Invacare®HomeFill®IIOxygen
System
ModelIOH200AW
en Compressor
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor
futurereference.
©2020 Invacare Corporation
All rights reserved. Republication, duplication or modication i n w hole o r i n p art i s prohibited
without prior written permission from Invacare. Trademarks are identied by ™and ® . A ll trademarks
are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
Platinum is a registered trademark in the U.S.A., Europe and Australia.
3-in-1 oil is a registered trademark of American Home Products Corporation.
WD-40 is a registered trademark of the WD-40 Company.
Updated versions of this manual are available on www.invacare.com.
Making Life's Experiences Possible is a registered trademark in the U.S.A.
Contents
1General.........................................4
1.1Symbols.....................................4
1.2IntendedUse.................................5
1.3IndicationsForUse.............................6
1.4ReplacementParts.............................6
1.5Contraindications...............................6
1.6Warranty.....................................6
2Safety..........................................7
2.1GeneralGuidelines.............................7
2.2LabelLocations................................10
2.3ProviderInformation............................12
2.4Double-InsulatedProducts........................12
2.5EMCInformation...............................12
3Setup...........................................14
3.1Unpacking....................................14
3.2CompressorInitialSetupChecklist..................14
3.3ChoosingaLocation............................15
3.4TransportingtheCompressor......................16
3.5InstallingthePowerCord........................16
3.6ConnectingtheCompressortotheConcentrator.......17
3.6.1ConnectingtheCompressortoInvacare
Concentrators..............................17
3.6.2ConnectingtheCompressortoNon-Invacare
Concentrators..............................17
4Usage..........................................18
4.1CompressorOperationChecklist...................18
4.2CylinderPrellInspection........................19
4.3ConnectingtheCylindertotheCompressor...........21
4.4DisconnectingtheCylinderfromtheCompressor.......22
4.5TurningtheCompressorOnandOff................24
5Maintenance.....................................26
5.1MaintenanceInfo..............................26
5.2WearandTear................................26
5.3ServiceLife...................................26
5.4CleaningtheCompressorFilter....................26
5.5CleaningtheCabinet............................27
5.6ReplacingtheFuse.............................28
6AfterUse........................................29
6.1Disposal.....................................29
7T roubleshooting...................................30
7.1IndicatorLights................................30
7.2Troubleshooting...............................32
8T echnicalData....................................35
8.1T echnicalDescription............................35
8.2CylinderFillTimes..............................35
8.3Specications.................................36
8.4ElectromagneticCompliance(EMC).................37
Invacare®HomeFill®IIOxygenSystem
1 General
1.1 Symbols
SymbolsintheManual
Signalwordsareusedinthismanualandapplytohazards
orunsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Seetheinformationbelowfordenitions
ofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin
deathorseriousinjury.
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin
deathorseriousinjury.
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin
propertydamageorminorinjuryorboth.
IMPORTANT
–Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformation
forefcient,trouble-freeuse.
SymbolsontheProduct
Thefollowingsymbolsarefoundonthecompressor,
compressorpackagingortheoxygencylinder .
ReadtheManual(Compressor)
Thecolorofthesymbol
backgroundisblueonproduct
labels.
ReadtheManual(Cylinder)
ElectricalHazard
AlternatingCurrent
TypeBFequipment—AppliedParts
Powerswitch“ON”
Powerswitch“OFF”
DONOTsmoke
4
60001253-D
General
NoOpenFlame
ClassII,DoubleInsulation.Referto
DoubleInsulatedproductsinthe
Safetysectionofthemanual.
Keepdryintransportorstorageand
duringuse
Protectedagainstsolidforeignobjects
of12.5mmdiameterandgreater.
Protectedagainstverticallyfalling
waterdrops.
Fuse
Oxidizer
Thecolorofthesymbol
backgroundisyellowon
productlabels.
Non-ammablegas,class2type
(oxygen)
Thecolorofthesymbol
backgroundisgreenon
productlabels.
TransportandStorageT emperature
Transport and Storage Humidity
Recycle this product. Refer to Disposal
of Equipment and Accessories in the
After Use section of the manual.
Complies With Applicable EU
Directives Including Medical Device
Directive (MDD) (93/42/EEC)
Authorized Representative in the
European Community
Manufacturer
1.2 IntendedUse
TheInvacareHomeFillcompressorwasdesignedtobe
usedonlywithoxygenconcentratorsthatmeettheinput
specicationslistedin8.3Specications,page36.
DONOTusetheInvacareHomeFillcompressorwith
anythingotherthananoxygenconcentratorthatmeetsthe
compressorinputrequirements.
TheintendeduseofanInvacareHomeFillcompressorisan
accessorydevicetoanoxygenconcentratorthatenables
thepatienttorellcustomportableoxygencylindersfor
personaluse.
60001253-D
5
Invacare®HomeFill®IIOxygenSystem
Thisproductisintendedforsinglepatientuseintheindoor
homecareenvironmentonlyandisnottobeusedforbulk
cylinderrelling.
Nospecicproductknowledgeortrainingisrequired,
otherthanwhatiscontainedinthismanual,tooperate
theproduct.
Itisnotintendedforsupplyingoxygentherapyduring
relling.
Thisproductisintendedtobeusedbyorunderadult
supervisiononlyafterreadingandunderstandingthe
instructionsandwarningsofthisusermanual.
WARNING!
–Theuseofthisdeviceislimitedtorellinga
specicpatient’sownoxygencylinders.The
designatedcylindersMUSTbeusedONL Yby
thespecicoxygenpatientandarenotto
bedistributedtoanyotherindividualforany
purpose.
–Nootheruseofthisdeviceisindicated.
1.3 IndicationsForUse
Indicatedtoenablethepatienttorellcustomportable
oxygencylindersforpersonaluse.
Thesepartsarebasemodels.PleasecontactInvacare
oryourproviderforcountryspecicpartnumbers.
1.5 Contraindications
Therearenoknowncontraindications.
1.6 Warranty
Termsandconditionsofthewarrantyarepartofthegeneral
termsandconditionsparticulartotheindividualcountriesin
whichthisproductissold.
1.4 ReplacementParts
ThefollowingreplacementpartsareavailablefromInvacare:
• SupplyLine,Partnumber1105177
• PowerCord,Partnumber1085605
6
60001253-D
Safety
2 Safety
2.1 GeneralGuidelines
DANGER!
RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,
injuryordamage.Thissectioncontainsimportant
informationforthesafeoperationanduseofthis
product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,providerortechnicalpersonnel
beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
–Inordertoensurethesafeinstallation,
assemblyandoperationoftheHomeFill
CompressortheseinstructionsMUSTbe
followed.
–THEINFORMA TIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
–DONOTSMOKEwhileusingthisdevice.
–DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonthecompressor
unlessrecommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
–Keepallmatches,lightedcigarettesorother
sourcesofignitionoutoftheroominwhich
thiscompressorislocatedandawayfrom
whereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andcompressor
outfromundersuchitemsasblankets,
bedcoverings,chaircushions,clothing,and
awayfromheatedorhotsurfacesincluding
spaceheaters,stoves,andsimilarelectrical
appliances.
60001253-D
7
Invacare®HomeFill®IIOxygenSystem
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
–DONOTallowoilfromyourhandsorother
sourcestocomeintocontactwiththe
compressorcouplingorcylinderttings.These
substancesmaybecomeammableinthe
presenceofoxygenandcauseinjury.Avoid
touchingthecylinderttingwheneverpossible.
Exampleofitemstoavoidarehandcreamsor
lotions,cookingoils,suntanoilsorsunblocks
andsimilaroilyproducts.Ifyouusethese
typesofproductsInvacaresuggestsyouwash
yourhandspriortousingtheHomeFillunit.
Ifthecylinderttingshouldcomeincontact
withsuchsubstancesitshouldbecleanedoff
withadampclothpriortoconnectingittothe
compressor .
–NEVERoilorlubricatethecompressorcoupler
orcylinderttings.NEVERuseanypenetrating
oilorlubricantsuchasWD-40®or3-in-1
Oil®.Ifconnectiondifcultyisexperienced
verifythatthecylinderttingisbeinginserted
straightdownintothecompressorcoupler .If
continueddifcultyisexperiencedcontactyour
supplierforassistance.
DANGER!
RiskofElectricShock
Thisunitcontainsmovingandhighpressure
components.Therearenouserserviceableparts
insidethecover .T oavoidelectricshock,injury
ordeath:
–DONOTremovethecover .
–Referservicingtoqualiedpersonnel.
WARNING!
RiskofSeriousInjuryorDamage
Useofnon-Invacareaccessoriesmayresultin
seriousinjuryordamage.
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunctionwith
Invacareaccessories.Accessoriesdesignedby
othermanufacturershavenotbeentestedby
Invacareandarenotrecommendedforuse
withInvacareproducts.
–Nomodicationsofthisequipmentisallowed.
–DONOTusenon-Invacareaccessories.
–T oobtainInvacareaccessories,contactInvacare
byphoneoratwww .invacare.com
8
60001253-D
Safety
WARNING!
Riskofinjuryordamage
Blockingtheairowintoandoutofthe
compressormaycauseinjuryordamage.
–NEVERblocktheairopeningsoftheproduct
orplaceitonasoftsurface,suchasabedor
couch,wheretheairopeningmaybeblocked.
–Keeptheopeningsfreefromlint,hairand
similarforeignitems.
WARNING!
RiskofInjuryorDamage
Adamagedproductmaymalfunctionandcause
injuryordamage.T oavoidinjuryordamage:
–DONOTgetthecompressorwet.
–Ifthecompressorhasadamagedcordorplug,
ifitisnotworkingproperly,ifithasbeen
droppedordamaged,orsubmersedinwater ,
callaqualiedtechnicianforexaminationand
repair .
WARNING!
–DONOTusethecompressoroutdoors.Indoor
useonly.
–Ifanyleakageoftheoxygencylinderisdetected
DONOTattempttousethecylinder.T urnthe
owselectorknobontheregulatorto“Off”.If
leakagepersists,placethecylinderoutdoors
andnotifyyourHomeHealthCareProvideror
ServiceRepresentativeofthiscondition.
–ChildrenshouldALWA YSbesupervisedaround
theHomeFillunit.Whenusedwiththe
HomeFilltableorsimilarelevatedsupport
surfaces,achildcouldpulloveranysuchtable
causingtheunittofall.Failuretoobserve
thiswarningmayresultinpersonalinjuryor
productdamage.
WARNING!
Riskofinjuryordamage
TheHomeFillisequippedwithahighpressure
reliefvalvetoensuretheuser’ssafety .When
activated,thissafetyfeaturewillmakean
extremelyloudnoise.Toavoidinjuryordamage:
–Ifthisnoiseoccurs,turntheunitOffand
contactyourHomeHealthCareProvideror
ServiceRepresentative.
60001253-D
9
Invacare®HomeFill®IIOxygenSystem
WARNING!
Riskofdeath,injuryordamage
Ifimproperlypositionedandsecured,thepower
cordandinterconnecthosecancauseinjurydue
totripping,falling,andstrangulation.Product
damagemayalsooccur.
–Properlystoreandpositionelectricalcords
and/ortubingtopreventtrippingand
strangulationhazards.
–Avoidpositioningpowercordsacrossareasof
hightrafc.
–Beforemovingorrepositioningeitherthe
compressororconcentrator ,ALWAYSdisconnect
theACpowercordsandtheinterconnecthose
betweenthecompressorandconcentrator .
CAUTION!
Statutorylawrestrictsthisdevicesaletooron
theorderofamedicalpractitionerlicensedbya
governmentalagencywherehe/shepractices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
2.2 LabelLocations
A
10
60001253-D
Safety
B
C
D
E
60001253-D
11
Invacare®HomeFill®IIOxygenSystem
2.3 ProviderInformation
Invacarerecommendsleavingafullcylinderof
oxygenwiththepatientaftersettinguptheHomeFill
compressor .
2.5 EMCInformation
BeforeleavingaHomeFillCompressorwithauser,the
followingchecklistMUSTbecompleted:
qMakesurethepressuretestoncylinderhasnotexpired.
qChecktheEMCsectionforoperationalinformationand
possibleRFissues.RefertoEMCInformation.
q Instructtheuseronthesafeoperationofthe
concentratorandreviewallwarnings.
q InstructtheuseronthesafeoperationoftheHomeFill
compressorandreviewallwarnings.
qLeaveafullcylinderofoxygenwiththeuser.
qLeaveacopyofallwarningsandllinginstructionswith
theuser.
2.4 Double-InsulatedProducts
CAUTION!
–DONOTunplugthepowercordbypullingon
thecord.T ounplug,grasptheplugnotthe
cord,otherwisedamagemayoccur.
