Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year
from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable
batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio.
You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service
facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for
the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or
misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from
battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS
INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE
OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY
PERSON, OR DAMAGE TO PROPERTY.
The print history shown belowlists the printing dates of %I1 Revisions and Addenda created for this manual. The Revision
Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revi-
sions, contain important change information that the user should incorporate immediately into the manual. Addenda are nun- ~~
bered sequentially. When a new Revision is Mzated, all Addenda associated with the previous Revision of the manual ax
incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 590-901-01) ........................... I ...............Y ... .I......1 ....................... _. ................... .1987
Revision B (Document Number 590-901-01) .................................... .I.................................................~.~ ...............
Revision C (Doc~umcnt Number 590-90 I-01). ............................................ ..~.........................................~..~....-..~ ..
Addendum C (Document Numbei 590-901-02) ......................................................................... ..~ ..- October 1995
Addendum C (Documenr Number 590-901-03) .................. ..~..~...~ ... I ..I._ ............. ..~ ....................... F&my 1996
Addendum C (Document Number 590.901-04 ................. ..~
Revision D (Documenr Number 590-90 I-01) .................... I ......_.....I. . I..............................- ........... February 1999
The following safety precautions should be observed before using
this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions
may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation,
operation, and maintenance information carefully before using the
product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with
hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating properly, for example, setting the line voltage
or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator
may perform them. Otherwise, they should be performed only by
service personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that
are rated Installation Category I and Installation Category II, as described in the International Electrotechnical Commission (IEC)
Standard IEC 60664. Most measurement, control, and data I/O signals are Installation Category I and must not be directly connected
to mains voltage or to voltage sources with high transient over-voltages. Installation Category II connections require protection for
high transient over-voltages often associated with local AC mains
connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise
marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal
voltage may be present on cable connector jacks or test fixtures. The
American National Standards Institute (ANSI) states that a shock
hazard exists when voltage levels greater than 30V RMS, 42.4V
peak, or 60VDC are present. A good safety practice is to expect
that hazardous voltage is present in any unknown circuit before
measuring.
Operators of this product must be protected from electric shock at
all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In
some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to
protect themselves from the risk of electric shock. If the circuit is
capable of operating at or above 1000 volts, no conductive part of
the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits.
They are intended to be used with impedance limited sources.
NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting
cables, test leads, and jumpers for possible wear, cracks, or breaks
before each use.
When installing equipment where access to the main power cord is
restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any
other instruments while power is applied to the circuit under test.
ALWAYS remove power from the entire test system and discharge
any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal
changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always
make measurements with dry hands while standing on a dry, insulated
surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its
specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or
switching card.
When fuses are used in a product, replace with same type and rating
for continued protection against fire hazard.
Chassis connections must only be used as shield connections for
measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a
lid interlock.
5/02
If or is present, connect it to safety earth ground using the
wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal
and common mode voltages. Use standard safety precautions to
avoid personal contact with these voltages.
The WARNING heading in a manual explains dangers that might
result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could
damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and
all test cables.
To maintain protection from electric shock and fire, replacement
components in mains circuits, including the power transformer, test
leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals,
may be used if the rating and type are the same. Other components
that are not safety related may be purchased from other suppliers as
long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments
to maintain accuracy and functionality of the product.) If you are
unsure about the applicability of a replacement component, call a
Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based
cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill
on the instrument. Products that consist of a circuit board with no
case or chassis (e.g., data acquisition board for installation into a
computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper
cleaning/servicing.
SAFETY PRECAUTIONS
The following safety precautions should be observed before using the Model 590,
This inshment is intended for use by qualified personnel who reco&iie shock hazards and are familiar
with the safety precautions required to avoid possible injury. Read over this manual carefully before operating
the instrument.
Exercise extreme caution when a shock hazard is present at the instrument’s test output. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 3OV RMS
or 424V peak are present. A good safety practice is to expect that a hazardous voltage is present in any
unknown circuit before measurement.
Do not exceed 3OV RMS (42.4V peak) between analog common and earth ground.
Inspct YOUI connecting cables for possible wear, sacks, or breaks before each use.
For maximum safety, do not touch the test leads or the ins&u.ment while power is applied to the circuit tider
test. Turn the power off and discharge all capacitors before connecting or disconnecting the instrument.
Do not touch any object which could provide a current path to the common side of the circuit under test
or power line (earth) ground. Always make ~easureme~nts with&y hands while standing on a dry, insulated
surface capable of withstanding the voltage being measured.
Do not exceed the instrument’s
section of this manual.
xnaximum allowable bias input as defined in the specifications and operation
SPECIFICATIONS
59011OOk and 5901100kllM FRONT PANEL SPECIFICATIONS
w/z Dii,
RANGE
2 PF
2&s OSIS O.lz% + I 50 x UC, + 2wl 4ns
2°F X0 fF 0.12% +~ (260 x GIG, + 5) SC‘7 fF
2ms loo ns 0.12% + ( 22 Y c/G, + 51 0.6s
590&i and 5901100WlM FRONT PANEL SPECIFICATIONS
Wh Diil
RANGE BESOLUTION
20 PF
ZCQ PS
Yr%
2°F
2oms
Aamy is
1 fF 0.29% + (SO x G/G,, + 10) 6fF
10 “s 0.29% + (I20 x UC, + 10)
10 fF 0.29% t Y x GIGS + 5)
x-m “s 0.29% * I20 x UC, + 5) %f:
I‘Q fF 0.29% + (SOI x GIG, + 51
1fi
maximum limit t-m Qz20; typical for Q<ZO.
ACCURACY u Y.zu)'
u)vs'C F-P NOISE5
iflads + counts)
0.25% + (260 x GIG, + 5js 9 fF
0.25% + ( 22 x UC, +-m-5)
*ccuRAcY 11 Years
lsswc F-F NOISE’
iwdg + umntsj
0.29% + (I20 x CIC,s +~ 5)
RLFRON
6,fi
FInm ON
4O”S
2WfF~~
1 PS
SHUNT CAPACZTANCT
LOADING EFFECt”
iwd8i-k)
0.1 % + (3 x G/G=)
0.1 % + (3 x UC”)
0.1 % + (3 x GIG,)
0.1 % + (3 x UC,)
0.02% + (20 x GlG,s)
Km6 + ( 7 x CGS) 0.1’ % + (7 x CKn)
0.02% + (20 x GIG,)
0.02% + ( 7 x UC,,) 0.02% * (3 x UC,,)
0.1 % + (30 x GIG=) 0.02% + (2 x GIG”)
0.1 % + (10 x CICrs) 0.02% + (2 x UC,)
TEMPERATURE COEFFxaENr wuNT cAP*aTANcE
W-WC & 28%50~c
*&rag + counb,I~c
0.02% + (20 x G,G,s) 0.5 % + (W x GIGm)
0.02% + ( 8 x cic,s) 0.5 % + (lo x CGr)
0.02% + (20 x GIG,s) 0.35% + (40 Y. GIG,s)
0.02% + ( 8 x CICPr) 0.35% + (16 x C/Cd
0.1 % + (3 x GIG,,.)
0.02% + (2 x G/G,)
LOADING EFFECT”
l (%rdg + counts1
0.35% +~
0.33% +
40 x G,G,)
16 x C/CR)
I
NOTES:
1. G = conductance reading; C = capacitance reading; Gm = fuU scale
cmductance; C, = f”U scale capacitme. Range and acc”racy designations based on parallel RC model.
2. Rant panel 2Jzarq is IeIative to C.&ration same
pamI accuracy and source accuracy for total accuracy. Factcry calibration
source accumcy is 0.06% for 1COkHz and 0.08% for lMHz. CAL is to be
used to cancei initial zero, gai”, and phase error terms withi” 8 hours of
measurement or whenever ambient temperature changes by more than
2oc.
accuracy. Add front
BIAS SOURCE
INTERNAL BIAS SOURCE OUTPUTz -20.00X’ to +2O.OOOV in 5mV
Step.
ACCURACY (1 Year, 18%8°0: i(O.ffi% setting + 10~1”) exdtive of
*(o.ws% + hv)IT.
MAXIMUM ODTPllT CuBREm f5OmA.
SETtTING TIME: <lms to 1% of f&l value.
NOISE: Typically <2OOpV F-p, O.lHz-IMHz; 3mV F-p to 75MHz
BIAS WAVEFORM:
~~.-STAIR:output~sinincrrmMtsofBIAsS?zFVfromFIRsT
BIAS V to LAST BIAS V, the” back to FIRST BIAS V.
PULSE Outputs pulse train; amplitude increments by BIAS STEP V
from FIRST BIAS V to LAST BIAS V (each pulse is from DKFAUIT
BIAS V to FIRST BIAS V for duration of SW TEwfE, then back to
DEFAULT BIAS V). Also Fqwnmable for single pulse.
EXTz AUows application of external bias sauce (via VOLTAGE BIAS
mw.
BIAS PARAMETERS: FIRST BIAS V, LAST BIAS V, DEFAULT BIAS V,
BIAS STEP V, START TIME, STOP TIME, STEP TIME, COIJNT.~
ts the programmed value.
4. “Shunt Capacitance Loading’ is additional accwxy with equal shunt
load cm Output and Input, per IOOpF shunt load.
5. Noise spedfied with SOOpF shunt loading on Output and Input. Noise
on 2pF and 2OpF ranges is tvpical with 1oOpF shunt load; 5oOpF wiU increase noise no more than x2. Measured at 10 rdS/s rate.
CAPAClTANCE NON-LlNEAlUTYt <OS% of range, for Q>20 or
D<O.OS, 18”-28°C.
TEST VOLTAGE: lSmV nns ilO%.
TEST FRBQUBNCY: 89011Wkz 1CQkHz. 54ollMz IMHz. Tolerance: +O.l%.
BIAS STEP V: Rogrammable in 5mV steps to 20V. Polarity selectable +
or -.
ST-T ~~ After transition from DEFALILT~ Be V to FIRST BIAS V,
START TIME must elapse before first meas-ent. Programmable i”
inaements of 1024ps from 1 to 65,536 increments.
Accuracy: *(O.l% + hs).
STEP TIME: The period between the ktitio” of BIAS STEP V and the
Start of the next meas”rwne”t. Programma
~frc!Qll to 65,536 fnaenlenis. Accuracy: ~f(O.l% +-lms).
STOP TIME: The period between the end of the final measurement and
the kawition from LAST BIAS V to DEFAULT BIAS V. Programmable
in increments of 1024us from 1 to 65,536 i”aeme”ts.
application of &emal bias source up to f2MIV, i5Gm.A.
Input Impedance: lwk0 paralkled by l$.
VOLTAGE BIAS MONITOR: Rear panel output terminals allow monitor
of the DC BIAS SOURCE output or extemauy applied VOLTAGE BIAS
INPUT; Level: IV = 1V wt. Output Resistance: Ikn.
VOLTAGE BIAS DISPLAY: Front panel &-digit display allows direct
readback of the DC BIAS SOURCE output or ater”aUy applied VOL
TAGE BJASINPLIT. Acnnacy *(O.O5% + 5 aunts). Temperature Caef-
ficient: *(0.005% +~ 0.1 count)/T.
ble in increments of 1024q
ANALYSIS CAPABILITY
(Rogrammiog and output an&able from front panel or IEEE-488 bus)
READING BUFFERS A and B: Two data buffers allow storage and math-
ematical manipulation on up to 450 mea.sluemerlt triplets: capacttance,
conductance, and voltage. In C vs. t, capxitmce and index only are
stored (up to 1350 points).
1I;pefforms the inverse of C on the capacitance data stored in reading
to a user-pmgammable reference value Co.
Cm: Searches reading buffer far the maaimum capacitance value.
G-6: Sequentiauy computes the difference between comes ondin
capacitance readings stored in reading buffer A and reading B # uffer
Iv,-V.1: Calculates the corresponding difference in applied voltage for
values of capatitance in reading buffer B equal to each valve in reading
bufferA.
C vs. t: Akws fast measurement of capacitance vs. time (loo0 rdgls).
CABLE COMPENSATION
Rip to 8 setups can be stored in non-volatile memory)
CALIBRM’ION CAPAQTOR COMPENSATION: Corrects for emxs due
tocablesorswitchingmatrixupto5metarseffe
length. nvmlleeanwmen ts are made with cables and matrix terminated
with pmision reference capacitors in place of the DUT. Model 5907
CablelMath Calibration Cqa.itor Set required. Bus programmable only.
MHz only. Accuracy: *(05% + appltcable front panel specification),
typical.
SINGLE-ENDED CABLE and S-PARAMETE R COMPENSATION can
also be made. See manual for detailed information.
r&m, Display, Status, Service Request, Self Test, Output Format.
TRANSLATOR: Up to 250 bytes of deSni!dore allow v&able pass@, def-
inttin decomposition and listing.
Bias Waveform, Bias Parameters, Plottin
Plotter Parameters
i!klibmiion, cable COG
GENERAL
DISPLAY: Three 4~-di@t displays for capadtance, conductance, and
voltage bias.
~~I~b~4md or autoranging (for rates up to 18 Id&; 10% wermnge
OVERRANGi INDICATION: Display reads OFLQ.
AVAILABLE MEASUREMENT RATES (to tntemal buff-):
4=&D&@ 1, 10, and 18 rdgls. 3yz-D&k 75 and loo0 rdgis.
FILTERc l-pole analog; pole at 37Hz. Filters t&b capacitance and conduc-
tame signals. For FlLTER off, multiply p-p noise specification by 5.
CAL: Initiates ti&ation to intm-d reference capacttor. Used to cancel
initial zero, gain, and phase errors.
ZRRO: Allows zeroing of on range readings. Allows relative readings to
be made with respect to a baseline value.
MAXTMUM OVERLOAD: OLn’PUT, Voltage Bias Input: 20X’ internal-
ly fused at ‘@. INFUT: Clamped by diodes to f0.N. M&mum current
2wmA. Analog output% w.
MAXIMUM COMMON MODE VOLTAGE (INPWI and OUTPUT,
Volt~;~ In nt):
ANALOG 0LTFli-B (Capacitance and Conductance): Level: 2V output
at full range. Initial offset: i25mv. wt Reststance: lw. Response
Time: lms to 1% of final value with fiiter off; 25ms maximum with
filter on.
FTOlTBR: Digital plotter output centrals IU’747OA plotter or equivalent
using HPGL via IEEE-488 for real-time plotting of all measure*e*ts as
well as results of math computations, with gdas and labels. Talks to
plotter on address 05. HPGL commands used are lN, Ip, IW, PA, PD,
Pu, SC, SI, SF.
PRONT PANEL SFFWS: Up to 7 fmnt panel setups can be stored in non-
voL@ile memory.
EXTERNAL TRIGGER: ‘lTL compatible Fxtenul Tr@ger Input and
output.
INPUT CONNECTORS: Isolated BNC f& m and Voltage Bias Input.
OUTPUT CONNECTORS: Isolated BNCs for OUTPUT, Voltige Bias Mont-
tar, and Analog Outputs. Non-isokd BNCs/or External Mgger.
sing up to 35T. storage: -250 to +wc.
WARMUP: 1 hour to rates3 accu.racy.
COOLING: Internal fan and filter for foxed air cooling.
POW!% lOEl25V or ZlC-w1V (external switch selected), 5OHz to 6oHz,
loOVA madmum. 9&llOV and ISXZOV version available upon request.
DIMENSIONS, WEIGHT: l33nvn high x 435mm wide x 44&m deep
(5’/i in. x 17% in. x 17% in.). Net weight 9.Ikg (20 lbs.).
ACCESSORIES SUPPLIED: Tiko Model 70515 BNC cables.
ACCESSORIES AVAILABE
Model 5904: 2OnF/2OmS Adapta
Model 5905:
Model 5906:
Model 5907: Cable/M&ix Calibration Sources
Model 7w7-1: Shielded IEEE-188 Digital cable, lm (3.3 ft.)
Model 7007-2: Shielded IEEE&?8 Digital Cable, 2m (6.6 ft.)
Model 7051-2: BNC to BNC Cable, 0.5m (2 ft.)
Model 7053.5: BNC to BNC Cable, 1.5m (5 ft.)
Model 7051-10: BNC to BNC C&k, 3m (l0 ft.)
speciacaaans sub++ to change wtdlout nolice
30V rms, dc to 6LlHz. Rear panel switch allows con-
lNF
LiT low to chasw.
Cal&ration Sources and Adaptas for 5901lM.
Cal&ration Source and Adapters for 5YO/lM,
59U/lOOk, 590/1OOk/lM, and 5904
5901lOOK ANALOG OUTPUT PERFORMANCE
ACCURACY ti Year)
180 -2vc
RANGE
20 pF 1%+~(5OxG/G,+l)
206
2LW pF 1% + (50 x G/G, + 0.5)
2OOps 1% + (20 x c/c,s + 0.5)
k(% reading + mv)
1% + (20 x UC, + 1 )
P-P NOISE’
ANALOG FILTER ON
TEMPERATURE coEFFIaEwI SHLNTCAPAClT.4NCZ
00 -w & 7.80 -WC LOADING EPFECl”
iph lwiblg + mv) iPh readbIg + mvl
0.2% i (l0 x GIG, + 0.1) 0.1 % + (0.3 x G/G&
0.2% + ( 4 x UC, + 0.1)
0.2% + (l0 x GIG, + 0.1) 0.1 % + (0.3 x GIG,)
0.2% + ( 4 x CGs + 0.1)
0.1 % + (0.3 x UC,)
0.1 % +~ (0.7 x UC,)
2nF
ZUIS 2% + (20 x UC, + 0.5) 0.6~5
2onF 3% + (50 x GIG, + 0.5)
24hS 3% + (20 x UC, + 0.5)
2% + (50 x G/6+ 0.5)
900 fF
Accuracy stated for Q 220 Typical enor for Q <20 .
Wing Model 5904 2onF12Om.S INPUT ADAFTFR.
NOTES:
1. G = conductance reading; C = capacitance reading; G, = fidl scale
conductance; C, = tidl scale capacitance.
2. Range and accuracy designations based on parallel RC model.
3v”Shunt Capacitance Loading’ is additional accumcy ermr with equal
shunt load on Test Output and Test Input, per lWpF load.
4 Noise specified with 5COpF shunt loading on Test Output and ‘I&t In-
put. Noise on 2pF and 20pF ranges is typical with loopF shunt loading;
ACCURACY (1 Year)
180 -WC P-P NOISE’
f(% reading + mvl ANALOG FIUER ON
2% + (75 x GIG, + 1)
2% C~( 30 x UC, + 1)
2% + (75 x GIG, + 1)
3% + (30 x UC, + 1) 0.9~ mv
L2mV
0.75mv
l.4~inV
20 pF
zoog
200 PF
2UiS
RANGE
a4% + (l0 x GIG, + 0.2) 0~2% + (0.2 x GIG,)
0.4% + ( 4 x UC,, + 0.2) 0.02% + (0.3 x UC,)
0.6% + (10 x GIG, + 0.1) 0.1% + (0.2 x GIG,)
0.4% C~( 4 x UC, + 0.1) 0.1 % + (0.2 x UC,)
5tXpF will increase noise no more than x2.
5.5904 must be caliimted with a particular 59011WK to achieve this ac-
clmxy level.
TESI VOIXAGE: l5mV rms fU)%.
TEST FREQUENCY: l~33kHz. Tolerance: tOI%.
TEMPERATURE COEFFIaENT SIIUNTCAPAClTANC??
