Integra 4015, 4013, 4014, 4016, 4621 Operating Instructions Manual

...
VIAFLO Electronic Pipettes Operating instructions
161950_V08
VOYAGER Adjustable Tip Spacing Pipettes
Declaration of conformity
7205 Zizers, Switzerland
declares on its own responsibility that the devices
comply with:
Zizers, March 12, 2018
Description Models
VIAFLO Pipettes 4011, 4012, 4013, 4014, 4015, 4016, 4621, 4622, 4623, 4624,
4626, 4631, 4632, 4633, 4634, 4636, 4641, 4642, 4646
VOYAGER Pipettes 4721, 4722, 4723, 4724, 4726, 4731, 4732, 4736, 4743, 4744,
4763, 4764
EU Directives (DoW: Date of Withdrawal) Before DoW DoW After DoW
Low Voltage Equipment 2006/95/EC 20.04.2016 2014/35/EU
Electromagnetic Compatibility 2004/108/EC 20.04.2016 2014/30/EU
Restriction of Hazardous Substances 2011/65/EU
Waste Electrical and Electronic Equipment 2012/19/EU
Battery Directive 2006/66/EC
EU Regulations
Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH)
1907/2006
Capacity Labelling of Portable Secondary Batteries 1103/2010
Standards for EU
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
EN 61010-1: 2010
Electrical equipment for measurement, control and laboratory use ­EMC requirements.
EN 61326-1: 2013
Standards for Canada and USA
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
CAN/CSA-C22.2
No. 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
UL 61010-1
Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interfer­ence received, including interference that may cause undesired operation.
Part 15 of the
FCC Rules
Class A
International Standards
Piston-operated volumetric apparatus - Part 2: Piston pipettes ISO 8655-2
Elmar Morscher CEO
Thomas Neher Quality Manager
www.integra-biosciences.com 3
VIAFLO/VOYAGER – Operating Instructions V08
Table of Contents
Chapter 1 Introduction
1.1 Symbols used...................................................................................7
1.2 Intended use.....................................................................................7
1.3 Safety notes .....................................................................................8
Chapter 2 Description of the device
2.1 Scope of delivery............................................................................10
2.2 Overview of the VIAFLO/VOYAGER Electronic Pipettes ...............11
2.2.1 Pipette parts........................................................................11
2.2.2 Back view............................................................................12
2.2.3 Display ................................................................................12
2.2.4 Touch wheel........................................................................13
2.2.5 Left and right arrow buttons ................................................13
2.2.6 PURGE button ....................................................................13
2.2.7 RUN key..............................................................................13
2.2.8 Tip ejector ...........................................................................14
2.2.9 Reset button........................................................................14
Chapter 3 Installation
3.1 Operating environment...................................................................15
3.2 Charging the battery.......................................................................15
3.2.1 Charging the battery on a stand..........................................16
3.2.2 Charging the battery with the mains adapter ......................16
3.3 Exchanging the battery...................................................................17
3.4 Toolbox - adapt your pipette...........................................................17
3.4.1 Preferences.........................................................................18
3.4.2 Calibration & Service ..........................................................20
3.4.3 Communications .................................................................21
3.4.4 Device Information..............................................................22
3.4.5 Language ............................................................................22
3.4.6 Write protect........................................................................23
4 INTEGRA Biosciences AG
VIAFLO/VOYAGER – Operating Instructions V08
Chapter 4 Operation
4.1 Turn on/off the device.....................................................................24
4.2 Attaching and removing GripTip pipette tips ..................................24
4.3 Start pipetting .................................................................................25
4.3.1 Pipetting..............................................................................25
4.3.2 Blowout modes ...................................................................25
4.3.3 Recommendations for pipetting ..........................................26
4.4 Pipetting options and settings ........................................................27
4.4.1 Edit option...........................................................................27
4.4.2 Volume selection.................................................................27
4.4.3 Speed selection ..................................................................28
4.4.4 Pace....................................................................................29
4.4.5 Count, Mix Cycle and Rows................................................29
4.4.6 Help.....................................................................................29
4.5 VOYAGER tip spacing ...................................................................30
4.5.1 Set the number of tip positions and the tip spacing ............30
4.5.2 Tip spacing operation..........................................................30
4.5.3 Homing the tips...................................................................31
