Invacare IOH200AW User guide

Invacare®HomeFill®IIOxygen System
ModelIOH200AW
enCompressor
UserManual
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor futurereference.
Contents
1General.........................................4
1.1Symbols.....................................4
1.2IntendedUse.................................5
1.3IndicationsForUse.............................6
1.4ReplacementParts.............................6
1.5Contraindications...............................6
1.6Warranty.....................................6
2Safety..........................................7
2.1GeneralGuidelines.............................7
2.2LabelLocations................................10
2.3ProviderInformation............................12
2.4Double-InsulatedProducts........................12
2.5EMCInformation...............................12
3Setup...........................................14
3.1Unpacking....................................14
3.2CompressorInitialSetupChecklist..................14
3.3ChoosingaLocation............................15
3.4TransportingtheCompressor......................16
3.5InstallingthePowerCord........................16
3.6ConnectingtheCompressortotheConcentrator.......17
3.6.1ConnectingtheCompressortoInvacare
Concentrators..............................17
3.6.2ConnectingtheCompressortoNon-Invacare
Concentrators..............................17
4Usage..........................................18
4.1CompressorOperationChecklist...................18
4.2CylinderPrellInspection........................19
4.3ConnectingtheCylindertotheCompressor...........21
4.4DisconnectingtheCylinderfromtheCompressor.......22
4.5TurningtheCompressorOnandOff................24
5Maintenance.....................................26
5.1MaintenanceInfo..............................26
5.2WearandTear................................26
5.3ServiceLife...................................26
5.4CleaningtheCompressorFilter....................26
5.5CleaningtheCabinet............................27
5.6ReplacingtheFuse.............................28
6AfterUse........................................29
6.1Disposal.....................................29
7T roubleshooting...................................30
7.1IndicatorLights................................30
7.2Troubleshooting...............................32
8T echnicalData....................................35
8.1T echnicalDescription............................35
8.2CylinderFillTimes..............................35
8.3Specications.................................36
8.4ElectromagneticCompliance(EMC).................37
Invacare®HomeFill®IIOxygenSystem

1General

1.1Symbols

SymbolsintheManual
Signalwordsareusedinthismanualandapplytohazards orunsafepracticeswhichcouldresultinpersonalinjuryor propertydamage.Seetheinformationbelowfordenitions ofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin deathorseriousinjury.
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin deathorseriousinjury.
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin propertydamageorminorinjuryorboth.
IMPORTANT
–Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformation forefcient,trouble-freeuse.
SymbolsontheProduct
Thefollowingsymbolsarefoundonthecompressor, compressorpackagingortheoxygencylinder .
ReadtheManual(Compressor)
Thecolorofthesymbol backgroundisblueonproduct labels.
ReadtheManual(Cylinder)
ElectricalHazard
AlternatingCurrent
TypeBFequipment—AppliedParts
Powerswitch“ON”
Powerswitch“OFF”
DONOTsmoke
4
60001253-D
General
NoOpenFlame
ClassII,DoubleInsulation.Referto DoubleInsulatedproductsinthe Safetysectionofthemanual.
Keepdryintransportorstorageand duringuse
Protectedagainstsolidforeignobjects of12.5mmdiameterandgreater.
Protectedagainstverticallyfalling waterdrops.
Fuse
Oxidizer
Thecolorofthesymbol backgroundisyellowon productlabels.
Non-ammablegas,class2type (oxygen)
Thecolorofthesymbol backgroundisgreenon productlabels.
TransportandStorageT emperature
Transport and Storage Humidity
Recycle this product. Refer to Disposal of Equipment and Accessories in the After Use section of the manual.
Complies With Applicable EU Directives Including Medical Device Directive (MDD) (93/42/EEC)
Authorized Representative in the European Community
Manufacturer

1.2IntendedUse

TheInvacareHomeFillcompressorwasdesignedtobe usedonlywithoxygenconcentratorsthatmeettheinput specicationslistedin8.3Specications,page36.
DONOTusetheInvacareHomeFillcompressorwith anythingotherthananoxygenconcentratorthatmeetsthe compressorinputrequirements.
TheintendeduseofanInvacareHomeFillcompressorisan accessorydevicetoanoxygenconcentratorthatenables thepatienttorellcustomportableoxygencylindersfor personaluse.
60001253-D
5
Invacare®HomeFill®IIOxygenSystem
Thisproductisintendedforsinglepatientuseintheindoor homecareenvironmentonlyandisnottobeusedforbulk cylinderrelling.
Nospecicproductknowledgeortrainingisrequired, otherthanwhatiscontainedinthismanual,tooperate theproduct.
Itisnotintendedforsupplyingoxygentherapyduring relling.
Thisproductisintendedtobeusedbyorunderadult supervisiononlyafterreadingandunderstandingthe instructionsandwarningsofthisusermanual.
WARNING!
–Theuseofthisdeviceislimitedtorellinga
specicpatient’sownoxygencylinders.The designatedcylindersMUSTbeusedONL Yby thespecicoxygenpatientandarenotto bedistributedtoanyotherindividualforany purpose.
–Nootheruseofthisdeviceisindicated.

1.3IndicationsForUse

Indicatedtoenablethepatienttorellcustomportable oxygencylindersforpersonaluse.
Thesepartsarebasemodels.PleasecontactInvacare oryourproviderforcountryspecicpartnumbers.

1.5Contraindications

Therearenoknowncontraindications.

1.6Warranty

Termsandconditionsofthewarrantyarepartofthegeneral termsandconditionsparticulartotheindividualcountriesin whichthisproductissold.

