Isotech microK-400, micro-800 User Manual

User Manual
Applies to:
Software version 1.4.16+
Firmware version 3.17+
Isothermal Technology Limited
Southport
Merseyside
PR9 9AG
T: +44 (0)1704 543830
F: +44 (0)1701 544799
E: info@microk.co.uk
W: www.microk.co.uk
CONTENTS
1 INTRODUCTION ............................................................................... 6
1.1 Unpacking ............................................................................................................................. 7
1.2 Safety ..................................................................................................................................... 7
1.3 Powering Up Your microK ................................................................................................. 9
1.4 A Quick Tour of Your microK ........................................................................................... 9
2 MAKING A MEASUREMENT (TUTORIALS) .................................. 12
2.1 Measuring Temperature with a Calibrated SPRT .......................................................... 12
2.2 Measuring Temperature with a Thermocouple .............................................................. 16
3 DRIVING YOUR MICROK ............................................................... 20
3.1 Introduction ........................................................................................................................ 20
3.2 The Startup Window ......................................................................................................... 20
3.3 The Main Window ............................................................................................................. 21
3.4 The Single Tab.................................................................................................................... 22
3.5 The Multi Tab .................................................................................................................... 25
3.6 The Settings Tab ................................................................................................................ 26
3.7 The Instrument Tab ........................................................................................................... 31
3.7.1 Edit Thermometers ........................................................................ 31
3.7.2 Edit Resistors ................................................................................ 34
3.7.3 Calibration .................................................................................... 35
3.7.4 Restart Software ............................................................................ 35
3.7.5 Update Software............................................................................ 36
3.7.6 Change Password .......................................................................... 36
3.7.7 Port Settings .................................................................................. 37
3.7.8 Set Date and Time ......................................................................... 38
3.7.9 Backup Data.................................................................................. 38
3.7.10 Clear Data Files ......................................................................... 39
3.7.11 Backup Configuration ............................................................... 39
3.7.12 Load Configuration ................................................................... 39
3.8 Using the Thermometer Database .................................................................................... 40
3.9 Using the Reference Resistors Database .......................................................................... 41
3.10 Saving Instrument Configurations .............................................................................. 41
3.11 Password Protection of Key Settings ........................................................................... 42
© Page 2 of 97 Issue: 1.04
4 CONNECTING THERMOMETER SENSORS ................................. 43
4.1 The “Cable-Pod” Connectors ........................................................................................... 43
4.2 Connecting PRTs (4-wire) ................................................................................................. 43
4.3 Connecting PRTs (3-wire) ................................................................................................. 44
4.4 Connecting PRTs (2-wire) ................................................................................................. 44
4.5 Connecting Thermocouples (external Ice Point or WTP) .............................................. 45
4.6 Connecting Thermocouples (using RJ compensation) ................................................... 46
4.7 Connecting Thermistors .................................................................................................... 46
5 GOOD PRACTICE GUIDELINES .................................................... 48
5.1 Looking After Your microK ............................................................................................. 48
5.2 Making a Good Electrical Measurement ......................................................................... 48
6 THE MICROK TECHNOLOGY ........................................................ 51
6.1 The ADC ............................................................................................................................. 51
6.2 Substitution Topology ....................................................................................................... 53
6.3 Minimising Thermal EMFs .............................................................................................. 55
6.4 Solid-State Switching ......................................................................................................... 56
6.5 Inherent Stability ............................................................................................................... 56
6.6 Eliminating Self-Heating Effects ...................................................................................... 57
7 CALIBRATION ................................................................................ 58
7.1 Checking Calibration ........................................................................................................ 58
7.1.1 Checking the Master Current Source ..............................................58
7.1.2 Resistance Measurements ..............................................................59
7.1.2.1 Zero Ohms Check ..................................................................59
7.1.2.2 400 Internal Reference Resistor..........................................60
7.1.2.3 100 Internal Reference Resistor..........................................61
7.1.2.4 25 Internal Reference Resistor ...........................................61
7.1.2.5 10 Internal Reference Resistor ...........................................61
