Isotech MEDUSA 511 Specifications

48
|
CALIBRATION SOLUTIONS
Dry Block Calibrator
30 to 700°C
510 Medusa & 511 Medusa 3
g
45 x 285mm Calibration Volume
g
Use for Comparison and Fixed Point Calibration
g
Use with very long thermometers
Isotech have a wide range of Dry Blocks to suit probes requiring a large immersion depth. These products feature large and deep calibration volumes. As such they are less portable than the earlier Dry Blocks, but have higher capacities and retain outstanding temperature uniformity, this uniformity is so good that these larger products are also apparatus for Secondary Laboratories to realize the Fixed Points of ITS-90.
Medusa 510 has a maximum operating temperature of 550°C. The Medusa Model 511 can be used to 700°C and features three zone control. In addition to the main heating zone there are additional top and bottom heaters which compensate for the end losses creating a constant temperature zone across the well.
For Comparison Calibration the Medusa should be used with an insert, the standard insert has six 8mm pockets 250mm deep. Also available is an insert 44mm diameter x 170mm deep which is suspended from the top of the block so that the height is user adjustable. For flexibility the Medusa can also be used with accessories for infrared thermometers and surface sensors. The Medusa is available in two models, the BASIC (B) and the SITE (S). The B model includes a sophisticated temperature controller with a dual display for Set Temperature and Dry Block Temperature.
The S model includes a built-in digital thermometer to which an external standard thermometer can be connected giving greater accuracy, eliminating temperature gradient and loading errors. Also included in the site model is a timer which can set the bath between two temperatures, and automate ITS­90 fixed point operation. For Surface Sensor and Blackbody use an external thermometer is recommended. For laboratory accuracy the Medusa can be used with a high-end temperature indicator such as an Isotech TTI model.
Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature. Increased resolution of ±0.01 available throughout the range via the PC interface and from 0.01 to +99.99 locally on the auto­ranging front display. The controller features multi-point block to display correction giving good absolute accuracy.
The S model has universal sensor input allowing Platinum Resistance Thermometers, Thermocouples (types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including 4-20mA current transmitters to be displayed on the in-built indicator. The indicator can be programmed with up to five calibration points to provide high accuracy digital probe matching. The
addressable over the communications link.
indicator and controller are both
Fixed Point Cells Available
Gallium 29.7646°C
Indium 156.5985°C
Tin 231.928°C
Zinc 419.527°C
Aluminium 660.323°C
49
CALIBRATION SOLUTIONS
|
Specification
0.3°C
0.25°C
0.2°C
0.15°C
0.1°C
0.05°C
0°C 100°C 200°C 300°C 400°C 500°C 600°C 700°C
0°C
510 Medusa
Medusa 3 Audit Calibration (Similar Sensors) S Model with 935-14-95 Senso and UKAS
Medusa 2 Audit Calibration (Similar Sensors) S Model with 935-14-95 Senso and UKAS
Medusa 3 Radial Homogeneity
Medusa 2 Radial Homogeneity
Model 510 Medusa 511 Medusa 3
Temperature Range 30°C to 550°C 50°C to 700°C
Absolute stability Metal Block Bath ±0.03°C over 30 minutes Blackbody Source ±0.1°C Surface Sensor Calibrator ±0.5°C ITS-90 Fixed Point ±0.001°C
Computer Interface Included with Software
Cools from 550°C to 30°C in 5 hours
Heats from 30°C to 550°C in 90 minutes
Uncertainties Refer to Uncertainties Graph
Calibration volume 45mm diameter by 285mm deep
Standard Insert Six 8mm pockets all 250mm deep
Display Resolution (0.01) to 99.99 (0.1) 100.0 to 650.0 PC can display 0.01 across whole range with the software included
Indicator units °C, °F, K
Power 108 to 130V or 208 to 240V 50 / 60Hz 1000 Watts 1800 Watts
Overall dimensions Height 430mm Width 310mm Depth 300mm Weight 17kg 25kg
Performance and Use
Features (Basic & Site) Medusa
Dry Block ü
Surface Sensor Option ü
Infrared Calibration Option ü
ITS-90 Fixed Point Cells ü
Additional 8mm Pre-heat Pocket ü
Configurable Units: °C, °F and K ü
Supply Voltage Power Correction ü
Calibration and Uncertainty
A certificate, traceable to National Standards, is included as standard. Recommended is an optional UKAS five-point calibration.
The accuracy of the Medusa will depend very much on the mode of use, see the Uncertainty Graph for typical uncertainties. NTPL calculate the uncertainties to UKAS requirements. The Medusa meets the Calibration Capacity requirements of EA-10/13, “EA Guidelines on the Calibration of Temperature Block Calibrators.”
Additional Features (Site) Medusa
Independant Temperature indicator ü
Universal Input Types PT100 ü
Thermocouples Types K,N,R,S,L,PL2,T,J,E ü
Linear Process Inputs Including 4-20 mA ü
Stand Alone Thermostat Testing ü
Thermostat Testing With PC ü
Five Point Digital Probe Matching ü
Configurable Units: °C, °F and K ü