Keithley 2601A, 2635A, 2602A, 2611A, 2612A, 2636A User Manual
Specifications and Main Features
Frequently Asked Questions
User Manual
www.keithley.com
Series 2600A System SourceMeter
®
User’s Manual
2600AS-900-01 Rev. B / September 2008
A GREATER MEASURE OF CONFIDENCE
WARRANTY
Keithley Instruments, Inc. warrant s this product to be free from defect s in materia l and workmanship fo r a per iod of
one (1) year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
software, rechargeable batteries, diskettes, and documentation.
During the warranty period, Keithley Instruments will, at its option, either repair or replace any product that proves
to be defective.
To exercise this warranty, write or call your local Keithley Instruments representative, or contact
Keithley Instruments headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions.
Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product
returned, transportation prepaid. Repaired or replaced products are warranted for the balan ce of the original
warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley Instruments’ express
written consent, or misuse of any product or part. This warranty also does not apply to fuses, software,
non-rechargeable batteries, damage from batter y leakage, or problems arising from normal wear or failure to follow
instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES
PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE
OF ITS INSTRUMENTS AND SOFTWARE, EVEN IF KEITHLEY INSTRUMENTS, INC. HAS BEEN ADVISED IN
ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT
ARE NOT LIMITED TO: COST OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF
INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part without the prior written
approval of Keithley Instruments, Inc. is strictly prohibited.
TSP, TSP-Link, and TSP-Net are trademarks of Keithley Instruments, Inc.
All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders
Cleveland, Ohio, U.S.A.
Document Number:
2600AS-900-01 Rev. B / September 2008
The following safety precautions should be observed before using this product an d any associated instrumentation. Although some
instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may
be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required
to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer
to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument.
They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or
replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the
operator may perform them. Otherwise, they should be performed only by service personnel.
Safety Precautions
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service
personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rate d Measurement Category I and Measurement
Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and
data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient
over-voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC
mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise
marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The
American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak,
or 60V DC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are
prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact.
Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of
operating at or above 1000V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER
connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current
and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting
cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
11/07
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power
!
disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test.
AL W AYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers,
installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always
make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information,
and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid
interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The symbol on an instrument indicates that the user should refer to the operating instructions located in the user documentation.
The symbol on an instrument shows that it can source or measure 1000V or more, including the combined effect of normal and
common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly
disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the
associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the
warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
T o maintain protection from electric shock and fire, replacement components in mains circuits - including the power transformer, test leads,
and input jacks - must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if
the rating and type are the same. Other components that are not safety-related may be purchased from other suppliers as long as they
are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain
accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley Instruments
office for information.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply cleaner
directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis
(for example, a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions.
If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Table of Contents
SectionT opicPage
1Front and Rear Panel Operation........................................................ 1-1
Front and rear panel familiarization............................................................ 1-2
Front panel summaries......................................................................... 1-2
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
2601A SYSTEM SourceMeter
®
POWER
AUTO
CURSOR
2602A SYSTEM SourceMeter
®
POWER
AUT O
CURSOR
EXIT ENTER
TRIG MENU
SPEED
DIGITS
REL
FILTER
SRCLIMITMEAS
MODE
0
1
0000
+ / -
LOAD RUNSTORE RECALL
6
7
2
3
4
5
89
OUTPUT
ON/OFF
LOCAL
DISPLAY
CONFIG
RANGE
RANGE
1
Model 2601A and Model 2611A
2
3
4
5
EXIT ENTER
TRIG MENU
SPEED
DIGITS
REL
FILTER
SRCLIMITMEAS
MODE
0
1
0000
+ / -
LOAD RUNSTORE RECALL
6
7
2
3
4
5
89
LOCAL
DISPLAY
CONFIG
RANGE
RANGE
SPEED
DIGITS
REL
FILTER
SRCLIMITMEAS
MODE
CHANNEL ACHANNEL B
CHAN
A
CHAN
B
OUTPUT
ON/OFF
Model 2602A and Model 2612A
KEITHLEY SourceMeter
KEITHLEY SourceMeter
5
4
3
2
1
Front and rear panel familiarization
Front panel summaries
The front panels of the Keithley Instruments Series 2600A System SourceMeter® instrument are
shown in
Rear panel summaries
The rear panels of the Models 2601A/2602A and 2611A/2612A are shown in Figure 1-2. The rear
panels of the Models 2635A and 2636A are shown in Figure 1-3. The descriptions of the rear
panel components follow Figure 1-2 and Figure 1-3.