7.1.2.6 1 Internal Reference Resistor .............................................61
7.1.3 Voltage Measurements ..................................................................62
7.1.3.1 Zero Voltage Offsets ..............................................................62
7.1.3.2 Voltage Span Check ...............................................................62
7.2 Adjusting Calibration ........................................................................................................ 63
7.2.1 Calibrate Master Current Source ....................................................64
7.2.2 Calibrating the 400 Internal Reference Resistor .........................65
© Page 3 of 97 Issue: 1.04
7.2.3 Calibrating the 100 Internal Reference Resistor ......................... 66
7.2.4 Calibrating the 25 Internal Reference Resistor ........................... 66
7.2.5 Calibrating the 10 Internal Reference Resistor ........................... 66
7.2.6 Calibrating the 1 Internal Reference Resistor ............................. 66
7.2.7 Calibrating Voltage Zeros ............................................................. 67
7.2.8 Calibrating Voltage Span .............................................................. 68
8 RS232 INTERFACE ........................................................................ 70
8.1 Establishing an RS232 Connection ................................................................................... 70
9 GPIB ................................................................................................ 73
9.1 GPIB Address ..................................................................................................................... 74
9.2 Establishing a GPIB Connection ...................................................................................... 74
10 SCPI COMMAND SET .................................................................... 75
10.1 Command Terminators ................................................................................................ 75
10.2 SCPI Command Structure ........................................................................................... 75
10.2.1 SCPI Numeric Suffices.............................................................. 77
10.2.2 Parameters ................................................................................. 77
10.2.3 Units ......................................................................................... 77
10.3 Making Measurements using SCPI Commands ......................................................... 79
10.3.1 Measuring Resistance using SCPI Commands ........................... 79
10.3.2 Measuring Voltage using SCPI Commands ............................... 80
10.4 SCPI Commands ........................................................................................................... 81
10.4.1 Command: *IDN? ..................................................................... 82
10.4.2 Command: *RST ....................................................................... 82
10.4.3 Command: SENSe:FUNCtion ................................................... 82
10.4.4 Command: SENSe:FUNCtion? ................................................. 82
10.4.5 Command: SENSe:CHANnel .................................................... 83
10.4.6 Command: SENSe:CHANnel? .................................................. 83
10.4.7 Command: SENSe:VOLTage:RANGe ...................................... 83
10.4.8 Command: SENSe:VOLTage:RANGe? .................................... 84
10.4.9 Command: SENSe:FRESistance:REFerence ............................. 84
10.4.10 Command: SENSe:FRESistance:REFerence? ............................ 84
10.4.11 Command: SENSe:FRESistance:RANGe .................................. 84
10.4.12 Command: SENSe:FRESistance:RANGe? ................................ 85
10.4.13 Command: SENSe:RATio:REFerence ....................................... 85
10.4.14 Command: SENSe:RATio:REFerence? ..................................... 85
10.4.15 Command: SENSe:RATio:RANGe ........................................... 86
10.4.16 Command: SENSe:RATio:RANGe? ......................................... 86
10.4.17 Command: INITiate .................................................................. 87
10.4.18 Command: FETCh? ................................................................... 87
10.4.19 Command: READ? ................................................................... 87
10.4.20 Command: READ#? ................................................................. 87
© Page 4 of 97 Issue: 1.04
10.4.21 Command: MEASure:VOLTage? ..............................................88
10.4.22 Command: MEASure:FRES:REF? ............................................88
10.4.23 Command: MEASure:RAT:REF? ..............................................89
10.4.24 Command: CURRent .................................................................89
10.4.25 Command: TEST:CURRent .......................................................90
10.4.26 Command: CALibrate:CURRent ...............................................90
10.4.27 Command: CALibrate:REFerence .............................................90
10.4.28 Command: CALibrate:REFerence? ............................................90
10.4.29 Command: CALibrate:OFFSet ...................................................91
10.4.30 Command: CALibrate:OFFSet? .................................................91
10.4.31 Command: CALibrate:GAIN .....................................................91
10.4.32 Command: CALibrate:GAIN? ...................................................92
10.4.33 Command: CALibrate:PASSword .............................................92
10.4.34 Command: CALibrate:UNLock .................................................92
10.4.35 Command: CALibrate:LOCK ....................................................93
11 SPECIFICATION ............................................................................. 94
12 APPROVALS .................................................................................. 96
12.1 CE Declaration .............................................................................................................. 96
12.2 FCC Statement .............................................................................................................. 96
12.3 Standards Applied ........................................................................................................ 97
© Page 5 of 97 Issue: 1.04
1 Introduction
The microK-400 and micro-800 are the only precision thermometry instruments
that can achieve sub mK precision and will work with all three common
thermometer sensors (PRTs, Thermocouples and Thermistors). They are based
on a completely new measurement technology, unique to these products, that
provides better accuracy and lower noise than comparables technologies (see
section 6). They are also the first instruments of their type to be completely solid-
state, making them highly reliable.
