Instek GSP-730 User Manual

Spectrum Analyzer
GSP-730
USER MANUAL
GW Part No. 82SP-73000M01
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Table of Contents
Table of Contents
SAFETY INSTRUCTIONS .................................................. 3
GETTING STARTED .......................................................... 7
GSP-730 Introduction ................................. 8
Appearance ............................................... 10
First Time Use Instructions ...................... 17
BASIC OPERATION ........................................................ 26
Frequency Settings .................................... 28
Span Settings ............................................ 31
Amplitude Settings ................................... 34
Autoset ..................................................... 36
Marker ...................................................... 38
Measurement ............................................ 50
Limit Line Testing ..................................... 57
Bandwidth................................................. 61
Trace ......................................................... 62
Display...................................................... 67
Save/Recall Files ....................................... 71
System Settings ........................................ 76
REMOTE CONTROL ....................................................... 78
Interface Configuration ............................. 80
Configure Remote Interface ...................... 80
Command Syntax ...................................... 82
Command List .......................................... 85
FAQ .............................................................................. 104
APPENDIX .................................................................... 106
GSP-730 Default Settings ........................ 106
1
GSP-730 User Manual
GSP-730 Specifications ........................... 108
GSP-730 Dimensions .............................. 110
Declaration of Conformity ....................... 111
INDEX .......................................................................... 112
2
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Earth (ground) Terminal
Frame or Chassis Terminal
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.

SAFETY INSTRUCTIONS

This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.

Safety Symbols

These safety symbols may appear in this manual or on the instrument.
3
GSP-730 User Manual
General Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Ensure signals to the RF input do not exceed
+30dBm/±25V DC.
Do not block the cooling fan opening. Do not disassemble the instrument unless you
are qualified.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The instrument falls under category II.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range: 100V~240V Frequency: 50/60Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.

Safety Guidelines

4
SAFETY INSTRUCTIONS
Cleaning
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Temperature: 5°C to 45°C Humidity: 90% @ 45°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The instrument falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -20°C to 60°C ; <60°C/70% RH
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
5
GSP-730 User Manual
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)

Power cord for the United Kingdom

When using the instrument in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
6
GETTING STARTED
GSP-730 Introduction .......................................................... 8
Main Features ......................................................................................... 8
Package Contents .................................................................................. 9
Appearance ....................................................................... 10
GSP-730 Front Panel ......................................................................... 10
Rear Panel ............................................................................................. 14
Display................................................................................................... 15
First Time Use Instructions ............................................... 17
Tilting the Stand .................................................................................. 17
Power UP and Down ......................................................................... 18
Software Update .................................................................................. 19
USB Driver Installation...................................................................... 20
Restoring Default Settings ................................................................. 21
Conventions ......................................................................................... 22

GETTING STARTED

This chapter provides a brief overview of the GSP-730, the package contents, instructions for first time use and an introduction to the front panel, rear panel and GUI.
7
GSP-730 User Manual
Performance
150kHz~3GHz bandwidth 100kHz resolution
Features
Autoset with automatic floor level and span. Marker table function Limit line testing Split window display ACPR measurement OCBW measurement Automatic resolution bandwidth mode.
Interface
480640 color LCD display On-screen menu icons VGA video output RS-232C USB 2.0 Host port for data storage USB 2.0 Device port for the virtual com port
communication

GSP-730 Introduction

The GSP-730 is a low-cost, basic spectrum analyzer. The GSP-730 has all the basic features of our more advanced models, but in a smaller package, designed especially for education.

Main Features

8
GETTING STARTED
Opening the box
Contents (single unit)
Main unit Quick Start Guide User Manual CD
Power cord x1 (region
dependent)
Calibration certificate

Package Contents

Check the contents before using the GSP-730.
9
GSP-730 User Manual
dB
Enter
F 1
F 2
F 3
F 4
F 5
F 6
0
BK SP
kHz /
Sec
µ
mSec
MHz /
GHz /
Sec
1
4
7
2
5
8
3
6
9
Frequency
Span
Amplitude
Autoset
Marker
Peak
Limit Line
Meas
Search
BW
Trace
Display
Memory
Preset
Hardcopy
System
Hardcopy
Setup
GSP-730
150 kHz 3 GHz
Spectrum Analyzer
RF INPUT 50
+30dBm MAX.
MAX.DC ±25V
LCD Display Function
Keys
Menu keys Hardcopy key
Arrow keys
RF Input terminal
Keypad and unit keys
USB A portPower
button
Scroll wheel
LCD display
640 X 480 color LCD display. The display shows the soft keys for the current function, frequency, amplitude and marker information.
Function keys
F 1
~
F6
The F1 to F6 function keys directly correspond to the soft keys on the right-hand side of display.
Frequency
Frequency
Sets the center frequency, start frequency, stop frequency and step frequency.
Span
Span
Sets the span, with options for full span, zero span and last span.

Appearance

GSP-730 Front Panel

10
GETTING STARTED
Amplitude
Amplitude
Sets the amplitude reference level, scale and amplitude units.
Autoset
Autoset
Automatically searches the peak signal with maximum amplitude and displays it with appropriate horizontal and vertical scales.
Marker
Marker
The Marker key is used to configure the markers, trace markers as well as other related functionality.
Peak Search
Peak
Search
Finds each maximum and minimum peak. Used with the Marker functions.
Meas
Meas
Configures ACPR and OCBW measurements.
Limit Line
Limit Line
Sets and tests Pass/Fail limit lines.
BW
BW
Sets the resolution bandwidth.
Trace
Trace
Sets traces and trace related functions.
Display
Display
The Display key configures the split-screen windowing mode and the basic display properties.
Memory
Memory
The memory key is used to save or recall setup, trace and limit line data.
11
GSP-730 User Manual
Preset
Preset
The Preset key will restore the spectrum analyzer to the factory settings.
Hardcopy
Hardcopy
The hardcopy key is a quick save key that will save a screen-shot of the display.
Hardcopy setup
Hardcopy
Setup
Configures the hardcopy options.
System
System
The System key is used to configure the RS232 interface, language, update the firmware as well as other system options.
Power key
Turns the instrument on/off.
Scroll wheel
Edit values, select listed items.
Arrow keys
Increment/decrement values (in steps), select listed items.
12
GETTING STARTED
RF input terminal
RF INPUT 50
+30dBm MAX.
MAX.DC ±25V
RF input port. Accepts RF inputs.
Maximum input: +30dBm
(+20dBm measurable)
Input impedance: 50Ω Maximum DC voltage: ±25V N-type: female
Numeric keypad
0
1
4
7
2
5
8
3
6
9
MHz / mSec
dB
Enter
GHz /
Sec
kHz /
µSec
BK SP
The numeric keypad is used to enter vales and parameters. It is often used in conjunction with the arrow keys and scroll wheel.
USB A, Micro SD
USB A port for saving/recalling settings/files to external memory. Only supports FAT/FAT32 formatting.
13

Rear Panel

Fan
RS232 port
VGA port
Power socket
USB B port
Security slot
RS232
RS232 9 pin DSUB port.
USB B
USB B Device port. USB 1.1/2.0
VGA
VGA video-out port. Supports SVGA (480X640)
Fan
Power Socket
Power Socket: 100~240V, 50/60Hz. 15W Max
Security Slot
Kensington-type security slot.
GSP-730 User Manual
14

Display

Reference level
Function menu
Marker information
Entry / Message area Trace icons
Traces and waveforms
Soft menu keys
Frequency/ Bandwidth settings
Reference level
Displays the reference level. For details, see page
34.
Marker information
Displays marker information. For details see page
38.
Function menu
Displays the current function menu.
Soft menu keys
The Soft menu keys are associated with the F1 to F6 function keys to the right of the display.
Trace Icons
Displays the color of each active trace and the trace mode of each active trace. See page 62 for more information about traces.
Entry/Message area
This area is used to show system messages, errors and input values/parameters.
GETTING STARTED
15
GSP-730 User Manual
Frequency/ Bandwidth settings
Displays the Start, Center and Stop frequencies, RBW, Span and Sweep settings.
Trace and waveforms
Main display showing the traces (page 62), limit lines (57) and marker positions (38).
16
GETTING STARTED
Description
The GSP-730 has two adjustable tabs at the front that can be used to position the instrument into two preset orientations.
Leaning Position
Set the tabs down to have the GSP-730 leaning backward.
Upright Position
Set the tabs flat to have the GSP-730 in an upright position.

