AVOXimeter® 4000
Whole Blood CO-Oximeter
Operator’s Manual
Manufacturing Company Location
International Technidyne Corp
8 Olsen Avenue, Edison, New Jersey USA
Phone: 1-732-548-5700 Fax: 1-732-248-1928
web site: www.itcmed.com
Copyright and Trademarks
Copyright© 2007 International Technidyne Corporation (ITC). All rights reserved. This material may not be reproduced or copied, in whole or in part, without the written permission of ITC.
ITC and AVOXimeter are registered trademarks of International Technidyne Corporation in the United States and other jurisdictions.
U.S. Patents. 5,430,542 and 6,262,798. Euro/UK Patent 0663070. Other patents pending.
Technical Support
Contact ITC Technical Support at (800) 631-5945 or (732) 548-5700, or by e-mail at techsupport@itcmed.com.
ii
Table of Contents |
|
|
1 |
INTRODUCTION ....................................................................................................... |
1 |
|
Intended Use of the ITC AVOXimeter 4000............................................................... |
1 |
|
Summary and Explanation of the Test....................................................................... |
1 |
|
Glossary of Abbreviation Equivalents ……………………………………………………2 |
|
|
Operating Precautions and Warnings ........................................................................ |
3 |
|
Limitations.................................................................................................................. |
4 |
2 |
DESCRIPTION .......................................................................................................... |
5 |
|
Front Panel ................................................................................................................ |
5 |
|
Keypad....................................................................................................................... |
6 |
|
Menus ....................................................................................................................... |
7 |
|
Test Cuvettes............................................................................................................. |
8 |
|
Connections ............................................................................................................... |
9 |
|
Automatic Standby and Shutdown............................................................................. |
9 |
|
Instrument Lockouts................................................................................................... |
9 |
|
Instrument Specifications......................................................................................... |
10 |
|
Reportable Range........................................................................................ |
10 |
|
Accuracy ...................................................................................................... |
10 |
|
Precision ...................................................................................................... |
10 |
|
Interference.................................................................................................. |
11 |
|
Calibration................................................................................................................ |
11 |
3 |
GETTING STARTED ............................................................................................... |
13 |
|
Unpacking and Inspection........................................................................................ |
13 |
|
Materials Provided ....................................................................................... |
13 |
|
Materials Required But Not Provided .......................................................... |
13 |
|
Optional Materials........................................................................................ |
14 |
|
Charging the Batteries ............................................................................................. |
14 |
|
Setting Up the Instrument ........................................................................................ |
15 |
|
Setting Display Backlighting ........................................................................ |
15 |
|
Specifying Units for Total Hemoglobin (THb) .............................................. |
15 |
|
Enabling or Disabling Display of [sO2], [O2Ct], and [O2Cap]...................... |
16 |
|
Enabling or Disabling Suppression of Negative Values............................... |
16 |
|
Changing the Date and Time....................................................................... |
17 |
|
Setting the Standby Delay ........................................................................... |
18 |
|
Specifying Entry of User ID and/or Patient ID.......................................................... |
19 |
|
Specifying Mandatory Entry of an Authorized User ID................................. |
19 |
|
Specifying Optional Entry of a User ID Whenever a Test is Run................. |
26 |
|
Specifying Optional Entry of a Patient ID Whenever a Test is Run ............. |
27 |
|
Specifying a Different Value for Hüfner’s Number ................................................... |
28 |
|
Calibration................................................................................................................ |
29 |
|
Cuvette Calibration Code............................................................................. |
29 |
|
Re-Calibration.............................................................................................. |
29 |
iii
iv
4 |
OPERATION.......................................................................................................... |
31 |
|
|
29 |
|
Startup ..................................................................................................................... |
31 |
|
Sample Collection and Preparation ......................................................................... |
32 |
|
Sample Collection........................................................................................ |
32 |
|
Sample Preparation ..................................................................................... |
32 |
|
Running a Test......................................................................................................... |
