Keithley 263 User Manual

Model 263
Quick Reference Guide
Calibrator/Source
INTRODUCTION
This quick reference guide contalne descriptions of parlous features and Information concerning the operation of the Model 283. Also included are programming examples using VBrlous COntr011wa.
Instruments Division
Document Number: 283-903-01 Rev. A
Cleveland. Ohio, U.S.A.
TABLE OF CONTENTS
SAFETY PRECAUTlONS 3 FRONT PANEL OPERAflON 4 CONNECTORS IRear Panell.. 9 PROGRAMMING OUTPUT VALUES 10 SOURCING METHODS 11 ; ~ GUARD 13 ‘~I FRONT PANEL PROGRAMS. 16 i IEEE-488 PROGRAMMING.. 17 PROGRAMS 23
IBM PC or AT ICEC PC < > 488 Interface1 24 IBM PC or AT ,National GPIS PC Interface,. 26 APPLE II 28
HP85................................ 30
HPSERlES300and.9816 ,,,,,,,,,,,,,.,,, 32 ; j I
HP9925A............................. 34 :
DECLSIll ,...,,.,,,,,,,,,,,o,,,,.,,, 36 i ,! _
: j
i i
SAFETY PRECAUTIONS
1. Before operation, ground the instrument through a prop­erly earth grounded power receptacle.
2. Before sevlcing, disconnect the instrument from the power line and all other equipment, and consult the Model 263 InstructIon Manual.
3. Do not touch any terminals while the Instrument Is oon­nectsd to any other test equipment.
3
FRONT PANEL OPERATION
I
POWER-Controls AC line power to the instrument.
SHIFT-Several controls have secondary functions that are selected by first pressing SHIFT and then pressing the ap- 1 : propdate button.
i ;
I
~I
1~~ >
: I
,
FUNCTION GROUP
FUNCTION
OHM9
o=
VOLTS EX
OmH
AMPS
f=-l
OYA V/R m
OB-'
OHMS-Pressing the OHMS button configures the Model 263 as a resistance source. The value of the reslstanoe will be displayed.
VOLTS-Pressing the VOLTS button conflgures the Model 263 88 a voltage 80”~~ The voltage VBIUB will be displsyed.
Pressing SHIFT VOLTS configures the Model 263 to output an external wurce applied to the EXT INPUT and COMMON terminals on the rear panel. The EXT indicator turns on and the message “USE, V” 18 displayed.
AMPS-Pressing the AMPS button configures the Model : 263 as an active current source. The programmed current value will be dlsplayed.
Pressing SHIFT AMPS turns on the V/R light indicating that the Model 263 is configured as a passive current aouroe.
COUL-Pressing the COUL button conflgures the Model 263 8s an active charge 8ouroe. The currently selected : range and programmed charge value will be displayed.
Pressing SHIFT COUL turns on the VIR light indicating that
the Model 263 is configured as a passive charge source.
CONTROLGROUP
I
’ ’
6
RANGE-These two buttons are used to swleot the range of the selected function. RANGE A upranges the Instru­ment and RANGE V downranges.
CURSOR-These two buttons are used to position the our­sot when using the Adjust Method to modify an output valuw. CURSOR C moves the cursor from left to right and conversely, CURSOR 4 moves the cursor from right to left. Not functional In OHMS.
Pressing SHIFT ON/OFF the first time displays the flashing cursor continually. Pressing It again disables the cursor. Not functional in OHMS.
ADJUST-These two buttons aro used to modify the display reading using the Adjust Method. ADJUST A increments the reading at the seleotnd dlglt and ADJUST T decrements the reading. Not funotional in OHMS.
KEYPAD/ENTER-Pressing this button enables the keypad allowing an output value to be keyed in using the number
10 through 91 buttons and the + button IKeypad Method). Pressing this button *gain enters the resdlng and disables the keypad. Not functional in OHMS.
SHIFT CANCEL-Pressing this button after keying in B reading (Keypad Method) canoels that reading and disables the keypad.
+/- Toggles the displayed reading (except OHMS1 bet­wsen posltlve (t) and negative l-l polarity. Not functional in OHMS.
ZERO-Toggles the display between zero and the previously displayed reading.
SHIFT GUARD-Toggles the output configuration between
guarded and unguarded.
7
MENU-Used to Sslsot front panel program& Program parameters are modified by the ADJUST buttons. Pressing MENU after modifying a program parsmeter enters the new parameter and returns the instrument to normal operation.