1.4ReplacementParts

ThefollowingreplacementpartsareavailablefromInvacare:
SupplyLine,Partnumber1105177
PowerCord,Partnumber1085605
6
60001253-D
Safety

2Safety

2.1GeneralGuidelines

DANGER! RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath, injuryordamage.Thissectioncontainsimportant informationforthesafeoperationanduseofthis product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,providerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–Inordertoensurethesafeinstallation,
assemblyandoperationoftheHomeFill CompressortheseinstructionsMUSTbe followed.
–THEINFORMA TIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–DONOTSMOKEwhileusingthisdevice. –DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonthecompressor
unlessrecommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Keepallmatches,lightedcigarettesorother
sourcesofignitionoutoftheroominwhich thiscompressorislocatedandawayfrom whereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andcompressor
outfromundersuchitemsasblankets, bedcoverings,chaircushions,clothing,and awayfromheatedorhotsurfacesincluding spaceheaters,stoves,andsimilarelectrical appliances.
60001253-D
7
Invacare®HomeFill®IIOxygenSystem
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–DONOTallowoilfromyourhandsorother
sourcestocomeintocontactwiththe compressorcouplingorcylinderttings.These substancesmaybecomeammableinthe presenceofoxygenandcauseinjury.Avoid touchingthecylinderttingwheneverpossible. Exampleofitemstoavoidarehandcreamsor lotions,cookingoils,suntanoilsorsunblocks andsimilaroilyproducts.Ifyouusethese typesofproductsInvacaresuggestsyouwash yourhandspriortousingtheHomeFillunit. Ifthecylinderttingshouldcomeincontact withsuchsubstancesitshouldbecleanedoff withadampclothpriortoconnectingittothe compressor .
–NEVERoilorlubricatethecompressorcoupler
orcylinderttings.NEVERuseanypenetrating oilorlubricantsuchasWD-40®or3-in-1 Oil®.Ifconnectiondifcultyisexperienced verifythatthecylinderttingisbeinginserted straightdownintothecompressorcoupler .If continueddifcultyisexperiencedcontactyour supplierforassistance.
DANGER! RiskofElectricShock
Thisunitcontainsmovingandhighpressure components.Therearenouserserviceableparts insidethecover .T oavoidelectricshock,injury ordeath:
–DONOTremovethecover . –Referservicingtoqualiedpersonnel.
WARNING! RiskofSeriousInjuryorDamage
Useofnon-Invacareaccessoriesmayresultin seriousinjuryordamage.
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunctionwith Invacareaccessories.Accessoriesdesignedby othermanufacturershavenotbeentestedby Invacareandarenotrecommendedforuse
withInvacareproducts. –Nomodicationsofthisequipmentisallowed. –DONOTusenon-Invacareaccessories. –T oobtainInvacareaccessories,contactInvacare
byphoneoratwww .invacare.com
8
60001253-D
Safety
WARNING! Riskofinjuryordamage
Blockingtheairowintoandoutofthe compressormaycauseinjuryordamage.
–NEVERblocktheairopeningsoftheproduct
orplaceitonasoftsurface,suchasabedor couch,wheretheairopeningmaybeblocked.
–Keeptheopeningsfreefromlint,hairand
similarforeignitems.
WARNING! RiskofInjuryorDamage
Adamagedproductmaymalfunctionandcause injuryordamage.T oavoidinjuryordamage:
–DONOTgetthecompressorwet. –Ifthecompressorhasadamagedcordorplug,
ifitisnotworkingproperly,ifithasbeen droppedordamaged,orsubmersedinwater , callaqualiedtechnicianforexaminationand repair .
WARNING!
–DONOTusethecompressoroutdoors.Indoor
useonly. –Ifanyleakageoftheoxygencylinderisdetected
DONOTattempttousethecylinder.T urnthe
owselectorknobontheregulatorto“Off”.If
leakagepersists,placethecylinderoutdoors
andnotifyyourHomeHealthCareProvideror
ServiceRepresentativeofthiscondition. –ChildrenshouldALWA YSbesupervisedaround
theHomeFillunit.Whenusedwiththe
HomeFilltableorsimilarelevatedsupport
surfaces,achildcouldpulloveranysuchtable
causingtheunittofall.Failuretoobserve
thiswarningmayresultinpersonalinjuryor
productdamage.
WARNING! Riskofinjuryordamage
TheHomeFillisequippedwithahighpressure reliefvalvetoensuretheuser’ssafety .When activated,thissafetyfeaturewillmakean extremelyloudnoise.Toavoidinjuryordamage:
–Ifthisnoiseoccurs,turntheunitOffand
contactyourHomeHealthCareProvideror
ServiceRepresentative.
60001253-D
9
Invacare®HomeFill®IIOxygenSystem
WARNING! Riskofdeath,injuryordamage
Ifimproperlypositionedandsecured,thepower cordandinterconnecthosecancauseinjurydue totripping,falling,andstrangulation.Product damagemayalsooccur.
–Properlystoreandpositionelectricalcords
and/ortubingtopreventtrippingand strangulationhazards.
–Avoidpositioningpowercordsacrossareasof
hightrafc.
–Beforemovingorrepositioningeitherthe
compressororconcentrator ,ALWAYSdisconnect theACpowercordsandtheinterconnecthose betweenthecompressorandconcentrator .
CAUTION!
Statutorylawrestrictsthisdevicesaletooron theorderofamedicalpractitionerlicensedbya governmentalagencywherehe/shepractices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.

2.2LabelLocations

A
10
60001253-D
Safety
B
C
D
E
60001253-D
11
Invacare®HomeFill®IIOxygenSystem

2.3ProviderInformation

Invacarerecommendsleavingafullcylinderof oxygenwiththepatientaftersettinguptheHomeFill compressor .

2.5EMCInformation

BeforeleavingaHomeFillCompressorwithauser,the followingchecklistMUSTbecompleted:
qMakesurethepressuretestoncylinderhasnotexpired. qChecktheEMCsectionforoperationalinformationand
possibleRFissues.RefertoEMCInformation.
qInstructtheuseronthesafeoperationofthe
concentratorandreviewallwarnings.
qInstructtheuseronthesafeoperationoftheHomeFill
compressorandreviewallwarnings.
qLeaveafullcylinderofoxygenwiththeuser. qLeaveacopyofallwarningsandllinginstructionswith
theuser.