Despite their sophistication, these instruments are very easy to use. There are no
knobs and dials, just a colour touch screen that makes use of the familiar
Windows CE™ operating system. If you use a Windows™ operating system on
your PC, you will find these instruments intuitive and easy to use. The USB port
on the front panel allows you to plug in and use other devices that work with
Windows CE™ such as a mouse, keyboard or flash drive.
Readings can be displayed in resistance ratio, ohms, volts or temperature units
(°C, °F or K). Algorithms for conversion to temperature include:
PRTs: ITS-90
Callendar-Van Dusen (IEC751 or custom
coefficients)
Thermocouples: IEC584-1 (for type B, E, J, K, N, R, S, T)
Type L polynomial
Gold Platinum Reference Function Polynomial
Thermistor: Steinhart-Hart
This manual provides a comprehensive guide to using the instrument. We
recognise that you will probably not wish to read it through at this time so
suggest you read the safety section below, take one of the tutorials in section 2
© Page 6 of 97 Issue: 1.04
and then refer back to the manual using the list of contents to find additional
information as required.
1.1 Unpacking
Your microK product should comprise the following items:
microK precision thermometer power lead suitable for your country this user manual
If any item is missing or damaged, please report this immediately to your
supplier. If possible, we recommend that you retain the packaging material in
case you need to return the instrument for calibration or service since it has been
!
designed to ensure that your microK is properly protected during transportation.
1.2 Safety
The microK is a precision instrument, designed for use in a laboratory. It
complies with the requirements of safety standard EN61010-1 (2001) and is
therefore safe to use in laboratory or light industrial environments. It is not
intended for use outdoors or in extreme environments (refer to specification in
Section 11).
The microK is likely to be connected to thermometer sensors in use and the
operator should take care to ensure that the complete system is safe. For
example, metal sheathed thermometers may be connected to the microK and then
placed in a furnace powered from a 230V electrical supply. Single fault
conditions in such a furnace could lead to the thermometer wires and therefore
the front terminals of the microK becoming electrically live and therefore a
hazard to the operator. Suitable precautions should be taken, such as using an
isolating transformer in the supply to such a furnace. If you require further advice
on safety issues, please contact Isothermal Technology or one of our appointed
distributors - we have extensive experience of thermometry and can provide
advice and equipment to help you.
© Page 7 of 97 Issue: 1.04
Retain these instructions. Use only as specified in these operating instructions or
the intrinsic protection may be impaired.
Please observe the following safety precautions:
Do not use your microK if it is damaged Only connect to an earthed supply socket. THIS UNIT IS CLASS 1
CONSTRUCTION AND MUST BE EARTHED!
Connect only to a power supply with a voltage corresponding to that on
the rating plate
This equipment is for indoor use and will meet its performance figures
within an ambient temperature range of 5°C to 40°C with maximum relative humidity of 80% for temperatures up to 31ºC decreasing linearly to 50% RH at 40ºC
Equipment is for operation at installation category II (transient voltages)
and pollution degree ll in accordance with IEC 664 at altitudes up to 2000 metres
Before replacing a fuse, DISCONNECT THE EQUIPMENT FROM THE
ELECTRICAL SUPPLY
The fuse is contained in the IEC socket on the rear panel. It must only be
replaced with a fuse of the type and rating marked on the rear panel
If a replacement fuse fails immediately, contact your local service agent.