4.6 Troubleshooting/FAQ .....................................................................32
4.6.1 General ...............................................................................32
4.6.2 Electronic ............................................................................33
Chapter 5 Pipetting modes
5.1 Overview pipetting modes ..............................................................34
5.2 Detailed description of pipetting modes .........................................35
5.2.1 Pipet mode..........................................................................35
5.2.2 Repeat dispense mode .......................................................36
5.2.3 Sample dilute mode ............................................................37
5.2.4 Pipet/Mix mode ...................................................................38
5.2.5 Manual pipet mode .............................................................39
5.2.6 Reverse pipet mode............................................................40
5.2.7 Variable dispense mode .....................................................41
5.2.8 Multi Aspirate mode ............................................................42
5.2.9 Sample Dilute/Mix mode .....................................................43
5.2.10 Serial Dilution mode..........................................................44
5.3 Custom step-based programming mode ........................................45
5.3.1 Create a custom program ...................................................45
5.3.2 Modify existing programs ....................................................47
5.3.3 Example of custom mode ...................................................48
www.integra-biosciences.com 5
VIAFLO/VOYAGER – Operating Instructions V08
Chapter 6 Maintenance
6.1 Cleaning .........................................................................................49
6.2 Assembly and Disassembly ...........................................................49
6.2.1 VIAFLO single channel lower end.......................................49
6.2.2 VIAFLO multichannel pipettes ............................................53
6.3 Sterilization.....................................................................................54
6.3.1 Autoclaving the disassembled components........................54
6.4 Servicing.........................................................................................55
6.4.1 Shipping to INTEGRA Biosciences.....................................55
6.4.2 Changing O-rings of tip fittings............................................55
6.4.3 Lubrication ..........................................................................56
6.5 Equipment disposal ........................................................................56
Chapter 7 Calibration
7.1 Definitions.......................................................................................57
7.2 Materials.........................................................................................58
7.3 Preparation.....................................................................................58
7.3.1 Test conditions and environment ........................................58
7.3.2 Cleaning..............................................................................58
7.3.3 Firmware .............................................................................58
7.3.4 Pre-wetting of tips ...............................................................59
7.3.5 Leak test .............................................................................59
7.4 Obtaining the Actual Volume..........................................................60
7.5 Calculation of accuracy and precision ............................................61
7.6 Adjusting electronic pipettes...........................................................62
Chapter 8 Technical Data
8.1 Environmental conditions ...............................................................64
8.2 Specification of the device..............................................................64
8.3 Pipetting speed...............................................................................65
8.4 Intellectual Property........................................................................66
8.5 Pipette Specifications .....................................................................67
8.6 Z Correction Factors.......................................................................70
Chapter 9 Accessories
9.1 Accessories ....................................................................................71
9.2 Consumables .................................................................................71
VIAFLO/VOYAGER – Operating Instructions V08
6 INTEGRA Biosciences AG
Imprint
© 2018 INTEGRA Biosciences AG
All rights to this documentation are reserved. In particular the rights of reproduction, processing, translation and the form of presentation lie with INTEGRA Biosciences AG. Neither the complete documentation nor parts thereof may be reproduced in any way, or stored and processed using electronic media or distributed in any other way without the written consent of INTEGRA Biosciences AG.
This operating instruction manual has part number 161950, the version is V08. It applies as of (see Toolbox - Device information):
of VIAFLO/VOYAGER Electronic Pipettes until a newer revision is released.
VIAFLO, VOYAGER, VIALINK, and GripTip are trademarks of INTEGRA Holding, Switzerland.
Manufacturer
Customer service
Please contact your local INTEGRA Biosciences representative. To find out name and address go to www.integra-biosciences.com
.
Further information and operating instructions in other languages are available on www.integra-biosciences.com
or on request info@integra-biosciences.com.
