Invacare Ocean VIP Ergo, Ocean Dual VIP Ergo User guide

Aquatec®OceanVIPErgo Aquatec®OceanDualVIPErgo
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor futurereference.
enMobileShower
UserManual
-andToiletChair
Contents
1General.........................................4
1.1Introduction..................................4
1.2SymbolsinthisDocument........................4
1.3WarrantyInformation...........................5
1.4Compliance...................................5
1.4.1Product-specicStandards.....................5
1.5ServiceLife...................................5
2Safety..........................................6
2.1SafetyInformation..............................6
2.2LabelsandSymbolsontheProduct.................7
3ProductOverview.................................10
3.1IntendedUse.................................10
3.2MainPartsoftheProduct........................10
4Setup...........................................12
4.1SafetyInformation..............................12
4.2AssemblingtheShower-andT oiletChair.............12
4.2.1AdjustingtheSeatHeight......................12
4.2.2InstallingtheBackrest........................13
4.2.3InstallingtheActivationHandle.................14
4.2.4InstallingandRemovingtheSeatPlate............15
4.2.5SettingArmrestHeight........................15
4.2.6InstallingtheFootrests........................16
4.2.7InstallingtheHeadrest........................17
4.2.8InstallingtheRearAntitippers..................18
4.2.9Installing/RemovingtheBrakeCovers/Clamping
ProtectionCovers...........................19
4.3AssemblingAccessoriesandOptions................20
4.3.1Installing/RemovingtheToiletPanHolderandToilet
Pan.....................................20
5Usage..........................................21
5.1SafetyInformation..............................21
5.2TurningArmrests...............................22
5.3SwivellingtheFootrests..........................22
5.4TiltingtheSeatFramewithBackrest................23
5.5TiltingtheBackrest(OceanDualVIPErgoOnly)........24
5.6Engaging/DisengagingtheCastorLocks..............25
6Transport........................................26
6.1SafetyInformation..............................26
7Maintenance.....................................27
7.1MaintenanceandInspection......................27
7.2CleaningandDisinfection........................27
7.2.1GeneralSafetyInformation.....................27
7.2.2CleaningIntervals............................28
7.2.3Cleaning..................................28
7.2.4DisinfectionInstructions.......................29
8AfterUse........................................30
8.1Storage......................................30
8.2Reconditioning................................30
8.3Disposal.....................................30
9Troubleshooting...................................31
9.1IdentifyingandRepairingFaults....................31
10T echnicalData...................................32
10.1DimensionsandWeight.........................32
10.2Materials....................................34
10.3EnvironmentalParameters.......................34
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

1General

1.1Introduction

Thisusermanualcontainsimportantinformationaboutthe handlingoftheproduct.Toensuresafetywhenusingthe product,readtheusermanualcarefullyandfollowthe safetyinstructions.
Onlyusethisproductifyouhavereadandunderstood thismanual.Seekadditionaladvicefromahealthcare professionalwhoisfamiliarwithyourmedicalconditionand clarifyanyquestionsregardingthecorrectuseandnecessary adjustmentwiththehealthcareprofessional.
Notethattheremaybesectionsinthisdocument,whichare notrelevanttoyourproduct,sincethisdocumentapplies toallavailablemodels(onthedateofprinting).Ifnot otherwisestated,eachsectioninthisdocumentreferstoall modelsoftheproduct.
Themodelsandcongurationsavailableinyourcountrycan befoundinthecountry-specicsalesdocuments.
Invacarereservestherighttoalterproductspecications withoutfurthernotice.
Beforereadingthisdocument,makesureyouhavethe latestversion.YoundthelatestversionasaPDFonthe Invacarewebsite.
Formoreinformationabouttheproduct,forexample productsafetynoticesandproductrecalls,contactyour Invacaredistributor .Seeaddressesattheendofthis document.
Incaseofaseriousincidentwiththeproduct,youshould informthemanufacturerandthecompetentauthorityin yourcountry .

1.2SymbolsinthisDocument

Symbolsandsignalwordsareusedinthisdocumentand applytohazardsorunsafepracticeswhichcouldresultin personalinjuryorpropertydamage.Seetheinformation belowfordenitionsofthesignalwords.
WARNING Indicatesahazardoussituationthatcouldresult inseriousinjuryordeathifitisnotavoided.
CAUTION Indicatesahazardoussituationthatcouldresult inminororslightinjuryifitisnotavoided.
NOTICE Indicatesahazardoussituationthatcouldresult indamagetopropertyifitisnotavoided.
TipsandRecommendations Givesusefultips,recommendationsand informationforefcient,trouble-freeuse.
Ifyoundthatthefontsizeintheprinteddocumentis difculttoread,youcandownloadthePDFversionfromthe website.ThePDFcanthenbescaledonscreentoafont sizethatismorecomfortableforyou.
4
OtherSymbols
(Notapplicableforallmanuals)
1652073-A
General
UKR P
UKResponsiblePerson Indicatesifaproductisnotmanufacturedin theUK.
Triman Indicatesrecyclingandsortingrules(only relevantforFrance).

1.3WarrantyInformation

Weprovideamanufacturer’swarrantyfortheproduct inaccordancewithourGeneralTermsandConditionsof Businessintherespectivecountries.
Warrantyclaimscanonlybemadethroughtheprovider fromwhomtheproductwasobtained.
supportusers)andrelatedstandards.Thisincludestesting concerningammability.Progressivesmoulderingignition andamingignitiondonotoccur .
Forfurtherinformationaboutlocalstandardsand regulations,contactyourlocalInvacarerepresentative.See addressesattheendofthisdocument.