DO NOT replace with a higher value
Always use the power cord supplied. Your sales outlet can provide a lead
suitable for your country
This equipment is for use in moderate climates only. NEVER use the
equipment in damp or wet conditions
Avoid excessive heat, humidity, dust & vibration Do not place liquid filled containers on the equipment Do not use where the equipment (or any associated accessories) may be
subjected to dripping or splashing liquids
Ensure that the power switch is easily accessible to allow the unit to be
switched off
The equipment weighs 13kg; use the handles provided. Always
disconnect the equipment from the electrical supply and any ancillary units before moving
Ensure that tabletop equipment is placed on a solid, level surface, which
is able to support its weight (and that of any attached accessories)
Ensure all cables and wires are routed safely to avoid tripping: also to
avoid sharp bends and pinches
Clean only with a damp cloth. Do not wet or allow moisture to penetrate
the unit. Do not use solvents. See section 5.1 for details of cleaning procedure
The product should be subjected to regular in-service inspections as
required by local regulations; a yearly interval is suggested
Verify that the supply cordset is undamaged and that the enclosure is
bonded to protective earth. Do not apply earth test currents to any front panel terminal nor to the shrouds of the USB, RS232 or GPIB connectors
© Page 8 of 97 Issue: 1.04
The product is designed to comply with EN 61010-1 and can be flash
tested. It is fitted with radio frequency interference suppressors. Therefore it is recommended that only a D.C. test be performed. Performing flash tests repeatedly can damage insulation
This equipment contains no user-serviceable parts. Refer all repairs to
qualified service personnel. Contact Isothermal Technology or one of our appointed distributors for details of approved service outlets
1.3 Powering Up Your microK
The microK operates on any standard AC electrical supply (88-264V RMS at 47-
63Hz) so unless your supply is unusual you can simply use the power cord
provided to connect your microK to a suitable electrical outlet.
The power switch is located at the rear of the instrument, immediately below the
IEC connector. When you turn your microK on it will go through a standard
!
Windows CE™ boot sequence and then display a Window containing the
Resume button. Pressing Resume restarts the microK in the configuration it was
in prior to the last power down, you will then see the Main Window and be ready
to operate your instrument.
Before pressing the Resume button, make sure that any thermometers attached
will not be damaged by the sense currents that will be applied (the last values
used). If there is a problem, disconnect your sensors before proceeding and then
reconnect them after changing the sense current to the required values.
1.4 A Quick Tour of Your microK
On the front panel of your microK you will find the input terminals for the three
measurement channels, the touch screen/colour display and a USB connector:
© Page 9 of 97 Issue: 1.04
USB
Input
Terminals
Touch Screen
& Display
Connector
Input Terminals: The input terminals accept 4mm plugs, spades or bare wires.
The current (I) and voltage sense (V) terminals are spaced on ¾” centres so that
standard BNC to 4mm adaptors (not supplied) can be used to connect to
thermometer sensors that have BNC terminations.
The contact material for the connectors is gold plated tellurium-copper, offering
extremely low thermal EMFs when connected to copper wires/connectors. This
is essential when using precision thermocouples.
Display: The display is a colour TFT VGA (640 x 480) LCD with a long-life
CCFL (cold-cathode fluorescent lamp) backlight. A touch panel is mounted in
front of the display so that you can control the instrument by simply touching the
buttons displayed on the LCD. The touch panel is an industrial grade components
offering good durability. It is intended to be operated with a finger. A stylus
intended for PDAs may also be used – never use a sharp object with the touch
panel as this will lead to premature failure.
The USB Connector: The USB connector is primarily intended to allow you to
connect a USB flash drive, store measurement results and transfer these to a PC.
The flash drive may also be used to backup the databases of thermometers,
reference resistors and instrument configurations for your microK.
© Page 10 of 97 Issue: 1.04
Connector
(unused)
GPIB
Connector
Other USB devices such as a mouse, keyboard, keypad or USB hub may be
connected to the USB port and used, provided they utilise the standard class
drivers provided with the Windows CE™ operating system incorporated in your
microK.
On the rear panel of your microK you will find the electrical supply
connector/power switch/fuse module plus the interface connectors that allow you
to connect your instrument to a PC. The RS232 and GPIB ports allow you to
control the instrument and take measurements from a PC with your own software
(see section 8). The command protocol employs the widely used SCPI format
(see section 9 for details). The USB connector is provided for future expansion
and should not be used on this model.
IEC 60320-1
Power Connector
On/Off
Switch
Supply Fuse
USB-B Connector
RS232
© Page 11 of 97 Issue: 1.04
2 Making a Measurement (Tutorials)
The operator interface is powered by the Windows CE™ operating system. With
the widespread use of Windows applications on PCs, driving your microK should
be fairly intuitive. As with most Windows™ applications, it is possible to
navigate your way through a process using a number of different paths, so rather
than describe each window and the function of each button and then leaving you
to work out how to operate the instrument, we provide you with a number of
tutorials that illustrate common applications. Even if these do not describe your
application, it is worth working through one of them since it will provide you
with a broad understanding of the wide range of features available on your
microK.
A traditional description of the function of all the features provided by the
operator interface is given in section 3.
