Iso-Tech IDM 98-III, IDM 99-III Instruction Manual

IDM 99-III
Instruction Manual
IDM 98-III & IDM 99-III
Digital Multimeters
Africa
Iso-Tech 1 & 2 Indianapolis Street Kyalami Business Park Kyalami, Midrand, South Africa
Asia
Iso-Tech 460 Alexandra Road, #15-01A PSA Building Singapore 119963
Europe
Iso-Tech PO Box 99 Corby Northamptonshire NN17 9RS United Kingdom
Japan
Iso-Tech West Tower (12th Floor) Yokohama Business Park 134 Godocho, Hodogaya Yokohama, Kanagawa 240-0005 Japan
USA
Iso-Tech 7410 Pebble Drive Fort Worth Texas 76118-6961
Canada
Iso-Tech 1701 Woodward Drive Ste 108 Ottawa Ontario K2C 0R4, Canada
South America
Iso-Tech Av. Pdte. Eduardo Frei M. 6001-71 Centro Empresas El Cortijo Conchali, Santiago, Chile
IDM 98III / 99III English IDM 98III / 99III English
2
3
SECTION 1 –SAFETY INFORMATION
Safety Information
Understand and follow operating instructions carefully.
Use the instrument only as specied in this manual; otherwise, the
protection provided by the instrument may be impaired.
WARNING
Identies hazardous conditions and actions that could cause BODILY HARM or DEATH
CAUTION
Identies conditions and actions that could DAMAGE the instrument or
equipment under test
WARNING
- Examine the instrument and probes before use. Do not use the instrument if it is wet or damaged
- When using test leads or probes, keep your ngers behind the nger guards.
- Remove the test lead from the instrument before opening the battery cover or instrument case.
- Always use the correct terminals, switch position and range for measurements.
- Never attempt a voltage measurement with the test leads inserted into the “A”input terminals.
- Verify the instrument is operating correctly by measuring a known voltage before use. If in doubt, have the instrument serviced.
- Do not apply more than the rated voltage, as marked on the instrument, between terminals or between any terminal and earth ground.
- Do not attempt a current measurement when the open-circuit voltage is above the fuse protection rating.
- Only replace a fuse with the correct type and rating as specied in this instruction manual.
- Use caution when measuring voltages above 30 Vac rms, 42 Vac
peak or 60 Vdc. These voltages pose a shock hazard.
- To avoid false readings that can lead to electric shock, replace the battery as soon as the low battery indicator appears in the display.
- Disconnect the circuit power and discharge all high-voltage capacitors before making resistance, current, continuity, diode or capacitance measurements.
- Do not use the instrument in a hazardous area or around explosive gasses or vapours.
- Wear suitable personal protective equipment when working around or
near hazardous live conductors which could be accessible.
Risk of electric shock
See instruction manual
DC measurement
Equipment protected by double or reinforced Insulation
Battery
Fuse
Earth
AC measurement
Conforms to EU directives
Dispose of in accordance with local regulations.
CAUTION
Disconnect the test leads from the test points before changing the position of the function rotary switch.
Never connect a source of voltage with the function rotary switch in
the
“Ω”, “
”, “°C” , “mA”, or “A” position.
Do not expose the instrument to extremes in temperature or high
humidity.
Never set the instrument in the “Ω”, “ ”, “°C” , “mA”, or “A” position
to measure the voltage of a circuit, this could result in damage the instrument and the equipment under test.
If possible, do not work alone, so assistance can be given if required
The following symbols may appear on the instrument and this instruction manual:
Unsafe Voltage
To alert you to the presence of a potentially hazardous voltage, when the
instrument detects a voltage 30 V or a voltage overload (OL) in V, mV . The
symbol is displayed.
Maintenance
Do not attempt to repair this instrument. It contains no user-serviceable parts.
Repair or servicing should only be performed by qualied personnel. This
instrument should be calibrated yearly, or more frequently if used in harsh conditions or if it is suspected of being inaccurate.
IDM 98III / 99III English IDM 98III / 99III English
4
5
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.
SECTION 2 - METER DESCRIPTION
Instrument Description
Front Panel Illustration
1. LCD display : 6000 counts
2. Push-buttons.
3. Rotary switch for turn the Power On / Off and select the function.
4. Input terminal for “A” (current measurement)
5. Input terminal for “V”, “Ω”, “ ”, Hz, and °C functions.
6. Common (ground reference) input terminal.
7. Input terminal for mA current measurement.
Features
6000 count digital display.
62 segment bar graph .
