Kanomax 3888, 3889 Specifications

Page 1
Handheld Particle Counter
Model 3888 & 3889
The Model 3888 (3-channel) and 3889 (6-channel) Handheld Particle Counters utilize the same, proven parts that made the original Handheld Particle Counters so reliable. The instruments meet/exceed new ISO 21501-4 Standards and are fully RoHS compliant. The instruments now have a base/stand for advanced communication capabilities (Wi-Fi, Ethernet, RS485), a larger, full color touchscreen display, and enables real time monitoring while connected to a PC.
Applications:
· Monitor and Verify Cleanrooms (Pharmaceutical, Semi Conductor, Food Industry)
· Monitor Surgical Rooms
· Filter Testing
· IAQ Investigation
· Data Center Cleaning
3888 measures particle sizes: 0.3µm, 0.5µm & 5.0µm
3889 measures particle sizes 0.3µm, 0.5µm, 1.0µm,
3.0µm, 5.0µm, 10.0µm
Easy-to-use interface is displayed on a 4.3 in. color
esistive touch panel screen
LCD, r
On-board data storage up to 10,000 measurements
Creates Microsoft Excel compatible .CSV file without the need for additional software and user can download it via USB memory
Kanomax Measuring Software displays and graphs the measurement data in real-time while connected to PC
Language options include English, Spanish, Japanese, and Chinese
Sleek/Modern
Instrument Design
Windows 10
Compatible Software
Page 2
HandheldParticleCounterModels3888/3889Specifications
Product
Handheld Particle Counter
3888 (3-channel model)
3 channels (0.3, 0.5, 5.0µm
0.1CFM (2.83L/min)
Maximum
50±20%for PSL particles near the minimum measurable size
1 particle/5 minutes
15% (for PSL particles near 0.3µm)
Display
4.3 inch color LCD, Resistive touch panel
USB (Host: for printer and USB ash memory, Stand (option)
RS485, Ethernet, and WiFi connection points in the base/stand
Media
Internal memory
Number and format
Up to 10,000 records in CSV format
Language
English, Japanese, Chinese, and Spanish
Internal supply
Li-ion rechargeable battery (Replaceable
External supply
AC adaptor Input 100 to 240 V
Continuous operation
Operating
Main body
From 10 to 40 , from 0 to 85%RH (With no condensation
Probe
From 0 to 50 , from 2 to 98%RHWith no condensation
Dimension
W213 X H100 X D55 mm
Weight
650g
portable
TemperatureFrom 0 to 50
Temperature0.1
Humidity accuracy
±3.0% (30 to 85%RH), ±5% (other range of humidity)
Temperature Response time
Approximately 60 seconds or less90 response
Model
Particle Measurement
Communication
Recording media
Sizes
Flow rate
measurable Concentration
Counting eciency
False count
Size resolution
Standard
3889 (6-channel model)
6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0µm
Accuracy: ±5% (Compliant with JIS B9921 and ISO21501-4
10% for 2,000,000 particles/CF compliant with JIS B9921 and ISO21501-4
100±10%for PSL particles of 1.5 to 2 times as large as the minimum measurable size
Compliant with JIS B9921 and ISO21501-4)
Compliant with JIS B9921 and ISO21501-4)
Compliant with JIS B9921 and ISO21501-4)
Device: for PC
Power
environment
Hydrothermal Measurement (Available as an option for the 6­channel particle counter)
time
Measurement range
Display resolution
accuracy
5 hoursSingle measurement
HumidityFrom 2.0 to 98.0RH
Humidity0.1
±0.5
Specications are subject to change without notice.
Distributed by:
Kanomax USA, Inc. 219 US Hwy 206, Andover, NJ 07821 U.S.A. TEL: 1-800-247-8887 (USA) / 1-973-786-6386 FAX: 1-973-786-7586 E-mail: sales@kanomax-usa.com
20170525Rev1
URL: www.kanomax-usa.com