00 -WC & ‘28” -WC WADING EFFECT
iv0 readlug +~ mv)
0.15% + @ x GIG,)
*P/o reading + m&l
0.5 % + (25 x GIG,)
O.l5% + ( 6 x CIC,) 0.5 % + (1.0 x CL)
~J!.E% +~ U5 x GIG,)
0.15% + ( 6 x UC,)
0.35% + (4.0 x GIG&
0.35% + ('I.6 x C/G)
2nF UptolnF
zolns up to IGms
2& AbaelnF
2OmS Above 1GmS
NOTE%
5% + (I50 x GIG, + 1)
5% l (40 x c/c,* + 1)
5% + 300 x GIG, + 1)
7% + 40 x UC, + 1)
t
l. G = eonduaance reading C = capacitance reading; G, =
conductance; 6
2. Range and accuracy designations based on parallel RC model.
3. “Sh&t Capacitance Loading” is additional accumcy error with equal
= full scale capacitance.
0.3 mv
0.2.mv
03 mv
0.2 mv
full scale Input. Noise cm 2pF and 2GpF ranges is typic.4 with l0lpF shunt loading;
shunt load on Test Output and Test Input, per lwpF load.
4. Noise specified with 5oOpF shunt loading on Test Output and Test
0.15% + (l5 x GIG,)
OS%.+ (6 x~CIC;s)
0.15% + (I.5 x GIG,)
0.35% + ( 6 x UC,)
5oOpF will iraease noise no more than X2.
TEST VOLTAGE: l5mV rms flO%.
TEST FREQUENCY: n4lj.z. ~&mnce: &u.%.
0.35% +~(40 x GIGm)
0.35% + 06 x gem)
0.35% + (4.0 x GIG,)
0.35% + 0.6 x UC,)
C~rttttins an overview of the instrument, incbxiing
features, unpacking b.tstn~&ons, as well as a brief
description of available accessories.
SECTION 1
General Information
Indudes an ov&ew of front and rear panel configuration and basic test procedures. Use the information in
this section to get your Model 590 up and running as
quickly as possibIe.
This section contains detailed information on operating
all available versions of the Model 590. Use this sec-
tion as a reference to all front panel operation.
Section 4 contains information on connecting the
Model 590 to the IEEE488 bus and programming the
instrument from a computer.
Outlines procedures necessary to verify &at the Model
590 and the 1OOlcHz and lMH.z modules are operating
within stated specifications.
SECTION 2
Getting Started
SECTION 3
Operation
,., ,.. ,”
SECTION 4 /
IEEE-488 Programming /
I
SECTION 5
Performance Verification
A complete description of operating principles for the
instrument is located in this section. Analog, digital,
microcomputer, and power supply circuits are described, as are the IEEE-488 interface, and the capacitame modules.
Details maintenance procedures for the Model 590, including fuse replacement, line voltage selection, calibration, and troubleshooting.
TEST CONNECTIONS .................................................
BNC Test Jacks ...........................................................................................................
Typical Test Configuration ..............................................................
Grounded and Floating Operation ...........................
READINGS AND HARDWARE CONTROL ..............................................................
Capacitance and Conductance Displays.. .................
Bias Voltage Display .............................................................................................................................................
Hardware Control Considerations
RANGE SELECTION ................................................................................................
Available Ranges ..........................................................
Manual Range Selection.. ................. ..t...................................~
Using Autoranging ...............................................................
Using the 20nF/20mS Range.. .................... . ...............
FREQUENCY SELECTION ...........................................................
Frequencies Available by Model.. ...................................................................................
Test Voltages .......................................................................................................................................................
Selecting a Frequency ...................................................
Frequency Error Messages ..................................................................................................................................
Disconnecting the Test Voltage
SERIESfPARALLEL MODEL ...............................................................................................................................
Measurement Model ~~~~~~~~~~~~...~......~............................................~~~~~~~~~~~~.~~~~~~~~~~.~~~~~~~~~~..~~.....................~....~~~ - 3-14
Model Selection
Conductance and Resistance Ranges.. .......................
Series and Parallel Equivalent Circuits.. ................
External C vs. t Measurements ............................................................................................................................
InputandOutputJackCoufiguratioti..
Typical Test Connections
EquivalentCircuitofTestConnections
Floating/Grounded Operation of Analog Common
Capacitors, Conductance f.Res&tance), and Bias Voltage Displays.
Flow Chat of Autorangiug Operation
Typical Test Connections Using Model 5904 20&20mS Adapter
Frequency Selection Flow Chart
Equivalent Citcuit of Parallel Capacitance and Conductance
Equivalent Series and PsmBeI Impedances
TypicalAnalogFiIterResponse.. ...........
Zero OperationFlowcharts...................I
FrontPanel CAL Sequence
Internal Drift Correction Sequence.
ExternaI C vs t Connections
cvstDisplayFomat..
Coimections for Cable Correction.
Two-Port Network
AnalogOutputComectionExample..
Multiple Ground Points Create a Ground Loop.
Elimina&gGroundLoop..
Parasitic Capacitance
General Data Format ................................
0 Command Data Format
UO Status Word Format (Hardware/Software Revision). ..........................................
UlErrorStatusWordFormat..
U2 Status Word Format (Buffer A Range Group)
U3 Status WOrd (Buffer A Trigger Group) .......................................................
U4 Status Word Format (Buffer A Zero Group). .......................................
U5 Status Word Format (Buffer A Bias Group) .........................................
U6~Statu5 Word Format (Buffer A Bias Voltage) .................................................
U7 Status Word Format @&er A Bias Times) ...................................................
US Status Word Format (Buffer A Position and Time). ...........................................
U9 Status Word Format (Buffer B Rang& Group).
UlOStatusWord(BufferBT$gerGro~.
Ull Status Word Format (8
U12 Status Word Format (Buffer B Bias Group). ........................................
U13 Status Word Format (Buffer B Bii Voltage).
U14 Status Word Format (Bufferg Bii Tie)
U15 Status Word Format (Buffer B Position and Tie) ...........................................
U16 Status Word Format (Buffer A M&mum and Minimum Capacitance)
U17StatusWordFormat(BufferAMaximumandhIinimumConductance).
U18 Status Word Format (Buffer A h4aximmn and Minimum Voltage).
Ul9StatosWordFormat(BufferBMaximmnandMinimumCapacitance).
U20 Status Word Format (Buffer B Maximum and Minimum Conductance)
U21 Status Word Format (Buffer B Maximum and Minimum Voltage)
U22 Status Word Format (Global Programmin
U23 Status Word Format (Plotter Progmmmin g Parameters)
U24 Status Word Format (IEEE Output Parameters). .............................................
U25 Status Word Format (lEEE Input Parameter) ................................................
U26StatusWordFormat
SRQMaskandStatusByteFonnat..
SRQ Timing with One-Shot Trigger Mode
This section contains information on Model 590 features,
warranty, manual addenda, specifications, and ssfety
terms and s bols. Also included are procedures for un-
packing and%specting the instrum ent, as well as a brief
description of available accessories.
The information in Section 1 is arranged as follows:
1.2 Features
1.3 Warranty Information
1.4 Manual Addenda
1.5 Safety Symbols and Terms
1.6 Specifications
1.7 Unpacking and inspection
1.8 Preparation for Use
1.10 Optional Accessories
1.2 FEATURES
The Model 590 CV Analyzer is a sophisticated instrument
designed as a complete solution for individuals requiring
capacitance and conductance versus voltage measurements
in semiconductor testing. The unit can test devices at either
1OOkHz or lMH.z, depending on installed modules. The
Model 59011OOk tests at loOkHz, while the Model 59011M
operates at lMH.z. The Model 59O/lOOk/lM can test at both
1OOkHz and lMHz. Test voltage for both frequencies is
15mv RMS.
The Model 5901100k measures capacitance and conduc-
tance on four ranges: 2pF/&S, 2OpF/20@, 200 FI200pS,
and 2nFQrnS (the optional Model 5904 Input A apter can
extend the 1OOkHz measurement range to 2OnF/20mS).
Shnilarly, the Model 5901lM measures capacitance and
conductance at lMHz on three ran es: 20 F/200&,
2OOpF/2mS, and 2nFl2OniS. The Mode 59011 OkllM in- 5 B
eludes both measurement capabilities.
1~
l A standard internal +2OV bias source that can generate
staircase, pulse imin, or DC waveforms. Provision to connect an external bias source of up to *2OOV DC are also
included.
l Two 450-word internal buffers to store capacitance (C),
conductance (G), and bias voltage (V) data taken during
testing. Two complete sets of C, G, V~d.ata can be stored;
one set can be saved for plotting while another test is
being performed.
l Standard plotter driver software allows the hIode 590
toTontro1 an intelligent digital plotter over the IEFE-488
bus, simplifying a variety of different plot types, including C vs V, G vs V, l/C= vs~ V, and Clc, vs V.
l Nominal reading rates of 1, 10, l&75, or loo0 readings
per second allow you to choose the best compromise between resolution, noise perfo-ce, and speed.
l Selectable analog filtering is included to minimize noise.
l External trigger inputand output capabilities are included
to synchronize the Model 590 with other equipment such
as external bias sources.
l Analog outputs of capacitance, conductance, and bias
voltage are included to allow the monitoring or analog
plotting of these readings with external equipment.
l Isolated analog and digital sections, which allow
measurements with common mode voltages up to 3OV
RMS.
l Internal calibration reference sources for maximum
accuracy.
*Built-in correction software to compensate for cable
transmission line effects that would otherwise degrade
accuracy. Up to seven sets of cable parameters can be
stored for later recall at the touch of a button. The unit
can also compensate for non-uniform transmission lines
with the aid of external standards.
l Internal math ability to simplify calculation of such para-
meters as parallel/series model, capacitance difference
and ratio, and n-n&mum and minimum capacitance
values.
l Up to seven different instrument configurations can be
stored and later recalled to simplify imimment conflguration. The factory configoration can also be recalled at the
touch of a button.
l A standard IEEE-488 interface is included, allowing the
instrument to be programmed from a computer. Enhanced Keithley Translator software simplifies pro-
gramming.
l-l
GENERAL INFORMATION
1.3 WARRANTY INFORMATION
Warranty information for your Model 590 may be found
inside the front cover of this manual. Should it become
necessary for you to use the warranty, contact your
Keithley representative or the factory for information on
obtaining warranty service. Keithley Instruments, Inc
maintains service facilities in the United States, West Germany, Frame, the Netherlands, Switzerland, and Austria.
Information concerning the operation, application, or service of your instrument may be directed to the applications
engineer at one of these locations.
1.4 MANUAL ADDENDA
Information coneming changes or improvements to the in-
strument which occur after this manual has been printed
will be found on an addendum sheet inchrded with the
instrument. Please be sure to read this information before
attempting to operate or service the instrument.
1.5 SAFETY TERMS AND SYMBOLS
The following safety terms are used in this manual or found
ontheinstrum
The symbol
should refer to the operating instructions in this manual
for further details.
The WARNING heading used in this manual explains
dangers that could result in personal injury or death.
Always read the associated information very carefully
before performing the indicated procedure.
The CAUTION heading used in this manual explains
hazards that could damage the instument. Such damage
may invalidate the warranty.
althea-~tindLatffthattheuser
immediately. Retain the original packing material in case
reshipment becomes necessary. 0
1.7.1 Shipment Contents
The following items are included with every Model 590
shipment:
Model 590 CV Analyzer
Model 590 Instruction Manual
Model 7051 RG-58, BNC test cables (2)
Additional accessories as ordered.
1.7.2 Module Complement
Modules ordered with the unit will be shipped already in-
stalled and calibrated. Available models include:
59011OOk
59o/lM
5ti/lOOk/lM
Note that the module complement is indicated by model
on the rear panel.
1OOkHz capacitance module only
lMfiz capacitance module only
Both 1OOkHz and lMHz capacitance
modules.
1.7.3 Additional Instruction Manuals
If an additional instruction manual is required, order the
manual package, Keithley Part Number 590-901-00. The
IMnual package indudes an instruction manual and all pertinent addenda.
1.8 PREPARATION iOR USE
1.6 SPECIFICATIONS
Detailed Model 590 specifications are located at the front
of this manual.
1.7 UNPACKING AND INSPECTION
The Model 590 was carefully inspected and packed~before
shipment. Upon receiving the instrument, carefully unpack
ail items from the shipping carton and inspect for any obvious signs of physical damage that might have occurred
during shipment. Report any damage to the shipping agent
l-2
1.8.1 Line Power
The Model 590 is intended to operate from 105-l25V or
210-25OV AC power sources. A special power transformer
may be installed for $61lOV and 180~220V ranges. The fac-
tory set voltage range is marked on the rear panel.
1.8.2 Line Voltage Selection
The operating voltage is selected by a switch located on
the rear panel. Before using the instrument, make sure that
the switch is in the correct position for the line voltage in
y0i.u area.
GENERAL INFORMATION
CAUTION
Do not attempt to operate the instrument on a
line voltage outside the indicated range, or instrument damage may occur.
1.8.3 Line Frequency
The Model 590 may be operated from either 50 or 6OHr
power sources.
1.8.4 IEEE-488 Primary Address
If the Model 590 is to be programmed over the IEEE-488
bus, it must be set to the correct primary address. The
primaryaddress has been set to 15 at the factory, but~it
can easily be changed from the front panel, as described
in Section 4.
1.9 REPACKING FOR SHIPMENT
Before shipment, the unit should be carefully packed in
its original packing carton using all original packing
materials.
Model 2289 Slide Rack Mount Kit-The Model 2289 Kit con-
sists of two sets of flanged brackets, equipment slides, and
hardware for mounting the Model 590 in a standard lP-inch
equipment rack or cabinet.
Model 5904 Adapter-The Model 5904 extends the 1OOkHz
measurement range of the instrument to 2OnF/2Oms. The
Model 5904 mountsdirectly on the INPUT and ODTFUT
jacks and includes BNC connectors for test cable connections. Note that the Model 5904 and Model 590 must be
calibrated as a matched pair for stated accuracy.
Model 7007-l IEEE-488 Cable-The Model 7007-l lm (3.3
ft.) shielded IEEE-488 interface cable is equipped with a
shielded IEEE-488 connector (metric) on each end.
Model 7007-2 IEEE-488 Cable-The Model 7007-2 2m (6.6
A;) shielded IEEE-488 interface cable is equipped with a
shielded IEEE-488 metric-saew connector on each end.
Model 7051 BNC to BNC Cables-The Model 7051 cables
are made up RG-58 503 cable terminated with a male BNC
connector on each end. Three lengths are available: The
Models 7051-2, 7051-5, and 7051-10 are 0.511-t (2 ft), 1.5m
(5 ft), and 3m (10 ft) in length respectively.
1.10.2 Calibration and Verification Sources
If the instrument is to be returned to Keithley Instruments
for repair, completes the following:
Write ATTENTION REF’AIR DEPARTMENT on the shipping label.
In&de the warranty status of the instrument.
Complete and h-&de the service form at the back of this
m.snual.
1 .lO OPTIONAL ACCESSORIES
The following accessories for the Model 590 are available
from Keithley Jnstruments, Inc. Contact your Keithley
representative or the factory for information on obtaining
these accessories.
1.10.1 General Accessories
Model2288 Fixed Rack Mount Kit-The Model 2288 Kit includes two ganged brackets and hardware for mounting
Eb$eyiel 590 in a standard 19-inch equipment rack or
The calibration sources listed below are intended for use
in field calibration or accoracy verification of the Model 590.
Each source is mounted in a shielded test Rxture, which
is equipped with BNC connectors. These fixtures are intended to connect directly to the front panel test INPUT
and OUTf’UT jacks to avoid cable errors (except for the
Model 5907 sources, which connect to cables through supplied adapters).
Sources used with each mode1 are summarhed in Table
l-l. Table l-2 summarizes nominal source values.
Table l-l. Calibration Source Sets by Model
Number
*Used for cable correction only; not needed for normal
calibration.
1-3
Model 5905 Calibration SOWC~S-The Model 5905 set con- Table l-2. Calibration Sources and Nominal Values
bins all the capacitance and conductance souxes necesmy
to calibrate or verify accuracy for the Model 59011M. See
Table l-2 for sources.
Model 5906 Calibration Sources-The Model 5906 sources
are necessary to calibrate or verify the Model 590 when
used with a Model 5904 ZO$/ZOmS Adapter, and are also
needed to complete calibration or accuracy verification of
a Model 59O/lOOk or a Model 59011OCkllM.
4.7pF, 18pF, 47pF, ISOpS, l.SmS, 18116
Model 5907 Calibration Sources-The Model 5907 sources
are intended for cable correction when using the calibra-
tion capacitor method of cable correction. The Model 5907
includes both 47OpF and 1.8nF capacitance sources, and
adapters for connecting the sources to the ends of the test
cables.
“Model 5905 and 5906 include right angle ada er and
BNC short for driving point-cable correction c ai? ‘bration.
**Model 5907 inchxdes two female BNC-to-BNC adapters
to connect source to cables.
tModeI5906 also includes all sources in Model 5905.
l-4
SECTION 2
GETTING STARTED
2.1 INTRODUCTION
This section contains intmductory information on operating
your instrument and is intended to help you get your Model
590 up and running as’quickly as possible. It includes a
brief desaiption of operating CO~IIK& and test connedions.
Once YOU are familiar with the material presented here,
refer to Section 3 for more detailed information.
Section 2 is organized as follows:
2.2 Front Panel Familiarization: Briefly describes each
front panel control and test connection, outlines display
operation, and lists where to find more detailed infor-
mation in Section 3.
2.3 Rear Panel Familiarization: Outlines each aspect of
switches.
2.4 Power Up Procedure: Describes how to connect the
ins&ument to line power, properly sekct line voltage, and
the typeof display messages to expect during the power
procedures for making simple one-point measurements,
CV measurements, plotting data, and performing C vs
t measurements.
2.2 FRONT PANEL FAMILIARIZATION
An overview of the Model 590 is given in the following
paragmphs. The front panel of the instrument is shown in
Figure Z-l, along with a brief desaiption of each item. Table
2-l is a cross reference to other sections of the man& where
more detailed information may be found.
2-l
Table 2-1. Model 590 Front Panel Cross Reference
IieXtl
1 LOCAL
2 POWERS
3 RANGE
4 Fi=Q
5 MODEL
6FILTBR
7 RATE
8 ZERO
16 GRID
17 SETUP/ABORT
18 A (Increment)
19 7 (Decrement)
20 ENTER
21 BUJZER
21 A-B
22 SHJlWQUlT
38 Bias Voltage Display
39 BUFFER and MEASURE
40 INPUT
41 OUTPUT
Description
Cancel remote, restore local operation.
Control AC ‘power.
Select mnge, auto, x10 attenuator.
Select 1OOkHz or IME test fre uency.
Select series R and C or paralle C and G.
Control single-pole analog low-pass filter.
Select 1, 10, 75, or 1000 reading per second rate.
Enable, disable baseline suppression.
Calibrate unit to internal standard.
Initiate reading or sweep.
Program trigger mode and source.
Turn DC bias on or off.
Program bias wavefom type.
Program bias voltages and times.
Plot over IEEE-488 bus.
Add second function to some keys, cancel menu, buffer, or
parameters.
Program numeric data.
Scroll cursor when progmmming parameters.
Calibrate cable for 1OOkHz or lMJ& use (driving point only).
Select cable correction parameter set.
Perform test of internal components.
Store up to seven instrument setups in NVRAM.
RecaIl up to eight instrument setups from NVRAM.
Program IEEE-488 primary address (O-30).
Invert C and square value.
Display normalized capacitance.
Display
Display differ&t between buffers A & B
Plot AV at constant C.
Display C as a function oft time (buffer index).~
Display capacitance reading.
Display conductance read@g.
Show IEEE-488 bus status
Display programmed or measured bias voltage.
Indicate display and reading status.
BNC connector to measure test signal.
BNC output ~applies lOLX& or lMHz test voltage and bias
voltage to circmt under test.
maximum capacitance.
?
Paragraph
4.7
2.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
E
3:14
3.14
3.14
3.16
3.16
3.16
3.13
3.13
3.13
3.15
3.15
Several
3.13
3.20
3.20
3.18
3.17
3.17
E9
3:19
3.19
3.19
3.19
3.20
3.4
3.4
4.7
3.14
2.2
3.3
3.3
Press SHIFT first to access these modes.
