QuickCal Range
Hand-Held Temperature Test Units
+350°C
to
+30°C
Isotech’s QuickCal range of testers are ideally suited to the less exacting applications where our larger, better specified
baths are unnecessary.
The High-Temperature QuickCal covers the temperature range from 30°C to 350°C.
Both have interchangeable inserts, so you don’t have to buy another calibrator each time you find a new diameter sensor
to calibrate, like our competitors models. The Low-Temperature QuickCal is particularly exciting since it can work from a
12-volt battery, making it truly portable.
The High-Temperature QuickCal has a single larger well and interchangeable inserts, 25mm in diameter by 120mm deep.
Blank and other special drillings are available to order.
The handbook which accompanies the QuickCal contains helpful advice on how to get the best results from your
calibrator. Accessories for the above include a carry case, alternative inserts and a rechargeable battery for the LowTemperature QuickCal.
SPECIFICATIONS
Model QuickCal-550
Temperature Range +30°C to +350°C in an ambient of 20°C
Set Point Resolution 0.1°C over range
Accuracy ±0.2 to ±0.4°C using Comparison
Techniques
<100°C ±0.4°C against display temperature
with a single probe in the 6.8mm diameter
sleeve
>100°C ±1.5°C against display temperature
with a single probe in the 6.8mm diameter
sleeve
INDUSTRIAL
Stability ±0.1°C
Time to Set Point 9 minutes from Ambient to 350°C
Additional Time Typically 5 minutes
for Best Stability
Calibration Volume 25mm diameter x 120mm deep
Supplied with insert, 5 pockets, 8mm,
6.5mm, 6.5mm, 4.5mm, 4.5mm, all
115mm deep
Power 110 VAC ±10% (50 / 60 Hz)
Overall Dimensions 65mm x 152mm x 175mm (HxWxD)
Weight 3.3 lbs (1.5kg)
Key Features
• Interchangeable Inserts
to suit different diameter sensors
• Wide Temperature Range
15°C to 350°C in two ranges
• Faster
9 minutes to temperature
• Compact
65mm x 142mm x 175mm not
including power entry
• Lightweight
weighs 3.3 lbs (1.5kg)
120mm
25mm
ACCESSORIES
Extra Inserts 550-06-01A: Standard Insert
Carrying Case 931-22-66
Fahrenheit Operation Unit can be supplied for °F Operation
Semi Standard 935-14-98
Platinum
Thermometer
HOW TO ORDER
QuickCal-550
Please specify Accessories required
84/industrial calibration
550-06-01B: Blank Insert
550, High-Temp Model