1.5ServiceLife

Theexpectedservicelifeofthisproductisveyearswhen useddailyandinaccordancewiththesafetyinstructions, maintenanceintervalsandcorrectuse,statedinthismanual. Theeffectiveservicelifecanvaryaccordingtofrequency andintensityofuse.

1.4Compliance

Qualityisfundamentaltothecompany’soperation,working withinthedisciplinesofISO13485.
ThisproductfeaturestheCEmark,incompliancewiththe MedicalDeviceRegulation2017/745ClassI.
ThisproductfeaturestheUKCAmark,incompliancewith PartIIUKMDR2002(asamended)ClassI.
Wearecontinuouslyworkingtowardsensuringthatthe company’simpactontheenvironment,locallyandglobally, isreducedtoaminimum.
WeonlyuseREACHcompliantmaterialsandcomponents.
1.4.1Product-specicStandards
Theproducthasbeentestedandconformstothestandard ISO17966(Assistiveproductsforpersonalhygienethat
1652073-A5
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

2Safety

2.1SafetyInformation

WARNING! RiskofSeriousInjuryorDamage
Improperuseofthisproductmaycauseinjury ordamage.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professionalorproviderbeforeattemptingto usethisequipment.
–Donotusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanual,servicemanualorinstruction sheetsuppliedwiththisproductoroptional equipment.
WARNING! RiskofInjury
–Beforeeveryuse,checktheproductfordamage
andensurethatitissecurelyassembled.
–Beforerstuse,cleantheproductthoroughly.
WARNING! RiskofInjury
Improperuseofthisproductmaycauseinjury ordamage.
–Donotuseifdefective. –Intheeventofamalfunction,contactyour
providerimmediately.
–Donotmakeanyunauthorisedchangesor
alterationstotheproduct.
–Onlyuseaccessories/optionsintendedfor
thisproduct.
WARNING! RiskofInjury
Improperuseofthisproductmaycauseinjury ordamage.
–Donotlifttheproductwhenunderload. –Donotstandontheproduct.
CAUTION! RiskofInjury
Non-originalorwrongpartsmayaffectthe functionandsafetyoftheproduct.
–Onlyuseoriginalpartsfortheproductinuse. –Duetoregionaldifferences,refertoyourlocal
Invacarecatalogueorwebsiteforavailable optionsorcontactyourInvacaredistributor . Seeaddressesattheendofthisdocument.
61652073-A
NOTICE!
A
B C
EF
G
H
I
J
S
N
L
D
(XX) XXXXXXXXXXXXXX( XX) XXXXXXXXX
X
XXXXXX XXXXXX
Whencombiningthisproductwithanother product,therestrictionsofbothproductsapply forthecombination.E.g.themaximumuser weightofoneproductinthecombinationmay belower .
–OnlyusecombinationsallowedbyInvacare.
ContactyourInvacaredistributorfor information.
–Beforeuse,readtheusermanualofeach
productandchecktherestrictions.Thefull compliancewiththeinformation,instructions andrecommendationscontainedinbothuser manualsisnecessaryforasafeuse.

2.2LabelsandSymbolsontheProduct

Safety
IdenticationLabel
Theidenticationlabelcontainsimportantinformation:
1652073-A
Manufactureraddress
A
UKConformityAssessed
B
EuropeanConformity
C
Foruseonlevelgroundonly
D
Readusermanual
E
Maximumuserweight
F
Dateofmanufacture
G
Serialnumber
H
UniqueDeviceIdentier
I
Medicaldevice
J
7
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
!
Referencenumber
K
Productdesignation
L
Theidenticationlabelisattachedtotheinsideof theleftsidepart.
LabelsontheProduct
Thefollowinglabelisattachedtothebacksideofthe headrest:
Label
Description
CAUTION!
Riskofimpact
Yourbodycanbehitbythe backrest/headrestwhenpulling theactivationhandle.
–Makesurethereisenough spacebetweenyouandthe backrest/headrestwhentiltingthe seatframewithbackrest.
Thefollowinglabelisattachedtothehandleofthe backrest:
Label
Description
WARNING!
Riskoftippingover
–Donotsteponthefootrest whengettingintooroutofthe shower-andtoiletchair.
–Pushthefootreststothesides orfoldupthefootplatesbefore gettingintooroutoftheshower­andtoiletchair .
–Beforesittingdownorstanding up,turnthecastorssothatthe lockispointingforwards.
–Donotleantoofarforwards whensittingintheshower-and toiletchair .
81652073-A
Thefollowinglabelsareattachedtotheactivationhandles:
Safety
Label
Description
Activationhandleforseattilting.
Activationhandleonlyforbackresttilting.
1652073-A9
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
L