2.1 Measuring Temperature with a Calibrated SPRT
From startup, press Resume to restart your microK:
© Page 12 of 97 Issue: 1.04
Before making any measurement, you will have to enter information about the
thermometer into the microK’s database. When the main window opens, press
the Instrument tab:
In the Instrument tab, press Edit Thermometers:
You will be prompted to enter the password (set to “1234” initially, but this
should be changed before using the microK in a real measurement/calibration
application to ensure security – see section 3.7.6). Enter the password to open the
thermometer database window. Press New to create a new thermometer entry in
the database and enter data by pressing the up/down or ellipses buttons by each
field:
© Page 13 of 97 Issue: 1.04
Press Coefficients to open the coefficients window. Press the ellipses buttons to
enter the calibration data for the thermometer from its calibration certificate:
Press OK to close the coefficients window and press OK again to close the
thermometer database window.
Press the Settings tab in the main window and disable channels 2 and 3 (this will
speed up measurements on channel 1, which we will be using) by using the
up/down buttons by the Channel box to select these channels and pressing
Disable. Use the up/down buttons to return to channel 1. Use the up/down (or
ellipses) buttons by each box to select the required settings:
© Page 14 of 97 Issue: 1.04
The sense current is assumed to be 1mA in this tutorial. It can be set to other
values using the ellipses button by the Current mA box (see section 3.6).
Connect the SPRT to channel 1 (see section 4.2 for details on how to connect 4-
wire PRTs). The tutorial assumes you are using the internal 100 reference
resistor. To use an external reference resistor, first enter information on the
resistor into the microK’s database (see section 3.7.2), then in the Settings tab
select the channel to which the reference is connected in the Reference Channel
box and the reference resistor entry in the Reference Resistor box.
Press the Single tab to see the measurements in numeric and graphical form. Set
the graph scales manually by making a note of the current Mean value and
pressing Set Scales. This opens the graph scales window. Ensure that autoscaling
is turned off (press Autoscale to toggle autoscaling on/off, the status of
autoscaling is shown by the adjacent indicator). Use the ellipses buttons to set the
y-axis limits to be 0.005°C above and below the current mean value. Use the
radio and up/down buttons to set the x-axis to 5 minutes:
© Page 15 of 97 Issue: 1.04
Press OK to return to the main window. The measurement system will now
accumulate data and after 5 minutes the graph will begin to scroll to show the
last 5 minutes of data:
2.2 Measuring Temperature with a Thermocouple
From startup, press Resume to restart your microK:
© Page 16 of 97 Issue: 1.04
This tutorial describes the use of an uncalibrated type-N thermocouple. The
microK can be used with calibrated thermocouples, in which case you will need
to enter information about the thermometer into the microK’s database and (see
section 3.7.1) and select this thermometer entry from the Thermometer box in the
Settings tab (see later in this tutorial). When the main window opens, press the
Settings tab:
In the Settings tab disable channels 2 and 3 (this will speed up measurements on
channel 1, which we will be using) by using the up/down buttons by the Channel
box to select these channels and pressing Disable. Use the up/down buttons to
return to channel 1. Use the up/down buttons by each box to select the required
settings:
© Page 17 of 97 Issue: 1.04
The reference junction is assumed to be an ice point in this tutorial. It can be set
to the water triple-point or to use a temperature measured by one of the other
channels (see sections 3.6 and 4.6) if using reference junction compensation.
Connect the N-type thermocouple to channel 1 (see section 4.5 for details on how
to connect thermocouples). Press the Single tab to see the measurements in
numeric and graphical form. Set the graph scales manually by making a note of
the current Mean value and pressing Set Scales. This opens the graph scales
window. Ensure that autoscaling is turned off (press Autoscale to toggle
autoscaling on/off, the status of autoscaling is shown by the adjacent indicator).
Use the ellipses buttons to set the y-axis limits to be 0.01°C above and below the
current mean value. Use the radio and up/down buttons to set the x-axis to 5
minutes:
© Page 18 of 97 Issue: 1.04
Press OK to return to the main window. The measurement system will now
accumulate data and after 5 minutes the graph will begin to scroll to show the
last 5 minutes of data:
© Page 19 of 97 Issue: 1.04
3 Driving Your microK
3.1 Introduction
The operator interface for the microK is provided through the touch screen and
colour display. The software is written in Microsoft’s new C# language and runs
under the Windows CE™ operating system. This provides a familiar and easy to
use interface.
