Kanomax Dif-Kit Specifications

Dif-Kit
Tracer Gas Hardware
Professional Fume Hood Diagnostic Tools
The ANSI/ASHRAE 110 test is a method of testing the performance of laboratory fume hoods. The Kanomax Dif-Kit is ideal for use in performing the Tracer Gas test in accordance with the ANSI/ASHRAE Standard 110-1995. Kanomax also o ers alternative tracer gas systems due to concerns to prevent greenhouse gas emissions.
Features & Benefi ts
Diff user is made to the design and specifi cations of
Standard drawing #110-83M
The internal critical orifi ce ensures a fl ow rate of 4
liters per minute
Alternative tracer gas systems are available - SF6,
NO2, or FM-200 (HFC-227)
Thermal anemometer for face velocity testing
Commercial fog machine for fl ow visualization
Set-up Diagram for Tracer Gas Test
Dif-Kit Speci cations
Critical Ori ce 4 liter per minute Regulator Dual Stage, Specialty Gas Flow Meter Calibrated 150 mm/200 psi, Glass Tube Pressure Gauge 0 to 60 psi
Speci cations are subject to change without notice.
Fog Machine Speci cations
Particle Size 0.25 to 60 microns Max Consumption 1 liter per hour Power Requirements 120 volts, 50/60 Hz Dimensions 14.3” x 5.6” x 7.5”
Speci cations are subject to change without notice.
DIF-KIT Parts & Pieces
8.7.18REV
Distributed by:
Kanomax USA, Inc.
P.O. Box 372 219 US Hwy 206, Andover, NJ 07821 U.S.A. TEL: 1-973-786-6386 E-mail: sales@kanomax-usa.com URL: www.kanomax-usa.com
Copyright © 2018 by KANOMAX USA, Inc.
Loading...