Keithley 2635B, 2634B, 2636B, 2601B, 2602B Quick Start Manual

...
Page 1
Safety
Introduction
Models 2634B, 2635B, and 2636B
System SourceMeter Instruments
Quick Start Guide
Page 2
Safety
Introduction
Safety precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualied personnel who recognize shock hazards and
are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product.
Refer to the user documentation for complete product specications.
If the product is used in a manner not specied, the protection provided by the product
warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance
of equipment, for ensuring that the equipment is operated within its specications and
operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety
procedures and proper use of the instrument. They must be protected from electric shock
and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating
properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair
products. Only properly trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient
overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to mains. These instruments will be
marked as category II or higher.
Unless explicitly allowed in the specications, operating manual, and instrument labels, do not
connect any instrument to mains.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test xtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30 V
RMS, 42.4 V peak, or
60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible
body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product
operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 V, no conductive part of the circuit
may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains.
When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear,
cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack
mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while
power is applied to the circuit under test. ALWAYS remove power from the entire test system
and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing
or removing switching cards, or making internal changes, such as installing or removing
jumpers.
Page 3
Safety
Introduction
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the
instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and re, replacement components in mains circuits — including the power transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national safety
approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the applicability of
a replacement component, call a Keithley Instruments ofce for information.
Unless otherwise noted in product-specic literature, Keithley instruments are
designed to operate indoors only, in the following environment: Altitude at or below
2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F) ; and pollution degree 1 or 2.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the
exterior of the instrument only. Do not apply cleaner directly to the instrument or allow
liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Safety precaution revision as of March 2016.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while
standing on a dry, insulated surface capable of withstanding the voltage being measured.
When fuses are used in a product, replace with the same type and rating for continued
protection against re hazard.
For safety, instruments and accessories must be used in accordance with the operating
instructions. If the instruments or accessories are used in a manner not specied in the
operating instructions, the protection provided by the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are dened in the specications and operating information and shown on the instrument panels, test xture panels, and switching cards.
Chassis connections must only be used as shield connections for measuring circuits, NOT
as protective earth (safety ground) connections.
If you are using a test xture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid interlock.
If a screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
The symbol on an instrument means caution, risk of danger. The user must refer to the
operating instructions located in the user documentation in all cases where the symbol
is marked on the instrument.
The symbol on an instrument means caution, risk of electric shock. Use standard
safety precautions to avoid personal contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
Introduction
Safety
Unpack
Connect Test
FAQs
Next Steps
Page 4
*See full specications for other pulse regions.
Carefully consider and congure the appropriate
output-off state, source levels, and compliance
levels before connecting the instrument to a device that can deliver energy. Failure to consider the output-off state, source levels, and compliance levels may result in damage to the instrument or to the device under test.
Power and environmental specications
For indoor use only.
Power Supply 100 VAC to 240 VAC, 50 Hz to 60 Hz
(auto sensing)
Maximum VA 240 VA
Operating altitude Maximum 2000 m (6562 feet) above
sea level
Operating
temperature
0 °C to 50 °C (32 °F to 122 °F), 70% relative humidity up to 35 °C. Derate 3% relative humidity, 35 °C to 50 °C
Storage temperature 25 °C to 65 °C (77 °F to 149 °F)
Pollution degree 1 or 2
DC source output 30.3 W maximum
±200 VDC maximum ±1.515 A at ±20 VDC ±101 mA at ±200 VDC
Pulse region 4* Region maximums: 10 A at 5 V
Maximum pulse width: 1 ms
Maximum duty cycle: 2.2%
Measurement input Measurement category I
Voltage: 200 VDC maximum HI to LO, 250 VDC LO to ground Current: 101 mA at 200 V;
1.515 A at 20 V Impedance: Variable
Page 5
Safety
Introduction
Software for the Series 2600B instruments is available for download
from the Keithley web page. Go to www.tek.com/keithley and
search for the specic software you need.
Test Script Builder (TSB) Integrated Development Environment: Provides an environment to develop a test program and ability to load the test program onto the instrument.
IVI instrument driver for National Instruments LabVIEW™ and related release notes.
J2SE™ Runtime Environment.
Keithley I/O layer and release notes
Introduction
The Keithley Instruments Series 2600B System SourceMeter® Instruments are medium-power source-measure units (SMUs)
that simplify test processes by combining source and measure
capabilities in a single instrument. A Series 2600B SourceMeter
Instrument is a scalable, high-throughput, cost-effective solution
for precision DC, pulse, and low-frequency AC source-measure
testing.
Complete documentation for the Series 2600B instruments is
available for download on the Keithley web page. Go to
www.tek.com/keithley and search for Series 2600B manual.
The Series 2600B documentation includes:
Quick Start Guide: Provides unpacking instructions, describes basic connections, and reviews basic operation information.