2-2
GETTING STARTED
El LOCAL-Pressing this key when the unit +s in
remote (REMOTE on) returns the instrument to
the local mode (REMOTE OFF) and restores
operation of other front panel controls unless
LLO (local lockout) is in effect.
POWER-POWER controls AC line power to the
El
instrument. The unit will be off when the switch
is out (0 position) and on when the switch is in
(1 position). Before applying power, make sure
the line selection switch on the rear panel is in
the correct position for the AC line voltage in your
area.
CONTROL GROUP
RANGE-Press RANGE Iniefly to manually
El
select rtige: 2pF (1OOkHz only), TOpF, 2OOpF, or
21-3. Pressing and holding RANGE for more than
a half secon;d places the-tit in autoranging, as
shown b
to cance auto and stay on present range.
RANGE switches in X10 attenuator to extend
lOOkI& measurement range to 2OnF with exter-
nal optional input transformer (Model 5904).
Press !ZHFJ? RANGE again to cancel X10.
Autoranging is included for convenience only
and should not be used for time critical
AUTO indicator. Press RANGE a ain
P
NOTE
S
AFT
Ranges Include:
1OOkHZ
2pFl2pS
2nF/2ms
2OnF/2Oms*
*Requires Model 5904 Input Transformer and X10
attenuator
q
FREQ-Press FREQ to select test frequency,
1oOkHz or IMHz at 15mV RMS. The excitation
voltage is applied to the circuit under test along
with the progr-ed bias voltage through the
OUTPUT jack. The associated indicator wiU in-
dicate the selected frequency. CONFLICT mes-
sage will be displayed if you attempts to use the
X10 attenuator at lh4H.z. FREQ cti also be used
to disconnect internal test voltages from the
device under test. The tit will display DISCON:
NECT in this case.
MODEL-MODEL selects series or parallel de-
El
vice model (series resistance and capacitance or
parallel conductance and capacitance). The iostru-
ment always measures and stores buffer data in
arsllel form, but the series equivalent is calcu-
ated and displayed when the series model is
P
selected. The selected model is shown by the
associated indicator.
2nF/20mS
Figure 2-1A. Model 590 Front Panel
2-3
GETTING STARTED
Full Scale Conductance/Resistance: Digital
Key# Rate Resolution Readings Fil&ing
lOOkI%? llMH2
G R G
2im
2OOkfl 2oo!.Ls
2OkQ
2ms
2Oms 2OOQ
FILTER-FILTER toggles the single-pole low-pass
El
analog filter on and off, as shown by the indicator
adjacent to the FILTER button. The approximate
-3dB point for the filter & 37H.2, and the effects
of the filter are reflected both at the analog out-
puts on the rear panel and on the display. Note
that the filter increases in&ument response time.
RATE-Press RATE then AN (or RATE) to saoll El CAL-J?ressing CAL performs an automatic one
El
through the rate selection menu: 1, 10, 75, or point calibration,of the selected module on the
1000 readings per second (or press the numeric current range using an internal 2OpF or 200pF
key indicated below). Press ENTER to seled~
RATE, or QTJIT to return to the previous rate. The
slower rates will provide more resolution and
quieter readings, as indicated below. frequencies for optimum accuracy.
2ki-l 2oms 2kQ
hS
R
2OOkQ
2OkQ calculated.
0 10001sec 3%*
pc g*
:
3 l/SW 4% C; G; V Yes
*Data displayed only after sweep is finished and
ZERO-ZERO provides means for supression of
El
a constant value from the readings, or it can be
used to cancel internal offsets to maximize ac-
ccuracy. Enabling ZERO stores the next reading
as the baseline value, which is then subtracted
from subsequent readings and stored in the bof-
fer header. Note that enabling zero reduces the
dynamic range of the measurement, and that the
zero value is carried from one mnge to another.
@~p+ling on range) capacitance source, and is
intended to corn
drift. CAL shoul
2 C. G. V Yes
nsate for shoe-term thermal
ge be used for each mnge at both
C only No
C, G, V No
2-4
NOTE
NOTE Do not press and hold CAL during the
Reading rates are slightly different than indicated because of the way the unit generates its
time base. See paragraph 3.9.
Figure 2-1A. Model 590 Front Panel (Cont.)
GETTING STARTED
TRIGGER GROUP
q
Ill
MANUAL-Pressing MANUAL will initiate a
one-shot or sweep sequence depending on the
selected trigger mode. This key is operational
regardless of the selected trigger source. Pressing MANUAL while a reading or sweep is in
progress will result in a trigger overrun error
message.
MODE/SOURCE-Press MODE then AN ,
MODE, or numeric key (see list below) to select
a trigger mode: one-shot or sweep, then press
ENTER. In One-shot, the instnunent will process
one reading per irigger, while in sweep the unit
will process a complete reading sweep (one complete reading sequence at all programmed bias
steps, with up to 450 readings stored in buffer
A). Both modes are available with all trigger
sources.
Press SOURCE (SHIFT MODE) then MODE or
A/Y to scroll through available trigger sources
(or press the appropriate numeric key in the list
below). Press ENTER to program displayed
source: front panel (MANUAL button), external
(a negative-going T%compatiile pulse applied
to the rear panel trigger input jack), tis well as
GET, X and talk commands sent over the
IEEE-488 bus.
Front panel trigger messages include:
NUllltiC
Key #
1
2
3
4
*Always enabled regardless of selected source.
Message Description
TRIGGER MODE One reading
l-SHOT
TRIGGER MODE voile sweeo
SWEEP
TRIGGER SOURCE Front Panel
Fr MANLJAL
TRIGGER SOURCE External trig-
EXT
TRIGGER SOURCE IEEE talk
TALK
TRIGGER SOURCE IEEE GET
GET command
TRIGGER SOURCE IEEE X
~~X~~
per trigger
per tiigge;
button*
ger pulse
command
command
Figure 2-1B. Model 590 Front Panel
2-5
9-z
huO3) Wed W’Jd 069 IaPW ‘Bl-Z aJn6!d
Programmed
Message Limits Resolution
SIARTTJME lms to 65sec
sToPT!mE lm to 65sec
lmsec
lmsec
STEP TIME lms to 65sec lmsec
FIRST BIAS V -20x7 to 2oV
LAST BJAS V -20x7 to 2oV
STEP BJAS V -20V to 2ov
5mv
5mv
5mv
DEFAULT -20V to 2oV ~5rnV
BIAS V COUNT* 1 to 450 (1,350
ate looo/sec
rate)
*Selects number of readings stored for external
and DC bias wavefom.
**Voltages can be programmed to lmV, but are
set in 5mV steps.
NOTES:
1. Multiply programmed times by 1.024 for actual time intervals.
2. Minimum stop time with pulse waveform is
50msec (IOlsec rate).
PLOTTER GROUP
Ed
El
El SETUP/ABORT-pressing SETUP enters the
PLOT-Pressing PLOT plots the data located in
the selected buffer (A or B) on an intelligent plotter over the IEEE-488 bus using the current
SETUP parameters.
NOTE
Disconnect the controller from the IEEE-488 bus
of the Model 590 before using PLOT or GRID.
GRID-Pressing GRID draws labels, axes, and
other parameters as appropriate for the selected
buffer and the SETUF’ parameters.
plotter setup menu which allows selection of the
parameters below. Use ~A or v to scroll through
menu selections then press the appropriate mmher (below) when desired selection is displayed,
then ENTER.
GETTING STARTED
Press ABORT (sm SETUP) to halt plqtting or
grid generation.
Parameters include:
Para-
Grid
meter Line Type Type Label Type
0
1
Dot at points FUJI grid Full labels
Spaced dots Axis only Labels axis
and divisions
2
Dashes Long dash
Dash dot
Long dash
-
-
Labels axis
OdY
No labels
-
-
short dash
Long dash
-
-
short dash,
long dash
7
meter Plot Type
0
1
Solid line
Type Buffer
c “S v
No Den A
G vs V* B
llc2 “S v ii -
:
c/co “S v
c “S t**
i
6
c*-c, “S v
[v.,-V,]C=CONST 1 1
- -
-
*R vs V with series model.
**Plots buffer index.
Para-
meter X Axis
0*
Auto scaling Auto scaling
1**
User-defined User-defined
scaling
YAxis
scaling
*X axis scaled tq minimum and nmximum values
Y axis scaled according to range ~of function.
**Use numeric keys to enter scaling factors, then
press ENTER.
Figure 2-18. Model 590 Front Panel (Cont.)
2-7
GETTING STARTED
--
--
IATA GROUP
Increment (&-Increment is used to scroll
through menu selections for other front panel
operating modes such as TRIGGER MODE,
PLOTTER SETUP, and BIAS WAVEFORM. In-
crement is also used to saoll through buffer locations when displaying buffer data.
Decrement (V-Like the incrementkey, decrement is used to scroll through parameter menus
and buffer locations, but in the opposite
diI&iOIL
ENTER-ENTER is used as the last step in the
menu and parameter selection process to a$ually
perform the operation being programmed.
BUFFER-Pressing BUFFER allows you to view
the contents of buffer A or buffer B on the front
panel displays. Once in this mode, select the
desired buffer (A or B) and use A or V to-se-quentially access various buffer locations. The
capacitance (C), conductance (G), and measured
bias voltage (I’) for each buffer location w-ill then
appear on the display as you access them. The
BUFFER LED will be on while the unit is displaying buffer data. Pressing BUFFER while accessing the buf?er displays the last valid buffer location. Pressing ENTER displays the first~valid buffer location (location Xl). Press QUIT to etit the
buffer.
tWHt
Ia
-
I-l-
A - B (SHIFT BUFFER)-A - B places the entire contents of buffer A into buffer B, including
capacitance, conductance, and bias voltage
values. Buffer A is the buffer into which AID
readings are stored, and buffer B is the plotter
buffer. Buffer A will be cleared after the data is
transferred.
SHLFUQlJl’-SHIFT adds a secondary function
to certain other front panel keys, in&ding
SETUP and RANGE (SHIFT RANGE~enables or
disables the X10 attenuator. The shifted modes
are marked below the keys in questions. While
shift is enabled, the indicator to the right of that
key will be on. To cancel shift, either select the
mode in question, or press SHIFT a second time.
Tf you press a key which has no second functior~
after enabling shift, the primary function of that
key will be performed. For example, SHlFlY
MODEL is the same as MODEL.
Press QUIT to retmn to normal operation while
selecting menu options, programming parameters, or viewing buffer data.
2-a
Figure 2-1C. Model 590 Front Panel
GETTING STARTED
q to El NUMERIC DATA KEYS (C-9, k,-) - Ia
These keys are used to enter numeric data when
programming such items asp bw~parmeters. If
you wish to restore the previously programmed
values, press the QUIT (SHIfT ENTER) key instead. Pressing the - key allows you to move the
display cursor to the right while programming
parameters.
Do not press - during power up, or insixument calibration may be altered if the
CAL switch is in the unlocked position.
q
Ed
q
CABLE CAL*-Pressing this key performs opencircuit cable correction. Note that the opposite~m
ends of the connecting cables must be left open
during the correction process. Once the correction is complete, you will be @en an opportunity to store the correction scheme for the particular cable (l-7) you are using at the update op-~~-- El
tion. TWO other forms of cable correction are
available only over the IEEE-483 bus, as discussed in Section 4.
CABLE #*-Use this key to select which of eight
previously stored cable correction set ups you
wish to use (O-7). Once selected, the unit will
automatically use the previously stored cable correction pammeters when making measurements.
Note that correction set up #O turns offcable car- Ed
rection and installs default values.
SELF TEST*-Use this key to perform a self test
on many internal components, including the
display. If no problems are found, ~the ins&ument will return to normal operation; however,
if an error occurs, an INVALID mess&&ili be
displayed.
NOTE
NOTE
Using cable correction can reduce the
dynamic range of the capacitance and
conductance readings.
m
H
SAVE*-SAVE allows you Jo save up to seven
complete instrument configuations in NVRAM.
To use this featue, simply select the operating
configuration and then press the SAVE button.
Key in the position (l-7) that you wish to save.
Note that state 1 is the con&oration the unit will
assume upon power up.
RECALL*-Use RECALL to assume machine
operating configurations that were stored with
the SAVE key, or the factory configuration.
Upon entering this mode, you will be prompted
for a configuration number. Key in the value
(O-7) and press ENTER. Note that state 0 is a factory default configuration p
ROM and cannot be altered.~ State 1 is the con-
figuration the ins@ument assumes upon power
up. RECALL can also be used to restore normal
buffer display after using a math function.
IEEE*-Press IFEE to verify or program the
lFEE4%3 primary address. Use the number keys
tom select a primary address value (O-30). Press
ENTER to program the new address. The programmed address will go into effecter
immediately.
MATHEMATICAL FUNCTIONS
The following calculations are performed on data
presently stored in the data buffers and are not
stored in memory. fin order to uses these functions, you must select buffer display with the
BLJFFER key. If reading normal instrument data,
pressing one of these keys will have no effect.
l/C-Pressing l/C2 inverts *e ~apaci@nce value
in each data~word of the selected buffer and then
squares it; the’ value for each pointy wilI be
displayed as you access that~ word location.
ermanently stored in
*SHIFT must be pressed first to access these modes.
Figure 2-1C. Model 590 Front Panel (Cont.)
2-Q
GEITING STARTED
C/Co-This feature allows you to display normalized capacitance data. Each capacitance value
in the selected buffer will then be divided by Co
and then displayed. The maximum capacitance
presently stored in the buffer is automatically us-
ed for Co.
Cm,-Pressing 6, displays the maximum
capacitance value stored in the selected buffer.
CA&-This key allows you to q&tract ea&
capacitance value in buffer B from the cor-
responding values in buffer A.
W., -V,l C =CONST.-This function rotates the
C-V plot axis by 90” and gives a display of the
change in voltage (AV) as a function of constanf
capacitance.
Figure 2-1C. Model 590 Front Panel (Cont.)
C vs t-This mode uses the bias waveform
q
parameters in effect when the readings were
taken to display capacitance versus time. While
in C vs t, you can use A/I to scroll through
vaious buffer locations. The buffer location
number will be shown in the L+z voltage display.
You can calculate the time at a specific location
as follows:
ti = tars + (tamp + l/R) (B)
Where: ta = time at a s@ecific buffer locatioti~~
t
t
R = reading rate (readings per second)*
B = buffer_location-cuq+r
*Use actual rate, not the nominal displayed value.-
Also multiply times by 1.024.
Programmed timesmust be multiplied by 1.024
to obtain actual times.
programmed start time
szan =
n.p
= prog-ed step time
2-l 0
GETTING STARTED
1
CAPACTMNCE DISPLAY-The normal capacitance display is a 4%-d@ 21,999, ~-19,999 co@
value. The decimal point along with the
associated engineering units indicate the range.
The displayed engineering units are pF and nF.
The display will show dashes when no valid
reading is available (for example, if the AID configuration is changed, or of no valid buffer data
is stored when accessing the buffer). Note that
display resolution is 3% digits at the 75 and 1000
reading per second rates.
CONDUCTANCE DISPLAY-The nominal conductance display is a 4%digit (3%‘~ digits at the
75 and 1000 reading per second rates), 21,999,
- 19,999 Count value, with engineering units also
displayed in pS or mS for conductance (parallel
model) or kQ or MQ for resistance (series model).
Again? the decimal point indicates the selected
range in conjunction with the displayed
engineering units. Like the capacitance display,
dashes will appear if no valid reading is
available.
@I
IEEE-488 STATUS INDICATORS-The TALK,
LISTEN, and REMOTE LEDs indicate when the
Model 590 has been placed in these modes when
the unit is being programmed over the DEEE-488
bus. The TALK and LISTEN indictor show when
the unit has been addressed to talk or listen,
respectively. These talk and listen commands are
derived from the unit’s primary address.
REMOTE will tmn on to indicate when the in-
strument is placed in remote by addressing the
unit to listen w.ith the REN line true. Note that
all front panel controls except LOCAL and
POWER will be inoperative when REMOTE is on.
Local operation can be restored by pressing
~LOCAL unless the IEEE-488 LLO (Local Lockout)
command is in effect.
BIAS VOLTAGE DISPLAY-The 4%-d@ bias
Ed
voltage display indicates the programmed or ac-
tual measured value of the internal *ZOV bias
source or the applied external bias voltage. While
progr&g bii parameters, the display will
show the progTammed value. When the unit is
displaying readings or stored buffer values, the
display will show the bias voltage as measued
by an internal A/D converter. This display will
also show buffer location in C vs t. Note that
display resolution is 3% digits at the 75 and 1000
reading per second rates.
Figure 2-1D. Model 590 Front Panel
2-11
GETTING STARTED
5J
BUFFER AND MEASURE-BUFFER will be on
while the insinunent is displaying buffer data.
MEASURE will turn on during a reading (one-
shot mode) or sweep (sweep mode) to indicate
a reading or sweep is in progress.
ZL -
INPUT INPUT is a BNC connector intended for
applying the measured test signal to the unit. The
center conductor of the jack is input high and the
shell of the jack, which is connected to analog
common, is input low (see below). The shell can
be floated up to 30V RMS above chassis ground,
assuming the groun&ng switch on the rear panel
is in the floating position.
OUTPUT-OUTPUT provides a composite of the
El
l$kHz or lMHz (depending on selected frequency), l5mV RhG test voltage and the DC bias
voltage, either internal or external (WARNING:
up to ZOOV DC may be present between output
high and output low when an external bii
source is connected and selected). The center
conductor is high, and the outer shell of the BNC
jack is low, and is connected to analog common
(see above). Maknmn common-mode voltage for
OWIJT is 30V RhJS v&en the rear panel grownding switch is in the floating position.
I4 HIGH
@
BNC
CONNECTOR
LOW
(ANALOG COMMON)
2-12
Figure 2-1 D. Model 590 Front Panel (Cont.)
The rear panel of the Model 590 is shown in Figure 2-2,
Table 2-2. Model 590 Rear Panel Cross Reference
GETTING STARTED
which also gives a brief description of each item. Table 2-2
lists oaraerauhs in this manual where more detailed informkio~o~ each subject may be found.
Apply 2OOV m2ucimum DC bias voltage.
Monitor internal or external bias voltage.
Select floating or grounded operation of analog common.
Scaled O-2V conductance value.
Scaled O-2V capacitance value.
Interface unit to IEEE-488 bus.
Input TTL pulses to trigger readings.
Output TTL pulses to trigger other instruments.
Select operating voltage range.
Protect AC line input.
Connection for AC input.
Filter cooling air.
Exhaust cooling air.
3.14
3.14
3.3
3.22
3.22
4.3
3.12
3.12
2.4
7.2
2.4
7.8
2.1
2-13
GETTING STARTED
El VOLTAGE BIAS INPUT-This BNC connector
is intended to apply external bias voltage up to
&!OOV DC, 5Gm4 maximum. Note that the input is internally fused to protect the insuument
from over current conditions. The voltage~apphed
toBL4SIhBXiTwillbeappBedtothecircuitunder
test through the OUTPUT jack only when exter-
nal biasing is selected with the WAVEFORM key
and when the bias voltage is turned on (BIAS
ON).
q ~~
VOLTAGE BIAS OUTPUT-This BNC output
jack provides a means to monitor the selected
bias voltage (external or internal) applied to the
circuit under test. The output resistance is lkQ;
thus, the input resistance of the monitoring in-
strument should be greater than lOOk to
minimize the effects of loading.
El
4
VOLTAGE B!AS AND ANALDG OUTPUT
BNC JACKS
HIGH
(%U)G COMMON1
Ed GROUNDING SWITCH-The position of this
switch controls floating or grounded operation of
the, following jacks: OUTPUT, INPUT,
VOLTAGE BIAS INPUT, VOLTAGE BIAS OUTPUT, and the two ANALOG OUTPUT jacks.