The buttons in the microK software have been sized and positioned to allow you
to operate the instrument with your finger. If numeric or alphanumeric data needs
to be entered, a ‘soft’ keyboard window appears. If you are unfamiliar with touch
screen devices such as PDAs, you may initially feel that the touch screen
interface is slow compared with using a mouse and keyboard, but with a little
practice you should quickly become comfortable with the technology and
appreciate its benefits. However, if you prefer to use a USB mouse or keyboard,
you can simply plug these into the USB connector on the front panel. They will
immediately become active provided they use the standard class drivers built into
the Windows CE™ operating system used in the microK. To use more than one
USB device, simply plug a USB hub into the front panel connector.
Calibration data entered into the microK’s database is password protected and
includes recalibration dates (the microK will alert you if you try to use a
thermometer or reference resistor after its recalibration date). This feature
ensures the integrity of your measurements and helps you to comply with the
requirements of accreditation bodies.
3.2 The Startup Window
When power is applied to the microK, it goes through a Windows CE™ boot
sequence that loads the operating system and then starts the microK application.
The Startup Window then appears showing the microK’s model and serial
number together with the version numbers for the measurement system firmware
and microK software used. The window also lists the details of any
© Page 20 of 97 Issue: 1.04
“microsKanner” multiplexers connected to the microK (the microsKanner
manual describes how to use “microsKanner” multiplexers with your microK
bridge). The Startup Window contains the Resume button:
!
Resume: Pressing the Resume button places the microK in the state it was in
prior to its last power down. Before pressing Resume, make sure that any
thermometers attached will not be damaged by the sense currents that will be
applied (the last values used). If there is a problem, disconnect your sensors
before proceeding and then reconnect them after changing the sense current to
the required values.
3.3 The Main Window
Having pressed Resume in the Startup Window, the user is presented with the
Main Window.
© Page 21 of 97 Issue: 1.04
displays all three ch
annels in numeric form
allows you to configure the channels to perform the
All functions to control the microK are available in this Window, which is
divided into four tabs:
Single
displays a single channel in numeric and graphical form
Multi
Settings
required measurement
Instrument
allows you to enter information on the thermometers and reference resistors used with the instrument into its database. It also allows you to save/load measurement data and instrument configurations in the database and to change instrument settings such as the GPIB address, time/date and security passwords
The Main Window opens with the Single tab selected.
3.4 The Single Tab
Select this tab to see a single channel in both numeric and graphical form:
© Page 22 of 97 Issue: 1.04
Channel box: Use the up/down buttons by the Channel box to select the channel
you wish to view. Alternatively press the ellipses button to open a numeric entry
window and select the channel directly. The ellipses button is particularly useful
when the microK is used with microsKanner multiplexers, which can provide up
to 90 expansion channels. You can only view channels that are enabled, so if a
channel is disabled in the Settings tab it will not appear in the sequence and will
not be measured. This means that the up/down buttons have no effect if only one
channel is enabled.
The numeric display shows the last reading together with the mean and standard
deviation of the most recent of readings (the number of readings in the statistics
is set in the Settings tab).
Clear Graph: Press the Clear Graph button to clear the graph and re-start the
autoscaling feature. This can be used to eliminate the effect of a large transient
result (perhaps caused when changing connections to the thermometer) on the
autoscaling function.
Clear button: Press the Clear button to clear (reset) the statistics.
Set Scales button: Use the Set Scales button to open a new window and set the x
and y axes for the graphical output:
© Page 23 of 97 Issue: 1.04
In this window, use the up/down buttons by the Channel box to select the
channel whose scales you wish to change (the window opens with the currently
displayed channel selected). Use the radio buttons to select either Seconds or
Minutes as the units. Use the up/down buttons by the Scale / Seconds box to
change the x-axis (time) scale (the same x axis scale is used for all channels). Or,
press the ellipses button to open a Numeric Keypad window and enter the
required scale directly. Any value between 1 second and 1000 minutes
(inclusive) may be used.
Use the ellipses by the Minimum and Maximum boxes to open a Numeric Keypad
window and set the y-axis scale. Alternatively, press the Autoscale button to
scale the y-axis so that it will show all the readings. The indicator by the
Autoscale button shows whether autoscaling is active.
The numeric and graphical outputs are updated as new readings are made. The
time between readings depends on the settings for that channel and any other
channels that are also enabled. This can vary between 2 seconds and 5 minutes.
The measurement sample time for a PRT or thermistor is <2s and for a
thermocouple it is <1s. The reading rate for three PRTs with one sample per
reading will therefore be <6s.