Extra large scale white backlit display
True RMS
0.08% basic DCV accuracy
Automatic AC/DC Voltage detect with low impedance (Auto-V LoZ)
VoltSense (Non-contact voltage detection)
Smart Data Hold
Peak Hold (1ms) (IDM99-III only)
Min/ Max function
AC+DC function
Frequency counter on AC mode
Capacitance measurement
Temperature measurement (IDM99-III only)
Low battery indicator with segments
Auto Power Off (20 minutes)
Shock proof from 4 feet drop
CAT IV 600V/CAT III 1000V Safety standard
Unpacking and Inspection
Upon removing the instrument from its packaging you should have the following items.
1. IDM98-III or IDM99-III digital multimeter.
2. Test leads set (one black, one red)
3. Temperature probe (IDM99-III only)
4. User manual
5. Protective holster
6. Battery (installed)
SECTION 3 – MEASUREMENTS
Preparation and caution before measurement
Observe all warnings and cautions
When connecting the test leads to the device under test connect
the common test lead before connecting the live test lead ; when
removing the test leads remove the test live lead before removing the common test lead.
The gures on the following pages show how to make basic
measurements.
IDM 98III / 99III English IDM 98III / 99III English
6
7
Measuring AC / DC Voltage
Rotate the switch and press the Function button to select the measuring function.
See “Using the Function Buttons”
Measuring AC/DC Current
Rotate the switch and press the Function button to select the measuring function. See “Using the Function Buttons”
Measuring Frequency
Rotate the switch and press the Function button to select the measuring function.
See “Using the Function Buttons”
Measuring Resistance/Capacitance
Rotate the switch and press the Function button to select the measuring function. See “Using the Function Buttons”
IDM 98III / 99III English IDM 98III / 99III English
8
9
Measuring Continuity / Diode
Rotate the switch and press the Function button to select the measuring function. See “Using the Function Buttons”
Measuring Temperature °C / °F (for IDM99-III only)
Rotate the switch and press the Function button to select the measuring function. See “Using the Function Buttons” (°C / °F)
IDM 98III / 99III English IDM 98III / 99III English
10
11
Using The Function Buttons
Function Button
Press the Function button to change the function on the same switch position.
RANGE Button
Auto Range Mode Manual Range Mode
MIN/MAX Button
The MAX/MIN mode records the min. and max. input values. When the input goes below the recorded min. value or above the recorded max. value, the meter records the new value . Press “HOLD” button to pause the recording.
Peak HOLD (for IDM99-III only)
In the Peak HOLD function, the meter records the peak min. value and the
peak max. value; when the inputs goes below the recorded peak min. value or
above the recorded peak max. value ,the meter records the new value . Press “HOLD” button to pause the recording.
IDM 98III / 99III English IDM 98III / 99III English
12
13
Smart HOLD
The meter will beep continuously and the display will ash if the measured
signal is larger than the display reading by 50 counts. Smart Hold will not be activated if the hold value is not scrolled over 50 counts when measuring AC or DC Voltage /Current with Auto-V(LoZ) Range.
VoltSense
Press and hold down the “VoltSense” button down to activate the VoltSense function.
CAUTION
The number of dashes displayed indicates the electric eld strength.
Be aware, if no indication, voltage could still be present
Backlight
Press the Backlight button to turn the backlight on/off.
Disable Auto Power Off
Reactivate the instrument by rotating the switch or by pressing any button.
IDM 98III / 99III English IDM 98III / 99III English
14
15
Auto Power Off Disable
Rotate the switch to the “OFF” position, Press and hold down the Function button and turn the instrument on.
Testing the LCD Display
Rotate the switch to the “OFF” position, Press and hold down the Smart­HOLD button and turn the instrument on.
SECTION 4 – BATTERY AND FUSE REPLACEMENT
Fuse Replacement
Caution
Use only fuses with the following fuse rating:
440mA,1000V IR 10KA Fast acting fuse (size 35 x 10mm)
DMM-B-44/100 – 10,000A, 1000 VAC, unity power factor and 10,000A, 1000 VDC with time constant of 2.2ms
11A, 1000V IR 20KA Fast acting fuse (size 38 x 10mm)
DMM-B-11A – 20,000A, 1000 VAC, power factor 0.2; 20,000A, 1000 VDC, time constant 10ms. (size 38 x 10mm)
CAUTION
The following safety information must be observed to ensure maximum personal safety during the operation of this instrument.
1.
To avoid electric shock, disconnect the test leads before replacing the instrument fuses or batteries.
2.
When replacing the instrument batteries, do not mix batteries of different types or old and new batteries.
3. Check the battery polarity carefully when inserting the batteries.
4.
Do not short-circuit used batteries, disassemble them, or throw them in a re.
Doing so may cause the batteries to explode.
5. Dispose of the used batteries in accordance with local regulations.
IDM 98III / 99III English IDM 98III / 99III English
16
17
Low battery and Battery Replacement
Replace the battery as soon as the low battery indicator appears to avoid false reading.