Inadouble-insulatedproduct,twosystemsofinsulation
areprovidedinsteadofgrounding.Nogroundingmeansis
requiredonadouble-insulatedproduct,norisameansfor
groundingtobeaddedtotheproduct.Adouble-insulated
productismarkedwiththewords“DOUBLEINSULATION”or
“DOUBLEINSULATED”orthesymbol:
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamage
frominterferencewithwirelessequipment:
–Keepthecompressoratleast9.8ft(3.0m)
awayfromwirelesscommunicationequipment
suchaswirelesshomenetworkdevices,mobile
phones,cordlessphonesandbasestations,
walkie-talkies,etc.
WARNING!
–Useofaccessoriesandcablesotherthanthose
speciedorprovidedbyInvacarecouldresult
inincreasedelectromagneticemissionsor
decreasedelectromagneticimmunityofthis
equipmentandresultinimproperoperation.
MedicalElectricalEquipmentneedstobeinstalledandused
accordingtotheEMCinformationinthismanual.
Thisequipmenthasbeentestedandfoundtocomply
withEMClimitsspeciedbyIEC/EN60601-1-2forType
BFequipment.Theselimitsaredeterminedtoprovide
reasonableprotectionagainstelectromagneticinterference
inatypicalhomehealthcareenvironment.Useofthis
equipmentinanenvironmentwithhigherlevelsof
interferencecouldaffecttheaccuracyoftheoxygen
12
60001253-D
concentrationinthecylinder.Thisequipmentisnotintended
foruseinaprofessionalhealthcareenvironment.
PortableandmobileRFcommunicationsequipmentcan
affecttheoperationofthisequipment.Thecompressor
maybestackedontoponanInvacareHomeFillcompatible
oxygenconcentratorduringsetup.Thecompressoroperation
shouldbeobservedtoverifynormalcylinderllingineither
situation.
Otherdevicesmayexperienceinterferencefromeven
thelowlevelsofelectromagneticemissionspermitted
bytheabovestandard.T odetermineiftheemissions
fromthecompressorarecausingtheinterference,turn
thecompressorOff.Iftheinterferencewiththeother
deviceoperationstops,thenthecompressoriscausing
theinterference.Insuchrarecases,interferencemaybe
reducedorcorrectedbyoneofthefollowingmeasures:
• Reposition,relocate,orincreasetheseparationbetween
thedevices.
• Connecteitherlinepowereddevicetoadifferent
electricalpowercircuit.
Foradditionalinformation,referto8.4Electromagnetic
Compliance(EMC),page37 .
Safety
60001253-D
13
Invacare®HomeFill®IIOxygenSystem
3 Setup
3.1 Unpacking
DESCRIPTION
CompressorAssembly
PackagingAssembly
SupplyLineAssembly
PowerCord
UserManual(notshown)
A
B
C
D
N/A
1. Checkforanyobviousdamagetothecartonorits
contents.Ifdamageisevident,notifythecarrier,or
yourlocalprovider.
2. Removeallloosepackingfromthecarton.
3. Carefullyremoveallofthecomponentsfromthecarton.
UnlessthecompressoristobeusedIMMEDIATELY ,
leavethecompressorinitspackagingforstorage
untiluseofthecompressorisrequired.
3.2 CompressorInitialSetupChecklist
ProvidersONL Y—RefertoDealerInformationinthe
Safetysectionofthismanualbeforeperformingthis
procedure.
UsersONLY—ContactyourInvacareprovideror
Invacareforassistancewiththesetupprocedureand
toreportunexpectedoperationorevents.
BeforeoperatingtheHomeFillCompressorforthersttime,
completethefollowingchecklist:
qChoosealocationforthecompressor .
qTransportthecompressortothedesiredlocation.
14
60001253-D
Setup
qConnectthepowercordtothecompressor.
qConnectthecompressortotheconcentrator .
3.3 ChoosingaLocation
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
–DONOTSMOKEwhileusingthisdevice.
–DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonthecompressor
unlessrecommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
–Keepallmatches,lightedcigarettesorother
sourcesofignitionoutoftheroominwhich
thiscompressorislocatedandawayfrom
whereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andcompressor
outfromundersuchitemsasblankets,
bedcoverings,chaircushions,clothing,and
awayfromheatedorhotsurfacesincluding
spaceheaters,stoves,andsimilarelectrical
appliances.
60001253-D
Thecompressorshouldbeatleastthreeinches(7.6cm)
awayfromwalls,draperiesorfurnituretoensuresufcient
airow.
Avoidplacingthecompressornexttoheaters,radiatorsor
hotairregisters.
DONOTuseinacloset.
Thecompressorshouldsitonahardatstablesurface,such
asatableortheReadyRack.
TheHomeFillReadyRackissoldseparatelyand
usedtopositiontheHomeFillcompressorontop
ofaPerfecto2orPlatinum®oxygenconcentrator .
RefertotheHomeFillReadyRackinstructionsheet
partnumbers1148363and1134824formore
information.
15
Invacare®HomeFill®IIOxygenSystem
3.4 TransportingtheCompressor
WARNING!
–NEVERtransportacompressorwithacylinder
connectedtoorstoredonthecompressor ,
otherwiseinjuryordamagecanoccur.
CAUTION!
–Whentransportingthecompressor ,becareful
nottobumpthecompressorortheconnectors
intoobstacles.Otherwise,damagetothe
compressormayoccur .
1. Therearetwomethodstosafelytransportthe
compressor:
• OneHand-GraspthehandgripAatthebottom
rearofthecompressor.
• TwoHands-GraspthehandgripsBonthesides
ofcompressor.
3.5 InstallingthePowerCord
1. Chooseaproperlocation.RefertoChoosingaLocation
inthissectionofthemanual.
2. MakesurethepowerswitchisintheOff(O)position.
3. PlugthepowercordintotheelectricconnectorAin
thebackofthecompressorB.
4. PlugtheoppositeendofthepowercordCintothe
walloutletD.
5. Leaveaminimumof3in(7.6cm)clearancetothe
powercordendsfordisconnectionaccess.
Thepowerinletactsasthemainsdisconnection
devicetoremoveallpowertotheunitwhenthe
linecordisdisconnected.
16
60001253-D
Setup
3.6 ConnectingtheCompressortothe
Concentrator
3.6.1 ConnectingtheCompressortoInvacare
Concentrators
1. InsertoneendoftheinterconnecthoseCintothe
outletttingBonthebackoftheconcentratorA.
