This manual contains proprietary information, which is
protected by copyright. All rights are reserved. No part of this
manual may be photocopied, reproduced or translated to
another language without prior written consent of the Good Will
company.
The information in this manual was correct at the time of
printing. However, Good Will continues to improve products
and reserves the right to change specifications, equipment, and
maintenance procedures at any time without notice.
LCR-6000 Series User Manual
Disclaimer
GWINSTEK assumes no liability for the
customer's failure to comply with these
requirements.
Ground The
Instrument
To avoid electric shock hazards, the instrument
chassis and cabinet must be connected to a safety
earth ground by the supplied power cable with
earth blade.
DO NOT Operate In
An Explosive
Atmosphere
Do not operate the instrument in the presence of
inflammable gasses or fumes. Operation of any
electrical instrument in such an environment
constitutes a definite safety hazard.
Safety Summary
Warning Dangerous
When you notice any of the unusual conditions listed below, immediately
terminate operation and disconnect the power cable.
Please contact the GWINSTEK service center representative for repair of the
instrument. If you continue to operate without repairing the instrument, there
is a potential fire or shock hazard for operators.
The instrument is operating abnormally if:
The instrument emits abnormal noise, smell, smoke, or a spark-like light
during the operation.
The instrument generates high temperature or electrical shock during
operation.
The power cable, plug, or receptacle on the instrument is damaged.
Foreign substances or liquid has fallen into the instrument.
The following general safety precautions must be observed during all phases
of operation, service, and repair of this instrument. Failure to comply with
these precautions or with specific WARNINGS elsewhere in this manual may
impair the protection provided by the equipment. In addition it violates safety
standards of design, manufacture, and intended use of the instrument.
i
Safety Summary
Keep Away From Live
Circuits
Operating personnel must not remove instrument
covers. Component replacement and internal
adjustments must be made by qualified
maintenance personnel. Do not replace
components with the power cable connected.
Under certain conditions, dangerous voltages may
exist even with the power cable is removed. To
avoid injuries, always disconnect the power and
discharge the circuits before touching them.
DO NOT Service Or
Adjust Alone
Do not attempt internal service or adjustment
unless another person, capable of rendering first
aid and resuscitation, is present.
DO NOT Substitute
Parts Or Modify
Instrument
Because of the danger of introducing additional
hazards, do not install substitute parts or perform
unauthorized modifications to the instrument.
Return the instrument to a GWINSTEK
authorized service center for service and repair to
ensure that safety features are maintained.
ii
LCR-6000 Series User Manual
Table of Contents
Safety Summary ............................................................................................ i
Table of Contents ......................................................................................... 1
Table of Figures ........................................................................................... 8
List of Table .................................................................................................... 10
1.UNPACKING AND PREPERATION .................................................... 11
Table 13-4 Correction factors for the cable length of test leads .................. 136
10
LCR-6000 Series User Manual
This chapter describes how to set up and start the LCR-6000 Series LCR
Meter.
Incoming Inspection
Environmental Requirements
Cleaning
How to Remove the Handle
After you receive the instrument, perform the following
checks during unpacking according to the following
procedure:
WARNING
If the external face of the instrument (such as
the cover, front/rear panel, VFD screen, power
switch, and port connectors) appears to have
been damaged during transport, do not turn on
the power switch. Otherwise, you may get an
electrical shock.
Make sure that the packing box or shock-absorbing
material used to package the instrument has not been
damaged.
Refer to the <Packing List> in the packing box, check
that all packaged items supplied with the meter have been
provided as per the specified optioned.
1. UNPACKING AND
PREPERATION
1.1 Incoming Inspection
11
UNPACKING AND PREPERATION
Note
If an abnormality is detected, contact the company and
transport the meter to your nearest GW INSTEK
Instruments sales or service office. Save the packing box,
shock-absorbing material, and packaged items as you
received them in case you need to mail the instrument to
an authorized GW Instek distributor or service center.
Set up the LCR-6000 Series where the following
environmental requirements are satisfied.
Operating Environment
Ensure that the operating environment meets the
following requirements:
Temperature:
0ºC to 50ºC
Temperature range
at calibration:
23ºC± 5ºC (<1ºC deviation from the
temperature when performing
calibration)
Humidity:
<70% at wet bulb temperature
≤40ºC (non-condensation)
Altitude:
0 to 2000m
To prevent electrical shock, disconnect the LCR-6000
Series power cable from the receptacle before cleaning.