2.4Double-InsulatedProducts

CAUTION!
–DONOTunplugthepowercordbypullingon
thecord.T ounplug,grasptheplugnotthe cord,otherwisedamagemayoccur.
Inadouble-insulatedproduct,twosystemsofinsulation areprovidedinsteadofgrounding.Nogroundingmeansis requiredonadouble-insulatedproduct,norisameansfor groundingtobeaddedtotheproduct.Adouble-insulated productismarkedwiththewords“DOUBLEINSULATION”or “DOUBLEINSULATED”orthesymbol:
WARNING! RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamage frominterferencewithwirelessequipment:
–Keepthecompressoratleast9.8ft(3.0m)
awayfromwirelesscommunicationequipment
suchaswirelesshomenetworkdevices,mobile
phones,cordlessphonesandbasestations,
walkie-talkies,etc.
WARNING!
–Useofaccessoriesandcablesotherthanthose
speciedorprovidedbyInvacarecouldresult
inincreasedelectromagneticemissionsor
decreasedelectromagneticimmunityofthis
equipmentandresultinimproperoperation.
MedicalElectricalEquipmentneedstobeinstalledandused accordingtotheEMCinformationinthismanual.
Thisequipmenthasbeentestedandfoundtocomply withEMClimitsspeciedbyIEC/EN60601-1-2forType BFequipment.Theselimitsaredeterminedtoprovide reasonableprotectionagainstelectromagneticinterference inatypicalhomehealthcareenvironment.Useofthis equipmentinanenvironmentwithhigherlevelsof interferencecouldaffecttheaccuracyoftheoxygen
12
60001253-D
concentrationinthecylinder.Thisequipmentisnotintended foruseinaprofessionalhealthcareenvironment.
PortableandmobileRFcommunicationsequipmentcan affecttheoperationofthisequipment.Thecompressor maybestackedontoponanInvacareHomeFillcompatible oxygenconcentratorduringsetup.Thecompressoroperation shouldbeobservedtoverifynormalcylinderllingineither situation.
Otherdevicesmayexperienceinterferencefromeven thelowlevelsofelectromagneticemissionspermitted bytheabovestandard.T odetermineiftheemissions fromthecompressorarecausingtheinterference,turn thecompressorOff.Iftheinterferencewiththeother deviceoperationstops,thenthecompressoriscausing theinterference.Insuchrarecases,interferencemaybe reducedorcorrectedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetween thedevices.
Connecteitherlinepowereddevicetoadifferent electricalpowercircuit.
Foradditionalinformation,referto8.4Electromagnetic Compliance(EMC),page37.
Safety
60001253-D
13
Invacare®HomeFill®IIOxygenSystem

3Setup

3.1Unpacking

DESCRIPTION
CompressorAssembly
PackagingAssembly
SupplyLineAssembly
PowerCord
UserManual(notshown)
A
B
C
D
N/A
1.Checkforanyobviousdamagetothecartonorits contents.Ifdamageisevident,notifythecarrier,or yourlocalprovider.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallofthecomponentsfromthecarton.
UnlessthecompressoristobeusedIMMEDIATELY , leavethecompressorinitspackagingforstorage untiluseofthecompressorisrequired.

3.2CompressorInitialSetupChecklist

ProvidersONL Y—RefertoDealerInformationinthe Safetysectionofthismanualbeforeperformingthis procedure.
UsersONLY—ContactyourInvacareprovideror Invacareforassistancewiththesetupprocedureand toreportunexpectedoperationorevents.
BeforeoperatingtheHomeFillCompressorforthersttime, completethefollowingchecklist:
qChoosealocationforthecompressor . qTransportthecompressortothedesiredlocation.
14
60001253-D
Setup
qConnectthepowercordtothecompressor. qConnectthecompressortotheconcentrator .

3.3ChoosingaLocation

DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–DONOTSMOKEwhileusingthisdevice. –DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonthecompressor
unlessrecommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Keepallmatches,lightedcigarettesorother
sourcesofignitionoutoftheroominwhich thiscompressorislocatedandawayfrom whereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andcompressor
outfromundersuchitemsasblankets, bedcoverings,chaircushions,clothing,and awayfromheatedorhotsurfacesincluding spaceheaters,stoves,andsimilarelectrical appliances.
60001253-D
Thecompressorshouldbeatleastthreeinches(7.6cm) awayfromwalls,draperiesorfurnituretoensuresufcient airow.
Avoidplacingthecompressornexttoheaters,radiatorsor hotairregisters.
DONOTuseinacloset.
Thecompressorshouldsitonahardatstablesurface,such asatableortheReadyRack.
TheHomeFillReadyRackissoldseparatelyand usedtopositiontheHomeFillcompressorontop ofaPerfecto2orPlatinum®oxygenconcentrator . RefertotheHomeFillReadyRackinstructionsheet partnumbers1148363and1134824formore information.
15
Invacare®HomeFill®IIOxygenSystem

3.4TransportingtheCompressor

WARNING!
–NEVERtransportacompressorwithacylinder
connectedtoorstoredonthecompressor , otherwiseinjuryordamagecanoccur.
CAUTION!
–Whentransportingthecompressor ,becareful
nottobumpthecompressorortheconnectors intoobstacles.Otherwise,damagetothe compressormayoccur .
1.Therearetwomethodstosafelytransportthe compressor:
OneHand-GraspthehandgripAatthebottom rearofthecompressor.
TwoHands-GraspthehandgripsBonthesides ofcompressor.

3.5InstallingthePowerCord

1.Chooseaproperlocation.RefertoChoosingaLocation inthissectionofthemanual.
2.MakesurethepowerswitchisintheOff(O)position.
3.PlugthepowercordintotheelectricconnectorAin thebackofthecompressorB.
4.PlugtheoppositeendofthepowercordCintothe walloutletD.
5.Leaveaminimumof3in(7.6cm)clearancetothe powercordendsfordisconnectionaccess.
Thepowerinletactsasthemainsdisconnection devicetoremoveallpowertotheunitwhenthe linecordisdisconnected.
16
60001253-D
Setup

3.6ConnectingtheCompressortothe Concentrator

3.6.1ConnectingtheCompressortoInvacare Concentrators
1.InsertoneendoftheinterconnecthoseCintothe
outletttingBonthebackoftheconcentratorA.
2.Inserttheoppositeendoftheinterconnecthoseinto
theinletttingDonthesideofthecompressor .
ThemetaltabsEontheconcentratoroutlet ttingandthecompressorinletttingwillpop outwithanaudible“click”whentheendofthe interconnecthoseisproperlyinstalled.
3.6.2ConnectingtheCompressortoNon-Invacare Concentrators
1.Refertotheoxygenconcentratorusermanualoutlet
gasspecicationstoverifycompatibilitywiththe HomeFillcompressorinputgasrequirementslistedin
8.3Specications,page36.
2.Refertotheoxygenconcentratorusermanualfor
connectinginstructions.
60001253-D
17
Invacare®HomeFill®IIOxygenSystem