When the switch is in the floating position, the
outer rings of these jacks can be floated up to 3oV
RMS above ground. When the switch is in the
grounded position, the rings are connected to
chassis ground.
dil E
!=LDATlNG
-EiiG
GROUNDED
q
CONDUCTANCE ANALOG OUTPUT-This
output jack provides a scaled voltage proportional
to the conductance reading. The range of the out-
put is 0-2V, full scale. For example, the nominal
output value ti be 1V with a 1OpS reading on
the 2GpF/2OpS range. Output impedance is lkR.
2-14
CAPACITANCE ANALOG OUTPUT-This
Ia
BNC jack provides a scaled output voltage that
is proportional to the capacitance reading. The
output range of the CAPACITANCE output is
0-2V full scale. For example, the nominal output
voltage with a 14OpF reading on the 2OOpF/2oO~S
range will be 1.4V. Output impedance is lkR.
Figure 2-2. Model 590 Rear Panel
GETTING STARTED
q IEEE-488~ INTERFACE-This c&n~tor pro-
vides a means to interface the Model 590 to the
IEEE488 bus. When connected to a controller, ’
inslnunent operating modes can be pro:
gmnmed over the bus. CV plots can also be generated via the bus when the instrument is used in
conjunction with an HP7470A or similar digital
plotter. IEEE-488 interface function codes are
marked adjacent to the connector.
EXTERNAL TRIGGER INPUT-EXTERNAL
TRlGGERlNPUT is aBNC jack to be used for ap-
plying a trigger pulse to initiate a one-shot or
sweep reading, depending on the trigser mode
programmed with the TRIGGER MODE key.
Note that external trigger must be enabled, also
with the MODE key. Inputs to this jack must be
TlScompatible, negative-going pulses with a
duration greater than 1~~. The center conductor is high and the outer ring, which is connected
to IEEE common is low, as shown below.
TRIGGERS ON
El
LEADING EDGE
@
LINE VOLTAGE SELECTION SWITCH-The
position of this switch determines the operating
voltage range of the instrument: lO~l25V or
ZlO-25OV (a special transformer is available for
90-11OV and X30-2ZOV ranges). The factory
voltage range is marked below the switch. CAU-
TION Do not operate the instrument on a line
voltage outside the indicated range, or ins&ument damage may occur.
•j ;‘NEW&
SO-1lOV OR 105~125V
LINE VOLTAGE
180-220V OR 210.250V
HIGH
LOW
(IEEE-488 COMMON)
--RIG&R 8NC JAcl$S’
I*
>l@C
q .
lhs BNC jack provides a ‘ITLcompa@&,
negative-going pulse when the instrument com-
pletes a one-shot reading or reading sweep,
depending on the selected trigger mode. The
center conductor is high, and the outer ring is low
(see above).
READING DONE (ONE SHOT)
Eil
STAYS HIGH DURINk
READING OR SWEEP
TTL HIGH
(3.4V TYPICALI
--4
(0.25V TYPICAL)
SWEEP DONE (SWEEP)
OR RE-TRIGGERED
LINE FUSE-The line fuse protects the AC power
line input of the instrument. When replacing the
fuse, use only the type and rating specified on
the rear panel of the unit. CAUTION: Replacing
then fuse with one that has a larger rating than
specified may cause instrument damage.
AC INPUT RECEPTACLE-This receptacle is the
El
AC power line input for the unit. Use only the
sumlied oower cord or the eauivslent with a pro-
p.&iy g&unded AC outlet t6 ensure conti~ed
protection against shock hazards.
FAN FILTER-The fan filter keeps dirt from be-
ing &awn into the instrument by the internal
cooling fan. The filter opening should be kept &se
of obstructions to ensure proper instrument cooling. Clean the filter periodically to assure proper
air flow (See Section 7).
H
EXHAUST VENTS-The exhaust vents direct air
from the inside of the instrument under pressure
generated by~tbe internal cooling faq. They too
must be kept free of obstructions to ensure pro-
lx3 cooling.
Figure 2-2. Model 590 Rear Panel (Cont.)
2-l 5
GETTING START6D
2.4 POWER UP PROCEDURE
The steps in the following paragraphs will take you
through the basic procedures for selecting the line voltage,
connecthg the instrument to line power, and turning on
the i.n8tTument.
2.4.1 Line Voltage Selection
The Model 590 can be operated on line voltages in the
range of 105l25V or 210-UOV, 50 or 6OHz (a speci+ p0y.e:
transformer can be installed for 90-11OV and 180-220V
ranges). Before connecting the unit to line power, make
sue the line voltage selection switch is in the correct position for the power line voltage in your area. See Figure 2-2
for the location of this switch.
CAUTION
Operating the instrument on a line voltage outside the indicated range may cause damage,
possibly voiding the warranty.
2.4.2 Line Power Connections
2.4.4 Power Up Self Test and Display
Messages
During the power up cycle, the instrument will perform
the following:
1. A RAM and ROM checksum test. Jf ti error is found
as the result of one of these tests the insbument will
display either all OS for a ROM failure, or alI As for a
RAM failure. Either type of error is considered fatal, and
the instrument will lock up. Refer to Section 7 for
troubleshootinzz tmxedures.
If the instrmnent is still under warranty (less than
otie year from the date of shipment), and a problem develops, it should be returned to Keithley
Instruments, Inc. for repair. See paragaph 1.9 for
information on retuning the unit.
Assuming the unit successfuUy passes the self test, it
2.
will then briefly display the model number and software
revision level, as in this example:
.,.
NOTE
590 REV D14
Using the supplied power cord, connect the instrument
to an appropriate 50 or 6OHz AC power source. The female
end of the cord connects to the AC receptacle on the rear
panel of the instnunent. The other end of the power cord
should be connected to a grounded AC outlet.
WARNING
The Model 590 must be connected to a grounded outlet in order to maintain continued protection against possible shock hazards. Failure to
use a grounded outlet may result in personal injury or death due to electric shock.
2.4.3 Power Switch
To tom on the power, simply push in the front panel
POWER switch. Power is on when the switch is at the in-
ner position. To turn power off, press POWER a seconds
time.
NOTE
Do not press and hold the CAL button during the
power up cycle, as doing so will cause the instrument to enter the diagnostic program. Refer to Section 7 for more information. Also, do not press and
hold the
cslibration may be compromised if the CAL switch
is in the unlocked position.
- key during power up as instrument
In this instance, the software revision level is Dl, but
your particulsx iostrmnent may be different. In any case,
the software revision level should be recorded in case
it becomes necessary to replace one of the ROMs in the
future.
Next, the programmed rimary address will be
3.
P displayed as in the examp e below:
IEEE ADDRESS 15
In this example, the factory default
15 is being &played. The adual disp
rimary address of
%
yed address will,
of course, depend on the programmed value.
Following these display messages, the unit will begin
4,
normal operation in accordance with the power up configuration discussed in the following paragraph.
2.4.5 Power Up Configuration
After the self testing and power up display messages are
completed, the Model 590 will assume specific operating
modes. The exact configuration is taken from save/recall
position 1. Table 2-3 summarizes the factory default iconfiguration for the unit. Note that many of these may be
different if you modify save/recall state 1. See paragraph
3.17 for more details.
2-16
GETTING STARTED
Table 2-3. Power Up Default Conditions
Mode
Range
Flequency”
Model
Filter
Rate
zero
Trigger Mode
Trigger Source
Bias Source
Bias Waveform
Start Time
Stop Time
Step Tie
First Bias
Last Bias
Bias Step
Default Bias
Count (#readings DC or
external)
Plotter Grid Type**
Plotter Pen Type**
plotter Line Type**
Plotter Label Type**
Plot Type**
Buffer to Plot**
XY s=*g**
IEEE Primary Address**
Cable #**
Condition
2nF
1oOkHZ
Parallel
On
10 readingslsec
Off
Sweep
Front Panel (MANUAL)
lmsec
ov
t;
ov
450
0 (Full Grid)
1 (Fen #l)
7 (Solid Line)
0 (ml Labels)
0 (C vs v)
0 (Buffer A)
Off
15
7 (to front panel)
2.4.6 Warm Up Period
The Model 590 can be used immediately when it is first
turned on. Note, however, that the unit must be allowed
to warm up for at least one hour to achieve rated accumcy.
Note, however, that you must use the CAL key to obtam
rated accomcy if the ambient temperature changes by more
than 2%
2.5 BASIC MEASUREMENT TECHNIQUES
The following paragraphs wilI take you through simple
step-by-step procedures to take one-point measurements,
obtain simple CV lots, and
measurements. TK s
ese proce ores are intended only to
erform fundamental C vs t
serve as a starting point, and they may not serve your
specific needs. Refer to Section 3 for detailed information
on making these type of measurements.
2.5.1 Test Connections
Use the basic test connections shown in Figure 2-3 for the
examples in tbis section. Paragraph 3.2 covers connecting
methods in more detail.
“59011OOk or 590/1OOk/lM units
NOTE: This configuration can be altered with SAVE 1 except **. To restore tbis configuration, use RECALL 0.
2-l 7
GETTING STARTED
PROBER SHIELDS-
CONNECTED TO
CABLE SHIELDS
___--------
_---
A. CONNECTIONS
---
--
COAXIALCABLE
2-18
S. SIMPLIFIED EQUIVALENT CIRCUIT
Figure 2-3. Typical Test Connections
Table 2-4. Initial Control Settings for One Point Measurements
GETTING STARTED
Mode
Range
~~d-yY
Reading Rate
Trigger
Analog Filter FILTER
Zero Baseline
Bias waveform
First Bias PARAMETER -20v to +2ov
Bias Voltage Status ON
Control Setting
Comments
RANGE As required Use most sensitive range possible
ZEL
1ookHz or lMH.2
Parallel Unit always measures using parallel model.
1 reading per second
Use cable correction at IMHz.
Maximum resolution/minimum noise.
%TE l-shot Allows display to freeze single measurement.
On Minimize noise
ZERO Off
Use only to subtract baseline.
WAVEFORM DC Static bias level for one-point measurement.
Select desired bias level.
On Enable bias voltage before measuring.
2.5.2 Basic One-Point Measurements parallel form, and the resulting data is internally converted
to serial form when that model is selected. The analog out-
Ordinarily, the Model 590 would be used to take a number
of readings with the resulting data plotted as a group of
points. In some instances however, you may wish to take
a singIg reading with or without a specific bii voltage and
display the result. Table 2-4 .mmnmizes recommended con-
trol settings for basic single-point measurements. Use the
&& proEe&,$ ~~~~f6~~~~~s~~l~-~o-m~~~~~~ on
put ‘w9’s refleas parallel mode1.
step 4: Se1ea a Readmg Rate
Since speed is not generally a requirement for single-point
~.~~~p+men+~
-------I ,
VW wordd~ p?&ly~~~se a rea&ngrafe of
one per second for maximum resolution and minimum
noise.
Step 1: Select Test Frequency
If your unit is equipped with both 1OOkHz and IMHz
modules (see rear panel), you can select the test frequency by pressing the FREQ button. Measurements made
through cables, at lMHz should.,use cable cow&on, as
discussed in paragraph 3.21. Cable correction is not
Step 5: Select the Trigger Mode
To display a single reading place the ins&m-tent in the one-
shot ixigger mode by pressing the TRJGGER MODE key
repeatedly until the I-SHOT message is displayed. Press
ENTER to select the new trigger mode.
necessary when the device under test is connected directly
to the front panel test jacks.
Step 2: Select a Range
Use the RANGE button to select a range consistent with
the anticipated measurement, or use autoranging, if and dataenq keys. Theprogr
desired. For best accuracy, select the most sensitive range
possible for the expected capacitance and conductance
readings.
Step 3: Select Parallel or Series Model
The test circuit can be modeled either as a parallel conduc-
tame and capacitance, or as a series resistance and
capacitance. You can select the display model with the
MODEL key. Note that the insfnunent always measures in
Step 6: Program the DC Bias Source
If you intend to ap~lv a DC biis voltaee to vour test cir-
cuit. use the WAVEF
r-d ~~~~ ~~~~
m~~:ORM key to select abCb&s waveform,
then program the first bias voltage with the Pm
nd bias source is f2OV. Before measuring,
source with the BIAS ON key, unless you are not using the
bias voltage, in which case you should leave it turned off.
Step 7: Bigger a Reading
Press the TRIGGER MANUAL key to trigger and display
a single set of capacitance, conductance, and biis voltage
readings. The reading set will remain on the display until
you press MANUAL again to trigger a new set of readings.
amma.bIe range of the inter-
turn on the bias
2-19
GETTING STARTED
Table 2-5. Initial Control Settings for Plotting
Mode
Range
poydw
1 control
1 RANGE
1 setting 1 comments
1 As rem&d 1 Use most sensitive
loo& or llvmz
Parallel
Reading Rate
Analog Filter
Zero Baseline
Trigger
Bll Waveform
Bii Start Time
Bias stop Time
Bii Step Time
First Bias
Last Bias
Step Biis
Default Bias
RATE
FILTER
ZERO
MODE
WAVEFORM
PARAMETER
PARAMETER
10 per second
Off
Off
Sweep
Single staircase
lmsec
lmsec
PARAMETER 10msec
PARAMETER -5V
PARAMETER +5V
PARAMETER O.lV
1 i%&&viETERI OV
*These values depend on requised bias parameters.
2.5.3 Basic Plotting Techniques
Use the basic procedure below to take a set of data points
and gaph the results on a plotter. Table 2-5 summarizes
control settings for a basic CV plot. This method is usable
only with a digitsl plotter. CV plots can also be obtained
by using an X-Y recorder with the analog outputs (see
paragraph 3.22).
ranee uossible.
Use cable correction at”*.
Unit measures pamIle model.
Best speed-resolution compromise.
Not necessary unless readings are noisy.
Use only to subtract baseline.
One complete reading sweep.
Often used waveform.
*
f
I *
Step 3: Select the Sweep Trigger Mode
Using the MODE key, select the sweep tigger mode. This
mode will allow you to take one complete set of data points
for later plotting.
Step 4: Select the Bias Waveform
Step I: Connect the Plotter
Connect an Hl’747OA plotter (or any other similar plotter
using HPGL) to the instrument with a suitable IEEE-488
cable.
NOTE
When performing stand-alone plotting, the plotter must be in the addressable mode using a
primary address of 5. Also, disconnect the controller from the bus when plotting eoom the front
pd.
Step 2: Select Control Fumtions
Using the appropriate front panel controls, select the range,
test frequency, and model. Select a 10 reading per second
rate with the RATE key.
U&the WAVEFORM key to select the required type of bias
waveform. Typically, you will probably use either the
single or dual staircase waveforms.
Step 5: Program Waveform Parameters
For most waveform types, you can program start and stop
hold times; step delay times; and start, stop, and default
voltage levels. Each of these parameters can be programm-
-- -c “--~?AR4h@,TER and data entry keys.
ed~ through .-
UJS “I ULS 1
Recommended values for the purposes of this demons&ation are listed in Table 2-5. After programming these
parameters, make sure the biis voltage is turned on.
Step 6: Trigger a Readiig Sweep
With your circuit connected to the test jacks, press the
MANUAL button to trigger a reading sweep. The instrument will cycle through the programmed bias steps,
measure the capacitance, conductance, and actual bias
voltage values, and store the data in the A/D buffer.
2-20
GETTING STARTED
I
.-
Step 7: Place the Data in the Plot Buffer
Before data can be plotted, it should be transferred from
the A/D buffer to the plot buffer. To do so, preSS the A
- B (SHIFJI BUFFER) button. Once the transfer is com-
plete, you can tri
writing your old 3
necessary (since you can plot directly from buffer A), but
it is a good idea to transfer data to avoid possibly overwriting it.
Step 8: View the Data
Jf desired, you can view data points before plotting by
pressing BUFFER. Select buffer B (if transferred in step 7),
then use A and V to scroll through data points. press
QUlT to cancel buffer access.
Step 9: Set Up Plotter Parameters
Use the SETUP key to select the following plotter
param@ers: grid type, pen type, line type, label type, plot
type, buffer, and XY scaling. Table 2-6 lists recommended
settings for simple plots. Use the numb
the appropriate parameter, then press
ger a new reading sweep without over-
ata. Note that this step is not absolutely
Table 2-6. lnitial Plotter Set Up
Description
Grid Type
Pen Type
Line Type
Label type
Plot Type
Buffer
X Scale
Y Scale
*Use SEXUP or A I V to scroll through modes.
**Press number key then ENTER to program value.
Step_ 10: Plot the Data
To plot your data, press the PLOT key. Data previously
placed in the plot buffer will then be graphed. Figure Z-4
sb.ows an example of a graph made in this manner. To stop
plotting, press the ABoRT key.
Step 11: Draw the Grid
0
1
0
0
Full Grid
Pen #l
Dash-dot
FUJI Labels
cvsv
Buffer B
I
Make sure that you have paper and proper pens installed
in the plotter, then press GRID. The instrument will then
command the plotter to draw the grid using previously
selected setup parameters.
2-21
GETTING STARTED
2-22
II-.01 X W/k%‘z!) 33NVlIWdV3
Figure 2-4. Plotting Example
GETTING STARTED
DEFINITIONS:
ts,m=START TIME (PROGRAMMED)
tm.=STOP TIME (PROGRAMMED)
b=STEP TIME (PROGRAMMED)
tr=READING INTERVAL=tm,+l/R
lR=READING RATE)
TIME COMPUTATION:
t~=tmrr+(tsmr+llRl B
WHERE: B=SUFFER LOCATION #
t=TIME AT BUFFER LOCATION
A. CAPACITANCE
START
SWEEP
TRIGGER
8. BIAS VOLTAGE
Figure 2-5. C vs t Waveform
2-23
GETTING STARTED
Table 2-7. Basic Settings for C vs t Measurements
IvIoae
Range
porgy
Reading Rate
Analog Filter
zero
Trigger
Waveform
Start Time
Stop Tie
Step Tie
Default Bias
First Bias
count
cunrrv, mxullg
RANGE
l+=Q
As required
ykJ& or lMHL
MODEL
RATE lOOO/sec
FILTER Off
ZERO Off
MODE Sweep
WAVEFORM DC
PARAMETER lmsec
PARAMETER lmsec
PARAMETER 10msec
PARAMETER OV
PARAMETER +5v
PARAMETER 100
C”UWLkZ‘lW
Do not overrange reading.
Use cable correction at lMHs.
Unit measures parallel model.
Fastest rate.
Filter increases response time.
Use only to subtract baseline.
One complete sweep.
Single level for C vs t.
Minimum start time.
Minimum stop time.
Nominal step time.
*
*
Select number of readings.
Yl-tese values depend on required bii levels.
2.5.4 Fundamental C vs t Measurements
Use the following basic setup procedure for simple C vs
t measurements. The procedure assumes that you have a
Step 4: Select a DC Bias Waveform
Press WAVEFORM repeatedly until you see the DC
waveform display message. Press ENTER to program the
test circuit already connected to the instrument. Table 2-7 waveform typesummarizes typical control settings for these measure-
ments. Figure 2-5 shows a typical C vs t waveform and also
defines certain terms. For complete details on C vs t Step 5: se1ed a Reading Rate
measurements, refer to paragraph 3.20.
Use the RATE key to program the desired reading rate:
1, 10, 75, or 1000 readings per second. Keep iu mind that
Step 1: Select Measurement Frequency
If your unit is equi ped to measure both at 1OOkHZ and
lMFJ.z, select the 2.
esrred frequency with the FREQ key.
~~~~ the interval between measurements is the sum of the
reading interval (reciprocal of the reading rate) and the pm
smmed step time. For the fastest possible C vs t measure-
gr
merits, select a rate-of 1000 readings per second .
oiep L: Select a Range
Use the RANGE key to select the desired measurement
range. Be sure to choose a range high enough to handle
the largest reading you expect. Autorangin is not recommended for C vs t measurements, espe
cialfy at the faster
reading rates.