Use a dry cloth or a cloth slightly dipped in water to clean
the casing. Do not attempt to clean the LCR-6000 Series
internally.
WARNING
Don’t Use Organic Solvents (such as alcohol or
gasoline) to clean the Instrument.
1.2 Environmental Requirements
1.3 Cleaning
12
LCR-6000 Series User Manual
Figure 1-1 How to remove the handle
1 Retracted
3 Carrying Position
2 Extended
Remove Handle
(Lift the handle perpendicular to the unit while pulling it in the direction of 1.)
1.4 How to Remove the Handle
A handle kit is attached to the LCR-6000 Series:
13
2. OVERVIEW
This chapter contains general information about LCR-6000 Series .The
information is organized as follows.
Introduction
Main Specifications
Feature overview
Thank you for purchasing an LCR-6000 Series LCR
meter.
The GW INSTEK LCR-6000 Series is a general-purpose
LCR meter for incoming inspection of components,
quality control, and laboratory use.
The LCR-6000 Series is used for evaluating LCR
components, materials, and semiconductor devices over a
wide range of frequencies (10 Hz to 300 kHz) and test
signal levels (10.00mV to 2.00V & 100.0uA to 20.00mA).
With its built-in comparator, the LCR-6000 Series can
output comparison/decision results for sorting
components into a maximum of ten bins. Furthermore,
by using the handler interface, the LCR-6000 Series can
be easily combined with a component handler, and a
system controller to fully automate component testing,
sorting, and quality-control data processing.
The list meas function permits entry of up to 10
frequencies or test signal levels points to be automatically
measured.
100.0uA- 20.00mA (± 10%)
CC mode:100.0uA- 20.00mA(± 6%) @2VMax
30Ω, 50Ωand 100Ω
2.3.1 Test Signal Frequency
2.3.2 Test Signal Level
LCR-6000 Series User Manual
2.3.3 Output impedance
17
2.4 Main Functions
OPEN/SHORT correction:
Eliminates measurement errors due to stray parasitic
impedance in the test fixtures.
Bin sort
The primary parameter can be sorted into BIN1-BIN9,
AUX, OUT and HI/IN/LO for each of the primary
measurement parameters.
The sequential mode or tolerance mode can be selected as
the sorting mode.
Limit Setup
Absolute value, deviation value, and % deviation value
can be used for setup.
BIN count
Countable from 0 to 999999
Points
There is a maximum of 10 points.
Sweep parameters
Test frequency, test voltage, test current.
Comparator function of list sweep
The comparator function enables you to set one pair of
lower and upper limits for each measurement point.
You can select from:
Judge with the first sweep parameter/Judge with the
second parameter/Not used for each pair of limits.
2.4.1 Correction Function
2.4.2 Comparator Function (Sorting)
Overview
2.4.3 List Sweep
18
LCR-6000 Series User Manual
Up to 10 setup conditions can be written to/read from
the built-in non-volatile memory.
The front panel keys can be locked.
Complies with SCPI.
Universal serial bus jack, type-A (4 contact positions,
contact 1 is on your left); female; for connection to USB
memory devices only.
After a USB disk has been inserted into the USB host
port of the LCR meter and if the meter detects that the
USB disk is usable, the meter will show “USB Disk
Ready. Press <Enter> to save screen.” on the lower part
of the LCD screen.
Figure 2-1 Disk Ready
2.5 Measurement Assistance Functions
2.5.1 Files
2.5.2 Key Lock
2.5.3 RS‐232
2.6 Options
2.6.1 USB host port
Save Screen
19
Overview
If the keypad is not needed to enter a numeric value, a
screenshot of the meter‟s LCD display can be captured
and saved to the USB disk by depressing the Enter key. If
the screen shot is saved successfully, “Screen saved.” will
be shown in the message area at the bottom of the
screen. The location and name of the screenshot will also
be shown, for example: F:\LCR6300\Screen\ScreenShot001.bmp.
Figure 2-2 Screen Saved
0V ~ ± 2.5V
Accuracy: 0.5%+0.005V
2.6.2 DC Bias Voltage
20
3. STARTUP
This chapter describes names and functions of the front panel, rear panel,
and screen display and provides the basic procedures for operating the LCR6000 Series.
Front panel summary
Rear panel summary
Power On/Off
Connect to Device under test
1
11
5
3
12
2
6
7
8
4
910
No.