First Time Use Instructions

Use the procedures below when first using the GSP-730 to tilt the stand, power up the instrument, update the firmware and restore the unit back to the default settings. Lastly, the Conventions sections will introduce you to the basic operating conventions used throughout the user manual.

Tilting the Stand

17
GSP-730 User Manual
Power Up
1. Insert the AC power cord
into the power socket.
2. Press the power button to
turn the GSP-730 on.
3. The GSP-730 will begin to boot up in under a
few seconds.
Note
If system fails to start, please see your local GW Instek distributor.
Power Down
1. Press the power button to
power down.

Power UP and Down

18
GETTING STARTED
Description
The GSP-730 allows the software to be updated by end-users. Before using the GSP-730, please check the GW Instek website or ask your local distributor for the latest software.
The update file, MAIN1.BIN, must be placed in the root directory of a USB flash drive.
System version
Before updating the software, please check the software version.
1. Press
System
>Information[F4].
2. The system version is displayed on the
SW Ver[F4] icon.
Software version
Update software
1. Place the update file, MAIN1.BIN, into the root
directory of a USB flash drive.
2. Insert the USB flash drive into the USB port on
the front panel.
3. Press
System
>Update From USB Flash[F5].
4. Press Update Now[F3] to execute the update
process.
When the message “Programmed Successful” is
displayed, the software has successfully completed the update procedure.

Software Update

19
GSP-730 User Manual
5. Reboot the system when the update procedure
has finished by cycling the power button.
6. Check the software version again to confirm
the update procedure.
Note
The upgrade process may take a few minutes.
Description
If the type B USB port on the rear panel is to be used for remote control, then the USB driver must be installed. The USB driver is located in the CD that accompanied this manual.
Driver installation
1. Ensure the GSP-730 is turned on.
2. Connect the USB cable from the PC
to the rear panel USB B port.
3. Windows will automatically detect the GSP-
730 as a new device.
4. Follow the instructions to locate the GSP-730
driver on the accompanying CD and install the driver.
5. To see if the driver has been successfully
installed, you can check to see if the GSP-730 is recognized by the Windows Device Manager when the GSP-730 in connected to the PC.
In Windows XP go to: Start>Control Panel>Device Manager.

USB Driver Installation

20
GETTING STARTED
The GSP-730 should be shown under the Ports (COM & LPT) node:
Note
If the USB driver installation fails, you can try to manually install the driver.
You can tell that the driver hasn’t been installed if
you see the AT91USBSerial icon in the ports node.
Right clicking this icon will allow you to manually install the GSP-730.ini driver.
Description
The factory default settings can be easily restored using the Preset key on the front panel. The default settings cannot be changed. See page 106 for a list of the factory default settings.
Steps
1. Press
Preset
.
The spectrum analyzer will load the default factory
settings.

Restoring Default Settings

21
GSP-730 User Manual
Soft Menu keys
The F1 to F6 function keys on the right side of the display correspond directly to the soft­menu keys on their left.
Soft-menu keys
F1 ~ F6 function keys
dB
Enter
F 1
F 2
F 3
F 4
F 5
F 6
0
BK SP
kHz / Sec
µ
mSec
MHz /
GHz / Sec
1
4
7
2
5
8
3
6
9
Frequency
Span
Amplitude
Autoset
Marker
Peak
Limit Line
Meas
Search
BW
Trace
Display
Memory
Preset
Hardcopy
System
Hardcopy
Setup
GSP-730
150 kHz 3 GHz
Spectrum Analyzer
RF INPUT 50
+20dBm MAX.
MAX.DC ±25V
W
Input Parameter Values
Selecting this type of menu key will allow you to enter a new value with the numeric keypad or increment/decrement the value using the scroll wheel or number pad. See the parameter input description below for more details.

Conventions

The following conventions are used throughout the user manual. Read the conventions below for a basic grasp of how to operate the GSP-730 menu system and front panel keys.
22
GETTING STARTED
Toggle State
Pressing this menu key will toggle the state. Notice that any soft-menu key that can be toggled will have the active parameter underlined.
Toggle State & Input Parameter
Pressing this menu key will allow you to toggle the state of the function between on and off. When in the on-state, the parameter value can be manually edited. Use the numeric keypad to enter the new value or use the scroll wheel to increment/decrement the current value. Again, the setting that is underlined is the active setting.
Sub Menu
Pressing the More menu key will enter a submenu.
Sub Menu to select parameter
Pressing this type of menu key will enter a submenu to select a parameter.
23
GSP-730 User Manual
Return to the Start of a Menu Tree.
When you have navigated down a menu tree and you wish to return to the start of the menu tree, simply press the same Menu again.
For example if you pressed:
Trace
> More[F6] > Min Hold[F2],
simply press
Trace
again to return to the start
of the Trace menu.
Active Function
Pressing this type of menu key will activate that function. The menu key will be highlighted to show it is the active function.
Parameter input
Numerical keypad
Scroll wheel
Directional arrow keys
Backspace, Enter keys and unit keys
0
1
4
7
2
5
8
3
6
9
MHz / mSec
dB
Enter
GHz /
Sec
kHz /
µSec
BK SP
Parameter values can be entered using the numeric keypad, the scroll wheel and the arrow keys.
24
GETTING STARTED
Using the numeric keypad
When prompted to enter a parameter, use the number keys (0~9), the decimal key (.) and the minus key (-) to enter a value. After a value has been entered, the unit keys can be used to select the units.
The value of the parameter is shown at the bottom of the screen as it is edited.
Edited parameter
Ref:-30dBm
Back Space
Use the backspace key to delete the last character or number entered.
Using the scroll wheel
Use the scroll wheel to alter the current value. Clockwise increases the value, anti-clockwise decreases the value. The scroll wheel is usually used for values that highly variable, such as the center frequency settings.
Directional arrows
Use the directional arrows to select discrete parameters or to alter values by a coarser resolution than the scroll wheel. Left/down decreases the value, right/up increases the value. The directional arrows are usually used for values that are of a discrete nature, such as selecting a memory location.
25
GSP-730 User Manual
Frequency Settings ............................................................ 28
Center Frequency ................................................................................ 28
Start and Stop Frequency ................................................................... 29
Center Frequency Step ....................................................................... 30
Span Settings .................................................................... 31
Span ....................................................................................................... 31
Full Span ............................................................................................... 32
Zero Span ............................................................................................. 32
Last Span ............................................................................................... 33
Amplitude Settings ............................................................ 34
Reference Level ................................................................................... 34
Amplitude Units .................................................................................. 35
Scale/Div .............................................................................................. 35
Autoset .............................................................................. 36
Using Autoset....................................................................................... 36
Limiting the Autoset Vertical Search Range ................................... 37
Limiting the Autoset Horizontal Search Range ............................. 37
Marker ............................................................................... 38
Activating a Marker ............................................................................. 39
Move Marker to Trace ........................................................................ 43
Show Markers in Table ....................................................................... 44
Peak Search ........................................................................................... 46
Peak Table ............................................................................................ 48
Measurement .................................................................... 50
Channel Analysis Overview ............................................................... 50
Limit Line Testing .............................................................. 57
Activate a Limit Line .......................................................................... 57
Creating a Limit (Point by Point) ..................................................... 58
Bandwidth ......................................................................... 61
Resolution Bandwidth Setting (RBW) ............................................. 61
Trace.................................................................................. 62
Selecting a Trace .................................................................................. 62
Trace Math............................................................................................ 64

BASIC OPERATION

26
BASIC OPERATION
Average Trace ...................................................................................... 65
Display .............................................................................. 67
Adjusting the LCD Brightness .......................................................... 67
Setting a Display Line (Reference Level Line) ............................... 67
Using the Video Out Port ................................................................. 68
Split Spectrum View ........................................................................... 68
Save/Recall Files ............................................................... 71
Save/Recall Setup................................................................................ 71
Save/Recall Trace Data ...................................................................... 72
Save/Recall Limit Lines ..................................................................... 73
Saving an Image File (Hardcopy) ..................................................... 73
Load Default Settings ......................................................................... 75
System Settings ................................................................. 76
System Information ............................................................................ 76
System Language ................................................................................. 77
27
GSP-730 User Manual
Description
The center frequency function sets the center frequency and centers the display to the center frequency.
Operation
1. Press
Frequency
>Center[F1] and enter the
frequency and unit.
Range: Default
0kHz~3GHz
1.5GHz
Display
Set center frequency

Frequency Settings

Center Frequency

28
BASIC OPERATION
Description
The start/stop frequency function will set the start and stop frequency of the span.
Operation
1. To set the start frequency, press
Frequency
>
Start[F2] and enter the frequency and unit.
2. To set the stop frequency, press
Frequency
>
Stop[F3] and enter the frequency and unit.
Range: Default Start frequency: Default Stop frequency:
0kHz~3GHz 0Hz 3GHz
Display
Start Frequency Stop Frequency
Note
The start and stop frequency can change when the span settings are used. The stop frequency must be set higher than the start frequency (for spans 0).

