 3.
l Typical applications for the Model 617 are included ‘in Sec-
tion 4. At least one application for each of the measuring
functions is included in this section.
l Performance verification procedures for the instrument
may be found in Section 5. This information will be helpful
if you wish to verify that the instrument is operating in
compliance with its stated specifications.
l Section 6 contains a complete description of operating
theory for the Model 617. Analog, digital, power supply,
and IEEE-488 interface operation is included.
l Should your instrument ever require servicing, refer to the
information located in Section 7. This section contains in-
formation on fuse replacement, line voltage selection,
calibration. and troubleshooting.
l Replacement parts may be ordered by using the information
contained in Section 8. Parts lists as well as schematic
diagrams and component layouts are located in this section.
1.8 UNPACKING AND INSPECTION
The Model 617 Programmable Electrometer was carefully in-
spected before shipment. Upon receiving the instrument,
carefully unpack all items from the shipping carton and check
for any obvious signs of physical damage that might have oc-
curred during shipment. Report any damage to the shipping
agent at once. Retain the original packing material in case
reshipment becomes necessary.
Model 617 Programmable Electrometer
Model 617 Instruction Manual.
Model 6011 Triaxial Input Cable
Additional accessories as ordered
If an additional instruction manual is required, order the
manual package (Keithley Part Number 617-901-W). The
manual package includes an instruction manual and all perti-
nent addenda.
1.9 GElTING STARTED
The Model 617 Programmable Electrometer is a highly
sophisticated instrument with many capabilities. Although
there are a number of complex aspects about the instrument,
you can use the following procedure to get your instrument
up and running quickly. For more detailed information, you
should consult the appropriate section of the manual.
1.
Carefully unpack your instrument as described in
paragraph 1.8.
2.
Locate the power cord and plug it into the rear panel
power jack. Plug the other end of the line cord into an appropriate power source. See Section 2 for more complete
information.
3.
Connect the supplied triaxial cable to the rear panel input
jack. Make sure the rear panel V, R GUARD switch is in
the off position.
4.
Press in the front panel POWER switch to apply power to
the instrument. The instrument will power up the the
autorange volts mode with zero check enabled. Thus, you
could simply connect the red and black input leads to a
voltage source and take a voltage reading at this point by
disabling zero check. Remember that the Model 617
measures DC voltages up to 2COV.
5.
To change to a different measuring function, simply press
the desired function button. For example, to measure
resistance. simply press the OHMS button.
6.
Complete detailed operation concerning Model 617 front
panel operation may be found in Section 2. If you wish to
control these functions over the IEEE-488 bus, consult
Section 3.
1.10 PREPARATION FOR USE
Once the instrument is unpacked, it must be connected to an
appropriate power source as described below.
l-2
Line Power-The Model 617 is designed to operate from
105-125V or 210-250V power sources. A special power transformer may be installed for 90-1lOV and 195-235V ranges.
The factory set range is marked on the rear panel of the in-
strurnent. Note that the line plug is designed to mate with the
supplied 3-wire power cord.
CAUTION
Do not attempt to operate the instrument
on a supply voltage outside the indicated
range, or instrument damage might occur.
Line Voltage Selection-The operating voltage of the instrument is internally selectable. Refer to Section 7 for the procedure to change or verify the line voltage setting,
Line Frequency-The Model 617 may be operated from either
50 or 60Hz power sources.
IEEE-488 Primary Address-If the Model 617 is to be programmed over the IEEE-488 bus, it must be set to the arrect
primary address. The primary address is set to 27 at the factory, but it may be programmed from the front panel as
described in Section 3.
1.11 REPACKING FOR SHIPMENT
Before shipping, the instrument should be carefully packed in
its original packing material.
Model 6011 and 6011-10 Triaxial Cables-The Model 6011 is
made up of 3 feet of triaxial cable that is terminated with a
trim plug on one end and 3 alligator clips on the other end.
The Model 6011-10 is a similar cable 10 feet in length. Note
that the Model 6011 is supplied with the Model 617.
Model 6012 Triax to UHF Adapter-The Model 6012 allows
the Model 617 to be used with accessories having UHF type
co”nectors.
Model 6lOlA Shielded Test Lead-The Model 610lA is a
straight through probe and shielded lead equipped with 0.8m
(3O”) of shielded low noise cable terminated by a Tefloninsulated UHF connector. The Model 6012 must be used to
adapt the Model 6101A to the Model 617 triaxial input.
Model 6103C Voltage Divider Probe-The Model 6103C extends Model 617 voltage measurement range to 30kV. The
Model 6103C has a division ratio of 1OOO:l with a nominal
accuracy of 5%. The probe has an input resistance of 4.5 x
10110 and is equipped with a UHF male plug. The Model 6012
adapter must be used to connect the Model 6103C to the
Model 617.