Reference Manual: Provides comprehensive information about the instrument’s features, operation, optimization, maintenance, troubleshooting, and programming commands
Safety
Introduction Unpack
Connect Test
FAQs
Next Steps
Page 6
Unpack and inspect the instrument
To unpack and inspect the instrument:
1. Inspect the box for damage.
2. Open the top of the box.
3. Remove the documentation and accessories.
4. Carefully lift the instrument out of the box.
5. Remove the packaging insert.
6. Inspect the instrument for any obvious signs of physical damage. Report any damage to the shipping agent immediately.
Page 7
Safety
Introduction
In addition to the Series 2600B System SourceMeter® Instrument,
you should have received:
1. Keithley USB ash drive containing the Model 2400 software
emulation personality script.
2. Interlock DB-25 male connector kit hardware.
3. One RJ-45 LAN crossover cable for the 2636B. Two RJ-45 LAN crossover cables for the 2634B and 2635B.
4. Power line cord.
5. One triaxial to alligator clip cable for the 3635B. Two triaxial
to alligator clip cables for the 2634B and 2636B.
6. 16 AWG black grounding wire.
7. 3-pin screw terminal connector.
8. UL safety supplement (not shown).
9. Safety Precautions (not shown).
10. 2634B, 2635B, and 2636B System SourceMeter Instruments Quick Start Guide (this document; not shown).
Refer to the packing list for additional items that might have
shipped with your instrument.
TestConnect
Safety
FAQs
Next Steps
Introduction
Unpack
Page 8
Make sure any test xture connected to the system protects the operator from contact with hazardous voltages, hot surfaces, and sharp objects. Use shields, barriers, insulation, and safety
interlocks to accomplish this.
Cover the device under test (DUT) to protect the operator from ying debris in the event of a system or DUT failure.
Double-insulate all electrical connections that an operator can touch. Double insulation ensures the operator is still protected even if one insulation layer fails. Refer to IEC 61010-1 for specic requirements.
Make sure all connections are behind a locked cabinet door or other barrier. This protects the system operator from accidentally removing a connection by hand and exposing hazardous
voltages. Use high-reliability fail-safe interlock switches to disconnect power sources when a test xture cover is opened.
Where possible, use automatic handlers so operators are not
required to access the DUT or other potentially hazardous
areas.
Provide training to all users of the system so they understand
all potential hazards and know how to protect themselves from
injury.
In many systems, during power up, the outputs may be in an
unknown state until they are properly initialized. Make sure the
design can tolerate this situation without causing operator injury or hardware damage.
Connect the instrument
Important test system safety information
This product is sold as a stand-alone instrument that may become part of a system that could contain hazardous voltages and energy sources. It is the responsibility of the test system designer, integrator, installer, maintenance personnel, and
service personnel to make sure the system is safe during use and is operating properly. You must also realize that in many
test systems a single fault, such as a software error, may output hazardous signal levels even when the system indicates that there is no hazard present. It is important that you consider the following factors in your system design and use:
The international safety standard IEC 61010-1 denes
voltages as hazardous if they exceed 30 V
RMS and
42.4 V peak, or 60 VDC for equipment rated for dry
locations. Keithley Instruments products are only rated for dry locations.
Read and comply with the specications of all instruments in the system. The overall allowed signal levels may be constrained by the lowest rated instrument in the system.
For example, if you are using a 500 V power supply with a 300 VDC rated switch, the maximum allowed voltage in the system is 300 VDC.
Page 9
Safety
Introduction
Install the instrument
The 2600B can be used on a bench or in a rack. Please see the instructions that came with your rack-mount kit if you are installing the 2600B in a rack.
Connect line power
The Series 2600B operates from a line voltage of 100 V to 240 V at a frequency of 50 Hz or 60 Hz. Line voltage is automatically sensed (there are no switches to set). Make sure
the operating voltage in your area is compatible.
To keep users safe, always read and follow all safety
warnings provided with each of the instruments in your system.
The power cord supplied with the Series 2600B contains a separate ground wire for use with grounded outlets. When proper connections are made, the instrument chassis is connected to power line ground through the ground wire in the power cord. Failure to use a grounded outlet may result in personal injury or death due to electric shock.
TestUnpack
Safety
FAQs
Next Steps
Introduction
Connect
Page 10
To connect line power:
1. Make sure the front panel power switch is in the off (0)
position.
2. Connect the socket of the supplied power cord to the power connection on the rear panel.
3. Connect the plug of the power cord to a grounded AC outlet.
Operating the instrument on an incorrect line voltage may cause damage to the instrument, possibly voiding the warranty.
Turn on the instrument
Turn on the instrument by pressing the front panel POWER switch to the on (|) position.
Page 11
Safety
Introduction
Power-up sequence
When the instrument is turned on, you should see:
A series of dots.
All segments of the display light.
A brief display showing the instrument model. For
example, if the instrument is a 2635B, the display will show “KEITHLEY Model 2635B.”
Line frequency detection and other startup checks.
The entire power-up process takes approximately 30 seconds to
complete. When initialization is complete, if the instrument is a
2634B or 2636B, you will see the default display screen, shown
below.
If the instrument is a 2635B, you will see the default display
screen, shown below.
TestUnpack
Safety
FAQs
Next Steps
Introduction
Connect
Page 12
Test the instrument
The following test veries basic operation of the 2634B, 2635B, and 2636B. In this test, you will use the front-panel controls
shown below to source voltage and measure the voltage output.
