Keithley 580 Datasheet

580 Micro-ohmmeter
with Test Leads & Test Lead Pouch
The Model 580 Micro-ohmmeter combines high accuracy, resolution, and sensitivity with three spe­cial capabilities that make micro-ohm measure­ments easier and more versatile than ever:
1. When measuring contact and connector resist­ances, it is important not to puncture oxides and films that may have formed. The Model 580 ensures this dry-circuit condition by clamping the open-circuit test voltage to 20mV on the 200m, 2, and 20ranges when the 20mV MAX button is pressed.
2. The Model 580 can test more devices by enabling both test current polarity and wave­form (DC or pulsed) to be selected. When
10µsensitivity
20mV voltage clamp
Selectable test current
waveforms
3 sets of test leads included
Optional IEEE-488 interface
Optional battery
Ordering Information
580 Micro-ohmmeter with
Test Leads and 5801 Test Lead Pouch
580-1 Micro-ohmmeter without
Test Leads and Pouch
580/5802
Micro-ohmmeter with Isolated Analog Output/
Combines high accuracy, resolution, and sensitivity for low resistance measurements
IEEE-488 Interface
580/1978
Micro-ohmmeter with Rechargeable Battery Pack
580/02/78
Micro-ohmmeter with Isolated Analog Output/ IEEE-488 Interface and Rechargeable Battery Pack
Model 580/5802 rear panel
Resistance measurements can be made from 10µto 200kon seven ranges, with 41⁄2-digit resolution and accuracy to within 0.04% of reading. Ranging can be performed either manually or automatically.
The Model 580 is supplied with three sets of test leads—standard leads, Kelvin probes, and Kelvin clip leads. For applications requiring portability or line power isolation, an optional battery pack is also avail­able. Digital calibration is possible through front panel controls or over the optional IEEE-488 bus interface.
ACCESSORIES AVAILABLE
TEST LEADS
5804 Test Lead Set 5805 Kelvin Probes, 0.9m (3 ft) 5805-12 Kelvin Probes, 3.6m (12 ft) 5806 Kelvin Clip Leads 5807-7 Helical Spring Point Test Leads
using pulsed test current, the Model 580 auto­matically compensates for thermoelectric EMFs. Tests on inductive devices are best performed using DC test current, as this avoids the effects of time constants on pulsed current resistance measurements. For temperature-sensitive components, a single trigger mode of operation minimizes power delivered to the device.
3. The optional IEEE-488 bus interface permits operation as a stand-alone unit as well as in a PC-based system.
RACK MOUNT KITS
1010 Single Rack Mounting Kit 1017 Dual Rack Mounting Kit
OTHER
1978 Rechargeable Battery Pack 5801 Test Lead Pouch 5802 Isolated Analog Output/IEEE-488 Interface
This product is available with an Extended Warranty.
Accessories Supplied
5801, 5804, 5805, and 5806 test leads; Operator’s and Service Manuals.
1.888.KEITHLEY (U.S. only)
LOW V/LOW R PRODUCTS
www.keithley.com
Model 5802: Isolated analog output/IEEE-488 interface for Model 580 (cover included).
A GREATER MEASURE OF CONFIDENCE
580 Micro-ohmmeter
with Test Leads & Test Lead Pouch
GENERAL
DISPLAY: 41⁄2 CONNECTORS: Measurement and rear panel EXTERNAL
RELATIVE: Display annunciator indicates REL. DRIVE: Selects either pulsed or DC SOURCE current. Pulsed
POLARITY: Selects either positive or negative SOURCE cur-
TRIGGER: Allows single pulsed measurements. EMC: Conforms to European Union Directive 89/336/EEC. SAFETY: Conforms to European Union Directive 73/23/EEC
OPERATING ENVIRONMENT: 0°–50°C, <80% relative humid-
POWER: 105–125V or 210–250V (switch selected), 90–110V
DIMENSIONS: 89mm high × 241mm wide × 300mm deep
-digit LCD.