OPERATE-Toggles the instrument between the standby and operate conditions.
8
CONNECTORS (Rear Panel)
OUTPUT-Output high is available at this triax connector
ion the center conductor). Output low is available here Ion the inner shield) or at the COMMON terminal depending an the configuration of the output kmguarded or guarded).
EXT INPUT-This banana plug is used to connect an exter-
nal input (2OOV. lOOmA peak1 to the Model 2133. The ex-
ternal Input is then available at ths output of the Model 283
when VOLTS EXT is selected.
COMMON-Output low Is always avsllable at this termlnsl
regardless of the oonflguration of the output.
PREAMP OUT-This terminal along wlth COMMON can be used to monitor the voltage drop across a load when the instrument is sourcing current or charge. This termlnsl can also be used as an external guard drive when sourcing AMPS or COUL. PREAMP OUT is inoperative in AMPS V/R. COUL V/R, VOLTS and OHMS.
IEEE CONNECTOR-Used to connect the instrument to the IEEE-488 bus. IEEE interface functions are marked above the connector.
AC RECEPTACLE-Power is applied through the supplied power cord to the 3-terminal AC receptacle.
9
PROGRAMMING OUTPUT VALUES
_
For all functions, except OHMS, the following two data entry methods can be used to program the instrument for 8” out­put value. In general, use the Adjust Method to modify B value and the Kevpad Method to enter a completelv new v.aIuB.
Adjust Method
1. Position Cursor. Using CURSOR 4 and CURSOR C, place the cursor on the digit to be changed. To continuallv : display the cursor location, press SHIFT ON/OFF.
2.Adjust Display Reading. Use thB ADJUST A and ADJUST
v to increment or decrement the reading. Polarity of the
reading is toggled by the +/- button.
Keypad Method
1. Enable Keypad. Enable the keypad by pressing KEYPAD/ ENTER. Keypad LED turns on.
2.Key In Reading. Key in the reading by pressing the BP­propriate number button (0 through 91 at each oursor location. Polarity is toggled by the +/- button.
Note: If it is desired to cancel the keyed in reading and return to the previously programmed value, press SHIFT CANCEL. The keypad will disable.
3.Enter Reading. Again press KEYPAD/ENTER to enter the displayed reading and disable the keeypsd. Keypad LED turns off.
: ;
,~ I I
I
~
‘-
:
:
10
SOURCING METHODS
Figures 1 through 6 show the basic methods that the Model
263 ums to source resistance, voltage, current and charge.
MODEL 263
_ _ _ _ _ _ - - _ _ _ - - - - - 1
Flgure 1. Ohms Source
Flgum 2. Volts Source
11
[ .j
Flgum 3. External Source
MODEL 263
r___-_____-_-_
L- _-__
---------,
Flgure 4. AMPS V/A and COUL V/A
Flgum 5. Amps and Coul Source
12
j !
l-=--t--
,
GUARD
In the normal unguarded output con‘iguratlan, o”tp”t high is connected to the center conductor of the triax OUTPUT connector and outp,” low is connected to the Inner shield Isee Figure 8). When the o”tput of the Mods1 203 18 plac­ed I” the guarded conflguration, by pressing SHIFT GUARD,
output low Is disconnected from the inner shield of the OUT.
PUT connector ellowlng a guard drive to be connected to it. In AMPS and COUL the guard drive is provided by the Model 283. On the other functions (AMPS V/R, COUL V/R, OHMS and VOLTS), the inner shield of the triax OUT­PUT connector Is floating unles8 a guard drive is supplied by the user. Figures 7 through 10 show the guarded out­put configurations for all functions.
Guarding 18 recommended for r~slstancw8 2 100Mg. Gusrd­ing is not necessary when the Model 263 is sourcing cur­rent to a feedback picoammeter since this oircult is already guarded. The low Input voltage burden of a feedback pico­mmeter results with virtually the some voltage 0” output high and low. Guarding Is recommended when sourcing cur­rent or charge to a high impedance load. Guarding Is un­“~CBSSB,~ when sourcing voltage.
