Invacare HF2POST9, 1148114, HF2POST6, HF2POSTD User Manual

HomeFill

Oxygen Post Valve

Cylinder

HF2POST6 HF2POST9

HF2POSTD

“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”

Dealer: This manual MUST be given to the end user. User:
Before using this cylinder, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE INJURY OR DAMAGE MAY RESULT.
WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare.
DANGER
DO NOT SMOKE while using this device. Keep all matches, lighted cigarettes, candles or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered. and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
Textiles

NOTICE

The information contained in this document is subject to change without notice.

HomeFill
2 Part No 1148114
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ............................................... 4
RECEIVING INSPECTION .................................. 4
INTENDED USE ................................................ 4
SECTION 1—GENERAL GUIDELINES ..................5
SECTION 2—FEATURES .................................... 8
SECTION 3—TYPICAL PRODUCT
PARAMETERS ................................................. 9
SECTION 4—INSPECTION AND COMPRESSOR
INTERFACE ................................................... 10
Cylinder Prefill Inspection ...............................................................10
Filling the Cylinder ............................................................................13
SECTION 5—MAINTENANCE .......................... 19
SECTION 6—CYLINDER FILL TIMES ................. 20
Troubleshooting ................................................................................20
Replacement Parts - Accessories ..................................................20
LIMITED WARRANTY .................................... 22
Part No 1148114 3 HomeFill™

SPECIAL NOTES

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.

SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in property damage.

RECEIVING INSPECTION

Remove product from package and inspect for damage. If there is any damage, DO NOT use. Contact your dealer/ Home Care equipment provider.

INTENDED USE

To store supplemental oxygen for use with a CGA 870 post valve compatible conserver or regulator to be used by patients who may have difficulty extracting oxygen from the air they breathe.

NOTE: Updated versions of this manual are available on www.invacare.com.

