Isotech Europa-6, Venus 2140, Calisto 2250 Data sheet

industrial calibration/69
Isocal-6 Series
Europa-6, Venus 2140, Calisto 2250
The Isocal-6® series offers an exceptionally wide operating range in an easy to use portable package. As discussed previously the Isocal-6 series can be used as a Dry Block, a Liquid Bath, a Blackbody Source for infrared thermometers, a Surface Sensor Calibrator and for performance to a few mK (0.001°C) ITS-90 Fixed Points. Isocal-6 calibrators are available in two models, BASIC and SITE. The BASIC model includes a sophisticated temperature controller with a dual display to show the desired Set Temperature and the actual Dry Block Temperature. The SITE model additionally includes a digital indicator to which an external temperature sensor can be connected providing an accuracy of ±0.1°C.
For Dry Block use the unit under test may be compared to the controller value, Isotech recommends the use of an external reference probe connected to the indicator of the SITE model which compensates for the temperature gradient and loading errors. This gives greater accuracy and traceability meeting the requirements of ISO 9000 and other quality standards.
For the best measurement results with liquid bath, surface sensor and blackbody configurations, an external standard probe should always be used, either with the SITE models’ indicator or a laboratory performance temperature indicator such as one of the Isotech True Temperature Indicators (TTI).
All Isocal-6 calibrators include as standard: Cal Notepad Windows Software, Serial Computer Interface, Ramp-to-Set-Point Feature. Increased resolution of ±0.01 available throughout the range via the PC interface and from -19.99 to +99.99 locally on the auto-ranging front display. The controller features multi-point block to display correction giving excellent absolute accuracy.
The SITE model indicator has a universal sensor input allowing Platinum Resistance Thermometers, Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including 4-20mA current transmitters to be displayed. The indicator can be programmed with up to five calibration points to provide high accuracy digital probe matching. The indicator and controller are both addressable over the communications link.
140°C
to
65°C
below
ambient
EUROPA
140°C
to
55°C
below
ambient
VENUS
New! The SITE model can now be used with the supplied Cal Notepad software to test thermostats.
Key Features
• Three Temperature Ranges to Choose From
Europa -45°C to 140°C* Venus -30°C to 140°C* Calisto +30°C to 250°C
* Based on an ambient of 20°C
• 6 Calibration Functions in One Portable Design
• Calibrate as a Dry or Liquid Bath
Thermocouples, platinum resistance thermometers, thermistors, liquid in glass thermometers, or thermostats.
• Simple to Use — Outstanding value for the money!
• SITE Model Includes Universal Input Temperature
Indicator allowing for up to five “correction” points
to be programmed
• Windows Software and PC Interface as Standard
• Free Evaluation Report
Ask for full data or visit www.isotechna.com
INDUSTRIAL
250°C
to 30°C
CALISTO
To see the ISOCAL-6 in use ask for our 27-minute video supplied on CD-Rom
industrial calibration/69
industrial calibration/71
Isocal-6, continued
Calibration and Uncertainty
A certificate, traceable to NIST, is included as standard.
The accuracy will depend very much on the mode of use and the types of sensor to be used. Please contact Isotech for tutorials and uncertainty calculations and comprehensive evaluation reports. The Isocal-6 meets the Calibration Capacity requirements of EA I0/13, “EA Guidelines on the Calibration of Temperature Block Calibrators”.
SPECIFICATIONS
Temperature Ranges Europa: -45°C to +140°C In an Ambient Venus: -30°C to +140°C of 20°C Calisto: +30°C to +250°C
Dry Block Accuracy 0.3°C Throughout entire temperature range
Absolute Stability Dry Block Bath ±0.03°C Over 30 Minutes Stirred Liquid Bath ±0.025°C
Ice/Water Bath ±0.001°C Blackbody Source ±0.3°C Surface Sensor Calibrator ±0.5°C ITS-90 Fixed Point Apparatus ±0.0005°C
Computer Interface Included with Windows Software
INDUSTRIAL
INDUSTRIAL
Cooling Time Europa: 140°C to 0°C in 15 min Venus: 140°C to 0°C in 15 min Calisto: 250°C to Ambient in 25 min
Heating Time Europa: -30°C to 140°C in 15 min Venus: -30°C to 140°C in 15 min Calisto: Ambient to 250°C in 25 min
Calibration Volume 35mm diameter by 160mm deep
Standard Insert Standard Metal Insert 1/8”; 3/16”; 1/4”;
3/8”; and 1/2” diameter holes, all 6.18” deep
Uniformity ±0.018°C
Display Resolution 0.01 -19.99 to 99.99
0.1 -55.0 to -20.0 and 100.0 to 140.0 PC can display 0.01 across whole range with the software included
Indicator Units °C, °F, K
Power Europa: 300 Watts Venus: 150 Watts Calisto: 300 Watts
Overall Dimensions 302mm x 176mm x 262mm (HxWxD)
Weight Europa: 30.9 lbs (14kg) Venus: 22.5 lbs (10.2kg) Calisto: 17.6 lbs (8kg)
ACCESSORIES
Metal Block Insert 951-06-INA: Standard insert type B
Type B
Alternative Metal 951-02-15: 6 pockets, 9.5mm, 8.0mm,