0ulPU-l HI LO
3
COMMON
Flgure 6. Normal (Unguarded) Output ConfIguration
13
Flgure 7. Guarded Ohms Output Conflgurstlon
OUTPUT
:----_HL-
r----
$t_l
It--
Flgum 8. Guarded AMPS WA and guarded COUL V/R
output Co”llguratlo”
L ~~’
Flgum S. Gusrdsd Amps and guarded Coul Output
Co”flguratlon
HI
OUTPUT
+
V.
c
I.0
Figure 10. Guarded VOLTS OUtput Conflgumtlon
16
FRONT PANEL PROGRAMS
The available front panel programs are summarized in Table 1. Perform the following steps to use B program:
1. Display the desired program using the MENU button.
2. Use the ADJUST A or ADJUST 7 button to change the program pammeter.
3. To enter the new program paramter, press the MENU but­ton again. The instrument will return to normal operation.
Note: The IEEE address fProgram IEEE1 and the display mode IProgram dlSPl are stored in memory. Thus, the :~~ i instrument will power-up to the programmed parameters. Temperature compensation status is not stored and will I always power up enabled Ill.
Table 1. Front Panel Programs
Check or change IEEE-488 bus address 10
Selects alternate display mode; engineer­ing units lpi or scientific notation l-61.
Check or change temperature compensa-
IEEE-488 PROGRAMMING
DEVICE-DEPENDENT COMMANDS
EKeCUte X
Execute other device-dependent commands
17
Ohms Amps
Auto on Auto on A”,0 0” A”,0 0” fluto cl”
to k”
lock”
lcJM” 20llA 20 v 200°C WGL
lOOM” 200 nA 20 v
IOG”
lOOGO 1000” IOOG” 20mA 20 V ZOOtiC ,OOGL
2 pA 2OLh” 2opc
1 k”
20 PA
200 PA 20 ” 2°C
tfm 2llA 20 ”
1G” 2pA 20 ”
2OpA 20 ” 20qk IOOGL
200 pA 20 V 200,X
2mA 20 ” ZOO& IOOGL
A”*0 Off A”,0 Off Auto Off Auto Off A”to Off
zero
Zero Off
:7
Zero on
Temperature Compensation co
c,
Temperature compensation off Tempsrsture ccmp~n~aticn on
Quml wo
Wl
Guard Off
Guard On
vcns CO”, Ladder
IML
2 v 2oopc
20°C
2O#C lOOGL
m”ll.
,CCML
1GL
2pc 1OOGL
1OOGL
~ I
( I
18
:, j ~
Self-Te@l JO Perform self-test
EOI KO
Kl
Enable EOI Disable EOI
SRO
Disable IYGi Ml6 M32
Charge done
Resdv
ENC,
Ccdlbratlon Value
An
1 Calibrate using value
19
Calfbmllon Value Lfl
1 Calibrate low temperature cal point
output Value V” Program output using value
PI&l GO
Gl
Reading with prefix Reading without prefix
operate 00 Standby
: 01 operate
20
DATA FORMAT
SRQ MASK end STATUS BYTE FORMAT
21
22
PROGRAMS
23
IBM PC or AT
(CEC PC < > 488 Interface)
The following program sends e command string to the Model 263 from an IBM PC or AT computer end displays the instrument reading string on the CRT. The computer tnuet be equipped with the CEC lnterfaos and the DOS 2.00 lx later revisicnl cpereting system.
DIRECTIONS
1. Using the front panel prcgrem feature, set the primary address of the Model 263 to 9.
2. With the power off, connect the Model 263 to the IEEE-499 interface installed in the IBM computer.
3. Type in SASICA on the computer keyboard to get into
the IBM interpretive BASIC Ienguage.
4. Enter the lines below using the return key after eech line.
6. Run the program and type in the desired command string. For example, to place the instrument in the current func-
tion end ZOO,& range, type in FIRSX and press the return
key.
6. The instrument reading string will then appear on the dlsplsy.
24
PROGRAM
COMMENTS
to OEFSEC.&HC408
’ Memory address. 20INtTIALIZE=O 30SEt40=9:ENTER=21 ’ Routine addresses. 40 MY. RDDRESS%=2I
’ PC < > 468 address. 50 INST. ADmESS%= ’ 263 address.
System ccntmller. ln1t1el1ze evetetn.
Prompt for string. Send commend string.
Meke room for dete.
Get readlna from 263.
NOTE: For conversion to numeric variable, make the fcllcw­lng changes:
26
IBM PC or AT
(National GPIB PC Interface)
The following program sends a command string to the
Model 263 from an IBM PC or AT computer and displays
the instrument reading string on the CRT. The computer
must be equipped with the National GPIB PC Interface and the DOS 2.00 (or later revision) operating system. Interface software must be Installed and configured as described in the instruction manual.