Start Save button: Press the Start Save button to record all the measurements
being made. You will then be offered the choice (buttons) of storing the data to
the Internal “DiskOnChip” (solid state drive) memory or an External USB flash
© Page 24 of 97 Issue: 1.04
drive. Data stored to the internal memory can subsequently be transferred to an
external USB flash drive using the Backup Data button in the Instrument tab (see
section 3.7.9). The internal memory used has a capacity of at least 32Mbytes; this
will accommodate approximately 1,300,000 PRT readings or about 30 days of
continuous logging at the fastest rate. Select the destination for the data to open
an Alpha-Numeric Keyboard window and enter a filename. An automatically
incrementing default name will appear in the entry box, press OK to accept this
or use the backspace and alpha-numeric keys to enter your preferred filename. If
you try to use a filename that has already been used, you will have to enter the
password to proceed (see section 3.11 for details on password protection). The
Shift key toggles the alpha keys between upper and lower case. The indicator by
the Start Save button in the Single tab shows whether readings are being recorded
to a file. The measurements are written to the file in comma delimited ASCII text
form allowing easy importation into Microsoft Excel™ or other applications.
3.5 The Multi Tab
Select this tab to see all three channels in numeric form only. The numeric
display shows the last reading together with the mean and standard deviation for
the most recent readings (the number of readings in the statistics is set in the
Settings tab) for each of the channels that are enabled.
© Page 25 of 97 Issue: 1.04
Clear buttons: Press the Clear button to clear (reset) the statistics for each
channel
The displayed values are updated as new readings are taken. The time between
readings depends on the settings for that channel and any other channels that are
also enabled. This can vary between 2 seconds and 5 minutes. The measurement
time for a PRT or thermistor is <2s and for a thermocouple it is <1s. The reading
rate for three PRTs with one sample per reading will therefore be <6s.
Start Save: Press the Start Save button to record all the measurements being
made. You will then be offered the choice (buttons) of storing the data to the
Internal “DiskOnChip” (solid state drive) memory or an External USB Flash
Drive. Data stored to the internal memory can subsequently be transferred to an
external USB Flash Drive using the Backup Data button on the Instrument tab
(see section 3.7.9). The internal memory used has a capacity to at least
32Mbytes; this will accommodate approximately 1,300,000 PRT readings or
about 30 days of continuous logging at the fastest rate. Select the destination for
the data to open an Alpha-Numeric Keyboard window and enter a filename. An
automatically incrementing default name will appear in the entry box, press OK
to accept this or use the backspace and alpha-numeric keys to enter your
preferred filename. If you try to use a filename that has already been used, you
will have to enter the password to proceed (see section 3.11 for details on
password protection). The Shift key toggles the alpha keys between upper and
lower case. The indicator by the Start Save button in the Multi tab shows whether
readings are being recorded to a file. The measurements are written to the file in
comma delimited ASCII text form allowing easy importation into Microsoft
Excel™ or other applications.
3.6 The Settings Tab
Select this tab to configure each channel for the required measurements.
© Page 26 of 97 Issue: 1.04
Channel box: Use the up/down buttons to select the channel you wish to
configure. Alternatively press the ellipses button to open a numeric entry window
and select the channel directly. The ellipses button is particularly useful when the
microK is used with microsKanner multiplexers, which can provide up to 90
expansion channels.
Enable/Disable button: Use this button to enable or disable a channel. Disable
unused channels to skip measurements on that channel and reduce the cycle time
through the channels. The nearby indicator shows whether the currently selected
channel is enabled.
Thermometer box: Use the up/down buttons to select the required thermometer.
Initially this box will contain only default PRT, Thermocouple or Thermistor.
You can use these entries for uncalibrated thermometer sensors. If you add
thermometers to the microK’s Thermometer Database (see section 3.7.1), these
will also appear in the list within this box. The ellipses button opens a list of all
available thermometers and is useful if you have a large number of entries in the
database.
Conversion box: If the selected thermometer is one of the default types, you can
use the up-down buttons to select a standard polynomial function for converting
the thermometer’s base measurement to temperature. If the selected thermometer
© Page 27 of 97 Issue: 1.04
is one that your have created in the Thermometer Database this box will be fixed
and will be the conversion type defined for that thermometer.
Range box: Use the up/down buttons to select the required range. The microK
has three ranges for each thermometer type. Use the 0.125V range for all
thermocouples as this provides the most suitable input voltage range. The
resistance ranges offered depend on the sense current (use the ellipses button by
the Current box in this tab to change the current). Choose a resistance range that
is greater than or equal to both the highest thermometer resistance and reference
resistor value you intend to use.
