Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-recharge
able batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT
OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC.,
HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED
DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLA
TION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes sub
sequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially. When a new Revision is created, all Addenda associated with the previous Revi
sion of the manual are incorporated into the new Revision of the manual. Each new Revision
includes a revised copy of this print history page.
Revision A (Document Number 2425-902-01)............................................................. August 2000
Revision B (Document Number 2425-902-01) ............................................................October 2004
-
-
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state
if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of products. Only properly
trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category
II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and
data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with
high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often as
sociated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to
Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures.
The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock.
If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting ca
-
-
5/03
bles or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The ! symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of
normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment frame.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Always read the associated
information very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer,
test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from
other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only
through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability
of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for
proper cleaning/servicing.
1-2Performance VerificationModel 2425 Service Manual
Introduction
Use the procedures in this section to verify that Model 2425 accuracy is within the limits
stated in the instrument’s one-year accuracy specifications. You can perform these verifi
cation procedures:
•When you first receive the instrument to make sure that it was not damaged during
shipment.
•To verify that the unit meets factory specifications.
•To determine if calibration is required.
•Following calibration to make sure it was performed properly.
WARNINGThe information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified
to do so. Some of these procedures may expose you to hazardous volt
ages, which could cause personal injury or death if contacted. Use
standard safety precautions when working with hazardous voltages.
NOTEIf the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the cor
rect course of action.
-
-
-
Verification test requirements
Be sure that you perform the verification tests:
•Under the proper environmental conditions.
•After the specified warm-up period.
•Using the correct line voltage.
•Using the proper test equipment.
•Using the specified output signals and reading limits.
Model 2425 Service ManualPerformance Verification1-3
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•An ambient temperature of 18 to 28°C (65 to 82°F).
•A relative humidity of less than 70% unless otherwise noted.
Warm-up period
Allow the Model 2425 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges
stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
The Model 2425 requires a line voltage of 100 to 240V and a line frequency of 50 or
60Hz. Verification tests must be performed within this range.
1-4Performance VerificationModel 2425 Service Manual
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equipment as long as that equipment has specifications comparable to those listed in Table 1-1.
Keep in mind, however, that test equipment uncertainty will add to the uncertainty of each
measurement. Generally, test equipment uncertainty should be at least four times better
than corresponding Model 2425 specifications. Table 1-1 lists the specifications of the
recommended test equipment, including maximum allowable uncertainty for alternate test
equipment (shown in parentheses).
Table 1-1
Recommended verification equipment
DescriptionManufacturer/ModelSpecifications
Digital MultimeterHewlett Packard
HP3458A
Resistance Calibrator Fluke 5450AResistance
Precision Resistor
1
90-day, full-range accuracy specifications of ranges required for various measurement points.
2
90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of alternate test
equipment shown in parentheses.
3
Required for verification of 3A current range. Characterize resistor to ±300ppm or better using recommended DMM before ver-
Model 2425 Service ManualPerformance Verification1-5
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before verifying the 3A current measurement range. (You need not characterize the resistor if you are
checking only the 3A current source range.) Use the 4-wire ohms function of the DMM
recommended in Table 1-1 to measure the resistance value. Then use that measured value
to calculate the current during the 3A current measurement range test procedure.
Verification limits
The verification limits stated in this section have been calculated using only the Model
2425 one-year accuracy specifications, and they do not include test equipment uncer
tainty. If a particular measurement falls outside the allowable range, recalculate new limits
based on Model 2425 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 20V
DC output range using a 20V output value. Using the Model 2425 20V range one-year
accuracy specification of ±(0.02% of output + 2.4mV offset), the calculated output limits
are:
When verifying the resistance measurement accuracy, it will probably be necessary to
recalculate resistance limits based on the actual calibrator resistance values. You can cal
culate resistance reading limits in the same manner described above, but be sure to use the
actual calibrator resistance values and the Model 2425 normal accuracy specifications for
your calculations.
As an example, assume you are testing the 20kΩ range, and that the actual value of the
nominal 19kΩ calibrator resistor is 19.01kΩ. Using the Model 2425 20kΩ range one-year
normal accuracy specifications of ±(0.063% of reading + 3Ω), the recalculated reading
limits are:
If the Model 2425 is not within specifications and not under warranty, see the calibration
procedures in Section 2 for information on calibrating the unit.
Test considerations
When performing the verification procedures:
•Be sure to restore factory front panel defaults as previously outlined.
•Make sure that the test equipment is properly warmed up and connected to the
Model 2425 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are
selected with the TERMINALS key.
•Make sure the Model 2425 is set to the correct source range (see below).
•Ensure that the Model 2425 output is turned on before making measurements.
Model 2425 Service ManualPerformance Verification1-7
•Ensure the test equipment is set up for the proper function and range.
•Allow the Model 2425 output signal to settle before making a measurement.
•Do not connect test equipment to the Model 2425 through a scanner, multiplexer,
or other switching equipment.
WARNINGThe maximum common-mode voltage (voltage between LO and chas-
sis ground) is 250V peak. Exceeding this value may cause a breakdown
in insulation, creating a shock hazard.
CAUTIONThe maximum voltage between INPUT/OUTPUT HI and LO or
4-WIRE SENSE HI and LO is 125V peak. The maximum voltage
between INPUT/OUTPUT HI and 4-WIRE SENSE HI or between
INPUT/OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these
voltage values may result in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output
value as outlined below.
1.Press either the SOURCE V or SOURCE I key to select the appropriate source
function.
2.Press the EDIT key as required to select the source display field. Note that the cursor will flash in the source field while its value is being edited.
3.With the cursor in the source display field flashing, set the source range to the lowest possible range for the value to be sourced using the up or down RANGE key.
For example, you should use the 20V source range to output a 20V source value.
With a 20V source value and the 20V range selected, the source field display will
appear as follows:
Vsrc: +20.0000 V
4.With the source field cursor flashing, set the source output to the required value
using either:
•The SOURCE adjustment and left and right arrow keys.
•The numeric keys.
5.Note that the source output value will be updated immediately; you need not press
ENTER when setting the source value.
1-8Performance VerificationModel 2425 Service Manual
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range
is coupled to the source range, and you cannot independently control the measure range.
Thus, it is not necessary for you to set the measure range when testing voltage or current
measurement accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 3.15A.
Conversely, when sourcing current, you can set the SourceMeter to limit voltage from
0.2mV to 105V. The SourceMeter output will not exceed the programmed compliance
limit.
Types of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on
which value is lower, the output will clamp at either the displayed compliance setting
(“real”) or at the maximum measurement range reading (“range”).
The “real” compliance condition can occur when the compliance setting is less than the
highest possible reading of the measurement range. When in compliance, the source out
put clamps at the displayed compliance value. For example, if the compliance voltage is
set to 1V and the measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible
reading of the selected measurement range. When in compliance, the source output
clamps at the maximum measurement range reading (not the compliance value). For
example, if the compliance voltage is set to 1V and the measurement range is 200mV, the
output voltage will clamp (limit) at 210mV.
-
Model 2425 Service ManualPerformance Verification1-9
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in
Table 1-2.
Table 1-2
Maximum compliance values
Measurement
range
200mV
2V
20V
100V
10μA
100μA
1mA
10mA
100mA
1A
3A
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e.,
“mA”) for the compliance display will flash.
210mV
2.1V
21V
105V
10.5μA
105 μA
1.05mA
10.5mA
105mA
1.05A
3.15A
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows.
They assume that the measurement function is the same as the compliance function.
Maximum
compliance value
•Compliance Setting < Measurement Range = Real Compliance
•Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compliance setting to the present measurement range. If the compliance setting is lower than the
maximum possible reading on the present measurement range, the compliance setting is
the compliance limit. If the compliance setting is higher than the measurement range, the
maximum reading on that measurement range is the compliance limit.
1-10Performance VerificationModel 2425 Service Manual
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance.
For purposes of the verification tests, the SourceMeter can be taken out of compliance by
going into the edit mode and increasing the compliance limit.
NOTEDo not take the unit out of compliance by decreasing the source value or chang-
ing the range. Always use the recommended range and source settings when
performing the verification tests.
Output voltage accuracy
Follow the steps below to verify that Model 2425 output voltage accuracy is within specified limits. This test involves setting the output voltage to each full-range value and measuring the voltages with a precision digital multimeter.
1.With the power off, connect the digital multimeter to the Model 2425 INPUT/
OUTPUT jacks, as shown in Figure 1-1.
2.Select the multimeter DC volts measuring function.
3.Press the Model 2425 SOURCE V key to source voltage, and make sure the source
output is turned on.
Figure 1-1
Connections for voltage verification tests
EDIT
DISPLAY
TOGGLE
POWER
V
1
LOCAL
REL
67
DIGITS SPEED
I
MEAS
Ω
230
FILTER
89
RECALL
STORE
FCTN
LIMIT
V
4
TRIG
SWEEP
+/-
CONFIG MENU
I
5
Model 2425
Digital Multimeter
2425 100 W SourceMeter
SOURCE
EDIT
EXIT ENTER
RANGE
AUTO
RANGE
4- WIRE
INPUT/
SENSE
OUTPUT
HI
125V
PEAK
®
ON/OFF
OUTPUT
125V
5V
PEAK
PEAK
LO
250V
PEAK
!
TERMINALS
FRONT/
REAR
Input HI
Input LO
Model 2425 Service ManualPerformance Verification1-11
4.Verify output voltage accuracy for each of the voltages listed in Table 1-3. For
each test point:
•Select the correct source range.
•Set the Model 2425 output voltage to the indicated value.