4Usage

4.1CompressorOperationChecklist

EachtimetheHomeFillCompressorisusedtollacylinder , completethefollowingchecklist:
qEnsuretheconcentratorhasbeenOnforatleastthirty
minutes.RefertotheconcentratorUserManual.
qPerformtheprellinspectiononthecylinder . qConnectthecylindertothecompressor . qPushthecompressorpowerswitchtotheOn(I)
position.
qExaminetheindicatorlightsonthecontrolpanel. qDisconnectandremovethefullcylinder . qPushthecompressorpowerswitchtotheOff(O)
position.
qIfllinganothercylinder,repeatthischecklist.
18
60001253-D
4.2CylinderPrellInspection
WARNING!
–ONL YusecylindersthathavetheInvacareHomeFillIIconnectionandHomeFillIIlabel. –AllcylindersMUSTbeinspectedbeforeattemptingtoll.Otherwise,injuryordamagemayoccur.
ToidentifyaHomeFillIIcompatiblecylinderA,lookforablackstripeBonthetopofthecylinderandlookfortheHomeFillII labelC.
Usage
60001253-D
19
Invacare®HomeFill®IIOxygenSystem
TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeverytenyears.
WARNING! RiskofInjury
Toreducetheriskofinjury:
–DONOTllcylindersthathavenotbeentestedinthepasttenyears.Otherwise,seriouspersonalinjurymay
result.Contactyourequipmentproviderforreplacement.
ExternalExamination
1.Examinetheoutsideofthecylinderforthefollowingconditions,andreplacethecylinderiftheyexist:
Dentsordings
Arcburns
Oilorgrease
Anyothersignsofdamagethatmightcauseacylindertobeunacceptableorunsafeforuse.
2.Examinethecylinderforevidenceofreorthermaldamage.Evidenceincludescharringorblisteringofthepaint,orother protectivecoatingorheatsensitiveindicator.Ifreorthermaldamageisfound,replacethecylinder.
3.Inspectthecylinder/regulatorassemblyforthefollowingandreplaceiffound:
Debris,oilorgrease
Noticeablesignsofdamage
Signsofcorrosioninsidethevalve
Signsofexcessiveheatorredamage
20
60001253-D

4.3ConnectingtheCylindertotheCompressor

WARNING!
–NEVERusetoolsofanykindto
connect/disconnectthecylinderandthe compressor .Otherwise,severeinjuryand/or damagemayoccur.
WARNING!
–DONOTdropoxygencylinders.Usetwohands
whenhandling/transportingoxygencylinders. Otherwise,injuryordamagemayoccur .
–DONOTusealiquidleakdetectortotestfor
leaks.
Usage
Fillportcoversnotshownforclarity.
60001253-D
21
Invacare®HomeFill®IIOxygenSystem
1.Priortoeachuse,inspectproductforvisibledamage. RefertotheExternalExaminationsection.DONOTuse ifanydamageisfound.
Ifforanyreason,anylabelbecomesillegibleor lost,contactyourequipmentprovider .
2.Examinethecylindergauge.Ifthecylinderpressureis lessthan1500psig(10342kPa),proceedtoSTEP3.
Forlocationofthecylindergauge,refertothe cylinder’smanual. Ifthecylinderpressureisgreaterthan1500psig (10342kPa),DONOTattempttotopoffthis cylinder.Itmaynotll.
3.SettheregulatorowdialonthecylinderAtoOff. Refertothecylinder’smanual.
4.Removethecylinderandconnectorllportcovers(if present).
5.MomentarilypushDOWNontheouterring(sleeve)B oftheconnectorllportCuntilGREENdotsDare visibletoresettheconnector .
Iftheouterring(sleeve)isintheUPposition (GREENdotsnotvisible),theconnectorllport willnotbeabletoacceptthecylinderllportE. PushingDOWNFmomentarilywillresetthe connectorllport(GREENdotsvisible)toaccept thecylinderllport.
8.Alignthecylinderllportwiththeconnectorllport.
9.PullUPontheouterring(sleeve)oftheconnector llportwhilepushingDOWNonthecylindertocouple thecylinderllportintotheconnectorllport.
Thecylinderisproperlyconnectedwhenan audible“click”isheard.
4.4DisconnectingtheCylinderfromthe
Compressor
6.Graspthecylinderintheareabehindthecylinder llport.
7.PositionthecylinderinthecompressorcradleG.
22
60001253-D
Usage
WARNING!
–NEVERusetoolsofanykindto
connect/disconnectthecylinderandthe compressor .Otherwise,severeinjuryand/or damagemayoccur.
1.PressthecompressorpowerswitchAtotheOff(O) position.
2.GraspthecylinderBintheareabehindthecylinder llportC.
3.Withtheotherhand,grasptheouterring(sleeve)of theconnectorllportEandpushDOWNF.
4.Liftuponthecylindertoremovefromtheconnector llport.
5.Whenthecylinderllportisdisconnectedfromthe connectorllport,releasetheouterring(sleeve)ofthe connectorllportandusetwohandstoremovethe cylinderfromthecompressorcradle.
WARNING!
–Thellportcoverontheconnectorllport
andthecylinderllportMUSTbereplaced afterllingandwhenevernotinuse.If eitherofthellportcoversaremissing contactyourhealthcareproviderbefore using.
60001253-D
23
Invacare®HomeFill®IIOxygenSystem
6.PlacethellportcoversontotheconnectorllportG andcylinderllportH.
Whenusingthepostvalvetypecylinder, attachanoxygenregulatorpermanufacturers instructions.
7.Examinethecylindergaugetoensurecylinderisfull (needlepointingintoGREENarea).Ifthecylinderisnot full,performthefollowing:
a.RepeatConnectingtheCylindertotheCompressor,
inthissectionofthemanual.
b.Ifcylinderisnotfullaftersecondattempttoll,
contacttheprovider .