2. Inserttheoppositeendoftheinterconnecthoseinto
theinletttingDonthesideofthecompressor .
ThemetaltabsEontheconcentratoroutlet
ttingandthecompressorinletttingwillpop
outwithanaudible“click”whentheendofthe
interconnecthoseisproperlyinstalled.
3.6.2 ConnectingtheCompressortoNon-Invacare
Concentrators
1. Refertotheoxygenconcentratorusermanualoutlet
gasspecicationstoverifycompatibilitywiththe
HomeFillcompressorinputgasrequirementslistedin
8.3Specications,page36 .
2. Refertotheoxygenconcentratorusermanualfor
connectinginstructions.
60001253-D
17
Invacare®HomeFill®IIOxygenSystem
4 Usage
4.1 CompressorOperationChecklist
EachtimetheHomeFillCompressorisusedtollacylinder ,
completethefollowingchecklist:
q EnsuretheconcentratorhasbeenOnforatleastthirty
minutes.RefertotheconcentratorUserManual.
qPerformtheprellinspectiononthecylinder .
qConnectthecylindertothecompressor .
qPushthecompressorpowerswitchtotheOn(I)
position.
qExaminetheindicatorlightsonthecontrolpanel.
qDisconnectandremovethefullcylinder .
qPushthecompressorpowerswitchtotheOff(O)
position.
q Ifllinganothercylinder,repeatthischecklist.
18
60001253-D
4.2 CylinderPrellInspection
WARNING!
–ONL YusecylindersthathavetheInvacareHomeFillIIconnectionandHomeFillIIlabel.
–AllcylindersMUSTbeinspectedbeforeattemptingtoll.Otherwise,injuryordamagemayoccur.
ToidentifyaHomeFillIIcompatiblecylinderA,lookforablackstripeBonthetopofthecylinderandlookfortheHomeFillII
labelC.
Usage
60001253-D
19
Invacare®HomeFill®IIOxygenSystem
TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeverytenyears.
WARNING!
RiskofInjury
Toreducetheriskofinjury:
–DONOTllcylindersthathavenotbeentestedinthepasttenyears.Otherwise,seriouspersonalinjurymay
result.Contactyourequipmentproviderforreplacement.
ExternalExamination
1. Examinetheoutsideofthecylinderforthefollowingconditions,andreplacethecylinderiftheyexist:
• Dentsordings
• Arcburns
• Oilorgrease
• Anyothersignsofdamagethatmightcauseacylindertobeunacceptableorunsafeforuse.
2. Examinethecylinderforevidenceofreorthermaldamage.Evidenceincludescharringorblisteringofthepaint,orother
protectivecoatingorheatsensitiveindicator.Ifreorthermaldamageisfound,replacethecylinder.
3. Inspectthecylinder/regulatorassemblyforthefollowingandreplaceiffound:
• Debris,oilorgrease
• Noticeablesignsofdamage
• Signsofcorrosioninsidethevalve
• Signsofexcessiveheatorredamage
20
60001253-D
4.3 ConnectingtheCylindertotheCompressor
WARNING!
–NEVERusetoolsofanykindto
connect/disconnectthecylinderandthe
compressor .Otherwise,severeinjuryand/or
damagemayoccur.
WARNING!
–DONOTdropoxygencylinders.Usetwohands
whenhandling/transportingoxygencylinders.
Otherwise,injuryordamagemayoccur .
–DONOTusealiquidleakdetectortotestfor
leaks.
Usage
Fillportcoversnotshownforclarity.
60001253-D
21
Invacare®HomeFill®IIOxygenSystem
1. Priortoeachuse,inspectproductforvisibledamage.
RefertotheExternalExaminationsection.DONOTuse
ifanydamageisfound.
Ifforanyreason,anylabelbecomesillegibleor
lost,contactyourequipmentprovider .
2. Examinethecylindergauge.Ifthecylinderpressureis
lessthan1500psig(10342kPa),proceedtoSTEP3.
Forlocationofthecylindergauge,refertothe
cylinder’smanual.
Ifthecylinderpressureisgreaterthan1500psig
(10342kPa),DONOTattempttotopoffthis
cylinder.Itmaynotll.
3. SettheregulatorowdialonthecylinderAtoOff.
Refertothecylinder’smanual.
4. Removethecylinderandconnectorllportcovers(if
present).
5. MomentarilypushDOWNontheouterring(sleeve)B
oftheconnectorllportCuntilGREENdotsDare
visibletoresettheconnector .
Iftheouterring(sleeve)isintheUPposition
(GREENdotsnotvisible),theconnectorllport
willnotbeabletoacceptthecylinderllportE.
PushingDOWNFmomentarilywillresetthe
connectorllport(GREENdotsvisible)toaccept
thecylinderllport.
8. Alignthecylinderllportwiththeconnectorllport.
9. PullUPontheouterring(sleeve)oftheconnector
llportwhilepushingDOWNonthecylindertocouple
thecylinderllportintotheconnectorllport.
Thecylinderisproperlyconnectedwhenan
audible“click”isheard.
4.4 DisconnectingtheCylinderfromthe
Compressor
6. Graspthecylinderintheareabehindthecylinder
llport.
7. PositionthecylinderinthecompressorcradleG.
22
60001253-D
Usage
WARNING!
–NEVERusetoolsofanykindto
connect/disconnectthecylinderandthe
compressor .Otherwise,severeinjuryand/or
damagemayoccur.
1. PressthecompressorpowerswitchAtotheOff(O)
position.
2. GraspthecylinderBintheareabehindthecylinder
llportC.
3. Withtheotherhand,grasptheouterring(sleeve)of
theconnectorllportEandpushDOWNF.
4. Liftuponthecylindertoremovefromtheconnector
llport.
5. Whenthecylinderllportisdisconnectedfromthe
connectorllport,releasetheouterring(sleeve)ofthe
connectorllportandusetwohandstoremovethe
cylinderfromthecompressorcradle.
WARNING!
–Thellportcoverontheconnectorllport
andthecylinderllportMUSTbereplaced
afterllingandwhenevernotinuse.If
eitherofthellportcoversaremissing
contactyourhealthcareproviderbefore
using.
60001253-D
23
Invacare®HomeFill®IIOxygenSystem
6. PlacethellportcoversontotheconnectorllportG
andcylinderllportH.
