Temperature control devices for research and industry are developed,
Congratulations!
You have made an excellent choice.
JULABO thanks you for the trust you have placed in us.
This operating manual has been designed to help you gain an understanding of the operation and
possible applications of our circulators. For optimal utilization of all functions, we recommend that you
thoroughly study this manual prior to beginning operation.
The JULABO Quality Management System
produced, and distributed according to the requirements of ISO 9001 and
ISO 14001. Certificate Registration No. 01 100044846
Unpacking and inspecting
Unpack the circulator and accessories and inspect them for possible transport damage. Damage should
be reported to the responsible carrier, railway, or postal authority, and a damage report should be
requested. These instructions must be followed fully for us to guarantee our full support of your claim
for protecting against loss from concealed damage. The form required for filing such a claim will be
provided by the carrier.
Printed in Germany Changes without prior notification reserved
13. Installation of electronic module with analog connectors ............................................. 89
14. JULABO Service – Online remote diagnosis ............................................................... 90
15. Cleaning / repairing the unit ......................................................................................... 91
4
Operating manual
JULABO circulators are not suitable for direct temperature control of foods, semi-
medium (bath fluid).
ICC
TCF
ATC
3
RS232
RS485
Pt100
PUMP
SMART
applications.
1. Intended use
JULABO circulators have been designed to control the temperature of specific fluids in a bath tank.
luxury foods and tobacco, or pharmaceutical and medical products.
Direct temperature control means unprotected contact of the object with the bath
1.1. Description
The circulators are operated via the splash-proof keypad. The implemented
microprocessor technology allows to set and to store different values that can be
indicated on the VFD COMFORT-DISPLAY and LCD DIALOG-DISPLAY. Three
menu keys facilitate adjusting setpoints, warning and safety functions and menu
functions.
The integrated programmer allows storing and running temperature and time-
dependent processes.
SL
The control electronics including “ICC - Intelligent Cascade Control“ automatically
adapts the heat supplied to the thermal requirements of the bath.
The TCF - Temperature Control Features allow the user to have access to all
important temperature control parameters. This means: Full control on the control
mode and the chance to manually adjust or adapt control to the specific application.
Absolute Temperature Calibration (ATC3) provides high temperature stability in the
bath. With the 3-point calibration an offset is adjusted at three temperatures to
ensure an accurate temperature pattern at the selected spot in the bath over the full
temperature range.
Electrical connections:
The serial interface, switchable from RS232 to RS485, allows modern process
technology without additional interface.
Connection for Pt100 external sensor for external temperature measurement and
control.
Alarm output for external alarm message or control of JULABO refrigerating baths
or solenoid valve (cooling water).
The electronic module (option) provides 3 further analog connections (alarm input,
standby input, recorder output, programmer input).
The excess temperature protection conforming to IEC 61010-2-010 is a safety
installation independent from the control circuit. This protection can be indicated
and set on the LCD DIALOG-DISPLAY.
The early warning system for low level signals that bath fluid needs to be refilled
before the low level protection conforming to IEC 61010-2-010 causes a complete
shut-down of the main functional elements.
Intelligent pump system: The pump capacity (electronically adjustable via the motor
speed) enables to adapt to varying conditions for internal and external temperature
The products of JULABO ensure safe operation when installed, operated, and maintained according to
common safety regulations. This section explains the potential dangers that may arise when operating
the circulator and also specifies the most important safety precautions to preclude these dangers as far
as possible.
The operator is responsible for the qualification of the personnel operating the units.
The personnel operating the units should be regularly instructed about the dangers involved with
their job activities as well as measures to avert these dangers.
Make sure all persons tasked with operating, installing, and maintaining the unit have read and
understand the safety information and operating instructions.
When using hazardous materials or materials that could become hazardous, the circulator may be
operated only by persons who are absolutely familiar with these materials and the circulator. These
persons must be fully aware of possible risks.
If you have any questions concerning the operation of your unit or the information in this manual, please
contact us!
Contact
Safety instructions for the operator:
You have received a product designed for industrial use. Nevertheless, avoid strikes to the
housing, vibrations, damage to the operating-element panel (keypad, display), and contamination.
Make sure the product is checked for proper condition regularly (depending on the conditions of
use). Regularly check (at least every 2 years) the proper condition of the mandatory, warning,
prohibition and safety labels.