Model 6104 Test Shield-The Model 6104 facilitates
resistance, voltage, or current measurements with either 2- or
3-terminal guarded connections at voltages up to 1200V. The
Model 6104 provides excellent electrostatic shielding and high
isolation resistance. Clips plug into banana jacks, allowing
custom connections. The Model 6104 has a BNC camector
on one side and binding posts on the other. The Model 6147
adapter (below) is required to connect the Model 6104 to the
Model 617.
If the instrument is to be returned to Keithley Instruments for
repair or calibration, include the following:
Write ATTENTION REPAIR DEPARTMENT on the shipping label.
Include the warranty status of the instrument.
Complete the service form at the back of this manual.
1.12 ACCESSORIES
The following accessories are available to enhance Model 617
capabilities.
Models 1019A and 1019s Rack Mounting Kits-The Model
1019A is a fiied or stationary rack mounting kit with two
front panels provided to enable either single or dual side-byside mounting of the Model 617 or other similar Keithley instrument. The Model 1019s is a similar rack mounting kit
with a sliding mount configuration.
Model 6105 Resistivity Chamber-The Model 6105 is a
guarded test fixture for measuring voltage and surface
resistivities. The unit assures good electrostatic shielding and
high insulation resistance. The complete system requires the
use of an external high-voltage supply such as the Model 247
as well as the Model 617. Volume resistivity up to 105Q/cm
and surface resistivity up to 1018Q can be measured in accordance with ASTM test procedures. Sheet samples 64 to
102mm (2Yz X 4”) in diameter and up to 6.4mm (IA”)
thickness can be accommodated. Excitation voltages up to
1OOOV may be used.
Model 6146 Triax Tee Adapter-The Model 6147 allows the
simultaneous connection of two triaxial cables to the single
triaxial input of the Model 617.
Model 6147 Triax to BNC Adapter-The Model 6147 allows
the Model 617 input to be connected to accessories having
BNC connectors.
Model 6171 and 6172 3 Lug-to-2 Lug Adapters-The Model
6171 is a 3 lug male-to-2 lug female triaxial adapter, while the
Model 6172 is a 2 lug male-to-3 lug female triaxial adapter.
1-3
Model 7008 IEEE-488 Cables-The Model 7008 cables are
designed to connect the Model 617 to the IEEE-466 bus and
are available in two similar versions. The Model 7008-3 is
0.9m (3 ft.) in length, while the Model 7008-6 is 1.&n (6 ft.)
long. Each cable is terminated with a standard IEEE-488 connector on each end, and each connector is equipped with two
metric SCTBWS.
Model 7024 Triaxial Cables-The Model 7024 cables are
similar units with male triaxial connectors on each end. The
Model 7024-l is 0.3m (1 ft:) in length, while the Models
7024-3 and 7024-10 are 0.9m (3 ft.) and 3.0m (10 ft.) long
respectively. These cables may be used to connect the Model
617 signal input to other equipment having similar triaxial
connections.
Model 7023 Female Triaxial Connector-The Model 7023 is a
chassis mount connector that mates with the Models 6011 and
7024 triaxial cables.
Model 8573 IEEE-488 Interface for the IBM PC-The Model
8573 allows the Model 617 to be connected to and controlled
from the IBM PC via the IEEE-488 bus.
1-4
SECTION 2
OPERATION
2.1 INTRODUCTION
Operation of the Model 617 may be divided into two general
categories: front panel operation and IEEE-488 bus operation.
This section contains information necessary to use the instrument on a front-panel basis. Note that many of these func-
tions can also be programmed over the IEEE-488 bus, as
described in Section 3.
The following paragraphs contain a complete description of
Model 617 front panel operation. First a complete description
of each front and rear panel function is presented. Next the
complete procedure for each of the measuring functions is
presented, followed by a description of the built in voltage
source. Finally, the analog output and guard functions are
described along with a method to apply external feedback.
2.2 POWER UP PROCEDURE
Use the procedure below to connect the Model 617 to line
power and power up the instrument.
1. Connect the female end of the power cord to the AC receptacle on the rear panel of the instrument. Connect the other
end of the cord to a grounded AC outlet.
switch. The switch will be at the inner most position when
the instrument is turned on.
3. The instrument will power up in the volts function, in the
autorange mode aqd with zero check enabled, as indicated
by the associated front panel LEDs. All other LEDs will be
off when the instrument is first turned on.