3ProductOverview

3.2MainPartsoftheProduct

3.1IntendedUse

AquatecOceanVIPErgoandAquatecOceanDualVIPErgo aremobileshower-andtoiletchairswithaseattilt-inspace function,stainlesssteelframe,individuallyadjustableseat height,tensionadjustablebackrest,ipuparmrestsand foldableangle-adjustabletexturedfootplates.Theproductis alsoavailablewithanadditionalbackrestrecliningfunction.
Thesemobileshowerandtoiletchairsareintended tosupportpersonalhygieneactivitieswithorwithout assistanceininstitutionalandhouseholdsurroundings.
Temperatureforuse(indoorsonly):10-65°C
IntendedUsers
Theintendeduseristhepersonsittingintheshower-and toiletchairoranassistant.Thedevicecanbeusedfor adolescentsoradultshavingaphysicalsizeequaltoormore than146cm,aweightequaltoormorethan40kganda bodymassindexequaltoormorethan16.
Forthemaximumuserweightoftheproductsee10 TechnicalData,page32.
Indications
Difcultyinstandingorkeepingbalance,
weaklegsorlowerlimbdisability ,
medicalconditionsthatimpairthebalance.
Therearenocontraindicationsknownwhenusingthe productasintended.
101652073-A
Headsupport
A
OceanVIPErgo:
B
1
Backrestwithbackrestcoverandarmrests
OceanDualVIPErgo:
B
2
Backrestwithbackrestcover ,armrestsand activationhandleforbackresttilting
Screws(4x)andwashers(4x)forbackrestxation
C
Seatframewithseatplateandsidepartswithbase
D
frameandcastors
ProductOverview
Clips(4x)forheightadjustment
E
Footrestswithheelstraps(2x)
F
Clips(2x)forfootrests
G
Activationhandleforseattilting
H
Usermanual
I
Allenkey
J
Rearantitipperwithscrew,washerandmounting
K
cap(2x)(OceanDualVIPErgoonly)
Brakecover(2x)
L
Brakelevercover(2x)
M
Clampingprotectioncover(2x)
N
Pushbar
O
ToiletPanHolderwithToiletPan(optional)
Toiletpanholder
A
Toiletpan
B
Lid
C
1652073-A
11
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

4Setup

4.1SafetyInformation

WARNING!
–Checkthepartsfortransportdamagebefore
commissioningandcontactyourproviderif required.
–Ensureduringassemblythatthepartsare
positionedcorrectlywithrespecttoone another.
–Self-lockingscrews/nutsorthread-locking
adhesiveareusedforseveralconnections.If theseareloosened,theymustbereplacedby newself-lockingscrews/nutsortheymustbe securedusingnewthread-lockingadhesive.
CAUTION! RiskofPinching
–T akecareduringassembly.

4.2AssemblingtheShower-andT oiletChair

Thissectiondescribestheassembly.Disassemblytakesplace inthereverseorder .
Thearmrestheight,footrestheight,footrestangle, headrest,heelstrap,tensionofbackrestcoverand accessories/optionsmustbeadjustedtooptimally ttheuser.
12

4.2.1AdjustingtheSeatHeight

Thissectiondescribestheassembly.Disassemblytakesplace inthereverseorder .
WARNING!
–Donotadjusttheseatheightoftheshower-and
toiletchairwhensomeoneissittingonit.
Theseatheightmustbeadjustedonbothsides simultaneously.
1.Slightlyraisetheshower-andtoiletchairattheside partA.
2.RemovethefourclipsC(twooneachside).
3.Setthecorrectheightbysimultaneouslymovingthe seatframeupwardsordownwardsonthefourtubes Bofthebaseframe.
4.Bringtheclipmountsonthesidepartinlinewitha holeinthebaseframe.
5.Insertthefourclipsfromtheoutsideintothesidepart, untiltheclipedgesenclosethebaseframetubeB.
1652073-A
Setup
WARNING! RiskofInjury
–Makesurethatallfourclipsaresecurely
fastened.
–Makesurethatallfourtubesaresettothe
sameheight.
Theseatheightcanbeadjustedtoaccommodate differentsituationsofusee.g.transferfromabed orawheelchairoruseabovetoiletetc.

4.2.2InstallingtheBackrest

1.Pushdownallfourfootpedalsofthecastorstolock them.
2.PushthebackrestDfrombehindontotheseatframe E,untilthecollarCtouchestheseatframetubeF.
NOTICE!
–FirmlytightenthescrewsusingtheAllenkey
supplied.
–Checkthescrewsregularlytoensuretheyare
fastenedsecurely.
3.PositionawasherBoneachscrewAandsecurethe backrestoneachside.
4.Tightenthescrewswithatorqueof5Nm.
AdjustingtheBackrestCover
Thetensionofthebackrestcanbeadjustedasneeded.
WARNING!
–Makesurethatatleastthreebeltsarealways
passedaroundthebackrestframe(onthe outside)andtheirbeltbucklesarefastened.
Tightening:
1.PullthebeltsAofthebackrestcovertighter.
1652073-A13
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
Loosening:
1.OpenthebucklesBbysqueezingtheretaininglugs together(arrows).
2.Pushthebeltsbackalittlethroughthebuckle.
3.Fastenthebeltbuckles.
4.Retightenthebeltsgently.
WARNING! RiskofInjuryduetoDamagedBackrestBuckles!
Checkforsecurefasteningofthebackrest buckleseachtimetheproductismaintainedor reconditioned.
–Checkallbuckleclipsforsecurefasteningby
closingthebucklesandpullingonbothends.
–Replacethebackrest/backrestbucklesifa
buckleopensundertheappliedforce.