4.5TurningtheCompressorOnandOff

1.MakesuretheconcentratorisOn.Refertothe concentratorUserManual.
Themaximumoxygenconcentratorowrate forpatientuseislistedinthetablebelow. Otherwise,theoxygenleveltothecompressor willbe<90%andthecompressorwillnotllthe cylinder.
OxygenConcentrator
MaximumOutputFlow forPatientUse
5Lmodels
9Lmodels
2.5L/min
5L/min
Patientswithahigherprescribedowshould notrellacylinderwhilereceivingsupplemental oxygentreatmentfromtheoxygenconcentrator . Ifyourconcentratorisnotintherangeslistedin thetable,refertoyourusermanualorcontact yourprovider.
2.Settheconcentratorowratetothepatienttooneof thefollowingsettings.RefertotheconcentratorUser Manual.
2.5L/minorlessfor5Lmodels
5L/minorlessfor9Lmodels
Non-Invacareconcentratorsmayrequirelower outputowratestothepatientinorderto provide≥90%oxygenlevelstotheHomeFill compressor .
24
60001253-D
3.Makesurethecylinderisconnectedtothecompressor . RefertoConnectingtheCylindertotheCompressorin thissectionofthemanual.
4.PushthepowerswitchAonthecontrolpaneltothe On(I)position.
5.ExaminethecontrolpanelB.Thefollowingsequence ofeventsshouldoccur:
a.0-3minutes-TheO
BelowNormal(YELLOW)light
2
Cwillbelitwhilethecompressorwarmsup.
b.Afterthreeminutes-TheFILLING(GREEN)lightD
willbeonwhilethecylinderislling.
c.TheFULL(GREEN)lightEwillbeonwhenthe
cylinderisnishedlling.ProceedtoSTEP6.
Usage
TheO
BELOWNORMAL(YELLOW)light
2
maycomeoniftheoxygenlevelfromthe concentratorhasnotyetreachedordropsbelow 90%.Thecompressorwillbeginorresume llingwhentheoxygenlevelreaches90%.If lightdoesnotgoOffwithin10minutes,referto Troubleshootinginthismanual. IftheA TTENTION(RED)lightandaudiblealarm FareonrefertoTroubleshootinginthismanual.
6.Removethefullcylinder .RefertoDisconnectingthe cylinderfromtheCompressorinthissectionofthe manual.
7.PushthepowerswitchtotheOff(O)position.
Forcylinderoperationanduse,refertothespecic usermanualforthecylinder/regulatorassembly.
60001253-D
25
Invacare®HomeFill®IIOxygenSystem

5Maintenance

5.1MaintenanceInfo

DANGER! RiskofElectricShock
Thisunitcontainsmovingandhighpressure components.Therearenouserserviceableparts insidethecover .T oavoidelectricshock,injury ordeath:
–DONOTremovethecover . –Referservicingtoqualiedpersonnel.
WARNING!
–Onlythepartsandproceduresidentiedinthis
sectionaretobeperformedbytheoperator . Allotherpartsandproceduresaretobe performedbyaqualiedtechnician.

5.2WearandTear

Invacarereservestherighttoaskforanyitembackthathas anallegeddefectinworkmanship.Seethewarrantyinthis manualforspecicwarrantyinformation.
RefertotheproceduresinthisUserManualforproper preventativemaintenancescheduleanduseoftheproduct.
Thisisjustageneralguidelineanddoesnotincludeitems damagedduetoabuseandmisuse.Usefullifeisdependent uponmanyfactors.
Alltypesoflters
Alltypesoftubing

5.3ServiceLife

Theexpectedservicelifeofthisproductis5yearsor5000 hoursofoperationwhenusedinaccordancewiththesafety instructions,maintenanceintervalsandcorrectuse,stated inthismanual.Theeffectiveservicelifecanvaryaccording tofrequencyandintensityofuse.Maintenanceshouldbe doneatleasteveryweek.

5.4CleaningtheCompressorFilter

WARNING!
–PushthepowerswitchtotheOff(O)position
andunplugthecompressorbeforecleaning.
–DONOToperatethecompressorwithoutthe
lterinstalled.
CAUTION!
–DONOTunplugbypullingoncord.T ounplug,
grasptheplugnotthecord,otherwisedamage mayoccur.
Allcompressorshaveonelteronthebackofthe cabinet.
Performthisprocedureatleastonceaweek.
Normalwearandtearitemsandcomponentsforthis productarelistedbelow.
26
60001253-D
1.UnplugthecompressorA.Grasptheplugnotthecord.
2.RemovethelterB.
3.CleanthelterwithavacuumcleanerCorwashin warmsoapywaterandrinsethoroughlyD.
4.Drythelterthoroughlybeforereinstallation.

5.5CleaningtheCabinet

DANGER! RiskofElectricShock
Thisunitcontainsmovingandhighpressure components.Therearenouserserviceableparts insidethecover .T oavoidelectricshock,injury ordeath:
–DONOTremovethecover . –Referservicingtoqualiedpersonnel.
Maintenance
1.Cleanthecabinetwithamildhouseholdcleanerand non-abrasiveclothorsponge.
60001253-D
27
Invacare®HomeFill®IIOxygenSystem

5.6ReplacingtheFuse 1.Unplugthepowercordfromtheelectricaloutlet.

2.RemovethepowercordfromthepowerinletAonthe compressorbaseC.
3.PulloutthefusedrawerBatthebottomofthepower inlet.
4.Usingafusepullerorasmallatbladescrewdriver , removethefrontfuseinsidethefusedrawerand discard.
Thefusedrawercontainsspacefortwofuses. Thefrontlocationisfortheoperationalfuse. Theunitissuppliedwiththefrontfuseonly. Therearlocationisforanoptionalsparefuse. Whenreplacingthefrontfuse,ensurethatthere isasparefusewiththesametypeandrating inthefusedrawer.
5.InstallreplacementfuseinfrontfuseDslot.
Usea5mmx20mm,timelag,lowbreaking capacity,250V ,2Amp,IEC60127styleonly (Invacarepartnumber1099820).
6.Pushfusedrawerbackintothepowerinlet.
7.Installpowercordintothepowerinletonthe compressorbase.
8.Plugpowercordintoelectricaloutlet.
9.TurnthepowerswitchtotheOn( toseethattheunitpowersupproperly .
10.Iftheunitdoesnotcomeon,turnthepowerswitch off,unplugthepowercord,andcallyourprovideror Invacareforservice.
)positionandcheck
28
60001253-D