Serial number 7000000 or higher
(Firmware) FW version 4.10 or higher
INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland T +41 81 286 95 30 F +41 81 286 95 33
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA T +1 603 578 5800
F +1 603 577 5529 info@integra-biosciences.com www.integra-biosciences.com
VIAFLO/VOYAGER – Operating Instructions V08 Introduction
www.integra-biosciences.com 7
1 Introduction
These operating instructions contain all the information required for installation, operation and maintenance of the VIAFLO/VOYAGER Electronic Pipettes. This chapter informs about the symbols used in these operating instructions, the intended use of the pipettes and the general safety instructions.
1.1 Symbols used
The operating instructions specifically advise of residual risks with the following symbols:
W
ARNING
This safety symbol warns against hazards that could result in injury. It also indicates hazards for machinery, materials and the environment. It is essential that you follow the corresponding precautions.
C
AUTION
This symbol cautions against potential material damage or the loss of data in a microprocessor controller. Follow the instructions.
N
OTE
This symbol identifies important notes regarding the correct operation of the device and labor-saving features.
1.2 Intended use
This is a general-purpose laboratory instrument for use in research only. Any use of this instrument in a medical or IVD setting is under the sole responsibility of the user.
If the VIAFLO/VOYAGER Electronic Pipettes are used in a manner not specified by INTEGRA Biosciences, the protection provided by the VIAFLO/VOYAGER Electronic Pipettes may be impaired.
VIAFLO/VOYAGER Electronic Pipettes are microprocessor controlled and stepper motor driven pipettes. They are used for aspirating and dispensing aqueous liquids in the volume range of 0.5–5000 µl using GripTip pipette tips. In addition, the VOYAGER adjustable tip spacing pipettes enable users to transfer multiple samples simultaneously between different labware.
VIAFLO/VOYAGER – Operating Instructions V08 Introduction
8 INTEGRA Biosciences AG
1.3 Safety notes
VIAFLO/VOYAGER Electronic Pipettes comply to the recognized safety regulations and are safe to operate. The pipettes should only be operated when in perfect condition and while observing these operating instructions.
The device may be associated with residual risks if it is used or operated improperly by untrained personnel. Any person operating the pipettes must have read and understood these operating instructions, and particularly, the safety notes, or must have been instructed by supervisors so that safe operation of the device is guaranteed.
W
ARNING
• Use only an original INTEGRA Li-ion battery (#4205) and charging device.
• Old Li-ion batteries may cause a safety risk. We recommend to replace the battery after 3 years of use. Also replace the battery if the charging intervals are unusually short or if the charging takes much longer than usual (4 hours or more). – These are indicators that the battery has reached the end of its life-cycle.
• Li-ion technology bears the risk of thermal runaway and cell rupture if the battery was damaged. Do not expose the battery to heat (> 60°C) and avoid mechanical stress. Batteries which were subject to deep dis­charges may develop internal short circuits, leading to an increased self­discharge rate and heating during battery charging. This may also result in thermal runaway and cell rupture.
C
AUTION
• To extend the battery life-cycle, it is recommended to charge the battery every 2 months if the pipette is not used regularly. If the pipette is not used for more than 6 months, unplug the battery.
W
ARNING
•Do not use the VIAFLO/VOYAGER Electronic Pipettes near flammable material or in explosive are a s . Also, do not pipette highly flammable liq­uids such as acetone or ether.
• When handling dangerous substances, comply with the material safety data sheet (MSDS) and with all safety guidelines such as the use of pro­tective clothing and safety goggles.
VIAFLO/VOYAGER – Operating Instructions V08 Introduction
www.integra-biosciences.com 9
CAUTION
•Do not immerse the VIAFLO/VOYAGER Electronic Pipettes in liquid. The fluid can damage internal parts. Avoid pipetting of liquids whose vapors could attack the materials PA (polyamide), POM (polyoxymethylene), FPM (fluor-rubber), NBR (nitrile-rubber), CR (chloroprene), silicone. Corrosive vapors could also damage metallic parts inside the device.
• Do not modify the VIAFLO/VOYAGER Electronic Pipettes in any way. Repairs may only be performed by INTEGRA Biosciences or by an authorized after-sales service member.
• Parts may be replaced with original INTEGRA Biosciences parts only.