CI_- r). TX-..-_- ‘L^ c _.._^ - .b-Lser Mode
Press the MODE key until the SWEEP trigger mode
~o~~sge is displayed. Press ENTER to program the sweep
2-24
Step 6: Program Bias Waveform Parameters
Using the PAR&EZER and data entry keys, program the
default bias, first bias, start; stop, and step times. Typically, the default bias is set to zero and the first bias is programmed to the amplitude of the pulse bias step, as shown
in Figue 2-5.
The progr-ed step time depends on the required time
interval between measurements (the total time interval is
the sum of the step time and the reciprocal of the reading
rate). For the fastest possible measurements, program a
minimum step time along with the 1000 reading per second rate;
Step 7: Trigger a Reading Sweep
Step 10: Display and Compute C vs t Information
Press MANUAL to trigger a re&ng sweep. The ins&u-~ :.cess C vs Ho display the reading buffer location numbers.
ment will then perform the sweep and make measurements at the programmed intervals. As measurements are
taken, readings will be placed in the A/D buffer for later
The buffer location will replace the bias voltage informa-
tion on the display. Use A/I to scroll through buffer loca-
tions and display location information at those points.
recall. Note that valid data will not be displayed until the
sweep has been completed at the 100 and 1000 reading per
second rates.
Cumulative time at a specific location can be computed as
follows:
Step 8: Transfer Buffer Contents
Press the A - B button (SHFT BUFFER) to place data just
taken into buffer B. Again, this step is not essential, but
it is recommended to avoid posslMe lost data.
Step 9: Access Buffer Data
Press BLFFER to access data taken durinzx the reading
sweep. Select buffer B then use A or V to &roll throu&
the various buffer locations. Note that data for each partic&x location includes a capacitance, conductance, and
bias voltage value (except fo;fhe 1000 reading per second
rate, which includes only capacitance data).
ta = 1.024 tshrt + WI24 tt, + 1 /R) (B)
Where: t, = time at a specific buffer location
t., = programmed St& time
Lp = pronammed step time
R = az reading rate‘
B = buffer location number
1.024 = multiplier to obtain actual times
Note that the actual (not~nominal) rates should be used
(see paragraph 3.9).
2-2512-26
SECT
ION 3
OPER,
3.1 INTRODUCTION
This section contains a complete, detailed desaiption of
each hont and rear panel aspect of the Model 590. The section is arranged as follows:
3.2 Display Messages: Lists display messages that may
be encountered during front panel operation of the
instrument.
3.3 Test Connections: Details operation of the test INPUT
and OUTPUT jacks on the front panel, and gives an example of typical test connections.
3.4 Readings and Hardware Control Aspects: Shows
how to titerpret both capacitance and conductance readings from the front panel display, and details some
aspects of hardware control.
tion, as well as use of the X10 attenuator tid optional
Model 5904 Adapter to extend the measurement~range
of the unit to 2OnF at 1OOkHz.
3.6 Frequency Selection: Details methods for 1oOkHz and
lh@lz test frequency selection, as well as some precau-
tions necessary when using each frequency.
3.7 Series/ Parallel Model: Describes parallel (G and C)
and series (C and R) model selection, discusses seriespamIle equivalents.
3.6 Filter: Covers enabling and disabling the single-pole
analog filter and gives a typical response curve.
3.9 Reading Rate: Describes selection of the 1, 10, ~100, and
1000 reading per second rates from the front panel.
4TION
3.13 Data Keys: Describes the operation of the numeric
keypad group for entering such parameters as bias
voltages and times.
3.14 Bias Voltage: Gives the basic procedure for progmmmiug bii waveforms, voltages, ar@ *es, as well as the
use of the rear panel external bias input and b&monitor
output jacks.
3.16 Buffer Operation: Outlines methods to access the two
450-word data buffers from the front panel, and how to
transfer the contents of buffer A to buffer B.
3.16 Plotting Data: Details use of an external intelligent
plotter to generate CV and other plots:
3.17 Save and Recall: Discusses procedures necessary to
save and recall instrument con6guratiow in NVRAM.
3.16 Self Test: Outlines the self test program that can help
determine if any internal problems are present.
3.19Mathematical Functions: Describes the many mathematical functions that can be used as an aid in analyzing
data located in one of the buffers.
3.20 C vs t Measurements: Details the procedure and
principles behind making capacitance versus time
measurements.
3.21 Cable Correction: Covers cable correction that Should
be used to optimize accuracy when making measurements at lMH2.
3.22 Analog Outputs: Details operation of the capacitance
and conductance analog outputs, and using an analog
plotter.
3.10 Zero: Gives the basic procedure for using z&o to store
a reading as a baseline value and then suppress that
value from subsequent readings.
3.11 Drift Correction: Covers use of the front panel CAL
key to perform drift correction using internal capacitance
reference standards.
3.12 Triggering: Details methods of selecting the trigger
source and mode and describes the operation of the rear
panel triggeT input and output jacks.
3.23 Measurement Considerations: Discusses some important considerations to take into account when making measurements with the Model 590.
3.2 DISPLAY MESSAGES
During Model 590 operation and programming, you will
encounter a number of messages on the front panel
display. Typical messages will be either of the informational or error variety, as discussed in the following
paragraphs.
3-1
OPER9TlON
Table 3-1. Error Messages
Message
-.--pF,-.---US,-.---~
OVFL
OVERLOAD
coNlwcT
INVALID
NEED 1OOkHz
NEEDlMHi
TRlGOVRRRUN
MULTiPLIERFAIL
Description
No valid reading available
J
C, G, or V ovenange
Module input overloaded
Mode selection conflict, or already plotting
Parameter iuvalid, or self test error
1OOkHz CV module not installed
lMHz CV module not installed
Unit triggered while processing reading or
sweep.
Self test indicates multiplier failure
3.2.1 Error Messages
Error messages are those messages which require some
form of Crective action on your part in order to properly
take a reading or program the instnxnent. For ekample,
the OWL message indicates that the capacitance or conductance value being measured is too high for the selected
measuring range.
Table 3-1 lists Model 590 error messages. Many bf these
messages are also covered in pertinent paragraphs of the
Cormxtive Action
Trigger reading
Move up range
Move up range
Do not use modes together
Program valid parameter
Do not select 1OOkHz
module
Do not select lMHz module
~ondil reading or sweep
See tro;bleshooting in
manual. Where applicable, the necessary corrective action
is also given in the table.
3.2.2 Informational Messages
lnfonnational messages are induded as an aid in programming the unit. No comective action is necessary in this case,
but you may still be required to enter a parameter at the
prompt. Table 3-2 lists Model 590 informational messages.
Again, most of these are covered in other parts of the
manual.
3-2
Table 3-2. informational Messages
OPERATION
Message
TRIGGER MODE l-SHOT
TRIGGER MODE SWEEP
TRIGGER SOURCE FP
TRIGGER J~OURCE EXT
TRIGGER SOURCE TALK
TRIGGER SOURCE GET
TRIGGER SOURCE X
BIAS WAVEFORM DC
BIAS WAVEFORM STAB7
BIAS WAVEFORM DSTAIR
BIAS WAVEFORM PULSE
BIAS WAVEFORM EXT
START TIME
STOP TIME
STEPTIME
1ST BIAS V
LAST BIAS V
STEP BIAS V
DEFAULT BIAS V
GRID TYPE O-l 0
PEN TYFE O-2
LINE TYPE o-7
LABEL TYPE O-3 G
PLOT TYPE O-6 0
BUFFER O=A l=B 0
X SCALE N=O Y=l
Y SCALE N=O Y=l
UPDATE N=O Y=l
BUFFER A=0 B=l
IEEE ADDRESS
SETUP NUMBER?
CABLE NUMBER?
BUSY
READING RATE 10
SELF TEST
CALCULATJNG DATA
DISCONNECT
Bias step size of each bias increments
Bias voltage before and after sweep
Plotter grid type
plot pen number
Plotter line type
Plotter label type
plot type
ye&p;
Y axis scaling
Update parameter?
Buffer selection
Display IEEE primary address
Setup position to save or recall
Cable # to save, recall
Unit performing calibration
Display/program reading rate
Unit running self test
Unit computing at end of sweep.
1 Test voltages disconnected from test jacks.
mmand triggering
3.3 TEST CONNECTIONS
The following paragraphs discuss methods for m&ng the
test connections necessary to measure capacitance and conductance with the Model 590. Grounded and floating
operation of the test jacks are also covered.
3.3.1 BNC Test Jacks
Both test INPUT and OUTPUT are BNC jacks, as shown
in Figure 3-l. The center conductor is high, and the outer
Fg or shell of the jack (connected to analog common) is
low.
WARNING
The INPUT and OUTPUT jacks may be floated
up to 30V RMS above chassis ground when the
rear panel grounding switch is in the floating
oosition. Exceedina this value may create a
ihock hazard. -
3-3
OPERATION
I
,-HIGH
Figure 3-1. input and Output Jack Configuration
3.3.2 Typical Test Configuration
Use the test INPUT along with the test OUTPUT to make
measurements, as shown in the typical example of Figure
3-2. Figure 3-3 shows the equivalent circuit of the test
setup.
When making measurements, keep the following points
inmind:
1. Use only RG-58 type of coaxial cable for both OUTPUT
and INMJT. Maximum recommended cable length is
I
five meters. The Keithley Model 7051 cables can be used for connections.
2. When measuring through cables at lMH.z, ou should
use cable cone&ion to compensate for cab e transmis- P
sion line effects. Paragraph 3.21 covers cable correction
programming in detail.
3. The maximum common mode voltage for both the test
OUTPUT and INPUT is 30V RMS, 42.4V peak when the,
rear paiwl grounding switch is in the floating position.
Analog common caMot be floated above ground when
the switch is in the grounded position.
4. Excessive shunt capacitance in the cable or test fixture.
may de
noise. ET3
noise figures.
ade accuracy of the measurement and increase
nsult the specifications for degradation and’
OPERATION
MODEL 590
4
RG-58 -
4XIAL CABLE
PROBER SHIELDS
CONNECTED TO
CABLE SHIELDS
r---------~---~
r----~
L---------I
i----~-~--T---l,X~*~C~~SS,S
IEST CONNECTION PROCEDURE:
,. Connect the RG-58 BNC cable between the test OUT-
PUT and the test input terminal of the test fixture (not
supplied).
!. Connect a second RG-58 BNC cable between test n\T-
PUT and the output of the test fixture.
;. Select grounded or floating operation with rear panel
switch. (WARNING: maximum common mode
voltage is 30V RMSL
RG-58 -
COAXIAL CABLE
----
=J , t ;
-
‘/=-I
l
I
FARADAY SHIELD (OPTIONAL)
4. Make sure probe shields are carried through as close
to the wafer as possible; These shields should be con-
necked to ~&e BNC connecting cable shields.
5. A faraday shield, surrounding the wafer and chuck,
incorporated into the test fixture, may be necessary
to minimize noise. This shield must be insulated from
the prober chassis and connected to analog common
via the BNC cable shields.
&The prober chassis should be connected to earth
ground as indicated above.
Figure 3-2. Typical Test Connections
3-5
OPERATION
____------
MODEL 590
Figure 3-3. Equivalent Circuit of Test Connections
3.3.3 Grounded and Floating Operation
The outer rings of the TEST INPUT and OLJTPUT jacks
are connected to analog common, which can either be con-~
netted to chassis ground or floated up to 30V R$fS above
ground potential.
WARNING
Do not exceed 30V RMS, 42.4V peak common
mode voltage, or a possible shock hazard may
result.
To selectgrounded or floating operation, simply place the
rear panel gro@ing switch in the appropriate position,
as shown in Figure 3-4. Note that the rear panel BLAS and
analog outputs will also be affected by this switch.
Grbuixded operation can be used in cases where it~is not
necessary to float analog common or if noise caused by
ground loops is not a problem. If analog co-on must
be floated above chassis ground potential, or if ground loop
problems occur (as may happen if other, grounded insinunents are connected to the test fixture), the instrument
should be operated with analog common floating. See
paragraph 323.1 for a detailed discussion of ground loops.
A. ANALOG COMMON FLDAnNG
I
R ANALOG COMMON GROUNDED
I
Figure 3-4. Floating/Grounded Operation of Analog
Common
I
I
3-6
OPERATION
CAPACITANCE
I
--
CONDUCTAN&? (PARALLEL1
, RESISTANY ISERIES),
I
--
Figure 3-5. Capacitance, Conductance (Resistance), and Bias Voltage Displays
3.4 READINGS AND HARDWARE CONTROL
The following paragraphs discuss capacitance and conductame readings and some hardware control notes.
BIAS
VOUAGE
,
--
,
,
--
tame, replacing the previously used mho). The display will
show conductance in S or mS when uarallel model is
selected. Note, however, that the &tit will display
resistance 5 this position when series model is selected,
& discussed in paragraph 3.7.
3.4.1 Capacitance and Conductance Displays
Note that conductance is simulv the recimocal of resistance
and is calculated as follows! ’
Capacitance and conductance readings 5ve shown on the
front panel display, as shown in Figure 3-5. The
‘capacitance reading appears in the left portion of the
display, and the conductance reading appears in the right
G + jwC =
1
R + JwC
portion of the display. Both readings are a 3%digit or
4%digit signed value, depending on the selected reading
rate. Where: G is the conductance in s&ens
R is the resistance in ohms
C is~the parallel capacitance
The capacitance and conductance displays will show either
C’ is the series capacitance.
the current reading, or a reading from one of the boffers,
depending on the selected mode. During normal operation, the current reading will be displayed; however,
NOTES:
capacitance and conductance readings from buffer A or
buffex B will be displayed when you select that option with
the buffer key.
1. The display will show dashes in place of numeric values
if no valid reading is available. To d&play readings, trigger the unit with an appropriate triggers stimulus, as
determined by the programmed trigger source.
The capacitance reading includes capacitance engineering
units in pF or nF. 1pF equals 10~‘* farads, while 1nF is lo+
farads. Conductance readings are in units of siemens (the
siemen is the internationally recognized unit of conduc-
2. The update rate of the displays in the sweep trigger
mode depends on the relative reading rate selected with
the RAE key. IvEASuRE indicates the relative -a&g
rate. Only one reading set per trigger stimulus will be
displayed in the one-shot tigger mode.
1
3-7
OPERATION
3. No valid data will be displayed until a sweep is
completed at the 75 and 1000 reading per second rates.
4. When BUFFER is on, the unit is displaying buffer location data instead of the current reading.
5. Display resolution is 4% digits (+20,000 counts
nominal) at 1 and 10 readings per second, and 3% digits
(+2,C!OO counts nominal) at the 75 and 1000 reading
per second rates.
6. The actual display count limits are +21,999, -19,999
counts. Note that accuracy above +ZO,ooO counts is
typical.
7. Shunt loading and cable correction reduce the dynamic
range of capacitance and conductance measurements.
3.4.2 Bias Voltage Display
The bias voltage display is located at the right of the front
panel, as shown in Figure 3-5. Depending on several factors, this display will show one of the following:
1. The current bias voltage: During normal operation, the
unit measxxes the adual bias voltage applied to the circuit under test through the test OUTPLJT jack. If the
internal bias source is selected, the display will show
the actual bias voltage at that particular waveform step.
If external bias is selected, the unit will measure and
display that voltage.
2. A buffer bias voltage value: When accessing buffer inf-tion (with the BUFFER key), the display will show
the voltage bias step that was applied to the test circuit
at that particular point in time.
3. Buffer location: When displaying C vs t information,
this display will show a particular buffer location
number. Time information can be computed from the
display as discussed in paragraph 3.21.
3.4.3 Hardware Control Considerations
a
Those keys which generally affect hardware operation include: RANGE, FILTER, ZERO, FREQ, RATE, and CAL.
When using these keys, keep in mind the following point%
1. Changing one of these modes will abort an active sweep.
2. The A/D buffer pointer will be reset and data will be
cleared from the A/D buffer (buffer A).
3.5 RANGE SELECTION
The following paragraphs discuss manual and auto range
selection, as well as the use of the X10 attenuator with optional in ut transformer to extend the measurement range
of the 1 Fil kHz module to 2OnF.
CONTROL
NOTES:
1. The voltage display will show dashes when no valid
reading is available.
2. The voltage dis
counts nominal
rates, and 3% digits (+2,ooO counts nominal) at the
and loo0 per second rates. The display update rate
depends on the reading rate (the relative reading rate
is indicated by MEASURE). Note however, that no
voltage data will be dis
when the 75 or 100 g .
selected.
3. The a&al display count limitation is +21,999, - 19,999
counts. Accuracy above +20,000 counts is typical.
4. The bias voltage must be turned on in order to read the
bias voltage.
3-8
lay resolution is 4% digits (*ZO,OOO
at the 1 and 10 reading per second
P
layed during a reading sweep
readmg per second rate IS
75
3.5.1 Available Ranges
The available ranges depend on the measurement frequency, e summarized in Table 3-3. Note that the 2OnF range
is not available when measuring at lMKz. The optional
Model 5904 input transformer must be used in conjunc-
tion with the X10 attenuator to extend the 1OOkHz measurement range to 2OnF. Also, there is no 2pF, lMH.z range.
Table 3-3 al+o shows full scale displayed values for.each
range. These values show 4%-d@ resolution, whxh 1s
available only at the 1, 10, and 18 per second reading rates
(the 18lsec rate is available only over the IEEE-488 bus).
OVERLOAD Module input overload Move up range or appCONFLICT X10 attenuator csnnot be Do not use conflicting
3.5.2 Invalid Reading Indications
Basically, there are two conditions that may cause an invalid reading indication. Fit, either the capacitance or
conductance’ reading (or possibly both) may exceed the
count capability of the associated display area. In this case,
the display for that parameter will display the following
message:
OVFL
To correct this condition, select a higher range.
A more serious situation exists iu cases where the input~m~
amplifier of the CV module is saturating. In thiscase, the
unit will display the following error message: ~~,-
OVERLOAD
Module saturation means that the test signal current is tooo
high~for the test input amplifier. Under these circum-
Description / Corrective Action
I I
Capacitance or conductance Move up range or
reading overrange
used at lMHz
stances, neither the capacitance nor conductance reading
is valid due to the non-linear characteristics of the input
circuits when saturated.
To correct this error, move the instrument up mnge until
a valid reading is noted by the absence of error messages.
Table 34 summaria es error messages assodated with improper range selection.
NOTES:
1. If an overload -occurs, the unit will cease waveform
&An overload condition is not flagged at the analog out-
3. An overload situation could be caused by an extraneous
apply smaller C or G
ly smaller C or G
modes
and buffer activity.
puts. An on-range reading may occur at the analog outputs under overload conditions.
signal appearing on the test INPUT jack. This signal
could come from external RFI or EMI sources not
associated with the 1OOkH.z or IMHz test frequencies,
or the DC bias voltage.
3-9
OPERATlON
3.5.3 Manual Range Selection
To select ranges manually, simply press the EANGE but-
ton briefly ( <‘/r second) to move the instrument up range.
Each time you 8 ress RANGE, the instrument will move
up one range. nce the highest range is reached, the unit
vdl switch to the lowest range the next time you press
RANGE briefly.
Pressing RANGE briefl
mode is Presently enab ed. Jn this case, the unit will stay
on the presently selected range.
NOTES:
1. Better overall accuracy and resolution can be obtained
by using the lowest range possible for the measured
capacitance and conductance.
2. Since capacitance and conductance ranges are paired
together, it may be necessary to measure the capacitance
or conductance on a less than optimum range in order
to keep both readings on scale.
will also cancel autorange, if that
sr
3.5.4 Using Autoranging
The Model 590 has a convenient autoranging feature which
simplifies range selection. To enable autoranging, simply
press and hold the RANGE button for more than one-half
second. The instrument will then go into the autorange
mode, as indictated by the AUTO LED. To cancel autoranging, briefly press the RANGE button a second time. The
insirument will then stay on the presently selected range.