2.3 POWER UP SELF TEST AND DISPLAY
MESSAGES
The RAM memory is automatically tested as part of the
power up procedure. If a RAM memory error occurs, the “rr”
message will remain on the display. If the instrument was not
able to read the stored calibration constants and configuration, the decimal points in the two exponent digits will flash.
If such errors occur, the instrument may be partially or completely inoperative. Refer to Section 7 for more complete
details.
A power up self test may be run and the software revision
level may be displayed by pressing and holding the TRIG but-
ton when the unit is first turned on. During the test, all front
panel LEDs and the display segments will turn on as in the example below:
WARNING
The Model 617 is equipped with a 3-wire
power cord that contains a separate ground
wire and is designed to be used with
grounded outlets. When proper connec-
tions are made, instrument chassis is con-
nected to power line ground. Failure to use
a grounded outlet may result in personal in-
jury or death because of electric shock.
CAUTION
Be sure that the power line voltage agrees
with the indicated range on the rear panel
of the instrument. Failure to observe this
precaution may result in instrument
damage. If necessary, the line voltage may
be changed as decribed in Section 7.
2. Turn on the power by pressing in the front panel POWER
The instrument will then display the software revision level
when TRIG is released, for example:
E.4
The instrument will then enter the diagnostic mode, which is
used as an aid in troubleshooting problems within the instrument. See Section 7 for details. The power must be turned off
to remove the instrument from the diagnostic mode.
NOTE
If the instrument is still under warranty (less
than one year from the date of shipment), and
problems develop, it should be returned to
Keithley Instruments for repair. See paragraph
1.11 for details on returning the instrument.
2-I
J
Figure 2-1. Model 617 Front Panel
\
PROGRAM PROGRAM
SELECT EXIT SELECT EXIT
0 0
i, i,
2.4 FRONT PANEL FAMILIARIZATION
The front panel layout of the Model 617 is shown in Figure
2-1. The front panel may be divided into two sections: controls and display indicators. The following paragraphs
describe each of these items in detail.
2.4.1 Controls
All front panel controls except POWER are momentary con-
tact switches. Many control buttons include an annunciator
light to indicate the selected mode. Some buttons have a
secondary function that is entered by pressing the SHIFT button before pressing the desired button. All such buttons (except ADJUST) are marked in yellow. The controls are color
coded into functional groups for ease of operation.
POWER-The POWER switch controls AC power to the in-
strument. Depressing and releasing the switch once turns the
power on. Depressing and releasing the switch a second time
turns the power off. The correct positions for on or off are
marked on the front panel immediately above the POWER
switch.
SHIFT-The SHIFT button adds a secondary function to
some of the other front panel controls, including VOLTS,
TRIG, OHMS, RECALL and PROGRAM SELECT. Note
that the shift function is entered by pressing SHIFT before the
second button rather than pressing the two simultaneously.
ELECTROMETER-The ELECTROMETER buttons control
the measuring functions, selection of instrument ranges, and
such items as zero check, zero and suppression, and front
panel triggering.
VOLTS-The VOLTS button places the instrument in the DC
volts measuring mode. When VOLTS is pressed, the indicator
next to the button turns on, showing that the instrument is set
for that mode. Note that the Model 617 will be in this mode
when it is first turned on. Pressing SHIFT/VOLTS will place
the instrument in the external feedback mode, as described in
paragraph 2.12.
OHMS--Pressing OHMS places the unit in the resistance
measuring function. The indicator next to the OHMS button
will be illuminated when the instrument is in this mode. Note
that there are two ways to measure resistance with the Model
617. Pressing OHMS alone will cause the instrument to
measure resistance using the constant current method. Pressing the SHIFT button before pressing OHMS places the in-
2-2
strument in the V/I mode of resistance measurement, as
described in paragraph 2.8. The V/I indicator will light when
the instrument is in this mode.
COUL-The Model 617 may be set up to measure charge by
pressing the COUL button. The indicator next to the COUL
button will illuminate when the instrument is set for this
mode.
AMPS-Pressing AMPS switches the instrument to the
current-measuring function. The AMPS indicator will turn
on when the instrument is in this mode.
RANGE-These two buttons allow you to increment or
decrement the range the instrument is in. Pressing the up ar-
row button will move the instrument up one range each time
it is operated, while the down arrow button will move the instrument down range one increment each time it is pressed.
Note that pressing either of these buttons will cancel
autorange if that mode was previously selected. The display
mantissa will remain blank until the first reading is ready to
be displayed.
AUTO-The AUTO button places the instrument in the auto
range mode. While in this mode, the Model 617 will switch to
the best range to measure the applied signal. Note that the instrument will be in the autorange mode when it is first turned
on. Autoranging is available for all functions and ranges.
