Applications
Cleanroom Monitoring
Cleanroom ISO Certification
USP 797 cerification
Facility Monitoring
Filter Testing
Features
Particle size: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 Ǵm
Sampling airflow: 28.3L/min
Air velocity range: 0 to 197 fpm
Temperature range: 32 to 122F *
Relative humidity range: 3 to 98%RH *
Differential pressure range: +/-100Pa *
Built-in printer
Power supply: Li-ion battery or AC 100 - 240V
Data logging: Store up to 10,000 measurements
PC communication via Ethernet
* Require option sensor
*
Benefits
Large touch screen
Small footprint
Display 6 particle sizes simultaneously in cf or m
Capable of remote control
Multi-parameter: Particle, Air velocity, Temperature, Humidity, Pressure
Multi-standard mode: ISO, FS 209E, BS 5295, EC GMP
Applications
3
Facility Monitoring
Airborne Contamination Control
Cleanroom Certification
Airborne Particle Counter Model 3900
Specifications
Light Source
Channel Sizes (Ǵm)
Flow Rate
Zero Count Level
Counting Efficiency
Coincidence Loss
Max Detectable Concentration
Sample Time
Delay Time
Sampling Cycles
Location Labels
Display
Data Storage
Printer
Hardware Interface
Operating Time (Battery)
Battery Type
Charging Time
Power
Dimensions
Weight
Enclosure Material
Operating Environment
Warranty
Standard Accessories
Laser Diode
0.3, 0.5, 1.0, 3.0, 5.0, 10.0
28.3 lpm (1.0 cfm)
1 count or less in 5 minutes (meets JIS B9921-1997)
50% at 0.3Ǵm (meets JIS B9921-1997)
Less than 5% at 500,000 particle/cf
500,000 counts/cf (17,667.8 counts/L)
1 second to 23 hrs 59 min 59 sec
0 second to 23 hrs 59 min 59 sec
1 - 9,999 while in repeat mode
0 - 999 appears on display and printout
6.4 inch color touch screen
5,000 samples (when using standard 512MB CF card)
Built in (compatible with Dust-free Paper)
USB, Ethernet
4 hours continuous operation
Removable Li-ion
6 hours
AC 100 - 240V, 50 to 60Hz
W 8.3 x D 8.6 x H 12.5 inches (21 x 22 x 32 cm)
17.5 lbs (8 kg)
Stainless Steel
50 to 104qF (10 to 40)
1 year (Instrument) / 2 years (Laser Diode)
Operation Manual, Isokinetic Probe, Power Cable, Battery & Charger
Printer Paper (2 rolls), Zero Filter, USB Cable, Data Download Software (Windows)
Optional Accessories
* For safe and trouble free operations, please read Operation Manual carefully before using Kanomax instruments.
Kanomax USA, Inc.
219 US Hwy. 206, Andover, NJ 07821 U.S.A.
TEL: 1-800-247-8887 (USA) ޓ1-973-786-6386
FAX: 1-973-786-7586
E-mail: info@kanomax-usa.com
URL: www.kanomax-usa.com
Carrying Case, Temperature and Humidity Probe, Air Velocity Probe
Differential Pressure Sensor
Distributed by:
Specifications subject to change without prior notice
20090630