Start and Stop Frequency

29
GSP-730 User Manual
Description
The Step function sets the step size of the center frequency when using the arrow keys.
When the arrow keys are used to alter the center frequency, each press will move the center frequency by the step size specified by the Step function.
By default, the center frequency step size is equal to 10% of the span.
Operation
1. Press
Frequency
> Step[F4] and set the center
frequency step size.
Range:
1Hz~3GHz
Display
Step size

Center Frequency Step

30
BASIC OPERATION
Description
The Span function will set the frequency range of the sweep. The sweep will be centered around the center frequency.
Setting the span will alter the start and stop frequencies.
Operation
1. Press

Span

> Span[F1] and enter the span
frequency range and unit.
Range: Default Span:
0kHz~3GHz 3GHz
Display
Span
Set Span

Span Settings

Span
31

Full Span

Description
The Full Span function will set the span to the full frequency range.
This function will set the start and stop frequencies to 0Hz and 3GHz respectively.
Operation
1. Press
Span
> Full Span[F2].
Description
The Zero Span function will set the frequency range of the sweep to 0Hz and fixes the start and stop frequencies to the center frequency. The Zero Span function measures the time domain characteristics of the input signal at the center frequency. The horizontal axis is displayed in the time domain.
Operation
1. Press
Span
> Zero Span[F3].
The span changes accordingly.

Zero Span

GSP-730 User Manual
32
Display
Time domain
0Hz Span
Example: Amplitude modulation
Note
The measurement functions such as ACPR and OCBW are not available with the zero span setting:

Last Span

Description
The last span function returns the spectrum analyzer to the previous span settings.
Operation
1. Press
Span
> Last Span[F4].
BASIC OPERATION
33
GSP-730 User Manual
Description
The reference level defines the absolute level of the amplitude on the top graticule in voltage or power.
Operation
1. Press
Amplitude
> Ref. Level[F1] and enter the
reference level amplitude.
Range: Resolution:
-20dBm ~ 20dBm 10dBm
Display
Ref Level reading

Amplitude Settings

The vertical display scale is defined by the reference level amplitude, attenuation, scale and external gain/loss.

Reference Level

34
BASIC OPERATION
Description
The amplitude units can be set from dBm, dBmV or dBuV.
1. Press
Amplitude
> Units …[F3] to change the
amplitude units.
Units:
dBm, dBmV, dBuV
Description
Sets the logarithmic units for the vertical divisions.
Operation
1. Press
Amplitude
> Scale[F2] repeatedly to select the
vertical division units.
Scale Range:
10, 5, 2, 1
Display
Scale

Amplitude Units

Scale/Div

35
Operation
1. Press

Autoset

> Autoset[F1].
Autoset Range
Amplitude: Span:
Over the full amplitude range. Over the full span range.
Example:
Before Autoset, default state
After Autoset
GSP-730 User Manual
Autoset
The Autoset function searches the peak signals and picks the signal peak with the maximum amplitude, and then shows it in the display.

Using Autoset

36
BASIC OPERATION
Note
RBW setting is reset to Auto when the Autoset function is used.
Description
You can set the amplitude floor so that the signals lower than the setting will be ignored by the Autoset search.
Operation
1. Press
Autoset
> Amp.Floor[F2] and switch the
range from Auto to Man.
1. Enter the amplitude limit with the number pad
and Enter key.
Range:
50 to +20dBm
Note
See page 35 for setting the amplitude units.
Description
You can change the frequency span limit in the display to get a better view of the Autoset result. By default, the frequency span after Autoset is set at 3MHz.
Operation
1. Press
Autoset
> Span[F3] and switch the range
from Auto to Man.
2. Enter the span frequency for the Autoset
search.
Manual Range:
full amplitude range.

Limiting the Autoset Vertical Search Range

Limiting the Autoset Horizontal Search Range

37
GSP-730 User Manual

Marker

A Marker shows the frequency and amplitude of a waveform point. The GSP-730 can activate up to 5 markers or marker pairs simultaneously.
The marker table and peak table functions help editing and viewing multiple markers in a single display.
The delta marker function allows you to see the frequency and amplitude differences between reference markers.
The GSP-730 can automatically move a marker to various locations including the peak signal, center frequency, and start/stop frequency. Other marker operations regarding signal peaks are available in the Peak Search function.
Activating a Marker → from page 39 Activate a Delta Marker→ from page 40 Move Marker Manually→ from page 41 Move Marker to Preset Locations→ from page 42 Move Marker to Trace→ from page 43 Turn All Markers On or Off→ from page 44 Show Markers in Table→ from page 44 Peak Search→ from page 46 Move Marker to Peak→ from page 46 Move Marker and Peak to Center→ from page 46 Search for Peaks→ from page 47 Peak Table→ from page 48
38
BASIC OPERATION
Operation
1. Press
Marker
> Marker[F1] and select a marker
number.
Marker:
1~5
2. Press [F2] to turn the selected marker on.
3. Press Mode[F3] and set the marker mode to
Normal.
4. The display will show the marker on the trace
(centered by default) with the marker measurement at the top of the display.
Maker No., Frequency, Amplitude
Marker

Activating a Marker

There are two basic marker types, normal markers and delta markers. Normal markers are used to measure the frequency/time or amplitude of a point on the trace. Delta markers are used to measure the difference between a reference point and a selected point on the trace.
Activate a Normal Marker
39
GSP-730 User Manual
Description
Delta markers are marker pairs that measure the difference in frequency and amplitude between a reference marker and a delta marker.
When delta markers are activated, the reference and delta marker appear at the position of the selected marker, or in the center of the display if the selected marker has not yet be activated.
The marker measurement is located at the top
of the display, under the “normal marker”
measurement.
Delta Markers
Ref:
Reference marker, designated as 1.
Delta:
Delta marker, designated as 11.
Operation
1. Press
Marker
> Marker[F1] and select a marker
number.
2. Press [F2] to turn the selected marker on.
3. Press Mode[F3] to set the mode to Delta to
activate the delta marker.
Activate a Delta Marker
40
BASIC OPERATION
Delta maker No., Frequency, Amplitude
Delta
Marker
Operation
1. Press
Marker
> Marker[F1] and select a marker
number.
2. Use the left/right arrow keys to
move the marker one screen division at a time or the use the scroll wheel to move the marker in fine increments (one pixel at a time).
3. Alternatively, the numeric keypad
can be used to directly enter the frequency of the marker position.
7 8 9
654
1 2 3
/0
Move Marker Manually
41
GSP-730 User Manual
Preset conditions
The currently selected marker (normal marker or delta marker) can be moved to a number of preset positions:
Center: Peak Start: Stop: Step: Ref. Level:
Move to center frequency. Move to the highest peak. Move to start frequency. Move to stop frequency. Move to step frequency. Move to reference level amplitude.
Note
When a marker is moved to a preset position the span and other settings may be automatically changed.
Move marker to peak:
Press
Marker
> To Peak[F4].
Move marker to center:
Press
Marker
> To Center[F5].
Move marker to other positions
Press
Marker
> More[F6]>Marker to ...[F4] and
select one of the preset positions:
Marker to Start[F2] Marker to Stop[F3] Marker to Step[F4] Marker to Ref. Level[F5]
Move Marker to Preset Locations
42
BASIC OPERATION
Description
The Marker Trace function moves the selected marker to the currently active trace.
Operation
1. Press
Marker
> Marker[F1] and select a marker
number.
2. Press More[F6]>Marker Trace and select a trace
to assign the selected marker to. If Auto is selected, the selected marker is automatically assigned a trace.
Marker Trace:
Auto, A, B, C
3. In the example below, marker 1 is set to trace
B.
Marker 1, Trace B