Autoranging may be cancelled either by pressing the AUTO
button or one of the two RANGE buttons.
ZERO CHECK-The zero check mode is used in conjunction
with the ZERO CORRECT control to cancel any offsets
within the instrument and is also used as a standby mode.
Pressing ZERO CHECK once will enable this mode, as shown
by the associated indicator light. When zero check is enabled,
the electrometer input circuit configuration changes (see
paragraph 2.11). No readings can be be taken with zero
check enabled. Pressing ZERO CHECK a second time will
disable this mode. Zero check should be enabled when
making connections or when changing functions.
ZERO CORRECT-The zero correct mode works with zero
check to cancel electrometer offsets. If zero check is enabled,
pressing ZERO CORRECT will store a new value that will be
used to cancel any offset. If the range is changed while zero
correct is enabled, the stored value will be scaled accordingly.
Zero correct may be cancelled by pressing the ZERO CORRECT button a second time. More information on using zero
correct may be found in paragraph 2.11.
may be disabled by pressing the SUPRESS button a second
time, and is cancelled if the function is changed.
TRIG-The TRIG button allows you to enter the one-shot
trigger mode and trigger single readings from the front panel.
To enter the one-shot mode, press SHIFT then TRIG. The
SGL indicator light will show that the instrument is in the
one-shot mode. Each time the TRIG button is pressed, a
single reading will be processed and displayed. The displayed
reading will flash when the TRIG button is pressed. The oneshot trigger mode can be cancelled by pressing SHIFT then
TRIG a second time. Additional information on triggering
may be found in paragraphs 2.13 and 3.10.14.
V-SOURCE-These buttons control the internal 2100V
source within the instrument. More information on the using
the voltage source is located in paragraph 2.8.
DISPLAY-The DISPLAY button toggles the front panel
display between the voltage source and the present display
mode (electrometer or data store). Pressing DISPLAY once
will switch the display from the present mode to the source
mode, as indicated by the LEDs adjacent to the display (more
information on the display is located in paragraph 2.4.2).
Pressing DISPLAY again will return the display to the
previous display mode.
ADJUST-These two buttons control the voltage source out-
put value. The up arrow button increases the voltage value in
50mV increments, while the down arrow decreases the
voltage source output in 5OmV increments. The values may
be scrolled by holding the desired ADJUST arrow in. The in-
strument will stop on the value currently displayed when the
button is released. The scrolling can be made more rapid by
pressing the SHIFT key before pressing the desire ADJUST
key. Note that the ADJUST keys are also used with certain
front panel programs, as described in paragraph 2.5. Note
that the maximum voltage values are +102.4V and
-102.35V.
OPERATE-The OPERATE button turns the actual voltage
source output on or off. Pressing the OPERATE button once
turns on the output. The LED next to the OPERATE button
will be illuminated when the source is turned on. Pressing the
OPERATE button a second time will turn off the output
(OO.oOV). Note that the OPERATE LED will flash when the
2mA current limit is exceeded.
SUPPRESS-The suppress mode allows you to cancel external offsets or store a baseline value to be subtracted from
subsequent readings. For example, if you applied 1OV to the
instrument and enabled suppress, that value would then be
subtracted from subsequent readings. Once suppress is enabled, the value is scaled when the range is changed. Suppress
DATA STORE-The two DATA STORE buttons control the
internal loo-reading data store mode of the instrument.
Through these two buttons, data storage may be enabled or
disabled, the storage rate may be selected, and readings may
be recalled to the front panel display. Paragraph 2.12 con-
tains a complete description of data store operation.
2-3
ON/OFF--This control enables or disables data store opera-
tion. In addition, reading rates can be selected by holding the
button in when first enabling data store. When data store is
enabled, the indicator light next to the ON/OFF button will
be on. Minimum and maximum values are stored and up-
dated as long as the ON/OFF LED is on.
RECALL/EXIT-This single button serves to recall readings
previously stored by data store. Pressing and holding this
button causes the instrument to scroll through the pointer ad-
dresses as indicated on the display. Once the desired reading
number is displayed, releasing the button causes the actual
reading to be displayed. To exit the recall mode, press SHIFT
EXIT.
PROGRAM-A single PROGRAM button controls such
modes as IEEE address, alpha or numeric display exponent,
and digital calibration. Paragraph 2.5 further describes front
panel programming.
SELECT/EXIT-This button enters the program mode to
allow access to parameters described above. Pressing
SELECT repeatedly causes the instrument to scroll through a
program menu. To cancel the program mode, press SHIFT
and then SELECT/EXIT in that order. Note that the program mode is cancelled by pressing SELECT/EXIT after a
program parameter change is made.