HomeFill
4 Part No 1148114
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a health care professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
DANGER
This product is not intended as a life-sustaining or life-supporting device.
WARNING
Only personnel instructed and trained in the use of the HF2POST6, HF2POST9, and HF2POSTD Cylinder Assembly should operate this product.
Product contains magnetic, ferrous material that may affect the results of an MRI.
Be sure all connections are tight and leak free.
DO NOT use a liquid leak detector to test for leaks.
DO NOT autoclave.
DO NOT gas sterilize with ethylene oxide.
DO NOT clean with aromatic hydrocarbons.
Part No 1148114 5 HomeFill
SECTION 1—GENERAL GUIDELINES
WARNING
DO NOT attempt to repair product.
DO NOT immerse product in any kind of liquid.
Store product in a clean area when not in use.
Only use an Invacare Corporation HF2POST6BAG or HF2POST9BAG carrying bag.
NEVER smoke in an area where oxygen is being administered.
ALWAYS confirm prescribed dose before administering to patient and monitor on a frequent basis.
DO NOT use if dirt or contaminants are present on or around cylinder, valve, product or connecting devices.
NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
ALWAYS follow ANSI and CGA standards for Medical Gas Products and Flowmeters (E-7) and Oxygen Handling (G-4).
The use of this device is limited to the oxygen patient. Cylinders MUST be used only by the oxygen patient and are not to be distributed to any other individual for any purpose.
DO NOT allow oil from your hands or other sources to come into contact with the cylinder fittings. These substances may become flammable in the presence of oxygen and cause injury. Avoid touching the cylinder fitting whenever possible. Example of items to avoid are hand creams or lotions, cooking oils, suntan oils or sunblocks and similar oily products. If you use these types of products Invacare suggests you wash your hands prior to using this product. If the cylinder fitting should come in contact with such substances it should be cleaned off with a damp cloth prior to connecting it to the compressor.
HomeFill
6 Part No 1148114
SECTION 1—GENERAL GUIDELINES
WARNING
NEVER oil or lubricate the cylinder fittings. NEVER use any penetrating oil or lubricant such as WD-40 or 3 in One. If connection difficulty is experienced verify that the cylinder fitting is being inserted straight down into the compressor coupler. If continued difficulty is experienced contact your supplier for assistance.
DO NOT allow cylinders to tip or fall. Secure gas cylinders so they cannot fall. Ensure cylinder is always in upright position.
DO NOT store cylinders near sources of heat or flame.
DO NOT use while sleeping without consulting your healthcare provider.
DO NOT place the unit under clothing while in use. When this product is in use a small amount of oxygen is vented. Wearing the unit under clothing may saturate fabrics with oxygen and cause them to burn rapidly if exposed to sparks or flame. It may take several hours for oxygen levels in fabrics to return to normal.
Part No 1148114 7 HomeFill
SECTION 2—FEATURES
SECTION 2—FEATURES
Toggle Valve
Cylinder Fill
Port
FIGURE 2.1 Product Diagram
HomeFill
8 Part No 1148114
SECTION 3—TYPICAL PRODUCT PARAMETERS
SECTION 3—TYPICAL PRODUCT PARAMETERS
DIMENSIONS:
HF2POST6: Weight 4.0 lbs (1.81 kg)
HF2POST6: Overall Length 11.1 in. (2.8 cm)
HF2POST6: Diameter 4.4 in. (11.1 cm)
HF2POST9: Weight 5.0 lbs (2.27 kg)
HF2POST9: Overall Length 14 in (35.5 cm)
HF2POST9: Diameter 4.4 in (11.1 cm)
HF2POSTD: Weight 6.30 lbs (2.85 kg)
HF2POSTD: Overall Length 20.8 in (52.8 cm)
HF2POSTD: Diameter 4.4 in (11.1 cm)
STORAGE CONDITIONS TEMPERATURE: -40°F (-40°C) to 140°F (60°C)
MAXIMUM HUMIDITY: 95%
Part No 1148114 9 HomeFill
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
NOTE: Refer to the HomeFill ™ Compressor Owner’s Manual for compressor operating instructions.
Cylinder Prefill Inspection
WARNING
All cylinders MUST be inspected before attempting to fill; otherwise, injury or damage may occur.
NEVER smoke in an area where oxygen is being administered.
NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
NOTE: For this procedure, refer to FIGURE 4.1 on page 12.
The Hydrostatic Testing Date
Aluminum cylinders MUST undergo testing every five years.
WARNING
DO NOT fill cylinders that have not been tested in the past five years. Otherwise, serious personal injury may result. Contact your dealer for replacement.
HomeFill
10 Part No 1148114
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
External Examination
NOTE: For this procedure, refer to FIGURE 4.1 on page 12.
1. Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist:
Dents or dings
Arc burns
Oil or grease
Any other signs of damage that might cause a cylinder
to be unacceptable or unsafe for use.
2. Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or blistering of the paint, or other protective coating or heat sensitive indicator. If fire or thermal damage is found, replace the cylinder.
3. Inspect the cylinder fillport for the following and replace if found:
Debris, oil or grease
Noticeable signs of damage
Signs of corrosion inside the valve
Signs of excessive heat or fire damage
Part No 1148114 11 HomeFill
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
DANGER
Improper filling or abuse of this cylinder, or failure to follow these safety instructions may cause serious bodily injury or death. High pressure oxidizing gas vigorously accelerates combustion. DO NOT alter or modify cylinder or related components. DO NOT use Oil or Grease. Keep cylinder, cylinder valve, all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death.
For medical applications. Use only practitioner. Uninterrupted use of over a long duration, without content of arterial blood, may be har mful.
Oxygen Cylinders filled by the Venture HomeFill shall be used for personal use only. "Not to be filled for resale or use by professional users".
This cylinder is to b e used with the Invacare Venture HomeFill system. This cylinder contains oxygen produced by an oxygen concentrator using the pressure swing adsorption process.
DO NOT handle cylinder or use contents until you are professionally trained, including emergency procedures.
Use in accordance Open valve slowly. S
ecure cylinder during s
area.
Keep away from heat, flame and spark. Keep out of
reach of children. DO NOT drop.
Transfilling of this gas is performed by Venture HomeFill Only.
Cylinders that have been refinished and/or exposed to elevated temperatures must be hydrostatically tested before filling.
Cylinder temperature Aluminum cylinders subject temperatures in excess of 350°F service and condemned by trained caustic paint strippers. Valve and safety relief devices must be removed and replaced by trained and authorized personnel. DO NOT alter or change valve assembly.
Invacare Corporation 2101 E. Lake Mary Blvd. Sanford, FL 32773 1-800-832-4707
!
!
WARNING
as directed by a high concentrations
monitoring its effect on
with Venture HomeFill Operator’s
Close valve after each use
torage and use. No smoking in cylinder
should not exceed 125°F (52°C). to the action of
must be
personnel. DO NOT use
DO NOT REMOVETHIS PRODUCT LABEL
Not for Commercial Resale
Part Number 1125090 Rev B - 11/04
licensed
oxygen
OXYGEN
+
93 + 3%
and when empty.
fire or heated to
withdrawn from
of oxygen
2
Manual.
FIGURE 4.1 External Examination
Checking Cylinder Pressure
WARNING
NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.
DO NOT drop oxygen cylinders. Use two hands when handling/transporting oxygen cylinders. Otherwise, injury or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 4.2 on page 14.
1. Install a conserver or regulator onto the cylinder according to the device manufacturer instructions.
HomeFill
12 Part No 1148114
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
2. Ensure toggle valve is open. Open the toggle valve by rotating counterclockwise.
3. Examine the gauge on the connected conserver or regulator. Perform one of the following:
If the cylinder pressure is less than 1500 psig, fill the
cylinder. Connect the cylinder to the compressor. Refer to Filling the Cylinder on page 13.
If the cylinder pressure is greater than 1500 psig., DO
NOT attempt to fill this cylinder. It may not fill. Use cylinder as normal.
Filling the Cylinder
NOTE: For this procedure, refer to FIGURE 4.2 thru FIGURE 4.7.
1. Check the cylinder pressure before attempting to fill cylinder. Refer to Checking Cylinder Pressure on page 12.
2. Close the toggle valve by rotating clockwise (FIGURE 4.2).
3. Remove the conserver or regulator according to the device manufacturer instructions (FIGURE 4.2).
Part No 1148114 13 HomeFill
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