Refer to the following gure to replace the batteries
SECTION 5 - SPECIFICATIONS
General Specications
Maximum voltage applied to any terminal : 1000 Vac rms or 1000 Vdc. rms
Display : 6000 counts. Polarity Indication : Automatic, positive implied, negative indicated. Overrange Indication : OL Batteries Life : 150hours alkaline battery (No backlight)
Low Batteries Indication : Voltage drops below operating voltage, “
will ash.
Power Requirement : PP3 9V battery Auto Power Off : 20 minutes. Operating Temperature : -10 ~10°C
10°C ~ 30°C (80% RH), 30°C ~ 40°C (75% RH), 40°C ~ 50°C (45%RH) Storage Temperature : -20°C to 60°C , 0 to 80% R.H. (batteries not
tted)
Temperature Coefcient : 0.15 x (Spec.Accy) / °C, < 18°C or > 28°C .
Measure : Samples 3 times per second . Altitude : 6561.7 ft (2000m) Safety : Complies with EN61010-1, UL61010-1,IEC 61010-1, CAT IV
600V, CAT III 1000V
Measurement
Category
Application
I
Measurements on circuits not directly connected to mains. Examples include: Measurements on battery powered equipment and specially protected (internal) mains-derived circuits.
II
Measurements on circuits directly connected to the low voltage installation. Examples include: Household appliances, portable tools and similar equipment.
III
Measurements performed in the building installation. Examples include measurements on distribution boards, junction boxes, socket-outlets and wiring and
cables in the xed installation.
IV
Measurements performed at the source of the low­voltage installation. Examples include measurements on primary overcurrent protection devices and electricity Instruments
Weight : 460g (including battery) Dimensions (W x H x D) : 94mm x190mm x 48mm with holster. Pollution degree : 2 EMC : EN 61326-1 Shock vibration : Sinusoidal vibration per MIL-PRF- 28800F
(5 ~ 55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on concrete oor. Indoor use only.
IDM 98III / 99III English IDM 98III / 99III English
18
19
Electrical Specications
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C < 80%RH.
(1) Voltage
Input Protection : 1000VDC or 1000VAC rms AC Frequency Response : 50Hz ~ 1KHz Input Impedance : 10MΩ, <100pF AC Conversion Type :
AC conversions are ac-coupled, true rms responding, calibrated to the sine wave input. For non-sine wave add the following Crest Factor corrections: For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy. For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy. For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
60.00mV
DC
0.01mV ± (0.08% + 10d)
600.0mV
DC
0.1mV ± (0.1% + 2d) ± (0.08% + 2d)
6.000V
DC
0.001V ± (0.09% + 2d) ± (0.08% + 2d)
60.00V
DC
0.01V ± (0.09% + 2d) ± (0.08% + 2d)
600.0V
DC
0.1V ± (0.09% + 2d) ± (0.08% + 2d)
1000V
DC
1V ± (0.09% + 2d) ± (0.08% + 2d)
60.00mV
AC
0.01mV ± (1.20% + 5d)
600.0mV
AC
0.1mV ± (1.5% + 5d) ± (1.20% + 5d)
6.000V
AC
0.001V ± (1.0% + 3d) ± (0.80% + 5d)
60.00V
AC
0.01V ± (1.0% + 3d) ± (0.80% + 5d)
600.0V
AC
0.1V ± (1.0% + 3d) ± (0.80% + 5d)
1000V
AC
1V ± (1.0% + 3d) ± (0.80% + 5d)
(2) Auto-V
Input Protection : 1000V
DC
or 1000VAC rms
AC Frequency Response : 50Hz ~ 1KHz Input Impedance : Approx. 3kΩ. AC Conversion Type :
Conversion type and additional specication are same as voltage
function.
(3) Current
Input Protection : Protected by high energy fuses.
440mA, 1000V AC/DC 10KA Fast acting fuse for mA input. 11A, 1000V AC/DC 20KA Fast acting fuse for A input
Max. Measuring T
ime : 3 minutes for “A” input, 10 minutes for “mA”
input. Rest time 20 minutes min.
AC Conversion Type : Conversion type and additional specication are
same as voltage function.
