This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Declaration of Conformity ....................... 111
INDEX .......................................................................... 112
2
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the instrument or to
other properties.
DANGER High Voltage
Attention Refer to the Manual
Earth (ground) Terminal
Frame or Chassis Terminal
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to insure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
3
GSP-730 User Manual
General
Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Ensure signals to the RF input do not exceed
+30dBm/±25V DC.
Do not block the cooling fan opening.
Do not disassemble the instrument unless you
are qualified.
(Measurement categories) EN 61010-1:2010 specifies the
measurement categories and their requirements as follows. The
instrument falls under category II.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range: 100V~240V
Frequency: 50/60Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
Safety Guidelines
4
SAFETY INSTRUCTIONS
Cleaning
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Temperature: 5°C to 45°C
Humidity: 90% @ 45°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees
and their requirements as follows. The instrument falls under
degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
5
GSP-730 User Manual
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the instrument in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
This chapter provides a brief overview of the
GSP-730, the package contents, instructions for
first time use and an introduction to the front
panel, rear panel and GUI.
7
GSP-730 User Manual
Performance
150kHz~3GHz bandwidth
100kHz resolution
Features
Autoset with automatic floor level and span.
Marker table function
Limit line testing
Split window display
ACPR measurement
OCBW measurement
Automatic resolution bandwidth mode.
Interface
480640 color LCD display
On-screen menu icons
VGA video output
RS-232C
USB 2.0 Host port for data storage
USB 2.0 Device port for the virtual com port
communication
GSP-730 Introduction
The GSP-730 is a low-cost, basic spectrum analyzer. The GSP-730
has all the basic features of our more advanced models, but in a
smaller package, designed especially for education.
Main Features
8
GETTING STARTED
Opening the box
Contents
(single unit)
Main unit
Quick Start Guide
User Manual CD
Power cord x1 (region
dependent)
Calibration certificate
Package Contents
Check the contents before using the GSP-730.
9
GSP-730 User Manual
dB
Enter
F 1
F 2
F 3
F 4
F 5
F 6
0
BK SP
kHz /
Sec
µ
mSec
MHz /
GHz /
Sec
1
4
7
2
5
8
3
6
9
Frequency
Span
Amplitude
Autoset
Marker
Peak
Limit Line
Meas
Search
BW
Trace
Display
Memory
Preset
Hardcopy
System
Hardcopy
Setup
GSP-730
150 kHz 3 GHz
Spectrum Analyzer
RF INPUT 50
+30dBm MAX.
MAX.DC ±25V
Ω
LCD Display Function
Keys
Menu keysHardcopy key
Arrow keys
RF Input
terminal
Keypad and unit
keys
USB A portPower
button
Scroll wheel
LCD display
640 X 480 color LCD display. The display shows
the soft keys for the current function, frequency,
amplitude and marker information.
Function keys
F 1
~
F6
The F1 to F6 function keys directly
correspond to the soft keys on the
right-hand side of display.
Frequency
Frequency
Sets the center frequency, start
frequency, stop frequency and
step frequency.
Span
Span
Sets the span, with options for full
span, zero span and last span.
Appearance
GSP-730 Front Panel
10
GETTING STARTED
Amplitude
Amplitude
Sets the amplitude reference level,
scale and amplitude units.
Autoset
Autoset
Automatically searches the peak
signal with maximum amplitude
and displays it with appropriate
horizontal and vertical scales.
Marker
Marker
The Marker key is used to
configure the markers, trace
markers as well as other related
functionality.
Peak Search
Peak
Search
Finds each maximum and
minimum peak. Used with the
Marker functions.
Meas
Meas
Configures ACPR and OCBW
measurements.
Limit Line
Limit Line
Sets and tests Pass/Fail limit lines.
BW
BW
Sets the resolution bandwidth.
Trace
Trace
Sets traces and trace related
functions.
Display
Display
The Display key configures the
split-screen windowing mode and
the basic display properties.
Memory
Memory
The memory key is used to save or
recall setup, trace and limit line
data.
11
GSP-730 User Manual
Preset
Preset
The Preset key will restore the
spectrum analyzer to the factory
settings.
Hardcopy
Hardcopy
The hardcopy key is a quick save
key that will save a screen-shot of
the display.
Hardcopy setup
Hardcopy
Setup
Configures the hardcopy options.
System
System
The System key is used to
configure the RS232 interface,
language, update the firmware as
well as other system options.
Power key
Turns the instrument on/off.
Scroll wheel
Edit values, select listed items.
Arrow keys
Increment/decrement values (in
steps), select listed items.
12
GETTING STARTED
RF input terminal
RF INPUT 50
+30dBm MAX.
MAX.DC ±25V
Ω
RF input port. Accepts RF inputs.
Maximum input: +30dBm
(+20dBm measurable)
Input impedance: 50Ω
Maximum DC voltage: ±25V
N-type: female
Numeric keypad
0
1
4
7
2
5
8
3
6
9
MHz /
mSec
dB
Enter
GHz /
Sec
kHz /
µSec
BK SP
The numeric
keypad is used to
enter vales and
parameters. It is
often used in
conjunction with
the arrow keys and
scroll wheel.
USB A, Micro SD
USB A port for saving/recalling
settings/files to external memory.
Only supports FAT/FAT32
formatting.
13
Rear Panel
Fan
RS232 port
VGA port
Power socket
USB B port
Security slot
RS232
RS232 9 pin DSUB port.
USB B
USB B Device port. USB 1.1/2.0
VGA
VGA video-out port. Supports
SVGA (480X640)
Fan
Power Socket
Power Socket:
100~240V, 50/60Hz.
15W Max
Security Slot
Kensington-type security slot.
GSP-730 User Manual
14
Display
Reference
level
Function
menu
Marker
information
Entry / Message areaTrace icons
Traces and
waveforms
Soft menu
keys
Frequency/
Bandwidth
settings
Reference level
Displays the reference level. For details, see page
34.