You do not need to connect a device under test (DUT) for this
test.
Step 1: Set source function, range, and level
1. Press the SRC key. You will see a blinking character in the SrcA value eld. Conrm that mV is displayed; if not, press the SRC key again.
2. While that character is still blinking, press the up or down RANGE keys until 20 V is displayed.
If the instrument has two channels (2634B or 2636B) and
is in dual-channel display mode, press the DISPLAY button
once to show only the settings for SMU channel A.
Page 13
Safety
Introduction
The main display screen reappears:
3. Press the CURSOR keys to move the cursor to the 10s digit.
4. Press the navigation wheel to enter EDIT mode. The EDIT indicator appears in the upper left corner of the display.
5. Turn the navigation wheel to set the source value to
20.0000 V, and then press the navigation wheel to enter
the selection and exit EDIT mode.
Step 2: Set the source limit
1. Press the LIMIT key. You will see a blinking character in the LimA value eld.
2. While that character is still blinking, press down RANGE keys as needed to select the 10 mA limit range. Verify that the source limit value in the LimA eld is 10.0000 mA.
ConnectUnpack
Safety
FAQs
Next Steps
Introduction
Test
Page 14
Step 3: Set measurement function and range
1. Press the MEAS key as many times as needed to select the V (voltage) measurement function. In the following gure, the measurement function has been set to “V.”
2. Press the AUTO key as many times as needed to select the AUTO range function. When AUTO is selected,
the Series 2600B automatically selects the best range for the measured value. You will briey see the display, shown below,
and then the main display screen reappears.
Step 4: Turn the output on
Turn the output on by pressing the OUTPUT ON/OFF control
(for 2634B and 2636B) or (2635B). The ON/OFF indicator LED
lights and measurements begin.
Step 5: Observe measurements
Observe the measured voltage on the main area of front panel
display. The readings should be very close to the 20 V source
value.
Page 15
Step 6: Turn the output off
When you nish making measurements, turn the output off by
pressing the OUTPUT ON/OFF control or control.
The output indicator LED turns off.
For a basic understanding of the of the Series 2600B functionality, we strongly recommend that rst-time users become familiar with
the Series 2600B Reference Manual.
These steps conrm basic functionality of your 2634B, 2635B, or 2636B instrument. Turn the instrument power OFF
now.
ConnectUnpack
Safety
FAQs
Next Steps
Introduction
Test
Page 16
FAQs
What should I do if I see an error message when I turn the
instrument on?
If an error message is displayed, press the EXIT (LOCAL) key. The Series 2600B will return to the default display screen. For
detailed information about error messages, see “Errors and
status messages” in the Series 2600B Reference Manual.
How do I use the included USB ash drive?
The USB ash drive included with the Series 2600B can be used
to load test scripts onto the instrument from the front panel. The
included USB ash drive contains a copy of the 2400 personality script, which allows the Series 2600B instrument to accept 2400 SCPI commands. For more information about loading and running the 2400 personality script, see “Model 2400 emulation” in the Series 2600B Reference Manual.
What do I do if I lost or formatted the included USB ash drive? How can I get another copy of the 2400 personality
script?
If you lose or format the USB drive or delete the 2400
personality script, you can download the latest version from the Keithley Instruments website. Go to www.tek.com/keithley and search for TSP Script for Series 2600B SMUs to Emulate Model 2400 SMUs.
Page 17
Safety
Introduction
Next steps
The Series 2600B Reference Manual
Refer to the Series 2600B Reference Manual for detailed
information about all features of the instrument.
Keithley Instruments website
See: www.tek.com/keithley for support and additional information about the instrument.
ConnectUnpack
Safety
Test
Introduction
FAQs
Next Steps
Page 18
Contact Information:
Australia* 1 800 709 465
Austria 00800 2255 4835
Belgium* 00800 2255 4835
Brazil +55 (11) 3759 7627
Canada 1 800 833 9200
Central East Europe / Baltics +41 52 675 3777
Central Europe / Greece +41 52 675 3777
Denmark +45 80 88 1401
Finland +41 52 675 3777
France* 00800 2255 4835
Germany* 00800 2255 4835
Hong Kong 400 820 5835
India 000 800 650 1835
Indonesia 007 803 601 5249
Italy 00800 2255 4835
Japan 81 (3) 6714 3010
Luxembourg +41 52 675 3777
Malaysia 1 800 22 55835
Middle East, Asia, and North Africa +41 52 675 3777
The Netherlands* 00800 2255 4835
New Zealand 0800 800 238
Norway 800 16098
People’s Republic of China 400 820 5835
Philippines 1 800 1601 0077
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea +82 2 6917 5000
Russia / CIS +7 (495) 6647564
Singapore 800 6011 473
South Africa +41 52 675 3777
Spain* 00800 2255 4835
Find more valuable resources at TEK.COM
Copyright © 2017, Tektronix. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published
material. Specication and price change privileges
reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies.
* European toll-free number. If not accessible, call: +41 52 675 3777
A Tektronix Company
Loading...