1. Using the front panel program feature, set the primary address of the Model 263 to 9.
2. With the power off, connect the Mods, 263 fo the
IEEE-489 interface installed in the IBM computer.
3. Type in BASICA on the computer keyboard to get into the IBM interpretive BASIC language.
4. Place the interface software disk in the default drive, type
LOAD”DECL”, and press the ret”,” key.
5. Add the lines below to lines l-6 which are now in memory. Modify the address in lines 1 and 2, as described in the National instruction manual.
8. Run the program and type in the desired command string. For example, to place the instrument in the current func­tion and 200@ range, type in FlR9X and press the return key.
7. The instrument reading string will then appear on the display.
8. Ta exit the program, type in EXIT at the command prompt and press the return key.
26
PROQRAM
COMMENTS
Set primary sddress to 8.
50U%.RH182’ChLL IBPOKE
Set timeouts.
<BRDB%!U%>
60 w=t :cfiLL IBSRE
Set REN We.
CBRDO%,U%>
70 INPUTS 1 CCNMRND
Prompt for command.
STRIIG” icm*
80 IF cm$=* 1 EXIT” THEN See If program Is to be
150
90 IF c1,**=
“ 3, THEN,‘B
halted. Check for null input.
108 CRLL IBHRT~W~~%~CMD~~ Address 263 to listen,
send string.
,,0RLw=SPRCE*~LBO>
Define reading input
buffer.
120 CALL IBRD<11263%,ROB~
Address 263 to talk, get
reading. 130PRINTRW 140 GOT0 70 150 I,%iBi CclLL IB”l,lL
Display the string.
Repeat.
Close the instrument file.
<P,263%,u%)
160 WLL ,BONL<BRDO%P u%> Close the board file.
L70 END
NOTE: For conversion to numerio varlsble, make the follow­ing changes:
APPLE II
(Apple II IEEE-488 interface)
The following program sends a command string to the ; Model 263 from an Apple II computer and displays the in­strument reading string on the computer CRT.
The computer must be equipped with the Apple II IEEE-466 Interface installed in slot 3.
DIRECTIONS
1. Using the front panel program feature. set the primary 1~ 1 ‘: address of the Model 263 to 6.
2,With the power off, connect the Model 263 to the j !
IEEE-466 interface installed in the Aoole II computer.
3. Enter the lines in ths program below, &ins the FliTURN key after each line.
4. Run the program and tvps in the desired command String at the command prompt. For example, to place the in­strumsnt in the current function and 200pA range. type in FtRgX and press the return key.
6. The instrument reading string will then appear on the CRT.
26
;
_~I
i,
PROORAM
COMMENTS
18Z*=CHRP<26>
20 INPUT ‘COPIIIH,~ID
STR,NG”iB* 30 PM3 40 IN”3
50PRINT“KA” 60PRINT”WT~“iZ*iB* Address 263 to listen,
70PRINT “LFI” 80PAlNT ~‘RDH”iZ*i: Address 263 to talk,
IIIPUT“ 3 ‘iA
98PRIHT1’UT”
160 PR#O 116 IN”6
120PRlHTAb
130 CDT0 20
Terminator. Prompt for and enter string.
Set output tom IEEE bus. $ne input from IEEE
Enable remote. send string.
Line Feed on.
G%: Untalk the 263.
Define o”tp”t to CRT. Define Input from
keyboard.
Display reading string. Repeat.
NOTE: If conversion to numeric variable is required. make the following changes:
29
HP 85
The following program sends a command string to the : ~ Model 263 fraom an HP-66 computer and displays the in­strmvwnt reading string on the computer CRT. The oom­puter must be equipped with the HP82937 GPIB Interface : : and an l/O ROM.
DIRECTIONS
I. Using the front panel progrsm feature, sot the primary
address of the Model 263 to 8.
2. With the power off, connect the Model 263 to the : ~. HP82937A GPl6 interface installed in the HP-66 oornputer,
3. Enter the lines in the program below, using the END LINE ‘~ key after each line.
4. Press the HP-86 RUN key and type in the desired corn- ’ ’ msnd string at the command prompt. For example, to place the instrument in the current function and 200~A ; ~ range, type in FlASX and press the END LINE key.
6. The instrument reading string will then appear on the in j CRT.
( 1
30
PROGRAM
COMMENTS
Prompt for command.
