Invacare IRC5PO2V User Manual

Invacare®Perfecto Concentrator
IRC5PO2V
enHomeFill®SystemCompatible
UserManual
₂ ₂
™VOxygen
©2016Invacare®Corporation Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. MakingLife’sExperiencesPossibleisatrademarkintheU.S.A. Pine-solandLestoilareregisteredtrademarksofTheCloroxCompany. DawnisaregisteredtrademarkofTheProctorandGambleCompany.
Contents
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
1.3IndicationsForUse.............................6
1.4Description...................................6
1.5Contraindications..............................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................8
2.3RadioFrequencyInterference......................11
3Components....................................12
3.1ComponentIdentification.........................12
3.2PneumaticDiagram.............................14
4Accessories.....................................15
4.1OptionalAccessories............................15
5Setup..........................................16
5.1Unpacking....................................16
5.2Inspection....................................16
5.3Storage......................................16
6Usage..........................................17
6.1Introduction..................................17
6.2SelectaLocation...............................17
6.3SettingUptheConcentrator......................18
6.3.1SettingUptheHumidifier......................18
6.3.2TurningtheConcentratorOn...................20
6.3.3Flowrate..................................20
6.3.4SensO
OxygenPurityIndicator.................21
2
6.3.5InitialStartupoftheConcentrator................21
6.3.6ExplanationofOxygenPurityIndicatorLights........22
7Maintenance....................................23
7.1ServiceLife...................................23
7.2CleaningtheCabinet............................23
7.3CleaningtheCabinetFilter........................24
7.4HumidifierCleaningandThermicDisinfection..........24
7.5CleaningandDisinfectionBetweenPatients............25
7.6PreventiveMaintenanceChecklist...................26
8AfterUse.......................................27
8.1RecyclingInformation............................27
8.2WearandTear................................27
9Troubleshooting.................................28
9.1Troubleshooting...............................28
10Technicaldata..................................30
10.1Specifications.................................30
11Warranty......................................33
11.1NorthAmericanLimitedWarranty.................33
Invacare®Perfecto™VOxygenConcentrator

1General

1.1Symbols

Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
SymbolsinDocumentation
GeneralWarningSign
Thebackgroundcolorinsidethe triangleisyellowonproductlabels.
Readmanual
Thecolorofthesymbolbackground isblueonproductlabels.
NoSmoking
Thecolorofthecirclewithdiagonal barisredonproductlabels.
NoOpenFlame
Thecolorofthecirclewithdiagonal barisredonproductlabels.
ClassII,DoubleInsulated
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
IndoorUseONLY
KeepDry
4
1193322-A-04
-20F MIN
150F MAX
TransportandStorageTemperature
15%
MIN
95% MAX
TransportandStorageHumidity
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
O
Indicators
2
SymbolO
O
2
PurityIndicatorLights(LED)
2
SYSTEM
GREENIndicatorLight
OKAY
O
over85%
2
O 73%to85%
between
2
YELLOWIndicatorLight
A.YELLOWSolid
B.YELLOWFlashingSensor
Failure.Callaqualifiedtechnician.
SYSTEM FAILURE
O
Below73%
2
REDIndicatorLight
RefertoTroubleshooting.
General
ElectricalHazard

1.2IntendedUse

TheintendedfunctionanduseoftheInvacare®Perfecto™V
Manufacturer
OxygenConcentratoristoprovidesupplementaloxygentopatients withrespiratorydisorders,byseparatingnitrogenfromroomair,by wayofamolecularsieve.Itisnotintendedtosustainorsupportlife.
Callsupplier
DANGER! RiskofInjuryorDeath
SymbolsonProduct
Unitrunning
Thisproductistobeusedasanoxygensupplementand isnotintendedtobelife-supportingorlife-sustaining. ONLYusethisproductifthepatientiscapableof spontaneousbreath,abletoinhaleandexhalewithout
Unitnotrunning
theuseofamachine.
–DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
1193322-A-045
Invacare®Perfecto™VOxygenConcentrator
DANGER! RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientsonflow ratesgreaterthan5L/minmaybeatanincreasedrisk forseriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
caresupplierBEFOREusingthisproductifyouare prescribedaflowrategreaterthan5L/min.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand specificationshasnotbeentestedandmayleadto productdamage,lossofproductfunction,orpersonal injury.
–DONOTusethisproductinanywayotherthan
describedintheSpecificationsandIntendedUse sectionsofthismanual.
DANGER! RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen concentratorinthemarkettoday,thisoxygen concentratorcanfailtoproduceoxygenduetopower failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient toswitchtotheirbackupsourceofoxygen.Referto Troubleshootingformoredetail.

1.3IndicationsForUse

TheintendedfunctionanduseoftheInvacare®Perfecto™V OxygenConcentratoristoprovidesupplementaloxygentopatients withrespiratorydisorders,byseparatingnitrogenfromroomair,by wayofamolecularsieve.Itisnotintendedtosustainorsupportlife.