AC Frequency Response : 50Hz ~ 1KHz
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
600.0V
DC
0.1V
± (1.0% + 3d) ± (0.80% + 3d)
1000V
DC
1V
600.0V
AC
0.1V
1000V
AC
1V
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
60.00mA
DC
0.01mA ± (1.0% + 3d) ± (0.8% + 3d)
600.0mA
DC
0.1mA ± (1.0% + 3d) ± (0.8% + 3d)
6.000A
DC
0.001A ± (1.0% + 3d) ± (0.8% + 3d)
10.00A
DC
0.01A ± (1.0% + 3d) ± (0.8% + 3d)
60.00mA
AC
0.01mA ± (1.5% + 3d) ± (1.2% + 3d)
600.0mA
AC
0.1mA ± (1.5% + 3d) ± (1.2% + 3d)
6.000A
AC
0.001A ± (1.5% + 3d) ± (1.2% + 3d)
10.00A
AC
0.01A ± (1.5% + 3d) ± (1.2% + 3d)
IDM 98III / 99III English IDM 98III / 99III English
20
21
(4) Peak HOLD (For IDM99-III only)
Specied accuracy ±150d.
(5) Resistance
Input Protection : 1000VDC or 1000VAC rms Max. Open Circuit Voltage :
Approx. 2.5V for 600Ω & 6kΩ Range.
Approx. 0.6V for others.
Max. Test Current
: Approximately 0.1mA.
* There is a little rolling less than ±50 digits when measuring 10.00MΩ.
(6) Continuity Check
Input Protection : 1000VDC or 1000VAC rms Max. Open Circuit Voltage : Approx. 2.5V Max. Test Current : Approx. 0.1mA Continuity Threshold : 30Ω Beep On.
100Ω Beep OFF.
Continuity Indicator : 2.7kHz Tone Buzzer
IDM98-III/IDM99-III
Range Resolution Accuracy
600.0Ω 0.1Ω ± (0.8% + 5d)
6.000KΩ 0.001KΩ
± (0.8% + 2d)
60.00KΩ 0.01KΩ
600.0KΩ 0.1KΩ
6.000MΩ 0.001MΩ
40.00MΩ* 0.01MΩ ± (1.0% + 5d)
IDM98-III/IDM99-III
Range Resolution Accuracy
600.0Ω 0.1Ω ± (0.8% + 5d)
(7) Diode Test
Input Protection : 1000V
DC
or 1000VAC rms
Max. Open Circuit Voltage : Approximately. 2.5V Max.Test Current : Approximately 0.4mA
(8) Capacitance
Input Protection : 1000V
DC
or 1000VAC rms
Max. Measuring Time : 0.7 sec for 1nF~1mF
3 sec for 1mF~10mF
(9) Frequency Counter
Input Protection : 1000V
DC
or 1000VAC rms
Min. Frequency : 1Hz Sensitivity :
> 5.0Vp-p (for ACV 1Hz ~ 10kHz).
> 10Vp-p (for ACV 10kHz ~ 100kHz).
2mAp-p (for AC mA). 0.2Ap-p (for ACA).
IDM98-III/IDM99-III
Range Resolution Accuracy
2.000V 1mV ± (1.5% + 2d)
IDM98-III/IDM99-III
Range Resolution Accuracy
1.000µF 0.001µF
± (1.2% + 2d)
10.00µF 0.01µF
100.0µF 0.1µF
1.000mF 0.001mF
10.00mF 0.01mF
IDM98-III/IDM99-III
Range Resolution Accuracy
100.00Hz 0.01Hz
± (0.1% + 2d)
1000.0Hz 0.1Hz
10.000KHz 0.001KHz
100.00KHz 0.01KHz
IDM 98III / 99III English IDM 98III / 99III English
22
23
(10) Temperature (For IDM99-III only)
Input Protection : 1000V
DC
or 1000VAC rms
* Does not include accuracy of the thermocouple probe.
* Accuracy specication assumes surrounding temperature stable to
± 1°C. For surrounding temperature changes of ± 5°C, rated accuracy applies after 2 hours.
(11) AC+DC Voltage
Note: When measuring AC+DC, the AC signal may contain a DC level,
the AC+DC True RMS value might not be accurate if the peak voltage of AC+DC value is over full scale.
Additional specication are same as voltage function.
(12) AC+DC mV
Additional specication are same as voltage function.
IDM99-III
Range Resolution Accuracy
-40.0°C ~ 400°C 0.1°C ± (1.0% + 10d)
-40.0°F ~ 752°F 0.1°F ± (1.0% + 18d)
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
6.000V 0.001V
± (2.5% + 5d) ± (2% + 5d)
60.00V 0.01V
600.0V 0.1V
1000V 1V
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
60.00mV 0.01mV – ± (2% + 10d)
600.0mV 0.1mV ± (2.5% + 5d)
(13) AC+DC mA
Additional specication are same as voltage function.
(14) AC+DC A
Additional specication are same as voltage function.
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
60.00mA 0.01mA ± (2.5% + 5d) ± (2% + 10d)
600.0mA 0.1mA
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
6.000A 0.001A ± (2.5% + 5d) ± (2% + 5d)
10.00A 0.01A