Isotech Oceanus 6 Specifications

45
CALIBRATION SOLUTIONS
|
-25 to 140°C
50 x 300mm Calibration Volume
Can also be used as a Dry Block, Liquid Bath...
-25°C to 140°C
The Oceanus-6 models but with a substantially larger calibration volume, 50mm diameter by 300mm deep. The Oceanus-6 can be used as a Dry Block, a Liquid Bath, a Blackbody Source for infrared thermometers, a Surface Sensor Calibrator and for performance to a few mK (0.001°C) ITS-90 Fixed Points. The Oceanus-6 (B) and the SITE (S). The B model includes a sophisticated temperature controller with a dual display for Set Temperature and Dry Block Temperature.
PLUS
has all the advantages of the Isocal-6
PLUS
is available in two models, the BASIC
ISOCAL - 6 Range
Oceanus-6
The S model includes a built in digital thermometer to which an external standard thermometer can be connected, for Dry Block use this will give greater accuracy eliminating temperature gradient and loading errors.
For Liquid Bath, Surface Sensor or Blackbody use an external thermometer should always be used - either with the S model or the B model and a separate stand-alone
PLUS
indicator. For Lab use the Oceanus-6 a laboratory performance temperature indicator such as one of the Isotech True Temperature Indicators with performance, for similar sensors, down to hundredths of a
PLUS
degree. The Oceanus-6 accuracies of ±0.0002°C (2 Sigma) at the Water triple point and the Gallium melt temperature of 29.7646°C and up to ±0.005°C in the stirred liquid bath option (by comparison).
offers unprecedented
can be used with
46
|
CALIBRATION SOLUTIONS
Specification
0.12°C
0.10°C
0.08°C
0.06°C
0.04°C
0.02°C
-20°C0°C 20°C 40°C 60°C 80°C 100°C 120°C
0°C
580 Oceanus 6
Audit Calibration (Similar Sensors) S model with UKAS option
Radial Homogeneity (difference between holes)
0.06
0.001
0.026
0.072
0.1
Model Oceanus-6
Temperature Range 45°C below ambient to +140°C
Absolute stability Dry Block Bath ±0.03°C over 30 minutes Stirred Liquid Bath ±0.025°C Ice/Water Bath ±0.001°C Blackbody Source ±0.3°C Surface Sensor Calibrator ±0.5°C ITS-90 Fixed Point ±0.0002°C
Computer Interface Included with Software
Cools from 20°C to -10°C in 90 minutes
Heats from -10°C to 80°C in 60 minutes
Calibration volume 50mm diameter by 300mm deep
Standard Insert 6 pockets, all 8.0mm diameter and
250mm deep
Display Resolution 0.01 -19.99 to 99.99
0.1 100.0 to 140.0 PC can display 0.01 across whole range with the software included
Indicator units °C, °F, K
Power 108 to 130V (50 / 60 Hz) or 208 to
240V (50 / 60 Hz) 300 Watts
Dimensions Height 430mm
Width 310mm Depth 300mm
Weight 17kg
Isocal-6 Performance and Use
Calibration and Uncertainty
A certificate, traceable to National Standards, is included as standard. Recommended is an optional UKAS five-point calibration.
The accuracy will depend very much on the mode of use and the types of sensor to be used. Please contact Isotech for tutorials and uncertainty calculations and comprehensive evaluation reports. The 580 Oceanus-
PLUS
6
meet the Calibration Capacity requirements of EURAMET/cg-13/v.01, “Guidelines on the Calibration of Temperature Block Calibrators”.
Features (Basic & Site) Oceanus 6+
Dry Block ü
Stirred Liquid bath Option ü
Stirred Ice Bath Operation ü
Surface Sensor Option ü
Infrared Calibration Option ü
ITS-90 Fixed Point Cells Water, Gallium
Additional 8mm Pre-heat Pocket ü
Configurable Units: °C, °F and K ü
Supply Voltage Power Correction ü
Additional Features (Site) Oceanus 6+
Independant Temperature indicator ü
Universal Input Types PT100 ü
Thermocouples Types K,N,R,S,L,PL2,T,J,E ü
Linear Process Inputs Including 4-20 mA ü
Stand Alone Thermostat Testing ü
Thermostat Testing With PC ü
Five Point Digital Probe Matching ü
Configurable Units: °C, °F and K ü
47
CALIBRATION SOLUTIONS
|
Isocal-6 Ultimate Flexibility - Oceanus
5
Calibrate all sensor types - Thermocouples, PRT’s, Thermistors, Thermostats, Infrared, Surface Sensors...
Metal Block Insert 580-06-03 Standard Insert included
Alternative Inserts
580-06-04 Blank Insert without
pockets for local machining. Includes M4 tapped hole for supplied extractor tool.
580-06-05 Custom insert. Isotech
can provide custom drilled pockets, minimum of 3mm separation between holes. Contact with your requirements.
Stirred Liquid Mode with Liquid Container Kit 580-06-07
Allows liquid bath use, includes container, magnetic stirrer, probe guide and sealing cap.
Stirred Ice Bath Mode with Liquid Container Kit
Uses same liquid kit to provide 0°C reference as a stirred ice bath
Surface Sensor Calibration with
Surface Sensor Kit
580-06-08 Includes angled
platinum resistance thermometer.
ITS-90 Fixed Point Cells
B12/46/210 Water Triple Point Cell
17401 Gallium Cell
UKAS Calibration
UKAS Calibration available to order, legally traceable in more than 70 countries.
Additional Accessories for use with the above kit
Thermometer Support Kit 580-06-00 Allows three thermo-
meters to be suspended in the bath, including liquid in glass types
C10 Oil 936-06-07 1 Litre
(-35°C to +140°C)
Infrared Calibration Mode with
Blackbody Target 580-02-12 Use optional Probe
935-14-85 placed in the
auxiliary block pocket for use as a reference.
Standard Probe
935-14-85/DB Platinum Resistance
Thermometer for use up to 250°C.
4.5mm diameter Angled head feature avoids sensors in block.
Carrying Case
931-22-58 Sturdy case
accommodates the unit with room for accessories
How To Order
Specify Model, Basic or Site, Supply Voltage, Accessories and if UKAS Calibration is required.
Loading...