OPERATION
HODEL 6OOS ELECTROMETER
ter zero pasit.ion.
1f not. adjust the meter zero
with the front panel ZERO Control. Normally, there
is no need co “se the COARSE ZERO Switch.
Range Switch to lo-’ AMPERES range.
Set the full-
scale voltage range with the ~“lci~lier Switch. Operaeing procedures are the same as subparagraph b
above.
d. After a
few moments increase
the voltage sensitivity by advancing the Multiplier Switch to .3, .1,
etc.
Continue zeroing with the ZERO Control.
e. After long periods of starage or after an over-
load, the Model 6008 nay drift excessively. The in-
put cransistars are insensitive to mechantcal shock;
however. B severe inpa overload may cause a zero
offset. This is corrected with the Zero Controls.
Drifting, though. can occur foe several hours.
f.
Although the offset current of the Electrom-
eter is much below that found in conventianal valt-
meters. it can be observed on the meter.
The c”rrent
charges ehe input caoacitance. and the Electrometer
appears to drift uhen the input is open.
Use the
ZERO CHECK S”tr~n co discharge the charge build-up.
Depressing the ZERO CHECK Button removes all signal
from the amolifier.
2-3.
“OLTXE \E‘\SUREEENTS.
a. The liodel 6~3011’s high input impedance allaws
circuit measurements wi’chaut causing circuit loading.
FOC law resistance in-circuit tests, the input resis-
tance can be lowerr<, 1,, -v,..:,.: ,,:, ‘k-t&p ~,i~c,b:,e,,is.
Make all voltage meas”rement~ with the FEED-
BACK Switch & in the NORMAL posieion.
b. High Impedance Eleasurements ( LO14 ohms, ‘0
picofarads), Follow the instr”ctions of paragraph
2-2.
Set the concrois as foliows:
ZERO CHECK Sutton
LOCK
Range Switch VOLTS
multiplier Switch
LO
FEEDBACK Switch
NORMAL
METER Switch
CENTER ZERO
connece the ““know” source co the INPUT Recepeacle
and unlock the ZERO CHECK Buteon. Set the METER
Switch to + or -, as necessary. Increase the sensi-
tivity with the Multiplier Switch. Recheck the zero
setting after increasing the sensitivity.
Low Impedance Measurements. TO decrease the
in;& resistance frnm
1OL4
nh”w +Pe *II,- nnn.y CWifCh
to one of the AMPERES ranges. The input resistance
is now the reciprocal of the current range. For instance, to obtain an inpur resistance of 10’ ohms -
which is normal for convantional “T”Ms - set the
d. To measure ~~“rces more than 10 volts, “se the
Model 6102A 1O:l Divider Probe or the 6103.4 1OOO:l
Divider Probe.
The Model 6102A extends the Model
6OOB’s
range to 100 volts; averall accuracy is *&%
and Input resistance is 1010 ohms.
The Node1 6103
extends the Model 6008’s range to 10 kilovolts’ overall accuracy is r6% and input resistance is 10
12
ohms,
Follow the same Operating procedures with the dividers as in subparagraph b. The riall-scale voltage
range is the divider ratio times the MultipLier Switch
setting.
2-4.
CURRENT MEASUREMENTS.
a. The Model 6008 can measure currents three ways.
1. In the normal method - used on any range -
the current is determined by measuring the voltage
drop across a resistor shunring che’amplifier in-
put. This method is useful when lower noise is more
important than faster response speeds or if some
damping is needed.
2. In the fasr method - far “se anly below the
10-Z ampere range - the shunt resistor is between.
the amplifier ““ep”t and input in the feedback loop.
This circuit largely neutralizes the effect af inpue capaciey and greatly increases the response
speed. Also, the inpue voltage drop is reduced to
less than One millivolt on any range.
3. For galvanometric current measurements, the
Model 6006 acts as B null indicator between a very
cccurate current source and the “nknawn current
SO”CCe.
b. Rise time varies primarily with the current
range, the input capacity and the method used. With
ehe FEEDBACK Switch in ehe FAST oosition. the rise
time on ehe ?ast sensitive range is less than 2.5
seconds and on the lo-6 ampere range is less than 3
milliseconds. Given a choice. it is beteer ta place
the Electrometer near to the current source than to
the data reading instrument. Transmitting the input
signal through long caaxial cables slightly decreases
the respanse speed and significantly increases noise
due to the cable capacitance.
c. Normal Methad (0.3 to lo-l3 ampere ranges)
1. hollow rhe instructions
of
paragraph 2-2
Set the controls as follows:
ZERO CHECK Sutton
Range Switch
Multiplier Switch
FEEDBACK Switch
METER Switch
LOCK
10-l AMPERES
1
NORMAL
CENTER ZERO.
Connece the unknown source to ehe INPUT Receptacle
and unlock the ZERO CHECK Sutton. Set the METER
Switch to + or -, as necessary. Increase the sen-
4
0872B