Isotech TTI-10 Specifications

Handheld Thermometer
-200 to 850°C
g
High Accuracy Handheld Thermometer
g
g
Perfect Standard for use with Isocal-6, Fast-Cal & Dry Blocks
The TTI-10 is a high accuracy handheld temperature indicator with two platinum resistance thermometer inputs. The high precision makes the instrument particularly suitable as a portable reference thermometer to use alongside Isotech temperature calibrators such as the Fast-Cal, Isocal-6 and Dry Block ranges. It is also suited for high accuracy measurements in industrial and scientific applications.
TTI-10 brings laboratory level performance of up to 10mK (0.01°C) and resolution up to 0.001°C in a portable handheld instrument. Battery life is typically 20 hours from a 9V PP3 battery and a protective rubber boot offers protection in field use.
TTI-10
The instrument can capture the minimum, maximum and average values over up to 4000 measurements with a logging rate selectable in the range of 1 second to 30 minutes.
The TTI-10 has an easy to use “learning calibration mode” that allows the TTI-10 to be system calibrated with a Platinum Resistance Thermometer simply by comparing it to a calibrated standard thermometer, no need to calculate coefficients or data, simply enter the reference probe temperature or temperatures and the TTI-10 does the work for you.
The USB interface allows connection to Isotech Cal Notepad software with its charting and logging features.
TTI-10 supports Isotech Semi Standard Platinum Resistance probes with system uncertainties (probe and instrument) as low as 20mK. We recommend the 935-14-61 and 935-14-16 probes detailed below and have special calibration deals available. Other probes and ranges are available, refer to Semi Standards – Platinum Resistance Thermometers in catalogue.
52
CALIBRATION SOLUTIONS
|
Input Connectors
Highest quality latching
metal ‘Lemo’ connectors.
Rubber Sleeve
The TTI-10 Handheld
Thermometer is supplied
with a protective
rubber boot.
Specifications
Sensing Element
Input Channels Two: 100 Ohm PRT, EN 60751
(Pt100), Four Wire
Range -200°C to +850°C
Units °C, °F and Ohms
Resolution 0.001°C from -199.999°C to
+199.999°C remaining range 0.01°C
Accuracy: ±0.012°C from -80°C to 199.999°C Instrument ±0.02°C ±0.0015% RDG from 200°C Only to 660°C
Logging Record Average, Min and Max over
4000 measurements
Measuring Adjustable: 1 second to 30 minutes interval
PC Interface USB - Cable Included
Options
Semi Standard Isotech Semi Standard Platinum PRT Resistance Thermometer: 935-14-112-TTI Fast Response, 2m Cable Length,
four wire with Lemo plug fitted
Connectors High Quality Latching Metal: Lemo:
Working 0°C to +40°C temperature
Display 2-line LCD
Display Single Channel or Dual Channels Simultaneously
Housing Plastic (ABS) supplied with protective
rubber boot
Weight 300g
Power Supply 9V battery PP3 (or via USB Cable)
Battery Life Typically 20 Hours
Dimensions 200 x 85 x 40 mm (LxWxH)
Semi Standard Isotech Semi Standard Platinum PRT Resistance Thermometer: General 935-14-116-TTI Purpose,2m Cable Length, four wire
with Lemo plug fitted
UKAS System Recommended: -50°C to 199.999°C Calibration Four Point System Calibration, TTI-10-14-112-SYST Uncertainty across range 0.025°C
(25mK)
Semi Standard Isotech Semi Standard Platinum PRT Resistance Thermometer: 935-14-61-TTI Fast Response, 2m Cable Length,
four wire with Lemo plug fitted
UKAS System Recommended: -50°C to 199.999°C Calibration Four Point System Calibration, TTI-10-14-61-SYST Uncertainty across range 0.02°C
(20mK)
n Recommended Probes (Fit TTI-10 Carry Case)
UKAS System Recommended: 0°C to 420°C Calibration Four Point System Calibration, TTI-10-14-116-SYST Uncertainty across range 0.04°C
(40mK)
Carrying Case 931-22-101
Model Maximum Diameter Length Sensing Handle Cable Application Range (A) Length (B) (C)
935-14-112/TTI -50°C to 250°C 3mm 225mm 6mm No Handle 2m PTFE Fast Response, Low Stem Conduction
935-14-61/TTI -50°C to 250°C 4mm 300mm 6mm 19 x 120mm 2m PTFE Fast Response, Low Stem Conduction
935-14-116/TTI -100°C to 450°C 6mm 350mm 25mm 19 x 120mm 2m PTFE General Purpose