4.2.3InstallingtheActivationHandle

1.RemovetheclampingpartCfromtheactivationhandle AbylooseningthescrewsD.
2.PositiontheactivationhandleAforseattiltingfrom theundersideattherightsideofthehandletubeB.
Thehandleshouldbeslightlyturnedinwardstothe backrest.
3.Installtheactivationhandletothehandletubeusing theclampingpartCandscrewsD.
FortheOceanDualVIPErgo,theactivationhandleE forbackresttiltingisalreadyinstalledtotheleftside ofthehandletube.Thepositionoftheactivation handlecanbechangedifrequired.
14
1652073-A
Setup
C
A
Borem ipsum
E
G
C
D
F

4.2.4InstallingandRemovingtheSeatPlate

Installing
1.Placetheseatplateontheseatframewiththecutout tothefront.
2.Presstheseatplatedownwiththeathandatthe sidesandthenatthefrontandbackuntiltheseatplate snapsontotheseatframe.
Removing
1.Pulltheseatplateupwardsattherear .

4.2.5SettingArmrestHeight

1.Openthebeltbuckleofthebackrestcoverwhichis passedontheinsideofthebackrestframe.
2.Openthebeltbuckleonthebackrestcover,whichis abovethearmrest.
Whenadjustingtoalowerpositionthebelt beneaththearmrestmustbeopened.
3.Passthisbeltontheinsideofthebackrestframeand closeitagain.
4.
Fig.4-1
RemovecapA.
5.FolduparmrestBtopullitoffthearmrestbracketC.
6.
Fig.4-2
1652073-A15
RemovescrewsDfromarmrestbracketCwithsupplied Allenkey .
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
C
A
B orem ipsum
7.RemovesecuringclampEwithnutsFfromtheinside ofthebackrestframe.
Thenutsareinsertedlooselyintothesecuring clampandmaydropout.
8.Toreinstallarmrestbracketandsecuringclampat desiredposition,insertscrewsDintoarmrestbracket andguidethemthroughthedesiredholesinthe backrestframe.
Thesmallerholesontheinsideofthebackrest framexthesecuringclamp.
9.PlacethesecuringclampinsuchwaythatthexingG ofsecuringclampengagesinsmallerhole.
10.TightenscrewsDwith5-6NmandnutsF.
11.
14.Passtheopenbeltofthebackrestcoverontheoutside ofthebackrestframe,closebeltbucklesandtighten allthebelts.
15.Theheightadjustmentprocedureisthesameforthe rightandleftarmrest.
Toinstallaccessories/options,thethirdbeltfrom thetopcanbeopenedandthebackrestcovercan bepassedbythebackrestframefromtheinside.

4.2.6InstallingtheFootrests

Fig.4-3
ReinstallarmrestBonbracketC.
12.Folddownarmrest.
13.ReinstallcapA.
161652073-A
Setup
1.InsertthepinoftheclipGfromtheoutsideintothe requiredholeonthefootrestB.Makesuretheclip engagesaroundthetube.
2.PushthefootrestintotheguideHatthefrontofthe sidepartuntilthepinofclipGengagesinthegroove A.Thefootrestcannolongerbeturnedtotheside.
Theassemblyprocedureisthesamefortheright andtheleftfootrest.
AdjustingtheHeightoftheFootrests
Thefootrestscanbeshortenedifrequired(referto theservicemanual).Shorteningthefootrestsmust beperformedbyaqualiedtechnician.
1.SlightlypullupthefootrestB.
2.PulltheclipGoutoftheholeinthefootrestand re-insertitintotherequiredhole.Pulloutthefootrest slightlyfurtherifnecessary.
3.PushdownthefootrestuntilthegrooveAcatchesthe pinoftheclip.
4.Settheotherfootresttotherequiredheightas described.
AdjustingtheInclinationoftheFootPlates
Theinclinationofthefootplatescanbeadjusted innitelyvariably.
1.LoosentheAllenscrewConthefootplateusingasize 4Allenkey.
2.TurnthefootplateDtotherequiredposition.
3.Re-tightentheAllenscrew.
1652073-A

4.2.7InstallingtheHeadrest

1.UsethestarhandleFtoinsertthescrewalittleinto thebracketG.
2.Pushthebracketwiththescrewontothehandleofthe backrestEfrombehindwiththescrewengaginginthe holeofthebackrest’shandle.
3.InserttheplateDintothebracketfromthefrontso thatbothlugsoftheplatefacedownandupandthe curvedplatesidetouchesthehandleofthebackrest.
4.PushtheheightadjustmenttubeCfromaboveinto therecessesofthebracket.
5.Usethestarhandletotightenthescrewuntiltheheight adjustmenttubeisxedinplace.
6.PushthepadJandthepadholderBintotheholder Hontheheightadjustmenttube.
7.UsethestarhandleAtotightentheupperscrewuntil thepadholderisxedinplace.
17
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
A
B
D
C
F
E
Toadaptittotheuser,theheightadjustmenttube canbettedwiththecurvefacingthefrontorthe back.
TheheadrestpadJcanberemovedandreplaced byotherpadshapes(optional).Toremoveit,loosen theretainingboltsI.
AdjustingtheLengthoftheHeadrest
TheheadresttubesCandBcanbeshortenedif required(refertotheservicemanual).Shortening theheadresttubemustbeperformedbyaqualied technician.