1.4Description

TheInvacarePerfecto respiratorydisorderswhorequiresupplementaloxygen.Thedevice isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87% to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse ofanasalcannula.
TheInvacarePerfecto pressureswingadsorptionmethodologytoproducetheoxygen gasoutput.Ambientairentersthedevice,isfilteredandthen compressed.Thiscompressedairisthendirectedtowardoneof twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe patient.
TheInvacarePerfecto patientinahomeenvironmentorinaninstitutionalenvironment. Deviceoperatesatanominal120VAC/60Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified technicalpersonnelonly.
concentratorisusedbypatientswith
2
concentratorusesamolecularsieveand
2
concentratoriscapableofoperationbythe
2

1.5Contraindications

Therearenoknowncontraindications.
61193322-A-04
Safety

2Safety

2.1LabelLocations

A
B
C
D
1193322-A-047
Serialnumberlabelislocatedontheresonator intakeassembly.
SpecificationLabelislocatedonthebackofthe concentratoratthebase.
Invacare®Perfecto™VOxygenConcentrator

2.2GeneralGuidelines

Inordertoensurethesafeinstallation,assemblyandoperationofthe concentratortheseinstructionsMUSTbefollowed.
DANGER! RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor damage.Thissectioncontainsimportantinformationfor thesafeoperationanduseofthisproduct.
–DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand understandingtheseinstructionsandanyadditional instructionalmaterialsuchasusermanuals,service manualsorinstructionsheetssuppliedwiththis productoroptionalequipment.
–Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional, dealerortechnicalpersonnelbeforeattemptingto usethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot workingcorrectly,contactatechnicianorInvacare forrepair.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy substancesandothercombustiblesareeasilyignited andburnwithgreatintensityinoxygenenrichedairand whenincontactwithoxygenunderpressure.Toavoid fire,death,injuryordamage:
DONOTSMOKEwhileusingthisdevice.DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
NOSMOKINGsignsshouldbeprominentlydisplayed.Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby anytypeoffriction.
–Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutofthe roominwhichthisconcentratorislocatedandaway fromwhereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings, chaircushions,clothing,andawayfromheatedorhot surfacesincludingspaceheaters,stoves,andsimilar electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor ontheorderofamedicalpractitionerlicensedbya governmentalagencywherehe/shepractices.
–ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
81193322-A-04
Safety
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor injurytopersons:
–DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable parts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the concentratormustbelocatedinanotherroomat least7ft(2.1m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
–DONOTusefrayedordamagedACpowercords.
WARNING! RiskOfInjuryOrDamage
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare accessories.Accessoriesdesignedbyother manufacturershavenotbeentestedbyInvacareand arenotrecommendedforusewithInvacareproducts.
–Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis device.Youshouldcontactyourlocalhomecare supplierforrecommendationsonwhichofthese deviceswillbebestforyou.Theyshouldalsogiveyou adviceontheproperusage,maintenance,andcleaning.
DANGER! RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway contamination:
–DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics, cleaningagents,orchemicalvapors.
WARNING! RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorimpairedindividuals.
–Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING! RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing entanglement:
–Keepchildrenandpetsawayfromnasalcannulaand
tubing.
–Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/orimpairedpersons.
WARNING! RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.Check
withyourequipmentsupplierorphysicianto determinehowoftenthecannulashouldbereplaced.
–DONOTsharecannulasbetweenpatients.
1193322-A-049
Invacare®Perfecto™VOxygenConcentrator
WARNING! RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto you.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings shouldbechanged.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorbypulling
onthecord.
–DONOTuseextensioncordswithACpowercords
provided.
–Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhenplugged
in.
–Makesureconcentratorisoffwhennotinuse.
WARNING! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned tominimizeroutinepreventivemaintenance.Toprevent injuryordamage:
–Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas factorytrainedpersonnelshouldperformpreventive maintenanceorperformanceadjustmentsonthe oxygenconcentrator,exceptfortasksdescribedin thismanual.
–UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION! RiskofDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto water,callequipmentsupplier/qualifiedtechnicianfor examinationandrepair.
–NEVERdroporinsertanyobjectorliquidintoany
opening.
–ForindooruseONLY.
101193322-A-04

2.3RadioFrequencyInterference

WARNING! RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrom interferencewithwirelessequipment:
–Keepconcentratoratleast9.8ft(3.0m)awayfrom
wirelesscommunicationequipmentsuchaswireless homenetworkdevices,mobilephones,cordless phonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits specifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovide areasonableprotectionagainstelectromagneticinterferenceina typicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels ofelectromagneticemissionspermittedbytheabovestandards. Todetermineiftheemissionsfromtheconcentratorarecausing theinterference,turntheconcentratorOff.Iftheinterference withtheotherdevice(s)stops,thentheconcentratoriscausingthe interference.Insuchrarecases,interferencemaybereducedor correctedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweenthe
equipment.
Connecttheequipmentintoanoutletonacircuitdifferentfrom
thattowhichtheotherdevice(s)isconnected.
Safety