N
OTE
Prolonged exposure of the VIAFLO/VOYAGER Electronic Pipettes to UV­light can cause discoloration and/or yellowing of the pipette housing. However, this will not affect the performance of the device in any way.
Regardless of the listed safety notes, additional applicable regulations and guidelines of trade associations, health authorities, trade supervisory offices, etc. must be observed.
Please visit our website www.integra-biosciences.com
on a regular basis for up to date
information regarding REACH classified chemicals contained in our products.
VIAFLO/VOYAGER – Operating Instructions V08 Description of the device
10 INTEGRA Biosciences AG
2 Description of the device
2.1 Scope of delivery
• VIAFLO or VOYAGER pipette
• Rechargeable battery (located inside the pipette, Li-ion, 3.7 V, 1050 mAh)
• Bag of spare O-rings (300 μl and 1250 µl volume ranges only)
• O-ring removal tool (300 μl and 1250 μl volume ranges only)
• Certificate of Performance
C
AUTION
Verify the scope of delivery when unpacking the device and check for potential transportation damage. Do not operate a device that is dama ged, instead contact your local dealer.
VIAFLO/VOYAGER – Operating Instructions V08 Description of the device
www.integra-biosciences.com 11
2.2 Overview of the VIAFLO/VOYAGER Electronic Pipettes
2.2.1 Pipette parts
1 Display
2 Back Button, to navigate backward
3 Touch Wheel, spin to scroll and move the cursor
4 OK Button, to make a selection
5 Left and Right Arrow Buttons, for selections
6 PURGE Button, to empty tips
7 RUN Key, to start operations
8 Tip Ejector
9 Finger Hook, facilitates easy operation
10 Volume Indicator Label, color matches GripTip box
insert.
11 Ejector Sleeve
12 Tip Fitting
1
2
5
6
7
8
3
4
9
11
12
10
VIAFLO/VOYAGER – Operating Instructions V08 Description of the device
12 INTEGRA Biosciences AG
2.2.2 Back view
2.2.3 Display
The Display shows all pipetting options.
13 Reset Button
14 Power Connector
15 Charging Stand Interface
16 Battery
Battery Life Indicator
Instruction Volumes
Options
13
14
15
16
Current mode
Active step
OK
VIAFLO/VOYAGER – Operating Instructions V08 Description of the device
www.integra-biosciences.com 13
2.2.4 Touch wheel
The Touch Wheel is fully operational with only one hand. Rotational finger movements translate into up and down cursor movement on the display. The Touch Wheel is fully functional with the use of latex gloves.
2.2.5 Left and right arrow buttons
On VOYAGER pipettes, these buttons are also used for tip spacing adjustments.
2.2.6 PURGE button
During pipetting, you can interrupt the current pipetting protocol and purge all remaining liquid currently in the GripTips. To do so, press PURGE (6
).
2.2.7 RUN key
Press and release the RUN Key (7
) to initiate aspiration, dispense, mix, purge, and special
pipetting operations. This button is centrally located for left- or right-handed actuation.
During dispense, you can press and hold the RUN Key to perform a two-step blowout, see “4.3.2 Blowout modes” on page 25
.
Move finger on the Touch Wheel to choose (and highlight) an option on the display. Press OK (4
) to
make the selection.
When a setting dial is displayed, spin the Touch Wheel to change the value and press OK.
At times, you will see and on a display screen. These buttons are used to select options.
Press to select the option indicated with the left arrow (HELP, in the example beside). Press to select the option indicated with the right arrow (RUN, in the example).
The pipette will display a prompt:
To proceed, press and release the RUN Key (7
). Upon completion of the dispense, the first step in the current program will be displayed.
VIAFLO/VOYAGER – Operating Instructions V08 Description of the device
14 INTEGRA Biosciences AG
2.2.8 Tip ejector
The Tip Ejector easily ejects tips from the tip fitting.
2.2.9 Reset button
The Reset Button (13
) is located on the back of the pipettes. It is used to reset the RAM in the pipette. The programs stored in memory are maintained. Once pressed, the startup screen will be displayed.
Press any key to continue and allow the instrument to initialize and home. The routine ends by displaying the Main Menu.