33 |
|
Running a Test on a Patient Sample ........................................................... |
33 |
|
Using the Printer ...................................................................................................... |
35 |
|
Printing the Current Test Results................................................................. |
35 |
|
Specifying Automatic Printing of Results ..................................................... |
35 |
|
Changing the Serial Port Baud Rate and Parity .......................................... |
35 |
|
Data Management ................................................................................................... |
36 |
|
Reviewing and/or Printing the Last Sample................................................. |
36 |
|
Locating, Reviewing, and/or Printing any Sample ....................................... |
37 |
|
Printing all Stored Data................................................................................ |
37 |
|
Aborting Printing of Results ......................................................................... |
37 |
|
Quality Control ......................................................................................................... |
38 |
|
Performing Optical Quality Control .............................................................. |
38 |
|
Running Liquid Controls .............................................................................. |
40 |
|
Entering Liquid Control Lot Numbers........................................................... |
42 |
|
Entering Cuvette Lot Numbers .................................................................... |
43 |
|
Shutdown ................................................................................................................. |
44 |
|
Troubleshooting ....................................................................................................... |
45 |
5 |
MAINTENANCE ...................................................................................................... |
49 |
|
Verification of Instrument Temperature.................................................................... |
49 |
|
Cleaning the Optical Detector .................................................................................. |
50 |
|
Replacing the Battery............................................................................................... |
55 |
6 |
QUALITY CONTROL LOGS ................................................................................... |
59 |
7 |
WARRANTY ............................................................................................................ |
63 |
|
Certification, Warranty and Service Warranty, and Service ..................................... |
63 |
8 |
SAFETY STANDARDS ........................................................................................... |
65 |
|
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions.................... |
66 |
|
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity ................. |
67 |
INDEX ............................................................................................................ |
69 |
v
Important Labels and Symbols
Before using the ITC AVOXimeter 4000, it is essential that the contents of this Operator’s Manual, any labels on the instrument or its packaging, and instructions accompanying ITC AVOXimeter 4000 cuvettes are read and understood by the operator. These materials make reference to additional symbols that are explained below:
Product Conforms to Directive 98/79, 27 October 1998 on In-Vitro Diagnostic Medical Devices
Serial Number of Instrument
Lot Number of Cuvettes
ITC Catalogue Number of Devices
Do Not Reuse– Single Use Only
Upper and Lower Temperature Limitations (For Storage or Use)
For in vitro Diagnostic Use
Attention - Read Accompanying Documentation or Instructions
Consult Instructions for Use
Input Port for DC Power Cord from AC/DC Power Module - Polarity, VDC and A Input
Serial Output Port for Data Transfer – RS232C
Temperature Probe Input
Name and Address of Manufacturer
Warning - Biohazard
Medical Equipment per Annex 1A, Item 8 Directive 2002/96/EC For Electronic Equipment Waste – Contact ITC Technical Support @ 1-800-631-5945
vi
vii
ITC AVOXimeter 4000 Operator’s Manual
The ITC AVOXimeter 4000 is a battery-operated desktop whole blood oximeter that performs individual point-of-care measurements of total hemoglobin (tHb), oxyhemoglobin saturation (%02Hb), carboxyhemoglobin (%COHb), and methemoglobin (%MetHb) on freshly-drawn or heparinor EDTA-anticoagulated whole blood samples. Oxygen content ([02Ct]), percent saturation (SO2) and oxygen carrying capacity (O2Cap) of the blood sample are automatically calculated from the %Hb02 and THb measurements.
No sample preparation is required, and analysis is quickly accomplished by injecting the sample into a disposable cuvette and inserting the cuvette into the instrument. The ITC AVOXimeter 4000 then illuminates the sample with multiple wavelengths, records the optical density of the sample at each of the wavelengths, and computes the results. In less than 10 seconds, the total hemoglobin concentration and the percentages of oxyhemoglobin, carboxyhemoglobin, and methemoglobin in the sample are shown in appropriate units on the liquid-crystal display on the front panel.
Data management capabilities are included with the instrument. These capabilities include storage of up to 100 patient or QC results, designation of quality control levels and lot numbers, tagging of test results with date, time, Patient ID and/or Operator ID, and printing of results.
The ITC AVOXimeter 4000 measures whole blood tHb, %02Hb, %COHb, and %MetHb using disposable single-use cuvettes. The operator inserts a whole blood sample into a cuvette, the cuvette is inserted into the test chamber on the instrument, and the results are displayed. The results will remain on the display after the cuvette is removed from the instrument until any key is pressed, clearing the screen.
The result can be automatically printed along with the time and date the test was run, the Patient ID, Operator ID, and other information entered. The result is also saved in an internal database, which has the capability to store up to 100 results.