Make sure that the mains power supply has low impedance to avoid any negative effects on
instruments being operated on the same mains.
This unit is designed for operation in a controlled electromagnetic environment. This means that
transmitting devices (e.g., cellular phones) should not be used in the immediate vicinity.
Magnetic radiation may affect other devices with components sensitive to magnetic fields
(e.g., monitors). We recommend maintaining a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C.
Permissible relative humidity: 50% (40 °C).
Do not store the unit in an aggressive atmosphere.
Protect the unit from contamination.
Do not expose the unit to sunlight.
Only qualified personnel is authorized to perform configuration, installation, maintenance and repairs of
the circulator.
Routine operation can also be carried out by untrained personnel who should however be instructed by
trained personnel.
6
SL
Use:
The bath can be filled with flammable materials. Fire hazard!
There might be chemical dangers depending on the bath medium used.
Observe all warnings for the used materials (bath fluids) and the respective instructions (safety data
sheets).
Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in well
ventilated areas.
Only use recommended materials (bath fluids). Only use non-acid and non corroding materials.
When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels to the front of the unit so they are highly visible:
When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible:
1
Warning label W00: Colors: yellow, black
Danger area. Attention! Observe instructions.
(operating manual, safety data sheet)
2
or
2
Mandatory label M018: Colors: blue, white
Carefully read the user information prior to beginning operation.
Scope: EU
Semi S1-0701 Table A1-2 #9
Carefully read the user information prior to beginning operation.
Scope: USA, NAFTA
Particular care and attention is necessary because of the wide operating range.
There are thermal dangers: Burn, scald, hot steam, hot parts and surfaces that can be touched.
Warning label W26: Colors: yellow, black
Hot surface warning.
(The label is put on by JULABO)
Observe the instructions in the manuals for instruments of a different make that you connect to the
circulator, particularly the respective safety recommendations. Also observe the pin assignment of
plugs and technical specifications of the products.
7
Operator responsibility – Safety recommendations
waste is not permitted!
2.1. Disposal
The circulator contains a back-up battery that supplies voltage to memory chips when the unit is
switched off. Do not dispose of the battery with household waste!
Depending on battery regulations in your country, you might be obliged to give back used or defect
batteries to gathering places.
The product may be used with oil as bath fluid. These oils fully or partially consist of mineral oil or
synthetic oil. For disposal, observe the instructions in the safety data sheets.
This unit contains refrigerants, which at this time are not considered harmful to the ozone layer.
However, over the long operating period of the unit, disposal rules may change. Therefore, only
qualified personnel should handle the disposal.
Valid in EU countries
See the current official journal of the European Union – WEEE directive.
Directive of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE).
This directive requires electrical and electronic equipment marked with a crossedout trash can to be disposed of separately in an environmentally friendly manner.
Contact an authorized waste management company in your country.
Disposal with household waste (unsorted waste) or similar collections of municipal
2.2. Warranty conditions
JULABO GmbH warrants its products against defects in material or in workmanship, when used under
appropriate conditions and in accordance with appropriate operating instructions
for a period of ONE YEAR.
Extension of the warranty period – free of charge
With the ‘1PLUS warranty’ the user receives a free of charge extension to the warranty of up to 24
months, limited to a maximum of 10 000 working hours.
To apply for this extended warranty the user must register the unit on the JULABO web site
www.julabo.com, indicating the serial no. The extended warranty will apply from the date of JULABO
GmbH’s original invoice.
JULABO GmbH reserves the right to decide the validity of any warranty claim. In case of faults arising
either due to faulty materials or workmanship, parts will be repaired or replaced free of charge, or a new
replacement unit will be supplied.
Any other compensation claims are excluded from this guarantee.