6AfterUse

6.1 Disposal

This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU.
This product may contain substances that could be harmful to the environment if disposed of in places (landlls) that are not appropriate according to legislation.
The 'crossed out wheelie bin' symbol is placed on this product to encourage you to recycle wherever possible.
Follow local governing ordinances and recycling plans regarding disposal of the device or components normally used in operation. The device does not generate waste or residue in operation. DO NOT dispose of the device in the normal waste stream. Any accessories not part of the device MUST be handled in accordance with the individual product marking for disposal.
Please be environmentally responsible and recycle this product at its end of life. Contact Invacare or your service provider for product recycling information.
AfterUse
60001253-D
29
Invacare®HomeFill®IIOxygenSystem

7Troubleshooting

7.1IndicatorLights

IndicatorLightColorCompressorStatusEffect
None
O2BELOWNORMAL-YELLOW
FILLING-GREEN
CompressorisOff.O
CompressorisOnandwarmingup(0-3 minutesafterpowerisOn).
O2fromconcentratoris<90%(anytime afterthreeminutewarmup).
CompressorisOnandcylinderislling. O2fromconcentratoris>90%(anytime afterthreeminutewarmup).
owingtouserifconcentratorison.
2
Cylinderllingnotstarted.
O
owingtouserifconcentratorisOn.
2
Cylinderllingstops,willrestartwhen O
>90%.
2
O
owingtouserifconcentratorisOn.
2
Cylinderislling.
O
owingtouserifconcentratorisOn.
2
30
60001253-D
IndicatorLightColorCompressorStatusEffect
FULL-GREEN
ATTENTION-RED(AudibleAlarm)
CompressorisOn.Cylinderisnished lling(full)-removecylinder(FULLlight isOnin>10mins.)
or
Cylinderisnotattachedproperly-check connection(FULLlightisOnin<10 mins.).
CompressorisOn.CylinderNOTlling. Causes:
Troubleshooting
Cylinderllingstops.
Cannotllanothercylinderuntil compressoristurnedOffandOnagain.
O
owingtouserifconcentratorisOn.
2
Compressordoesnotbegintollor stopslling.
60001253-D
Fullcylinderattachedwhenunitturned On-replacewithemptycylinder
or
Internalfailure-contactprovider.
UnitmustbeturnedOffandOnbefore llingcanoccur.
O
owingtouserifconcentratorisOn.
2
31
Invacare®HomeFill®IIOxygenSystem

7.2Troubleshooting

ProblemSolution
NoindicatorlightsonwhencompressorturnedOn.
REDLightOn1.Checkthecylindergaugetoseeifitisfull:
1.TurncompressorOff(O).
2.Checkthatpowercordisproperlypluggedinto compressorandwalloutlet.
3.TurnthecompressorOn(I).
4.Cleanairlters.RefertoCleaningtheCompressorFilter intheMaintenancesectionofthemanual.
5.IfperformingSteps1-4doesnotcauseanyindicatorlights tocomeon,contactyourhomehealthcareprovideror Invacareforservice.
Cylinderfull-Removethecylinderandturntheunit Off.PerformtheCylinderPrellInspectioninthe Usagesectionofthemanualwithacylinderthatis notfull.
Cylindernotfull-ProceedtoStep2.
2.Checkthatthecylinderisconnectedproperly .
TurnthecompressorOff(O).
Removethecylinderandreinstall.
TurnthecompressorOn(I).Ifcylinderisnotfull andisconnectedproperly,thecompressormayhave internalfailure.Contactyourhomehealthcare providerorInvacare.
32
60001253-D
ProblemSolution
CompressorcyclesonandOff
or
O
BelowNormal
2
(yellow)andlling(green)lightsarealternating
or
O2belownormal(yellow)lightgoesOnandOff
O
belownormal(YELLOW)lightstaysOn
2
Troubleshooting
1.Checkthattheconcentratorliterowsettingisat orbelowtheowsettinglistedin4.5Turningthe CompressorOnandOff,page24. Ifnecessary,adjusttheliterowsetting(refertothe concentratorUsermanual).Waitatleast20minutesfor theGREENlighttoappear .Ifitdoesnotappear,proceed tostep2.
2.Theconcentratormayneedservice,contactyourhome healthcareproviderorInvacare.
1.Ensuretheconcentratorhaswarmedupforatleast 20minutes.Ifithasnotbeen20minutes,turnthe compressorOffuntiltheconcentratorwarmsup.Ifthe YELLOWlightstillremainsonproceedtostep2.
2.Inspecttheconnectionbetweenthecompressorandthe concentratorfordamageorpinching.
Ifdamaged,replace.
Ifpinched,straightentubing.Turnthecompressor Off(O)foratleast10seconds.T urnthecompressor On(I).IftheYELLOWlightstillremainsonafter10 minutes,proceedtostep3.
3.Theconcentratormayneedservice,contactyourhome healthcareproviderorInvacare.
60001253-D
33
Invacare®HomeFill®IIOxygenSystem
ProblemSolution
Filltimesexceedmaximumas
recommendedlltimesORGREENllinglightstaysOn.
1.EnsurethecylinderregulatorissettoOff(O).
2.Checktherecommendedlltimesshownintheindividual cylinder’smanual.
3.Checkthatthecylinderisconnectedproperly.Turnthe compressorOff(O).Removethecylinderandreinstall. TurnthecompressorOn(I).
4.Checkthecompressor/concentratorlineforproper connectionand/ordamage(i.e.-leaksorkinks).
5.Checkthattheconcentratorliterowsettingisat orbelowtheowsettinglistedin4.5Turningthe CompressorOnandOff,page24.Ifnecessary,adjustthe literowsetting(refertotheconcentratorUsermanual). IftheGREENllinglightdoesnotgoOff,thecompressor and/orconcentratormayneedservice.Contactyour homehealthcareproviderorInvacare.
ExcessivenoisewhencompressorisOn1.CallhomehealthcareproviderorInvacareforservice.
HissingsoundfromRegulatorwhenowselectorisOff1.TurnowselectorOff.
2.Ifhissingsoundscontinue,callhomehealthcareprovider orInvacare.
RegulatorisOnbutO2isnotbeingdelivered1.Rellcylinder
2.Checkthatcannulaisstraightandnotpinched.If damaged,replace.
3.CallhomehealthcareproviderorInvacareforservice.
Refertothecylinder’smanualforproperlltimes.
Fillingtimesareapproximateandmayvaryon environmentalconditions.
34
60001253-D
TechnicalData