•Verify that the multimeter reading is within the limits given in the table.
5.Repeat the procedure for negative output voltages with the same magnitudes as
those listed in Table 1-3.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-3
Output voltage accuracy limits
Model 2425
source range
200mV
2V
20V
100V
Model 2425 output
voltage setting
200.000mV
2.00000V
20.0000V
100.000V
Voltage measurement accuracy
Follow the steps below to verify that Model 2425 voltage measurement accuracy is within
specified limits. The test involves setting the source voltage to full-range values, as mea
sured by a precision digital multimeter, and then verifying that the Model 2425 voltage
readings are within required limits.
1.With the power off, connect the digital multimeter to the Model 2425 INPUT/
OUTPUT jacks, as shown in Figure 1-1.
2.Select the multimeter DC volts function.
3.Set the Model 2425 to both source and measure voltage by pressing the SOURCE
V and MEAS V keys, and make sure the source output is turned on.
Output voltage limits
(1 year, 18°C to 28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
19.9936 to 20.0064V
99.968 to 100.032V
-
1-12Performance VerificationModel 2425 Service Manual
4.Verify output voltage accuracy for each of the voltages listed in Table 1-4. For
each test point:
•Select the correct source range.
•Set the Model 2425 output voltage to the indicated value as measured by the
digital multimeter.
•Verify that the Model 2425 voltage reading is within the limits given in the
table.
NOTEIt may not be possible to set the voltage source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
5.Repeat the procedure for negative source voltages with the same magnitudes as
those listed in Table 1-4.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key select the rear
panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Model 2425 source
1
and measure range
200mV
2V
20V
100V
1
Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2
As measured by multimeter. Use closest possible value and modify reading limits accordingly.
Source voltage
200.000mV
2.00000V
20.0000V
100.000V
2
Model 2425 voltage reading
limits (1 year, 18°C to 28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
19.9960 to 20.0040V
99.982 to 100.018V
Model 2425 Service ManualPerformance Verification1-13
Output current accuracy
Follow the steps below to verify that Model 2425 output current accuracy is within specified limits. The test involves setting the output current to each full-range value and measuring the currents with a precision digital multimeter.
10μA to 1A range accuracy
1.With the power off, connect the digital multimeter to the Model 2425 INPUT/
OUTPUT jacks, as shown in Figure 1-2.
Figure 1-2
Connections for 10μA to 1A range current verification tests
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
REL
67
DIGITS SPEED
125V
PEAK
5
SWEEP
SOURCE
I
2425 100 W SourceMeter
EDIT
EXIT ENTER
MEAS
Ω
230
FILTER
89
RECALL
STORE
FCTN
LIMIT
V
4
TRIG
+/-
CONFIG MENU
I
1
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
Model 2425
PEAK
5V
LO
TERMINALS
125V
PEAK
250V
PEAK
!
FRONT/
REAR
Input LO
Amps
Digital Multimeter
2.Select the multimeter DC current measuring function.
3.Press the Model 2425 SOURCE I key to source current, and make sure the source
output is turned on.
4.Verify output current accuracy for the 10μA-1A range currents listed in Table 1-5.
For each test point:
•Select the correct source range.
•Set the Model 2425 output current to the correct value.
•Verify that the multimeter reading is within the limits given in the table.
1-14Performance VerificationModel 2425 Service Manual
5.Repeat the procedure for negative output currents with the same magnitudes as
those listed in Table 1-5.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
3A range accuracy
1.With the power off, connect the digital multimeter and the 1Ω resistor to the Model
2425 INPUT/OUTPUT jacks, as shown in Figure 1-3.
2.Select the multimeter DC volts measuring function.
3.Press the Model 2425 SOURCE I key to source current, and make sure the source
output is turned on.
Figure 1-3
Connections for 3A range current verification tests
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
1
LOCAL
REL
67
DIGITS SPEED
I
MEAS
Ω
230
FILTER
89
RECALL
STORE
FCTN
LIMIT
2425 100 W SourceMeter
SOURCE
I
V
4
5
TRIG
SWEEP
+/-
CONFIG MENU
EXIT ENTER
Model 2425
Digital Multimeter
125V
PEAK
®
EDIT
RANGE
AUTO
RANGE
ON/OFF
OUTPUT
PEAK
5V
LO
!
TERMINALS
FRONT/
REAR
PEAK
PEAK
125V
250V
1Ω Resistor
Input HI
Input LO
Model 2425 Service ManualPerformance Verification1-15
4.Verify output current accuracy for the 3A range. Be sure to:
•Select the 3A source range.
•Set the Model 2425 output current to the correct 3A output value.
•Verify that the multimeter reading is within the 3A range limits given in Table 1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal
value, the DMM voltage reading is the same as the sourced current.)
5.Repeat the procedure for a negative 3A current output value.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2425
source range
10mA
100mA
1mA
10mA
100mA
1A
1
3A
1
See separate procedure for 3A range. DMM voltage reading is same as sourced current.
Model 2425
output current setting
10.0000mA
100.000mA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
Output current limits
(1 year, 18°C to 28°C)
9.9947 to 10.0053mA
99.949 to 100.051mA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99893 to 1.00107A
2.99543 to 3.00457A
1-16Performance VerificationModel 2425 Service Manual
Current measurement accuracy
Follow the steps below to verify that Model 2425 current measurement accuracy is within
specified limits. The procedure involves applying accurate currents from the Model 2425
current source and then verifying that Model 2425 current measurements are within
required limits.
10μA to 1A range accuracy
1.With the power off, connect the digital multimeter to the Model 2425 INPUT/
OUTPUT jacks, as shown in Figure 1-2.
2.Select the multimeter DC current function.
3.Set the Model 2425 to both source and measure current by pressing the SOURCE I
and MEAS I keys, and make sure the source output is turned on.
4.Verify measure current accuracy for the 10μA-1A range currents listed in Table
1-6. For each measurement:
•Select the correct source range.
•Set the Model 2425 source output to the correct value as measured by the digital mul-
timeter.
•Verify that the Model 2425 current reading is within the limits given in the table.
NOTEIt may not be possible to set the current source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
5.Repeat the procedure for negative calibrator currents with the same magnitudes as
those listed in Table 1-6.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
3A range accuracy
NOTEThe 1Ω resistor should be characterized to within ±300ppm before verifying the
3A current measurement range. Use the 4-wire ohms function of the DMM to
measure the resistance value, and then use that measured value to calculate the
current during the measurement procedure.
1.With the power off, connect the 1Ω resistor and digital multimeter to the Model
2425 INPUT/OUTPUT jacks, as shown in Figure 1-3.
2.Select the multimeter DC volts function.
3.Set the Model 2425 to both source and measure current by pressing the SOURCE I
and MEAS I keys, and make sure the source output is turned on.
Model 2425 Service ManualPerformance Verification1-17
4.Verify measurement current accuracy for the 3A range as follows:
•Select the 3A source range.
•Set the Model 2425 source output to the correct 3A value as measured by the digital
multimeter.
•Note the DMM voltage reading, and then calculate the current from the voltage reading and characterized 1Ω resistance value as I = V/R, where V is the DMM voltage
reading and R is the characterized resistance value.
•Verify that the Model 2425 current reading is within the 3A limits given in Table 1-
6.
NOTEIt may not be possible to set the current source to the specified 3A value. Use the
closest possible setting, and modify reading limits accordingly.
5.Repeat the procedure for a negative 3A current.
6.Repeat the procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select
the rear panel jacks with the front panel TERMINALS key.
Table 1-6
Current measurement accuracy limits
Model 2425 source
1
and measure range
10μA
100μA
1mA
10mA
100mA
1A
3A
1
Measure range coupled to source range when simultaneously sourcing and measuring current.
2
As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
3
Current calculated as follows: I = V/R, where V is the DMM voltage reading, and R is the
characterized value of the 1Ω resistor.
Source current
10.00000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
Model 2425 current reading lim-
2
its (1 year, 18°C to 28°C)
9.9966 to 10.0034μA
99.969 to 100.031μA
0.99967 to 1.00033mA
9.9959 to 10.0041mA
99.939 to 100.061mA
0.99928 to 1.00072A
2.99673 to 3.00327A
3
1-18Performance VerificationModel 2425 Service Manual
Resistance measurement accuracy
Use the following steps to verify that Model 2425 resistance measurement accuracy is
within specified limits. This procedure involves applying accurate resistances from a
resistance calibrator and then verifying that Model 2425 resistance measurements are
within required limits.
CAUTIONBefore testing the 2Ω and 20Ω ranges, make sure your resistance cali-
brator can safely handle the default test currents for those ranges (see
Model 2425 and calibrator specifications). If not, use the CONFIG
OHMS menu to select the MANUAL source mode, then set the source
current to an appropriate safe value. When using the manual source
mode, total resistance reading uncertainty includes both Source I and
Measure V uncertainty (see specifications), and calculated reading
limits should take the additional uncertainty into account.
1.With the power off, connect the resistance calibrator to the Model 2425 INPUT/
OUTPUT and 4-WIRE SENSE jacks, as shown in Figure 1-4. Be sure to use the
4-wire connections as shown.
Figure 1-4
Connections for resistance accuracy verification
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
REL
67
DIGITS SPEED
4- WIRE
INPUT/
SENSE
OUTPUT
HI
125V
PEAK
5
SWEEP
SOURCE
I
2425 100 W SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
1
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
Model 2425
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
Output HI
125V
5V
PEAK
PEAK
LO
250V
PEAK
!
TERMINALS
FRONT/
REAR
Sense HI
Resistance Calibrator
Output LO
Sense LO
Model 2425 Service ManualPerformance Verification1-19
2.Select the resistance calibrator external sense mode.
