45
CALIBRATION SOLUTIONS
|
-25 to 140°C
50 x 300mm Calibration Volume
Use with Water and Gallium Fixed Points
Can also be used as a Dry Block, Liquid Bath...
-25°C to 140°C
The Oceanus-6
models but with a substantially larger calibration volume,
50mm diameter by 300mm deep. The Oceanus-6 can be
used as a Dry Block, a Liquid Bath, a Blackbody Source for
infrared thermometers, a Surface Sensor Calibrator and for
performance to a few mK (0.001°C) ITS-90 Fixed Points.
The Oceanus-6
(B) and the SITE (S). The B model includes a sophisticated
temperature controller with a dual display for Set
Temperature and Dry Block Temperature.
PLUS
has all the advantages of the Isocal-6
PLUS
is available in two models, the BASIC
ISOCAL - 6 Range
Oceanus-6
The S model includes a built in digital thermometer to
which an external standard thermometer can be connected,
for Dry Block use this will give greater accuracy eliminating
temperature gradient and loading errors.
For Liquid Bath, Surface Sensor or Blackbody use an
external thermometer should always be used - either with
the S model or the B model and a separate stand-alone
PLUS
indicator. For Lab use the Oceanus-6
a laboratory performance temperature indicator such as
one of the Isotech True Temperature Indicators with
performance, for similar sensors, down to hundredths of a
PLUS
degree. The Oceanus-6
accuracies of ±0.0002°C (2 Sigma) at the Water triple point
and the Gallium melt temperature of 29.7646°C and up to
±0.005°C in the stirred liquid bath option (by comparison).
offers unprecedented
can be used with

46
|
CALIBRATION SOLUTIONS
Specification
0.12°C
0.10°C
0.08°C
0.06°C
0.04°C
0.02°C
-20°C0°C 20°C 40°C 60°C 80°C 100°C 120°C
0°C
580 Oceanus 6
Audit Calibration (Similar Sensors)
S model with UKAS option
Radial Homogeneity
(difference between holes)
0.06
0.001
0.026
0.072
0.1
Model Oceanus-6
Temperature Range 45°C below ambient to +140°C
Absolute stability Dry Block Bath ±0.03°C
over 30 minutes Stirred Liquid Bath ±0.025°C
Ice/Water Bath ±0.001°C
Blackbody Source ±0.3°C
Surface Sensor Calibrator ±0.5°C
ITS-90 Fixed Point ±0.0002°C
Computer Interface Included with Software
Cools from 20°C to -10°C in 90 minutes
Heats from -10°C to 80°C in 60 minutes
Calibration volume 50mm diameter by 300mm deep
Standard Insert 6 pockets, all 8.0mm diameter and
250mm deep
Display Resolution 0.01 -19.99 to 99.99
0.1 100.0 to 140.0
PC can display 0.01 across
whole range with the
software included
Indicator units °C, °F, K
Power 108 to 130V (50 / 60 Hz) or 208 to
240V (50 / 60 Hz)
300 Watts
Dimensions Height 430mm
Width 310mm
Depth 300mm
Weight 17kg
Isocal-6 Performance and Use
Calibration and Uncertainty
A certificate, traceable to National Standards, is included
as standard. Recommended is an optional UKAS
five-point calibration.
The accuracy will depend very much on the mode of use
and the types of sensor to be used. Please contact
Isotech for tutorials and uncertainty calculations and
comprehensive evaluation reports. The 580 Oceanus-
PLUS
6
meet the Calibration Capacity requirements of
EURAMET/cg-13/v.01, “Guidelines on the Calibration of
Temperature Block Calibrators”.
Features (Basic & Site) Oceanus 6+
Dry Block ü
Stirred Liquid bath Option ü
Stirred Ice Bath Operation ü
Surface Sensor Option ü
Infrared Calibration Option ü
ITS-90 Fixed Point Cells Water, Gallium
Additional 8mm Pre-heat Pocket ü
Configurable Units: °C, °F and K ü
Supply Voltage Power Correction ü
Additional Features (Site) Oceanus 6+
Independant Temperature indicator ü
Universal Input Types PT100 ü
Thermocouples Types K,N,R,S,L,PL2,T,J,E ü
Linear Process Inputs Including 4-20 mA ü
Stand Alone Thermostat Testing ü
Thermostat Testing With PC ü
Five Point Digital Probe Matching ü
Configurable Units: °C, °F and K ü