Whenusingthepostvalvetypecylinder,
attachanoxygenregulatorpermanufacturers
instructions.
7. Examinethecylindergaugetoensurecylinderisfull
(needlepointingintoGREENarea).Ifthecylinderisnot
full,performthefollowing:
a.RepeatConnectingtheCylindertotheCompressor,
inthissectionofthemanual.
b.Ifcylinderisnotfullaftersecondattempttoll,
contacttheprovider .
4.5 TurningtheCompressorOnandOff
1. MakesuretheconcentratorisOn.Refertothe
concentratorUserManual.
Themaximumoxygenconcentratorowrate
forpatientuseislistedinthetablebelow.
Otherwise,theoxygenleveltothecompressor
willbe<90%andthecompressorwillnotllthe
cylinder.
OxygenConcentrator
MaximumOutputFlow
forPatientUse
5Lmodels
9Lmodels
2.5L/min
5L/min
Patientswithahigherprescribedowshould
notrellacylinderwhilereceivingsupplemental
oxygentreatmentfromtheoxygenconcentrator .
Ifyourconcentratorisnotintherangeslistedin
thetable,refertoyourusermanualorcontact
yourprovider.
2. Settheconcentratorowratetothepatienttooneof
thefollowingsettings.RefertotheconcentratorUser
Manual.
• 2.5L/minorlessfor5Lmodels
• 5L/minorlessfor9Lmodels
Non-Invacareconcentratorsmayrequirelower
outputowratestothepatientinorderto
provide≥90%oxygenlevelstotheHomeFill
compressor .
24
60001253-D
3. Makesurethecylinderisconnectedtothecompressor .
RefertoConnectingtheCylindertotheCompressorin
thissectionofthemanual.
4. PushthepowerswitchAonthecontrolpaneltothe
On(I)position.
5. ExaminethecontrolpanelB.Thefollowingsequence
ofeventsshouldoccur:
a.0-3minutes-TheO
BelowNormal(YELLOW)light
2
C willbelitwhilethecompressorwarmsup.
b.Afterthreeminutes-TheFILLING(GREEN)lightD
willbeonwhilethecylinderislling.
c.TheFULL(GREEN)lightEwillbeonwhenthe
cylinderisnishedlling.ProceedtoSTEP6.
Usage
TheO
BELOWNORMAL(YELLOW)light
2
maycomeoniftheoxygenlevelfromthe
concentratorhasnotyetreachedordropsbelow
90%.Thecompressorwillbeginorresume
llingwhentheoxygenlevelreaches90%.If
lightdoesnotgoOffwithin10minutes,referto
Troubleshootinginthismanual.
IftheA TTENTION(RED)lightandaudiblealarm
F areonrefertoTroubleshootinginthismanual.
6. Removethefullcylinder .RefertoDisconnectingthe
cylinderfromtheCompressorinthissectionofthe
manual.
7. PushthepowerswitchtotheOff(O)position.
Forcylinderoperationanduse,refertothespecic
usermanualforthecylinder/regulatorassembly.
60001253-D
25
Invacare®HomeFill®IIOxygenSystem
5 Maintenance
5.1 MaintenanceInfo
DANGER!
RiskofElectricShock
Thisunitcontainsmovingandhighpressure
components.Therearenouserserviceableparts
insidethecover .T oavoidelectricshock,injury
ordeath:
–DONOTremovethecover .
–Referservicingtoqualiedpersonnel.
WARNING!
–Onlythepartsandproceduresidentiedinthis
sectionaretobeperformedbytheoperator .
Allotherpartsandproceduresaretobe
performedbyaqualiedtechnician.
5.2 WearandTear
Invacarereservestherighttoaskforanyitembackthathas
anallegeddefectinworkmanship.Seethewarrantyinthis
manualforspecicwarrantyinformation.
RefertotheproceduresinthisUserManualforproper
preventativemaintenancescheduleanduseoftheproduct.
Thisisjustageneralguidelineanddoesnotincludeitems
damagedduetoabuseandmisuse.Usefullifeisdependent
uponmanyfactors.
• Alltypesoflters
• Alltypesoftubing
5.3 ServiceLife
Theexpectedservicelifeofthisproductis5yearsor5000
hoursofoperationwhenusedinaccordancewiththesafety
instructions,maintenanceintervalsandcorrectuse,stated
inthismanual.Theeffectiveservicelifecanvaryaccording
tofrequencyandintensityofuse.Maintenanceshouldbe
doneatleasteveryweek.
5.4 CleaningtheCompressorFilter
WARNING!
–PushthepowerswitchtotheOff(O)position
andunplugthecompressorbeforecleaning.
–DONOToperatethecompressorwithoutthe
lterinstalled.
CAUTION!
–DONOTunplugbypullingoncord.T ounplug,
grasptheplugnotthecord,otherwisedamage
mayoccur.
Allcompressorshaveonelteronthebackofthe
cabinet.
Performthisprocedureatleastonceaweek.
Normalwearandtearitemsandcomponentsforthis
productarelistedbelow.
26
60001253-D
1. UnplugthecompressorA.Grasptheplugnotthecord.
2. RemovethelterB.
3. CleanthelterwithavacuumcleanerCorwashin
warmsoapywaterandrinsethoroughlyD.
4. Drythelterthoroughlybeforereinstallation.
5.5 CleaningtheCabinet
DANGER!
RiskofElectricShock
Thisunitcontainsmovingandhighpressure
components.Therearenouserserviceableparts
insidethecover .T oavoidelectricshock,injury
ordeath:
–DONOTremovethecover .
–Referservicingtoqualiedpersonnel.
Maintenance
1. Cleanthecabinetwithamildhouseholdcleanerand
non-abrasiveclothorsponge.
60001253-D
27
Invacare®HomeFill®IIOxygenSystem
5.6 ReplacingtheFuse 1. Unplugthepowercordfromtheelectricaloutlet.
2. RemovethepowercordfromthepowerinletAonthe
compressorbaseC.
3. PulloutthefusedrawerBatthebottomofthepower
inlet.
4. Usingafusepullerorasmallatbladescrewdriver ,
removethefrontfuseinsidethefusedrawerand
discard.
Thefusedrawercontainsspacefortwofuses.
Thefrontlocationisfortheoperationalfuse.
Theunitissuppliedwiththefrontfuseonly.
Therearlocationisforanoptionalsparefuse.