3.Configure the Model 2425 ohms function for the 4-wire sense mode as follows:
•Press CONFIG then MEAS Ω. The instrument will display the following:
•Select SENSE-MODE, then press ENTER. The following will be displayed:
SENSE-MODE
2-WIRE 4-WIRE
•Select 4-WIRE, then press ENTER.
•Press EXIT to return to normal display.
4.Press MEAS Ω to select the ohms measurement function, and make sure the source
output is turned on.
5.Verify ohms measurement accuracy for each of the resistance values listed in Table
1-7. For each measurement:
•Set the resistance calibrator output to the nominal resistance or closest available
value.
NOTEIt may not be possible to set the resistance calibrator to the specified value. Use
the closest possible setting, and modify reading limits accordingly.
•Select the appropriate ohms measurement range with the RANGE keys.
•Verify that the Model 2425 resistance reading is within the limits given in the table.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE
SENSE jacks. Be sure to select the rear panel jacks with the front panel TERMI
NALS key.
Table 1-7
Ohms measurement accuracy limits
Calibrator
Model 2425 range
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
1
Nominal resistance values.
2
Reading limits based on Model 2425 normal accuracy specifications and nominal resistance values. If
actual resistance values differ from nominal values shown, recalculate reading limits using actual calibrator
resistance values and Model 2425 normal accuracy specifications. See Verification limits earlier in this
section for details.
resistance
1.9Ω
19Ω
190Ω
1.9kΩ
19kΩ
190kΩ
1.9MΩ
19MΩ
1
Model 2425 resistance reading limits
(1 year, 18°C to 28°C)
1.89649 to 1.90351Ω
18.9784 to 19.0216Ω
189.824 to 190.176Ω
1.89845 to 1.90155kΩ
18.9850 to 19.0150kΩ
189.847 to 190.153kΩ
1.89861 to 1.90139MΩ
18.9517 to 19.0483MΩ
-
2
1-20Performance VerificationModel 2425 Service Manual
2
Calibration
2-2CalibrationModel 2425 Service Manual
Introduction
Use the procedures in this section to calibrate the Model 2425. These procedures require
accurate test equipment to measure precise DC voltages and currents. Calibration can be
performed either from the front panel or by sending SCPI calibration commands over the
IEEE-488 bus or RS-232 port with the aid of a computer.
WARNINGThe information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified
to do so. Some of these procedures may expose you to hazardous volt
ages.
Environmental conditions
Temperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18 to 28°C (65 to 82°F)
with relative humidity of less than 70% unless otherwise noted.
-
Warm-up period
Allow the Model 2425 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges
stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
Model 2425 requires a line voltage of 100 to 240V at line frequency of 50 or 60Hz. The
instrument must be calibrated while operating from a line voltage within this range.
Model 2425 Service ManualCalibration2-3
Calibration considerations
When performing the calibration procedures:
•Ensure that the test equipment is properly warmed up and connected to the Model
2425 front panel INPUT/OUTPUT jacks. Also be certain that the front panel jacks
are selected with the TERMINALS switch.
•Always allow the source signal to settle before calibrating each point.
•Do not connect test equipment to the Model 2425 through a scanner or other
switching equipment.
•If an error occurs during calibration, the Model 2425 will generate an appropriate
error message. See Appendix B for more information.
WARNINGThe maximum common-mode voltage (voltage between LO and chas-
sis ground) is 250V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
CAUTIONThe maximum voltage between INPUT/OUTPUT HI and LO or
4-WIRE SENSE HI and LO is 125V peak. The maximum voltage
between INPUT/OUTPUT HI and 4-WIRE SENSE HI or between
INPUT/OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these
voltage values may result in instrument damage.
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its specifications.
2-4CalibrationModel 2425 Service Manual
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use
alternate equipment as long as that equipment has specifications comparable to those
listed in the table. For optimum calibration accuracy, test equipment specifications should
be at least four times better than corresponding Model 2425 specifications.
Table 2-1
Recommended calibration equipment
DescriptionManufacturer/ModelSpecifications
Digital MultimeterHewlett Packard
HP3458A
DC Voltage
1
1V:
10V:
100V:
±5.6ppm
±4.3ppm
±6.3ppm
DC Current
1
10μA:
100μA:
1mA:
10mA:
100mA:
1A:
Precision Resistor
1
90-day, full-range accuracy specifications of ranges required for various measurement points.
2
Necessary for calibration of 3A current range. Resistor must be characterized to ±300ppm or better using
recommended DMM before calibrating 3A range.
2
Isotec RUG-Z-1R00-0.11Ω, ±0.1%, 100W
1Ω resistor characterization
The 1Ω resistor must be characterized to ±300ppm or better before calibrating the 3A current range. Use the 4-wire ohms function of the DMM recommended in Table 1-1 to measure the resistance value, and then use that measured value to calculate the current during
the 3A current range calibration procedure.
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
Model 2425 Service ManualCalibration2-5
Unlocking calibration
Before performing calibration, you must first unlock calibration by entering or sending the
calibration password as explained in the following paragraphs.
Unlocking calibration from the front panel
1.Press the MENU key, then choose CAL, and press ENTER. The instrument will
display the following:
3.Use the up and down RANGE keys to select the letter or number, and use the left
and right arrow keys to choose the position. (Press down RANGE for letters; up
RANGE for numbers.) Enter the present password on the display. (Front panel
default: 002425.)
4.Once the correct password is displayed, press the ENTER key. If the password was
correctly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5.Press EXIT to return to normal display. Calibration will be unlocked and assume
the states summarized in Table 2-2. Attempts to change any of the settings listed
with calibration unlocked will result in an error +510, “Not permitted with cal
un-locked.”
NOTEWith calibration unlocked, the sense function and range track the source func-
tion and range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC
setting will be 'VOLT:DC'. When :SOUR:FUNC is set to CURR, the
:SENS:FUNC setting will be 'CURR:DC'. A similar command coupling exists
for :SOUR:VOLT:RANG/:SENS:VOLT:RANG and
SOUR:CURR:RANG:SENS:CURR:RANG.
2-6CalibrationModel 2425 Service Manual
Table 2-2
Calibration unlocked states
ModeStateEquivalent remote command
Concurrent Functions
Sense Function
Sense Volts NPLC
Sense Volts Range
Sense Current NPLC
Sense Current Range
Filter Count
Filter Control
Filter Averaging
Source V Mode
Volts Autorange
Source I Mode
Current Autorange
Autozero
Trigger Arm Count
Trigger Arm Source
Trigger Count
Trigger Source
OFF
Source
1.0
Source V
1.0
Source I
10
REPEAT
ON
FIXED
OFF
FIXED
OFF
ON
1
Immediate
1
Immediate
:SENS:FUNC:CONC OFF
:SENS:FUNC <source_ function>
:SENS:VOLT:NPLC 1.0
:SENS:VOLT:RANG <range>
:SENS:CURR:NPLC 1.0
:SENS:CURR:RANG <range>
:SENS:AVER:COUN 10
:SENS:AVER:TCON REPeat
:SENS:AVER:STAT ON
:SOUR:VOLT:MODE FIXED
:SOUR:VOLT:RANG:AUTO OFF
:SOUR:CURR:MODE FIXED
:SOUR:CURR:RANG:AUTO OFF
:SYST:AZERO ON
:ARM:COUNT 1
:ARM:SOUR IMMediate
:TRIG:COUNT 1
:TRIG:SOUR IMMediate
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE '<password>'
For example, the following command uses the default password:
:CAL:PROT:CODE 'KI002425'
Model 2425 Service ManualCalibration2-7
Changing the password
The default password may be changed from the front panel or via remote as discussed
next.
Changing the password from the front panel
Follow the steps below to change the password from the front panel:
1.Press the MENU key, then choose CAL and press ENTER. The instrument will
display the following:
4.Using the range keys and the left and right arrow keys, enter the new password on
the display.
5.Once the desired password is displayed, press the ENTER key to store the new
password.
Changing the password by remote
To change the calibration password by remote, first send the present password, and then
send the new password. For example, the following command sequence changes the pass
word from the 'KI002425' remote default to 'KI_CAL':
:CAL:PROT:CODE 'KI002425'
:CAL:PROT:CODE 'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTEIf you change the first two characters of the password to something other than
“KI”, you will not be able to unlock calibration from the front panel.
-
2-8CalibrationModel 2425 Service Manual
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the
CAL pads, which are located on the display board. Doing so will also reset the password
to the factory default (KI002425).
See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the
display board component layout drawing at the end of Section 6 for the location of the
CAL pads.
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be
accessible in the calibration menu. To view calibration dates and calibration count at any
time:
1.From normal display, press MENU, select CAL, and then press ENTER. The unit
will display the following:
2.Select VIEW-DATES, then press ENTER. The Model 2425 will display the next
and last calibration dates and the calibration count as in the following example:
NEXT CAL: 12/15/1999
Last calibration: 12/15/1999 Count: 0001
Calibration errors
The Model 2425 checks for errors after each calibration step, minimizing the possibility
that improper calibration may occur due to operator error.
Front panel error reporting
If an error is detected during comprehensive calibration, the instrument will display an
appropriate error message (see Appendix B). The unit will then prompt you to repeat the
calibration step that caused the error.
Model 2425 Service ManualCalibration2-9
Remote error reporting
You can detect errors while in remote by testing the state of EAV (Error Available) bit (bit
2) in the status byte. (Use the *STB? query to request the status byte.) Query the
instrument for the type of error by using the appropriate :SYST:ERR? query. The Model
2425 will respond with the error number and a text message describing the nature of the
error. See Appendix B for details.
