Isotech PRT Probe Data sheet

Semi-Standard PRTs
Platinum Resistance Thermometers
670°C
to
-50°C
In most industrial laboratories the costly and mechanically fragile SPRT is unnecessary. For these applications the uncertainties may be larger, and hence the Isotech range of semi-standards may be a more practical choice. Usually made in stainless steel sheaths 6mm in diameter, with an Ro value of 100Ω and an (R100-R0)/R0 of 0.385. A number of configurations are available since particular characteristics might be required such as short sensing length, fast response, or applicability to a particular temperature range.
In this collection of semi-standards Isotech has examined not only the uses that customers put them through but also the equipment that is used to calibrate them. Most units are now a little longer and have a slightly wider temperature range than our previous products. The stability of the semi-standard will depend upon its range of usage; your semi­standard should remain within its uncertainties between successive annual calibrations, unless you drop it or otherwise mechanically shock it. Our semi-standards are now fitted with a handle and strain relief, except where otherwise stated. All are made with our own high quality 100Ω detectors and completely assembled and calibrated in-house in our UKAS accredited and supervised Laboratory. Calibration is available with fixed points for the best uncertainties and to give the highest confidence possible.
Please note that stem conduction accounts for the biggest sources of errors during calibration. Remember that for a given immersion depth stem conduction is dependent on the temperature difference between the sensing element and ambient temperature, and on the total conductance of the thermometer assembly. It is sometimes thought (mistakenly) that a shorter thermometer will be subject to less stem conduction.
Performance
Isotech has generated a long history for many of our semi-standards. Here are some documented facts:
The 935-14-95 model has the widest temperature range and consequently is likely to suffer the largest changes in characteristics.
Guy Snelling sent this email (right) about the 935-14-95.
INDUSTRIAL
Advantages
• Ideal for All Industrial Applications
• High Stability
• Low Drift
• Selection of Temperature Ranges
• Choice of Lengths and Diameters
92/industrial calibration
industrial calibration/93
Semi-Standards PRTs, continued
Cable
SPECIFICATIONS
Model Number
Diameter (A)
Length (B)
Sensing Length
Handle (C)
Cable (D)
Temperature Range
Application
Features
Annealing Drifts
Handle Length
935-14-61
4mm
300mm
6mm
Yes
2m PTFE 4 wire
-50 to +250°C
Fast Response
Small Stem Conduction
935-14-13
6mm
350mm
25mm
Yes
2m PTFE 4 wire
-196 to +250°C
Low-Temp. Use
Copper Earth Screen
935-14-16
6mm
450mm
25mm
Yes
2m PTFE 4 wire
-100 to +450°C
General Use
316 stainless steel sheath
935-14-72
6mm
375mm
25mm
No
2m PTFE 4 wire
-50 to +670°C
Gemini Jupiter
Fits lid of Carry-case
935-14-82
4mm
210mm
6mm
No (Angled)
1.5m PTFE 4 wire
-50 to +250°C
Europa Venus Calisto
General Purpose
935-14-85
6mm
420mm
25mm
No (Angled)
540mm PTFE 4 wire
-50 to +250°C
Oceanus-6
General Purpose
Diameter
935-14-95
6mm
450mm
25mm
Yes
2mPTFE 4 wire
-100 to +670°C
General Use Medusa
Metal Alloy Sheath
INDUSTRIAL
The most severe test of a thermometer is to cycle it from its minimum temperature of use to its maximum temperature.
Below is tabulated the annealing drifts of a typical 935-14-95.
Annealing Range — 0°C to +670°C
with 4 hour dwell at 670°C
Cycle Number Shift (mK)
1 -0.8
2 -2.2
3 +0.5
4 -0.5
5 -2.3
6 -1.2
7 -0.2
8 +0.6
Total Shift over 8 Cycles: -6.1 mK
Mean Shift per Cycle: -0.76 mK
SPECIFICATIONS — GENERAL
Ro 100Ω ±0.05Ω
Alpha 0.003850 ±0.000005
Stability 0.010 ohm/year
Calibration A NIST Calibration Certificate can be
provided at extra cost
Recommended 1mA Current
Self-Heating Error 0.004°C at 1mA
ACCESSORIES
Termination Accessories BW: Bare Wire
TTI: Lemo Connector: suits TTI-5,
TTI-6, and TTI-7 DB: Connector for Dry Block Calibrator SITE Indicator, or TTI-8
HOW TO ORDER
Please Specify Model Type and Termination Option
(for example 935-14-13/BW)
industrial calibration/93
Loading...