TRIGGER inputs: Banana jacks.
drive provides automatic cancellation of thermal offsets using 50% duty cycle pulse. Display annunciator indicates drive selected.
rent in either drive.
(meets EN61010-1/IEC 1010).
ity up to 35°C; linearly derate 3% R.H./°C from 35° to 50°C.
available. 50–60Hz, 12VA. Optional 6-hour battery pack, Model 1978.
1
2 in × 91⁄2 in × 113⁄4 in). Test lead pouch adds 76mm (3
(3 in) in height.
IEEE-488 BUS IMPLEMENTATION
MULTILINE COMMANDS: DCL, SDC, GET, GTL UNT, UNL,
SPE, SPD, LLO.
UNILINE COMMANDS: IFC, REN, EOI, SRQ, ATN. INTERFACE FUNCTIONS: SH1, AH1, T5, TE0, L4, LE0, SR1,
RL0, PP0, DC1, DT1, C0, E1.
PROGRAMMABLE PARAMETERS: Range, DRY CIRCUIT
TEST, Operate, RELative, POLARITY, DRIVE, TRIGger, Calibration, EOI, SRQ, Status, Data Format, Terminator.
NON DRY DRY CIRCUIT
CIRCUIT TEST TEST
MAXIMUM 1 Year, 18°–28°C POWER 1 Year, 18°–28°C
RANGE RESOLUTION CURRENT PULSED DC IN SAMPLE PULSED
200 m 10 µ 100 mA 0.04 + 2 0.04 + 3 500 µW 0.05 + 2
2 100 µ 10 mA 0.04 + 2 0.04 + 3 50 µW 0.05 + 2
20 1m 1 mA 0.04 + 2 0.04 + 3 5 µW 0.05 + 2
200 10 m 1 mA 0.04 + 2 0.04 + 2
2k 100 mΩ 1 mA 0.04 + 2 0.04 + 2
20 k 1 10 µA 0.05 + 2 0.05 + 2
200 k 10 10 µA 0.075 + 2 0.075+ 2
CONFIGURATION: 4-wire (two sense, two source). MAXIMUM SOURCE VOLTAGE: 20mV in Dry Circuit Test, 1V oth-
erwise.
MAXIMUM TEST LEAD RESISTANCE 200mand 2Ranges: Up to 5in each SOURCE lead and
10in each SENSE lead with Non Dry Circuit Test; up to the selected full range resistance in each SOURCE lead and 10Ω in each SENSE lead with Dry Circuit Test.
20through 200kRanges: Up to half of the selected range in
each test lead.
TEST ±(%rdg + counts) DISSIPATION ±(%rdg + counts)
ACCURACY MAXIMUM ACCURACY
CONVERSION RATE: 3 readings/second typical. RANGING: Auto or manual. AUTORANGING TIME: 200ms per range change, average. SETTLING TIME: <1 second to within 10 counts on range. MAXIMUM INPUT OVERLOAD: 10V limited to 10A. MAXIMUM COMMON MODE VOLTAGE: 30V rms at DC, 50 or
60Hz.
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C):
±(0.1 × applicable accuracy specification)/°C.
ISOLATED ANALOG OUTPUT/ IEEE-488 INTERFACE (Model 5802 Option)
ANALOG OUTPUT LEVEL: 1V = 10,000 counts on ×1 gain.
1V = 100 counts on ×100 gain. Maximum output voltage = ±4V.
ACCURACY: ±(0.25% of displayed reading + 2mV). In ×100,
2mV output = 0.2 displayed counts.
RESPONSE TIME: Follows display conversion rate. OUTPUT RESISTANCE: 1000. ISOLATION: ANALOG OUTPUT LO is connected to IEEE COM-
MON. Maximum common mode voltage from IEEE COMMON to earth ground is 30V rms at DC, 50 or 60Hz.
Model 580 Specifications
1.888.KEITHLEY (U.S. only)
www.keithley.com
A GREATER MEASURE OF CONFIDENCE
LOW I/HIGH R PRODUCTS
Loading...