Front panel calibration
The following front panel calibration procedure calibrates all ranges of both the current
and voltage source and measure functions. Note that each range is separately calibrated by
repeating the entire procedure for each range.
Step 1: Prepare the Model 2425 for calibration.
1.Turn on the Model 2425 and the digital multimeter, and allow them to warm up for
at least one hour before performing calibration.
2.Press the MENU key, then choose CAL and press ENTER. Select UNLOCK, and
then press ENTER. The instrument will display the following:
3.Use the up and down range keys to select the letter or number, and use the left and
right arrow keys to choose the position. Enter the present password on the display.
(Front panel default: 002425.) Press ENTER to complete the process.
4.Press EXIT to return to normal display. Instrument operating states will be set as
summarized in Table 2-2.
2-10CalibrationModel 2425 Service Manual
Step 2: Voltage calibration.
Perform the steps below for each voltage range, using Table 2-3 as a guide.
1.Connect the Model 2425 to the digital multimeter, as shown in Figure 2-1. Select
the multimeter DC volts measurement function.
Figure 2-1
Voltage calibration test connections
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
REL
67
DIGITS SPEED
125V
PEAK
5
SWEEP
SOURCE
I
2425 100 W SourceMeter
EDIT
EXIT ENTER
MEAS
Ω
230
FILTER
89
RECALL
STORE
FCTN
LIMIT
V
4
TRIG
+/-
CONFIG MENU
I
1
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
125V
5V
PEAK
PEAK
LO
250V
PEAK
!
TERMINALS
FRONT/
REAR
Model 2425
Input HI
Digital Multimeter
2.From normal display, press the SOURCE V key.
3.Press the EDIT key to select the source field (cursor flashing in source display
field), and then use the down RANGE key to select the 200mV source range.
4.From normal display, press MENU.
5.Select CAL, then press ENTER. The unit will display the following:
8.Note and record the DMM reading, and then adjust the Model 2425 display to
agree exactly with the actual DMM reading. Use the up and down arrow keys to
select the digit value, and use the left and right arrow keys to choose the digit posi
tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is
within ±10% of the present range.
9.After adjusting the display to agree with the DMM reading, press ENTER. The
instrument will then display the following:
V-CAL
Press ENTER to Output +000.00mV
10.Press ENTER. The Model 2425 will source 0mV and at the same time display the
following:
11.Note and record the DMM reading, and then adjust the Model 2425 display to
agree with the actual DMM reading. Note that the display value adjustment limits
are within ±1% of the present range.
12.After adjusting the display value to agree with the DMM reading, press ENTER.
The unit will then display the following:
V-CAL
Press ENTER to Output -200.00mV
13.Press ENTER. The Model 2425 will source -200mV and display the following:
14.Note and record the DMM reading, and then adjust the Model 2425 display to
agree with the DMM reading. Again, the maximum display adjustment is within
±10% of the present range.
15.After adjusting the display value to agree with the DMM reading, press ENTER
and note that the instrument displays:
V-CAL
Press ENTER to Output +000.00mV
16.Press ENTER. The Model 2425 will source 0mV and simultaneously display the
following:
17.Note and record the DMM reading, and then adjust the display to agree with the
DMM reading. Once again, the maximum adjustment is within ±1% of the present
range.
18.After adjusting the display to agree with the DMM reading, press ENTER to complete calibration of the present range.
19.Press EXIT to return to normal display, then select the 2V source range. Repeat
steps 2 through 18 for the 2V range.
20.After calibrating the 2V range, repeat the entire procedure for the 20V and 100V
ranges using Table 2-3 as a guide. Be sure to select the appropriate source range
with the EDIT and RANGE keys before calibrating each range.
21.Press EXIT as necessary to return to normal display.
Table 2-3
Front panel voltage calibration
Source range1Source voltageMultimeter voltage reading
0.2V+200.00mV
+000.00mV
-200.00mV
+000.00mV
2V+2.0000V
+0.0000V
-2.0000V
+0.0000V
20V+20.000V
+00.000V
-20.000V
+00.000V
100V+100.00V
+000.00V
-100.00V
+000.00V
1
Use EDIT and RANGE keys to select source range.
2
Multimeter reading used in corresponding calibration step. See procedure.
8.Note and record the DMM reading, and then adjust the Model 2425 display to
agree exactly with the actual DMM reading. Use the up and down arrow keys to
select the digit value, and use the left and right arrow keys to choose the digit posi
tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is
within ±10% of the present range.
9.After adjusting the display to agree with the DMM reading, press ENTER. The
instrument will then display the following:
I-CAL
Press ENTER to Output +00.000μA
10.Press ENTER. The Model 2425 will source 0mA and at the same time display the
following:
11.Note and record the DMM reading, and then adjust the Model 2425 display to
agree with the actual DMM reading. Note that the display value adjustment limits
are within ±1% of the present range.
12.After adjusting the display value to agree with the DMM reading, press ENTER.
The unit will then display the following:
I-CAL
Press ENTER to Output -10.000μA
13.Press ENTER. The Model 2425 will source -10μA and display the following:
14.Note and record the DMM reading, and then adjust the Model 2425 display to
agree with the DMM reading. Again, the maximum display adjustment is within
±10% of the present range.
15.After adjusting the display value to agree with the DMM reading, press ENTER
and note that the instrument displays:
I-CAL
Press ENTER to Output +00.000μA
16.Press ENTER The Model 2425 will source 0μA and simultaneously display the
following:
17.Note and record the DMM reading, and then adjust the display to agree with the
DMM reading. Once again, the maximum adjustment is within ±1% of the present
range.
18.After adjusting the display to agree with the DMM reading, press ENTER to complete calibration of the present range.
19.Press EXIT to return to the normal display, and then select the 100μA source range
using the EDIT and up RANGE keys. Repeat steps 2 through 18 for the 100μA
range.
20.After calibrating the 100μA range, repeat the entire procedure for the 1mA through
1A ranges using Table 2-4 as a guide. Be sure to select the appropriate source
range with the EDIT and up RANGE keys before calibrating each range.
21.After calibrating the 1A range, connect the 1Ω characterized resistor and DMM to
the Model 2425 INPUT/OUTPUT jacks, as shown in Figure 2-3.
22.Select the DMM DC volts function.
23.Repeat steps 2 through 18 for the 3A range using Table 2-4 as a guide. When entering the DMM reading, use the calculated current as follows: I = V/R, where V is
the DMM voltage reading, and R is the characterized value of the 1Ω resistor.
3.Select LOCK, then press ENTER to lock out calibration.
2-18CalibrationModel 2425 Service Manual
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands
over the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters
are separately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
these commands, refer to Appendix B.
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed. (For example, omitting a negative full-scale step.)
Ranges that calibrated successfully will be saved if calibration is unlocked. Ranges that failed will not be saved.
Calibration subsystem.
Cal commands protected by password.
Unlock cal; changes password if cal is already unlocked.
(Default password: KI002425.)
Query number of times 2425 has been calibrated.
Save calibration data to EEPROM.*
Lock calibration, inhibit SAVE command operation.
Request cal lock status.
Program calibration year, month, day.
Query calibration year, month, day.
Program calibration due year, month, day.
Query calibration due year, month, day.
Calibrate active measure range. (See Table 2-6 parameters.)
Query measurement cal constants for active range.
Calibrate active source range. (See Table 2-7 parameters.)
Query source cal constants for active range.
Model 2425 Service ManualCalibration2-19
Recommended calibration parameters
The maximum calibration command parameter ranges are: 75% to 150% of full scale for
positive and negative full scale calibration points; ± zero calibration steps have ±50% of
full scale for valid entry ranges. However, for optimum calibration, it is recommended that
you use calibration points within the ranges listed in Table 2-6 and Table 2-7. Note that
each sense range requires three parameters: zero, negative full scale, and positive full
scale. Similarly, each source range requires four parameters: two zero parameters, a posi
tive full-scale parameter, and a negative full-scale parameter.
Note: Parameter steps for each range may be performed in any order, but all parameter steps for each range must be completed. For
optimum calibration, use parameters within recommended limits.
(negative full scale)
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-90 to -110
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -1E-3
-90E-3 to -110E-3
-0.9 to -1.1
-2.7 to -3.15
Second parameter
(negative zero)
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-1 to +1
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to +3E-2
Third parameter
(positive full scale)
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+90 to +110
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+2.7 to +3.15
Fourth parameter
(positive zero)
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-1 to +1
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to +3E-2
Remote calibration procedure
Step 1: Prepare the Model 2425 for calibration.
1.Connect the Model 2425 to the controller IEEE-488 interface or RS-232 port using
a shielded interface cable.
2.Turn on the Model 2425 and the test equipment, and allow them to warm up for at
least one hour before performing calibration.
3.If you are using the IEEE-488 interface, make sure the primary address of the
Model 2425 is the same as the address specified in the program you will be using
to send commands. (Use the MENU key and the COMMUNICATION menu to
access the IEEE-488 address.)
Step 2: Voltage calibration.
1.Connect the Model 2425 to the digital multimeter, and select the multimeter DC
volts function. (Figure 1-2).
2.Send the commands summarized in Table 2-8 in the order listed to initialize voltage calibration. (When the :CAL:PROT:CODE command is sent, the instrument
will assume the operating states listed in Table 2-2.)