Figure 1-1
Models 2601A, 2611A, 2602A, 2612A, 2635A, and 2636A front panels
Figure 1-1. The descriptions of the front panel controls follow Figure 1-1.
1-2Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
NOTE The Models 2601A, 2611A, and 2635A have one SourceMeter
channel (Channel A) and the Models 260 2A, 2612A, and 2636A have
two SourceMeter channels (Channel A and Channel B).
1.Special keys and power switch:
DISPLAYToggles between the various source-measure is displayed and the user message
mode.
Selects Models 2602A/2612A/2636A single or dual-channel display.
CONFIGUsed to configure a function or operation.
POWERPower switch: “In” position (I) turns SourceMeter on, “out” position (O) turns
SourceMeter off (O).
Number keysThe number keys (0-9, +/-, 0000) allow direct numeric entry in the EDIT mode.
2. Source-measure setup, performance control, and special operation:
Top Row: Source-measure setup
Model 2601A/2611A/2635A and Model 2602A/2612A/2636A:
SRCChannel A: Selects the source function (V or A) and places cursor in the source field
for editing.
MEASChannel A: Cycles through measure fun c tions (V, A, Ω or W).
LIMITChannel A: Places the cursor in the compliance limit field for editing.
MODEChannel A: Directly chooses the measurement function (V, A, Ω or W).
Model 2602A/2612A/2636A only:
SRCChannel B: Selects the source function (V or A) and places cursor in the source field.
MEASChannel B: Cycles through measure fun c tions (V, A, Ω or W).
LIMITChannel B: Places the cursor in the compliance limit field for editing.
MODEChannel B: Directly chooses the measurement function (V, A, Ω or W).
Middle Row: Source-measure setup
Model 2601A/2611A/2635A and Model 2602A/2612A/2636A:
DIGITSChannel A: Changes display resolution to 4-1/2, 5-1/2, or 6-1/2 digits.
SPEEDChannel A: Sets the measurement speed by controlling the A/D converter measurement
aperture.
RELChannel A: Controls relative, which allows a baseline value to be subtracted from a
reading.
FILTERChannel A: Controls the digital filter, which can be used to reduce reading noise.
Model 2602A/2612A/2636A only:
DIGITSChannel B: Changes display resolution to 4-1/2, 5-1/2, or 6-1/2 digits.
SPEEDChannel B: Sets the measurement speed by controlling the A/D converter measurement
aperture.
RELChannel B: Controls relative, which allows a baseline value to be subtracted from a
reading.
FILTERChannel B: Controls the digital filter, which can be used to reduce reading noise.
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-3
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
Bottom Row: Source-measure setup
LOADLoads factory or user-defined scripts for execution.
RUNRuns last selected factory or user-defined scripts.
STOREStores readings, source values, and timestamp values in one of two internal buffers per
channel for later recall.
RECALLRecalls stored readings, source values, and timestamp values from either of the two buffers.
TRIGTriggers readings.
MENUAccesses the Main Menu for saving and recalling setups, selecting remote interface, line
frequency, self-tests, serial number and beeper control.
EXITCancels selection, and backs out of menu structure. Used as a LOCAL key to take the unit
out of remote.
ENTERAccepts selection, moves to next choice or exits menu.
3. Range keys:
and Selects the next higher or lower source or measure range.
AUTOEnables or disables source or measure auto range.
4. Output control and LED status indicator:
OUTPUT ON/OFFTurns source output on or off.
LED indicatorIlluminated when output is on.
5. Navigation wheel, USB port, and CURSOR keys:
When in source edit, use the CURSOR keys for cursor control and then turn the navigation wheel to
change a source or compliance value. The can also be used to enable or disable the source edit mode.