Up to three liquid control lot numbers for each level of Liquid Quality Control (LQC) can be stored in the ITC AVOXimeter 4000 and must be tagged to the stored or printed records. Cuvette lot numbers can also be tagged to the LQC records. The instruments can be configured so that only authorized operators can operate the system and that patient IDs can be entered for each test run.
The ITC AVOXimeter 4000 measures oxygenated hemoglobin (O2Hb), reduced hemoglobin (HHb), methemoglobin (MetHb), and carboxyhemoglobin (COHb) directly, using novel optics and multiple wavelengths. This reduces interference from dyshemoglobins and other interfering substances such as fetal hemoglobin and bilirubin and minimizes the effects of hemolysis.
1
ITC AVOXimeter 4000 Operator’s Manual
The measured values are used to calculate total hemoglobin (tHb) and percent oxyhemoglobin saturation [%02Hb] of the sample, using the fractional method described below:
(tHb) |
= |
(O2Hb) + (HHb) + (MetHb) + (COHb) |
||
%O2Hb |
= |
(O2Hb) x 100 |
|
|
(tHb) |
||||
|
|
Oxygen content [O2Ct] of the sample is then calculated:
O2Ct = |
1.39 x tHb x % O2Hb |
|
100 |
||
|
where 1.39 is the amount of oxygen assumed to be carried by one gram of oxygenated hemoglobin (Hϋfner’s Number). Depending on your facility protocols, the Hϋfner’s Number stored in the ITC AVOXimeter 4000 can be set at any value in the range of 1.30 to 1.39 (see page 28).
In some cases, the display screen of the ITC AVOXimeter 4000 does not show the most commonly used format for a specific constituent’s abbreviation. Below is a glossary of abbreviation equivalents to ensure that users fully understand each reading on the ITC AVOXimeter 4000 display.
Constituent |
AVOXimeter 4000 |
Conventional |
|
Display |
Equivalent |
||
|
|||
Total Hemoglobin |
tHb |
tHb |
|
Fractional O2 Saturation |
sO2 |
SO2 |
|
Oxygen Content |
O2Ct |
O2CT |
|
Carboxyhemoglobin |
COHb |
COHb |
|
Methemoglobin |
MetHb |
MetHb |
|
Oxyhemoglobin Saturation |
%O2Hb |
%O2Hb |
2
ITC AVOXimeter 4000 Operator’s Manual
●For in vitro Diagnostic use.
●Do not allow blood, water, or other liquids to enter the instrument.
●The ITC AVOXimeter 4000 instrument is designed for use only with ITC AVOXimeter 4000 cuvettes.
●Do not re-use test cuvettes.
●Always keep cuvettes in sealed bag with desiccant, and replace desiccant if the indicator dot on the desiccant pack indicates the presence of mositure.
●When filling cuvette, do not use excessive pressure on the syringe or cause the vent patch to bulge outward by overfilling the cuvette.
●For proper calibration and calibration verification, use only the controls recommended in this manual. Controls from other sources may yield erroneous results.
●The ITC AVOXimeter 4000 instrument is designed to be used for testing in a stationary position. DO NOT perform testing while carrying or holding the instrument.
●In order to charge the ITC AVOXimeter 4000 instrument, the AC power cord should be plugged into an electrical service outlet and the AC/DC power module while the DC power cord from the AC/DC power module is plugged into the DC port in the back of the instrument.
●DO NOT expose the ITC AVOXimeter 4000 instrument to extreme temperature (above 35°C, 95°F). Such exposure could affect the performance of any type of electronic instrumentation.
●DO NOT drop the ITC AVOXimeter 4000 instrument, and do not use the results if the instrument is dropped during a test.
●Only properly qualified personnel should attempt to open and perform work on the ITC AVOXimeter 4000 instrument as identified in this manual.
●DO NOT remove the AC/DC power module from the ITC AVOXimeter 4000 instrument by pulling on the cord.
●The use of accessory equipment (e.g., printers, etc.) not identified in this manual either in the patient vicinity, or that does not comply with either the equivalent safety requirements of this equipment or UL/IEC 60601-1 or IEC 60601-1-2, may lead to a reduced level of safety with the resulting system.