Marker
information
Displays marker information. For details see page
38.
Function menu
Displays the current function menu.
Soft menu keys
The Soft menu keys are associated with the F1 to
F6 function keys to the right of the display.
Trace Icons
Displays the color of each active trace and the trace
mode of each active trace. See page 62 for more
information about traces.
Entry/Message
area
This area is used to show system messages, errors
and input values/parameters.
GETTING STARTED
15
GSP-730 User Manual
Frequency/
Bandwidth
settings
Displays the Start, Center and Stop frequencies,
RBW, Span and Sweep settings.
Trace and
waveforms
Main display showing the traces (page 62), limit
lines (57) and marker positions (38).
16
GETTING STARTED
Description
The GSP-730 has two adjustable tabs at the
front that can be used to position the
instrument into two preset orientations.
Leaning Position
Set the tabs down to
have the GSP-730
leaning backward.
Upright Position
Set the tabs flat to
have the GSP-730 in
an upright position.
First Time Use Instructions
Use the procedures below when first using the GSP-730 to tilt the
stand, power up the instrument, update the firmware and restore
the unit back to the default settings. Lastly, the Conventions
sections will introduce you to the basic operating conventions used
throughout the user manual.
Tilting the Stand
17
GSP-730 User Manual
Power Up
1. Insert the AC power cord
into the power socket.
2. Press the power button to
turn the GSP-730 on.
3. The GSP-730 will begin to boot up in under a
few seconds.
Note
If system fails to start, please see your local GW
Instek distributor.
Power Down
1. Press the power button to
power down.
Power UP and Down
18
GETTING STARTED
Description
The GSP-730 allows the software to be updated
by end-users. Before using the GSP-730, please
check the GW Instek website or ask your local
distributor for the latest software.
The update file, MAIN1.BIN, must be placed in
the root directory of a USB flash drive.
System version
Before updating the software, please check the
software version.
1. Press
System
>Information[F4].
2. The system version is displayed on the
SW Ver[F4] icon.
Software version
Update software
1. Place the update file, MAIN1.BIN, into the root
directory of a USB flash drive.
2. Insert the USB flash drive into the USB port on
the front panel.
3. Press
System
>Update From USB Flash[F5].
4. Press Update Now[F3] to execute the update
process.
When the message “Programmed Successful” is
displayed, the software has successfully completed
the update procedure.
Software Update
19
GSP-730 User Manual
5. Reboot the system when the update procedure
has finished by cycling the power button.
6. Check the software version again to confirm
the update procedure.
Note
The upgrade process may take a few minutes.
Description
If the type B USB port on the rear panel is to be
used for remote control, then the USB driver
must be installed. The USB driver is located in
the CD that accompanied this manual.
Driver installation
1. Ensure the GSP-730 is turned on.
2. Connect the USB cable from the PC
to the rear panel USB B port.
3. Windows will automatically detect the GSP-
730 as a new device.
4. Follow the instructions to locate the GSP-730
driver on the accompanying CD and install the
driver.
5. To see if the driver has been successfully
installed, you can check to see if the GSP-730 is
recognized by the Windows Device Manager
when the GSP-730 in connected to the PC.
In Windows XP go to: Start>Control
Panel>Device Manager.
USB Driver Installation
20
GETTING STARTED
The GSP-730 should be shown under the
Ports (COM & LPT) node:
Note
If the USB driver installation fails, you can try to
manually install the driver.
You can tell that the driver hasn’t been installed if
you see the AT91USBSerial icon in the ports node.
Right clicking this icon will allow you to manually
install the GSP-730.ini driver.
Description
The factory default settings can be easily
restored using the Preset key on the front
panel. The default settings cannot be changed.
See page 106 for a list of the factory default
settings.
Steps
1.Press
Preset
.
The spectrum analyzer will load the default factory
settings.
Restoring Default Settings
21
GSP-730 User Manual
Soft Menu keys
The F1 to F6 function keys on the right side of
the display correspond directly to the softmenu keys on their left.
Soft-menu keys
F1 ~ F6 function keys
dB
Enter
F 1
F 2
F 3
F 4
F 5
F 6
0
BK SP
kHz /
Sec
µ
mSec
MHz /
GHz /
Sec
1
4
7
2
5
8
3
6
9
Frequency
Span
Amplitude
Autoset
Marker
Peak
Limit Line
Meas
Search
BW
Trace
Display
Memory
Preset
Hardcopy
System
Hardcopy
Setup
GSP-730
150 kHz 3 GHz
Spectrum Analyzer
RF INPUT 50
+20dBm MAX.
MAX.DC ±25V
W
Input Parameter
Values
Selecting this type of menu key will allow you
to enter a new value with the numeric keypad
or increment/decrement the value using the
scroll wheel or number pad. See the parameter
input description below for more details.
Conventions
The following conventions are used throughout the user manual.
Read the conventions below for a basic grasp of how to operate the
GSP-730 menu system and front panel keys.
22
GETTING STARTED
Toggle State
Pressing this menu key will toggle the state.
Notice that any soft-menu key that can be
toggled will have the active parameter
underlined.
Toggle State &
Input Parameter
Pressing this menu key will allow you to toggle
the state of the function between on and off.
When in the on-state, the parameter value can
be manually edited. Use the numeric keypad to
enter the new value or use the scroll wheel to
increment/decrement the current value. Again,
the setting that is underlined is the active
setting.
Sub Menu
Pressing the More menu key will enter a
submenu.
Sub Menu to
select parameter
Pressing this type of menu key will enter a
submenu to select a parameter.
23
GSP-730 User Manual
Return to the Start
of a Menu Tree.
When you have navigated down a menu tree
and you wish to return to the start of the menu
tree, simply press the same Menu again.
