Stirred Liquid Baths
-40 to 300°C
g
Can accept large Water Triple Point Cells
g
Variable Cooling Power
g
Low Filling Cost
The Isotech Hydra range of calibration baths have proved
very successful since their introduction in 2003. The Orion
range has been introduced to fit alongside the Hydra models.
The Orion calibration baths offer deeper calibration
volumes than the Isotech Hydra series with world class
temperature stability. There is a high temperature model
(30°C to 300°C) and two models with built in refrigeration
(-40°C to 125°C) and (-80°C to 125°C).
These deep stirred liquid baths are suitable for the low
uncertainty calibration of long thermometers including
liquid in glass and extended length SPRTS.
Orion 796
The Orion unit has sufficient depth, 470mm (18.5”), to
accommodate Isotech’s largest Water Triple Point Cells
providing a cost effective sub mK calibration system.
There are accessories to hold Mercury, Water, Gallium
and Indium ITS-90 Fixed Point Cells.
Isotech baths have always been designed for thermometer
calibration, setting the standard for immersion depth versus
filling costs. As the Orion requires just 8.5L of fluid it offers
significant cost savings, both initial and the on going
running costs, when compared to other larger volume baths.
In operation the liquid level is within 10mm of the top lid
allowing calibration of liquid in glass thermometers
without the need for complicated accessories.
In addition to using a monocular to aid reading the scale
I-Cal Easy automatic calibration software can capture
images with low cost digital webcams.
The Isotech Orion range provides many features that
laboratories need but which other manufacturers often
cannot supply.
g
Deep immersion depth (470mm /
18.5”) with low filling costs
g
Variable Speed Stirring - allowing
optimum performance for a wide
range of liquid viscosities
g
Modular design for easier
maintenance
g
Over and Under Temperature
Protection - protect expensive
Water Triple Point Cells as well as
over temperature conditions
g
Connections for easy drain and
overflow
g
Automatic Power Up - no manual
intervention needed - easier
automation - energy saving
g
Liquid Level to suit Liquid in Glass
Thermometers
- no level adaptors required
SOLUTIONS FOR PRIMARY & SECONDARY LABORATORIES
65
|
66
|
SOLUTIONS FOR PRIMARY & SECONDARY LABORATORIES
Calibration
Area
90mm
150mm
The diagram (left) shows a section
view of the Orion showing its ability
to house even the largest of cells
The cell shown is a Jarrett- Isotech B-11,
450mm total length and 65mm diameter.
Model 796L 796M 796H
Temp Range -80°C - 125°C -40°C - 125°C 30°C - 300°C
Volume 150mm Diameter, 470mm Deep (8.5 litres)
Absolute Stability ±0.006°C ±0.006°C ±0.006°C
(Methanol, -80°C) (Methanol, -40°C) (C10 Oil, 125°C)
Vertical Uniformity ±0.005°C 0.005°C ±0.007°C
(Methanol, 0°C) (Methanol, 0°C) (C10 Oil, 125°C)
Heating time 25°C to 125°C < 50 mins (C10 Oil) 30°C to 300°C < 120 mins (VH Oil)
Cooling time 0°C to -80°C < 240 mins 10°C to -40°C < 180 mins (Methanol) (Methanol)
Communications Includes Serial Interface, PC Cable and Software
Dimensions 485mm wide, 525mm deep, 1150mm high
Weight 78kg 68kg 48kg
Safety Compliant to CE Regulations
Power 2.8kW 1kW 800W
108-130 or 208-240V, 50/60Hz
How to Order 796L 796M 796H
Refer to Evaluation Report for Full Details
Accessories:
The Orion can be supplied with both copper and
aluminium equalising blocks. Aluminium is recommended
for use only with oils. Copper, with it’s better corrosion
resistance, is recommended for oils, alcohol and water.
Refer to ‘Liquid Selection Guide’ on page 73 for
information on liquids and important Health and Safety
Information. The standard probe holder can support
RTD’s, Thermocouples and Liquid in Glass
Thermometers. A monocular is available to aid reading
the scale of a L.I.G. Thermometer.
The liquid level is typically within only 10mm of the lid, so
complicated fluid level adaptors are not essential.
There are a range of supports to hold various fixed point
cells, Slim and Optimal, allowing the bath to realise ITS-90
fixed point cells to sub mK uncertainties.
Isotech has a full range of fixed point cells, for more
details visit our web-site: www.isotech.co.uk