8
2.3. EC Conformity
SL
9
Operator responsibility – Safety recommendations
10
SL
11
Operator responsibility – Safety recommendations
Calibration Baths
FK30-SL
FK31-SL
Working temperature range
°C
-30 ... 200
-30 ... 200
Temperature stability
(1
°C
±0.005
±0.005
Temperature uniformity
°C
<±0.01
<±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20
0.46 0.34 0.15
+20 0 -20
0.46 0.34 0.15
Refrigerant
R134a
R134a
Overall dimensions (WxDxH)
cm
32x45x79
32x45x91
Bath opening (WxL)
cm
∅12
∅12
Bath depth
cm
17
31
Filling volume
Liter
14
24
Weight
kg
48
51
Mains power connection
V/ Hz
230 / 50
230 / 50
A
5 (Cooling machine)
4 (Cooling machine)
Mains power connection
V/ Hz
115 / 60
115 / 60
A
7 (Cooling machine)
7 (Cooling machine)
Circulator (SL-C)
SL
Temperature selection
digital
via keypad
indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY (°C/°F)
Resolution
°C
0.01
ATC3 INT / EXT
°C
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
kW
2.0
Heater wattage (at 115 V)
kW
1.0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate at 0 bar
l/min
22 ... 26
Max. pressure at 0 liters
bar
0.7 Max. suction at 0 liters
bar
0.4
Ambient temperature
°C
5 ... 40
Mains power connection
V/ Hz
208-230 / 50/60
Current draw (at 208 V / 230 V)
A
8 / 9
Mains power connection
V/ Hz
100-115 / 50/60
Current draw (at 100 V / 115 V)
A
9 / 10
2.4. Technical specifications
Current draw (at 230 V)
Current draw (at 230 V) <GB>
A
A
Current draw (at 100 V / 115 V) A
(1
±0.02 °C@+200 °C
14
9 (Circulator)
9/10 (Circulator)
14
10 (Circulator)
9/10 (Circulator)
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C Technical changes without prior notification reserved.
12
Electrical connections:
Caution:
Notice:
compatibility may be impacted.
External alarm device 24-0 V DC / max. 25 mA
Computer interface RS232 or RS485
External Pt100 sensor
Optional for HL, SL
(Order No. 8900100 Electronic module with analog connections)
Programmer input -100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA
Input for the signal of a flow meter or external manipulated variable
Temperature recorder outputs 0 - 10 V (0 V = -100 °C, 10 V = 400 °C)
0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)
4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C)
Standby input for external emergency switch-off
Alarm output for external alarm signal
Safety installations according to IEC 61010-2-010:
Excess temperature protection adjustable from 0 °C ... 320 °C
Low liquid level protection float switch
Classification according to DIN 12876-1 class III
Supplementary safety installations
Early warning system for low level float switch
High temperature warning function optical + audible (in intervals)
Low temperature warning function optical + audible (in intervals)
Supervision of working sensor plausibility control
Reciprocal sensor monitoring between
working and safety sensors difference >35 K
Alarm message optical + audible (permanent)
Warning message optical + audible (in intervals)
Environmental conditions according to IEC 61 010-1:
Use indoors only.
Altitude up to 2000 m - normal zero.
Ambient temperature: see Technical specifications
Humidity:
Max. relative humidity 80% for temperatures up to +31 °C,
linear decrease down to 50% relative humidity at a temperature of +40 °C
Max. mains voltage fluctuations of ±10% are permissible.
Protection class according to IEC 60 529 IP21
The unit corresponds to Class I
Overvoltage category II
Pollution degree 2
SL
EMC requirements
The device is an ISM device of group 1 per CISPR 11 (uses HF for internal purposes) and is classified
in class A (industrial and commercial sector).
The unit is not suitable for use in explosive environment
•Devices of class A are intended for the use in an industrial electromagnetic
environment.
•When operating in other electromagnetic environments, their electromagnetic
13
Operator responsibility – Safety recommendations
Information about the used refrigerants
The Regulation (EU) No. 517/2014 on fluorinated greenhouse gases applies to all systems which
contain fluorinated refrigerants and replaces (EC) 842/2006.
The aim of the Regulation is to protect the environment by reducing emissions of fluorinated
greenhouse gases.
Among other things it regulates the emission limits, use and recovery of these substances. It also
contains requirements for operators of systems which require / contain these substances to function.
Under Regulation 517/2014, the operator of a system of this nature has the following duties:
• The operator must ensure that the equipment is checked at regular intervals for leaks.
• These intervals depend on the CO
refrigerant fill volume and type of refrigerant. The CO
equivalent of the system. This is calculated from the
2
equivalent of your system is shown on
2
the model plate.
• The operator undertakes to have maintenance, repair, service, recovery and recycling work
carried out by certified personnel who have been authorized by JULABO.
• All such work must be documented. The operator must keep records and archive them for at
least five years. The records must be submitted to the relevant authority on request.