The serial number can be found beneath the tip ejector. Press and hold the tip ejector in the down position to see the 7-digit serial number.
VIAFLO/VOYAGER – Operating Instructions V08 Installation
www.integra-biosciences.com 15
3 Installation
3.1 Operating environment
The VIAFLO/VOYAGER Electronic Pipettes have been designed for use in a laboratory. They shall be operated in a dry and dust-free location with a temperature of 5– 40 °C and a maximal (non-condensing) relative humidity of 80 %.
3.2 Charging the battery
All VIAFLO/VOYAGER Electronic Pipettes share the same rechargeable Li-ion battery. Charge the battery completely before first use. A full charge takes 2.5 hours (max. 4 hours) and will provide approximately 3000 pipetting cycles for single channel pipettes and 1500 cycles for multichannel pipettes.
C
AUTION
To extend the battery life-cycle, it is recommended to charge the battery every 2 months if the pipette is not used regularly. If the pipette is not used for more than 6 months, unplug the battery.
A battery indicator is provided on the pipette display (2.2.3) indicating different states:
• Blinking red icon: Battery is low and needs to be recharged. If not connected to the mains adapter, the pipette will turn off soon.
• Green icon: Pipette is connected to the mains adapter.
• Blinking bars: Pipette is recharging.
• Two static and one blinking bar on the right: Pipette is recharging and battery is 80% charged.
• Three static bars: Battery is fully charged.
C
AUTION
Use only the approved INTEGRA battery, mains adapter or charging stand, see (“9.1 Accessories” on page 71). Use of an incompatible power transformer can damage the pipette.
VIAFLO/VOYAGER – Operating Instructions V08 Installation
16 INTEGRA Biosciences AG
3.2.1 Charging the battery on a stand
Use one of the different charging stands (#4210, #4211, #4215, #3215 with #3217 or #3218, see “9.1 Accessories” on page 71) - to charge the battery.
C
AUTION
Always use the correct mains adapter for the charging stand.
The pipette will turn on when placed on the stand and turn off when the Turn Off Time is reached, also see 3.2.2
. For disconnecting simply lift the pipette up from the stand.
3.2.2 Charging the battery with the mains adapter
With the optional mains adapter (#4200), you can use the pipette while charging through the line cord.
If the pipette is idle while charging, the display may dim or show the start up screen (see “4.1
Turn on/off the device” on page 24), but it will continue to show the battery life indicator. When the Standby Time is reached, the display is turned off. When the Turn Off Time is reached, the instrument will shut off.
Place the pipette on the charging stand by fitting the power receptacle (15
, on the top back of the pipette)
over the prong connector on the top of the stand.
Plug the cable of the adequate power supply to the socket.
Insert the mains adapter connector into the receptacle on the top back of the pipette (14). Plug the mains adapter into a wall outlet.
The pipette will turn on when the line cord is connected.
VIAFLO/VOYAGER – Operating Instructions V08 Installation
www.integra-biosciences.com 17
3.3 Exchanging the battery
After exchanging the battery, a protective switch is active. The pipette can only be started after connecting it to the mains power supply.
W
ARNING
INTEGRA VIAFLO/VOYAGER Electronic Pipettes use Li-ion batteries, see “1.3 Safety notes” on page 8
.
3.4 Toolbox - adapt your pipette
The Toolbox provides options to adapt the device to appropriate applications, setting personal preferences, calibration, computer connectivity and storing device information.
Help information is available for each mode.
To exchange the battery, loosen the screw to detach the pipette backing. Disconnect the power plug of battery (16
). Connect the power plug of the new battery to the pipette's socket and reassemble the pipette backing.
Toolbox mode Description
ASSIST Additional pipetting modes available in combination with
ASSIST (hidden by default, see Preferences - Main Menu).
ASSIST PLUS Additional pipetting modes available in combination with
ASSIST PLUS (hidden by default, see Preferences - Main Menu).
Preferences Customizes the system parameters.
Calibration & Service Sets calibration and service history options.
Communications Enables communication between your electronic pipette and a
PC.
Device Information View your pipette’s serial number and set a personal ID.
Language Sets language.