For example if you pressed:
Trace
> More[F6] > Min Hold[F2],
simply press
Trace
again to return to the start
of the Trace menu.
Active Function
Pressing this type of menu key will activate
that function. The menu key will be
highlighted to show it is the active function.
Parameter input
Numerical keypad
Scroll wheel
Directional arrow keys
Backspace, Enter keys
and unit keys
0
1
4
7
2
5
8
3
6
9
MHz /
mSec
dB
Enter
GHz /
Sec
kHz /
µSec
BK SP
Parameter values can be entered using the
numeric keypad, the scroll wheel and the
arrow keys.
24
GETTING STARTED
Using the numeric
keypad
When prompted to enter a parameter, use the
number keys (0~9), the decimal key (.) and the
minus key (-) to enter a value. After a value has
been entered, the unit keys can be used to
select the units.
The value of the parameter is shown at the
bottom of the screen as it is edited.
Edited parameter
Ref:-30dBm
Back Space
Use the backspace key to delete the last
character or number entered.
Using the scroll
wheel
Use the scroll wheel to alter the current value.
Clockwise increases the value, anti-clockwise
decreases the value. The scroll wheel is usually
used for values that highly variable, such as the
center frequency settings.
Directional arrows
Use the directional arrows to select discrete
parameters or to alter values by a coarser
resolution than the scroll wheel. Left/down
decreases the value, right/up increases the
value. The directional arrows are usually used
for values that are of a discrete nature, such as
selecting a memory location.
25
GSP-730 User Manual
Frequency Settings ............................................................ 28
Center Frequency ................................................................................ 28
Start and Stop Frequency ................................................................... 29
Center Frequency Step ....................................................................... 30
The start and stop frequency can change when the
span settings are used.
The stop frequency must be set higher than the
start frequency (for spans 0).
Start and Stop Frequency
29
GSP-730 User Manual
Description
The Step function sets the step size of the center
frequency when using the arrow keys.
When the arrow keys are used to alter the
center frequency, each press will move the
center frequency by the step size specified by
the Step function.
By default, the center frequency step size is
equal to 10% of the span.
Operation
1. Press
Frequency
> Step[F4] and set the center
frequency step size.
Range:
1Hz~3GHz
Display
Step size
Center Frequency Step
30
BASIC OPERATION
Description
The Span function will set the frequency range
of the sweep. The sweep will be centered
around the center frequency.
Setting the span will alter the start and stop
frequencies.
Operation
1.Press
Span
> Span[F1] and enter the span
frequency range and unit.
Range:
Default Span:
0kHz~3GHz
3GHz
Display
Span
Set Span
Span Settings
Span
31
Full Span
Description
The Full Span function will set the span to the
full frequency range.
This function will set the start and stop
frequencies to 0Hz and 3GHz respectively.
Operation
1.Press
Span
> Full Span[F2].
Description
The Zero Span function will set the frequency
range of the sweep to 0Hz and fixes the start
and stop frequencies to the center frequency.
The Zero Span function measures the time
domain characteristics of the input signal at the
center frequency. The horizontal axis is
displayed in the time domain.
Operation
1.Press
Span
> Zero Span[F3].
The span changes accordingly.
Zero Span
GSP-730 User Manual
32
Display
Time domain
0Hz Span
Example: Amplitude modulation
Note
The measurement functions such as ACPR and
OCBW are not available with the zero span
setting:
Last Span
Description
The last span function returns the spectrum
analyzer to the previous span settings.
Operation
1.Press
Span
> Last Span[F4].
BASIC OPERATION
33
GSP-730 User Manual
Description
The reference level defines the absolute level of
the amplitude on the top graticule in voltage or
power.
Operation
1. Press
Amplitude
> Ref. Level[F1] and enter the
reference level amplitude.
Range:
Resolution:
-20dBm ~ 20dBm
10dBm
Display
Ref Level reading
Amplitude Settings
The vertical display scale is defined by the reference level
amplitude, attenuation, scale and external gain/loss.
Reference Level
34
BASIC OPERATION
Description
The amplitude units can be set from dBm,
dBmV or dBuV.
1. Press
Amplitude
> Units …[F3] to change the
amplitude units.
Units:
dBm, dBmV, dBuV
Description
Sets the logarithmic units for the vertical
divisions.
Operation
1. Press
Amplitude
> Scale[F2] repeatedly to select the
vertical division units.
Scale Range:
10, 5, 2, 1
Display
Scale
Amplitude Units
Scale/Div
35
Operation
1.Press
Autoset
> Autoset[F1].
Autoset Range
Amplitude:
Span:
Over the full amplitude range.
Over the full span range.
Example:
Before Autoset, default state
After Autoset
GSP-730 User Manual
Autoset
The Autoset function searches the peak signals and picks the signal
peak with the maximum amplitude, and then shows it in the
display.
Using Autoset
36
BASIC OPERATION
Note
RBW setting is reset to Auto when the Autoset
function is used.
Description
You can set the amplitude floor so that the
signals lower than the setting will be ignored
by the Autoset search.
Operation
1.Press
Autoset
> Amp.Floor[F2] and switch the
range from Auto to Man.
1. Enter the amplitude limit with the number pad
and Enter key.
Range:
−50 to +20dBm
Note
See page 35 for setting the amplitude units.
Description
You can change the frequency span limit in the
display to get a better view of the Autoset
result. By default, the frequency span after
Autoset is set at 3MHz.
Operation
1.Press
Autoset
> Span[F3] and switch the range
from Auto to Man.
2. Enter the span frequency for the Autoset
search.
Manual Range:
full amplitude range.
Limiting the Autoset Vertical Search Range
Limiting the Autoset Horizontal Search Range
37
GSP-730 User Manual
Marker
A Marker shows the frequency and amplitude of a waveform point.