Model 2425 Service ManualCalibration2-21
3.Perform the range calibration steps listed in Table 2-9 for all ranges. For each
range:
•Send the :SOUR:VOLT:RANG command to select the source and sense range
being calibrated. For example, for the 2V range, the following command
would be sent:
:SOUR:VOLT:RANG 2
•Program the source to output the negative full-range value using the
:SOUR:VOLT command. For example:
:SOUR:VOLT -2
•Note and record the multimeter reading.
•Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 2V range
would be:
:CAL:PROT:SOUR -1.998
:CAL:PROT:SENS -1.998
•Program the voltage source for 0V output using the :SOUR:VOLT 0.0 command.
•Note the multimeter reading.
•Send the source and sense calibration commands using the multimeter reading
for the parameter. For example:
:CAL:PROT:SOUR1E-3
:CAL:PROT:SmENS1E-3
•Set the source to the positive full-range value using the :SOUR:VOLT command. For example:
:SOUR:VOLT 2
•Note and record the multimeter reading.
•Send the source and sense commands using the multimeter reading as the
parameter. For example:
:CAL:PROT:SOUR 1.997
:CAL:PROT:SENS 1.997
•Send the SOUR:VOLT 0.0 command to set the source voltage to 0V.
•Note and record the multimeter reading.
•Send the :CAL:PROT:SOUR command using the multimeter reading as the
command parameter. For example:
:CAL:PROT:SOUR -1.02E-3
2-22CalibrationModel 2425 Service Manual
Table 2-8
Voltage calibration initialization commands
CommandDescription
*RST
:SOUR:FUNC VOLT
:SENS:CURR:PROT 0.1
:SENS:CURR:RANG 0.1
:SOUR:VOLT:PROT MAX
:SYST:RSEN OFF
:CAL:PROT:CODE 'KI002425'
:OUTP:STAT ON
Restore GPIB defaults.
Activate voltage source.
Current limit when voltage source is active.
Make sure 1A range is not active.
Maximum allowable source voltage.
Disable remote sensing.
Unlock cal.
Turn source on.
Step 3: Current calibration.
1.Connect the Model 2425 to the digital multimeter (Figure 2-2), and select the multimeter DC current function.
2.Send the commands summarized in Table 2-10 in the order listed to initialize current calibration.
Model 2425 Service ManualCalibration2-23
Table 2-9
Voltage range calibration commands
StepCommand/procedure
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
:SOUR:VOLT:RANGE <Range>
:SOUR:VOLT -<Source_value>
Take DMM reading.
2
:CAL:PROT:SOUR <DMM_Reading>
Check 2425 for errors.
3
:CAL:PROT:SENS <DMM_Reading>
Check 2425 for errors.
:SOUR:VOLT 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2425 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2425 for errors.
:SOUR:VOLT +<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2425 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2425 for errors.
:SOUR:VOLT 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
1
Description
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0V.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0V.
Read actual output value.
Calibrate source positive zero.
1
Perform complete procedure for each range, where <Range> = 0.2, 2, 20, and 100, and<Source_value> = 0.2, 2, 20, and 100.
2
<DMM_Reading> parameter is multimeter reading from previous step.
3
Use :SYST:ERR? query to check for errors.
3.Calibrate the 10μA to 1A current ranges using the procedure summarized in Table
2-11. For each range:
•Send the :SOUR:CURR:RANG command to select the source and sense range
being calibrated. For example, for the 1mA range, the command is:
:SOUR:CURR:RANG 1E-3
•Program the source to output the negative full-range value using the
:SOUR:CURR command. For example:
:SOUR:CURR -1E-3
•Note and record the multimeter reading.
2-24CalibrationModel 2425 Service Manual
•Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 1mA
range would be:
:CAL:PROT:SOUR -1.025E-3
:CAL:PROT:SENS -1.025E-3
•Program the current source for 0A output using the :SOUR:CURR 0.0 command.
•Note the multimeter reading.
•Send the source and sense calibration commands using the multimeter reading
for the parameter. For example:
:CAL:PROT:SOUR 1E-6
:CAL:PROT:SENS 1E-6
•Set the source to the positive full-range value using the :SOUR:CURR command. For example, for the 1mA range:
:SOUR:CURR 1E-3
•Note and record the multimeter reading.
•Send the source and sense commands using the multimeter reading as the
parameter. For example:
:CAL:PROT:SOUR 1.03E-3
:CAL:PROT:SENS 1.03E-3
•Send the :SOUR:CURR 0.0 command to set the source current to 0A.
•Note and record the multimeter reading.
•Send the :CAL:PROT:SOUR command using the multimeter reading as the
command parameter. For example:
:CAL:PROT:SOUR 1E-6
4.Connect the 1Ω resistor and DMM to the Model 2425 INPUT/OUPUT jacks, as
shown in Figure 2-3. Select the DMM DC volts function.
5.Repeat step 3 for the 3A range using the calculated current as follows: I = V/R,
where V is the DMM voltage reading, and R is the characterized value of the 1Ω
resistor.
Table 2-10
Current calibration initialization commands
CommandDescription
:SOUR:FUNC CURR
:SENS:VOLT:PROT 20
:SENS:VOLT:RANG 20
:OUTP:STAT ON
Select source current mode.
Voltage limit when current source is active.
Make sure 100V range is not active.
Turn source on.
Model 2425 Service ManualCalibration2-25
Table 2-11
Current range calibration commands
StepCommand/procedure
1
:SOUR:CURR:RANGE <Range>
2
:SOUR:CURR -<Source_value>
3
Take DMM reading.
4
:CAL:PROT:SOUR <DMM_Reading>
5
Check 2425 for errors.
6
:CAL:PROT:SENS <DMM_Reading>
7
Check 2425 for errors.
8
:SOUR:CURR 0.0
9
Take DMM reading.
10
:CAL:PROT:SOUR <DMM_Reading>
11
Check 2425 for errors.
12
:CAL:PROT:SENS <DMM_Reading>
13
Check 2425 for errors.
14
:SOUR:CURR +<Source_value>
15
Take DMM reading.
16
:CAL:PROT:SOUR <DMM_Reading>
17
Check 2425 for errors.
18
:CAL:PROT:SENS <DMM_Reading>
19
Check 2425 for errors.
20
:SOUR:CURR 0.0
21
Take DMM reading.
22
:CAL:PROT:SOUR <DMM_Reading>
1
Perform complete procedure for each range, where <Range> and <Source_value> = 10E-6, 100E-6, 1E-3, 10E-3, 100E-3, 1, or 3.
2
<DMM_Reading> parameter is multimeter reading from previous step.
3
Use :SYST:ERR? query to check for errors.
2
3
1
Description
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A.
Read actual output value.
Calibrate source positive zero.
Step 4: Program calibration dates.
Use following commands to set the calibration date and calibration due date:
:CAL:PROT:NDUE <year>, <month>, <day> (Next calibration due date)
Note that the year, month, and date must be separated by commas. The allowable range for
the year is from 1998 to 2097, the month is from 1 to 12, and the date is from 1 to 31.
2-26CalibrationModel 2425 Service Manual
Step 5: Save calibration constants.
Calibration is now complete, so you can store the calibration constants in EEROM by
sending the following command:
:CAL:PROT:SAVE
NOTECalibration will be temporary unless you send the SAVE command. Also,
calibration data will not be saved if (1) calibration is locked, (2) invalid data
exists, or (3) all steps were not completed.
Step 6: Lock out calibration.
To lock out further calibration, send the following command after completing the calibration procedure:
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the
entire instrument is properly calibrated. In some instances, however, you may want to
calibrate only certain ranges. To do so, complete the entire procedure only for the range(s)
to be calibrated. Keep in mind, however, that you must complete all parameter steps for
each source or sense range. Also be sure to set calibration dates and save calibration after
calibrating the desired range(s).
3
Routine Maintenance
3-2Routine MaintenanceModel 2425 Service Manual
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Introduction
The information in this section deals with routine type maintenance that can be performed
by the operator.
Line fuse replacement
WARNINGDisconnect the line cord at the rear panel, and remove all test leads
connected to the instrument (front and rear) before replacing the line
fuse.
The power line fuse is accessible from the rear panel, just above the AC power receptacle
(Figure 3-1).
Figure 3-1
Rear panel
WARNING: NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
5V
HI
125V
PEAK
4-WIRE
SENSE
WITH FRONT PANEL MENU)
5V
PEAK
LO
INPUT/
!
OUTPUT
IEEE-488
(ENTER IEEE ADDRESS
125V
PEAK
PK
250V
PEAK
PEAK
5V
V, Ω,
GUARD
GUARD
SENSE
RS232
TRIGGER
LINK
MADE IN
U.S.A.
LINE FUSE
SLOWBLOW
3.15A, 250V
LINE RATING
100-240VAC
50, 60HZ
250VA MAX
INTERLOCK-
DIGITAL I/O
Perform the following steps to replace the line fuse:
1.Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to
the fuse holder.
2.Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
Model 2425 Service ManualRoutine Maintenance3-3
CAUTIONTo prevent instrument damage, use only the fuse type specified in
Table 3-1.
3.Re-install the fuse carrier.
NOTEIf the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional
information.
This section of the manual will assist you in troubleshooting and repairing the Model
2425. Included are self-tests, test procedures, troubleshooting tables, and circuit descrip
tions. Note that disassembly instructions are located in Section 5, and component layout
drawings are at the end of Section 6.
-
Safety considerations
WARNINGThe information in this section is intended for qualified service
personnel only. Do not perform these procedures unless you are
qualified to do so. Some of these procedures may expose you to
hazardous voltages that could cause personal injury or death. Use
caution when working with hazardous voltages.
Repair considerations
Before making any repairs to the Model 2425, be sure to read the following
considerations.