Write Protect Protects programs or menu options from modification.
Time/Date Sets current time and date.
VIAFLO/VOYAGER – Operating Instructions V08 Installation
18 INTEGRA Biosciences AG
3.4.1 Preferences
Preferences customizes your system parameters. Select a preference and press OK to access.
Preference Description Range
Sound Simple tones indicate completion of operations and
errors. Select an option and press OK to change the status of the beep tone between On and Off:
Step Complete: At the end of a program step
Program Complete: At program completion
Purge Key: When Purge is pressed.
Messages: When a message appears.
Error Message: When an error message appears or when illegal data entry is attempted.
Touch Wheel: When using the Touch Wheel.
Last Dispense: Before the last dispense in Repeat Dispense and Variable Dispense.
/ (On/Off)
Display Customizes your display. Press OK to select an option,
use the Touch Wheel to display the desired value.
Start up Screen: Select the startup display at one of the following: None or Custom (up to two, uploaded with VIALINK). Press to save your selection.
Brightness: Use the Touch Wheel to change the brightness: 1 (dim) to 10 (bright). Press OK.
Dim Time: The display will dim after a set number of minutes. A shorter dim time helps preserve battery life. Press OK to save your selection.
Standby Time: The display will turn off after a set number of minutes and allows you to continue work­ing where you have left. A shorter standby time helps preserve battery life.
Turn Off Time: The pipette will turn off after a set number of hours. You can change this setting. Press OK to save your selection.
None, Custom 1 or 2
1-10
Never, 1-20 min
5-60 min (default 5 min)
1-24 hours (default 8 h)
Main Menu Select a function to be hidden from the main menu (Off)
and press OK, e.g. ASSIST, Pipet, Repeat Dispense, Sample Dilute, Pipet/Mix, Tip spacing.
/ (On/Off)
Touch Wheel
Adjust your Touch Wheel sensitivity. Press to save. Low, Medium,
High
VIAFLO/VOYAGER – Operating Instructions V08 Installation
www.integra-biosciences.com 19
After changing desired settings, press to save.
N
OTE
Extending the volume range is not recommended. Full functionality and specified accuracy/precision cannot be guaranteed. The default values of the speed table are optimized for th e pipette motor. Changing the values may cause a more noisy operation of the pipette but does not do harm to the pipette.
Preference Description Range
Pipetting Select an option and press OK.
Purge Key Speed: Choose the desired purge speed and press OK to save your selection.
Blowin Delay: Choose a timed delay between the blowout and the blowin (delay before the piston homes) at the end of a dispense, if no two step blow­out is performed, see “4.3.2
Blowout modes” on
page 25.
Extend Volume: For pipetting below or above the specified volume range: 50 μl pipette: (1.0)–2–50 μl 125 μl pipette: (2.0)–5–125 μl 300 μl pipette: (5.0)–10–300–(310) μl 1250 μl pipette: (25)–50–1250 μl The volumes in brackets refer to extended volumes, e.g. extend the minimal pipetting volume on a 125 μl pipette from 5 μl to 2 μl. Not available for 12.5 μl and 5000 μl pipettes.
Speed Table: Adapts the speed steps 1-10 to user defined pipetting speeds in μl/s. Select a speed step, press OK and use the Touch Wheel to change the value. Press OK and save your selections.
1-10
None/
0.5-5.0 s
/ (On/Off)
μl/s values depend on pipette size
Communi­cation Menu
Select the communication to be shown under communications (, On) and press OK, e.g. Bluetooth, USB, Via ASSIST.
/ (On/Off)
VIAFLO/VOYAGER – Operating Instructions V08 Installation
20 INTEGRA Biosciences AG
3.4.2 Calibration & Service
These options enable you to set calibration features, review service history and check the number of pipetting moves performed.
Calibration & Service
Description Range
Calibration Allows for re-calibration of the pipette to restore
accuracy. The calibration factors for Pipette and Repeat type are displayed. To edit the calibration volumes, press Edit.
Target Volume: This is the volume you are interested in using for the calibration.
Actual Volume: This is the measured volume obtained when dispensing the target volume.
Current Factor: Displays the factor currently in use.