The GSP-730 can activate up to 5 markers or marker pairs
simultaneously.
The marker table and peak table functions help editing and
viewing multiple markers in a single display.
The delta marker function allows you to see the frequency and
amplitude differences between reference markers.
The GSP-730 can automatically move a marker to various locations
including the peak signal, center frequency, and start/stop
frequency. Other marker operations regarding signal peaks are
available in the Peak Search function.
Activating a Marker → from page 39
Activate a Delta Marker→ from page 40
Move Marker Manually→ from page 41
Move Marker to Preset Locations→ from page 42
Move Marker to Trace→ from page 43
Turn All Markers On or Off→ from page 44
Show Markers in Table→ from page 44
Peak Search→ from page 46
Move Marker to Peak→ from page 46
Move Marker and Peak to Center→ from page 46
Search for Peaks→ from page 47
Peak Table→ from page 48
38
BASIC OPERATION
Operation
1.Press
Marker
> Marker[F1] and select a marker
number.
Marker:
1~5
2. Press [F2] to turn the selected marker on.
3. Press Mode[F3] and set the marker mode to
Normal.
4. The display will show the marker on the trace
(centered by default) with the marker
measurement at the top of the display.
Maker No., Frequency, Amplitude
Marker
Activating a Marker
There are two basic marker types, normal markers and delta
markers. Normal markers are used to measure the frequency/time
or amplitude of a point on the trace. Delta markers are used to
measure the difference between a reference point and a selected
point on the trace.
Activate a Normal Marker
39
GSP-730 User Manual
Description
Delta markers are marker pairs that measure
the difference in frequency and amplitude
between a reference marker and a delta
marker.
When delta markers are activated, the
reference and delta marker appear at the
position of the selected marker, or in the center
of the display if the selected marker has not yet
be activated.
The marker measurement is located at the top
of the display, under the “normal marker”
measurement.
Delta Markers
Ref:
Reference marker, designated as 1.
Delta:
Delta marker, designated as 11.
Operation
1.Press
Marker
> Marker[F1] and select a marker
number.
2. Press [F2] to turn the selected marker on.
3. Press Mode[F3] to set the mode to Delta to
activate the delta marker.
Activate a Delta Marker
40
BASIC OPERATION
Delta maker No., Frequency, Amplitude
Delta
Marker
Operation
1.Press
Marker
> Marker[F1] and select a marker
number.
2. Use the left/right arrow keys to
move the marker one screen
division at a time or the use the
scroll wheel to move the marker in
fine increments (one pixel at a
time).
3. Alternatively, the numeric keypad
can be used to directly enter the
frequency of the marker position.
789
654
123
/0
Move Marker Manually
41
GSP-730 User Manual
Preset conditions
The currently selected marker (normal marker
or delta marker) can be moved to a number of
preset positions:
Center:
Peak
Start:
Stop:
Step:
Ref. Level:
Move to center frequency.
Move to the highest peak.
Move to start frequency.
Move to stop frequency.
Move to step frequency.
Move to reference level
amplitude.
Note
When a marker is moved to a preset position the
span and other settings may be automatically
changed.
Move marker to
peak:
Press
Marker
> To Peak[F4].
Move marker to
center:
Press
Marker
> To Center[F5].
Move marker to
other positions
Press
Marker
> More[F6]>Marker to ...[F4] and
select one of the preset positions:
Marker to Start[F2]
Marker to Stop[F3]
Marker to Step[F4]
Marker to Ref. Level[F5]
Move Marker to Preset Locations
42
BASIC OPERATION
Description
The Marker Trace function moves the selected
marker to the currently active trace.
Operation
1.Press
Marker
> Marker[F1] and select a marker
number.
2. Press More[F6]>Marker Trace and select a trace
to assign the selected marker to. If Auto is
selected, the selected marker is automatically
assigned a trace.
Marker Trace:
Auto, A, B, C
3. In the example below, marker 1 is set to trace
B.
Marker 1, Trace B
Move Marker to Trace
43
GSP-730 User Manual
Description
All markers that have been activated, both
normal and delta markers, can be turned off at
the same time with the All Mrk Off function.
Operation
Press
Marker
> Marker[F1] >More[F6]>All Mrk
Off[F3] and turn all the markers off.
Description
The GSP-730 has a Marker Table function to
show all the active markers and measurements
at once.
Operation
1.Press
Marker
> Marker[F1] >More[F6]>Marker
Table[F2] and turn the marker table on.
2. The display will split into two screens. The
bottom half will show the Marker Table with
the marker No. (normal, reference or delta),
frequency and the amplitude of the marker.
Marker Table
Turn All Markers On or Off
Show Markers in Table
44
BASIC OPERATION
Description
While the Marker Table function is the active
function, the position of each marker and delta
marker can be edited within the marker table.
1. Use the arrow keys to move the cursor to the
frequency column of the desired marker.
Cursor
2. Enter the new position of the marker using the
keypad and units keys.
Edit Markers in Marker Table
45
GSP-730 User Manual
Description
Move the active marker to the highest peak.
The highest peak can be either found once or
continuously.
Operation
1.Press
Marker
> Marker[F1] and select a marker
number.
2. Press
Peak
Search
>Peak Search[F1]. The marker will
move to the highest signal peak.
3. To continually search for the peak each sweep,
press,
Peak
Search
>More[F6]>Peak Track[F1] and set
Peak Track to ON.
Description
The Peak to Center function moves the marker
to the highest signal peak and moves the center
frequency to that peak.
Operation
1.Press
Marker
> Select Marker[F1] and select a
marker number.
2. Press
Peak
Search
>Peak to Center[F5].
Note
The span will not be changed.
Peak Search
The Peak Search key is used to find trace peaks. The currently
active marker is used in conjunction with the peak functions to
mark the peaks that are found. Peaks can be sorted by frequency or
amplitude in the peak table.