CONSERVER/REGULATOR

INSTALLED
Gauge
Toggle
CONSERVER/REGULATOR
REMOVED
Valve
Conserver/
Regulator
Cylinder
FIGURE 4.2 Component Identification
4. Remove the post valve cover (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.3).
Connector
Post Valve Cover
Fillport Cover
FIGURE 4.3 Fillport Covers
5. On the compressor unit, momentarily push DOWN on the outer ring (sleeve) of the connector fillport to reset the connector (FIGURE 4.4).
HomeFill
14 Part No 1148114
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport.
PUSH
DOWN
Outer Ring (Sleeve)

Connector Fillport

GREEN Dots
BEFORE COUPLING CYLINDER, PUSH DOWN
ON SLEEVE
UNTIL GREEN DOTS ARE VISIBLE.
FIGURE 4.4 Resetting Connector Fillport
6. Grasp the cylinder assembly as shown in Detail “A” in FIGURE 4.5.
7. Position the cylinder in the compressor cradle as shown in Detail “C” in FIGURE 4.5.
8. Align the cylinder fillport with the connector fillport as shown in Detail “C” in FIGURE 4.5.
9. Pull UP on the outer ring (sleeve) of the connector fillport (Detail “B”) while pushing DOWN on the cylinder assembly to couple the cylinder fillport into the connector fillport as shown in Detail “B” in FIGURE 4.5.
NOTE: The cylinder is properly connected when an audible “click” is heard.
10. Press the compressor power switch to the On ( ) position.
Part No 1148114 15 HomeFill
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
NOTE: Refer to Cylinder Fill Times on page 20 for the length of time it will take to fill the cylinder.
DETAIL “A”

Cylinder

Assembly
Compressor
DETAIL “C”
Cylinder
Fillport
DETAIL “B”
Outer Ring
(Sleeve)
Power Switch
Cylinder
Fillport
Outer Ring
(Sleeve)
Compressor
Cradle
Cylinder
Connector
Fillport
PULL
FIGURE 4.5 Connecting the Cylinder to the Compressor
HomeFill
16 Part No 1148114
UP
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
11. After the cylinder is full, press the compressor power switch to the Off (O) position (FIGURE 4.6).
12. Grasp the cylinder assembly (FIGURE 4.6).
13. With the other hand, grasp the outer ring (sleeve) of the connector fillport and push DOWN (FIGURE 4.6).
14. Lift up on the cylinder assembly to remove from the connector fillport (FIGURE 4.6).
PUSH DOWN
Outer Ring
(Sleeve)
Cylinder
Assembly
Compressor
Cylinder
Fillport
Power Switch
Connector
Fillport
FIGURE 4.6 Disconnecting the Cylinder from the Compressor
15. When the cylinder fillport is disconnected from the connector fillport, release the outer ring (sleeve) of the connector fillport (FIGURE 4.6).
16. Remove the cylinder from the compressor cradle (FIGURE
4.6).
Part No 1148114 17 HomeFill
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
WARNING
The fillport cover on the connector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of the fillport covers are missing contact your healthcare provider before using.
17. Place the fillport covers onto the connector fillport (located on the compressor) and post valve fillport (located on the cylinder) (FIGURE 4.7).
18. Install a CGA870 post valve compatible conserver or regulator according to the manufacturer’s instructions.
19. Open the toggle valve by turning counterclockwise and examine the conserver or regulator gauge to ensure cylinder is full (needle pointing into GREEN area). If the cylinder is not full, perform the following:
Repeat Filling the Cylinder on page 13.
If cylinder is not full after second attempt to fill, contact
the provider.
Toggle
Valve
Connector
Fillport Cover
Fillport Cover
FIGURE 4.7 Connector Fillport
HomeFill
18 Part No 1148114
SECTION 5—MAINTENANCE
SECTION 5—MAINTENANCE
1. After each use, clean exterior of the product with a dry, lint free cloth only.
2. Store product in a clean area free from grease, oil, and other sources of contamination.