Whenreplacingthefrontfuse,ensurethatthere
isasparefusewiththesametypeandrating
inthefusedrawer.
5. InstallreplacementfuseinfrontfuseDslot.
Usea5mmx20mm,timelag,lowbreaking
capacity,250V ,2Amp,IEC60127styleonly
(Invacarepartnumber1099820).
6. Pushfusedrawerbackintothepowerinlet.
7. Installpowercordintothepowerinletonthe
compressorbase.
8. Plugpowercordintoelectricaloutlet.
9. TurnthepowerswitchtotheOn(
toseethattheunitpowersupproperly .
10.Iftheunitdoesnotcomeon,turnthepowerswitch
off,unplugthepowercord,andcallyourprovideror
Invacareforservice.
)positionandcheck
28
60001253-D
6 AfterUse
6.1 Disposal
This product has been supplied from an environmentally
aware manufacturer that complies with the Waste Electrical
and Electronic Equipment (WEEE) Directive 2012/19/EU.
This product may contain substances that could be harmful
to the environment if disposed of in places (landlls) that
are not appropriate according to legislation.
The 'crossed out wheelie bin' symbol is placed on this
product to encourage you to recycle wherever possible.
Follow local governing ordinances and recycling plans
regarding disposal of the device or components normally
used in operation. The device does not generate waste or
residue in operation. DO NOT dispose of the device in the
normal waste stream. Any accessories not part of the device
MUST be handled in accordance with the individual product
marking for disposal.
Please be environmentally responsible and recycle this
product at its end of life. Contact Invacare or your service
provider for product recycling information.
AfterUse
60001253-D
29
Invacare®HomeFill®IIOxygenSystem
7 Troubleshooting
7.1 IndicatorLights
IndicatorLightColor CompressorStatus Effect
None
O2BELOWNORMAL-YELLOW
FILLING-GREEN
CompressorisOff. O
CompressorisOnandwarmingup(0-3
minutesafterpowerisOn).
O2fromconcentratoris<90%(anytime
afterthreeminutewarmup).
CompressorisOnandcylinderislling.
O2fromconcentratoris>90%(anytime
afterthreeminutewarmup).
owingtouserifconcentratorison.
2
Cylinderllingnotstarted.
O
owingtouserifconcentratorisOn.
2
Cylinderllingstops,willrestartwhen
O
>90%.
2
O
owingtouserifconcentratorisOn.
2
Cylinderislling.
O
owingtouserifconcentratorisOn.
2
30
60001253-D
IndicatorLightColor CompressorStatus Effect
FULL-GREEN
ATTENTION-RED(AudibleAlarm)
CompressorisOn.Cylinderisnished
lling(full)-removecylinder(FULLlight
isOnin>10mins.)
or
Cylinderisnotattachedproperly-check
connection(FULLlightisOnin<10
mins.).
CompressorisOn.CylinderNOTlling.
Causes:
Troubleshooting
Cylinderllingstops.
Cannotllanothercylinderuntil
compressoristurnedOffandOnagain.
O
owingtouserifconcentratorisOn.
2
Compressordoesnotbegintollor
stopslling.
60001253-D
Fullcylinderattachedwhenunitturned
On-replacewithemptycylinder
or
Internalfailure-contactprovider.
UnitmustbeturnedOffandOnbefore
llingcanoccur.
O
owingtouserifconcentratorisOn.
2
31
Invacare®HomeFill®IIOxygenSystem
7.2 Troubleshooting
Problem Solution
NoindicatorlightsonwhencompressorturnedOn.
REDLightOn 1. Checkthecylindergaugetoseeifitisfull:
1. TurncompressorOff(O).
2. Checkthatpowercordisproperlypluggedinto
compressorandwalloutlet.
3. TurnthecompressorOn(I).
4. Cleanairlters.RefertoCleaningtheCompressorFilter
intheMaintenancesectionofthemanual.
5. IfperformingSteps1-4doesnotcauseanyindicatorlights
tocomeon,contactyourhomehealthcareprovideror
Invacareforservice.
• Cylinderfull-Removethecylinderandturntheunit
Off.PerformtheCylinderPrellInspectioninthe
Usagesectionofthemanualwithacylinderthatis
notfull.
• Cylindernotfull-ProceedtoStep2.
2. Checkthatthecylinderisconnectedproperly .
• TurnthecompressorOff(O).
• Removethecylinderandreinstall.
• TurnthecompressorOn(I).Ifcylinderisnotfull
andisconnectedproperly,thecompressormayhave
internalfailure.Contactyourhomehealthcare
providerorInvacare.
32
60001253-D
Problem Solution
CompressorcyclesonandOff
or
O
BelowNormal
2
(yellow)andlling(green)lightsarealternating
or
O2belownormal(yellow)lightgoesOnandOff
O
belownormal(YELLOW)lightstaysOn
2
Troubleshooting
1. Checkthattheconcentratorliterowsettingisat
orbelowtheowsettinglistedin4.5Turningthe
CompressorOnandOff,page24 .
Ifnecessary,adjusttheliterowsetting(refertothe
concentratorUsermanual).Waitatleast20minutesfor
theGREENlighttoappear .Ifitdoesnotappear,proceed
tostep2.
2. Theconcentratormayneedservice,contactyourhome
healthcareproviderorInvacare.
1. Ensuretheconcentratorhaswarmedupforatleast
20minutes.Ifithasnotbeen20minutes,turnthe
compressorOffuntiltheconcentratorwarmsup.Ifthe
YELLOWlightstillremainsonproceedtostep2.
2. Inspecttheconnectionbetweenthecompressorandthe
concentratorfordamageorpinching.
• Ifdamaged,replace.
• Ifpinched,straightentubing.Turnthecompressor
Off(O)foratleast10seconds.T urnthecompressor
On(I).IftheYELLOWlightstillremainsonafter10
minutes,proceedtostep3.
3. Theconcentratormayneedservice,contactyourhome
healthcareproviderorInvacare.
60001253-D
33
Invacare®HomeFill®IIOxygenSystem
Problem Solution
Filltimesexceedmaximumas
recommendedlltimesORGREENllinglightstaysOn.
1. EnsurethecylinderregulatorissettoOff(O).
2. Checktherecommendedlltimesshownintheindividual
cylinder’smanual.
3. Checkthatthecylinderisconnectedproperly.Turnthe
compressorOff(O).Removethecylinderandreinstall.
TurnthecompressorOn(I).