Move Marker to Peak
Move Marker and Peak to Center
46
Search for Peaks
Description
The
Peak
Search
key can be used to search for a
number of different peaks.
Peak Search
Next Peak:
Searches for next highest
peak visible on the display.
Next Peak Right:
Searches for the next peak to
the right of the marker.
Next Peak Left:
Searches for the next peak to
the left of the marker.
Min Search:
Searches for the lowest peak.
Operation
1.Press
Marker
> Select Marker[F1] and select a
marker number.
2. Press
Peak
Search
and select the type of peak you
wish to find.
Example:
Next Peak
BASIC OPERATION
47
GSP-730 User Manual
Example:
Next Peak Right
Example:
Next Peak Left
Description
The Peak Table function will display up to 5
peaks. The amplitude and frequency for each
peak is listed.
Operation
1.Press
Peak
Search
>More [F6]>Peak Table[F1] and turn
the peak table on.
Peak Table
48
BASIC OPERATION
2. Press Peak Sort[F2] and set the sorting type:
Freq:
Amp:
Sort by frequency in
ascending order.
Sort by amplitude in
ascending order.
Display
The bottom-half of the screen shows the peak
table with the peak marker no., frequency and
amplitude.
Peak Table
Peak markers
49
GSP-730 User Manual
Description
Channel analysis measurement includes ACPR
(adjacent channel power) and OCBW
(occupied bandwidth) measurements.
Parameters
Channel bandwidth
The frequency
bandwidth the target
channel occupies.
Range:
Between 0Hz~3GHz
(0Hz excepted)
Channel Space
The frequency distance
between each main
channel.
Range:
Between 0Hz~3GHz
Adjacent channel
bandwidth 1 & 2
The frequency
bandwidth the adjacent
channels occupy.
Range:
Between 0Hz~3GHz
(0Hz excepted)
Measurement
This section describes how to use the automatic measurement
modes. The GSP-730 includes the following measurements:
ACPR → from page 51.
OCBW → from page 54.
Channel Analysis Overview
50
Adjacent channel offset
1 ~ 2
The frequency distance
between the adjacent
channels and main
channel.
Range: 1
Between 0Hz~3GHz
(0Hz excepted)
OCBW%
The ratio of occupied
bandwidth to the
amount of power
consumed.
Range: 0% to 100%,
0.1% resolution.
ACPR
Description
Adjacent channel power refers to the amount
of power leaked to the adjacent channel from
the main channel. This measurement is a ratio
of the main channel power to power in the
adjacent channel.
Example
Main
CHBW
ADJ
CH1
ADJ
CH2
ADJ
CH1
ADJ
CH2
Offset 1
Offset 2
Channel spacing
To next main
channel
Operation:
Setting up the
main channel
1.Press
Meas
> ACPR[F2] and turn ACPR on.
Any other measurement mode will automatically be
BASIC OPERATION
51
GSP-730 User Manual
disabled.
2. The display splits into two screens. The top
screen shows the sweep waveform. The bottom
screen shows the ACPR settings and
measurement results in real time.
Turn ACPR off to return back to the normal mode.
Channel setup
Measurements
3. Press Channel Setup…[F1] and set the
following:
Main CH BW[F1]
Set the bandwidth of the
main channel.
Main CH Space[F2]
Specify the channel
spacing.
Note
The main channel bandwidth and space
settings are shown in the setup area at the
bottom of the screen, not on the soft-key icon.
Channel BW
Channel space
52
BASIC OPERATION
Operation:
Setting up the
adjacent
channel(s)
1. Press ADJCH Setup…[F3] to setup the adjacent
channels:
Adj CH BW 1[F1]
Sets the bandwidth of
the 1st adjacent channel.
Adj CH Offs 1[F2]
Sets the channel offset
of the 1st adjacent
channel.
Adj CH BW 2[F3]
Sets the bandwidth of
the 2nd adjacent channel.
Adj CH Offs 2[F4]
Sets the channel offset
of the 2nd adjacent
channel.
Note
The adjacent channel bandwidth and space
settings are shown in the setup area at the
bottom of the screen, not on the soft-key icons.
Adj CH1 settings
Adj CH2 settings
Move Channels
Up/Down
1.Press the
Meas
again or press Return[F6]
repeatedly to return to the start of the Measure
menu tree.
2. Press CH Up[F5] to go to the next main
channel.
3. Press CH Down[F6] to go to the previous main
channel.
Note
The channel space (Main CH Space) setting
determines where the next main channel is
located.
53
OCBW
Description
Occupied bandwidth measurements are used
to measure the power of the occupied channel
as a percentage to the power of the channel.
Example
CH BW
OCBW
Operation:
Setting up the
main channel
1.Press
Meas
> OCBW %[F3] and turn OCBW
on.
Any other measurement mode will automatically be
disabled.
2. The display splits into two screens. The top
shows the channel bandwidth. The bottom
screen shows the OCBW measurement results
in real time.
Turn OCBW off to return back to the normal mode.
GSP-730 User Manual
54
BASIC OPERATION
Channel power and OCBW
power results
CH BW
3. Press Channel Setup…[F1] and set the
following:
Main CH BW[F1]
Set the bandwidth of the
main channel.
Main CH Space[F2]
Specify the channel
spacing.
Note
The main channel bandwidth and space
settings are shown in the setup area at the
bottom of the screen, not on the soft-key icon.
Channel BW
Channel space
55
GSP-730 User Manual
Move Channels
Up/Down
1.Press
Meas
again or press Return[F6]
repeatedly to return to the start of the Measure
menu tree.
2. Press CH Up[F5] to go to the next main
channel.
3. Press CH Down[F6] to go to the previous main
channel.
Note
The channel space (Main CH Space) setting
determines where the next main channel is
located.