4.2.8InstallingtheRearAntitippers

WARNING! RiskofTipping
ThereisariskofsevereinjuryontheOcean DualVIPErgowhenthebackrestistilted.The completechaircantiltbackwards.
–BeforeusingtheOceanDualVIPErgo,make
surethatbothrearantitippersareinstalled.
1.RemovethecovercapBfromthebaseframeA.
2.AttachthemountingcapDtothebaseframe.
3.InserttheantitippertubeC.
4.Fixtheantitipperthroughitssecondholeusingthe screwEwiththewasherF.
5.Installthesecondantitipperontheotherside.
181652073-A
WARNING! RiskofInjury
–Neverstandontheantitippers.
Setup

4.2.9Installing/RemovingtheBrakeCovers/Clamping ProtectionCovers

A
B
C
Brakecover
Brakelevercover
Clampingprotectioncover
BrakeCover
1.AttachthebrakelevercoverAbypinninguptheslot BtothesmallshaftC.
2.PositionthelowerpartofthebrakecoverDatanangle frombelowoverthebrakehousingE.
3.Foldtheupperpartofthebrakecovertothebrake housingsothattherecessesFsitushwiththeedges G.
4.Installthesecondbrakecoverontheotherside.
ClampingProtectingCover
1.InstalltheclampingprotectingcoverAbysnappingit ontotheframetubesB.MakesurethatallclipsC arefullyengaged.
2.Installthesecondclampingprotectingcoveronthe otherside.
1652073-A19
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

4.3AssemblingAccessoriesandOptions

WARNING! Riskofinjury
–Theassemblyofaccessoriesandoptionsmust
becarriedoutaccordingtothecorresponding instructions.

4.3.1Installing/RemovingtheT oiletPanHolderandT oilet Pan

ThetoiletpanFcanbeinstalled/removedwithor withoutthelidG.
5.Placethelidonorremoveitfromthetoiletpan.
6.HoldthetoiletpanbyitshandleHandinsertitfrom therearsidebetweenthetoiletpanholderandtheseat.
Slightlypushdownthetoiletpanwheninsertingor removing.
Thelidcanbehungontheplastichookatthe undersideofthetoiletpanhandle.
1.RemovetheseatplateBfromtheframeA.
2.PositiontheendsofthetoiletpanholderCintothe holesoftherearcrosstubeE.
3.Engagethefrontpartofthetoiletpanholderwiththe frontcrosstubeD.
4.Reinstalltheseatplatetotheframe.
201652073-A
Usage

5Usage

5.1SafetyInformation

WARNING!
–Onlyusetheshower-andtoiletchaironaat
surfaceandindoors.
–Observethespecicationsontheidentication
label.Donotoverloadtheshower-andtoilet chair.
–Observethesafetylabelsonthebacksideofthe
headrestandonthehandleofthebackrest,2.2 LabelsandSymbolsontheProduct,page7.
–Donotplaceexcessiveweightontheshower-
andtoiletchair .
–Alwaysperformavisualinspectionofthe
productforexternaldamagebeforeeachuse.
WARNING! RiskofSlippingandRolling
–Beforeuse,makesurethattheseatissecurely
fastenedtotheseatframe
–Lockallcastorswithleversusingthefootlevers
beforesittingdownontheshower-andtoilet chairorstandingupfromit,aswellasinthe parkingposition.
WARNING! RiskofTipping
–Swivelthefootreststothesideorfoldupthe
footplatesbeforesittingdownorstandingup.
–Beforesittingdownorstandingup,turnthe
frontcastorssothatthelockispointing
forwards. –Neverstandonthefootrests. –Donotleanexcessivelyforwards. –Onlypushtheseateduserontheshower-and
toiletchairwhentheseatframeandbackrest
aresettoanuprightposition. –Onlytransfertheuserfromortotheshower-
andtoiletchairwhentheseatframeand
backrestaresettoanuprightposition. –Neverremovethepneumaticsprings(Ocean
DualVIPErgoonly).
WARNING! RiskofInjury!
–BeforeusingtheV ariableSoftSeat(optional),
thelateralcovershavetobeinstalled.
WARNING! Riskoftippingbackwards!
–TheOceanDualVIPErgomustonlybeusedif
therearantitippersareinstalled.
1652073-A
21
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
WARNING! RiskofTipping
–Theshower-andtoiletchairmustonlybe
movedbyusingthepushbar .
CAUTION! RiskofburningyourSkin
Theproductcomponentscanheatupwhen exposedtoexternalsourcesofheat.
–Considerheatedupsurfaceswhenusingthe
shower-andtoiletchairinthesaunaorwhen itisexposedtohotwater .
–Donotexposetheproducttodirectsunlight
beforeuse.
CAUTION! RiskofcatchingduetoTiltingMechanism
–Whentiltingtheseatframeorthebackrest
makesurethattheuser'sarmsareplacedon thearmrestsorfoldedinfrontofthebody.
–Makesurethattheclampingprotectioncover
isinstalledtotheframe,see4.2.9Installing/
RemovingtheBrakeCovers/Clamping ProtectionCovers,page19.
NOTICE! RiskofbreakingtheHeadrest
–Onlysteerorpushtheshower-andtoiletchair
usingthehandleonthebackrest,neverwith theheadrest.
NOTICE! RiskofDamagetotheProduct
–Donotlifttheshower-andtoiletchairusing
thefootrests,armrestsoranyaccessories/
options.
Foruserswithhighriskofstandingonthefootrests (e.g.dementia),frontantitippers(optional)are availabletopreventtheshower-andtoiletchair fromtippingforward.