Factory Reset: Resets the correction factor back to the original factory setting. Press to apply the factory setting.
-
Calibration Reminder
Time or Cycles
Sets a calibration reminder based on a specified time frame or number of pipetting cycles. When the calibration reminder is displayed, press any key to confirm. However, the reminder will reappear every time the pipette is turned on until you change the reminder time or use the reset option.
Reminder: Press OK to turn the reminder timer On or Off.
Days/Cycles: Use the Touch Wheel to set a reminder interval for calibration (time in days or in thousands of cycles). Press to set the timer to the defined calibration interval.
Remind in/Total Cycles: Displays the residual time or amount of cycles respectively before calibration is required.
Reset: Resets the timer to the defined calibration interval. Press to enable.
Press to save.
/ (On/Off) 1-365 days or 1k - 240k cycles
Service History
Displays notes of any service that took place on the pipette listed newest entry first.
-
VIAFLO/VOYAGER – Operating Instructions V08 Installation
www.integra-biosciences.com 21
After changing desired settings, press to save.
3.4.3 Communications
The VIAFLO/VOYAGER Electronic Pipettes can be programmed from a PC via a Charging/Communication Stand (#4211), Charging/Communication Station for Linear Stand (#3218) or via wireless Bluetooth connection.
VIALINK is a pipette management software for the PC. It allows the user to upload/ download custom programs, images, firmware updates and service history to and from VIAFLO/VOYAGER Electronic Pipettes. It can be downloaded from the INTEGRA website in the product section. A detailed description of the software, along with the operating instructions, can be found on the website as well.
Select one connection type and press OK to allow communication with the PC. VIALINK will detect the pipette automatically.
Move Counter
Displays the number of completed moves.
• Piston Moves: One piston move is defined as one full start and stop cycle of the piston.
• Spacing Moves (VOYAGER only): One tip spacing move is defined as one full start and stop cycle of the spacing motor.
-
Communi­cations
Description
USB Place the pipette on a Charging/Communication Stand (#4211) and
connect it to the USB port of your PC. To exit the communications mode follow screen prompt or press the disconnect button in VIALINK.
Bluetooth Each pipette needs its own Bluetooth module (#4221). Bluetooth
communication has an open field connectivity of about 10 meters. If a paring code is required: 12345.
Via ASSIST Activate this communication for live position teaching on ASSIST or
ASSIST PLUS.
Remote Ctrl (Bluetooth)
To control the pipette by an external device via Bluetooth, e. g. to integrate the pipette in an automation system.
Remote Ctrl (Wire)
To control the pipette by an external device, e. g. to integrate the pipette in an automation system.
Calibration & Service
Description Range
VIAFLO/VOYAGER – Operating Instructions V08 Installation
22 INTEGRA Biosciences AG
3.4.4 Device Information
This menu displays information about your pipette, such as pipette size, number of channels, serial number, firmware (FW) and hardware (HW) version.
Go to the next pages with Next .
3.4.5 Language
Device Information
Description
Edit Owner • Press Edit Owner to enter the user name for your pipette. Use the
Touch Wheel to highlight a character and press OK. You can press to Delete the last character entered. After entering the desired text, press to Save.
The last page displays the compliance screen, which is also shown during start up. It lists the standards the pipettes comply with.
Language Description
Language You can choose the language in which all screens are displayed. Scroll to
the desired language, press OK and to Save.
VIAFLO/VOYAGER – Operating Instructions V08 Installation
www.integra-biosciences.com 23
3.4.6 Write protect
Select this option to protect programs and menu options from inadvertent modification. The pipetting programs can still be used.
Keep the password in a safe place. Should you lose your password, contact INTEGRA Biosciences to retrieve your password.
Write Protect
Description Range
Select an option and press OK to switch protection on or off:
Standard Programs
Custom Programs
Calibration
Tool box
Tip Spacing (VOYAGER only)
Password Protection: Protect the access to the write protect menu by selecting “”.
Edit Password, if password protect is switched on. To enter a password use the Touch Wheel to high- light a character and press OK. Press to save the password. The password must be entered before you can access the write protect menu.
/ On/Off