Invacare HF2POST9, 1148114, HF2POST6, HF2POSTD User Manual

HomeFill

Oxygen Post Valve

Cylinder

HF2POST6 HF2POST9

HF2POSTD

“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”

Dealer: This manual MUST be given to the end user. User:
Before using this cylinder, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE INJURY OR DAMAGE MAY RESULT.
WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare.
DANGER
DO NOT SMOKE while using this device. Keep all matches, lighted cigarettes, candles or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered. and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
Textiles

NOTICE

The information contained in this document is subject to change without notice.

HomeFill
2 Part No 1148114
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ............................................... 4
RECEIVING INSPECTION .................................. 4
INTENDED USE ................................................ 4
SECTION 1—GENERAL GUIDELINES ..................5
SECTION 2—FEATURES .................................... 8
SECTION 3—TYPICAL PRODUCT
PARAMETERS ................................................. 9
SECTION 4—INSPECTION AND COMPRESSOR
INTERFACE ................................................... 10
Cylinder Prefill Inspection ...............................................................10
Filling the Cylinder ............................................................................13
SECTION 5—MAINTENANCE .......................... 19
SECTION 6—CYLINDER FILL TIMES ................. 20
Troubleshooting ................................................................................20
Replacement Parts - Accessories ..................................................20
LIMITED WARRANTY .................................... 22
Part No 1148114 3 HomeFill™

SPECIAL NOTES

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.

SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in property damage.

RECEIVING INSPECTION

Remove product from package and inspect for damage. If there is any damage, DO NOT use. Contact your dealer/ Home Care equipment provider.

INTENDED USE

To store supplemental oxygen for use with a CGA 870 post valve compatible conserver or regulator to be used by patients who may have difficulty extracting oxygen from the air they breathe.

NOTE: Updated versions of this manual are available on www.invacare.com.

HomeFill
4 Part No 1148114
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a health care professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
DANGER
This product is not intended as a life-sustaining or life-supporting device.
WARNING
Only personnel instructed and trained in the use of the HF2POST6, HF2POST9, and HF2POSTD Cylinder Assembly should operate this product.
Product contains magnetic, ferrous material that may affect the results of an MRI.
Be sure all connections are tight and leak free.
DO NOT use a liquid leak detector to test for leaks.
DO NOT autoclave.
DO NOT gas sterilize with ethylene oxide.
DO NOT clean with aromatic hydrocarbons.
Part No 1148114 5 HomeFill
SECTION 1—GENERAL GUIDELINES
WARNING
DO NOT attempt to repair product.
DO NOT immerse product in any kind of liquid.
Store product in a clean area when not in use.
Only use an Invacare Corporation HF2POST6BAG or HF2POST9BAG carrying bag.
NEVER smoke in an area where oxygen is being administered.
ALWAYS confirm prescribed dose before administering to patient and monitor on a frequent basis.
DO NOT use if dirt or contaminants are present on or around cylinder, valve, product or connecting devices.
NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
ALWAYS follow ANSI and CGA standards for Medical Gas Products and Flowmeters (E-7) and Oxygen Handling (G-4).
The use of this device is limited to the oxygen patient. Cylinders MUST be used only by the oxygen patient and are not to be distributed to any other individual for any purpose.
DO NOT allow oil from your hands or other sources to come into contact with the cylinder fittings. These substances may become flammable in the presence of oxygen and cause injury. Avoid touching the cylinder fitting whenever possible. Example of items to avoid are hand creams or lotions, cooking oils, suntan oils or sunblocks and similar oily products. If you use these types of products Invacare suggests you wash your hands prior to using this product. If the cylinder fitting should come in contact with such substances it should be cleaned off with a damp cloth prior to connecting it to the compressor.
HomeFill
6 Part No 1148114
SECTION 1—GENERAL GUIDELINES
WARNING
NEVER oil or lubricate the cylinder fittings. NEVER use any penetrating oil or lubricant such as WD-40 or 3 in One. If connection difficulty is experienced verify that the cylinder fitting is being inserted straight down into the compressor coupler. If continued difficulty is experienced contact your supplier for assistance.
DO NOT allow cylinders to tip or fall. Secure gas cylinders so they cannot fall. Ensure cylinder is always in upright position.
DO NOT store cylinders near sources of heat or flame.
DO NOT use while sleeping without consulting your healthcare provider.
DO NOT place the unit under clothing while in use. When this product is in use a small amount of oxygen is vented. Wearing the unit under clothing may saturate fabrics with oxygen and cause them to burn rapidly if exposed to sparks or flame. It may take several hours for oxygen levels in fabrics to return to normal.
Part No 1148114 7 HomeFill
SECTION 2—FEATURES
SECTION 2—FEATURES
Toggle Valve
Cylinder Fill
Port
FIGURE 2.1 Product Diagram
HomeFill
8 Part No 1148114