4. Autoranging cannot be used with the 75 and 1,000 per
second reading rates. The unit will generate ~a CONFLICT error under these conditions.
3.5.5 Using the 20nF/20mS Range
By using the internal X10 attenuator in conjunction with
the optional Model 5904 2Ofi12OmS Adapter, the 1OOkHz
measurement ran e of the Model 590 can be extended to
2013, as describe ~below.
X10 Attenuator
Use the procedure below to enable the X10 attenuator.
1. Connect the Model 5904 2Oti/2OmS Adapter to the test
jacks (see below).
2. Using the RANGE key, place the instrument on the 2nF
(highest range).
3. Select a measurement frequency of 1OOkHz with the
=Q key.
4. Press SHIFT RANGE to enable the attenuator. The X10
LED next to the RANGE key will illuminate to indicate
that the instrument is in that mode.
5. Take the readings from the display. The instrument will
automatically scale the readings and display the proper
values.
6. To disable the X10 attenuator, press SHIFT RANGE a
second time.
NOTES:
%
Keep in mind that autoranging is included for convenience
only and should not be used for critical measurements
because of possible effects on the readings.
Figure 3-6 shows a flow chart of autoranging operation.
NOTES:
1. Accuracy with other ran es than 20nF with the X10
adapter is not specified.
innn,$ the Model 5904 X10 adapter is not recom-
2. When taking data with rapidly changing bias waveforms, manual rangin should be used to ensure con-
sistent timing for ea ci! pomt. Measurement time can
vary widely during autoranging.
3. The instrument will not autorange into the X10 mode.
3-10
d erefore, the useof autorang-
1. The X10 attenuator is intended for use with the optional
Model 5904 2O.eF/2OmS Adapter (see below). Since the
instrument has no way of sensing if the adapter is connected, incorrect readings will result if you enable the
X10 attenuator without connecting the adapter, or use
the input adapter without the X10 attenuator enabled.
2. The X10 attenuator may be used with other ranges, if
desired, but accuracy for those ranges is not specified.
The instrument will automatically scale the reading to
reflect the X10 attenuation factor. In this situation, the
~available ranges will be 20pF/20pS, 2OOpF/200pS,
2nFQmS, and 2OnF/2OmS. The tit can be o&rated for
Model 5905 used on the 2OpF through 2nF ranges, ii
desired. See paragraph 7.3.
3. The X10 attenuator is not available for use at lMHz. The
following message will be displayed if you attempt to
enable the attenuator with a lMH!z test frequency
selected, or if you attempt to enable to select a lMHr
test frequency with the X10 attenuator enabled:
CONFLICT
DWIAY
ERROR
MESSAOE
END
d5
OPERATION
Figure 3-6. Flow Chart of Autoranging Operation
4. The Model 5904 must be calibrated with a particular
Model 590 to achieve stated front panel accuracy.
Input Transformer Connections
Figure 3-7 shows typical connections when &ng the
1. Use only RGSS type coaxial cable to make the test connections. Maximum recommended cable length is five
meters.
2. The maz&u.un common-mode voltage for tloating operation is 3OV RMS. 42.4V oeak.
3. Excessive shunt ~pacitake will degrade accuracy and
increase noise.
3-1;
OPERATION
--
RG-58
OAXIAL CABLE
--I
PROBER SHIELDS
CONNECTED TO
CABLE SHIELDS
L!
-- ,- -~~ --~
MODEL 590
r--
I
/
- -.-~-~-~-~.- - - _
r---
-----1
PROBERS
I 7
: ===s\ +!===j. 1 ’
WAFER 4
I
I cHucK-I
L---
-~--t---T---’
TEST CONNBCTION PROCEDURE:
1. Mount the Model 5904 adapter on the test OUTPUT
and INPUT jacks. Be sure not to install the adapter
upside down.
2. Connect an RG-58 BNC cable between the test output of the adapter and the test input of the test t%ture.
3. Connect a second RG-58 BNC cable between the test
fixture output and the test input of the adapter
mounted on the instrument.
4. Selects grounded or floating operating with the rear
panel switch. (WARMNG: maximum common mode
voltage is 3OV RMS.)
1 II
WAFER
I
l
I
Jj
1
FARADAY SHIELD (OPTIONAL1
h
L PROBER TEST FIXTURE CHASSIS
5. Probe shields, connected to cable shields, should be
carried through as close to the wafer as possible.
6. A faraday shield may be necessary to minimme noise.
This shield must be insulated from the prober chassis
and connected to analog common via the cable
shields.
7.The prober shield should be connected to earth
ground as indicated above.
NOTE
Enable the X10 attenuator (SHB?T RANGE)
when using the input adapter or else reading
scaling will be off by a factor of 10.
3-12
Figure 3-7. Typical Test Connections Using Model 5904 20pFl20mS Adapter
3.6.2 Test Voltages
OPERATION
An intend signal somce supplies a 1OOkHz or lMH2,
15mV RMS test voltage. Available frequencies as well as
the frequency selection procedure are covered below.
CONTROL
RANGE
FREQ
RATE
q lOOK
•~MHZ
MoDEL urn
ol,,,
The nominal test outputLvoltage for both 1OOkHz and
lMHz is 15mV RhE, with a tolerance of f 10%. The fre-
quzmm~~;~~pf~ both the 1OOkHz and lhG+ test
3.6.3 Selecting a Frequency
If your instrument is equipped for 1OOkH.z and lMH.z
operation, you can select the desired operating frequency
simply by pressing the FREQ button. Doing so will cause
the unit to change to the other frequency, as indicated by
the associated LED. Pressing FREQ again will sele~ct the
disconnect~mode, as discussed in paragraph 3.6.5.
Figure 3-8 shows a flowchart outlining frequency selection.
NOTES:
1. Cable correction should be used when measuring
through cables at lMH.z. See paragraph 3.21 for more
information.
2. The 2pF rtige is not available at IMHz.
There are three available models of the 590. The Models
59011oOk and 5901lM supply test voltages at frequencies
of 1oOkHz and lMHz res
will operate at either 1 &
maizes the available models, installed modules, and test
frequencies. Available test frequencies are marked on the
rear panel.
3.6.4 Frequency Error Messages nored, but the second trigger will cause a trigger over-
ruu condition.
Error messages associated with frequency selection are
summarized in Table 34.
3. No valid data will be stored or be made available over
lhe bus while in the disconnect mode.
Table 3-6. Frequency Error Messages
Message Description
NEED loOk 1OOkHz module not installed
NFED lM IMHz module not installed
CONFLICT X10 attenuator enabled when selecting
lMHz, or CAL or CABLE CAL pressed
in disconnect
3.6.5 Disconnecting the Test Voltage
A second feature of the FREQ key allows you to disconnect the test and bias voltages from the device under test
without havin to remove the connectin cables attached
to the test EW
voltages simply press FREQ until the DISCONNECT
message is displayed. Internal relays will then disconnect
the test and bias voltages from the front panel test INlWT
and OLJTPUT jacks, allowing the center conductors of
these jacks to float. In addition to the DISCONNECT
messages, both the lOOkJ3z and lMHz LEDs will turn off
while the unit is in the disconnect mode. To return to nor-
mal operation, simply press FREQ again.
NOTES:
and OUTPUT jacks.
5JT
+ . o &sconnect the
3.7 SERIES/PARALLEL MODEL
The following paragraphs cover measurement model, how
to select the model, and discuss series-parallel eqoivslents.
I
CONTROL
3.7.1 Measurement Model
1. Calibration, pressing, CAL, or CABLE CAL are illegal
when the unit is in disconnect; the instrument will
The Model 590 measures the capacitance, C, and the con-
ductance, G, of an equivalent parallel circuit connected be-
display a CONFLICT message under these conditions. tween the test OUTPUT and test INPUT jacks. Figure 3-9
2. The fir.4 trigger received while in disconnect will be ig-
--------I
vv
-__---~---
ANALOG COMMON
shows an equivalent circuit of the test configuration.
f) OUTPUT
-I
Figure 3-9. Equivalent Circuit of Parallel Capacitance and Conductance
3-14
OPERATlON
3.7.2 Model Selection
To alternate between series and parallel models, press the
MODEL key on the front panel. One of the associated
LEDs will indicate whether series or parallel model is in
effect. When the series model is III effect, the unit
mathematically converts parallel measured data to serial
form (data is always stored internally in parallel form).
NOTES:
1. Buffer data is always stored in parallel form. Series con-
version is performed when buffer data is displayed, if
that model is selected.
2. The analog outputs are always in parallel form.
3.7.3 Conductance and Resistance Ranges
With a psmllel model, the unit displays conductance. With
a series model, however, the unit displays resistance.
Equivalent full range conductance and resistance ranges
for both frequencies are shown in Table 3-7.
3.7.4 Series and Parallel Equivalent Circuits
In a similar manner, the resistance and reactance of the
series form of (b) are represented by R and X, respective-
ly. The impedance of the series &cuit is 2.
G
Y=G+JB
6 = oC, ICAPACITIVEI
OR
B = 1 IINDUCTIVE)
&
iA1 PARALLEL CIRCUIT
Z=R+JX
X = 1 (CAPACITIVE)
z
OR
X = w4 IINDUCTIVE)
(61 SERIES CIRCUIT
Figure 3-10. Series and Parallel Impedances
The net impedances of the equivalent series and parallel
tits at a given frequency are equal. However, the indi-
vidual components are not. We can demonstrate this relationship mathematically as follows:
A complex impedance can be represented by a simple
series or parallel equivalent circuit made up of a single
resistive element and a single reactive element, as shown
in Figure 3-10. In the parallel form of (a), the resistive element is represented as the conductance, G, while the reactance is represented by the susceptance, 8. The two
together mathematically combine to give the admittance,
Y, which is simply the reciprocal of the circuit impedance.
Table 3-7. Resistance and Conductance Ranges
100kHz
Parallel Range
2pF/2pS 2pF/2Ma
2OpF/20fiS
2 2 F/20W1
E~ops
2% 232ko 2n%2Oms
2OnF/2oms*
Series Range Par+llel Range
2OpF/200kQ
2OnF/2OOn*
1
R+JX=
G + jB
To eliminate the imaginary form in the denominator of the
right-hand term, we can multiply both the denominator
and numerator by the conjugate sofa the denominator as
follows:
1
R + jX =
G + jB
G - jB
x
G - jB
lhmz
Series Range
2OpF/200kS
20 F/2&
2OpF/2OOkQ
&O$$;kQ
*5904 and X10 attenuator required.
3-15
OPERATION
Performing the multiplication and combining terms, tie:For the series circuit, the dissipation factor is defined as:
have:
G - jB
R+jX=
G= + B=
By using the dissi
If we assume the reactance is capacitive, we can substitute
- l/oC, fci? the reactance and UC, for the susceptance (C,
is the equivalent series capacitance, and C, is the equivalent
parallel capacitance). The above equation then becomes:
R-j ~1 G - joC,
Jn a lossless circuit (R and G both 0), C, and C, would be and 3Ofi respectively. From these values, we can calculate
equal. A practical circuit, however, does have loss because
of the finite values of R or G. Thus, C, and C, are not
equal-the greater the circuit loss, the larger the disparity 30 x 10-c
between these two values.
Series and parallel capacitance values can be converted to
their equivalent forms by taking into account a dissipation
factor, D. D is simply the reciprocal of the Q of the circuit.
For a parallel circuit, the dissipation factor is:~
-=
WC
G= + o=Cp=
summarized in Ta
to another. Note that C, and C, are virtually identical for
very small values of D. For example, if D is 0.01 C and
C, are within 0.01% of one another.
Example:
Assume that we make a measurement on a partid
equivalent circuit and obtain values for C, and G of 16opF
the dissipation factor, D, as follows:
The eouivalent series cw&tance is then calculated as
follows’:
The analog filter can be used to minimize the amounts of
noise appearing in the displayed readings, and at the
anal0
trol 0 f
also given.
outputs. The following paragraphs describe con-
the analog titer; a typxal filter response curve is
CONTROL
RANGE
FREQ
q lOOK
0 0
01~~2
FILTER
RATE CAL~P
ZERO
0
MoDEL UC-J
0
L-
a*+
NOTES:
1. The analog filter increases instrument response time to
changes in input signals for the conductance and
capacitance readings. Thus, inaccurate readings may
result if the filter is used while measuring with rapidly
changing bias waveforms. Table 3-9 summarizes
nominal response times to various percentage of final
values.
2. The effects oft the analog filter are reflected at the
capacitance and conductance analog outputs on the rear
panel of the instrument.
3. The malog filter has no effect on the voltage bias
reading.
4. Pressing FILTER will abort a sweep.
Table 3-8. Typical Filter Response Times
Percent of Fii Reading 1 Typical Filter Response
10%
1%
0.1%
0.01%
1Omsec
2Omsec
3Omsec
4Omsec
3.8.2 Typical Filter Response
3.8.1 Filter Control
A typical response curve for the single-pole analog low-
To &able 01 &able the m&g filter, simply press fie pass filter is shown in Figure 3-11. Note that~ the filter
FlLTER key on the front panel. The on/off status of the
filter is indicated by the associated LED.
0
-10
E
=z
2 0
z
-20
-30
10 100 1000 loo00
FREQUENCY IHZ)
respmse rolls off at 6dB per octave (2OdB per decade) above
the -3dB point of approximately 37Hz.
Figure 3-11. Typical Analog Filter Response
3-17
OPERATlOM
3.8.3 Using the Filter
Noise in the reading is usually seen as an unsteady display
value that jumps around. In this situation, it is generally
benefidal to leave the filter enabled to stabilize the
readings. However, using the filter with rapidly changing
waveforms can degrade accuracy because of increased
response time, as indicated above.
CONTROL
If additional filterin
iucorporates digital
is require<, use a reading rate that
ki .
tenng, as dxussed in paragraph 3.9.
3.9 READING RATES
From the front panel, you can select reading rates of 1, 10,
75; or 1000 readings per second. A fifth rate, which is
available only over the’IE?X488 bus, produces 18 readings
per second with 4%-d@ display resolution.
NOTE
Because of the way the unit generates its time
base, the actual reading rates are slightly ti:~
ferent than indicated. Table 3-10 lists actual
intervals along with other pertinent informa-
tiOiL
3.9.1 Selecting a Reading Rate
Display or select the reading rate as follows:
1. Press the RATE key. The instrument will display the
reading rate now in effect. For example, the display
might show:
READING RATE 10
MoDEL cl@
0
In this case, a rate of 10 readings per second is in effect.
2. To return to normal operation without changing the
rate, press QUIT. The instrument will then return to
normal operation with the active sweep (if any)
unaffected.
3. To scroll through available reading rates, press and hold
~the RATE or AIT key, or press the numeric key.
associated with that rate, as summarized in Table 3-10.
4. When the desired rate is shown, press the ENTER key
to program the displayed reading rate. The instrument
will return to normal operation.
NOTES:
1. Re-progr
Any stored data from the previous sweep will be lost.
2. No valid data will be displayed until the sweep is completed and calculations are performed at the 75 and loo0
per second rates.
cl*+
amming the rate will abort an active sweep..
Table 3-10. Reading Rate Summary
Display
Readings ‘Resolution Period
3% digits’ l2OpeC
cc “G”‘yv
C: G: V
C, G V
*Data displayed after sweep is completed.
3-18
3% digits* 24Opec
41% digits 2.4msec
4% digits 16.7ms.e~
Integration Integrations
Averaged
1 ~~~~~~~~
1
i
Actual Readmg Effective
Interval (msec) Rate
1.024
13.3 75X/set
102.3
1024 0.977/set
976.56/&c
9.77/set
sweep (7
5 and 1006/set rates onl?), the fIont panel keys
will 6e in0 erative. The amount of time necessary for
cur
these cal
ations depends on the the number of data
uoints taken in the sweep. A CALCULATING DATA
&sage will be display&d.
4. Autoranging cannot be Fed at the 75~ 0: J,OO~&c
reading rates. The unit will di
if you attempt to program ert
lay a CONFLICT error
Ti
er of these rates with
autoranging enabled.
5. A CONFLICT error will o&u if you attempt to select
l- 75 kc rates with more than 450 readings per sweep
programmed.
3.9.2 Display Resolution
The display resolution for the 1 and 10 readingper second
rates is 41% digits. Display resolution at the 75 and 1000
er second rate is 3% digits. Note, however, that data will
t .
e &played only when the sweep is finished at the two
fastest rates.
3.9.3 Digital Filtering
OPERATION
at the expense of resolution. At the other extreme, you
would opt for the 1 reading per second rate in situations
requiring
rrxs4mu.m resolution and minimum noise.
3.9.5 1000 Reading Per Second Rate
~Consideraticns
When using the 1000 reading per second rate, the following points should be kept in mind.
1. Only capacitance data is taken; neither conductance nor
bias voltage data is taken.
2. Data is placed intq the buffer in raw form and will be
made available only when the reading sweep has been
completed.
3. The only available bii waveforms are DC and external.
A CONFLICT error will occur if you attempt to select
other waveforms.
4. MEASURE will be on during integration.
5,~Attempting to program control the Model 590 during the
sweep may destroy the timing integrity of the waveform.
6. Up to 1,350 readings can be stored in the buffer at the
10001sec rate.
to minim&x noise. A basic averaging scheme is used in
both cases, with 4 integrations averaged at the llsec rate,
and two integations averaged at the lO/sec rate. Since the
degree of filtering depends on the amount of averaging,
the best noise performance can be expected at the slowest
rate.
3.9.4 General Rate Selection Considerations
The primary factors affected by the reading rate (other than
the absolute number of readings Peru second) are the integration period and the amount of resolution. Since
reading noise is affected both by the integration period and
the amount of ~digital filtering, the reading will have $he
least noise at the 1 per second rate (longest integration
period and most digital f&ring), and the most n&se at
the 75 and 1000 per second rates (shortest integration
period and no digital filtering).
Optimum rate selection, then, for your particular application will depend on required resolution and speed, as well
as the amount of noise tolerable in the readings. For example,~ if speed is your primary requirement and you requjre capacitance, conductance, and bias voltage readings,
& tam C, G, and,V measureme@), but
3.9.6 1000 Readings Per Second Instrument
Settings
To obtain a rate of 1000 readings per second, you must se=~~
lect~the following instrument settings:
l Trigger mode: Sweep (paragraph 3.12)
l Trigger souye:~ internal (paragrap-t 3.12)
l Bias waveform: DC or external (paragraph 3.14.1)
l Step time: Ims (paragraph 3.14.2)
3.9.7Typical Reading Rates
Typical reading rates are
Internal Trigger Rate External Trigger Rate
Rate ImsDelay ZmsDelay lmsDelay 2msDela:
75 75
18
16.4
10 10
/ 1 1 1
Note: All rates include both start and stop times of lms.
summarized below.
64.6 68
16.1 16
9.5 9.6
1 0.97
63
16
9.5
3-19
OPERATION
3.10 USING ZERO
Zero can be used to store a set of capacitance and conductance readmgs as baseline values. Once stored, the baseline
values are then subtracted from subsequent readings.
CONTROL
FILTER
ZERO
1. Make sure that zero is disabled (ZERO off).
2. Use the FRRQ key select the test frequency.
3. Select the reading rate and range, as required.
4. Connect the capacitance and conductance values to the
instrument via the front-panel INPUT and OUTPUT
jacks. See paragraph 3.3 for detailed information on testy
connections.
5. Press ZERO to enable zero.
6. Trigger the reading or sweep by applying the appropriate trigger stimulus. From the front panel, you
can do so simply by pressing the MANUAL button. The
process of triggering the instrumentwiU store the firsts
set of capacitance and conductance readings as baseline
values.
7. Connect the test circuit to be measured to the test jacks.
Jfnecessa$ &i&e&e instiu%ent.-The stored baseline
values will be subtracted from the actual readings, and
the result will be displayed.
8. To disable zero, press ZERO a second time. Stored
baseline values will be lost once zero is disabled.