CAUTIONThe PC-boards are built using surface mount techniques and require
specialized equipment and skills for repair. If you are not equipped
and/or qualified, it is strongly recommended that you send the unit
back to the factory for repairs or limit repairs to the PC-board
replacement level. Without proper equipment and training, you could
damage a PC-board beyond repair.
•Repairs will require various degrees of disassembly. However, it is recommended
that the Front Panel Tests be performed prior to any disassembly. The disassembly
instructions for the Model 2425 are contained in Section 5 of this manual.
•Do not make repairs to surface mount PC-boards unless equipped and qualified to
do so. (See previous CAUTION.)
•When working inside the unit and replacing parts, adhere to the handling
precautions and cleaning procedures explained in Section 5.
•Many CMOS devices are installed in the Model 2425. These static-sensitive
devices require special handling as explained in Section 5.
•Whenever a circuit board is removed or a component is replaced, the Model 2425
must be recalibrated. See Section 2 for details on calibrating the unit.
Model 2425 Service ManualTroubleshooting4-3
Power-on self-test
During the power-on sequence, the Model 2425 will perform a checksum test on its
EPROM and test its RAM. If one of these tests fails, the instrument will lock up.
Front panel tests
There are three front panel tests: one to test the functionality of the front panel keys and
two to test the display. In the event of a test failure, refer to Display board checks for
details on troubleshooting the display board.
KEYS test
The KEYS test lets you check the functionality of each front panel key. Perform the following steps to run the KEYS test.
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select KEYS, and press ENTER to start the test. When a key is pressed, the label
name for that key will be displayed to indicate that it is functioning properly. When
the key is released, the message “No keys pressed” is displayed.
5.Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT
aborts the test and returns the instrument to the SELF-TEST MENU. Continue
pressing EXIT to back out of the menu structure.
4-4TroubleshootingModel 2425 Service Manual
DISPLAY PATTERNS test
The display test lets you verify that each pixel and annunciator in the vacuum fluorescent
display is working properly. Perform the following steps to run the display test:
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select DISPLAY-PATTERNS, and press ENTER to start the display test. There
are five parts to the display test. Each time a front panel key (except EXIT) is
pressed, the next part of the test sequence is selected. The five parts of the test
sequence are as follows:
•Checkerboard pattern (alternate pixels on) and all annunciators.
•Checkerboard pattern and the annunciators that are on during normal operation.
•Horizontal lines (pixels) of the first digit are sequenced.
•Vertical lines (pixels) of the first digit are sequenced.
•Each digit (and adjacent annunciator) is sequenced. All the pixels of the
selected digit are on.
5.When finished, abort the display test by pressing EXIT. The instrument returns to
the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the
menu structure.
CHAR SET test
The character set test lets you display all characters. Perform the following steps to run the
character set test:
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select CHAR-SET, and press ENTER to start the character set test. Press any key
except EXIT to cycle through all displayable characters.
5.When finished, abort the character set test by pressing EXIT. The instrument
returns to the FRONT PANEL TESTS MENU. Continue pressing EXIT to back
out of the menu structure.
Model 2425 Service ManualTroubleshooting4-5
Principles of operation
The following information is provided to support the troubleshooting tests and procedures
covered in this section of the manual. Refer to the following drawings:
Figure 4-1 — Overall block diagram
Figure 4-2 — Analog circuitry block diagram
Figure 4-3 — Power supply block diagram
Figure 4-4 — Output stage simplified schematic
Figure 4-5 — Digital circuitry block diagram
Overall block diagram
Figure 4-1 shows an overall block diagram of the Model 2425. Circuitry may be divided
into three general areas:
•Analog circuits — includes sourcing circuits such as the DACs, clamps, output
stage, and feedback circuits, as well as measurement circuits such as the A/D con
verter.
•Digital circuits — includes the microcomputer that controls the analog section,
front panel, and GPIB and RS-232 ports, as well as associated interfacing circuits.
•Power supplies — converts the AC line voltage into DC voltages that supply the
power for the digital and analog circuits, and the output stage.
-
4-6TroubleshootingModel 2425 Service Manual
Figure 4-1
Overall block diagram
Analog Section
Display,
Keyboard
Digital
I/O
Trigger
Link
Front
Panel
Controller
Trigger,
Digital
I/O
DACs
Clamps
Feedback
A/D
Converter
Microcomputer
Output
Stage
RS-232
GPIB
Interface
Guard
Buffer
Output
Guard
RS-232 I/O
GPIB I/O
To Analog
Circuits
±15V
Analog
Power
Supply
Digital Section
Output Stage
+5V
±42V
Power Supply
To
±150V
Output
Stage
Power
Supply
Line In
To
Digital Circuits
+5V +12V
Digital
Power
Supply
Model 2425 Service ManualTroubleshooting4-7
Analog circuits
Figure 4-2 shows a block diagram of the analog circuits.
D/A converters control the programmed voltage and current, or voltage compliance and
current compliance. Each DAC has two ranges, a 10V full-scale output or a 1V full-scale
output. The DAC outputs are fed to the summing node, FB. Either the V DAC or the I
DAC has the ability to control the main loop. If the unit is set for SV (source voltage), it
will source voltage until the compliance current is reached (as determined by the I DAC
setting), and the current loop will override the voltage loop. If, however, the unit is set for
SI (source current), it will source current until the compliance voltage is reached (as
determined by the V DAC setting), and the voltage loop will override the current loop. A
priority bit in the Vclamp/I clamp circuit controls these functions.
The error amplifier adds open-loop gain and slew-rate control to the system to assure
accuracy and provide a controllable signal for the output stage, which provides the
necessary voltage and current gain to drive the output. Sense resistors in the HI output lead
provide output current sensing, and a separate sense resistor is used for each current range.
The 1A and 3A ranges use 0.2V full-scale for a full-range output, while all other ranges
use 2V output for full-scale current. Voltage feedback is routed either internally or
externally.
Figure 4-2
Analog circuitry block diagram
V DAC
I DAC
Control
FB
VFB
A/D
There are four voltage ranges: 0.2V, 2V, 20V, and 100V. The feedback gain changes only
for the 20V and 100V ranges, resulting in three unique feedback gain values. A multi
IFB
V Clamp
I Clamp
+7
MUX
Stage
+42
+150
+
-
Sense
Resistors
Protection
Protection
O
Output
HI
S+
Output
LO
S-
Guard
Out
Guard
Sense
-42
-150
Error
Amp
VFB
IFB
S
Output
O
Remote
O
-
4-8TroubleshootingModel 2425 Service Manual
plexer directs the voltage feedback, current feedback, reference, or ground signal to the
A/D converter. An opto-isolated interface provides control signals for both DACs, analog
circuit control, and A/D converter communication to the digital section.
Power supply
Figure 4-3 shows a block diagram of the Model 2425 power delivery system.
The offline switching power supply provides all power for the instrument while providing
universal inputs and power factor correction for the 120/240V line. The digital board runs
off of 5V and 12V supplies derived from the switcher. The +12VD supply is set to pro
gram the flash ROM. (See Digital circuitry below.)
A constant-frequency switching supply runs off the +24VD supplies and generates all the
floating and output supply voltages for the analog board: +5V and ±15V, ±42V and
±150V.
Figure 4-3
Power supply block diagram
-
Line
Neutral
Analog Board
+15V+5V
FF
Constant Frequency
Low Noise Floating
Switching Supply
PFC
D
DC/DC
Converter
F
+24Vdc
-15V
+24dc
F
DC
Regulators
Output Stage
-150V -42V+42V +150V
+12Vdc
+5Vdc
O
Digital
Circuits
D
D
Model 2425 Service ManualTroubleshooting4-9
Output stage
Figure 4-4 shows a simplified schematic of the output stage.
The Model 2425 output stage serves two purposes: (1) it converts signals from floating
common to output common, and (2) it provides both voltage and current amplification.
The output stage drive transistors are biased in class B configuration to prevent the possi
bility of thermal runaway with high-current output values.
Output transistors Q518 and Q521 are cascoded with output MOSFETs Q516 and Q523.
All other MOSFETs and transistors are slaves, and the voltages across these devices are
determined by the resistor-capacitor ladder circuits shown. High-current drive capability
is provided by Q500-Q511. Coarse current limits are built into the output stage.
-
4-10TroubleshootingModel 2425 Service Manual
Figure 4-4
Output stage simplified schematic
+85V
Maindrive
O
+15Vf
300K
F
HI Drive
+
-
Q500
+42V
Q504
Q514
Q516
Q518
F
Q521
-15Vf
F
Q523
Q525
Q507
-42V
Q503
HI Drive
-85V
Model 2425 Service ManualTroubleshooting4-11
A/D converter
The SourceMeter unit uses a multi-slope charge balance A/D converter with a single-slope
run-down. The converter is controlled by gate array U610. Commands are issued by the
MPU on the digital board through communications opto-isolators to U610, and U610
sends A/D reading data back through opto-isolators to the digital board for calibration and
processing.
Active guard
The Model 2425 has an active guard or “six-wire ohms” circuit used to measure complex
devices. This circuitry provides a low-current (50mA) equivalent of the voltage on output
HI. If the unit is in the SV mode, the low-current equivalent of the source voltage will
appear on the guard terminal. If the unit is in the SI mode, the voltage on output HI is
equal to the source current multiplied by the external resistance value. An equivalent
voltage will be generated by the guard circuit, and a guard sense terminal is provided to
sense around the voltage drop in the guard leads since significant current can flow
(50mA).
Digital circuitry
Refer to Figure 4-5 for the following discussion on digital circuitry.
The core digital circuitry uses a Motorola 68332 microcontroller running at 16.78MHz.