56
BASIC OPERATION
Operation
1. Press
Limit Line
to enter the Limit Line menu.
2. Press H Limit[F1] or L Limit[F2] and turn the
high or low limit on or off.
H Limit/L Limit:
On, Off
The H Limit is shown as a blue line.
The L Limit is shown as a pink line.
High limit line
Low Line
Limit Line Testing
The Limit Line function is used to set the upper or lower amplitude
limits over the entire frequency range. The limit lines can be used to
detect whether the input signal is above, below or within the limit
lines.
The limit lines can be manually edited using 10 frequency points
from the start to the stop frequencies.
To save and recall limit lines, please see page 73.
Activate a Limit Line
57
GSP-730 User Manual
Description
Create a limit manually, point by point. Ten
manually selected frequency points can be
used to create the upper or lower limit line.
Operation
1. Press
Limit Line
> Limit[F1] and select the limit line
you wish to edit.
Limit:
High, Low
2. Press Edit Table[F2], and turn the edit table on.
The GSP-730 is split into two screens. The top
screen shows the trace and the selected limit
line (high or low) and the bottom screen shows
the limit line table.
Spectrum display
Limit Line Table
All 10 points will be displayed in a limit line
table at the bottom of the display. By default,
each point is set to 0dBm.
3. Use the arrow keys to move the cursor to the
Creating a Limit (Point by Point)
58
BASIC OPERATION
frequency column of the desired point.
Cursor
4. Enter the new frequency and amplitude of the
point using the keypad and the unit keys.
5. Repeat steps 3-5 for the remaining points (A
maximum of ten points).
6. To delete the selected point, press Delete [F3].
7. To delete all the points, press Delete All… [F4].
The points will revert to their default frequency and
amplitude values.
8. To delete a point from the editing table, press
Delete.
The whole points, including the frequency value and
the amplitude are removed. This command reduces
the number of points used in the limit line.
9. Press Undelete[F5] the restore the last point that
was deleted.
Description
The Pass/Fail testing uses the limit lines as
boundaries. When the input signal escapes the
boundary of the limit lines, then the test is
judged as a FAIL, if the signal stays within the
boundary, the test is judged as a PASS.
Pass/Fail Testing
59
GSP-730 User Manual
Pass:
Upper limit
Lower limit
Fail:
Upper limit
Lower limit
Note
Before pass/fail testing can begin, limit lines
for the upper and/or lower limits must first be
saved and activated. See the page 57.
Operation
1. Press
Limit Line
>Pass/Fail[F4] to turn the testing
on or off.
2. The test result is updated in real-time at the
bottom of the display.
Pass:
Fail:
Note
At least one limit line (high or low) must be
turned on to enable testing.
If only high limit line is on, each trace point has to
be lower than the high limit line to get a PASS
judgment, otherwise get the judgment will be
FAIL.
Conversely, if only low limit line is on then each
trace point has to be higher than the low limit line
to get a PASS, otherwise get the test will be judged
a FAIL.
60
BASIC OPERATION
Description
The RBW (Resolution Bandwidth) defines the
width of the IF (intermediate frequency) filter
that is used to separate signal peaks from one
another. The narrower the RBW, the greater
the capability to separate signals at close
frequencies. But it also makes the sweep time
longer under specific frequency spans (the
display is updated less frequently).
Operation
1. Press
BW
> RBW[F1] and set the RBW to
Auto or Man.
2. Set the resolution bandwidth and unit for Man
mode.
Mode:
Frequency Range:
Auto, Man
1MHz, 300kHz, 100kHz
Note
The manual RBW is only available when the
span≥10MHz. If the span is greater than 10MHz
then the RBW is automatically set to Auto.
Bandwidth
BW key sets the resolution bandwidth (RBW). The resolution
bandwidth and the sweep time are related. Please take into account
how the sweep time is effected by the resolution bandwidth.
Resolution Bandwidth Setting (RBW)
61
GSP-730 User Manual
Description
Each trace (A, B, C) is represented by a
different color. Trace A is green, trace B is
orange and trace C is yellow. When activated,
an icon for each trace color and function is
shown at the bottom of the display. When a
trace is selected, parameters can be set/edited
from the Trace menu.
Display Icons
Trace A, B, C
Trace Type & Icon
The type of trace used determines how the
trace data is stored or manipulated before
being displayed. The analyzer updates each
trace according to the type of trace used.
Clear &
Write
The GSP-730 continuously
updates the display with each
sweep. This is the default trace
type.
Trace
The GSP-730 is able to set the parameters of up to 3 different traces
on the display at once. Each trace is represented by a different color
and is updated with each sweep.
To save or recall traces to/from memory, see page 71.
Selecting a Trace
62
BASIC OPERATION
Peak Hold
Min Hold
The maximum or minimum
points are maintained for the
selected trace. The trace points
are updated each sweep if new
maximum or minimum points
are found.
View
View will hold the selected
trace and stop updating the
trace data for the selected
trace. Pressing View[F4] will
display the trace data that was
cleared using the Blank[F5]
key.
Blank
Clears the selected trace from
the display and stores trace
data. The trace data can be
restored by pressing View[F4].
Performs trace math from two traces (A, B) and
stores the results in trace A or swaps the data
from trace A to trace B.
Math functions
A <--> B
Swaps the data from trace A to
B and vice versa.
A + B -> A
Adds trace A and B and stores
the result in trace A.
A – B -> A
Subtracts trace B from trace
and stores the result in trace A.
A + const ->A
Adds an offset to trace A.
A – const ->A
Subtracts an offset from trace
A.
Operation
1. Press
Trace
>Trace Math..[F3] and select a
trace math function.
A <--> B
[F1]
A + B -> A
[F2]
A – B -> A
[F3]
A + const ->A
[F4]
A – const ->A
[F5]
2. If A + const ->A or A + const ->A was selected,
set the constant (offset value).