Move Marker to Trace

43
GSP-730 User Manual
Description
All markers that have been activated, both normal and delta markers, can be turned off at the same time with the All Mrk Off function.
Operation
Press
Marker
> Marker[F1] >More[F6]>All Mrk
Off[F3] and turn all the markers off.
Description
The GSP-730 has a Marker Table function to show all the active markers and measurements at once.
Operation
1. Press
Marker
> Marker[F1] >More[F6]>Marker
Table[F2] and turn the marker table on.
2. The display will split into two screens. The
bottom half will show the Marker Table with the marker No. (normal, reference or delta), frequency and the amplitude of the marker.
Marker Table
Turn All Markers On or Off

Show Markers in Table

44
BASIC OPERATION
Description
While the Marker Table function is the active function, the position of each marker and delta marker can be edited within the marker table.
1. Use the arrow keys to move the cursor to the
frequency column of the desired marker.
Cursor
2. Enter the new position of the marker using the
keypad and units keys.
Edit Markers in Marker Table
45
GSP-730 User Manual
Description
Move the active marker to the highest peak. The highest peak can be either found once or continuously.
Operation
1. Press
Marker
> Marker[F1] and select a marker
number.
2. Press
Peak
Search
>Peak Search[F1]. The marker will
move to the highest signal peak.
3. To continually search for the peak each sweep,
press,
Peak
Search
>More[F6]>Peak Track[F1] and set
Peak Track to ON.
Description
The Peak to Center function moves the marker to the highest signal peak and moves the center frequency to that peak.
Operation
1. Press
Marker
> Select Marker[F1] and select a
marker number.
2. Press
Peak
Search
>Peak to Center[F5].
Note
The span will not be changed.

Peak Search

The Peak Search key is used to find trace peaks. The currently active marker is used in conjunction with the peak functions to mark the peaks that are found. Peaks can be sorted by frequency or amplitude in the peak table.
Move Marker to Peak
Move Marker and Peak to Center
46
Search for Peaks
Description
The
Peak
Search
key can be used to search for a
number of different peaks.
Peak Search
Next Peak:
Searches for next highest peak visible on the display.
Next Peak Right:
Searches for the next peak to the right of the marker.
Next Peak Left:
Searches for the next peak to the left of the marker.
Min Search:
Searches for the lowest peak.
Operation
1. Press
Marker
> Select Marker[F1] and select a
marker number.
2. Press
Peak
Search
and select the type of peak you
wish to find.
Example: Next Peak
BASIC OPERATION
47
GSP-730 User Manual
Example: Next Peak Right
Example: Next Peak Left
Description
The Peak Table function will display up to 5 peaks. The amplitude and frequency for each peak is listed.
Operation
1. Press
Peak
Search
>More [F6]>Peak Table[F1] and turn
the peak table on.

Peak Table

48
BASIC OPERATION
2. Press Peak Sort[F2] and set the sorting type:
Freq:
Amp:
Sort by frequency in ascending order. Sort by amplitude in ascending order.
Display
The bottom-half of the screen shows the peak table with the peak marker no., frequency and amplitude.
Peak Table
Peak markers
49
GSP-730 User Manual
Description
Channel analysis measurement includes ACPR (adjacent channel power) and OCBW (occupied bandwidth) measurements.
Parameters
Channel bandwidth
The frequency bandwidth the target channel occupies. Range: Between 0Hz~3GHz (0Hz excepted)
Channel Space
The frequency distance between each main channel. Range: Between 0Hz~3GHz
Adjacent channel bandwidth 1 & 2
The frequency bandwidth the adjacent channels occupy. Range: Between 0Hz~3GHz (0Hz excepted)

Measurement

This section describes how to use the automatic measurement modes. The GSP-730 includes the following measurements:
ACPR → from page 51. OCBW → from page 54.

Channel Analysis Overview

50
Adjacent channel offset 1 ~ 2
The frequency distance between the adjacent channels and main channel. Range: 1 Between 0Hz~3GHz (0Hz excepted)
OCBW%
The ratio of occupied bandwidth to the amount of power consumed. Range: 0% to 100%,
0.1% resolution.
ACPR
Description
Adjacent channel power refers to the amount of power leaked to the adjacent channel from the main channel. This measurement is a ratio of the main channel power to power in the adjacent channel.
Example
Main
CHBW
ADJ
CH1
ADJ
CH2
ADJ
CH1
ADJ
CH2
Offset 1 Offset 2
Channel spacing
To next main
channel
Operation: Setting up the main channel
1. Press
Meas
> ACPR[F2] and turn ACPR on.
Any other measurement mode will automatically be
BASIC OPERATION
51
GSP-730 User Manual
disabled.
2. The display splits into two screens. The top
screen shows the sweep waveform. The bottom screen shows the ACPR settings and measurement results in real time.
Turn ACPR off to return back to the normal mode.
Channel setup
Measurements
3. Press Channel Setup…[F1] and set the
following:
Main CH BW[F1]
Set the bandwidth of the main channel.
Main CH Space[F2]
Specify the channel spacing.
Note
The main channel bandwidth and space settings are shown in the setup area at the bottom of the screen, not on the soft-key icon.
Channel BW
Channel space
52
BASIC OPERATION
Operation: Setting up the adjacent channel(s)
1. Press ADJCH Setup[F3] to setup the adjacent
channels:
Adj CH BW 1[F1]
Sets the bandwidth of the 1st adjacent channel.
Adj CH Offs 1[F2]
Sets the channel offset of the 1st adjacent channel.
Adj CH BW 2[F3]
Sets the bandwidth of the 2nd adjacent channel.
Adj CH Offs 2[F4]
Sets the channel offset of the 2nd adjacent channel.
Note
The adjacent channel bandwidth and space settings are shown in the setup area at the bottom of the screen, not on the soft-key icons.
Adj CH1 settings
Adj CH2 settings
Move Channels Up/Down
1. Press the
Meas
again or press Return[F6] repeatedly to return to the start of the Measure menu tree.
2. Press CH Up[F5] to go to the next main
channel.
3. Press CH Down[F6] to go to the previous main
channel.
Note
The channel space (Main CH Space) setting determines where the next main channel is located.
53
OCBW
Description
Occupied bandwidth measurements are used to measure the power of the occupied channel as a percentage to the power of the channel.
Example
CH BW
OCBW
Operation: Setting up the main channel
1. Press
Meas
> OCBW %[F3] and turn OCBW
on.
Any other measurement mode will automatically be
disabled.
2. The display splits into two screens. The top
shows the channel bandwidth. The bottom screen shows the OCBW measurement results in real time.
Turn OCBW off to return back to the normal mode.
GSP-730 User Manual
54
BASIC OPERATION
Channel power and OCBW
power results
CH BW
3. Press Channel Setup[F1] and set the
following:
Main CH BW[F1]
Set the bandwidth of the main channel.
Main CH Space[F2]
Specify the channel spacing.
Note
The main channel bandwidth and space settings are shown in the setup area at the bottom of the screen, not on the soft-key icon.
Channel BW
Channel space
55
GSP-730 User Manual
Move Channels Up/Down
1. Press
Meas
again or press Return[F6] repeatedly to return to the start of the Measure menu tree.
2. Press CH Up[F5] to go to the next main
channel.
3. Press CH Down[F6] to go to the previous main
channel.
Note
The channel space (Main CH Space) setting determines where the next main channel is located.
56
BASIC OPERATION
Operation
1. Press
Limit Line
to enter the Limit Line menu.
2. Press H Limit[F1] or L Limit[F2] and turn the
high or low limit on or off.
H Limit/L Limit:
On, Off
The H Limit is shown as a blue line. The L Limit is shown as a pink line.
High limit line
Low Line

Limit Line Testing

The Limit Line function is used to set the upper or lower amplitude limits over the entire frequency range. The limit lines can be used to detect whether the input signal is above, below or within the limit lines.
The limit lines can be manually edited using 10 frequency points from the start to the stop frequencies.
To save and recall limit lines, please see page 73.