Any items exposed to human blood, plasma or serum must be handled cautiously as a biohazardous material in accordance with laboratory safety practices and federal and local regulations. Federal, state and local laws and regulations require that hazardous waste be disposed of in a specific manner. Waste material from the ITC AVOXimeter 4000 which may be classified as biohazardous include used cuvettes. It is important that steps be taken to dispose of these materials in accordance with the prevailing regulations in your location.
3
ITC AVOXimeter 4000 Operator’s Manual
Do not disturb the instrument while a test is in progress.
As with all diagnostic tests, ITC AVOXimeter 4000 test results should be scrutinized in light of a specific patient’s condition and therapy. Any results exhibiting inconsistency with the patient’s clinical status should be repeated or supplemented with additional test data.
4
ITC AVOXimeter 4000 Operator’s Manual
The ITC AVOXimeter 4000 (Figure 1) is a tabletop device for use at the bedside. It contains a test chamber which performs all operations to measure the concentrations of reduced oxyhemoglobin (HHb), oxyhemoglobin saturation (%02Hb), carboxyhemoglobin (COHb), and methemoglobin (MetHb) of a whole blood sample after the operator inserts a test cuvette containing the sample into the test chamber.
The concentration of total hemoglobin (tHb), the relative fractions of carboxyhemoglobin (%COHb) and methemoglobin (%MetHb), and the oxygen content (02) of the blood sample are then automatically calculated and reported.
Each ITC AVOXimeter 4000 is calibrated at the factory. The ITC AVOXimeter 4000 can be operated either from its internal batteries or from the AC adapter. The batteries are charged whenever the AC Adapter is connected.
Printer (Optional) |
Test Chamber |
Temperature |
Probe |
||
Display |
Keypad |
|
Panel |
|
|
Finger Grip
Test Cuvette
Optical QC |
|
Filters |
Test Cuvettes |
Figure 1. ITC AVOXimeter 4000 Oximeter
The front panel (Figure 2) contains the test chamber, a keypad with the key, action and menu keys, number keys, and a display panel. Operator instructions are shown on the display panel, and the operator enters commands and information using the keypad.
When the test is completed, the results are shown on the display panel and stored in system memory for current or future printing.
The display panel is illuminated to enhance visibility in low light conditions. The illumination can be adjusted (or turned off) to conserve power during battery operation.
5
ITC AVOXimeter 4000 Operator’s Manual
Test Chamber Number Keys (Blue)
|
|
|
|
|
|
|
|
Action Keys |
|
|
|
|
|
|
|
|
(White) |
Display |
|
|
|
Enter/On Key |
||||
Panel |
|
|
|
|
|
|
|
(Red) |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
Menu Keys |
|
|
|
|
|
|
|
|
(White) |
Figure 2. Front Panel
The routine analysis of blood samples does not require the use of menus or the numeric keypad. However, these enable the user to take advantage of many useful features.
The purpose of each key is summarized below:
Key
and
to
Purpose
Switch the instrument on. Select a command.
Display a menu of commands for calibration, printing, stored data, and shutdown.
Display a menu of commands for entering hemodynamic variables, entering device settings, entering the time and date, viewing battery status and temperature, and managing data.
Print the results that are displayed.
Respond to questions that are displayed.
Backspace over a numerical entry (such as a QC lot number) so that it can be corrected.
Return to the previous menu.
Enter characters for Operator IDs or Patient IDs. Enter a character for selection of a command.
6
ITC AVOXimeter 4000 Operator’s Manual
The principal menus, their commands, and the procedure to access each menu are summarized below:
Note: Press the button at any time to return to the previous menu.