When in a menu, use the CURSOR keys or navigation wheelfor menu item cursor control. When
displaying a menu value, use the CURSOR keys for cursor control and turn the navigation wheel to
change the value. Pressing the navigation wheel opens a menu item or selects a menu option or value.
Use the USB port to connect a USB flash drive. The USB flash drive stores reading buffer data, scripts,
and user setup options.
6. Display indicators (not shown):
EDITUnit is in the source editing mode.
ERRQuestionable reading or invalid calibration step.
REMUnit in remote mode.
TALKUnit addressed to talk.
LSTNUnit addressed to listen.
SRQService request.
RELRelative mode enabled.
FILTDigital filter is enabled.
AUTOAuto source or measure range selected.
ARMUnit armed and ready to run.
* (asterisk)Readings are being stored in buffer.
1-4Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LOHIHIGGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL B
!
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S. A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
LAN
TSP-Link
R
CAT I
SS
LO LOHIHIGGGG
CHANNEL A
!
!
!
Model 2601A/2611A
Model 2602A/2612A
1
2
3
4578
9
1
610
1
2
3
4578
9
6
10
Figure 1-2
Model 2601A/2602A/2611A/2612A rear panels
Model 2601A/2602A and Model 261 1A/2612A rear panels
1. CHANNEL A and CHANNEL B (Channel B on Model 2602A/2612A only)
Input/output connections for source, sense, and guard.
2. DIGITAL I/O
Female DB-25 connector. Fourteen pins for digital input or output, one pin for output enable (2601A/
2602A) or safety interlock (2611A/2612A); +5V and GND pins are also provided.
Use a cable equipped with a male DB-25 connector (Keithley Instruments part number CA-126-1CA).
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-5
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
3. IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Keithley Instruments
Model 7007-1 or Model 7007-2.
4. Cooling exhaust vent
Exhaust vent for internal cooling fan. Keep vent free of obstructions to prevent overheating.
5. Chassis ground
Ground screw for connections to chassis ground.
6. Low noise chassis ground
Ground jack for connecting Output HI or LO to chassis.
7. RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null modem) DB-9 shielded
cable (Keithley Instruments Model 7009-5) for connection to a PC.
8. TSP-Link
Expansion interface that allows a Series 2600A and other TSP-enabled instruments to trigger and
communicate with each other. Use a category 5e or higher LAN crossover cable (Keithley Instruments
part number CA-180-3A).
9. Power module
Contains the AC line receptacle and power line fuse. The instrument can operate on line voltages of
100V to 240V AC at line frequencies of 50 Hz or 60 Hz. See the Series 2600 A Reference Manual for
line fuse replacement instructions.
10. LAN
Use this RJ-45 connector to connect the instrument to the local area network. The RJ-45 connector
connects a network card, a network switch, a router or a hub. When connecting directly to a PC, a
crossover cable (included) must be used. When connecting to a network switch, router, or hub, a normal
CAT-5 cable (not provided) should be used unless your equipment has Auto-MDIX capabilities. If it does
have Auto-MDIX, the crossover cables may be used.
1-6Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
LAN
TSP-Link
R
SENSE
LOLOHI
CHANNEL A
SENSE
HI
GUARD
A LO
!
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
SENSE
LOLOHI
CHANNEL A
SENSE
HI
GUARD
A LO
LAN
NO AUTO-MDIX
SENSE
LO
LOHI
CHANNEL B
SENSE
HI
GUARD
!
Model 2635A
Model 2636A
1
2
3
4
5611
8
910
1
2
3
4
5611
8
910
1
Figure 1-3
Model 2635A and 2636A rear panels
Model 2635A/2636A rear panels
1. CHANNEL A and CHANNEL B (Channel B on Model 2636A only)
Triax connectors for Input/output, guard, and sense connections. Use only low-noise triax cables such
as the Keithley Model 7078-TRX (available in several lengths). Connector terminals and associated
triax cable conductors are as follows:
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-7
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
2. DIGITAL I/O
Female DB-25 connector. Fourteen pins for digital input or output, one pin for safety inte rlock. Use a
cable equipped with a male DB-25 connector (Keithley Instruments part number CA-126-1CA).
3. IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Model 7007-1 or Model
7007-2.
4. Cooling exhaust vent
Exhaust vent for internal cooling fan. Keep vent free of obstructions to prevent overheating.
5. Chassis ground
Ground screw for connections to chassis ground.
6. RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null modem) DB-9 shielded
cable (Keithley Instruments Model 7009-5) for connection to the PC.
7. TSP-Link
Expansion interface that allows a Series 2600A and other TSP-enabled instruments to trigger and
communicate with each other. Use a category 5e or higher LAN crossover cable (Keithley Instruments
part number CA-180-3A).
8. Power module
Contains the AC line receptacle and power line fuse. The instrument can operate on line voltages of
100V to 240VAC at line frequencies of 50Hz or 60Hz. See the Series 2600A Reference Manual for line
fuse replacement instructions.
9. Ground / LO (Triax connector)
Channel A and Channel B low noise chassis ground triax connectors. Use only low-noise triax cables
such as the Keithley Model 7078-TRX. Connector terminals and associated triax cable connectors are
as follows:
Channel A and Channel B low-noise chassis ground Phoenix connector.
11. LAN
Use this RJ-45 connector to connect the instrument to the local area network. The RJ-45 connector
connects a network card, a network switch, a router or a hub. When connecting directly to a PC, a
crossover cable (included) must be used. When connecting to a network switch, router, or hub, a normal
CAT-5 cable (not provided) should be used unless your equipment has Auto-MDIX capabilities. If it does
have Auto-MDIX, the crossover cables may be used.
1-8Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
What are the source-measure capabilities?
Model 2601A and Model 2602A
Basic DC source-measure capabilities for the Models 2601A and 2602A are listed in Table 1-3.
Table 1-3
Model 2601A and 2602A source-measure capabilities
FunctionCapabilities
Source ±DC voltage
Source ±DC current
Measure ±DC voltage
Measure ±DC current
1. 10A maximum available in pulse mode only.
Model 261 1A and Model 2612A
Basic DC source-measure capabilities for the Models 2611A and 2612A are listed in Table 1-4.
Table 1-4
Model 2611A and 2612A source-measure capabilities
1μV to 40.4V
1pA to 3.03A
1μV to 40.8V
1pA to 3.06A
1
1
FunctionCapabilities
Source ±DC voltage
Source ±DC current
Measure ±DC voltage
Measure ±DC current
1. 10A maximum available in pulse mode only.
1μV to 202V
1pA to 1.515A
1μV to 204V
1pA to 1.53A
Model 2635A and Model 2636A
Basic DC source-measure capabilities for the Models 2635A and 2636A are listed in Table 1-5.
Table 1-5
Model 2635A and 2636A source-measure capabilities
FunctionCapabilities
Source ±DC voltage
Source ±DC current
Measure ±DC voltage
Measure ±DC current
1μV to 202V
20fA to 1.515A
1μV to 204V
1fA to 1.53A
1
1
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-9
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
How do I power-up the instrument?
Step 1: Connect to line power
CAUTIONOperating the instrument on an incorrect line voltage may cause damage
to the instrument, possibly voiding the warranty.
1.Before you plug the power cord into the Series 2600A, make sure that the front panel powe r
switch is in the off (O) position.
2.Connect the female end of the supplied power cord to the AC receptacle on the rear panel.
Connect the other end of the power cord to a grounded AC outlet.
WARNINGFailure to use a grounded outlet may result in personal injury or
death due to electric shock.
Step 2: T urn on power
Turn on the instrum ent by pressing the front panel power switch to the on (I) position. Assuming no
errors occur, the Series 2600A will power-up as follows:
1.The OUTPUT indicators and display pixels flash briefly.
2.The following items are shown in sequen ce :
•The firmware revision number.
•The line frequency.
•The TSP-Link node.
•The enabled command interface(s) and address (GPIB/LAN/RS-232).