Activate a Limit Line

57
GSP-730 User Manual
Description
Create a limit manually, point by point. Ten manually selected frequency points can be used to create the upper or lower limit line.
Operation
1. Press
Limit Line
> Limit[F1] and select the limit line you wish to edit.
Limit:
High, Low
2. Press Edit Table[F2], and turn the edit table on.
The GSP-730 is split into two screens. The top screen shows the trace and the selected limit line (high or low) and the bottom screen shows the limit line table.
Spectrum display
Limit Line Table
All 10 points will be displayed in a limit line table at the bottom of the display. By default, each point is set to 0dBm.
3. Use the arrow keys to move the cursor to the

Creating a Limit (Point by Point)

58
BASIC OPERATION
frequency column of the desired point.
Cursor
4. Enter the new frequency and amplitude of the
point using the keypad and the unit keys.
5. Repeat steps 3-5 for the remaining points (A
maximum of ten points).
6. To delete the selected point, press Delete [F3].
7. To delete all the points, press Delete All [F4].
The points will revert to their default frequency and
amplitude values.
8. To delete a point from the editing table, press
Delete.
The whole points, including the frequency value and
the amplitude are removed. This command reduces the number of points used in the limit line.
9. Press Undelete[F5] the restore the last point that
was deleted.
Description
The Pass/Fail testing uses the limit lines as boundaries. When the input signal escapes the boundary of the limit lines, then the test is judged as a FAIL, if the signal stays within the boundary, the test is judged as a PASS.
Pass/Fail Testing
59
GSP-730 User Manual
Pass:
Upper limit
Lower limit
Fail:
Upper limit
Lower limit
Note
Before pass/fail testing can begin, limit lines for the upper and/or lower limits must first be saved and activated. See the page 57.
Operation
1. Press
Limit Line
>Pass/Fail[F4] to turn the testing on or off.
2. The test result is updated in real-time at the
bottom of the display.
Pass:
Fail:
Note
At least one limit line (high or low) must be turned on to enable testing.
If only high limit line is on, each trace point has to be lower than the high limit line to get a PASS judgment, otherwise get the judgment will be FAIL.
Conversely, if only low limit line is on then each trace point has to be higher than the low limit line to get a PASS, otherwise get the test will be judged a FAIL.
60
BASIC OPERATION
Description
The RBW (Resolution Bandwidth) defines the width of the IF (intermediate frequency) filter that is used to separate signal peaks from one another. The narrower the RBW, the greater the capability to separate signals at close frequencies. But it also makes the sweep time longer under specific frequency spans (the display is updated less frequently).
Operation
1. Press
BW
> RBW[F1] and set the RBW to
Auto or Man.
2. Set the resolution bandwidth and unit for Man
mode.
Mode: Frequency Range:
Auto, Man 1MHz, 300kHz, 100kHz
Note
The manual RBW is only available when the span10MHz. If the span is greater than 10MHz then the RBW is automatically set to Auto.

Bandwidth

BW key sets the resolution bandwidth (RBW). The resolution bandwidth and the sweep time are related. Please take into account how the sweep time is effected by the resolution bandwidth.

Resolution Bandwidth Setting (RBW)

61
GSP-730 User Manual
Description
Each trace (A, B, C) is represented by a different color. Trace A is green, trace B is orange and trace C is yellow. When activated, an icon for each trace color and function is shown at the bottom of the display. When a trace is selected, parameters can be set/edited from the Trace menu.
Display Icons
Trace A, B, C
Trace Type & Icon
The type of trace used determines how the trace data is stored or manipulated before being displayed. The analyzer updates each trace according to the type of trace used.
Clear &
Write
The GSP-730 continuously updates the display with each sweep. This is the default trace type.

Trace

The GSP-730 is able to set the parameters of up to 3 different traces on the display at once. Each trace is represented by a different color and is updated with each sweep.
To save or recall traces to/from memory, see page 71.

Selecting a Trace

62
BASIC OPERATION
Peak Hold
Min Hold
The maximum or minimum points are maintained for the selected trace. The trace points are updated each sweep if new maximum or minimum points are found.
View
View will hold the selected trace and stop updating the trace data for the selected trace. Pressing View[F4] will display the trace data that was cleared using the Blank[F5] key.
Blank
Clears the selected trace from the display and stores trace data. The trace data can be restored by pressing View[F4].
Operation
1. Press
Trace
> Trace[F1] and choose a trace.
Trace:
A, B, C
2. Select the trace type:
Clear & Write[F2] Peak Hold[F3] View[F4] Blank[F5] More[F6]>Min Hold[F1]
Note
Traces B and C are set to
Blank
by default.
63

Trace Math

Description
Performs trace math from two traces (A, B) and stores the results in trace A or swaps the data from trace A to trace B.
Math functions
A <--> B
Swaps the data from trace A to B and vice versa.
A + B -> A
Adds trace A and B and stores the result in trace A.
A – B -> A
Subtracts trace B from trace and stores the result in trace A.
A + const ->A
Adds an offset to trace A.
A – const ->A
Subtracts an offset from trace A.
Operation
1. Press
Trace
>Trace Math..[F3] and select a
trace math function.
A <--> B
[F1]
A + B -> A
[F2]
A – B -> A
[F3]
A + const ->A
[F4]
A – const ->A
[F5]
2. If A + const ->A or A + const ->A was selected,
set the constant (offset value).
Constant:
-40dBm ~ 40dBm
GSP-730 User Manual
64

Average Trace

Description
The Average function averages the currently selected trace for a user-defined number of times before it is displayed. This feature smoothes the noise level, but has the drawback of slowing down the display update rate.
Operation
1. Press
Trace
> More[F6] and toggle Avg on.
2. Set the number of averages.
Range: Default:
4 ~ 100 4
BASIC OPERATION
65
Example:
Average:Off
Average: On (8)
GSP-730 User Manual
66
BASIC OPERATION
Description
The LCD brightness levels can be adjusted to five pre-set levels.
Operation
1. Press

Display

> LCD Dimmer[F1] and use either
the number pad, the scroll wheel or arrow keys to set the brightness.
Description
The Display Line function is used to super­impose a reference level line over the traces.
Operation
1. Press
Display
> Display Line[F3] to turn the
display line on.
2. Set the display line level and press Enter.
Example:
Display line
Display line set at -50dBm
Display
The Display key configures the basic display settings as well as the split screen modes.

Adjusting the LCD Brightness

Setting a Display Line (Reference Level Line)

67
GSP-730 User Manual
Description
The GSP-730 has a dedicated VGA terminal to output the display to an external monitor. The video output is always on.
Output resolution
480 x 640 (fixed)
Operation
1. Connect an external monitor to
the rear panel VGA terminal.
Description
The split spectrum view is able to view two different sweep ranges on the display at the same time using a split screen view. The top and bottom view can have independent sweep ranges, amplitudes, spans and other settings. However only one split screen (top or bottom) can be swept each time.
Example
Lower
spectrum
Upper
spectrum

Using the Video Out Port

Split Spectrum View

68
BASIC OPERATION
Split spectrum functions
Half-Upper
Half-Upper will put the spectrum analyzer into split screen mode. It will make the top sweep the active sweep and pause the bottom sweep. When Half-Upper is on, only the upper sweep parameters can be edited.
Half-Lower
Half-Lower will put the spectrum analyzer into split screen mode. It will make the bottom sweep the active sweep and pause the top sweep. When Half-Lower is on, only the lower sweep settings can be edited.
Alternate Sweep
This setting will alternate the sweep between the bottom and top spectrums. If alternate sweep is turned on, only the upper sweep parameters can be edited.
69
GSP-730 User Manual
Operation
1. Press
Display
>Half-Upper[F4] or Half-Lower[F5]
or Alternate Sweep[F6] to enable the split spectrum view.
Turning Half-Upper on will automatically turn
Half-Lower off.
Turning Half –Lower on will automatically turn
Half-Upper off .
If Alternate Sweep is turned on, each sweep will
alternate, but only the upper sweep parameters can be edited.
2. To return to a full-screen, single spectrum
display, press Full Display[F2].
Note
After exiting the split spectrum view, the analyzer will use the settings from the active window. The settings for the inactive screen will be retained for the next time that split spectrum view is used.
If the spectrum analyzer was in the Alternate mode, then the upper sweep settings will be returned.
70
BASIC OPERATION
Description
Setup data contains all the data necessary to recall the state of the GSP-730 to known state.
Setup data contains the following data:
Center frequency, Start frequency, Stop frequency, Step frequency, Ref. Level, Scale, Units, RBW
Save
1. To save the current settings, press
Memory
> Setup To[F1] and choose a memory location to save to with the arrow keys.
Setup To:
1~5
2. Press
Enter
kHz /
µSec
to execute the save.
Recall
1. To recall a setup, press
Memory
> Setup From[F2] and choose a memory location to recall from with the arrow keys.
Setup From:
1~5
2. Press
Enter
kHz /
µSec
to execute the recall.

Save/Recall Files

The GSP-730 can save and recall setup data, trace data and limit line data to and from internal memory. There are five memory locations for each save file type. These files cannot be saved to USB.
The Hardcopy key can be used to save image files to a USB flash drive.