3.10.3 Using Zero to Optimize Instrument
Accuracy
The accuracy specifications given at the tiont of this manual
assume that the instruments has been properly zeroed us-
Zero is desaibed in the following paragraphs.
3.10.1 Enabling and Disabling Zero
To enable or disable zero, press the ZERO button; the in- lMHz modules, select the desired test trequency wrtn
dicator to the right of the ZERO key indicates the state of the FREQ key.
zero.
When zero is first enabled, the first capacitance and con-
ductance readings triggered after zero is enabled will be
stored as the baseline values,
from subsequent readings as long as zero is enabled. Theses
zero vahres will also be stored in the buffer header to be
used when accessing buffer data.
which will then ba
3 subtracted
3.10.2 Storing Capacitance and Conductance
Baseline Values
Use the following basic procedure to store capacitance and
conductance parameters ss your baseline.
ing the fundamental procedure below. Jn order to optimize
accuracy, it is recommended that you repeat the procedure
below every hour, especially in situations where the ambient temperature varies considerably.
1. Ifyour instrument is equipped with both lOOkHa and
2. Select the desired reading rate and measurement range
usine the RATE and RANGE keys.
3. Che& to see that zero is initiaUyWdisabled (ZERO off).
4. Connect the cables and test fixtore to the test INPUT
and OUTPUT jacks, but leave the DUT (device under
test) disconnected at~this time.
5. Press ZERO and trigger a reading or sweep by pressing
MANUAL. The capacitance and conductance displays
should then show a zero reading.
6. Corii%t the DUT to be measured to the fixture probes
(or other similar test fixture connections).
7. Readings using the zeroed value can now be triggered
and read in the normal manner. For optimum accuracy,
it is recommended that the instrument be m-zeroed using the above procedure whenever the range, reading
rate, or frequency is changed.
. . .
3-20
OPERATION
tiep~ the following points in mind when using zero:
1. Zero reduces the dynamic range of the displayed
readings. For example, if a 1OOpF baseline value is in
effect on the 200pF range, a capacitance of 1OOpF will
overrange the reading even though a 10OpF reading is
well within the limits of the 20OpF range.
2. Zero calculations are performed on a parallel or series
model, as appropriate.
3. Zero affects only displayed data; it does not modify the
way it is measured and stored within the insinxnent.
4. Zero is not affected by pressing other keys.
5. Changing range or frequency leaves the same zero
value active.
6. If the buffer is enabled (BUFFER KEY),~the zero values
will be obtained from the buffer header. Figure 3-12
shows a flow chart of zero operation both with and
without buffer operation.
7. The accuracy specifications listed at the front of this
manual assume that the instrument has been properly
zeroed using the procedure in paragraph 3.10.2 above.
8. Zero stores baseline values for capacitance and conductance or& no~bias~vo~fa~eobaseline values~are~storedl
changed.
START
Q
ZERO
ON?
YES
NO
Q-7
TURN
ZERO OFF
3.10.5 Examples of Zero Operation
Table 3-11 lists some examples to help clarify zero operation. This table lists stored baseline values, applied signals,
and resulting front panel display values for those combina-
tiOll.5.
Table 3-11. Examples of Zero Operation
Stored Baseline 1 Applied Signal
5.OOpF/2.6&
1m/1oms
llOpF115OmS
19.0OOpF/4.6@
680pF115mS
18OpF/lOOmS
Figure 3-12. Zero Operation Flowchart
1 Displayed Readings
14.OOOpF/2.00@S
-0.52OnF/5.00OmS
7O.CKlpF/-5OmS
3-21
OPEFIATlON
3.11 DRIFT CORRECTION
The front panel CAL button provides a means to perform
internal drift correction, as described in the following
paragraphs.
CONTROL
CAL
MooEL urn
0
3.11.1 Correction Procedure
Perform internal drift correction using the procedure
below. A flowchart outlining the basic sequence is shown
in Figure 3-13.
1. Select the 1 reading per second rate, or the slowest rate
you intend to use (seenote below).
2. If your instrument is equipped with both 1OOkHz and
lMHi mod&s, use FREQ to select the desired measurement frequency.
3. Select the first mnge to be c&rated with the RANGE
key.
cl++
4. Press the CAL key. The instrument will cyde through
the internal correction. sequence, a procedure that could
take up to 10 seconds, depending on the selected rate.
During the calibration process, the following message
will be displayed:
BUSY
5. Repeat steps 3 and 4 for all ranges you intend to use.
6. Select the other test frequency, if installed, and repeat
steps 3 through 5 above for that frequency.
NOTES:
1. The unit should be corrected using the above procedure
for each range and frequency you intend to use.
2. For optimum measurement acaracy, the correction pmcedure outlined above should be repeated at least once
per hour, especially in situations where the ambient
temperature varies widely.
3. During the correction sequence, the unit takes 10
readings of the @emal transfer standards. Thus, at the
l-reading per second rate, up to 10 seconds must be
allowed for each correction. If greater speed is a requirement, use the 10 reading per second rate, which wiIl
degrade correction accuracy only slightly.
4. Correction should be performed at a reading rate that
yields 4M-digit resolution (1 or 10 readings per second)
in order to achieve good accuracy. For maximum accuracy, always correct the unit at a 1 reading per second
rate, unless speed is a requirement as indicated above.
5. It is not necessary to disconnect external circuits from
the test INPUT and OUTPUT, as the instrument
automatically does so during correction.
6. New correction constants obtained by this method is only temporary and will be lost when the power is turned
off.
7. Pressing CAL will abort an active measurement sweep.
8. Upon power up, default values of 0 and 1 are installed
for drift correction coefficents, effectively resulting jn no
correction until CAL is used.
3-22
9. The CAL key is inoperative in autoranging or if cable
correction constants are in effect. A CONFLICT error
will occur iFyou press CAL under these conditions.
OPERATION
.3.11.2 Internal Correction Sequence
The general internal sequence the instrument follows when
performing drift correction is outline below. Figures 3-14
shows an outline of the sequence.
The internal sequence is as follows:
1. The insizmnent recalls correction constants for the
selected range and frequency from NVRAM. These con-
stants are derived during full instrument calibration, as
discussed in paragraph 7.3.
2. The unit disconnects the test INPUT and OLJTPIJT jacks
from internal circuitry. An open circuit measurement
is then made to be used as a zero offset value.
3. The instrument then measures an internal 2OpF or
2OOpF cap%itor (depending on the selected range). Ten
measurements per reading are made and then averaged in order to obtain a more accurate value.
4. The correction constant recalled in step 1 is divided by
the results of the measurement in step 3. The resulting
value is then stored in RAM and used by the instm-
ment as a compensating factor when taking normal
readings. This new correction values will then remain
in effect for the selected range and frequency for alI
subsequent measurements until the power is turned off,
or until CAL is pressed again.
Figure 3-13. Front Panel CAL Sequence
3-23
OPERATION
MEASURE
INTERNAL
CAPACITOR
TRIGGER
MANUAL
0
MODE
0
SOURCE
C~%+ANT
4.
REFERENCE
STORE NEW
Co%EoN
i
EXIT
c4
Figure 3-14. Internal Drift Correction Sequence
3.12 TRIGGERING THE INSTRUMENT
A trigger stimulus is used to initiate either one ieadjng,
or a groups of readings (called a sweep), depending on the
selected trigger mode. Basically, two bigger modes are
available: one-shot and sweep. The trigger stimuhrs itself
depends on the programmed trigger source: front panel,
external trigger input, or IEEE-488 GET, X , I
mand triggers.
The following paragraphs discuss trigger source and mode,
as well as trigger ov- conditions, and external trigger
input and output pulses.
3.12.1 Selecting a Trigger Mode
Available Trigger Modes
There tie two basic trigger modes available, including:
1. One-shot: with each trigger, the instrument takes a
single set of capacitance, conductance, and bias voltage
readings, and stores the resulting data at a single word
location in the A/D buffer. Data is also made available
to the display and J.EEE-433 bus when ready. A pulse
occurs at the external trigger output after each single
set of readings is completed. When the waveform is
complete (or the programmed number of points are
taken), the next trigger will reset the buffer and begin
storing readings at location one. For IEEE bus operation, an SRQ can be generated to signal to the controller
that no more triggers should be sent. See Section 4 for
IEEE488 bus information.
2. Sweep: a single trigger stimulus causes the insbxment
to cycle through the programmed bias waveform sequence. A single set of capacitance, conductance, and
bias voltage readings will be taben and stored in the AD
buffer at each bias steu. As each stenoccurs. the
resulting data will app& on the front panel and will
be made available for transmission over the IEEE-488
bus. The instrument will intitiate a pulse at the exter-
MI trigger output jack at the end of the sweep. The
display will continue to update after the sweep has been
nn,..-,e+arl
3-24
OPERATION
More information on the how the trigger modes affect the
Selection Procedure
Use the procedure below to select the desired &igger mode.
1. Press the MODE button and note the following message
is displayed:
TRIGGER MODE
2. Now press MODE or A/I repeatedly until the desired
trigger mode message is displayed, or press the
associated numeric key, as summarized in Table 3-12.
For example, for the sweep mode, the front panel
display message is:
TRIGGER MODE SWEEP
Similarly, the instrument will display the following
message for the one-shot mode:
TRIGGER MODE 1 SHOT
3. Once the desired mode is displayed, press the ENTER
key. The instrument will then return to the previous
display with the new trigger mode in effect.
4. If you wish to cancel trigger mode selection without
changing the selected mode, press QUIT.
NOTES:
2. Pressing RANGE, FREQ, FILTER, ZERO, GAL, TRIG-
GER MODE, any BIAS key, or re-programming the
reading rate will abort an active sweep without re-
triggering a sweep. In either case, any data stored in the
A/D buffer will be cleared. To avoid loosing data in this
situation, always transfer data to buffer B with the A
- B key before pressing other keys.
3. Triggering a sweep will clear data presently stored in
the A/D buffer (buffer A).
4. Only capacitance data is taken at the 1880 reading per
second rate.
5. At the 75 and 1CWsec rates, a CALCULATING DATA
message will be displayed at the end of a sweep before
the unit returns to normal operation.
3.12.2 Programming the Trigger Source
Available Trigger Sources
The programmed trigger source provides the stimulus to
initiate a one-shot or sweep depending on the selected trigger mode. Trigger sources include:
1. Front panel MANUAL button. Note that this button is
always operational regardless of the selected source
(unless the unit is placed in remote over the IEEE-488
bus).
2. External trigger pulse. An appropriate pulse, applied to
the external tigger input jack on the rear panel, provides the tigger stimuhrs.
3. IEEE command triggers. IEEE-488 GET, X, or talk com-
mands provide the stimulus when the appropriate
source is selected.
1. To =boa m =cfive swew P*= the MODE key. me
presently active sweep will be aborted, and the in&u-
ment will display the TRIGGER MODE message. Press
ENTER to return to normal display, and initiate another
trigger to begin another sweep, if desired.
Table 3-l 2. Trigger Modes
Numeric
Key # Display Message
0 TRIGGER MODE 1 SHOT
1 TRIGGER MODE SWEEP
Note that all trigger sources are available with all trigger
modes, as summarized in Table 3-13.
NOTE
In order to ensure rapid response to a new trigger
after a sweep, press MODE ENTER to turn off the
A/D converter.
Description
One-shot (one reading per trigger)
Single-sweep (one sweep per trigger)
3-25
OPERATION
Table 3-13. Trigger Mode and Source
Combinations
Front External IEEE IEEE IEEE
Source/Mode
One-shot X
Sweep
Selection Procedure
Select the desired trigger source as follows:
1. Press SOURCE (SHlFI MODE) and note that the following message is displayed:
Z.Press the MODE or A/I key repeatedly until the
desired trigger source is displayed (or press the
associated numeric key, as indicated in Table 3-14). For
example, for external triggering, the display w+ll show:
Panel Trigger Talk GET X
:: X ::
TRIGGER SOURCE
::
::
3.12.3 Front Panel Triggering
To trigger the instrument from the front panel, simply
press the MANUAL button. Note that this button is always
operational regardless of the selected trigger source (unless
the unit is placed in remote over the IEEE-+-, in which
case all front panel buttons except LOCAL will be locked
out). Thus, front panel trigg!r~ source selection provides
a me- to lock out-all other trigger sources when only
front-panel triggering is desired.
The number of ~feadings the ‘instrument takes after
MANUAL is pressed will depend on the selected trigger
mode. In the one-shot mode, you must press MANUAL
for each reading. Pressing MANUAL with the sweep mode
active performs a complete reading sequence, with a group
of capacitance, conductance, and bias voltage readings
taken and stored with each biis step. If you press
MANUAL and the instrument is not ready, an err01
message will be displayed, as discussed in the following
paragaph.
3.12.4 Trigger Overrun Conditions
TRIGGER SOURCE EXT
3. Once the desired trigger source is displayed, press the
ENTER key. The instrument will then return to the
previous display with the new trigger source in effect.
At this point, the selected trigger stimulus must be ap-
plied to initiate the reading or sweep.
4. To return the display to normal without changing the
previously selected source, press the QUIT button
instead.
TRIGGER SOURCE TALK IEEE tslk command.
TRIGGER SOURCE GET
TRIGGER SOURCE X IEEE X command.
Once the instrument is triggered, it will begin a reading
or sweep, as discussed above. If another trigger is received while the unit is processing a reading or sweep (depen-
db,
g on trigger mode), a trigger overrun condition will oc-
cur, in which case the instrument will display the followjng error message:
TRIG-OVERRUN
Fiie 3-15 shows a flowchart of trigger overrun operation.
Trigger Source
TRIGGER JNPUT.
IEE GET command.
3-26
*hJ.ANLJAL key is always operationa&
OPERATION
NOTES:
1. A reading or sweep is not aborted by a trigger overrun
condition.
2. Only an active trigger source can create a trigger over- itiates a complete sweep sequence. -- nm situation. For example, if you select the externsl tng-
ger source, either an external trigger pulse or pressing
MANUAL can create a trigger overron (recall that
MANUAL is always active).
3. The instrument will not generate an overrun error if triggered during the stop time. At the end of the sweep,
a new sweep will be started in this situation.
YES
As previously described, the effect of the external trigger
pulse depends on the selected trigger mode. In the oneshot mode, a separate pulse is required for each reading
set;mIn the sweet mode, however, one trizzer pulse in-
I
TRIGGERS ON
LEADING EDGE
I
NOTE
TRIGGER INITIATES
ONE READING. IONE-SHOT)
OR ONE SWEEP 6WEEP).
Figure 3-16. External Trigger Input Pulse
Specification
Figure 3-15. Trigger Overrun Operation
3.12.5 External Trigger Input
To use external triggering, first select that source with the
MODE key as described in paragraph 3.12.2. The instru-
ment will then be triggered when an input pulse with the
spec%ications shown in Figure 3-16 is applied to the EXTERNAL TRIGGER INPUT jack. The unit is triggered on
the leading edge of the pulse, as shown on the diagram.
Note that the center conductor of the jack is high, and the
outer ring, which is connected to IEEE common is low, as
shown in Figure 3-16.
3.12.6 External Trigger Output
The externsl trigger output provides a negative-going, TKcompatible pulse as shown in Figure 3-18. The leading edge
of the trigger pulse indicates end of reading or end of
sweep as the case may be.
me occurrence of this pulse occurs depends on the selected
trigger mode as follows:
1. At the end of each reading in the one-shot mode.
2. At the end of the syeep in the sweep mode (1, 10, and
18 lsec reading rates).
3. In the 751sec and lOOO/sec rates, the pulse occurs after
alI cakulations are done in the sweep mode.
The center conductor of the external trigger output is high
and the outer ring, which is connected to IEEE common
is low, as shown in Figm.e 3-l7,
3-27
OPERATION
HIGH
Low (IEEE488 COMMONI
Figure 3-17. Trigger Input and Output Jack
Configuration
Figure 3-16. External Trigger Output Pulse
Specification
3.12.7 External Triggering Example
- Figure 3-19 shows the basic circuit configuration for using
the Model 590 along with Model 230 as an external bias
voltage source. Connect~and program the instruments as
follows:
1. Using suitable coaxial cable, connect the 230 OUTPUT
to the 590 VOLTAGE BIAS INPUT on the rear panel.
2. Using a BNC coaxial cable, connect the 230 EXTERNAL
TRIGGER OUTPUT to the 590 EXTERNAL TRIGGER
~~~~~~ ~INMJT.
3. Select the STEP program mode on the Model 230 and
program the various memory locations for voltage
levels, current limits, and required dwell times. The
dwell time should be programmed to the desired step
zisonsolt the Model 230 Instruction Manual for
4. Press the 590 MODE key and program the unit for the
one-shot trigger mode. Press SHIFT MODE and select
the external higger source.
5. Press the 590 WAVEFORM key repeatedly until the the
EXT display message is shown, then press the ENTER
key. Tlus step sets up the unit for use with the external bias souxe.
6. Connect the circuit under test to the 590 test INPUT
and OurnuT jacks.
7. Select the 590 range, frequency, model, and reading
rate, as required. -
- -
8. Press the 230 RESET key and turn on its output with
the OPERATE key.
WARNING
Up to + 1OlV may be present at the 590 TEST
OUTPUT after the next step.
e
As an example of external triggering operation, let us
assume that the Model 590 is to be used with a Keithley
Model 230 Programmable Voltage Source. The Model 230
is~ capable of output voltages as high as * 101V. Thus, this
instrument is ideal for testing devices which require higher
bias levels than the nominal *2OV level of the Model 590.
3-20
9. Press the BIAS ON key to turn on the voltage applied
through the test OUTPUT jack.
~10. Start the measurement sequence by pressing the 230
START/STOP key. This action will trigger a 590
reading; Ate the end of the dwell time for location, a
590 reading will be triggered and stored.
NOTE
See paragraph 4.12 for an example program that
demonstrates this process.
OPERATION
MQDEL 230
VOLTAGE SOURCE
Figure 3-19. External Triggering Connections
3.12.8 IEEE-488 Bus Triggering
To trigger a reading or sweep with an IEEE-488 trigger
source, you must send the appropriate IEEF&8 command
over the bus: X, talk, or GET, depending on the selected
source See Section 4 for complete details.
WARNING:
HAZARDOUS VOLTAGE
MAY BE PRESENT
AT TEST OUTPUT
Jf one of these commands has been selected as the trigger
source, you can also trigger the instrument by pressing the
MANUAL button unless the insimment is in remote.
3-29
OPERATION
T(j Eacj e b
QUIT
DATA
CABLE CAL
0 c”7 ti b
SAVE RECALL
ENTER
b cl3 ti
l/C2
BUFFER
0
A-B
\
3.13 DATA KEYS
The following paragraphs desaiie the operation of the O-9.
+I-, A and V keys, the SHIFT key, and ENTER and
QUIT. BUFFER and A operation in paragraph 3.15.9.
B are discussed under buffer
+I-
0 r”l I
CA-CB
CABLE #
c/c,
WA-V&CONST.
press the SHIFT key a second time. If you press a key
which has no shifted function with shift enabled, the
primary function of that key will be performed. For example, pressing SHIFT FILTER performs the same operation
as simply pressing FILTER.
SELF TEST
IEEE
CMAX
C vs T
3.13.1 Increment and Decrement
One purpose of inaement (A) and deaement (‘I) is to
scroll through various buffer locations when viewing buffer data on the front panel. To use these keys with the buffer, simply press BUFFER, select the buffer to access at the
prompt, then use A or V to sequentially scroll through
buffer locations
A second purpose of A/V is to provide an alternate
method of scrolling through menus when using RATE,
WAVEFORM, PARAMETER, MODE, and SETUF’.
3.13.2 SHIFT/QUIT Key Operation
The SHIFI key is used to add secondary functions to some
other front panel keys such as MODE (SOURCE). To access one of these shifted modes, simply press the SHIFT
button followed by the desired key. Note that the indicator
adjacent to the SJ%FT key will be on when shift is active.