Refer to the text of the Regulation for further information.
14
In addition to the safety warnings listed, warnings are posted throughout the operating
Warning:
Caution:
Notice:
in its surroundings can be damaged.
Note!
Draws attention to something special.
Important!
Indicates usage tips and other useful information.
This icon is used in the operating instructions to indicate flashing values or parameters
which have to be set or confirmed.
Operating instructions
3. Safety notes for the user
3.1. Explanation of safety notes
manual. These warnings are designated by an exclamation mark inside an equilateral
triangle. “Warning of a dangerous situation (Attention! Please follow the
documentation).”
The danger is classified using a signal word.
Read and follow these important instructions for averting dangers.
Describes a possibly highly dangerous situation. If these instructions are not
followed, serious injury and danger to life could result.
Describes a possibly dangerous situation. If this is not avoided, slight or minor
injuries could result. A warning of possible property damage may also be contained in
the text.
SL
Describes a possibly harmful situation. If this is not avoided, the product or anything
3.2. Explanation of other notes
15
Safety notes for the user
these parts.
3.3. Safety recommendations
Follow the safety instructions to avoid personal injury and property damage. Also, the
valid safety instructions for workplaces must be followed.
• Only connect the unit to a power socket with an earthing contact (PE – protective
earth)!
• The power supply plug serves as a safe disconnecting device from the line and
must always be easily accessible.
• Place the unit on an even surface on a base made of nonflammable material.
• Do not stay in the area below the unit.
• Make sure you read and understand all instructions and safety precautions listed in
this manual before installing or operating your unit.
• Adjust excess-temperature safety device below the flash point of the bath fluid.
• Observe the limited working temperature range when using plastic bath tanks.
• Never operate the unit without bath fluid in the bath.
• Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the fluid.
• Prevent water from entering the hot bath oil.
• Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (e.g., by switching the unit
on for a short moment).
• Use suitable connecting tubing.
• Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
• Make sure that the tubing is securely attached.
• Regularly check the tubing for material defects (e.g., for cracks).
• Never operate damaged or leaking units.
• Always turn off the unit and disconnect the mains cable from the power source
before performing any service or maintenance procedures, or before moving the
unit.
• Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
• Always empty the bath before moving the unit.
• Transport the unit with care.
• Sudden jolts or drops may cause damage in the interior of the unit.
• Observe all warning labels.
• Never remove warning labels.
• Never operate units with damaged mains power cables.
• Repairs are to be carried out only by qualified service personnel.
16
Some parts of the bath tank and the pump connections may become extremely hot
during continuous operation. Therefore, exercise particular caution when touching
SL
Caution:
determine the method of decontamination.
example.
The temperature controlling i.e. of fluids in a reactor constitutes normal circulator
practice.
We do not know which substances are contained within these vessels.
Many substances are:
• inflammable, easily ignited or explosive
• hazardous to health
• environmentally unsafe
i.e.: dangerous
The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to reduce the
risks to a minimum.
• Are all tubes and electrical cables connected and installed?
Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.
• Do dangerous steams or gases arise when heating?
Is an exhaust needed when working?
• What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance and
Notice: Check the safety installations at least twice a year!
• Excess temperature protection according to IEC 61010-2-010.
With a screwdriver turn back the adjustable excess temperature protection until the
shut-down point (actual temperature).
• Low level protection according to IEC 61010-2-010.
To check the function of the float, it can be manually lowered with a screwdriver for
17
Operating controls and functional elements
1a
2
PI Capacity in % - with manipulated variable set to >control<*
4. Operating controls and functional elements
Front view Rear view
1b
18
Mains power switch, illuminated for circulator
Mains power switch, illuminated for cooling machine
VFD COMFORT-DISPLAY
Header: Control indicators
Line 1: Actual value internal or external
The display is depending on the selected control mode in the
menu > Control < (internal or external).
or
Actual value of the precision reference sensor „C“ xxx.xx
Line 2: Working temp. setpoint, constantly S xxx.xx
Line 3: Actual value (E = external or I = internal or C= reference sensor)
Alternating with the display in line 1
Use the keys
to indicate further values in line 3
PS Capacity in % - with manipulated variable set to >serial<* or
2.1
R
2.2
3
LCD DIALOG-DISPLAY
Navigation aid in MENU - Window
>eprog<*
H Heater capacity in Watts
U Mains voltage Volts
F Flow rate in liters/minute
(providing EPROG input set to >Flow rate<)
*refer to >MENU / CONFIGIGURATIO> page 34
SL
2.3
Control indicators in the header:
Heating / Cooling / Alarm /
emote control
Control indicators in the header:
Temperature indication Internal or External actual value
Temperature indication in °C or °F
Display for the adjusted pump pressure stage in the -OFF- mode.