Save/Recall Setup

71
GSP-730 User Manual
Description
The trace data can be saved/recalled for any of the A, B or C traces to/from one of 5 pre-set internal memory locations. The trace data cannot be recalled or saved to USB.
When saving or recalling trace data from a split spectrum, only the active spectrum is saved/recalled.
Save
1. To save the current trace data, press
Memory
>
Save Trace Data..[F3]
2. Press Source Trace[F1] and select the source:
Source:
A, B, C
3. Press Destination[F2] and select the memory
location to save to:
Destination:
1~5
4. Press Start[F5] to save the selected trace data.
Recall
5. To recall trace data, press
Memory
> Recall Trace
Data..[F3]
6. Press Source Trace[F1] and select the memory
location to recall from:
Source:
1~5
7. Press Destination[F2] and select the destination
trace
Destination:
A, B, C

Save/Recall Trace Data

72
BASIC OPERATION
8. Press Start[F5] to recall the selected trace data.
Description
Upper and lower limit lines can be saved to one of 5 pre-set internal memory locations. The limit line data cannot be saved to USB.
Save
1. To save the current upper and lower limit
lines, press
Memory
> Limitln to[F5] and choose a memory location to save to with the arrow keys.
Limit line:
1~5
2. Press
Enter
kHz /
µSec
to execute the save.
Recall
3. To recall pre-saved upper and lower limit
lines, press
Memory
> Limitln from[F6] and choose a memory location to recall from with the arrow keys.
Limit line:
1~5
4. Press
Enter
kHz /
µSec
to execute the recall.
Description
The Hardcopy key can be used to save a screenshot of the display to a USB flash drive. The screen shot is saved as a bitmap file.

Save/Recall Limit Lines

Saving an Image File (Hardcopy)

73
Operation
1. Insert a USB flash drive into the USB port.
2. Press
Hardcopy
and the image file will begin
saving.
Wait a few moments for the file to save. When the
file has finished saving, “Screen Saved OK” will appear at the bottom of the display.
Note
The file name will be automatically created in the following format:
File name: SCRXX.bmp
Where XX is a number that is incremented each time the file is saved.
Warning
Do not remove the USB drive until the file has completed saving.
Hardcopy Setup
Description
The Hardcopy Setup key is used set the image file properties of the bitmap file that is created when the Hardcopy key is pressed.
Ink Normal:
This is the normal, default image setting.
Ink Saving:
This will invert all the colors on the display so that the file will conserve ink when printed.
Operation
1. Press
Hardcopy
Setup
and choose the image type:
Ink Normal[F1]
Ink Saving[F2]
GSP-730 User Manual
74
BASIC OPERATION
Note
The next time the Hardcopy key is pressed, the image will be saved using the settings above.
Description
The Preset key is used to load the default settings. The default settings are listed in the appendix on page 106.
Operation
1. Press
Preset
.
The system will load the preset settings and the
screen will update with the new settings.
Note
The default settings cannot be changed.

Load Default Settings

75
GSP-730 User Manual
Description
The System Information displays the following:
Serial Number:
XX digit serial number
HW Version:
Hardware version
FW Version:
Firmware version
SW Version:
Software version
Language:
Shows the language number as seen in the System>Language menu.
Operation
1. Press
System
>Information[F4] to display the
system information.
The system information will be displayed on the
system menu soft-keys.
Example
System
information

System Settings

System Information

76
BASIC OPERATION
Description
The language option sets the icon display language.
Operation
1. Press
System
>Language…[F3] to bring up the
Language menu.
2. Choose a system language. The language
number is the number that will be displayed in the system information.
Language 1 English
Sets the language to English
Language 2 Chinese S
Sets the language to simplified Chinese

System Language

77
GSP-730 User Manual
Interface Configuration ...................................................... 80
Configure Remote Interface ............................................................... 80
Remote Control Function Check ...................................................... 81
Command Syntax ............................................................... 82
Command List ................................................................... 85
IEEE488.2 Standard Commands ...................................................... 87
*IDN? ..................................................................................................... 87
Sweep Commands ................................................................................ 87
si............................................................................................................... 87
sn ............................................................................................................. 87
ts .............................................................................................................. 88
Frequency Commands ......................................................................... 88
meas:freq:cen ......................................................................................... 88
meas:freq:st ............................................................................................ 88
meas:freq:stp.......................................................................................... 89
Span Commands ................................................................................... 89
meas:span ............................................................................................... 89
meas:span:full ........................................................................................ 90
Amplitude Commands ........................................................................ 90
meas:refl:unit ......................................................................................... 90
meas:refl ................................................................................................. 90
Marker and Peak Search Commands ................................................ 91
meas:mark:on ........................................................................................ 91
meas:mark:off ........................................................................................ 92
meas:mark:norm ................................................................................... 92
meas:mark:norm:freq? ......................................................................... 92
meas:mark:norm:level? ........................................................................ 93
meas:mark:delta .................................................................................... 93

REMOTE CONTROL

This chapter describes basic configuration of IEEE488.2 based remote control. For a command list, refer to the programming manual, downloadable from GW Instek website, www.gwinstek.com
78
REMOTE CONTROL
meas:mark:delta:freq? .......................................................................... 93
meas:mark:delta:level? ......................................................................... 94
meas:mark:tomin .................................................................................. 94
meas:mark:topeak ................................................................................. 94
meas:mark:tonp .................................................................................... 95
meas:mark:trace .................................................................................... 95
Trace Commands ................................................................................. 95
meas:tra:val1:val2 .................................................................................. 95
meas:tra:avg:on ..................................................................................... 96
meas:tra:avg:off ..................................................................................... 96
meas:tra:read.......................................................................................... 97
Power Measurement Commands ...................................................... 97
meas:acpr ............................................................................................... 97
meas:acpr:lower? ................................................................................... 98
meas:acpr:upper? .................................................................................. 98
meas:ocbw ............................................................................................. 98
meas:ocbw:bw? ..................................................................................... 99
meas:ocbw:chpw? ................................................................................. 99
Limit Line Commands ........................................................................ 99
meas:lmtline:passfail ............................................................................ 99
meas:lmtline:on ................................................................................... 100
meas:lmtline:off .................................................................................. 100
BW Commands .................................................................................. 100
con:rbw:auto........................................................................................ 100
con:rbw? ............................................................................................... 101
con:rbw:man ........................................................................................ 101
con:rbw:mode?.................................................................................... 101
con:swt? ................................................................................................ 102
Display Commands ............................................................................ 102
con:disp:split:upper ............................................................................ 102
con:disp:split:lower ............................................................................ 102
con:disp:split:alt .................................................................................. 102
con:disp:split:full ................................................................................. 103
Preset Commands .............................................................................. 103
con:preset ............................................................................................. 103
System Commands............................................................................. 103
con:sys:ser? .......................................................................................... 103
79
GSP-730 User Manual
USB configuration
PC side connector
Type A, host
GSP side connector
Rear panel Type B, slave
Speed
1.1/2.0 (full speed)
RS232 configuration
PC side connector
RS232 male port
GSP side connector
RS232 female port
Baud Rate:
9600, 19200, 38400, 57600,
115200.
Parity:
None, Even, Odd, Space, Mark, Multidrop.
Stop bit:
1, 1.5, 2.
Data bit:
5, 6, 7, 8
Description
The GSP-730 can use either the type B USB port or the RS232 on the rear panel for remote control.
When using the USB B port, the GSP-730 uses a USB driver to simulate an RS232 connection with a PC via USB. It is these RS232 settings that are configured for remote control.
Before using the USB B port for remote control, please install the USB driver. See page 20 for details.

Interface Configuration

Configure Remote Interface

80
REMOTE CONTROL
Panel operation
1. USB Connection:
Connect a USB cable from the PC to the rear panel USB B port.
RS232 Connection: Connect an RS232C cable from the PC to the rear panel RS232 port.
2. Press
System
>Serial Port…[F1]> Serial[F1] to
enter the remote configuration.
3. Set the following RS-232 settings using the
arrow keys:
Baud Rate[F1]:
9600, 19200, 38400, 57600,
115200.
Parity[F2]:
None, Even, Odd, Space, Mark, Multidrop.
Stop Bit[F3]:
1, 1.5, 2.
Data[F4]:
5, 6, 7, 8
Functionality check
Invoke a terminal application such as MTTTY (Multi-Threaded TTY).
To check the COM port No., see the Device Manager in the PC. For WinXP go to; Control panel → System → Hardware tab.
Run this query command via the terminal after the instrument has been configured for remote control (page 80).
*idn?

