Iso-Tech IDM19 Operating Instructions Manual

Page 1
OPERATING INSTRUCTIONS
Overload Protection
Ω
Open circuit volts
Accuracy
10mV
100mV 1V Resolution
Ω
34V 340V 600V Range
Ω
POCKET
DIGITAL MULTIMETER
DC AC
0 2010 30
RANGE HOLD
Ω
DC/AC
Ω
VOLT
OFF
AUTO POWER OFF
600V
MAX
600V
SAFETY INFORMATION
The following safety information must be observed to insure maximum personal safety during the operation at this meter:
Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly.
Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating material.
Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Small amounts of current can be dangerous.
Use caution when working above 60V dc or 30V ac rms. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger guards on the probes.
Measuring voltage which exceeds the limits of the multimeter may damage the meter and expose the operator to a shock hazard. Always recognize the meter voltage limits as stated on the front of the meter.
SPECIFICATIONS
Display: 3½digit liquid crystal display (LCD) with
a maximum reading of 3400, with a 34-segment analog bar-graph. Polarity: Automatic, (-) negative polarity indication.
Overrange: "OL" mark indication. Low battery indication: The " " is displayed
when the battery voltage drops below the operating level.
Measurement rate: 2 times per second, nominal. Bar graph: 20 times per second. Auto power off: approx. 8 minutes. Operating environment: 0°C to 40°C (32°F to
144°F) at <70% R.H. Storage temperature: -20°C to 60°C (-4°F to 140°F) at < 80% R.H. with battery removed from meter. Temperature Coefficient: 0.1 x (specified accuracy) per °C. (0°C to 18°C, 28°C to 40°C). Power: Two 1.5 button-type batteries (IEC # LR-44, NEDA # 1166A).
Battery life: 70 hours continuous operation. Dimensions: 125mm (H) x 60mm (W) x 24mm
(D). Weight: Approx. 3.9 oz. (110g) including batteries and case.
(Accuracy at 23°C ± 5°C, <70% R.H.)
Overload Protection
100MΩ
Input Impedance
ACCURACY
± 2.0% rdg + 2 dgts
100µV
Resolution
Range
340mV
10M
± 1.0% rdg + 2 dgts
1mV
3.4V
600V DC or AC rms
10MΩ
9.1MΩ
± 2.0% rdg + 2 dgts
10mV
100mV
34V
340V
DCV
9.1MΩ
± 4.0% rdg + 5 dgts
1V
1mV
3.4V
600V
ACV
(50/60Hz)
-1.2Vdc
± 2.0% rdg + 5 dgts
100m
340
450V DC or AC rms
± 2.0% rdg + 4 dgts
1Ω
10Ω
3.4kΩ
34kΩ
OHM
450V DC or AC rms
450V DC or AC rms
-0.45Vdc Test current 1.0mA
Response time: 500ms
± 3.0% rdg + 4 dgts
± 5.0% rdg + 5 dgts
± 3.0% rdg + 3 dgts
Audible indication: < 35Ω
1kΩ
1mV
100Ω
10kΩ
100mΩ
2V
340kΩ
340Ω
34MΩ
3.4MΩ
Diode Test
Continuity
Page 2
OPERATION
Before attempting to replace the battery,
first disconnect the Test Leads from any
Before taking any measurements, read the Safety Information Section. Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements.
Autoranging
The meter defaults to autorange when you turn it on. In autorange, the meter selects the range automatically.
Manually Selecting a Range
The meter also has a manual range mode. In manual range, you select and lock the meter in a range. To manually select a range: Press [RANGE] button to hold the selected range. Subsequently pressing the [RANGE] button will select each range in sequence from the lowest to highest range. Hold the button for 2 seconds to return to the Autorange Mode.
Data Hold Feature
Press [HOLD] button to toggle in and out of the Data Hold mode. In the data hold mode, the " H" annunciator is displayed and the last reading is held on the display. Press [HOLD] button again to release the hold and current readings are once again displayed.
Mode Switch (DC/AC), (Ω / / )
Press this switch to toggle between DC and AC in the voltage measurements. Press this switch to toggle between the continuity/diode and ohms modes, if the function switch is set to Ω / / position.
Measuring Voltage
1. Set the Function Switch to " VOLT" position.
2. Press Mode switch to toggle between "DC" and "AC". The "DC" to "AC" mark is displayed.
3. Touch the probes to the test points. The range will change automatically to the level that will display the input voltage with best resolution.
4. The value indicated in the display window is the measured value of voltage with proper decimal point and annunciator indication.
Measuring Resistance and Testing Continuity
1. Set the function switch to Ω / / position.
2. Turn off power to the circuit under test. External voltage across the components causes invalid readings.
3. To toggle between the ohms/continuity/diode modes, press Mode Switch.
4. Touch the probes to the test points. In ohms, the value indicated in the display is the measured value of resistance with proper decimal point and annunciator indication. In continuity test, the beeper sounds continuously, if the resistance is less than approximately 35Ω.
Testing Diodes
1. Set the Function switch to Ω / / position.
2. Turn off power to the circuit under test. External voltage across the components causes invalid readings.
3. Press Mode Switch to toggle between the ohms/continuity/diode modes
4. Touch probes to the diode. A forward -voltage drop is about 0.6V (typical for a silicon diode).
5. Reverse probes. If the diode is good, "OL" is displayed. If the diode is shorted, a value near 0mV will be displayed.
6. If the diode is open, "OL" is displayed in both directions.
7. Audible Indication: Less than 0.25V.
Auto Power off
1. Auto power off: approx. 8 minutes.
2. After auto power off, press any button to restart the meter, and the reading of measurement will be maintained in the display.
Cancellation Of Auto Power Off Feature: Press and hold the (RANGE) button while rotating function switch
from off to any position to turn the meter on. The auto power off feature is disabled.
Safety: Conforms to IEC1010-1 (EN61010-1), Rev-2; CATII 600V, CATIII 300V.
EMC: Conforms to EN61326.
The symbols used on this instrument are:
Caution, refer to accompanying documents
Equipment protected throughout by Double
insulation (Class II) Alternating current Direct current Ground
BATTERY REPLACEMENT
Power is supplied by two button-type batteries (NEDA 1166A, IEC LR -44). The " " appears on the LCD display when replacement is needed.
WARNING
energized circuit.
1. Disconnect the test leads from any energized circuit.
2. Set the Function Switch to OFF.
3. Remove battery cover screw.
4. Slide off battery cover and change batteries.
5. Replace battery cover and screw.
SCREW
Loading...