Constant:
-40dBm ~ 40dBm
GSP-730 User Manual
64
Average Trace
Description
The Average function averages the currently
selected trace for a user-defined number of
times before it is displayed. This feature
smoothes the noise level, but has the drawback
of slowing down the display update rate.
Operation
1. Press
Trace
> More[F6] and toggle Avg on.
2. Set the number of averages.
Range:
Default:
4 ~ 100
4
BASIC OPERATION
65
Example:
Average:Off
Average: On (8)
GSP-730 User Manual
66
BASIC OPERATION
Description
The LCD brightness levels can be adjusted to
five pre-set levels.
Operation
1. Press
Display
> LCD Dimmer[F1] and use either
the number pad, the scroll wheel or arrow keys
to set the brightness.
Description
The Display Line function is used to superimpose a reference level line over the traces.
Operation
1. Press
Display
> Display Line[F3] to turn the
display line on.
2. Set the display line level and press Enter.
Example:
Display line
Display line set at -50dBm
Display
The Display key configures the basic display settings as well as the
split screen modes.
Adjusting the LCD Brightness
Setting a Display Line (Reference Level Line)
67
GSP-730 User Manual
Description
The GSP-730 has a dedicated VGA terminal to
output the display to an external monitor. The
video output is always on.
Output resolution
480 x 640 (fixed)
Operation
1. Connect an external monitor to
the rear panel VGA terminal.
Description
The split spectrum view is able to view two
different sweep ranges on the display at the
same time using a split screen view. The top
and bottom view can have independent sweep
ranges, amplitudes, spans and other settings.
However only one split screen (top or bottom)
can be swept each time.
Example
Lower
spectrum
Upper
spectrum
Using the Video Out Port
Split Spectrum View
68
BASIC OPERATION
Split spectrum
functions
Half-Upper
Half-Upper will put the
spectrum analyzer into split
screen mode. It will make the
top sweep the active sweep
and pause the bottom sweep.
When Half-Upper is on, only
the upper sweep parameters
can be edited.
Half-Lower
Half-Lower will put the
spectrum analyzer into split
screen mode. It will make the
bottom sweep the active
sweep and pause the top
sweep. When Half-Lower is
on, only the lower sweep
settings can be edited.
Alternate
Sweep
This setting will alternate the
sweep between the bottom
and top spectrums. If alternate
sweep is turned on, only the
upper sweep parameters can
be edited.
69
GSP-730 User Manual
Operation
1. Press
Display
>Half-Upper[F4] or Half-Lower[F5]
or Alternate Sweep[F6] to enable the split
spectrum view.
Turning Half-Upper on will automatically turn
Half-Lower off.
Turning Half –Lower on will automatically turn
Half-Upper off .
If Alternate Sweep is turned on, each sweep will
alternate, but only the upper sweep parameters can
be edited.
2. To return to a full-screen, single spectrum
display, press Full Display[F2].
Note
After exiting the split spectrum view, the analyzer
will use the settings from the active window. The
settings for the inactive screen will be retained for
the next time that split spectrum view is used.
If the spectrum analyzer was in the Alternate
mode, then the upper sweep settings will be
returned.
70
BASIC OPERATION
Description
Setup data contains all the data necessary to
recall the state of the GSP-730 to known state.
>
Setup To[F1] and choose a memory location to
save to with the arrow keys.
Setup To:
1~5
2. Press
Enter
kHz /
µSec
to execute the save.
Recall
1. To recall a setup, press
Memory
> Setup From[F2]
and choose a memory location to recall from
with the arrow keys.
Setup From:
1~5
2. Press
Enter
kHz /
µSec
to execute the recall.
Save/Recall Files
The GSP-730 can save and recall setup data, trace data and limit
line data to and from internal memory. There are five memory
locations for each save file type. These files cannot be saved to USB.
The Hardcopy key can be used to save image files to a USB flash
drive.
Save/Recall Setup
71
GSP-730 User Manual
Description
The trace data can be saved/recalled for any of
the A, B or C traces to/from one of 5 pre-set
internal memory locations. The trace data
cannot be recalled or saved to USB.
When saving or recalling trace data from a
split spectrum, only the active spectrum is
saved/recalled.
Save
1. To save the current trace data, press
Memory
>
Save Trace Data..[F3]
2. Press Source Trace[F1] and select the source:
Source:
A, B, C
3. Press Destination[F2] and select the memory
location to save to:
Destination:
1~5
4. Press Start[F5] to save the selected trace data.
Recall
5. To recall trace data, press
Memory
> Recall Trace
Data..[F3]
6. Press Source Trace[F1] and select the memory
location to recall from:
Source:
1~5
7. Press Destination[F2] and select the destination
trace
Destination:
A, B, C
Save/Recall Trace Data
72
BASIC OPERATION
8. Press Start[F5] to recall the selected trace data.
Description
Upper and lower limit lines can be saved to
one of 5 pre-set internal memory locations. The
limit line data cannot be saved to USB.
Save
1. To save the current upper and lower limit
lines, press
Memory
> Limitln to[F5] and choose
a memory location to save to with the arrow
keys.
Limit line:
1~5
2. Press
Enter
kHz /
µSec
to execute the save.
Recall
3. To recall pre-saved upper and lower limit
lines, press
Memory
> Limitln from[F6] and
choose a memory location to recall from with
the arrow keys.
Limit line:
1~5
4. Press
Enter
kHz /
µSec
to execute the recall.
Description
The Hardcopy key can be used to save a
screenshot of the display to a USB flash drive.
The screen shot is saved as a bitmap file.
Save/Recall Limit Lines
Saving an Image File (Hardcopy)
73
Operation
1. Insert a USB flash drive into the USB port.
2. Press
Hardcopy
and the image file will begin
saving.