Display for the effective pump pressure stage (rotation speed) after start.
Four stages, adjustable via the
Line 1: Setpoint and origin of setpoint programming
(Key / RS232 or RS485 / ext. Pt100 / EProg)
Line 2: Actual value - internal or external,
identical to line 1 of the VFD-COMFORT-DISPLAY
Line 3: Heating capacity in %.
Line 4: Control type: internal / external
button, in the menu >PUMP<.
Keys / :
Selection of menu items / parameters
- Setting in line 3:
Actual value / parameter
Orientation aid in MENU - Window
Line 1 - name with allocation to key
T – Setpoint
2 Submenu Level 2
Example: CONFIGURATION
3 Submenu Level 3
Safety values
1 Main menu Level 1
19
Operating controls and functional elements
A beep signals the end of setting
The above graph “front side” shows an example for standard display.
5
Keypad 2
Numeric keypad: numerals 0 to 9
4
4.1
4.2
Keypad 1 Navigation keys
1. Key: >OK< Start / Stop (pump / heater )
2. >OK< in the menu Menu item / select submenu for setting
Save set value
Save selected parameter
After the actions Start, Stop and change from VFD Display to standard
display the key
1. Key: >Return< Stop (pump / heater)
Special display on VFD Display leave line 3.
2. >Return< in the menu one menu level down
Correction function for parameters or values (prior to OK)
immediately back to standard display
- icon for „keep key pressed down“.
is locked for a short time.
4.3
4.4
4.5
4.6
..
1. Key: >Up / Down <temperature – increase/decrease setpoint
Push key quickly for single steps,
Keep key pressed for fast change.
2. >Up/Down< in the menu selection of menu items / parameters
Menu keys
Key: start the menu > warning and safety values<
Key: start the menu >temperature setpoints<
Key: display of MENU structure
minus / decimal point
Backspace key
on LCD DIALOG-DISPLAY.
6
20
Adjustable excess temperature protection according to IEC 61010-2-010
SL
Rear view
7
Socket for external measurement and control sensor
8
9
Socket: control cable of JULABO refrigerated circulator
The circulator automatically recognizes the connected electronic module.
10
ALARM
Alarm output (for external alarm signal)
11
STAND-BY
Standby input (for external emergency switch-off)
12
REG+E-PROG
16a
17a
Mains power cable with plug for circulator
18
Built-in mains outlet for connection of circulator (230 V / 16 A)
19
20
ext Pt100
SERIAL
or external setpoint programming
Interface RS232 / RS485: remote control via personal computer
or Socket for a precision reference sensor
or output for alarm messages
Option: Electronic module Order No. 8 900 100
Programmer input and temperature recorder output
16b
17b
Mains fuses for Circulator: T16A
Mains fuses for cooling machine: T 10,0 A
Mains power cable with plug cooling machine
Drain screw with drain connection
Venting grid, removable
21
Preparations
wait about one hour after setting
The bath cover (1) is mounted on the bath tank with 4 captive screws (2).
Required tool: SW3 hexagonal socket wrench with ball-shaped head
Part No.: 7.901.1088 Gasket for bath cover FK30, FK31
5. Preparations
5.1. Installation
• Place the unit on an even surface on a pad made of non-flammable
material.
• Cooling machine, pump motor and electronics produce intrinsic heat that
is dissipated via the venting openings! Never cover these openings!
• Keep at least 20 cm of open space on the front and rear venting grids.
• The place of installation should be large enough and provide sufficient
air ventilation to ensure the room does not warm up excessively because
of the heat the instrument rejects to the environment. (Max. permissible
ambient temperature: 40 °C).
For a fault (leakage) in the refrigeration system, the standard EN 378
prescribes a certain room space to be available for each kg of
refrigerant.
The refrigerant quantity is specified on the type plate.
> > For 0.25 kg of refrigerant R134a, a room space of 1 m
3
is required.