8TechnicalData

8.1 Technical Description

The HomeFill II Oxygen System (IOH200AW) is an electro-mechanical prescription system for use in the home. Its intended function and use is to supply pressurized oxygen to ll I nvacare p roprietary g as c ylinders f or ambulatory use while the patient receives therapeutic oxygen from an oxygen concentrator. The IOH200AW gas system is designed for use with oxygen concentrators that meet the input specications f or pressure,  ow, an d oxygen purity.
The HomeFill II Oxygen system consists of IOH200AW gas compressor, connection tubing, and the Invacare supplied portable gas cylinders with the integral ow r egulator and interlocking ll tting.
The IOH200AW compressor consists of a multistage compressor, internal reservoir, control electronics with an integral oxygen concentration sensor, interlocking ll tting, i nlet  tting, an d dr ive mo tor. Th e compressor operation is controlled by the electronics assembly. The motor/compressor is allowed to ll a c ylinder o nly w hen the oxygen concentration of the inlet gas stream is greater than 90% and the cylinder pressure is not already greater than 1500 PSI (10342 kPa). The motor/compressor ll operation is completed when the gas in a cylinder reaches the maximum ll p ressure o f 2 000 P SI ( 13786 kPa).
cylinderalsocontainsanintegralowregulatorwithxed, selectableowsettings.

8.2CylinderFillTimes

CYLINDERFILLING TIMES
NORMAL
MAXIMUM
Typicallltimesassumetheinputrequirementsare met.
Fillingtimesmayvaryonenvironmentalconditions.
ConcentratorFlowRatetoPatient:
5Lmodelsupto2.5L/min.
9Lmodelsupto5L/min.
HF2PCE9AHF2PCE6
2hr20min1hr25min
2hr45min1hr50min
The IOH200AW compressor can ll c ylinders f or portable use by ambulatory patients. The Invacare supplied portable gas cylinders have an interlocking ll tting th at ma tes only with the Invacare IOH200AW compressor assembly. The gas
60001253-D
35
Invacare®HomeFill®IIOxygenSystem
8.3Specications
RegulatoryListings
RegulatoryListings:
InputSpecications
InputPressureRequired:
InputFlowRequired:
InputOxygenPurity:>90%O
GeneralSpecications
ElectricalRatings:230V~,1.0A,50Hz
OperatingEnvironmental Conditions:
AtmosphericPressure:700to1060hPa
StorageandTransport:
Width:
Height:
Depth:
EN/IEC60601–1
EN/IEC60601–1–2
5–25psi(34.5–172.4 kPa)
2L/min
2
41to104°F(5to40°C)at 15to60%non-condensing humidity
-25to158°F(-23to 70°C)at15to95% non-condensinghumidity.
20¼inches(51.44cm)
15inches(38.1cm)
16inches(40.64cm)
Weight:
ShippingWeight:
33lbs(14.97kg)
39lbs(17.69kg)
36
60001253-D

8.4ElectromagneticCompliance(EMC)

Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice shouldassurethatitisusedinsuchanenvironment.
TechnicalData
Emissionstest
RFemissions CISPR11
RFemissions CISPR11
Harmonic emissions IEC61000-3-2
ComplianceElectromagneticenvironment-guidance
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissions
GroupI
areverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic equipment.
ClassB
ClassA
TheDeviceissuitableforuseinallestablishmentsincludingdomestic establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes. Voltageuctuations /ickeremissions
Complies
IEC61000-3-3
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice shouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic discharge(ESD)
IEC61000-4-2
IEC60601testlevelCompliancelevelElectromagneticenvironment–guidance
±2,4,8kVcontact
±2,4,8kVcontact
Floorsshouldbewood,concreteorceramictile. Ifoorsarecoveredwithsyntheticmaterial,the
±2,4,8,15kVair
±2,4,8,15kVair
relativehumidityshouldbeatleast30%.
60001253-D
37
Invacare®HomeFill®IIOxygenSystem
ElectricalFast transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltagedips
IEC61000-4-11
Voltage Interruptions
IEC61000-4-11
Powerfrequency (50/60Hz) magneticeld
IEC61000-4-8
±2kVmains,100kHz±2kVmains,100kHz
±1kVline(s)toline(s)
±2kVline(s)toground
0%UT;1/2cycle@0, 45,90,135,180,225, 270,and315
and
0%UT;1cycle
and
70%UT;25/30cycles Singlephase:at0°
0%U
;250/300cycle0%UT;250/300cycle
T
±1kVline(s)toline(s)
±2kVline(s)toground
0%UT;1/2cycle@0, 45,90,135,180,225, 270,and315
and
0%UT;1cycle
and
70%UT;25/30cycles Singlephase:at0°
30A/m,50/60Hz30A/m,50/60Hz
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.Iftheuser oftheDevicerequirescontinuedoperationduring powermainsinterruptions,itisrecommendedthat theDevicebepoweredfromanun-interruptible powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication ofthetestlevel.
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.Iftheuser oftheDevicerequirescontinuedoperationduring powermainsinterruptions,itisrecommendedthat theDevicebepoweredfromanun-interruptible powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication ofthetestlevel.
Powerfrequencymagneticeldsshouldbeat levelscharacteristicofatypicallocationinatypical commercialorhospitalenvironment.
38
60001253-D
ConductedRF IEC61000-4-6
3Vrms,6Vrms3Vrms,6Vrms
TechnicalData
PortableandmobileRFcommunicationsequipment shouldbeusednoclosertoanypartofthe Deviceincludingcables,thantherecommended separationdistancecalculatedfromtheequation applicabletothefrequencyofthetransmitter .
Recommendedseparationdistance:
d=1.2√P150kHzto80MHz
RadiatedRF IEC61000-4-3
10V/m10V/m
d=0.35√P d=0.7√P
80MHzto800MHz 800MHzto2,5GHz
wherePisthemaximumoutputpowerrating ofthetransmitterinwatts(W)accordingto thetransmittermanufactureranddisthe recommendedseparationdistanceinmetres(m).
FieldstrengthsfromxedRFtransmitters,as determinedbyanelectromagneticsitesurvey, shouldbelessthanthecompliancelevelineach frequencyrange.
b
Interferencemayoccurinthevicinityofequipment markedwiththefollowingsymbol:
a
Field strengths from xed t ransmitters, s uch a s b ase s tations for r adio ( cellular/cordless) t elephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
60001253-D
a
39
Invacare®HomeFill®IIOxygenSystem
measuredeldstrengthinthelocationinwhichtheDeviceisusedexceedstheapplicableRFcompliancelevelabove,theDevice shouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary , suchasreorientingorrelocatingtheDevice.
b
Overthefrequencyrange150kHzto80MHz,eldstrengthsshouldbelessthan[V1]V/m.
At80MHzand800MHz,thehigherfrequencyrangeapplies.
RecommendedseparationdistancesbetweenportableandmobileRFcommunicationsequipmentandtheDevice
TheDeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.The customerortheuseroftheDevicecanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetween portableandmobileRFcommunicationsequipment(transmitters)andtheDeviceasrecommendedbelow ,accordingtothe maximumoutputpowerofthecommunicationsequipment
Separationdistanceaccordingtofrequencyoftransmitter[m]
Ratedmaximumoutput oftransmitter[W]
150kHzto80MHz
d=1.2√P
80MHzto800MHz
d=0.35√P
800MHzto2,5GHz
d=0.7√P
0.010.120.040.07
0.10.370.110.22
11.170.350.70
103.691.112.21
10011.673.507.00
40
60001253-D
TechnicalData
Fortransmittersratedatamaximumoutputpowernotlistedabovetherecommendedseparation,distancedinmeters(m) canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter ,wherePisthemaximumoutputpower ratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.
At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreection fromstructures,objectsandpeople.
60001253-D
41
Notes
Belgium&Luxemburg
Invacarenv ,Autobaan22,B-8210Loppem•Tel:(32)(0)50831010•Fax:(32)(0)508310 11•belgium@invacare.com•www.invacare.be
Danmark
InvacareA/S,Sdr.Ringvej37,DK-2605Brøndby•Tel:(45)(0)36900000•Fax:(45)(0)36900001
•denmark@invacare.com•www.invacare.dk
Deutschland
InvacareGmbH,Alemannenstraße10,D-88316Isny•Tel:(49)(0)75627000•Fax:(49)(0)75627 0066•kontakt@invacare.com•www.invacare.de
EasternEurope,Middle East&CIS España
InvacareEUExport•Kleiststraße49•D-32457PortaWestfalica•Germany•T el:(49)5731754540• Fax:(49)5731754541•webinfo-eu-export@invacare.com•www.invacare-eu-export.com InvacareSA,c/Arenys/n,PolígonIndustrialdeCelrà,E-17460Celrà(Girona)•Tel:(34)(0)9724932 00•Fax:(34)(0)972493220•contactsp@invacare.com•www .invacare.es
France
InvacarePoirierSAS,RoutedeStRoch,F-37230Fondettes•Tel:(33)(0)247626466•Fax:(33)(0)2 47421224•contactfr@invacare.com•www.invacare.fr
IrelandInvacareIrelandLtd,Unit5SeatownBusinessCampus•SeatownRoad,Swords,CountyDublin–
Ireland•Tel:(353)18107084•Fax:(353)18107085•ireland@invacare.com•www .invacare.ie
Italia
InvacareMeccSans.r.l.,ViadeiPini62,I-36016Thiene(VI)•Tel:(39)0445380059•Fax:(39)0445 380034•italia@invacare.com•www.invacare.it
Nederland
InvacareBV,Galvanistraat14-3,NL-6716AEEde•Tel:(31)(0)318695757•Fax:(31)(0)318695758
•nederland@invacare.com•csede@invacare.com•www .invacare.nl
Norge
InvacareAS,Grensesvingen9,Postboks6230,Etterstad,N-0603Oslo•Tel:(47)(0)22579500•Fax: (47)(0)22579501•norway@invacare.com•island@invacare.com•www.invacare.no
Österreich
InvacareAustriaGmbH,HerzogOdilostrasse101,A-5310Mondsee•T el:(43)623255350•Fax: (43)623255354•info@invacare-austria.com•www.invacare.at
Portugal
InvacareLda•RuaEstradaVelha,949,P-4465-784LeçadoBalio•T el:(351)(0)225105946/47•Fax: (351)(0)225105739•portugal@invacare.com•www.invacare.pt
Sverige
InvacareAB•Fagerstagatan9•S-16353Spånga•Tel:(46)(0)87617090•Fax:(46)(0)876181 08•sweden@invacare.com•www.invacare.se
Suomi
CampMobility•Patamäenkatu5,33900Tampere•Tel:09-35076310•info@campmobility.• www.campmobility .
Schweiz/Suisse/Svizzera
InvacareAG•Benkenstrasse260•CH-4108Witterswil•Tel.:(41)(0)614877080•Fax.:(41)(0)61 4877081•switzerland@invacare.com•www.invacare.ch
UnitedKingdom
InvacareLimited,PencoedT echnologyPark,Pencoed,BridgendCF355HZ•T el:(44)(0)1656776222• Fax:(44)(0)1656776220•UK@invacare.com•www.invacare.co.uk
EC
EU Representative
REP
Invacare GmbH Am Achener Hof 8 88316 Isny Germany
Manufacturer
60001253-D 2020-06-30
In vacareCorporation
OneInvacareWay
Elyria,OhioUSA
440-329-6000 800-333-6900
TechnicalServices
440-329-6593 800-832-4707
www.invacare.com
MakingLife’sExperiencesPossible®
Invacare Corporation One Invacare Way
USA
Elyria, Ohio 44035 USA
44035