CAUTION

DO NOT use cleaning solutions.

DO NOT immerse product in any kind of liquid.

All repairs MUST be done by Invacare Corporation.

Part No 1148114 19 HomeFill
SECTION 6—CYLINDER FILL TIMES
SECTION 6—CYLINDER FILL TIMES

NOTE: All filling times are approximate and may vary on environmental conditions.

*
Cylinder
Filling
Times
NORMAL 1 hr 25 min 2 hr 20 min 3 hr 20 min
MAXIMUM 1 hr 50 min 2 hr 45 min 3 hr 50 min
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
L/min)
HF2POST6 HF2POST9 HF2POSTD

Troubleshooting

If post valve cylinder fails to function, try the following solutions. If problem cannot be corrected, consult your dealer/home care equipment provider.
PROBLEM POSSIBLE CAUSE SOLUTION
No flow. 1. Regulator or
conserver is in “Off” position.
2. Cylinder empty. 2. Replace/fill cylinder.
3. Toggle Valve closed. 3. Open toggle valve.
4. Conserver not sensing breath.
1. Set to prescribed setting.
4a. Check position of cannula in nose. Do not breathe through mouth.
4b. Check to see if the cannula is kinked.

Replacement Parts - Accessories

HF2POST6BAG or HF2POST9BAG Carrying Bag.
HomeFill
20 Part No 1148114

NOTES

NOTES
Part No 1148114 21 HomeFill

LIMITED WARRANTY

LIMITED WARRANTY

PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.

This warranty is extended only to the original purchaser/user of our products.

This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state.

Invacare warrants this product to be free from defects in materials and workmanship for a period of three years from date of purchase. If within such warranty period any such product shall be proven to Invacare’s satisfaction to be defective, such product shall be repaired or replaced, at Invacare’s option. This warranty does not apply to problems arising from normal wear and tear or failure to adhere to these instructions, nor does it include shipping charges incurred in replacement part installation or repair of any such product. Invacare’s sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or replacement.

For warranty service, please contact the Invacare service department during normal business hours at the toll-free telephone number identified on the back cover of this manual.

Upon receiving notice of an alleged defect in a product, Invacare will issue a serialized return authorization. It shall be the responsibility of the purchaser to pack the product(s) or part(s) in an appropriate manner to avoid shipping damage and return the product(s) or part(s), at the purchaser’s expense, to either Invacare’s plant or service center as specified by Invacare in advance. The defective product(s) or component part(s) must be returned for warranty inspection using the serial number as identification within 30 days of the return authorization date. DO NOT return any product(s) or part(s) to our factory or service center without prior consent. C.O.D. shipments will be refused – please pre-pay shipping charges.
HomeFill
22 Part No 1148114
LIMITED WARRANTY
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE, OR STORAGE; PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS); PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE; PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE’S CONTROL; OR PRODUCTS REPAIRED BY ANYONE OTHER THAN INVACARE OR AN INVACARE SERVICE CENTER. SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.

SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT BE APPLICABLE.

THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.

Part No 1148114 23 HomeFill
Invacare Corporation
USA
One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707
www.invacare.com
All rights reserved. Trademarks are identified by the symbols ™ and ®. All trademarks are owned by or licensed to Invacare Corporation unless otherwise noted. © 2007 Invacare Corporation

Part No 1148114 Rev B - 9/07