Step 3: Set line frequency
The Series 2600A will operate at line frequencies of either 50Hz or 60Hz. For best measurement
noise performance, the unit should be con fig ur ed to match the actual line frequency used by
selecting MENU > LINE-FREQ > 50Hz,60Hz, or AUTO (when set to AUTO, the SourceMeter will
automatically detect the power line frequency at power-up).
1-10Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
How do I make measurements?
See the Series 2600A Reference Manual for more information on the following:
•Basic source-measure procedure.
•Ohms measurements.
•Power measurements.
The measurement procedure below demonstrates how to set up the Series 2600A to source 10V
and measure current through a 10kΩ resistor.
WARNINGHazardous voltages may be present on the output and guard
terminals. To prevent electrical shock that could cause injury or
death, NEVER make or break connections to the Series 2600A
while the output is on. Power off the equipment from the front
panel or disconnect the main power cord from the rear of the
SourceMeter before handling cables connected to the outputs.
Putting the equipment into standby does not guarantee the outputs
are not powered if a hardware or software fault occurs.
Step 1: Connect the DUT
Connect a 10kΩ resistor to the SourceMeter Channel A HI and LO terminals as shown in
Figure 1-4. See the Series 2600A Reference Manual for complete connection information.
Step 2: Select source and set source level
Perform the following steps to select the voltage source and set the source value to 10V:
1.Press Channel A SRC to select the V-Source as indicated by the “V” units in the source
field on the display. The flashing cursor indicates which value is presently selected for
editing.
2.Press the up RANGE key, and select the 20V source range (Model 2611A/2612A/2635A/
2636A) or 40V source range (Model 2601A/2602A). Use the lowest possible source range
for best accuracy.
3.Use the CURSOR keys to move the cursor to the 10s digit, then press the navigation wheel to enter the EDIT mode (EDIT indicator is displayed).
4.Using the navigation wheel, set the source value to 10.0000V, then press the ENTER key
or the navigation wheel.
Step 3: Set compliance limit
Perform the following steps to set the current compliance limit value to 10mA:
1.Put the Model 2602A/2612A/2636A in the single-channel display mode with the DISPLAY
key.
2.Press the LIMIT key, then press the ENTER key or the navigation wheel.
3.Move the cursor to the 10s digit, then press the navigation wheel to enter the EDIT mo de.
4.Using the navigation wheel, set the limit value to 10.000mA, then press the ENTER key or
the navigation wheel.
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-11
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
Step 4: Select measurement function and range
1.Put the Model 2602A/2612A/2636A in the single-channel display mode with the DISPLAY
key, then select the current measurement function by pressing MEAS or MODE.
2.Select the measurement range with the RANGE keys. For the purposes of this example,
press AUTO RANGE, and note that the AUTO indicator is displayed. The instrument will
automatically select the best range based on the measured value. You can also use manual
ranging by pressing the up or down RANGE key, but be sure to use the lowest possible
range for best accuracy.
Step 5: T urn output on
Turn the output on by pressing the appropriate OUTPUT ON/OFF key. The
ON/OFF indicator LED will turn on when the channel output is on.
NOTE The Models 2611A, 2612A, 2635A, and 26 36A are equipped with a
safety interlock circuit that prevents operation on the 200V range if
the circuit is not energized. This circuit is illustrated in
to the Series 2600A Reference Manual for more information.
Figure 1-5. See
Step 6: Make measurements
1.Observe the readings on the display (press TRIG if necessary to trigger the unit to begin
taking readings). For the single-channel display mode, the readings will appear on the top
line, while source and limit values are on the bottom line. For the 10kΩ resistor under test,
typical display values are:
1.00000mA
SrcA: +10.0000 V LimA:010.0000mA
2.Use the DISPLAY key to cycle through the various display modes shown in Figure 1-6 (the
User State display messages are defined with specific display commands; see the Series
2600A Reference Manual for more information).
3.Press the MEAS key several times to display measured voltage, resistance, and power.
Typical values for the 10kΩ resistor are: 10.0000V, 10.0000kΩ, and 10.0000mW.
Step 7: T urn output off
When finished making measurements, turn the output off by pressing the OUTPUT ON/OFF key.
The OUTPUT indicator light will turn off.