4. Checkthecompressor/concentratorlineforproper
connectionand/ordamage(i.e.-leaksorkinks).
5. Checkthattheconcentratorliterowsettingisat
orbelowtheowsettinglistedin4.5Turningthe
CompressorOnandOff,page24 .Ifnecessary,adjustthe
literowsetting(refertotheconcentratorUsermanual).
IftheGREENllinglightdoesnotgoOff,thecompressor
and/orconcentratormayneedservice.Contactyour
homehealthcareproviderorInvacare.
ExcessivenoisewhencompressorisOn 1. CallhomehealthcareproviderorInvacareforservice.
HissingsoundfromRegulatorwhenowselectorisOff 1. TurnowselectorOff.
2. Ifhissingsoundscontinue,callhomehealthcareprovider
orInvacare.
RegulatorisOnbutO2isnotbeingdelivered 1. Rellcylinder
2. Checkthatcannulaisstraightandnotpinched.If
damaged,replace.
3. CallhomehealthcareproviderorInvacareforservice.
Refertothecylinder’smanualforproperlltimes.
Fillingtimesareapproximateandmayvaryon
environmentalconditions.
34
60001253-D
TechnicalData
8 TechnicalData
8.1 Technical Description
The HomeFill II Oxygen System (IOH200AW) is an
electro-mechanical prescription system for use in the home.
Its intended function and use is to supply pressurized oxygen
to ll I nvacare p roprietary g as c ylinders f or ambulatory
use while the patient receives therapeutic oxygen from an
oxygen concentrator. The IOH200AW gas system is designed
for use with oxygen concentrators that meet the input
specications f or pressure, ow, an d oxygen purity.
The HomeFill II Oxygen system consists of IOH200AW gas
compressor, connection tubing, and the Invacare supplied
portable gas cylinders with the integral ow r egulator and
interlocking ll tting.
The IOH200AW compressor consists of a multistage
compressor, internal reservoir, control electronics with
an integral oxygen concentration sensor, interlocking ll
tting, i nlet tting, an d dr ive mo tor. Th e compressor
operation is controlled by the electronics assembly. The
motor/compressor is allowed to ll a c ylinder o nly w hen the
oxygen concentration of the inlet gas stream is greater than
90% and the cylinder pressure is not already greater than
1500 PSI (10342 kPa). The motor/compressor ll operation is
completed when the gas in a cylinder reaches the maximum
ll p ressure o f 2 000 P SI ( 13786 kPa).
cylinderalsocontainsanintegralowregulatorwithxed,
selectableowsettings.
8.2 CylinderFillTimes
CYLINDERFILLING
TIMES
NORMAL
MAXIMUM
Typicallltimesassumetheinputrequirementsare
met.
Fillingtimesmayvaryonenvironmentalconditions.
ConcentratorFlowRatetoPatient:
5Lmodelsupto2.5L/min.
9Lmodelsupto5L/min.
HF2PCE9A HF2PCE6
2hr20min 1hr25min
2hr45min 1hr50min
The IOH200AW compressor can ll c ylinders f or portable
use by ambulatory patients. The Invacare supplied portable
gas cylinders have an interlocking ll tting th at ma tes only
with the Invacare IOH200AW compressor assembly. The gas
60001253-D
35
Invacare®HomeFill®IIOxygenSystem
8.3 Specications
RegulatoryListings
RegulatoryListings:
InputSpecications
InputPressureRequired:
InputFlowRequired:
InputOxygenPurity: >90%O
GeneralSpecications
ElectricalRatings: 230V~,1.0A,50Hz
OperatingEnvironmental
Conditions:
AtmosphericPressure: 700to1060hPa
StorageandTransport:
Width:
Height:
Depth:
EN/IEC60601–1
EN/IEC60601–1–2
5–25psi(34.5–172.4
kPa)
2L/min
2
41to104°F(5to40°C)at
15to60%non-condensing
humidity
-25to158°F(-23to
70°C)at15to95%
non-condensinghumidity.
20¼inches(51.44cm)
15inches(38.1cm)
16inches(40.64cm)
Weight:
ShippingWeight:
33lbs(14.97kg)
39lbs(17.69kg)
36
60001253-D
8.4 ElectromagneticCompliance(EMC)
Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
TechnicalData
Emissionstest
RFemissions
CISPR11
RFemissions
CISPR11
Harmonic
emissions
IEC61000-3-2
Compliance Electromagneticenvironment-guidance
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissions
GroupI
areverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic
equipment.
ClassB
ClassA
TheDeviceissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Voltageuctuations
/ickeremissions
Complies
IEC61000-3-3
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic
discharge(ESD)
IEC61000-4-2
IEC60601testlevel Compliancelevel Electromagneticenvironment–guidance
±2,4,8kVcontact
±2,4,8kVcontact
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
±2,4,8,15kVair
±2,4,8,15kVair
relativehumidityshouldbeatleast30%.
60001253-D
37
Invacare®HomeFill®IIOxygenSystem
ElectricalFast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltagedips
IEC61000-4-11
Voltage
Interruptions
IEC61000-4-11
Powerfrequency
(50/60Hz)
magneticeld
IEC61000-4-8
±2kVmains,100kHz ±2kVmains,100kHz
±1kVline(s)toline(s)
±2kVline(s)toground
0%U T;1/2cycle@0,
45,90,135,180,225,
270,and315
and
0%U T;1cycle
and
70%U T;25/30cycles
Singlephase:at0°
0%U
;250/300cycle 0%U T;250/300cycle
T
±1kVline(s)toline(s)
±2kVline(s)toground
0%U T;1/2cycle@0,
45,90,135,180,225,
270,and315
and
0%U T;1cycle
and
70%U T;25/30cycles
Singlephase:at0°
30A/m,50/60Hz 30A/m,50/60Hz
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheDevicerequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthat
theDevicebepoweredfromanun-interruptible
powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication
ofthetestlevel.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheDevicerequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthat
theDevicebepoweredfromanun-interruptible
powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication
ofthetestlevel.
Powerfrequencymagneticeldsshouldbeat
levelscharacteristicofatypicallocationinatypical
commercialorhospitalenvironment.
38
60001253-D
ConductedRF
IEC61000-4-6
3Vrms,6Vrms 3Vrms,6Vrms
TechnicalData
PortableandmobileRFcommunicationsequipment
shouldbeusednoclosertoanypartofthe
Deviceincludingcables,thantherecommended
separationdistancecalculatedfromtheequation
applicabletothefrequencyofthetransmitter .