Description
1
LCD Display
2
Soft-key
3
Number Key
4
Cursor Key
5
Power Switch
6
Measure key
LCR-6000 Series User Manual
3.1 Front panel
Figure 3-1 Front panel
Table 3-1 Front panel description
21
7
Setup Key
8
BNC Terminal
9
Trigger Key
10
ESC Key
11
USB Disk Port (USB-Host)
12
System Soft-key
3.2 Rear Panel
21
43
No.
Description
1
Power Cable Receptacle(Outlet) (to LINE)
2
Frame Terminal
3
RS-232C Interface
4
Handler Interface
Press the power key for at least 2 seconds. Release power
key when the POWER LED is lit.
Press the power key at least 2 seconds. LCR-6000 Series
will shut down when you release the power key.
Figure 3-2 Rear Panel
Startup
Table 3-2 Rear panel description
3.3 Power On/Off
3.3.1 Power On
3.3.2 Power Off
22
LCR-6000 Series User Manual
LCR-6000 Series is ready to be used as soon as the
power-up sequence has completed. However, to achieve
the specification accuracy, first warm up the instrument
for 30 minutes.
The LCR-6000 Series uses a four terminal measurement
configuration that provides easy, accurate and stable
measurements and avoids mutual inductance, interference
from measurement signals, noise and other factors
inherent with other types of connections.
Figure 3-3 Connect to DUT
L ForceL SenseH SenseH Force
- DUT +
WARNING
Do not apply DC voltage or current to the
UNKNOWN terminals. Applying a DC voltage or
current may lead to device failure. Connect the
measurement sample (DUT) to the test port (or
the test fixture, cables, etc. after the DUT has
been completely discharged.
+
-
3.4 Warm‐up Time
3.5 Connect to the Device Under Test
(DUT)
23
4. MEASURE KEY
This section includes the following information:
MEAS DISPLAY page
OPEN/SHORT page
LIST SETUP page
LIST MEAS page
When the [Measure] key is pressed, the
[MEAS DISPLAY] page appears. The following
measurement controls can be set.
FUNC – Measurement Function
RANGE – Impedance range
FREQ – Test Frequency
TRIG – Trigger Mode
LEVEL – Test Signal Level
SPEED – Meas Speed
4.1 [MEAS DISPLAY] Page
Measure Key
24
LCR-6000 Series User Manual
Figure 4-1 [MEAS DISPLAY] Page
The LCR-6000 Series simultaneously measures four
components of complex impedance (parameters) in a
measurement cycle. These include a primary parameter, a
secondary parameter and two monitor parameters.
NOTE
The monitor parameters can be set in the
[SETUP] page. The monitor parameters are
initially set to OFF.
Table 4-1 Measurement parameter combinations
Cs-Rs
Cs-D
Cp-Rp
Cp-D
Lp-Rp
Lp-Q
Ls-Rs
Ls-Q
Rs-Q
Rp-Q
R-X
DCR
Z-θr
Z-θd
Z-D
Z-Q
Table 4-2 Monitor parameters
Z D Q
Vac
Iac Δ Δ%
θr
θd R X
G B Y
4.1.1 Measurement Function [FUNC]
Types of measurement parameters
Monitor parameters
25
Measure Key
Table 4-3 Measurement and Monitor parameter descriptions
Parameter
Description
Cs
Capacitance value measured using the
series equivalent circuit model
Cp
Capacitance value measured using the
parallel equivalent circuit model
Ls
Inductance value measured using the
series equivalent circuit model
Lp
Inductance value measured using the
parallel equivalent circuit model
Rs
Equivalent series resistance measured
using the series equivalent circuit
model (ESR)
Rp
Equivalent parallel resistance
measured using the parallel equivalent
circuit model
Z
Absolute value of impedance
Y
Absolute value of admittance
G
Conductance
B
Susceptance
R
Resistance (=Rs)
X
Reactance
D
Dissipation factor
Q
Quality factor(=1/D)
θr
Phase radian
θd
Phase angle
Vac
Test signal Voltage
Iac
Test signal Current
Δ
Absolute deviation value
Δ%
Relative deviation value
DCR
Direct Current Resistance
Step 1. Press the [Measure] key.
Step 2. Press the [MEAS DISPLAY] soft-key if
[MEAS DISPLAY] is not already the active
mode.
Step 3. Use the cursor key to select [FUNC] field.
Step 4. Use the soft-keys to select measurement
function.
Procedure for setting the measurement function
[FUNC]
26
Loading...
+ 111 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.