1-12Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
FVR
Chassis
Series 2600A
Signal
Ground
Chassis
Signal Ground is a local signal ground and defined as
the Low-Noise Chassis Ground Banana Jack.
Chassis is defined as the metal chassis of the Series 2600.
1) Frequency Variable Resistor (FVR) Isolates
the SMUs from high frequencies on the chassis.
For DC to 60Hz, the FVR is a virtual short (zero
ohms).
2) DO NOT use the Chassis Screw terminal to
make signal connections to external circuitry.
High frequency (>1MHz) on the chassis may
result in higher noise at the output.
Chassis Screw terminal is connected to the metal chassis of the
Series 2600.
1
Chassis
Screw
2
Low-Noise
Chassis Ground
Banana Jack
Signal
Ground
Model 2636A
Figure 1-4
Model 2602A/2612A low-noise chassis ground banana jack and chassis screw
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-13
Section 1: Front and Rear Panel OperationSeries 2600A System SourceMeter® Instruments User’s Manual
To output stage
10kW
Rear panel
Chassis ground
Coil resistance
145W +/- 10%
-220V supply
+220V supply
INTERLOCK pin
(on DIGITAL I/O connector)
Read by firmware
W
User State
1.00000mA . V
SrcA:+10.0000V SrcB:+000.000mV
1.00000mA
SrcA:+10.0000V LimA:10.0000mA
. V
SrcB:+000.000mV LimB:100.000mA
Press DISPLAY key
Press DISPLAY key
Press DISPLAY key
Press DISPLAY key
Source-Measure display for SMU A and SMU B:
Top line displays the measure function (V, A, W or W).
Bottom line displays the source function (V or A)
and level.
Source-Measure and Compliance Limit display for SMU A:
Top line displays the measure function (V, A, W or W)
Bottom line displays the source function (V or A) and level,
and the compliance limit (A or V).
Source-Measure and Compliance Limit display for SMU B:
Top line displays the measure function (V, A, W or W)
Bottom line displays the source function (V or A) and level,
and the compliance limit (A or V).
Display for user-defined messages and prompts.
Indicates that a measured reading has not been
triggered.
Figure 1-5
Interlock circuit
Figure 1-6
Display modes
1-14Return to Section Topics2600AS-900-01 Rev. B / September 2008
Series 2600A System SourceMeter® Instruments User’s ManualSection 1: Front and Rear Panel Operation
How do I use the reading buffer?
The instrument has two dedicated reading buf fers per channe l that can store over 60,00 0 readings
with the time stamps and source values options enabled. Disable the time st amps and source
values options to store over 140,000 readings.
The following example shows how to store 100 readings, source values, and timestamps in
Channel A, Buffer 1 and recall them from the front panel.
Step 1: Connecting the device under test (DUT)
Connect a 10kΩ resistor to the Channel A HI and LO terminals (see Figure 1-4).
Step 2: Set up source and measure functions
Using the procedure described in How do I make measurements?, set up source and measure
functions:
•Source function: Volts.
•Source range: 20 V (2611A/2612A/2635A/2636A) or 40V (2601A/2602A).
•Source value: 10 V.
•Measure function: Current.
•Measure range: Auto.
Step 3: Configuring the reading buffer
Complete the following steps to configure reading buffers from the front panel:
1.Press CONFIG > STORE and then choose one of the following:
•CHANA-BUFF
•CHANB-BUFF
2.Select the DEST option and then choose one of the following :
•CHANX-BUFF1
•CHANX-BUFF2
•NONE
3.Select BUFFER1 or BUFFER2.
4.(Optional) To clear the buffer, turn the navigation wheel to select CLEAR > YES.
5.Turn the navigation wheel to select ELEMENTS.
NOTE You must clear the reading buffer before you enable or disable the
source value or the time stamp options.
6.(Optional) Push the navigation wheel to select TSTAMP, then select OFF or ON.
7.(Optional) Turn the navigation wheel to select SRC-VAL, then select OFF or ON.
8.Press the EXIT key to return to the main menu.
2600AS-900-01 Rev. B / September 2008Return to Section Topics1-15
Loading...