|
Menu |
|
Commands |
Access |
|
|
|
|
|
|
|
|
|
|
|
|
Main Menu |
|
|
|
Press the |
key when a test is not |
|
|
|
|
|
||||
|
|
|
|
|
running or another menu is not active. |
||
|
Calibration |
|
|
|
Press |
followed by |
while the |
|
|
|
|
||||
|
|
|
|
||||
|
Submenu |
|
|
|
main menu is displayed. |
|
|
|
Printer Mode |
|
|
|
Press |
followed by |
while the |
|
|
|
|
||||
|
|
|
|
||||
|
Submenu |
|
|
|
main menu is displayed. |
|
|
|
Stored Data |
|
|
|
Press |
followed by |
while the |
|
|
|
|
||||
|
|
|
|
||||
|
Submenu |
|
|
|
main menu is displayed. |
|
|
|
|
|
|
|
|
||
|
Computer Menu |
|
|
|
Press the |
key when a test is not |
|
|
|
|
|
||||
|
|
|
|
||||
|
|
|
|
|
running or another menu is not active. |
||
|
Data Management |
|
|
|
Press |
followed by |
while the |
|
|
|
|
||||
|
|
|
|
||||
|
Submenu |
|
|
|
computer menu is displayed. |
||
|
Device Settings |
|
|
|
Press |
followed by |
while the |
|
|
|
|
||||
|
|
|
|
||||
|
Submenu (Page 1) |
|
|
|
computer menu is displayed. |
||
|
Device Settings |
|
|
|
Press |
followed by |
while |
|
|
|
|
||||
|
|
|
|
||||
|
Submenu (Page 2) |
|
|
|
Page 1 of the device settings menu is |
||
|
|
|
|
|
displayed. |
|
|
|
Time, Date, and |
|
|
|
Press |
followed by |
while the |
|
|
|
|
||||
|
Temperature |
|
|
|
computer menu is displayed. |
||
|
Submenu |
|
|
|
|
|
|
7
ITC AVOXimeter 4000 Operator’s Manual
Tests are performed with single-use disposable test cuvettes (Figure 3). Each test cuvette contains a finger grip, filling port, optical window, and a vent patch.
Filling Port
Finger Grip
Optical
Window
Light Path
Vent Patch
Figure 3. Test Cuvette
A whole blood sample is inserted into a test cuvette by connecting a small syringe containing the whole blood sample to the filling port and then gently pressing the syringe plunger to dispense approximately 50 µL of whole blood into the test cuvette. Air escapes from the vent patch at the end of the test cuvette while the whole blood sample is being inserted. The test cuvette (with the syringe still attached) is then inserted into the test chamber of the instrument (see page 32 for details).
Note: Be sure to handle the cuvette either by the edges or by the finger grip. Refer to the package insert accompanying the test cuvettes for storage and handling instructions.
●Remove any blood or debris from the exterior of the test cuvette before inserting it into the test chamber.
●After filling the cuvette with blood, inspect the vent patches to ensure they are not bulging out. If a vent patch protrudes, discard the cuvette. Do not insert a cuvette with a protruding vent patch into the test chamber.
BIOHAZARD WARNING: Any items exposed to human blood, plasma or serum must be handled cautiously as a biohazardous material in accordance with laboratory safety practices and federal and local regulations. Federal, state and local laws and regulations require that hazardous waste be disposed of in a specific manner. Waste material from the ITC AVOXimeter 4000 which may be classified as biohazardous include used cuvettes. It is important that steps be taken to dispose of these materials in accordance with the prevailing regulations in your location.
8
ITC AVOXimeter 4000 Operator’s Manual
BIOHAZARD WARNING: Any items exposed to human blood, plasma or serum must be handled cautiously as a biohazardous material in accordance with laboratory safety practices and federal and local regulations. Federal, state and local laws and regulations require that hazardous waste be disposed of in a specific manner. Waste material from the ITC AVOXimeter 4000 which may be classified as biohazardous include used cuvettes. It is important that steps be taken to dispose of these materials in accordance with the prevailing regulations in your location.
Connections to the power supply and an optional printer (or a computer) are made at the rear of the instrument (Figure 4).
Use only the power supply provided with the instrument.
Serial Number
Label
Carrying
Handle
AC Adapter
Connector
Printer or Computer |
Temperature Sensor |
Connector |
Connector |
Figure 4. Rear Panel Components
The ITC AVOXimeter 4000 enters a low-power standby mode after the instrument has been idle for a specified period of time (the standby delay). The instrument is factory preset for a standby delay time of 60 minutes, but a time of 10 to 180 minutes can be specified (see page 18). To resume normal operation when the instrument is in standby, press and hold down any key for one second.
Note: The ITC AVOXimeter 4000 also enters standby if the battery charge becomes critically low.
The ITC AVOXimeter 4000 shuts down after it has been in standby for 4 hours.
The instrument can be configured to allow use only by authorized operators and/or to allow use only if optical QC has been performed.