5.2TurningArmrests

CAUTION! RiskofPinchingFingers
–Donotputngersbetweenarmrestand
backrestwhenfoldingbackthearmrests.

5.3SwivellingtheFootrests

22
1652073-A
Usage
EntryPosition/ExitPosition:
1.SlightlypullupthefootrestAintheguideBuntilthe pinoftheclipDnolongerengagesinthegrooveC.
2.Swivelthefootresttotheside.
DrivingPosition:
CAUTION! RiskofAccidentduetolooseFootrest
–Fixfootrestsinthedrivingposition.
1.TurnthefootrestAinwards.
2.Pushthefootrestdownward,untilthepinofclipD engagesinthegrooveC.
Alternatively,thefootplatesGcanbeswivelled upwardstoenter/exittheshower-andtoiletchair .
ReplacingtheHeelStrap
1.OpenthehookandloopstripEontheheelstrap.
2.RemovethescrewFfromtheundersideofthefoot plate.
3.Removetheheelstrapfromtheexibleplasticpin.
Toattachthestrap,carryoutthesameprocedure inthereverseorder .

5.4TiltingtheSeatFramewithBackrest

NOTICE!
–Neverreachintheareabetweenthesideparts
andthetubularframewhentilting.
TiltingBackwards/Forwards
Theseatframecanbetiltedapprox.40°backwards togetherwiththebackrest.
Tiltstoppers(optional)areavailabletopreventthe productfromtiltingforwardtoaloweranglethan0°.
Forthetwolowestseatheights,thecomplete adjustmentrangeofthefootrestcannotbeused. Beforeuse,checkthatthefootrestsdonottouch theoorwhentiltingtheseatforwards.Ifnecessary re-adjusttheseatheightorthefootrestheight accordingly.
Wheninstallingensurethatthehookandloopstrip oftheheelstrapisfacingbackwards.
1652073-A23
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
CAUTION! RiskofCrushing
Yourngersorhandscangetcaughtand squeezedbythetiltingmechanism.
–Beforeeveryuse,makesurethatthesidepart
coversCofthetiltingmechanismarecorrectly mounted.
4.Settherequiredinclinationbypushingdownorpulling upthehandleB.
5.Releasetheactivationhandleandletthepinsofthe tiltingmechanismengageinthenearestholes.Theseat frameissetatthecurrentlysetinclination.
5.5TiltingtheBackrest(OceanDualVIPErgo
Only)
CAUTION! RiskofGettingFingersCaughtBetweenArmrest BracketandSidePart
–Neverreachintotheareabetweenthe
armrestbracketandthesidepartwhentilting backwardstheseatframeorwhenlowering thebackrest.
–Whentiltingtheseatframeorthebackrest
makesurethattheuser’sarmsareplacedon thearmrestsorfoldedinfrontofthebody.
1.Pushdownallfourfootpedalsforthecastorstolock them.
2.HoldtheshowerchairatitspushbarB.
3.PullandholdtheactivationhandleA.
24
1652073-A
1.Pushdownallfourfootpedalsforthecastorstolock them.
2.HoldtheshowerchairatitspushbarB.
3.PullandholdtheactivationhandleA.
Usage
Increasedforceisneededfortiltingbackwards withoutseateduser.Automatictiltingupwards withoutseateduserwhenactivationhandleispulled.

5.6Engaging/DisengagingtheCastorLocks

CAUTION! RiskofTrapping/Squeezing
Allcastorlocksarefoot-operated.
–Donotoperatethecastorlocksbyhand.
1.Tolockthecastorlocksstepontherearendofthe pedalAuntilitengages.
Theactivationhandleisequippedwithalocking device.Foractivatingthehandle,thelockingdevice
Thecastorisnowlockedinpositionandcannot swivelanylonger .
hastobepulledinadvance.
2.Tounlockthecastorlockssteponthefrontendofthe
4.Tiltthebackresttotherequiredposition.
5.Releasetheactivationhandletoxthebackrestinthe currentposition.
1652073-A25
pedaluntilitdisengages.
Castorwithbluepedal(optiondirectionallock) engagestheswivellockforeasierstraightmovement.
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

6Transport

6.1SafetyInformation

Theproductcanbedismantledfortransportintothe partsofthescopeofdelivery,seechapter3.2Main PartsoftheProduct,page10.
1.Disassembletheproduct,seechapter4Setup,page12 byfollowingtheassemblystepsinreverseorder .
NOTICE! RiskofDamagetotheProduct
–Transporttheproductinadryarea. –Donotexposetheproducttodirectsunlight. –Refertothestorageandshippingconditions,
seechapter10.3EnvironmentalParameters, page34.
NOTICE!
–Self-lockingscrews/nutsorthread-locking
adhesiveareusedforseveralconnections.If theseareloosened,theymustbereplacedby newself-lockingscrews/nutsortheymustbe securedusingnewthread-lockingadhesive.
261652073-A
Maintenance

7Maintenance

7.1MaintenanceandInspection

Theproductismaintenance-free,providedthecleaningand safetyinstructionsoutlinedbelowareadheredto.
Toensuregoodperformanceandsafetythefollowingchecks shouldregularlybecarriedout:
Checktheproductandaccessories/optionsregularly fordamage.Donotusetheproductifdefective.
Checktiltingfunctionoftheseatframeandbackrest (OceanDualVIPErgo).Easytiltingandsecurelocking ofthepositionismandatory.
Iftiltingfunctiontoaloweranglethan0°isnot possible,removethetiltstoppers(optional)ifneeded.
Incaseofdamageormalfunctionitisrecommendedto contacttheprovider .
Ensurethattheproductissecurelyassembledincl. allprotectioncoversandthatscrewsareproperly tightened.
WARNING! RiskofInjuryandDamage
Iftheproductisimproperlyadjusteditcancause accidentsandsevereinjuries.
–Theproductmustberepairedbyaqualied
technician.
–Damagedpartshavetobereplacedbya
qualiedtechnician.
–OnlyuseoriginalInvacarespareparts.