2.4.2 Display and Indicators
The operation of the 4% digit display and various indicators
is described below. The display updates at about three
readings per second.
STATUS Indicators-These three indicators apply to operation of the Model 617 over the IEEE-488 bus. The REMOTE
indicator shows when the instrument is in the IEEE-488
remote state, while the TALK and LISTEN indicators show
when the instrument is in the talk and listen states respect-
ively. See Section 3 for more information on using the Model
617 over the IEEE-486 bus.
2.4.3 Tilt Bail ’
The tilt bail, which is located on the bottom of the instru-
ment, allows the front panel to be elevated to a convenient
viewing height. To extend the bail, rotate it out 90” from the
bottom cover and latch it into place. To retract the bail, pull
out until it unlatches and rotate it against the bottom cover.
2.5 FRONT PANEL PROGRAMS
The Model 617 has three front panel programs that can be
used to set the primary address, set the display exponent
mode (alpha or numeric), or calibrate the instrument from the
front panel. To select a program, press PROGRAM SELECT
button repeatedly while bbserving the display. The instrument will scroll through the available programs with identifying messages. as shown in Table 2-2. When in the program
mode, the DISPLAY and DATA STORE RECALL buttons
are inoperative; the data store mode may be turned off, but
not on. The operation of the various programs is described in
the following paragraphs. To exit a program, press SHIFT
EXIT. If a change was made, pressing SELECT alone will exit
the program.
Display-The Model 617 has a display made up of a 4% digit
signed mantissa as well as a two-digit signed exponent. The
exponent can be represented either in scientific notation, or
with an alphanumeric subsript such as nA. The exponent dis-
play mode can be changed with a front panel program, as
described in paragraph 2.5. Note that, when scientific nota-
tion is used, the decimal point remains fixed as in 1.9999. The
range is indicated by the exponent. In addition, the display
has a number of front panel error messages, as shown in
Table 2-l.
Display Indicators-The METER, SOURCE, and DATA
LEDs indicate what the display is actually showing. When the
METER LED is on, the display represents an electrometer
reading. When the SOURCE LED is illuminated, the voltage
source value is being displayed. A data store reading is
displayed when the DATA LED is turned on. Normally, the
display will be the the meter mode, but the DISPLAY and
RECALL buttons will switch the display to the source and
data modes respectively.
2-4
2.5.1 IEEE-488 Address
Selection of the IEEE-488 address program is indicated by the
following message:
IEEE 27
Along with the message, the presently programmed IEEE-488
address (27 in this example) will be displayed. To select a new
address, use the V-SOURCE ADJUST keys. When the desired
value is shown in the display, press SHIFT then SELECT
EXIT to return to normal operation (or if a change was made,
simply press SELECT). For complete information on using the
Model 617 over the IEEE-488 bus, refer to Section 3.
Table 2-1. Display Error Messages
Message
OL
b Err
Description
Overrange input applied (- for negative valueI.
Bus Error: Instrument programmed while npt in remote; ~
or illegal command or command option sent.*
n Err
(Number Error: Calibration or voltage source value out of
; limits.*
‘t Err
Trigger Overrun Error: Instrument triggered while processing reading from previous trigger.
“See Section 3.
Table 2-2. Front Panel Program Messages
Displays/sets IEEE primary address.
IdlSP
Sets numeric or alpha exponent.
I
I
The display in the alpha mode appears as:
dISI’m
Once the desired exponent mode is selected, press SHIFT then
SELECT EXIT to return to normal operation (or simply PRO-
GRAM SELECT if a change was made).
2.5.3 Calibration
CAL
Allows calibration of instrument.
2.5.2 Exponent Mode (Alpha or Numeric)
The display exponent of the Model 617 can be operated in
either the alpha mode or the numeric mode. In the alpha
mode, the exponent is given in actual units such as mA. In the
numeric mode, the exponent is given in scientific notation.
Table 2-3 gives typical examples. including units.
To select the exponent program, scroll through the program
menu until the following message is displayed:
dISI’
Use either of the V-SOURCE ADJUST buttons to set the ex-
ponent to the desired mode. In the numeric mode, the display
might show:
dISP -3
An advanced feature of the Model 617 is its digital calibration
program. The instrument can be calibrated from the front
panel or over the IEEE-488 bus. To use the front panel
calibration program, refer to the calibration procedures outiined in Section 7.
2.6 REAR PANEL FAMILIARIZATION
The rear panel of the Model 617 is shown in Figure 2-2.
2.6.1 Connectors and Terminals
AC Receptacle-Power is applied through the suppiied power
cord to the AC receptacle. Note that the supply voltage is
marked adjacent to the receptacle.