9
ITC AVOXimeter 4000 Operator’s Manual
Size
Weight
Operating Temperature
Battery Type
Operating Time On Battery
Anticipated Battery Life
Power Supply/Chargers
Serial Data Port
Sample Type
Sample Volume
Analysis Time
Analysis Wavelengths
20.3 cm (8.0 in) x 25.4 cm (10.0 in) x 9.5 cm (3.8 in) 1.8 kg (4 lbs)
Room temperature (15°C to 30°C, 59°F to 86°F) Nickel Cadmium (NiCad)
Approximately 8 hours (constant run) or 10 complete test cycles per charge. Tests may also be run while the ITC AVOXimeter 4000 is plugged into the AC/DC power module.
Approximately 500 charge / discharge cycles
Input: 100 / 240 VAC, 50 / 60 Hz Output: 12 VDC, 800 mA
RS232C Whole blood 50 µL
7 to 10 seconds per sample
5
%O2Hb |
0 to 100% |
%COHb |
0 to 75% |
%MetHb |
0 to 85% |
[O2] |
0 to 35 mL O2/dL |
tHb (>10 g/dL) |
±0.45 g/dL |
tHb (<10 g/dL) |
±0.35 g/dL |
%O2Hb |
±1.6% |
%COHb |
±2% |
%MetHb |
±1.5% |
tHb |
0.3 g/dL |
%O2Hb |
0.8 % O2Hb |
%COHb |
1 % COHb |
%MetHb |
0.7 % MetHb |
Reportable Range |
4 to 25 g/dL |
tHb |
|
Accuracy |
|
Precision |
|
10
ITC AVOXimeter 4000 Operator’s Manual
|
|
|
tHb |
|
%O2Hb |
|
|
%COHb |
|
%MetHb |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Bilirubin |
|
None |
|
None |
< 1% |
|
< 1% |
|
||
|
(11 mg/dL) |
|
|
|
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Hemolysis |
|
None |
|
None |
< 1% |
|
< 1% |
|
||
|
Fetal Hemoglobin |
|
< 0.45 g/dL |
< 1% |
|
|
< 0.6% per |
< 1% |
|
||
|
(tHb = 13.5 g/dL, HbF = 100% |
|
|
|
%HbF |
|
|||||
|
|
|
|
|
|
|
|
|
|||
|
Indocyanine Green Dye |
|
< 0.45 g/dL |
< 1% |
|
< 1% |
|
< 1% |
|
||
|
(<10 mg/L) |
|
|
|
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Methemoglobin |
|
< 0.2 g/dL |
< 1% |
|
|
|
|
|
|
|
|
(tHb = 16 g/dL, MetHb <10%, |
|
|
|
|
|
|
|
|||
|
7.1 < pH < 7.8) |
|
|
|
|
|
|
|
|
|
|
The ITC AVOXimeter 4000 is factory-calibrated and employs highly stable state-of-the-art light sources. Should recalibration be required please contact an ITC technical support representative.
Proper calibration also requires entry of the correct cuvette pathlength by the user (see page 26) and use of a customary value for Hüfner’s number (see page 26).
CAUTION: If quality control results are not acceptable, erroneous results are encountered, or error messages are displayed, the most likely cause may be contamination of the optical detector by blood or debris, which cannot be resolved by re-calibration. Consult the Troubleshooting section for additional information.
11
ITC AVOXimeter 4000 Operator’s Manual
12
ITC AVOXimeter 4000 Operator’s Manual
Note: Inspect each component for damage when unpacking. If damage is observed, contact your shipping representative immediately.
1.Remove any protective packaging that may be present around the instrument.
2.Examine the packaging material to be sure that the AC Adapter, temperature probe, connecting cables, or other components have been removed. The materials that are provided are listed below.
Note: Do not discard the packaging material. It should be preserved for future use, in the event that it is necessary to re-pack the instrument for shipping or transport.
Item |
Quantity |
|
|
ITC AVOXimeter 4000 Instrument |
1 |
Power Supply |
1 |
Temperature Probe |
1 |
Operator’s Manual |
1 |
Quality Control Filters |
2 |
Note: An AC power cord is supplied only for the 110VAC version of the US/Canada/Japan instrument. For all others, the customer must obtain a 3 conductor AC power cord that is compatible with an IEC 320 connection at the power supply AC inlet and any other local requirements.
