Invacare 4000, 2000 User Manual

I.V. ROD TELESCOPING KIT 9000XT RECLINER - KIT NO. 1074656 2000, 4000 AND 9000 RIDELITE RECLINER - KIT NO. 1029598
Installation, Assembly and Operating Instructions
NOTE: Check ALL parts for shipping damage. In case of shipping damage, DO NOT use. Contact Carrier/Dealer for further instruction.
The following recommendations are made for safe instal­lation and adjustment of the I.V . Rod T elescoping Kit:
GENERAL WARNINGS
DO NOT install or use this equipment with­out first reading and understanding this instruction sheet. If you are unable to un­derstand the Warnings, Cautions or Instruc­tions, contact a healthcare professional, dealer or technical personnel before at­tempting to install this equipment - other­wise, injury or damage may occur.
DO NOT pull/push the wheelchair by the I.V. rod assembly. Injury to the assistant/ occupant and/or damage to the I.V. rod assembly may occur.
DO NOT place any other objects on the I.V. rod assembly other than those deter­mined by a healthcare professional.
INSTALLATION WARNINGS
After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is securely tightened - other­wise injury or damage may result.
Use EXTREME caution when the I.V. rod holder for MODELS 2000, 4000 and 9000 Ridelite is installed. When installed, the rear arm release lever is removed and will NOT lock the arm in place.
NOTE: If installing the telescoping I.V . rod holder onto a 2000, 4000 or 9000 RIDELITE, refer to INSTALL-
ING THE TELESCOPING I.V . ROD HOLDER - 2000, 4000 or 9000 RIDELITE in this instruction sheet.
INSTALLING THE TELESCOPING I.V. ROD HOLDER - 9000XT ONLY (FIGURE 1)
NOTE: Refer to INSTALLATION WARNINGS in the SAFETY SUMMARY in this instruction sheet.
1. Remove the rear wheel from the wheelchair . Refer to
REMOVING/INSTALLING THE REAR WHEELS
in this instruction sheet.
2. Remove the arm from the wheelchair. Refer to RE- MOVING/INST ALLING THE ARMRESTS in this in­struction sheet.
3. Remove the locknut and hex screw that secure the arm socket to the wheelchair frame.
4. Remove the locknut that secures the back tube to the clevis bracket.
5. Position the I.V. holder on the existing pivot bolt as shown in FIGURE 1.
Arm Socket
(OUTSIDE of
Wheelchair
Frame)
Locknut
Outside of
Wheelchair
Frame
Locknut
(STEP 8,9,11)
Side Frame
I.V. Holder
(STEP 5)
Mounting
Screw
(STEP 7)
Front
Cylinder
(STEPS 4,6,11 )
Hex
Screw
STEP 3
Back Tube
(STEP 4)
Clevis
Bracket
(STEP 4)
Pivot Bolt
(STEP 5)
Flat Portion
(STEP 5)
Locknut
FIGURE 1 - INSTALLING THE
TELESCOPING I.V. ROD HOLDER
1
6. Reinstall locknut onto pivot bolt. DO NOT tighten.
2. Adjust I.V . Rod to desired height.
7. Install the provided mounting screw through the lower hole on the front cylinder section of the I.V . holder and into the side frame.
8. Install locknut onto mounting screw.
9. Turn locknut until snug fit is obtained. DO NOT tighten.
10. Reinstall the arm to ensure proper fit. Refer to RE- MOVING/INST ALLING THE ARMRESTS in this in­struction sheet.
1 1. Securely tighten both locknuts of STEPS 6 AND 9.
12. Install the I. V . rod support tube. Refer to INSTALL- ING THE I.V . ROD SUPPORT TUBE in this instruc­tion sheet.
INSTALLING THE I.V. ROD SUPPORT TUBE (FIGURE 2)
NOTE: Refer to INSTALLATION WARNINGS in the SAFETY SUMMARY in this instruction sheet.
NOTE: The I.V. holder must already be mounted to the side frame of the wheelchair.
3. Securely tighten adjustment knob.
Adjustment
Knob
Steel
Washer
(STEP 4)
Locknut
(STEP 5)
Small Nylon
Washer
(STEP 4)
Large
Nylon
Washer
(STEPS 3
AND 4)
Support Tube
(STEPS 1,2)
I.V. Rod
I.V. Rod
I.V. Rod
Holder
(STEP 1)
1. Insert the I.V . rod support tube into the rear cylinder of the I.V . rod holder.
NOTE: When installing the I.V . rod support tube, ensure the adjustment knob is pointing out away from the wheel­chair.
2. Turn the I.V . rod support tube until the holes in the I.V . rod support tube align with the holes in the rear cylin­der of the I.V . rod holder.
3. Press the large nylon washer into the back hole of the rear cylinder so that it is flush with outside surface.
4. Insert mounting screw through rear cylinder, large nylon washer, small nylon washer , and steel washer .
5. Install locknut onto mounting screw. Securely tighten.
6. Reinstall the rear wheel onto the wheelchair. Refer to
REMOVING/INSTALLING THE REAR WHEELS
in this instruction sheet.
7. Adjust I.V. rod to desired height. Refer to ADJUST- ING THE I.V. ROD in this instruction sheet.
ADJUSTING THE I.V. ROD (FIGURE 2)
NOTE: Refer to INSTALLATION WARNINGS in the SAFETY SUMMARY in this instruction sheet.
Rear
Cylinder
Mounting Screw
(STEP 4)
NOTE: Attaching hardware enlarged 3X scale for clarity.
FIGURE 2 - INSTALLING I.V. ROD
SUPPORT TUBE/ADJUSTING THE I.V. ROD
REMOVING/INSTALLING THE REAR WHEELS
NOTE: Refer to INSTALLATION WARNINGS in the SAFETY SUMMARY in this instruction sheet.
Recliners With Patient Operated Wheel Locks (FIGURE 3)
REMOVING.
1. Remove the axle mounting screw and locknut that secure the patient operated wheel lock, spacers, and rear wheel to the wheelchair frame.
1. Loosen adjustment knob.
2
Loading...
+ 2 hidden pages