IEEE-488 Connector-This connector is used to connect the
instrument to the IEEE-488 bus. IEEE-488 function codes are
marked above the connector.
2-5
Table 2-3. Typical Display Exponent Values
I Display
IPA
nC
PA
mV
kR
MQ
GQ
Tfl
ingineering
Units
PA
!lC
PA
mV
kCI
MO
GR
TQ
icientific
Uotation
10-77-A
10-w
lo-GA
lo-3v
103Q
lO@l
low
1012sl
f
N,
anocoulombr
rv
licroamperes
Millivolts
Kilohms
1
Megohms
Gigohms
Teraohms
PQ
INPUT-The INPUT connector is a 2-lug triax connector to
be used for all electrometer signal inputs. Note that you
should not confuse a triaxial connector with the BNC type
that is used for the EXTERNAL TRIGGER and ELEC-
TROMETER COMPLETE connections. Also, do not attempt
to force a 3-lug triaxial connector onto the INPUT Connector. The Models 6171 and 6172 adapters are available to
make the necessary conversion.
2V ANALOG OUTPUT-The 2V ANALOG OUTPUT prc-
vides a scaled O-2V output from the electrometer (2V output
for full range input). The output uses a standard S-way bind-
ing post and is inverting in the volts and ohms modes.
PR!ZAMP OUT-The PREAMI’ OUT provides a guard cutput for voltage and resistance measurements. This output can
also be used as an inverting output or with external feedback
when measuring current or charge. The PREAMP OUT has a
maximum output value of k3C0V and uses a standard s-way
binding post.
WARNING
Hazardous voltage may be present at the
PREAMP OUT, depending on the input
signal.
PO
101522
COM Terminal-The COM terminal is a 5way binding post
that provides a low connection for both the 2V ANALOG
OIJ’IT’LT and the PREAMP OUT This terminal is also used
for input low connection when in guarded mode; COM
is internally connected to input low through a lC0Q resistor.
Do not connect PREAMP OUT, COM, or 2V ANALOG
OUTPUT to earth when floating input.
V-SOURCE OUTPUT-The HI and LO outputs are the connections for the internal voltage source. This source can be
used as a stand-alone source or in conjunction with the elec-
trometer section to make resistance measurements as high as
1owL
EXTERNAL TRIGGER INPUT-This BNC connector can be
used to apply external trigger pulses to the Model 617 to trigger the instrument to take one or more readings, depending
on the selected trigger mode.
Petaohms
i
2-6
Figure 2-2. Model 617 Rear Panel
METER COMPLETE OUTPUT-This BNC connector prcvides an output pulse when the Model 617 has completed a
reading; it is useful for triggeling other instrumentation.
Chassis Ground-This jack is a s-way biding post that is
connected to instrument chassis ground. It is intended for use
in situations requiring an accessible chassis ground terminal.
A shorting link is supplied and connected to the CHASSIS
GROUND terminal.
2.6.2 V, !I GUARD Switch
The Model 617 has provisions for connecting a guard to the
inner shield of the input cable. Guarding is useful in the volts
and ohms modes to speed up response time and minimize the
effects of leakage resistance. Note that guarded operation is
not recommended in amps or coulombs. The V, Q GUARD
switch allows easy selection of the guarded mode of cperaticn. See paragraph 2.7.4 for more information on guarded
operation.
2.6.3 Line Fuse
The LINE FUSE, which is accessible on the rear panel, pro
vides protection for the AC power line output. For infcrmaticn on replacing this fuse, refer to Section 7.
2.7 BASIC MEASUREMENT TECHNIQUES
The paragraphs below describe the basic procedures for using
the Model 617 to make voltage, resistance, charge, and current measurements.
2.7.1 Warm Up Period
The Model 617 is usable immediately when it is first turned
on. However, the instrument must be allowed to warm up for
at least two hours to achieve rated accuracy.
NOTE
While rated accuracy is achieved after the two hour
wan-n up period, input bias current may require additional time to come to its optimum level. Allow
two hours for input bias current to settle to less
than 1OfA and eight hours to less than 5fA. It is
preferable in sensitive applications to leave the unit
on continuously.
2-7
2.7.2 Input Connections
The rear panel IivPIJT connector is a Teflon-insulated receptacle intended for all input signals to the Model 617. As
shown in Figure 2-3, the center terminal is high, the inner ring
or shield is low, and the outer shield is connected to instrument chassis ground. In the guarded mode, the inner shield is
driven at guard potential, while the Outer shield is chassis
ground.
NOTE
The input connector must be kept clean to main-
tain high input impedance.