Item |
Quantity |
|
|
ITC AVOXimeter 4000 Cuvettes |
As Needed |
Liquid Controls (Manufactured by IL or RNA) |
As Needed |
See page 39 for additional information |
|
13
ITC AVOXimeter 4000 Operator’s Manual
Item |
Quantity |
|
|
Dymo Printer |
1 |
• 110 V (Part No. SE300-110) |
|
• 220 V (Part No. SE300-220) |
|
• UK (Part No. SE300-UK) |
|
Printer Paper (Part No. 30270) |
As Needed |
Charge the batteries before the system is used for the first time.
1.Plug the AC Adapter into an electrical service outlet.
2.Connect the AC Adapter cord to the power connector on the rear of the instrument.
3.Allow the battery to charge for at least eight hours.
Note: The AC Adapter can remain connected all the time.
4.To ensure adequate charge, leave the instrument connected to the AC Adapter for a minimum of eight hours. This eliminates the risk of the instrument powering down during a test.
Fully charged batteries will allow the ITC AVOXimeter 4000 to analyze blood samples continuously for up to 8 hours when the display is set at medium brightness.
Battery power can be conserved by:
●Reducing (or turning off) display backlighting (see page 15).
●Reducing the standby delay (see page 18).
Note: The batteries can suffer from a “memory effect” if they are charged before being completely discharged. For optimal battery performance, discharge completely when possible before charging them. The message “Battery Critical – Connect Charger” will be displayed when the battery is completely discharged.
The message “Battery Critical – Connect Charger” is displayed and the instrument reverts to the standby mode if the battery power is insufficient to complete the test. The AC Adapter must be used for additional tests until the battery is recharged.
Checking the Battery:
1.Display the “Time, Date, and Temperature” menu (a submenu of , see page 7).
2.Press followed by to display the battery status:
14
ITC AVOXimeter 4000 Operator’s Manual
3.Press to display the time, date, and battery menu again.
4.Press to return to the previous menu, if desired.
The user can specify the display brightness, specify the units that are used for reporting total hemoglobin (THb), change the date and time, and specify the length of time that the instrument is idle before it enters the standby mode.
Lighting of the display can be reduced to conserve battery power or increased to improve visibility.
1.Display the “Device Settings” menu (a submenu of , see page 7).
2.Press followed by to display the screen for changing the LCD backlighting:
3.Select the backlighting level (or turn it off) by pressing the corresponding number key followed by . The backlighting changes accordingly.
4.Press to display the “Device Settings” menu again.
Measured values of total hemoglobin can be expressed in units of mmol/L or gm/dL.
1.Display the “Device Settings” menu (a submenu of , see page 7).
2.Press followed by to display the screen for changing the units for total hemoglobin:
3.Press followed by to change the units, or press followed by when the desired units are displayed. The “Device Settings” menu is again displayed.
4.Press to return to the previous menu, if desired.
15
ITC AVOXimeter 4000 Operator’s Manual
Display of calculated oxygen content (02Ct), percent saturation (SO2) and oxygen carrying capacity (O2Cap) of the blood sample can be enabled or disabled.
Oxygen content (02Ct), percent saturation (sO2), and oxygen carrying capacity Note: (O2Cap) are automatically calculated for each test.
When display of these values is enabled, these values are displayed on the second page (of three pages) when results are reviewed in the ‘stored data’ menu (see page 36), and they are included in printed test results.
Only measured values of tHb, %HbO2, %COHb, and %MetHb are displayed on the instrument display panel after a test is run, regardless of whether display of calculated oxygen content (02Ct), percent saturation (SO2), and oxygen carrying capacity (O2Cap) of the blood sample is enabled.
1.Display the second page of the “Device Settings” menu (a submenu of ).
2.Press followed by to display the screen for enabling or disabling display of calculated (02Ct), (SO2), and (O2Cap) for each test:
3.Select the appropriate option by pressing the corresponding number key followed by . A corresponding confirmation prompt is displayed.
4.Press to return to the previous menu, if desired.
Users can choose to display or suppress negative values.
Negative values may occur because in some cases, due to instrument precision Note: ranges, a very low Metor Carboxyhemoglobin value may read as a negative
value. Users can choose to suppress negative values, and have them read as ‘0’.
1.Display the second page of the “Device Settings” menu (a submenu of , see page 7).
2.Press followed by to display the screen for enabling or disabling display of negative values:
3.Select the appropriate option by pressing the corresponding number key followed
by . A corresponding confirmation prompt is displayed. Press to return to the previous menu, if desired.
16