Remote Control Function Check

81
GSP-730 User Manual
This should return the Manufacturer, Model number, Serial number, and Firmware version in the following format.
GW-INSTEK, GSP-730, XXXXXXXX, V.VV
Manufacturer: GW-INSTEK Model number : GSP-730 Serial number : XXXXXXXXXXXX Firmware version : V.VV
Note
For further details or if you have trouble running this function check, please see the programming manual, available on the GW Instek web site @ www.gwinstek.com.
Compatible Standard
IEEE488.2
Partial compatibility
SCPI, 1999
Partial compatibility
Command Structure
SCPI (Standard Commands for Programmable Instruments) commands follow a tree-like structure, organized into nodes. Each level of the command tree is a node. Each keyword in a SCPI command represents each node in the command tree. Each keyword (node) of a SCPI command is separated by a colon (:).
For example, the diagram below shows an SCPI sub-structure and a command example.
meas
:freq
:stp?
:cen?
:st?
meas:freq:cen?

Command Syntax

82
REMOTE CONTROL
Command types
There are a number of different instrument commands and queries. A command sends instructions or data to the unit and a query receives data or status information from the unit.
Command types
Single Command
A single command with/without a parameter
Example
meas:freq:cen 100 MHz
Query
A query is a simple or compound command followed by a question mark (?). A parameter (data) is returned.
Example
meas:freq:cen?
Command Format
100 k h z
1 2 3 4 5
Meas:f r eq:cen
1. Command header
2. Space
3. Parameter 1
4. Optional space
5. Unit or suffix.
Common Input/Return Parameters
Type
Description
Example
<Boolean>
Boolean logic
0, 1
<NR1>
integers
0, 1, 2, 3
<NR2>
decimal numbers
0.1, 3.14, 8.5
<NRf>
any of NR1, 2
1, 1.5
83
GSP-730 User Manual
<freq>
<NRf> + unit
2.5 mhz
Unit = kHz, MHz, GHz. Note: The unit can be omitted (defaults to currently set unit).
<refl>
<NRf> + unit
-30 dBm
Unit = dBm, dBmV, dBuV Note: The unit can be omitted (defaults to currently set unit).
<ampl>
NR3 +unit
30.0 dBm
Note: The unit can be omitted. (Unit defaults to current unit)
<trace data>
{ -92, -91, ……., -89, -92, -92, -91 }
CSV data that represents each point in a trace.
<string>
ASCII string data.
Message Terminator
LF
Line feed code (0x0A)
84
REMOTE CONTROL
IEEE488.2 Standard Commands
*IDN? ....................................................................................... 87
Sweep Commands
si ................................................................................................. 87
sn ................................................................................................ 87
ts ................................................................................................. 88
Frequency Commands
meas:freq:cen ........................................................................... 88
meas:freq:st............................................................................... 88
meas:freq:stp ............................................................................ 89
Span Commands
meas:span.................................................................................. 89
meas:span:full ........................................................................... 90
Amplitude Commands
meas:refl:unit ............................................................................ 90
meas:refl .................................................................................... 90
Marker Commands
meas:mark:on ........................................................................... 91
meas:mark:off .......................................................................... 92
meas:mark:norm ...................................................................... 92
meas:mark:norm:freq? ............................................................ 92
meas:mark:norm:level? ........................................................... 93
meas:mark:delta ....................................................................... 93
meas:mark:delta:freq? ............................................................. 93
meas:mark:delta:level? ............................................................ 94
meas:mark:tomin ..................................................................... 94
meas:mark:topeak .................................................................... 94
meas:mark:tonp ....................................................................... 95
meas:mark:trace ....................................................................... 95
Trace commands
meas:tra:val1:val2 .................................................................... 95
meas:tra:avg:on ........................................................................ 96
meas:tra:avg:off ........................................................................ 96
meas:tra:read ............................................................................ 97
Power measurement
meas:acpr .................................................................................. 97
meas:acpr:lower? ..................................................................... 98

Command List

85
GSP-730 User Manual
commands
meas:acpr:upper? ..................................................................... 98
meas:ocbw ................................................................................ 98
meas:ocbw:bw? ........................................................................ 99
meas:ocbw:chpw? .................................................................... 99
Limit Line commands
meas:lmtline:passfail................................................................ 99
meas:lmtline:on ......................................................................100
meas:lmtline:off .....................................................................100
BW commands
con:rbw:auto ...........................................................................100
con:rbw? ..................................................................................101
con:rbw:man ...........................................................................101
con:rbw:mode? .......................................................................101
con:swt? ...................................................................................102
Display commands
con:disp:split:upper ...............................................................102
con:disp:split:lower................................................................102
con:disp:split:alt .....................................................................102
con:disp:split:full ....................................................................103
Preset commands
con:preset ................................................................................103
System commands
con:sys:ser? .............................................................................103
86
REMOTE CONTROL
*IDN? ....................................................................................... 87
*IDN?
Query
Description
Queries the manufacturer, model number, serial number, and firmware version of the instrument.
Query Syntax
*IDN?
Return parameter
<string>
Returns the instrument identification as a string in the following format:
GW-INSTEK, GSP-730, XXXXXXXX, V.VV
Manufacturer: GWINSTEK
Model number : GSP-730
Serial number : XXXXXXXX Firmware version : V.VV
si ................................................................................................. 87
sn ................................................................................................ 87
ts ................................................................................................. 88
si
Set
Description
Stops the sweep.
Example
si
sn
Set
Description
Continues a stopped sweep.
Example
sn

IEEE488.2 Standard Commands

Sweep Commands

87
GSP-730 User Manual
ts
Set
Description
Resets the sweep and starts it once (i.e., sweeps one time).
Example
ts
meas:freq:cen............................................................................ 88
meas:freq:st ............................................................................... 88
meas:freq:stp ............................................................................ 89
meas:freq:cen
Set
Query
Description
Sets or queries the center frequency.
Syntax
meas:freq:cen <freq>
Query Syntax
meas:freq:cen?
Parameter
<freq>
Center frequency.
Return parameter
<freq>
Returns the frequency and unit.
Example
meas:freq:cen 100 khz
Sets the center frequency to 100kHz.
Query example
Meas:freq:cen?
>100 kHz
meas:freq:st
Set
Query
Description
Sets or queries the start frequency.
Syntax
meas:freq:st <freq>
Query Syntax
meas:freq:st?

Frequency Commands

88
REMOTE CONTROL
Parameter
<freq>
Start frequency
Return parameter
<freq>
Returns the start frequency and unit
Example
meas:freq:st 100 mhz
Sets the start frequency to 100MHz
Query Example
meas:freq:st?
> 100000 kHz
meas:freq:stp
Set
Query
Description
Sets or queries the stop frequency.
Syntax
meas:freq:stp <freq>
Query Syntax
meas:freq:stp?
Parameter
<freq>
Stop frequency
Return parameter
<freq>
Returns the stop frequency and unit
Example
meas:freq:stp 100 mhz
Sets the stop frequency to 100MHz
Query Example
meas:freq:stp?
> 100000 kHz
meas:span.................................................................................. 89
meas:span:full ........................................................................... 90
meas:span
Set
Query
Description
Sets or queries the frequency span.
Syntax
meas:span <freq>
Query Syntax
meas:span?

Span Commands

89
GSP-730 User Manual
Parameter
<freq>
Span frequency range
Return parameter
<freq>
Returns the span and unit
Example
meas:span 10 mhz
Sets the span to 10MHz
Query Example
meas:span?
> 10000.0 kHz
meas:span:full
Set
Description
Sets the span to the full span.
Syntax
meas:span:full
meas:refl:unit ............................................................................ 90
meas:refl .................................................................................... 90
meas:refl:unit
Set
Query
Description
Sets the reference level unit.
Syntax
meas:refl:unit {1|2|3}
Query Syntax
meas:refl:unit?
Parameter/ Return parameter
1
dBm
2
dBmV
3
dBuV
Query Example
Meas:refl:unit?
>1
The reference level units are dBm.
meas:refl
Set
Query
Description
Sets or queries the reference level.