The supplied Model 6011 input cable is designed to mate with
the input connector. The other end of the Model 6011 is terminated with three alligator clips. Input high is color coded in
red, input low is colored black, and chassis ground is color
coded in green. Keep in mind that these connections are for
the unguarded mode. In the guarded mode, red is high, black
is guard, and green is chassis ground. The COM binding post
provides a connection to input low through loOn for use in
the guarded mode.
GROUND
GROUND
mm
CAUTION
The maximum voltage between input high
and input low is ZXlV rms. DC to 60Hz sine
wsve (10 seconds maximum in mA ranges).
Exceeding this value may cause damage to
the instrument.
2.7.3 Making Voltage Measurements
The Model 617 can be used to measure voltages in the range
of *lOpV to +2COV. In principle, the instrument operates
much like an ordinary DMM, but its special characteristics
allow it to make measurements in cases where an ordinary
DMM would be unable to perform well. In particular, the
very high input resistance of 2oOTSl (2 X 1OW) allows it to
accurately measure voltage sources with high internal
resistances. In contrast, an ordinary DMh4 may have an input resistance of only loMa. resulting in inaccurate
measurements because of instrument loading.
Use the procedure below to make voltage measurements.
1. Turn on instrument power and allow it to warm up for two
hours to reach rated accuracy.
2. Check to see that the voltage function is selected by pressing the VOLTS button. Use the autorange mode, or select
the desired range with the ranging pushbuttons.
3. To achieve specified accuracy, especially on the lower
ranges, it is recommended that you zero the instrument. To
do so, first enable zero check and then press the ZERO
CORRECT button. Correcting zero in the lowest range of
any function will correct all ranges because of internal scaling.
A. UNGUARDED 6. GUARDED
I”, ii GUARD OFF)
Figure 2-3. Input Connector Configuration
NOTE
W, R GUARD ON,
It is recommended that zero check be enabled
when connecting or disconnecting input signals.
WARNING
The maximum common-mode input
voltage (the voltage between input low and
chassis ground1 is 5OOV peak. Exceeding
this value may create a shock hazard.
CAUTION
Connecting PREAMP OUT, COM, or 2V ANALOG
C
XJTPUT to earth while floating input may
damage the instrument.
2-8
NOTE
The input circuit configuration changes with
zero check enabled. See paragraph 2.11.1.
4. Connect the Model 6011 triaxial input cable or other
similar cable to the rear panel INPUT jack on the instrument. For sources with high output resistance, the cable
should be kept as short as possible to minimize cable
capacitance.
5. If response time and leakage resistance are considerations,
place the instrument in the guarded mode as described in
the next paragraph.
6. Connect the other end of the cable to the voltage to be
measured, as shown in Figure 2-4. Disable zero check.
7. The reading may be obtained directly from the display.
The exponent can be placed either in the alpha or numeric
mode, as described in paragraph 2.5.
I+ I
vs= j
T- I
Figure 2-4. Connections for Voltage Measurements
TO A/D CONVERTER
>
Voltage Measurement Considerations: Two primary considerations come to mind when making voltage
measurements. especially for voltage sources with high output resistances. For one thing, the loading effects of the
measuring instrument come into play at the high resistance
levels involved. Secondly, the distributed capacitance of the
source, the input cable, and the input circuit of the instrument
itself come into play when making these measurements.
To see how meter loading can affect accuracy, refer to Figure
2-5. In this figure there is a voltage source with a value ES and
an output RS connected to the input of the electrometer,
which has its input resistance represented by RIN. The percent
error due to loading can be calculated as follows:
100 RS
% ERROR = -
Rs + RIN
Thus, to keep the error under 0.1%. the input resistance must
be about 1000 times the value of the source resistance. R.
At very high resistance levels, the very large time contants
created by even a minimal amount of capacitance can slow
down response time considerably. For example, measuring a
wurce with an internal resistance of 1OOGQ would result in an
RC time constant of one second when measured through a
cable with a nominal capacitance of 1OpF. If 1% accuracy is
required, a single measurement would require at least five
seconds.
Basically, there are two ways to minimize this problem: (1)
keep the input cable as short as possible, and (2) use guarding.
With the first method, there is a limit as to how short the
cable can be. Using guarding can reduce these effects by up to
a factor of 1000. The Model 617 has a rear panel switch to
allow guarding to be easily applied to the input circuit: see the
next paragraph for details.
2-9
At low signal levels, noise may affect accuracy. Shielding of
the unknown voltage can reduce noise effects substantially.