Reference Channel box (PRTs and thermistors only): Use the up/down
buttons to select the reference resistor used in the resistance measurement; either
one of the internal reference resistors or the channel to which an external
resistance standard is connected. Alternatively press the ellipses button to open
the Reference Channel window and select the channel directly. The External
button in this window allows you to select one of the microK’s three external
channels or any expansion channel provided by a microsKanner.
Reference Resistor box (PRTs and thermistors only): If an internal reference
resistor was selected in the Reference Channel box, this box will show the
calibrated value of that resistor. If one of the input channels was selected in the
Reference Channel box, this box will contain a list of the reference resistors in
the microK’s database (entered using the Edit Resistors button in the Instrument
tab – see section 3.7.2) and “Uncalibrated”. Select the required resistor using the
up/down buttons. Select “Uncalibrated” if you only want to make a resistance
ratio measurement between two channels. The ellipses button opens a list of all
available reference resistors and is useful if you have a large number of entries in
the database.
Reference Junction box (thermocouples only): This contains 0°C, 0.01°C and
a list of any input channels that have been configured to measure temperature.
Choose 0°C or 0.01°C if you are using an external reference junction in an ice-
point bath or water triple point respectively. Choose one of the measurement
channels if you want to measure the temperature of the reference junction with a
© Page 28 of 97 Issue: 1.04
thermometer sensor on that channel. The software will not allow you to use a
thermocouple channel as the reference junction sensor (to avoid circular
dependencies). You must specify a thermometer sensor for the reference junction
channel that includes temperature conversion otherwise it will not appear in the
Reference Junction box.
Units box: Use the up/down buttons to select the required units. The base units
(resistance ratio “Rt/Rs” for PRTs or thermistors and “Volts” for thermocouples)
are always available. “Ohms” will also be available (for PRTs and thermistors) if
the value of the reference resistor is available (Reference Resistor not set to
“Uncalibrated”). If the thermometer sensor chosen has a temperature conversion
specified, then temperature units °C, °F and K are also available.
Current box (PRTs and thermistors only): Press the ellipses button to change
the sense current. Use the buttons in the window that then opens to select the
required sense current:
The standard PRT sense currents are available directly through the buttons
provided. Press the Custom button to specify a different value (range 0 to 10mA),
this will then open a Numeric Keypad window.
Buttons for ÷2 and ×2 are included to make extrapolation to zero power easier
since changing the current in a resistor by a factor of 2 changes the power
dissipated by a factor of 2. A PRT’s calibration may be defined at a specified
current. Alternatively, it may be specified at zero power. In this case, you need to
© Page 29 of 97 Issue: 1.04
determine what the resistance of the PRT would be at zero power. This is most
easily done by measuring its resistance at the normal sense current and then increasing the current by a factor of 2 and repeating the measurement. If you
subtract the change in resistance from the first resistance value (at nominal
current) this gives you the resistance of the PRT at zero power. Alternatively, the
second measurement can be made at a current that has been reduced by a factor of 2. In this case the change in resistance should be subtracted from the second
resistance value (current scaled by 2) to determine the zero power resistance.
The accuracy of the current source is 0.4%, so the uncertainty associated with
this extrapolation back to zero power is only 0.8% of the measured resistance
change, which should be very small. Other factors such as the fact that the PRT
is not a constant resistance and so the power dissipated has changed between the
two measurements are insignificant.
Samples per Reading box: Use the up/down buttons to select the number of
samples you want to average together to form a single reading. Alternatively,
press the ellipses button to open a Numeric Keypad window and enter the
required number directly. The number of samples may be between 1 and 100
inclusive.
Increasing the number of samples will slow down the rate at which readings are
made, but will also reduce the uncertainty associated with these readings (as
shown by the standard deviations reported in the “Single” and “Multi” tabs). This
uncertainty reduction is proportional to the square-root of the number of samples
(for example: selecting 9 samples per reading will reduce the uncertainty
associated with each reading by a factor of 3).
Readings in Rolling Statistics box: Use the up-down buttons to select the
number of samples in the rolling statistics. Alternatively press the ellipses button
to open a Numeric Keypad window and enter the value directly. Values between
1 and 1000 (inclusive) may be used.
© Page 30 of 97 Issue: 1.04
Loading...
+ 67 hidden pages