Amplitude Commands

90
REMOTE CONTROL
Syntax
meas:refl <refl>
Query Syntax
meas:refl?
Parameter
<refl>
Reference level in the currently selected unit (from the meas:refl:unit command).
Return parameter
<refl>
Returns reference level and unit.
Example
meas:refl 10
Sets the reference level to 10 dBm (for unit = dBm).
Query Example
Meas:refl?
>10 dBm
meas:mark:on ........................................................................... 91
meas:mark:off .......................................................................... 92
meas:mark:norm ...................................................................... 92
meas:mark:norm:freq? ............................................................ 92
meas:mark:norm:level? ........................................................... 93
meas:mark:delta ....................................................................... 93
meas:mark:delta:freq? ............................................................. 93
meas:mark:delta:level? ............................................................ 94
meas:mark:tomin ..................................................................... 94
meas:mark:topeak .................................................................... 94
meas:mark:tonp ....................................................................... 95
meas:mark:trace ....................................................................... 95
meas:mark:on
Set
Query
Description
Sets or queries which markers are turned on.
Syntax
meas:mark:on {<NR1>|all}
Query Syntax
meas:mark:on <NR1>?
Parameter
<NR1>
Marker number 1~ 5.
all
All markers.
Return parameter
ON
The selected marker is on.
OFF
The selected marker is off.

Marker and Peak Search Commands

91
GSP-730 User Manual
Example
meas:mark on 1
Turns marker 1 on.
Query Example
Meas:mark 1?
>OFF
meas:mark:off
Set
Description
Sets which markers are turned off.
Syntax
meas:mark:off {<NR1>|all}
Parameter
<NR1>
Marker number 1~ 5.
All
All markers.
Example
meas:mark off 1
Turns marker 1 off.
meas:mark:norm
Set
Description
Sets the selected marker to normal mode.
Syntax
meas:mark:norm <NR1>
Parameter
<NR1>
Marker number 1~ 5.
Example
meas:mark:norm 1
Sets marker 1 to normal mode.
meas:mark:norm:freq?
Set
Query
Description
Queries the frequency of the selected normal marker.
Query syntax
meas:mark:norm:freq <NR1>?
Parameter
<NR1>
Marker number 1~ 5.
92
REMOTE CONTROL
Return parameter
<freq>
Returns the frequency and unit of the selected marker.
Example
meas:mark:norm:freq 1?
>1.5GHz.
meas:mark:norm:level?
Set
Query
Description
Queries the amplitude of the selected normal marker.
Query syntax
meas:mark:norm:level <NR1>?
Parameter
<NR1>
Marker number 1~ 5.
Return parameter
<amp>
Returns the amplitude and unit of the selected marker.
Example
meas:mark:norm:level 1?
>10.0dBm.
meas:mark:delta
Set
Description
Sets the selected marker to delta mode. It also sets the relative frequency of the delta marker (in relation to the normal marker frequency).
Syntax
meas:mark:delta <NR1> <freq>
Parameter
<NR1>
Marker number 1~ 5.
<freq>
Relative frequency of the delta marker.
Example
meas:mark:freq 1 10 MHz
Turns delta marker 1 on and sets its offset to 10MHz.
meas:mark:delta:freq?
Set
Query
Description
Queries the (relative) frequency of the selected delta marker.
Query syntax
meas:mark:delta:freq <NR1>?
93
GSP-730 User Manual
Parameter
<NR1>
Marker number 1~ 5.
Return parameter
<freq>
Returns the relative frequency and unit of the selected delta marker.
Example
meas:mark:norm:freq 1?
>12.0kHz.
meas:mark:delta:level?
Set
Query
Description
Queries the amplitude of the selected delta marker.
Query syntax
meas:mark:delta:level <NR1>?
Parameter
<NR1>
Marker number 1~ 5.
Return parameter
<amp>
Returns the amplitude and unit of the selected delta marker.
Example
meas:mark:delta:level 1?
>10.0dBm.
meas:mark:tomin
Set
Description
Sets the selected marker to the minimum peak.
Syntax
meas:mark:tomin <NR1>
Parameter
<NR1>
Marker number 1~ 5.
Example
meas:mark:tomin 1
Sets marker 1 to the minimum peak.
meas:mark:topeak
Set
Description
Sets the selected marker to the peak.
Syntax
meas:mark:topeak <NR1>
Parameter
<NR1>
Marker number 1~ 5.
Example
meas:mark:topeak 1
Sets marker 1 to the peak.
94
REMOTE CONTROL
meas:mark:tonp
Set
Description
Moves the selected normal or delta marker to the next peak.
Syntax
meas:mark:tonp <NR1>
Parameter
<NR1>
Marker number 1~ 5.
Example
meas:mark:tono 1
Moves marker 1 to the next peak.
meas:mark:trace
Set
Description
Sets the selected marker to the selected trace.
Syntax
meas:mark:topeak <NR1> <trace>
Parameter
<NR1>
Marker number 1~ 5.
<trace>
0
Auto (auto assign a trace)
1
Trace A
2
Trace B
3
Trace C
Example
meas:mark:trace 1 2
Sets marker 1 to trace B.
meas:tra:val1:val2 .................................................................... 95
meas:tra:avg:on ........................................................................ 96
meas:tra:avg:off ........................................................................ 96
meas:tra:read ............................................................................ 97
meas:tra:val1:val2
Set
Description
Sets the mode for the selected trace.
Syntax
meas:tra <trace><mode>

Trace Commands

95
GSP-730 User Manual
Parameter
<trace>
1
Trace A
2
Trace B
3
Trace C
<mode>
1
Clear and write mode
2
Peak hold mode
3
View mode
4
Blank mode
5
Minimum hold mode
Example
meas:tra 1 1
Sets trace A to clear and write mode.
meas:tra:avg:on
Set
Description
Turns the average function on and sets the number of averages for the slected trace.
Syntax
meas:tra:avg:on <trace> <NR1>
Parameter <trace>
1
Trace A
2
Trace B
3
Trace C
<NR1>
4~20
Number of averages.
Example
meas:tra:avg:on 1 4
Sets the number of averages used for Trace A to 4.
meas:tra:avg:off
Set
Description
Turns the average function off for the slected trace.
Syntax
meas:tra:avg:on <trace>
Parameter
<trace>
1
Trace A
2
Trace B
3
Trace C
all
All traces
Example
meas:tra:avg:off all
Turns the average function off for all the traces.
96
REMOTE CONTROL
meas:tra:read
Set
Query
Description
Returns the all the trace data for the selected trace.
Query syntax
meas:tra:read? <trace>
Parameter
<trace>
1
Trace A
2
Tr ace B
3
Trace C
all
All traces
Return parameter
<trace data>
Comma separated data values encapsulated in brackets. i.e., {-92, -91, -90, ………-81}
Example
meas:tra:read? 1 >{ -92, -91, -90, -90, -90, -88, ……., -89, -92, -92, -91 }
Returns the trace data for the selected trace(s). A total of 501 trace points are returned, from the start
frequency to the stop frequency. If “all” is selected, the
trace data is returned in three lots, {trace A}{trace B}{traceC}. The units are in decibels. If the selected trace is not active, 0s will be returned for each trace point.
meas:acpr .................................................................................. 97
meas:acpr:lower? ..................................................................... 98
meas:acpr:upper? ..................................................................... 98
meas:ocbw ................................................................................ 98
meas:ocbw:bw? ........................................................................ 99
meas:ocbw:chpw? .................................................................... 99
meas:acpr
Set
Query
Description
Turns the ACPR function on or off, or queries its status.
Syntax
meas:acpr {on|off }

Power Measurement Commands

97
GSP-730 User Manual
Query Syntax
meas:acpr?
Parameter/ Return parameter
on
ACPR mode = on
off
ACPR mode = off
Example
meas:acpr on
Turns the ACPR function on.
meas:acpr:lower?
Set
Query
Description
Returns the lower ACPR measurement result for the selected channel offset (offset 1 or 2).
Query syntax
meas:acpr:lower? {1|2}
Parameter 1
Channel offset 1
2
Channel offset 2
Return parameter
<NR2>
Returns the ACPR measurement result.
Example
meas:acpr:lower? 1
>6.0
meas:acpr:upper?
Set
Query
Description
Returns the upper ACPR measurement result for the selected channel offset (offset 1 or 2).
Query syntax
meas:acpr:upper? {1|2}
Parameter 1
Channel offset 1
2
Channel offset 2
Return parameter
<NR2>
Returns the ACPR measurement result.
Example
meas:acpr:upper? 1
>-11.8
meas:ocbw
Set
Query
Description
Turns the OCBW function on or off, or queries its status.
98
Loading...