The memory configuration includes two 256K × 8-bit flash EEPROMs and two 128K × 8bit RAMs used in parallel to utilize the 16-bit data bus of the MPU. The RAM is battery
backed-up, providing continued storage of data buffer information during power-down
cycles, and flash ROM support allows internal firmware upgrades using either the serial or
GPIB port for downloading new firmware. All calibration constants and the save 0 setup
are stored in a separate serial EEPROM. Setups 1 through 4 are stored in battery backedup RAM.
External communication is provided via GPIB and serial interfaces. A 9914 GPIB IEEE488 standard interface IC is used for the GPIB, and a 68332 Queued Serial Module (QSM)
provides the serial UART. For internal communications, the Time Processing Unit (TPU)
is used for serial communications with the front panel display module, and both the TPU
and QSM handle digital-to-analog interfacing.
4-12TroubleshootingModel 2425 Service Manual
Figure 4-5
Digital circuitry block diagram
A/D
Control/Data
Reset
2
E PROM
U17
A/D Interface
U9, U25
Voltage Source
Control
ROM
U15, U16
Microprocessor
U3
RAM
U12, U14
16.78MHz
Serial
Interface
U4
GPIB
U6, U13
U20
To Display
Board Controller
Trigger
U23
RS-232
Interface
IEEE-488
Interface
Trigger
Display board circuit theory
Display board components are shown in the digital circuitry block diagram in Figure 4-5.
U902 is the display microcontroller that controls the VFD (vacuum fluorescent display)
and interprets key data. The microcontroller has four peripheral I/O ports that are used for
the various control and read functions.
Display data is serially transmitted to the microcontroller from the digital board via the
TXB line to the microcontroller PD0 terminal. In a similar manner, key data is serially
Digital I/O
U7
Digital
I/O
Model 2425 Service ManualTroubleshooting4-13
sent back to the digital board through the RXB line via PD1. The 4MHz clock for the
microcontroller is generated on the digital board.
DS901 is the VFD (vacuum fluorescent display) module, which can display up to 49 characters. Each character is organized as a 5 × 7 matrix of dots or pixels and includes a long
under-bar segment to act as a cursor.
The display uses a common multiplexing scheme with each character refreshed in
sequence. U903 and U904 are the grid drivers, and U901 and U905 are the dot drivers.
Note that dot driver and grid driver data is serially transmitted from the microcontroller
(PD3 and PC1).
The front panel keys (S901-S931) are organized into a row-column matrix to minimize the
number of microcontroller peripheral lines required to read the keyboard. A key is read by
strobing the columns and reading all rows for each strobed column. Key down data is
interpreted by the display microcontroller and sent back to the main microprocessor using
proprietary encoding schemes.
Troubleshooting
Troubleshooting information for the various circuits is summarized below.
Display board checks
If the front panel display tests indicate that there is a problem on the display board, use
Table 4-1. See “Principles of operation” for display circuit theory.
Table 4-1
Display board checks
StepItem/componentRequired conditionRemarks
1
Front panel test
2
J1033
3
U902, pin 1
4
U902, pin 43
5
U902, pin32
6
U902, pin 33
Verify that all segments operate.
+5V, ±5%
Goes low briefly on power up, and then
goes high.
4MHz square wave.
Pulse train every 1 ms.
Brief pulse train when front panel key is
pressed.
Use front panel display test.
Digital +5V supply.
Microcontroller RESET.
Controller 4MHz clock.
Control from main processor.
Key down data sent to main
processor.
4-14TroubleshootingModel 2425 Service Manual
Power supply checks
Power supply problems can be checked using Table 4-2. See “Principles of operation” for
circuit theory on the power supply.
Table 4-2
Power supply checks
StepItem/componentRequired conditionRemarks
1
2
3
4
5
6
7
8
9
Line fuse
Line power
TP502
TP503
TP504
TP505
TP507
TP508
TP510
Check continuity.
Plugged into live recepta-
cle, power on.
+150V, ±5%
-150V, ±5%
+38V, ±10%
-38V, ±10%
+15V, ±5%
-15V, ±5%
+5V, ±5%
Remove to check.
Check for correct power-up
sequence.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
+15VF, referenced to
TP500.
-15VF, referenced to TP500.
+5VF, referenced to TP500.
Model 2425 Service ManualTroubleshooting4-15
Digital circuitry checks
Digital circuit problems can be checked out using Table 4-3. See “Principles of operation”
for a digital circuit description.
Table 4-3
Digital circuitry checks
Step Item/componentRequired conditionRemarks
1
Power-on test
2
U3 pin 19
3
U3 pin 7
4
U3 pin 68
5
U3, lines A0-A19
6
U3, lines D0-D15
7
U3 pin 66
8
U4 pin 7
9
U4 pin 8
10
U13 pins 34-42
11
U13 pins 26-31
12
U13 pin 24
13
U13 pin 25
14
U3 pin 43
15
U3 pin 44
16
U3 pin 45
17
U3 pin 47
RAM OK, ROM OK.
Digital common.
+5V
Low on power-up, then goes high.
Check for stuck bits.
Check for stuck bits.
16.78MHz.
Pulse train during RS-232 I/O.
Pulse train during RS-232 I/O.
Pulse train during IEEE-488 I/O.
Pulses during IEEE-488 I/O.
Low with remote enabled.
Low during interface clear.
Pulse train.
Pulse train.
Pulse train.
Pulse train.
Verify that RAM and ROM are functional.
All signals referenced to digital common.
Digital logic supply.
MPU RESET line.
MPU address bus.
MPU data bus.
MPU clock.
RS-232 RX line.
RS-232 TX line.
IEEE-488 data bus.
IEEE-488 command lines.
IEEE-488 REN line.
IEEE-488 IFC line.
D_ADDATA
D_DATA
D_CLK
D_STB
>100V voltage protection
SOURCE +10V
SOURCE +10V (SVMI)
SOURCE +10V
SOURCE +10V
OUTPUT COM
OUTPUT COM
SVMI, OUTPUT ON, 20V
Bench defaults
-13V ±1V
-5V ±0.5V
-10V ±1V
-10.5V ±1V
0V ±0.1V
7V ±0.7V
7V ±0.7V
20V ±0.5V
6.4V ±0.6V
1
2
3
4
5
6
7
8
9
1
Measured with respect to FCOM (TP500).
2
Measured with respect to OCOM (TP501).
Battery replacement
WARNINGDisconnect the instrument from the power line and all other equip-
ment before changing the battery.
The volatile memories of the Model 2425 are protected by a replaceable battery when
power is off. Typical battery life is approximately 10 years. The battery should be sus
pected if the instrument no longer retains buffer data or user-defined operating parameters
such as instrument setups, source memory, and math expressions. If the battery is absent
or totally exhausted, the display will show the “Reading buffer data lost” message shortly
after the Model 2425 is switched on.
The battery is a 3V wafer-type lithium cell, Panasonic type BR-2330 or equivalent
(Keithley part number BA-46), which is located on the digital board. Replacement of the
battery requires removal of the case cover and analog board assembly. (See Section 5.)
-
Model 2425 Service ManualTroubleshooting4-17
Battery replacement precautions
WARNINGThe following precautions must be followed to avoid personal injury.
1.Wear safety glasses or goggles when working with lithium batteries.
2.Do not short the battery terminals together.
3.Keep lithium batteries away from all liquids.
4.Do not attempt to recharge lithium batteries.
5.Observe proper polarity when installing the battery.
6.Do not incinerate or otherwise expose the battery to excessive heat (>60°C).
7.Bulk quantities of lithium batteries should be disposed of as hazardous waste.
Battery replacement procedure
1.Remove the case cover and analog board assembly as covered in Section 5.
2.Locate the battery on the digital board.
3.Carefully unsolder and remove the old battery.
4.Install and solder the new battery in place.
5.Re-assemble the instrument, and turn it on. The “Reading buffer data lost” error
message will be displayed.
6.Send the :SYST:MEM:INIT command via remote to perform the following:
•Clear the reading buffer.
•Initialize instrument setups 1 through 4 to present instrument settings.
•Initialize all 100 source memory locations to present instrument settings.
•Delete user math expressions.
4-18TroubleshootingModel 2425 Service Manual
No comm link error
A “No Comm Link” error indicates that the front panel processor has stopped communicating with the main processor, which is located on the digital board. This error indicates
that one of the main processor ROMs may require re-seating in its socket. ROMs may be
reseated as follows:
1.Turn off the power, and disconnect the line cord and all other test leads and cables
from the instrument.
2.Remove the case cover as outlined in Section 5.
3.Remove the analog board assembly as outlined in Section 5.
4.Locate the two firmware ROMS, U15 and U16, located on the digital board. These
are the only ICs installed in sockets. (Refer to the component layout drawing at the
end of Section 6 for exact locations.)
5.Carefully push down on each ROM IC to make sure it is properly seated in its
socket.
CAUTIONBe careful not to push down excessively; digital board could crack.
6.Connect the line cord, and turn on the power. If the problem persists, additional
troubleshooting will be required.
5
Disassembly
5-2DisassemblyModel 2425 Service Manual
Introduction
This section explains how to handle, clean, and disassemble the Model 2425. Disassembly
drawings are located at the end of this section.
Handling and cleaning
To avoid contaminating PC board traces with body oil or other foreign matter, avoid
touching the PC board traces while you are repairing the instrument. Motherboard areas
covered by the shield have high-impedance devices or sensitive circuitry where contami
nation could cause degraded performance.
Handling PC boards
Observe the following precautions when handling PC boards:
•Wear cotton gloves.