When using shielding, the shield should be connected to input
IOW.
rI
/
To approach the concept of guarding, let us first review the
unguarded circuit shown in Figure 2-6. The measured signal is
represented by the voltage source ES and the source resistance
RS. Cable leakage impedance is represented by 2~ The
source resistance and leakage impedance form a voltage
divider that attenuates the source voltage as follows:
ZLES
E, =
ZL + Rs
Thus, to keep the error due to leakage resistance under O.l%,
the leakage resistance must be at least 1000 times the source
resistance value.
T ; / N”!
t-
---- J
G-lJ
L-----J
J
Figure 2-5. Meter Loading Considerations
2.7.4 Guarded Operation
Guarding consists of using a conductor supplied by a low impedance source to totally surround the leads carrying a highimpedance signal. The output of this low-impedance source is
kept at the same potential as the signal itself, resulting in
drastically reduced leakage currents.
Guarding the circuit miminizes these effects by driving the
shield at signal potential, as shown in Figure 2-7. Here, a uni-
ty gain amplifier with a high input impedance and low output
impedance is used. The input of the amplifier is connected to
the signal, while the output is used to drive the shield. Since
the amplifier has unity gain, the potential across ZLis essentially zero, so no leakage current flows. Leakage between
the cable shield and ground may be considerable, but it is of
no consequence since that current is supplied by the low impedance source, rather than by the signal itself.
2-10
Figure 2-6. Unguarded Circuit
1
Figure 2-7. Guarded Circuit
-I
low input offset current. The low voltage burden is achieved
because the Model 617 measures current as a feedback type
picoammeter, rather than the shunt method used by many
DMMs.
NOTE
After measuring high voltage in volts, or following an overload condition in ohms, it may take a
number of miriutes for input current to drop to
within specified limits. Input current can be
verified by placing the protection cap on the INPUT jack and then connecting a jumper between
the COM and chassis ground terminals, With
the instrument on the 2pA range and zero check
disabled, allow the reading to settle until the instrument is within specifications.
NOTE
Safe operation and good measurement practice
dictates the use of an external resistor when
necessary to limit currents to less than 3011~4.
When the rear panel V, Q GUARD switch is placed in the ON
position, guard potential is placed on the inner shield of the
triaxial cable. The other shield remains at chassis ground.
Thus, it is necessary to use the COM terminal for low signal
connections, as shown in Figure 2-0. For very critical measurements, a shielded, guarded enclosure should be used.
WARNING
Hazardous voltage lup to 3OOV) may be present on the inner shield when V, Q GUARD
is on, depending on the input signal. A safety shield, connected to chassis ground is
recommended when making voltage
measurements over 30V or guarded
resistance measurements.
NOTE
The use of guarding is not recommended in
amps or coulombs.
The PREAMI’ OLJT terminal may be used for guarding in the
volts and ohms modes in a similar manner. In this mode, the
preamplifier acts as a unity gain amplifier with low output
impedance.
WARNING
Hazardous voltage (up to 3WV) may be
present at the PREAMP OUT terminal,
depending on the input signal.
2.7.5 Making Current Measurements
The Model 617 can resolve currents as low as 0.1 fA
(lo--lbA), and measure as high as 2011~4 in 11 ranges. The
Model 617 exhibits low input voltage burden and extremely
To measure current with the Model 617, use the following
procedure.
I. Turn on the power and allow the instrument to warm up
for at least two hours to obtain rated accuracy.
2. Select the current mode by pressing the AMPS button on
the front panel. Set V, Q GUARD to OFF.
3. To achieve rated accuracy, select the 2pA range, zero the
instrument by enabling zero check and then pressing the
ZERO CORRECT button. Select the desired range, or use
autoranging if desired.
4. Connect the Model 6011 or other similar cable to the rear
panel INPUT jack. Connect the other end of the circuit to
be measured as shown in Figure 2-9. Shielding will be required for low-level measurements. Connect the shield to
input low.
5. Disable zero check.
6. Read the current value directly from the display. The exponent may be placed either in the alpha or numeric modes,
as described in paragraph 2.5.
Current Measurement Considerations: At very low levels (in
the picoampere range), noise currents generated in the cable
or from other sources can affect measurements. Currents
generated by triboelectric effects are a primary cause of noise
currents generated in connecting cables. These currents are
generated by charges created at the junction between a conductor and an insulator because of friction. Coaxial and triaxial cables are especially prone to such noise currents, which
are generated by cable flexing. To minimize these effects, the
cable should be tied down firmly to minimize any flexing.
Also, special low-noise cable, constructed with graphite be-
tween the shield and insulator, is available to minimize these
effects. However, even with low-noise cables, several tens of
femtoamps of noise currents can be generated by cable move-
ment.
2-11
Loading...
+ 153 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.