Recommendedseparationdistance:
d=1.2√P 150kHzto80MHz
RadiatedRF
IEC61000-4-3
10V/m 10V/m
d=0.35√P
d=0.7√P
80MHzto800MHz
800MHzto2,5GHz
wherePisthemaximumoutputpowerrating
ofthetransmitterinwatts(W)accordingto
thetransmittermanufactureranddisthe
recommendedseparationdistanceinmetres(m).
FieldstrengthsfromxedRFtransmitters,as
determinedbyanelectromagneticsitesurvey,
shouldbelessthanthecompliancelevelineach
frequencyrange.
b
Interferencemayoccurinthevicinityofequipment
markedwiththefollowingsymbol:
a
Field strengths from xed t ransmitters, s uch a s b ase s tations for r adio ( cellular/cordless) t elephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
60001253-D
a
39
Invacare®HomeFill®IIOxygenSystem
measuredeldstrengthinthelocationinwhichtheDeviceisusedexceedstheapplicableRFcompliancelevelabove,theDevice
shouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary ,
suchasreorientingorrelocatingtheDevice.
b
Overthefrequencyrange150kHzto80MHz,eldstrengthsshouldbelessthan[V1]V/m.
At80MHzand800MHz,thehigherfrequencyrangeapplies.
RecommendedseparationdistancesbetweenportableandmobileRFcommunicationsequipmentandtheDevice
TheDeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.The
customerortheuseroftheDevicecanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetween
portableandmobileRFcommunicationsequipment(transmitters)andtheDeviceasrecommendedbelow ,accordingtothe
maximumoutputpowerofthecommunicationsequipment
Separationdistanceaccordingtofrequencyoftransmitter[m]
Ratedmaximumoutput
oftransmitter[W]
150kHzto80MHz
d=1.2√P
80MHzto800MHz
d=0.35√P
800MHzto2,5GHz
d=0.7√P
0.01 0.12 0.04 0.07
0.1 0.37 0.11 0.22
1 1.17 0.35 0.70
10 3.69 1.11 2.21
100 11.67 3.50 7.00
40
60001253-D
TechnicalData
Fortransmittersratedatamaximumoutputpowernotlistedabovetherecommendedseparation,distancedinmeters(m)
canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter ,wherePisthemaximumoutputpower
ratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.
At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreection
fromstructures,objectsandpeople.
60001253-D
41
Notes
Belgium&Luxemburg
Invacarenv ,Autobaan22,B-8210Loppem•Tel:(32)(0)50831010•Fax:(32)(0)508310
11•belgium@invacare.com•www.invacare.be
Danmark
InvacareA/S,Sdr.Ringvej37,DK-2605Brøndby•Tel:(45)(0)36900000•Fax:(45)(0)36900001
•denmark@invacare.com•www.invacare.dk
Deutschland
InvacareGmbH,Alemannenstraße10,D-88316Isny•Tel:(49)(0)75627000•Fax:(49)(0)75627
0066•kontakt@invacare.com•www.invacare.de
EasternEurope,Middle
East&CIS
España
InvacareEUExport•Kleiststraße49•D-32457PortaWestfalica•Germany•T el:(49)5731754540•
Fax:(49)5731754541•webinfo-eu-export@invacare.com•www.invacare-eu-export.com
InvacareSA,c/Arenys/n,PolígonIndustrialdeCelrà,E-17460Celrà(Girona)•Tel:(34)(0)9724932
00•Fax:(34)(0)972493220•contactsp@invacare.com•www .invacare.es
France
InvacarePoirierSAS,RoutedeStRoch,F-37230Fondettes•Tel:(33)(0)247626466•Fax:(33)(0)2
47421224•contactfr@invacare.com•www.invacare.fr
Ireland InvacareIrelandLtd,Unit5SeatownBusinessCampus•SeatownRoad,Swords,CountyDublin–
Ireland•Tel:(353)18107084•Fax:(353)18107085•ireland@invacare.com•www .invacare.ie
Italia
InvacareMeccSans.r.l.,ViadeiPini62,I-36016Thiene(VI)•Tel:(39)0445380059•Fax:(39)0445
380034•italia@invacare.com•www.invacare.it
Nederland
InvacareBV,Galvanistraat14-3,NL-6716AEEde•Tel:(31)(0)318695757•Fax:(31)(0)318695758
•nederland@invacare.com•csede@invacare.com•www .invacare.nl
Norge
InvacareAS,Grensesvingen9,Postboks6230,Etterstad,N-0603Oslo•Tel:(47)(0)22579500•Fax:
(47)(0)22579501•norway@invacare.com•island@invacare.com•www.invacare.no
Österreich
InvacareAustriaGmbH,HerzogOdilostrasse101,A-5310Mondsee•T el:(43)623255350•Fax:
(43)623255354•info@invacare-austria.com•www.invacare.at
Portugal
InvacareLda•RuaEstradaVelha,949,P-4465-784LeçadoBalio•T el:(351)(0)225105946/47•Fax:
(351)(0)225105739•portugal@invacare.com•www.invacare.pt
Sverige
InvacareAB•Fagerstagatan9•S-16353Spånga•Tel:(46)(0)87617090•Fax:(46)(0)876181
08•sweden@invacare.com•www.invacare.se
Suomi
CampMobility•Patamäenkatu5,33900Tampere•Tel:09-35076310•info@campmobility.•
www.campmobility .
Schweiz/Suisse/Svizzera
InvacareAG•Benkenstrasse260•CH-4108Witterswil•Tel.:(41)(0)614877080•Fax.:(41)(0)61
4877081•switzerland@invacare.com•www.invacare.ch
UnitedKingdom
InvacareLimited,PencoedT echnologyPark,Pencoed,BridgendCF355HZ•T el:(44)(0)1656776222•
Fax:(44)(0)1656776220•UK@invacare.com•www.invacare.co.uk
EC
EU Representative
REP
Invacare GmbH
Am Achener Hof 8
88316 Isny
Germany
Manufacturer
60001253-D 20 20-0 6-30
In vacareCorporation
OneInvacareWay
Elyria,OhioUSA
440-329-6000
800-333-6900
TechnicalServices
440-329-6593
800-832-4707
www.invacare.com
MakingLife’sExperiencesPossible®
Invacare Corporation
One Invacare Way
USA
Elyria, Ohio 44035
USA
44035