• Do not set up the unit in the immediate vicinity of heat sources and do
not expose to sun light
• Before operating the unit after transport,
it up. This will allow any oil that has accumulated laterally during
transport to flow back down thus ensuring maximum cooling
performance of the compressor.
The circulating pump transports the
bath fluid to the temperature chamber
(4) at the bottom (3). The overflow on
top of the chamber ensures a constant
fluid level during operation.
The bath cover (5) provides 6
measuring places (6) that can be fitted
®
with different Viton
sleeves..
Order-No.: Viton
®
sleeves
8 930 602 2 mm inner dia.
8 930 603 3 mm inner dia.
8 930 604 4 mm inner dia.
8 930 605 5 mm inner dia.
8 930 605 6 mm inner dia.
8 930 608 8 mm inner dia.
Take off the complete insert for cleaning. The gasket below the bath
cover should be replaced in this case.
22
5.2. Bath fluids
Caution:
corrosion in the bath, even in stainless steel.
Bath fluid
Temperature range
soft/decalcified water
5 °C to 80 °C
mixture water/glycol, mixture 1:1
-20°C to 50°C
Description
G
HY
H5
Order Number
10 liters
8 940 124
8 940 104
8 940 106
5 liters
8 940 125
8 940 105
8 940 107
Temperature range
°C
-30 ... 80
-80 ... 55
-50 ...105
Flash point
°C
--
78
124
Fire point
°C
--
80
142
Color
light yellow
clear
clear
Description
H10
H20S
Order Number
10 liters
8 940 114
8 940 108
5 liters
8 940 115
8 940 109
Temperature range
°C
-20 ... 180
0 ... 220
Flash point
°C
190
230
Fire point
°C
216
274
Color
clear
light brown
Unsuitable bath fluids are fluids which, e.g.,
Carefully read the safety data sheet of the bath fluid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of ≤65 °C is used, only supervised operation is possible.
Water: The quality of water depends on local conditions.
•Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to calcification in the bath.
• Ferrous water can cause corrosion - even on stainless steel.
• Chloric water can cause pitting corrosion.
• Distilled and deionized water is unsuitable. Their special properties cause
Recommended bath fluids:
SL
JULABO bath fluids
JULABO
JULABO
Thermal
Thermal
Thermal
Thermal
Thermal
See website for list of recommended bath fluids.
ATTENTION: The maximum permissible viscosity is 70 mm
Caution:
Fire or other dangers when using bath fluids that are not recommended:
Use only non-acidic and non-corroding bath fluids.
JULABO assumes no liability for damage caused by the selection of an unsuitable
bath liquid.
2
/s.
23
Preparations
• are highly viscous
• No liability for use of other bath fluids!
protection.
19
(much higher than recommended at the respective working temperature)
• have a low viscosity and have creep characteristics
• have corrosive characteristics or
• tend to crack.
5.3. Filling / draining
Notice:
•Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
•Do not drain the bath fluid while it is hot!
Recommendation: Temperature range 5 °C to 40 °C
Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment, for example).
•Store and dispose the used bath fluid according to the laws for environmental
Filling
Take care that no liquid enters the interior of the
circulator.
The recommended filling level is marked with an
angle plate (M) inside the bath.
When the filling level does not exceed the marker
(M) and providing a suitable bath fluid is used for
the specified working temperature range of the unit,
neither low liquid level alarm should be triggered
nor overflowing of the bath fluid should be caused.
Draining:
• Turn off the unit and disconnect the mains cable
from the power source.
• Hold the venting grid, pull out and remove.
24
• Slide a short piece of tube onto the drain
connection and hold it into a container.
• Loosen the drain screw a few turns and drain the
unit completely.
• Tighten the drain tap.
6. Operating procedures
Caution:
•We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply voltage
the mains socket.
deutsch / english
6.1. Power connection
• Only connect the unit to a power socket with earthing contact (PE – protective
earth)!
• The power supply plug serves as safe disconnecting device from the line and must
be always easily accessible.
• Never operate equipment with damaged mains power cables.
• Regularly check the mains power cables for material defects (e.g. for cracks).
specified on the identification plate.
• Connect the circulator with mains power cable (17a) to the mains
outlet (18).
• Connect the control cable (9) to the sockets (9a, 9b).