7.2CleaningandDisinfection

7.2.1GeneralSafetyInformation

CAUTION! RiskofContamination
–T akeprecautionsforyourselfanduse
appropriateprotectiveequipment.
NOTICE!
Wronguidsormethodscanharmordamage theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand mustprotectthematerialstheyareusedto clean.
–Neverusecorrosiveuids(alkalines,acidetc.)
orabrasivecleaningagents.Werecommend anordinaryhouseholdcleaningagentsuchas dishwashingliquid,ifnotspeciedotherwisein thecleaninginstructions.
–Neveruseasolvent(cellulosethinner,acetone
etc.)thatchangesthestructureoftheplastic ordissolvestheattachedlabels.
–Alwaysmakesurethattheproductis
completelydriedbeforetakingitintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term careenvironments,followyourin-houseprocedures.
1652073-A
27
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

7.2.2CleaningIntervals

NOTICE!
Regularcleaninganddisinfectionenhancesmooth operation,increasestheservicelifeandprevents contamination. Cleananddisinfecttheproduct:
–regularlywhileinuse, –beforeandafteranyserviceprocedure, –whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser .

7.2.3Cleaning

Theproductcanbecleanedusingcommerciallyavailable agents.
CleaninginaDomesticEnvironment
1.Removeanyaccessories/optionstted(onlyaccessories /optionswhichdonotrequiretools).
2.Pullthefootrestsoutoftheguides.
3.Removetheseatplateandthebackrestcover .
4.Removethebrakecoversandclampingprotection covers,seechapter4.2.9Installing/RemovingtheBrake Covers/ClampingProtectionCovers,page19.
5.Wipedowntheindividualpartsusingaclothor brushandcommerciallyavailablecleaningagentsand disinfectants.
6.Rinsethepartswithwarmwater .
7.Drythepartswithacloth.
8.Reassembletheproduct,seechapter4Setup,page12.
CleaninginaClinicalEnvironment
Inadditiontocleaningasdescribedabovewe recommendcleaningtheproductpriortotheuseby anotherpatient,afterendingoftheperiodofuse oronaregularbasis(dependingonwhichoccurs rst)at
90°Cfor1minute
85°Cfor3minutes
80°Cfor10minutes
intheautoclave.
CAUTION! RiskofburningyourSkin
–Allowtheproducttocoolaftercleaninginthe
washingmachineorautoclave.
CleaningtheBackrestCover
1.Removethebackrestcover .
2.Machine-washthecoveratmax.60°Cusingamild detergent.
NOTICE! RiskofDamage
–Washingathighertemperaturescancause
damage.
–Donottumbledry.
NOTICE!
Excessivecleaningcandamagethetextile.
–Werecommendreplacingthebackrestcover
everytwoyearsforhygienereasons.
281652073-A
CleaningtheT oiletPan(optional)
Thetoiletpancanbecleanedinanautoclaveatfollowing temperatures/times:
90°Cfor1minute
85°Cfor3minutes
80°Cfor10minutes.

7.2.4DisinfectionInstructions

InDomesticCare
Method:Followtheapplicationnotesfortheused disinfectantandwipe-disinfectallaccessiblesurfaces.
Disinfectant:Ordinaryhouseholddisinfectant.
Drying:Allowtheproducttoair-dry.
InClinicalandLongTermCare
Followyourin-housedisinfectionproceduresandonlyuse thedisinfectantsandmethodsspeciedtherein.
Maintenance
1652073-A29
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

8AfterUse

8.1Storage

WARNING! RiskofDamagetotheProduct
–Storetheproductinadryarea. –Donotstoretheproductnearheatsources. –Donotexposetheproducttodirectsunlight. –Refertothestorageandshippingconditionsin
chapter10.3EnvironmentalParameters,page
34.
Afterlongtimestorage,ahigherforcemaybe neededtotriggerthegasspring(OceanDualVIP Ergoonly).

8.2Reconditioning

Thisproductissuitableforreuse.Toreconditiontheproduct foranewuser,carryoutthefollowingactions:
Inspection
Cleaninganddisinfection
Fordetailedinformation,see7Maintenance,page27 theservicemanualforthisproduct.

