Isotech Gold, Platinum Specifications

37
Isotech North America - www.isotechna.com - ph: +802-863-8050 - email: sales@isotechna.com - SOLUTIONS FOR PRIMARY & SECONDARY LABORATORIES
|
0 to 1000°C
g
g
Best Homogeneity
g
Economic Attention to HTSPRTs
Since 1995 Isotech have been producing various designs of special Pt/Au, Pt/Pd, Pd/Au thermocouples for researchers. From our experience we can now offer the most popular of these, the Pt/Au thermocouple in a standard form.
All wires are 99.999+% pure and are fully annealed according to the recommendations of McLaren. Assembly also follows his prescriptions which have never been bettered.
After final assembly and annealing the Pt/Au thermocouples will conform to IEC 26460, Edition 1 2008-07.
For the smallest uncertainties we calibrate the thermocouple at the Zinc, Aluminium and Silver Fixed Points.
Thermocouple
Platinum/Gold
Economic alternative to HTSPRTs
Construction allows for differential expansion
Accuracy of ±0.05°C over the whole temperature range
We achieve these results because:
1. All materials are selected for their purity and high quality.
2. All parts are pre-aged and annealed prior to construction.
3. The construction allows for differential expansion of the Gold and the Platinum by having a coil of platinum bridge the two thermo elements at their measuring junction.
4. There are no joins between the measuring and reference junctions.
5. The reference junction is also researched and we use thermally pure copper wire of selected diameter which has been pre-annealed in inert gas to maintain the accuracy of the measuring junction.
6. The reference junction needs to be placed in an accurate reference system such as a Water Triple Point Cell or an Isotech Zeref.
7. An article describing in detail the construction, handling and operation of the thermocouple is provided free with each unit.
Temperature Range 0°C to 1000°C
Sheath materials Measuring Junction Quartz Reference Junction Stainless Steel
Thermo-element Purities Platinum 99.999% Pure Gold 99.999% Pure
Calibration Zinc 0.05°C Uncertainties Aluminium 0.05°C Silver 0.05°C
0-1000°C 0.10°C Including uncertainty of interpolation / extraction
Dimensions Refer to drawing
Carrying Case Included as standard
How to order
Model type: Pt/Au Thermocouple
Including emf vs. temperature traceable calibration certificate and carrying case.
UKAS Calibration optional.
Loading...