Wait a few moments for the file to save. When the
file has finished saving, “Screen Saved OK” will
appear at the bottom of the display.
Note
The file name will be automatically created in the
following format:
File name: SCRXX.bmp
Where XX is a number that is incremented each
time the file is saved.
Warning
Do not remove the USB drive until the file has
completed saving.
Hardcopy Setup
Description
The Hardcopy Setup key is used set the image
file properties of the bitmap file that is created
when the Hardcopy key is pressed.
Ink Normal:
This is the normal, default
image setting.
Ink Saving:
This will invert all the
colors on the display so that
the file will conserve ink
when printed.
Operation
1. Press
Hardcopy
Setup
and choose the image type:
Ink Normal[F1]
Ink Saving[F2]
GSP-730 User Manual
74
BASIC OPERATION
Note
The next time the Hardcopy key is pressed, the
image will be saved using the settings above.
Description
The Preset key is used to load the default
settings. The default settings are listed in the
appendix on page 106.
Operation
1. Press
Preset
.
The system will load the preset settings and the
screen will update with the new settings.
Note
The default settings cannot be changed.
Load Default Settings
75
GSP-730 User Manual
Description
The System Information displays the
following:
Serial Number:
XX digit serial number
HW Version:
Hardware version
FW Version:
Firmware version
SW Version:
Software version
Language:
Shows the language
number as seen in the
System>Language
menu.
Operation
1. Press
System
>Information[F4] to display the
system information.
The system information will be displayed on the
system menu soft-keys.
Example
System
information
System Settings
System Information
76
BASIC OPERATION
Description
The language option sets the icon display
language.
Operation
1. Press
System
>Language…[F3] to bring up the
Language menu.
2. Choose a system language. The language
number is the number that will be displayed in
the system information.
This chapter describes basic configuration of
IEEE488.2 based remote control. For a command
list, refer to the programming manual,
downloadable from GW Instek website,
www.gwinstek.com
The GSP-730 can use either the type B USB port
or the RS232 on the rear panel for remote
control.
When using the USB B port, the GSP-730 uses a
USB driver to simulate an RS232 connection
with a PC via USB. It is these RS232 settings
that are configured for remote control.
Before using the USB B port for remote control,
please install the USB driver. See page 20 for
details.
Interface Configuration
Configure Remote Interface
80
REMOTE CONTROL
Panel operation
1. USB Connection:
Connect a USB cable from the PC
to the rear panel USB B port.
RS232 Connection:
Connect an RS232C cable from the
PC to the rear panel RS232 port.
2. Press
System
>Serial Port…[F1]> Serial[F1] to
enter the remote configuration.
3. Set the following RS-232 settings using the
arrow keys:
Baud Rate[F1]:
9600, 19200, 38400, 57600,
115200.
Parity[F2]:
None, Even, Odd, Space,
Mark, Multidrop.
Stop Bit[F3]:
1, 1.5, 2.
Data[F4]:
5, 6, 7, 8
Functionality
check
Invoke a terminal application such as MTTTY
(Multi-Threaded TTY).
To check the COM port No., see the Device
Manager in the PC.
For WinXP go to;
Control panel → System → Hardware tab.
Run this query command via the terminal after
the instrument has been configured for remote
control (page 80).
*idn?
Remote Control Function Check
81
GSP-730 User Manual
This should return the Manufacturer, Model
number, Serial number, and Firmware version
in the following format.
GW-INSTEK, GSP-730, XXXXXXXX, V.VV
Manufacturer: GW-INSTEK
Model number : GSP-730
Serial number : XXXXXXXXXXXX
Firmware version : V.VV
Note
For further details or if you have trouble running
this function check, please see the programming
manual, available on the GW Instek web site @
www.gwinstek.com.
Compatible
Standard
IEEE488.2
Partial compatibility
SCPI, 1999
Partial compatibility
Command
Structure
SCPI (Standard Commands for Programmable
Instruments) commands follow a tree-like
structure, organized into nodes. Each level of
the command tree is a node. Each keyword in a
SCPI command represents each node in the
command tree. Each keyword (node) of a SCPI
command is separated by a colon (:).
For example, the diagram below shows an
SCPI sub-structure and a command example.
meas
:freq
:stp?
:cen?
:st?
meas:freq:cen?
Command Syntax
82
REMOTE CONTROL
Command types
There are a number of different instrument
commands and queries. A command sends
instructions or data to the unit and a query
receives data or status information from the
unit.
Command types
Single
Command
A single command
with/without a parameter
Example
meas:freq:cen 100 MHz
Query
A query is a simple or
compound command
followed by a question mark
(?). A parameter (data) is
returned.
Example
meas:freq:cen?
Command
Format
100 k h z
12 3 4 5
Meas:f r eq:cen
1. Command header
2. Space
3. Parameter 1
4. Optional space
5. Unit or suffix.
Common
Input/Return
Parameters
Type
Description
Example
<Boolean>
Boolean logic
0, 1
<NR1>
integers
0, 1, 2, 3
<NR2>
decimal
numbers
0.1, 3.14, 8.5
<NRf>
any of NR1, 2
1, 1.5
83
GSP-730 User Manual
<freq>
<NRf> + unit
2.5 mhz
Unit = kHz, MHz, GHz.
Note: The unit can be omitted
(defaults to currently set unit).
<refl>
<NRf> + unit
-30 dBm
Unit = dBm, dBmV, dBuV
Note: The unit can be omitted
(defaults to currently set unit).
<ampl>
NR3 +unit
30.0 dBm
Note: The unit can be omitted.
(Unit defaults to current unit)
Returns the trace data for the selected trace(s). A total
of 501 trace points are returned, from the start
frequency to the stop frequency. If “all” is selected, the
trace data is returned in three lots, {trace A}{trace
B}{traceC}. The units are in decibels. If the selected
trace is not active, 0s will be returned for each trace
point.