Dif-Kit
Tracer Gas Hardware
Professional Fume Hood Diagnostic Tools
The ANSI/ASHRAE 110 test is a method of testing the
performance of laboratory fume hoods. The Kanomax
Dif-Kit is ideal for use in performing the Tracer Gas test in
accordance with the ANSI/ASHRAE Standard 110-1995.
Kanomax also o ers alternative tracer gas systems due
to concerns to prevent greenhouse gas emissions.
Features & Benefi ts
• Diff user is made to the design and specifi cations of
Standard drawing #110-83M
• The internal critical orifi ce ensures a fl ow rate of 4
liters per minute
• Alternative tracer gas systems are available - SF6,
NO2, or FM-200 (HFC-227)
• Thermal anemometer for face velocity testing
• Commercial fog machine for fl ow visualization
Set-up Diagram for Tracer Gas Test
Dif-Kit Speci cations
Critical Ori ce 4 liter per minute
Regulator Dual Stage, Specialty Gas
Flow Meter Calibrated 150 mm/200 psi, Glass Tube
Pressure Gauge 0 to 60 psi
Speci cations are subject to change without notice.
Fog Machine Speci cations
Particle Size 0.25 to 60 microns
Max Consumption 1 liter per hour
Power Requirements 120 volts, 50/60 Hz
Dimensions 14.3” x 5.6” x 7.5”
Speci cations are subject to change without notice.