8.3Disposal

Beenvironmentallyresponsibleandrecyclethisproduct throughyourrecyclingfacilityatitsendoflife.
Disassembletheproductanditscomponents,sothedifferent materialscanbeseparatedandrecycledindividually.
Thedisposalandrecyclingofusedproductsandpackaging mustcomplywiththelawsandregulationsforwaste handlingineachcountry.Contactyourlocalwaste managementcompanyforinformation.
and
Makesurethattheusermanualishandedoverwiththe product.
Ifanydamageormalfunctionisdetected,donotreuse theproduct.
301652073-A

9Troubleshooting

9.1IdentifyingandRepairingFaults

Thefollowingtableprovidesinformationonmalfunctionsandtheirpossiblecauses.Ifyoucannotcorrectthemalfunction usingthesolutionsnamed,contactyourproviderdirectly.
MalfunctionPossibleCauseSolution
DifcultyinadjustingtheseatheightDirtbetweentheseatframeandthe
sidepart
Thecastorsaredifcultorimpossibleto turn
Difcultypushingthefootreststothe side
Theseatframewithbackrestcannotbe tilted
Higherforceneededtotriggerthegas spring(DualVIP)
Thebackrestcannotbetilted(DualVIP)
Rattlingnoiseduringtiltingoftheseat framewithbackrest.
DirtbetweentheforkandtherimThoroughlycleanthecastors
DepositsintheguidesThoroughlycleanthefootrestsandthe
DepositsintiltingmechanismRemovebrakecoverandbrakelever
BowdencablenotcorrectlyadjustedTheBowdencablemustbeadjustedby
Long-timestorage
BowdencablenotcorrectlyadjustedTheBowdencablemustbeadjustedby
BowdencablenotcorrectlyadjustedTheBowdencablemustbeadjustedby
Theactivationhandleisnotpulled completely
Disassemblethebaseframeandclean baseframeandsidepartthoroughly
guides
coverandthoroughlycleanthetilting mechanism
aqualiedtechnician
Repeatthetiltingofthebackrestseveral times
aqualiedtechnician
aqualiedtechnician
Pulltheactivationhandlecompletely.
Troubleshooting
1652073-A31
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo

10TechnicalData

10.1DimensionsandWeight

OceanVIPErgo
A
B
321652073-A
OceanVIPErgoXL*
480mm
480mm
OceanDualVIPErgo
TechnicalData
OceanVIPErgo
C
1
C
2
D
E
F
565mm645mm565mm
ca.1052mm–1340mm(inneutralseatposition)
ca.1052mm–1202mm(inneutralseatposition)
G
OceanVIPErgoXL*
475mm–600mm
533mm–658mm
ca.1000mm
OceanDualVIPErgo
Tilted:ca.1520mm
Tiltedandbackrestreclined(OceanDualVIPonly):ca.1700mm
H
I
J
K
L
475mm535mm475mm
M
N
213mm/251mm/289mm
319mm–469mm(instepsof25mm)
350mm(top)/300mm(bottom)
250mm(top)/182mm(bottom)
O
P
Q
R
S
295mm
230mm
ca.742mm
160mm
400mm–525mm
380mm
385mm
T∅5"(127mm)
1652073-A33
Aquatec®OceanVIPErgoAquatec®OceanDualVIPErgo
U
Angleofinclinationseat
Angleofinclinationbackrest
Turningdiameter
Reversingwidth
OceanVIPErgo
OceanVIPErgoXL*
172mm
-5°–40°
2200mm
1140mm
OceanDualVIPErgo
Productweight23.5kg28.3kg
Maximumuserweight150kg
*XL-armrestsareoptionallyavailable.

10.2Materials

Frame
SeatplatePolypropylene,ber-glassreinforced
BackrestcoverNylon,PVC-coated
Footrests
Armrest
Castors
Allproductcomponentsarecorrosionresistantanddonot containnaturalrubberlatex.
Stainlesssteel,powder-coated
Polypropylene,ber-glassreinforced
Polyamid,ber-glassreinforced
Plasticsandstainlesssteel

10.3EnvironmentalParameters

OperatingConditions
Ambienttemperature
Relativehumidity15%-93%,non-condensing
Atmosphericpressure795-1060hPa
StorageandShippingConditions
Ambienttemperature
Relativehumidity30%-75%,non-condensing
Atmosphericpressure795-1060hPa
10-65°C
0-65°C
5°–35°
1290mm
341652073-A
Notes
UnitedKingdom&Ireland:
UKR P
InvacareLimited PencoedT echnologyPark,Pencoed BridgendCF355AQ Tel:(44)(0)1656776200 uk@invacare.com www.invacare.co.uk
Australia:
InvacareAustraliaPty .Ltd. Unit18/12StantonRoad, SevenHills,NSW2147, Australia Phone:1800460460 Fax:1800814367 orders@invacare.com.au www.invacare.com.au
EUExport:
InvacarePoirierSAS RoutedeStRoch F-37230Fondettes Tel:(33)(0)247626980 serviceclient_export@invacare.com www.invacare.eu.com
NewZealand:
InvacareNewZealandLtd 4WesteldPlace,MtWellington1060 NewZealand Phone:0800468222 Fax:0800807788 sales@invacare.co.nz www.invacare.co.nz
USA:
InvacareCorporation OneInvacareWay Elyria,OhioUSA 44035 800–333–6900 www.invacare.com
Canada:
InvacareCanadaL.P . 84CitationDr#14 Concord,ONL4K3C1 Phone:8006685324 www.pro.invacare.ca
InvacareGmbH AmAchenerHof8 D-88316Isny Germany
InvacareUKOperationsLimited Unit4,P encoedT echnologyPark,Pencoed BridgendCF355AQ UK
1652073-A2023-08-18