Input command string. Address 283 to listen. send strln
Address 2
put reading.,
:: 3 to talk, In-
R;;isp;y readlng string.
31
HP SERIES 300 and 9818
The following program sends a command string to the
Model 263 from a Hewlett-Packard Ssries 300 or Model
9816 computer and displays the inwument reading strina on the computer CRT. The computer mwt be equipped with the HP62937 GPIB Interface and SASICA 4.0.
DIRECTIONS
1. Using the front panel program feature, set the primary address of the Model 283 to 8.
2. With the power off, connect the Model 203 to the HPS2937A GPIB interface installed in the computer.
3. Enter the lines In the program below, using the ENTER key after each line.
4. Press the RUN key end type in the desired command
string at the command prompt. For example, to place tha
instrument in the current function and 200~A range, type I” FlA9X and press the ENTER key.
6. The instrument resdlng string will then appew on the CRT.
32
PROGRAM
COMMENTS
Place 263 I” r*rnct*.
Prompt for and Input command. Address 203 to listen, send string.
Address 263 tc talk. In-
put readlng.
Display reading string.
Repeat.
NOTE: For ccnverslon tc a numeric variable, change the prc­gram as follows:
33
HP 9825A
Use the fcllowlng~prcgrsm to send a command string to the Model 263 from a Hewlett-Packard Model 9826A and display the instrument reading string on the computer printer. The computer must be equipped with the HP99034A HP16 Interface and a 9972A extended I/O ROM.
DIRECTIONS
1. From the front panel, set the primary address of the Model 263 to 13.
2.With the power off, connect the Model 263 to the 99034A HPIB interface installed in the 992SA.
3.Enter the lines in the program below, using the STORE
key after each line Line numbers are automatically assign­ed by the 9926A.
4.Press the 982SA RUN key and type in the desired ccm-
msnd string at the command prompt. For example, to place the Instrument in the current function and ZOOpA range. type in FlR9X and press the CONT key.
&The instrument reading string will then appear on the
computer print cut.
34
PROGRAM
COMMENTS
0 dim AX251 9 LW201
1dev”263”,708
2 rernl1263’)
3 env ~COMIIAND STRING” I
7 gtc 3
NOTE: For conversIon to numerlo variable, modify the prc­gram 88 follows:
Olmensicn data strings. De,,ne 283 at address Place 263 in remote. Prompt for command string.
Address 263 to lIsten, send 8trlng. .44ry 263 to talk, input
Print data strlnp on printer. ­Repeat.
8.
6 prt val(00[51)
36
DEC LSI 11
The following program sends a command string to the : Model 283 from a DEC LSI 11 minicomputer and displays the instrument reading string on the DEC CRT terminal. The
LSI 11 must be configured with IBK words of RAM and 8” 1~ IBV 11 IEEE-489 interface. The software must be configured with the IB software as well 88 FORTRAN and the AT 11 i
operating system.
1. Using the front panel program feature, set the primary i address of the Model 263 to 8.
2,With the power off, connect the Model 263 to the IBV I
11 IEEE-488 interface cable.
3. Enter the program below, using the editor under FIT 11 i and the name IEEE.FOR.
4.Comcile using the FORTRAN ccmeilsr as follows: FOR­TRAN IEEE. -
6.Link with the svstern and IB libraries as follows: LINK IEEE,IBLIB.
&Type RUN IEEE and press the RETURN key.
7.The display will read “ENTER ADDRESS”.
B.Typs in B and press the RETURN key.
9.The display will read “TEST SETUP’:
10. Type in the desired command string and press the RETURN key. For example, to program the instrument
for the current function and 200~A range, type in FlRSX and press RETURN.
11. The Instrument data string will appear an the computer i
display.
~
36
PROGRAM
COMMENTS
DO2I=I,lB CALLIBSTER<IIFl)
2 CONTINUE
CALLIBSTER<i5,5> C*LLIBTtMc~t20~
PRIADR)
18 I=,8REC”<IHPlJT,80r
PRIADR)
IHPUT~I+L~=O
CRLLPUTSTR<7rINPUTl
‘6’) COLL lBUNT GOT0 12 END
Turn off IE errors. Allow 6 error 16’s.
Allow 1 second bus dmsc”t. Set line feed as terminator. lbrn on remote.
Input primary address.
send 8trlng. Get data from instrument.
Untslk the 263. Repeat.
37
Loading...