• Connect the refrigerated circulator with mains power cable (17b) to
SL
6.2. Switching on / Selecting the language
Switching on:
• Turn on the mains power switch (1a).
The unit performs a self-test.
Then the software version (example: V 1.12-S008C) appears. The
display „OFF“ or „R OFF“ indicates the unit is ready to operate.
The circulator enters the operating mode activated before switching
the circulator off:
keypad control mode (manual operation) or
remote control mode (operation via personal computer).
Selecting the language:
There are two options for the language of the LCD DIALOG-DISPLAY:
German or English. Select the desired language in the menu
>Configuration< under the submenu >Language/Sprache<.
Press the respective keys in the following order:
3x up to menu Configuration
6x up to submenu >Language/Sprache<
25
Operating procedures
The VFD COMFORT-DISPLAY indicates the message "OFF".
6.2.1. Switching on the cooling machine
• Switch on the cooling machine using the switch (1b) .
6.3. Start - Stop
Control of the cooling machine:
With the mains switch (1b) turned on, the circulator automatically
switches the cooling machine off and on.
To save energy, turn off the cooling machine with the mains switch (1b)
whenever cooling is not required.
Start:
• Press
The actual bath temperature is displayed on the VFD COMFORTDISPLAY. The circulating pump starts with a slight delay.
Stop:
• Press key.
or
Keep
key.
key pressed.
6.4. Adjusting the temperature chamber
Adjusting the temperature chamber after set
up:
• If bath fluid overflow in the temperature
chamber is not equal in every direction, an
adjustment can be made with the 3 screws
that are accessible from above.
Required tool: SW3 hexagonal socket wrench
with ball-shaped head
26
7. Setting of temperatures
SL
Factory settings:
Press the
3 different working temperatures can be adjusted. Their values are freely
selectable within the operating temperature range.
key to call up the menu for temperature selection.
The temperatures can be set in start or stop mode.
Setting of working temperature in the
1. Press the key
2. Select SETPOINT 1 or 2 or 3 using the key
3. Confirm by pressing the
The circulator uses the new working temperature value for
temperature control.
Example:
Adjustment/modification of the pre-setting of "SETPOINT 3"
If the active Setpoint is changed, the new value is immediately used
for the control of the working temperature.
The heater control indicator flashes.
. The value flashes
key.
menu
or .
Setpoint3: Change the value from 70.00°C to 80.00°C.
See standard display line 1
• Use the numeric keypad to enter
Then press enter
Notice: Refer to SETPOINT MAX / MIN in chapter „LIMITS“
to store the value.
and .
8. Safety installations, warning functions
Check the safety installations at least twice a year! Refer to page 17.
Settings for the excess temperature protection > SafeTemp< and for the
warning functions > OverTemp< and > SubTemp< are made in a menu
which is called up by pressing the key
.
Menu item > Type< allows choosing between a warning and an
alarm cut-off for the menu items > OverTemp< and > SubTemp<.
27
Safety installations, warning functions
Rough setting can be effected by using the
For setting proceed as follows.
Warning:
Adjust excess-temperature safety device below the flash point of the bath fluid.
In case of wrong setting there is a fire hazard!
We disclaim all liability for damage caused by wrong settings!
SafeTemp:
Setting range: 20 °C ... 320 °C
temperature scale.
SafeTemp:
Type:
OverTemp :
SubTemp:
Settings:
• Press the key
menu item and confirm the by pressing the key .
• Set the new cut-out value within 30 seconds with a screw driver. Exact
display and setting on LCD display
Recommendation:
Set over temperature-protective setting >SafeTemp< 5 °C to 10 °C
above the set point of the working temperature.
Switch-over from warning to shutdown function
• Change the parameter by pressing the key
pressing the key
or
pressing the key
Possible parameters: warning or alarm
• Set value on numeric keypad and confirm by pressing
Example: 83.00 °C
Press the keys
and by pressing the key or select the
and confirm by
.
if the parameter is to be retained.
and and confirm by pressing
Sensor:
28
Indicated is the temperature value of the safety sensor.
SafeTemperature:
Setting range: 20 °C ... 320 °C
Indicated is the cut-out value set with a screwdriver on the excess
temperature protection device.
This safety installation is independent of the control circuit. When the
temperature of the bath fluid has reached the safety temperature, a
complete shutdown of the heater and pump is effected.