•Only handle PC boards by the edges and shields.
•Do not touch any board traces or components not associated with repair.
•Do not touch areas adjacent to electrical contacts.
•Use dry nitrogen gas to clean dust off PC boards.
-
Solder repairs
Observe the following precautions when you must solder a circuit board:
•Use an OA-based (organic activated) flux, and take care not to spread the flux to
other areas of the circuit board.
•Remove the flux from the work area when you have finished the repair by using
pure water with clean, foam-tipped swabs or a clean, soft brush.
•Once you have removed the flux, swab only the repair area with methanol, then
blow-dry the board with dry nitrogen gas.
•After cleaning, allow the board to dry in a 50°C, low-humidity environment for
several hours.
Model 2425 Service ManualDisassembly5-3
Static sensitive devices
CMOS devices operate at very high impedance levels. Therefore, any static that builds up
on you or your clothing may be sufficient to destroy these devices if they are not handled
properly. Use the following precautions to avoid damaging them:
CAUTIONMany CMOS devices are installed in the Model 2425. Handle all
semiconductor devices as being static sensitive.
•Transport and handle ICs only in containers specially designed to prevent static
build-up. Typically, you will receive these parts in anti-static containers made of
plastic or foam. Keep these devices in their original containers until ready for
installation.
•Remove the devices from their protective containers only at a properly grounded
work station. Also, ground yourself with a suitable wrist strap.
•Handle the devices only by the body; do not touch the pins.
•Ground any printed circuit board into which a semiconductor device is to be
inserted to the bench or table.
•Use only anti-static type desoldering tools.
•Use only grounded-tip solder irons.
•Once the device is installed in the PC board, it is normally adequately protected,
and you can handle the boards normally.
Assembly drawings
Use the assembly drawings located at the end of this section to assist you as you disassemble and re-assemble the Model 2425. Also, refer to these drawings for information about
the Keithley part numbers of most mechanical parts in the unit. Assembly drawings
include:
Follow the steps below to remove the case cover to gain access to internal parts.
WARNINGBefore removing the case cover, disconnect the line cord and any test
leads from the instrument.
1.Remove handle — The handle serves as an adjustable tilt-bail. Adjust its position
by gently pulling it away from the sides of the instrument case and swinging it up
or down. To remove the handle, swing the handle below the bottom surface of the
case and back until the orientation arrows on the handles line up with the orienta
tion arrows on the mounting ears. With the arrows lined up, pull the ends of the
handle away from the case.
2.Remove mounting ears — Remove the screw that secures each mounting ear. Pull
down and out on each mounting ear.
NOTEWhen re-installing the mounting ears, make sure to mount the right ear to the
right side of the chassis, and the left ear to the left side of the chassis. Each ear
is marked “RIGHT” or “LEFT” on its inside surface.
-
3.Remove rear bezel — To remove the rear bezel, loosen the two screws that secure
the rear bezel to the chassis, then pull the bezel away from the case.
4.Remove grounding screws — Remove the two grounding screws that secure the
case to the chassis. They are located on the bottom of the case at the back.
5.Remove chassis — To remove the case, grasp the front bezel of the instrument,
and carefully slide the chassis forward. Slide the chassis out of the metal case.
NOTETo gain access to the components under the analog board shield, remove the
shield, which is secured to the analog board by a single screw.
Model 2425 Service ManualDisassembly5-5
Analog board removal
Perform the following steps to remove the analog board. This procedure assumes that the
case cover is already removed.
1.Remove the small pulse board before removing the analog board.
2.Disconnect the front and rear input terminals.
You must disconnect these input terminal connections for both the front and rear
inputs:
•INPUT/OUTPUT HI and LO
•4-WIRE SENSE HI and LO
•V, Ω, GUARD and GUARD SENSE (rear panel only)
Remove all the connections by pulling the wires off the pin connectors, then
remove the ferrite noise filters from the chassis. During re-assembly, use the fol
lowing table to identify input terminals:
Input terminalsFront wire color Rear wire color
-
INPUT/OUTPUT HI
INPUT/OUTPUT LO
4-WIRE SENSE HI
4-WIRE SENSE LO
V, Ω, GUARD
GUARD SENSE
3.Unplug cables.
•Carefully unplug the ribbon cables at J1027, J1028, and J1029.
•Unplug the ON/OFF cable at J1034.
4.Remove screws.
•Remove two fastening screws that secure the analog board assembly to the
chassis. These screws are located on the side of the board opposite from the
heat sink.
•Remove two screws that secure the heat sink to the chassis.
Red
Black
Yellow
Gray
-
-
White/Red
White/Black
White/Yellow
White/Gray
White
Blue/White
5-6DisassemblyModel 2425 Service Manual
5.Remove analog board assembly.
After all screws have been removed, carefully lift the analog board assembly free
of the main chassis.
6.Disassemble analog board assembly.
•Remove the screws that secure the analog board and heat sink to the analog
board subchassis.
•Carefully remove the heat sink by sliding the clips off the power transistors.
CAUTIONBe careful not to damage the heat sink insulation layer.
•Remove the analog board from the subchassis.
•Remove four screws that secure the bottom cover, then remove the cover from
the bottom of the PC board.
NOTEWhen re-installing the heat sink, make sure that all clips are properly installed
and centered on each pair of output transistors.
Digital board removal
Perform the following steps to remove the digital board. This procedure assumes that the
analog board assembly is already removed.
NOTEIn order to remove the digital board, the display board must first be removed.
1.Remove IEEE-488, Digital I/O, and RS-232 fasteners.
The IEEE-488, Digital I/O, and RS-232 connectors each have two nuts that secure the
connectors to the rear panel. Remove these nuts.
2.Remove POWER switch rod.
At the switch, place the edge of a flat-blade screwdriver in the notch on the pushrod.
Gently twist the screwdriver while pulling the rod from the shaft.
3.Unplug cables:
•Unplug the display board ribbon cable.
•Unplug the cables going to the power supply.
•Unplug the rear panel power module cable.
•The fan may need to be removed.
4.Remove digital board.
Slide the digital board forward until it is free of the guide pins, then remove the board.
During re-assembly, replace the board, and start the IEEE-488, Digital I/O, and RS-232
connector nuts and the mounting screw. Tighten all the fasteners once they are all in
place and the board is correctly aligned.
Model 2425 Service ManualDisassembly5-7
Front panel disassembly
Use the following procedures to remove the display board and/or the pushbutton switch
pad:
1.Unplug the display board ribbon cables.
2.Remove front panel assembly.
This assembly has four retaining clips that snap onto the chassis over four pem nut
studs. Two retaining clips are located on each side of the front panel. Pull the
retaining clips outward and, at the same time, pull the front panel assembly for
ward until it separates from the chassis.
3.Using a thin-bladed screwdriver, pry the plastic PC board stop (located at the bot-
tom of the display board) until the bar separates from the casing. Pull the display
board from the front panel.
4.Remove the switch pad by pulling it from the front panel.
-
Removing power components
The following procedures to remove the power supply and/or power module require that
the case cover and motherboard be removed, as previously explained.
Power module removal
Perform the following steps to remove the rear panel power module:
WARNINGTo avoid electrical shock, which could result in injury or death, the
ground wire of the power module must be connected to chassis
ground. When installing the power module, be sure to re-connect the
green and yellow ground wire to the threaded stud on the chassis.
1.Remove the analog board.
2.Unplug the cable connecting the power module to the digital board.
3.Disconnect the power module’s ground wire. This green and yellow wire connects
to a threaded stud on the chassis with a kep nut.
4.Squeeze the latches on either side of the power module while pushing the module
from the access hole.
5-8DisassemblyModel 2425 Service Manual
Instrument re-assembly
WARNINGTo ensure continued protection against electrical shock, verify that
power line ground (green and yellow wire attached to the power mod
ule) is connected to the chassis.
Also make sure the two bottom case screws are properly installed to
secure and ground the case cover to the chassis.
Re-assemble the instrument by reversing the previous disassembly procedures. Make sure
that all parts are properly seated and secured, and that all connections are properly made.
To ensure proper operation, replace the analog signal wire ferrite noise filters, and
securely fasten the shield.
-
6
Replacement Parts
6-2Replacement PartsModel 2425 Service Manual
Introduction
This section contains replacement parts information and component layout drawings for
the Model 2425.
Parts lists
The electrical parts lists for the Model 2425 are shown in tables at the end of this section.
For part numbers to the various mechanical parts and assemblies, use the Miscellaneous
parts list and the assembly drawings provided at the end of Section 5.
Ordering information
To place an order, or to obtain information concerning replacement parts, contact your
Keithley representative or the factory (see inside front cover for addresses). When ordering parts, be sure to include the following information:
•Instrument model number (Model 2425)
•Instrument serial number
•Part description
•Component designation (if applicable)
•Keithley part number
Factory service
If the instrument is to be returned to Keithley Instruments for repair, perform the
following:
•Call the Repair Department at 1-800-552-1115 for a Return Material Authorization
(RMA) number.
•Complete the service form at the back of this manual, and include it with the
instrument.
•Carefully pack the instrument in the original packing carton.
•Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label.
Component layouts
The component layouts for the various circuit boards are provided on the following pages.
Drawings include:
IRF9630
R117-R128,R133-R136,R141-R144RES, 10K, 5%, 250MW, METAL
R-376-10K
FILM
R129,R155,R211,R224-R229,R234-R237RES, 10K, 1%, 100MW, THICK FILM R-418-10K
R130RES, 10K, 1%, 100MW, THICK FILM R-418-10K
R137-140,R145-149,R713,R714RES, 10K, 5%, 250MW, METAL