To cancel shift without selecting a corresponding mode,
When progr
you can return to normal display without making a change
by pressing QUIT. QUIT can also be used to exit buffer
display.
amming parameters or selecting menu options,
3.13.3 Numeric Input
The numeric keys, which include O-9 and f are used to
program numeric data when programming such
parameters as bias wavefonnvoltages. Also, some plotter
setups are entered in numeric form.
To use these keys, simply press the desired mode button
and then enter your data with the number keys.
3.13.4 Display Cursor
During the process of programming numeric parameters,
the display digit affeaed by~a key press will brighten to
act as a display qursor. As you type in digits, the cursor
will move to the right, and it will wrap around to the left
after passing through the right most digit. The - key can
be used to move the cursor to the right.
l
3-30
e
Use the ENTER key to update the active variables to the
new values. A single press of ENTER wiU save all
parameters just modified, and the unit will return to normal display.
3.13.8 Multiple Parameter Entry
When progmmmin
not press ENTER after each entry. Instead, press the A
or V key to scroll to the next parameter to be programmed. After changing the last parameter, press the ENTER
key to modify all parameters just entered. If you press
QUlT at any point, none of the parameters will be changed.
g two or more parameters, you need
3.13.7 Invalid Parameter Entry
Parameters are check for validity when you scroll to the
next pammete: set and when the ENTER key is pressed.
If a p&meter outside the allowed range is programmed,
the unit will briefly display the following message:
INVALID
The unit wiu then return to the erroneous value.
3.13.8 Data Key Examples
OPERATION
hers in the voltage display area.
6. Press and hold A and note that the instrument scrolls
through buffer locations in ascending order.
7. Release A, then press and hold v. Note that the in.ss.ent scrolls through buffer locations in descending
8. Press QUlT to return to the normal display.
Example 2: Lk~k~InaementlDeaement to Saoll Through
As discussed previously, the second purpose for A/I is
~for saolliqg through parameter menus, as in the follow-
ing example.
1. Press the WAVEFORM key and note that a bias
waveform type is shown on the front panel display. For
example, the display might show:
BIAS WAVEFORM DC
2. Press and hold the A key and note that you can saoll
through bias waveform types.
3. Now press and hold the v key and note that you can
scroll through available bias waveforms in the opposite
direction.
4. Press QUlT to return to normal display without changing the previously programmed waveform.
5. Press ENTER to return to the normal display with the
newly selected waveform in effect.
The examples below will help to clarify data key operation.
Example 1: Using Inaement and Decrement to Access the
Buffer
One purpose of A/V is to scroll through buffer locations
and display data stored there. The following procedure
demonstrates the basic process.
1. Using the MODE key, select the sweep trigger mode.
2. Press MANUAL to trigger a reading sweep.
3. Press BUFFER and select the A/D buffer at the prompt
by pressing 0. Note that the BUFFER LED turns on to
indicate that you are reading buffer data.
4. The instmment will then show data stored at the first
buffer location. A typical display is:
1.8OOOpF 00.11~& +5.1OOV
5. Press the C vs t button to display buffer location num-
Example 3: Programming Single Numeric Parameters
To program a numeric parameter, simply press
PARAMETER and then key in the desired value, as in the
example below.
1.
Using the WAVEFORM key, select a single staircase
WaVefOlTll.
2.
Press the PARAMETER key repeatedly u&l the following message is displayed:
IST BIAS V +CHl.OOO
3.
Now key in the desired voltage with the numeric keys.
For example, to program a value of 4V, press: 0 4 0 0
0. Note that the digit affected by a keypress is
highlighted on the display.
4.
To complete programming, press the ENTER key. The
voltage value wiIl be programmed, and the unit will
return to the normal display mode.
3-31
OPERATlON
Example 4: Programming Multiple Parameters
when pro*
necesw to press ENTER af& ea?h modification; instead,
you can scroll through the parameter menu, stopping at
each point to make the desired changes. The example
below demonstrates this process by progr
stop, and step times.
1. Use the WAVEFORM key to select a single staircase
waveform.
2. Press PARAMETER and note that the following prompt
for start time is displayed:
In this instance, the default time of lmsec is displayed.
3. Key in the desired start time using the numeric keys.
For example, to program a 2.5sec start time, press the
following keys: p 2~ 5 0 0.
4. Press the PARAMETER or A key. The unit will now
display the programmed stop time:
Again, the default lmsec value is displayed in this
example.
5. Key in the desired value with the numeric keys. For ex-
ample, to program a 5.04secstop time press the following keys: 0 5 0 4 0.
6. Press PARAMETER nor A again to display the pro-
grammed step time, as in the example below:
amming more thsn one parameter, it is not
amming itart;
START TIME +OO.OOl
STOP TIME +OO.OOl
4. Press the + I- key a number of times and note that the
display alternates between a positive and negative value.
Example 6: Demonstrating an INVALID Error Message
lf you attempt to program a~psae$er outside the allow-
ed range, the instrument will display the INVALID error
message, as in the example below.
1. Press PARAMETER repeatedly until the following
message is displayed:
1ST BIAS +05.000
The 5V value is due to Example 3 above.
2. Attempt to program a bias voltage above the allowed
+ 20V value by pressing the following keys in sequence:
30000.
3. Now press ENTER to complete progmmming. Note that
the instrument briefly displays the message below and
then returns to the previously programmed parameter:
lNVALlD
4. Press A or V to attempt to scroll to the next p-eter
item. Again, the instrument briefly displays the INVALID error message and returns to the previously pro-
grammed parameter.
a
STEP TJMJZ +OO.OOl
7. Key in the desired value; for example, to program a
5Omsec step time, press the following keys: 0 0 0 5 0.
8. Press the ENTER key to complete programming of all
three values. The unit will then return to normal
display.
Example 5: Using the +I- Key
The + I-pokey is used to select positive or negative bias
voltage parameters, as in the following example.
1. Select a single staircase waveform using the
WAVEFORM key.
2. Press the PARAMETER key until the following message
is displayed:
1ST BIAS V +00X00 ~~~
3. Program a 5V voltage as follows: 0 5 0 0 0.
Example 7: Using ENTER and QUIT
The ENTER key is the last step in the parameter or menu
selection process. ln contrast, the QUlT key allows you to
cancel a mistake in parameter programming or menu selection w$hout act+ly changing previously programmed
values. The example below demonstrates operation of
ENTER and QUIT.
1. ~Press the MODE key and note the programmed trigger
mode. For example, the display might show:
TRIGGER MODE SWEEP
2. Press MODE again to change the trigger mode.
3. Press ENTER to invoke the trigger mode change.
4. Press MODE and note that the new trigger mode pm-
grammed in steps 2 and 3 is now in effect. For example, if you changed from fie sweep to one-shot mode,
the display will show.
TRIGGER MODE 1 SHOT
3-32
3.14.1 Selecting a Bias Waveform
OPERATION
7. Now press MODE again and note that the trigger mode
was not changed because you used the QUlT button.
Example 8: Using the - Key to Move Cursor
- can be used during parameter programming to move
the cursor. The example below will demonstrate +I&
process.
1. Press PARAMETER tid note that the instrument
displays the start~time, as in this example:
START TIME +OO.OOl
2. Press the
right.
- key and note that the cursor moves to the Selecting a Waveform
3.14 BIAS VOLTAGE
The following paragraphs contain information on selecting
a bii waveform programming bias waveform parameters,
and turning on the bias voltage. Details on waveform
definitions, external bias jnput, and bias voltage monitor
Available Waveforms
Bias waveforms available include:
DC: any DC level in the range of -20V to +2OV.
Single staircase: stepping either up or down.
Dual staircase: stepping up then down or down then up.
Pulse train: step up or down.
External: external voltage applied to the BIAS INPUT jack.
Table 3-15 shows the general configuration of these
waveforms. Waveforms and parameters are defined in
detail in paragraph 3.14.4.
To select the required bias waveform, press and hold the
WAVEFORM or AIV key until the desired waveform is
displayed, as summarized in Table 3-15. Press ENTER to
select the displayed waveform. type.
If you wish to return to the previously programmed
waveform, press QUIT instead.
WAVEFORM
0
PARAMETER
0
Typical Waveform Uses
Typical uses for the various waveforms include:
1. DC
A. One-shot: external time base C-t.
B. Sweep: internal time base C-t, or “dumb” C-meter.
2. Single staircase
A. One-shot: external time base C-V.
B. Sweep: internal’time base C-V.
3. Dud staircase
A. One-shot: external time base C-V with hysteresis.
8. Sweep: internal time base C-V with hysteresis.
4. Pulse
A. One-shot:~extemal time base C-V, return to default
between points.
B. Sweep: internal time base C-V, retumto default be-
tween points.
5. External
A. One-shot: external time base, external bias control.
B. Sweep: internal time base, external bias control.
3-33
Table 3-15. Bias Waveform Summary
BIAS WAVEFORM DC Static DC levels
BIAS WAVEFORM PULSE
BIAS WAVEFORM EXT
External Bias Voltage
3.14.2 Programming Waveform Parameters
Parameter Types
Programmable parameters in&de such values as start and
stop hold times, as well as first andlast bias values. Table
3-16 summakes programmable parameters, display
messages, and the allowable range for each parameter.
NOTE
1. All parameters are note programmable for every waveform type. The instrument will prompt only for those
parameters which apply to the particular waveform.
Table 3-16 indicates which parameters apply to the
various waveforms.
2. The buffer can store a maximum of 450 readings (1,350
at lOOD/sec rate). Programming reading sweeps longer
than this will result in lost data. If more than450 (1,350)
readings are programmed (a function of fkst and last
voltage, as well as bias step size), only the first 450
(1,350) will be stored in the buffer.
3. The number of readings stored as part of the sweep for
all waveforms except DC and external de
stop, and step voltages (see discussion b ef
ends on start,
ow). The numher of readings stored with external and DC waveforms
depends on the COUNTparameter (COUNT 5450 or
s1,350).
4. Volta e parameters can be rogr-ed to lmV resolution,& achdy have 5m$ minimum steps. @
3-34
Table 3-16. Programmable Bias Voltage Parameters
STOP TIME
STEP TIME
ET BIAS V
LAST BIAS V
BIAS STEP V
DEFAULT BIAS V
Description Liiw
Start time*
Eg p-:
Fit bias voltage
Last bias voltage
Step voltage
Default bias
lmsec to 65sec
lmsec to 65sec
lmsec to 65sec
-20v to +2ov
-20v to +2oJ7
-20v to +2ov
-20v to +2ov
lesolution*
lmsec
lmsec
lmsec
5mv
5mv
5mV
5mv
JC
X
c
X
X
iticas
::
X
X
X
X
X
Single
SS
‘taircasc
X
::
X
X
X
X
Dud
%llSl
-
::
X
X
X
;
3xtemal
::
X
voltage
COUNT
#Readings per
sweep
1 to 450 (1,350
at l,OOO/sec rate:
X
-
X
t
NOTE: Voltage parameters can be programmed to lmV, but are set in 5mV steps.
* Programmed times must be multiplied by 1.024 to obtain actual times.
X Indicates parameter applies to waveform.
5. Bias voltages of -2O.OOOV can be programmed, but the
read-back display limit is -19.999V. Buffer data will
reflect voltage values above -19.999V, however.
6. Once you have programmed all parameters, press the
ENTER key. The instrument will then return to normal
display.
6. The unit wiIl display a CONFLICT erzor if you attempt 7. To return to normal display without modifying parato program more than 450 readings per sweep. meters, press QUIT instead of ENTER.
Selecting Parameters
Number of Readings in a Sweep
Select your bias parameters as follows: The number of readings in a sweep for DC and external
waveforms is determined solely by the COUNT parameter.
1. Press and hold WAVEFORM or A (or press the
associated numeric key, as indicated in Table 3-15) to
select the desired waveform. For example, for a single
Thus, to control the size of the sweep, you should pro-
am COUNT for the desired number in the range of l-450
1,350 at l,oOO/sec rate).
i?
staircase, the display will show:
BIAS WAVEFORM STAIR
In contrast, the number of readings in the sweep for the
staircase or pulse waveforms will depend on the pro-
2. Press ENTER to program the waveform.
3. Press and hold PARAME
TEE or A/l until the readings in a given sweep with a single staircase waveform
gr-ed first and last bias levels, as well as the ste
with a maximum of 450. For example, the total num E
size
er of
parameter you wish to program is displayed. For ex- - be computed a.~ fokws:
ample, to program start time, the displq will show:
START TIME +OO.OOl
n=
(vz - VP)
+ *
V*
4. Using the number keys, program the desired value. For
example, to program a step delay of O.lsec,~press: 0 0 where: n = number of *e=hgs
1 0 0.
5. Press PARAMETER or A to advance to the next
parameter, then key in the desired value. If ou key
m an invalid parameter, the instrument
wd briefly
display the following error message: ~~~~~~~~~~~ ;
(Jf(V, - V,)/Vs is not an integer, add one to the above
Vr = last bias
v, = first bias
V, = step bias.
calculation).
INVALID
3-35
OPERATION
Example: Assume that first and last bias values are -5V
and +5V respectively, and that the bias step is O.lV. The
number of readings in the sweep is:
(+5 - (-5)) +1
II=
“.I
n = 101 readings per sweep
Minimum Voltage Steps
Althou h bias voltage parameters can be entered in 1mV
steps, t& actual minimum step size is 5mV due to hardwsxe constraints. The actual bias voltage is calculated as
follows:
v,= Lv” _I x5
5
Where: V, = Actual bias voltage (mv)
V, = Progmnuned bias voltage (mV)
L -1
= -p&e the &,tegei of x,r,/s
The exact configmaiion of internal circuitzy differs depending on whether the internal or external bias source is used,
as shown in Figure 3-21(a). Figure 3-21(b) shows an
~~~;o~ent circuit of software implementation of bias
NOTES
1. Pressing ON will abort any active sweep and reset the
AID buffer.
2. When external bias is turned o?, and the unit is not triggered, the following will be &played:
W.-V
ENTER
9
Example: Assume a 1023V voltage is programmed. The
actual bias voltage is:
10,223
v,= L- _I x 5 = 10,22omv
5
vs = 10.22v
3.14.3 Controlling the Bias Voltage (BIAS ON)
Use the BIAS ON key to turn the bias voltage on or off.
The indicator to the right of the ON key shows the state
of the bias source. Note that this key controls both internal and external bias sources (when selected). Figure 3-20
shows a flowchart of ON key operation.
WARNING
Up to 200V DC may be present at the test OUTPUT jack when the bias is turned on while using
an external bias source.
Figure 3-20. Bias ON Key Operation
3-36
l”%H=
:ME!Lz 3f
VOLTAGE
OPERATION
3.
BIAS
MONITOR
OUTPUT
BIAS
INPUT
Ah
s9-
aD’....,.. ~~~ ,,,
MAX
3.9fl
>a
0
‘“v
-h
INTEXT BIAS ON
”
&
16
..,
1OOM
Figure 3-21. Bias Switching Network (a) Software Implementation of Bias ON (b)
3.14.4 Bias Waveform and Parameter
Definitions
DefinXoi~s for the various bias waveforms and associated
parameters are located in Fi
applicable, a sepzate de~~%t%?%%3kr?~
wavefoorm and trigger mode combination.
Parameters are further defined as follows:
Start Time: The time period, occurring on the f&t bias step,
from the point the instrument is fkst triggered until the
fit step iime.
Stoo Time: The time oeriod after the read& taken dur-
in ihat last bias step before the instrument r&ms to the
dsauIt bias voltage.
Step Tie: The time period after a transition to a new bias
step before the instrument begins a measurement (except
for the, firsts step, which also includes the start time).
First Bias: The initial voItage setting of the bias waveform.
Last Bias: The final voltage setig of the bias waveform.
Step Bias: The incremental change of each step in the bias
voltage waveform.
Default Bias: The bias voltage value both before and after
a sweep, or between pulses (pulse train only).
Count: This uarameter sets the number of readinzs Der
sweep for thi DC and external bias waveforms o$y:
3-3j
OPERATlON
NOTES: First Bias: +5v
Last Bias: -5V
1. To program a negative going staircase (or negative &en
positive dual staircase) set the first bias more positive
than the last bias, and use negative step bias values. For
example, to step from +SV to -5V in 1OOmV increments, program the following values:
Step Bias: -O.lV
2. The final step bias value may be smaller than the progmmmed Step Bias, depending on First Bias, Last Bias,
and Step Bias Values.
r
------
I
DEFAULT BIAS
- NOTES:
TRIGGER EVENTS
FIRE-T BIAS
1. ONE TRIGGER PER READING IS REQUIRED.
2. NUMBER OF READI NGS IS DETERMINED BY COUNT PARAMETER.
3. BUFFER IS .RESET BY NEXT TRIGGER AFTER COUNT # OF READINGS ARE TAKEN.
:iiF&R
DEFAULT BIAS
-----
3-36
Figure 3-22. DC, One-Shot
DEFAULT WAS
OPERATION
TRIGGER EVENTS
+
FIRST BIAS
NOTE:
1. UNIT TAKES COUNT PARAMETER NUMBER OF READINGS PER TRIGGER EVENT
2. THE A/D BUFFER IS CLEARED BY A TRIGGER
Figure 3-23. DC, Single Sweep
3-39
OPERATION
TRIGGER EVENTS
1 f + I + 1 t c +
LAST BIAS
DEFAULT BIAS
r
L
NOTES
1. ONE TRIGGER PER READING REQUIRED.
2 NEXT TRIGGER AFTER AU STEPS TAKEN RESETS SUFFER.
J‘ k-L
BUF$
DEFAUU BIAS
DEFINITION OF LAST STEP
-LAST VDLTAGE STEP MAY ND7 BE FULL OC A a STEP BIAS
Figure 3-24. Single Staircase, One-Shot
3-40
OPERATION
E
z
E
1
- DEFAULT BIAS DEFAULT BIAS
FIRST BIAS FIRST BIAS
LA’.3 BIAS LA’.3 BIAS
NOTES:
1. ONE SWEEP PER TRIGGER.
2. A TRIGGER RESETS THE BUFFER.
3. SEE FIGURE 3-5.4 ?t%k LAST STEP DEFINITION.
E
E
E
1
DEFAULT BIAS DEFAULT BIAS
Figure 3-25. Single Staircase, Sweep
3-41
OPERATION
t t 1 +
7 DEFAULT-El& ,
j FIRST B,AS.$$d
TRIGGER EVENTS
+ 1 t t
LAST BIAS
I.7
r-----J-+
NOTES:
1. ONE TRIGGER PER READING REWIRED.
2. TRIGGER AFTER ALL STEPS TAKEN RESETS BUFFER.
-+
%F
DEFAULT BIAS
1 FlRST BlA$
Figure 3-26. Dual Staircase, One-Shot
3-42
DEFAULT BIAS
OPERATION
NOTES:
1. ONE TRIGGER PER SWEEP REQUIRED.
2. TRIGGER RESETS BUFFER.
Figure 3-27. Dual Staircase, Single Sweep
3-43
OPERATION
BIAS
NOTES:
1. ONE TRIGGER PER READING REQUIRED.
2 FlRST TRIGGER AFTER LAST STEP RESETS BUFFER.
Figure i-28. bulse Train, One-Shot
3-44
OPERATION
TRIGGER
1 I
FIRST BIAS
‘7 STEP BIAS
LAST BIAS
DEFAULT BIAS
TRIGGER
~
DEFAULT BIA
NOTES:
1. ONE SWEEP PER TRIGGER
2. BUFFER IS RESET BY TRIGGER.
Figure 3-29. Pulse Train, Single Sweep
345
OPERATION
NOTE.9
1. COUNT # OF READINGS ARE STORED IN BUFFER
2. FIRST CYCLE ONLY INCLUDES START TIME.
3. LAST CYCLE ONLY INCLUDES STOP TIME.
4. FIRST TRIGGER AFTER COUNT # READINGS RESETS THE BUFFER.