Iso-Tech ICM 134 Instruction Manual

Instruction Manual ICM 134 Clamp Meter
EN DE ES IT FR
EN - 2
Read and understand this Instruction Manual completely before using this instrument. Failure to observe the warnings and cautions in this Instruction Manual may result in injury or death, or damage to the instrument and other equipment or property. If this instrument is used in a manner not specified in these instructions, the protec­tion provided by the instrument may be impaired.
WARNING
- Examine the instrument and probes and
leads before use. Do not use the instru­ ment if it is wet or damaged, or if you suspect it is not operating correctly.
- When using the instrument, test leads or probes, keep your fingers behind the finger guards.
- Remove the test lead from the instrument before opening the battery cover or instru­ ment case.
- Always use the correct terminals, switch position and range for measurements.
- Verify the instrument is operating correctly by measuring a known voltage before use. If in doubt, have the instrument serviced.
- Do not apply more than the rated voltage as marked on the instrument between terminals, or between any terminal and earth ground.
- Use caution when measuring voltages above 30 Vac rms or 60 Vdc. These voltages pose a shock hazard.
- To avoid incorrect readings that can lead to electric shock, replace the battery as
soon as low battery indicator appears in the display.
- Disconnect the circuit power and discharge all high-voltage capacitors before making resistance, continuity, or diode measurements.
- Do not use the instrument in a Hazardous Area or around explosive gasses or vapours.
EN - 3
- Wear suitable Personal Protective Equipment when working around or near hazardous Live conductors which could be accessible.
- Do not use the thermocouple to measure the temperature of Hazardous Live conductors or equipment.
Caution
- Disconnect the test leads from the test
points before changing the position of the function rotary switch.
- Never connect the instrument to a source of voltage with the function rotary switch in /
</ A position.
- If possible, do not work alone, so
assistance can be given if required.
- Do not expose the instrument to extremes of temperature or high humidity.
- If this instrument is used in the vicinity of equipment which generates electro magnetic interference, the display may become unstable or the measurements may be subject to large errors.
The following symbols may appear on the instrument and in this Instruction Manual:
Risk of electric shock
Refer to Instruction Manual

Direct Current (dc)
Alternating Current (ac)
Battery
Earth
Equipment protected throughout by double or reinforced insulation
Conforms to EU directives
Dispose of this equipment in accordance with local regulations.
F
Application around and removal from hazardous Live conductors is permitted.
EN - 4
Maintenance
Do not attempt to repair this Instrument. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel. This instrument should be calibrated yearly, or more frequently if used in harsh conditions or if it is suspected of being inaccurate. For calibration and repair contact RS Components - the address is given at the end of these instructions.
Cleaning
Periodically wipe the case with a damp cloth and detergent. Do not use abrasives or solvents.
EN - 5
To conserve battery power, the Auto Power
-off function will turn the instrument off approximately 10 minutes after the last operation.
10min
To disable the Auto Power-off function:
Turn the instrument off, press and hold the “Backlight” button and turn the rotary switch to the required function. The auto Power-off is disabled until the next time the instru­ment is turned off and on again.
Power On / Off
Auto Power Off
EN - 6
ACV / DCV
Resistance
Loading...
+ 11 hidden pages