Block Inserts 6.4mm, 6.4mm, 4.5mm,
4.5mm, all 157mm deep
951-02-15A: Blank Insert 951-06-08: Special Insert type C 8mm,
6 x 6.5mm diameter holes,
Type A
all 157 deep
Stirred Liquid Bath 951-06-01: Includes a container, Water/Ice Bath magnetic stirrer and probe guide
Thermometer 951-06-03: Allows three thermometers Support Kit to be suspended in the bath, including liquid in glass types
Oil 951-06-06: 0.1L (20°C to 200°C) 953-04-01: 0.1L (+150°C to +250°C) (Calisto Only)
Blackbody Target 951-06-04: Use with Standard Probe
Surface Sensor Kit 951-06-02: Includes an Insert and an
Fixed Point Cells ITL-M-7724-M: Mercury Slim Cell
Standard Probe 935-14-82: Platinum Resistance
Carrying Case 931-22-64: Sturdy case accommodates
NIST Calibration NIST Calibration provided
free with purchase
Type C
(935-14-82)
angled PRT
D8: Water Slim Cell ITL-M-17401-M: Gallium Slim Cell ITL-M-17156-M: Indium Slim Cell
Thermometer
the unit with room for accessories
HOW TO ORDER
EUROPA-6
EUROPA-6-SITE
VENUS 2140
VENUS 2140-SITE
CALISTO 2250
CALISTO 2250-SITE
Please specify accessories required
70/industrial calibration
70/industrial calibration
Introduction to the Isocal-6 Calibration System
What is Isocal-6?
The Isocal-6 is a flexible range of calibration systems that can calibrate all types of temperature sensors and share common features.
What is special about the Isocal-6 range?
Previously temperature calibration baths were specified by type, for example, Dry Block Bath, Liquid Bath, Blackbody, Fixed-Point Apparatus, etc.
Each different calibrator type has unique advantages over another, for example, Portable Stirred Liquid Baths have a large well which is filled with liquid. The temperature sensors are placed directly into the liquid avoiding the need for specially drilled blocks. They are ideal for awkward shaped and short sensors. Angled probes will not readily fit into a metal block but can be placed into the liquid.
Accuracies are in general better than dry blocks due to lack of air gaps and the temperature uniformity of the stirred liquid.
If a laboratory standard temperature indicator is added much greater accuracies than those from Dry Blocks alone can be achieved, due to better axial and radial uniformity and with suitable reference thermometers performance of up to 0.005°C is achievable.
However dryblocks typically heat and cool much faster, are less messy and more portable. For more information on Dryblocks refer to our dedicated section on page 73.
With the Isocal-6 different inserts are added to the bath to allow it to be used in 6 different modes providing a solution for all temperature sensor calibration requirements.
INDUSTRIAL
What are these different modes?
The Isocal-6 includes a removable metal insert as standard for use as a Dry Block Calibrator, with optional accessories the Isocal-6 can perform as: A Stirred Liquid Bath A Stirred Ice Bath 0°C Reference (Cooling models only) A Blackbody Source for IR thermometers A Calibrator for Surface Sensors An ITS-90 Fixed Point Apparatus
What if only one option is required?
The advantage of the Isocal-6 is that you need only purchase the main apparatus initially, and add other accessories at any time, to increase your accuracy, stability and overall capability.
What temperature range is covered?
Different models cover the temperature range from -45°C to 250°C, above this maximum temperature it is not practical to use a stirred liquid bath. For higher temperatures see the other Dry Block Calibrators that go to 1200°C, many with accessories for Surface Sensor and Blackbody use.
The Isocal-6 range is divided into two groups. The first group, including the Europa, Venus and Calisto models are supplied standard in the Dry Block configuration and have a large 35mm x 160mm calibration volume. The second group includes the Hyperion and Drago which are supplied standard in the liquid bath configuration and have a larger calibration volume of 65mm x 190mm for ultimate performance and versatility. Each group has all the Isocal-6 accessories available to it.
What performance can be achieved?
This varies on the way it is used, for an ITS-90 fixed point apparatus uncertainties of better than 0.001°C can be achieved — see the uncertainty graphs for other modes of operation.
Local support?
Isotech has an extensive network of authorized representatives throughout North America who are skilled in temperature metrology and eager to provide both before and after sales support.
Made to last?
All Isotech products are engineered for long life. If you have ever tried to get inside the dashboard of a motor car you will appreciate that cheapness goes with inaccessibility for repair and maintenance. Isotech’s products are designed for long life, easy repair, accessibility and maintenance.
The ISOCAL-6 is currently subject to a patent application (reference 9900158.8).
Metal Block
Bath
68/industrial calibration
Stirred Liquid
Bath
Stirred Ice/Water
Bath
Blackbody
Source
Surface Sensor
Calibrator
ITS-90 Fixed-Point
Apparatus