INFICON BCG450 ATM User Manual

Operating Manual
TripleGauge™
Bayard-Alpert Pirani Capacitance Diaphragm Gauge
BCG450 BCG450-SD BCG450-SP
tina40e1-b (2011-04) 1
Product Identification
In all communications with INFICON, please specify the information on the product nameplate. For convenient reference copy that information into the space provided below.
INFICON AG, LI-9496 Balzers
Model: PN: SN: V W
Validity
Intended Use
This document applies to products with the following part numbers:
BCG450 (without display)
353-550 (vacuum connection DN 25 ISO-KF) 353-551
(vacuum connection DN 40 CF-R)
353-561 (vacuum connection DN 25 ISO-KF, with baffle)
BCG450 (with display)
353-552 (vacuum connection DN 25 ISO-KF) 353-553
(vacuum connection DN 40 CF-R)
BCG450-SD (with DeviceNet interface and switching functions)
353-557 (vacuum connection DN 25 ISO-KF) 353-558
(vacuum connection DN 40 CF-R)
353-562 (vacuum connection DN 25 ISO-KF, with baffle)
BCG450-SP (with Profibus interface and switching functions)
353-554 (vacuum connection DN 25 ISO-KF) 353-556
(vacuum connection DN 40 CF-R)
The part number (PN) can be taken from the product nameplate.
If not indicated otherwise in the legends, the illustrations in this docu-
ment correspond to gauge with part number 353-552. They apply to the other gauges by analogy.
All BCG450 versions are shipped with an instruction sheet ( [8]). BCG450-SD and BCG450-SP come with a supplementary instruction sheet describing the field­bus interfaces and the switching functions ( [9]).
We reserve the right to make technical changes without prior notice.
The BCG450, BCG450-SD and BCG450-SP gauges have been designed for vacuum measurement of gases in the pressure range 5×10
-10
… 1500 mbar.
They must not be used for measuring flammable or combustible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within the explosion range.
The gauges can be operated in connection with the INFICON Vacuum Gauge Controller VGC401, VGC402 or VGC403 or with other control devices.
2
tina40e1-b (2011-04)
Functional Principle
Trademarks
Due to the combination of three sensor technologies incorporated in the gauge (Capacitance diaphragm sensor, Pirani sensor and hot cathode ionisation sensor (BA)), a minimized gas type dependence is achieved.
Between 10 mbar and atmospheric pressure, the capacitance diaphragm sensor operates without any gas type dependence. Below 1 mbar, the Pirani sensor and the hot cathode ionisation sensor take over with only a small gas type dependence.
-3
Between 1 … 10 mbar and 5×10
… 2×10-2 mbar the gauges built in electronic circuits take care of continuous and smooth crossovers between the ranges. Over the whole measurement range, the measurement signal is output as a logarithm of the pressure.
The hot cathode is switched on by the Pirani measurement system only below the switching threshold of 2.4×10 off when the pressure exceeds 3.2×10
-2
mbar (to prevent filament burn-out). It is switched
-2
mbar.
Gauge adjustment is carried out automatically, no manual adjustment is required.
A user programmable atmospheric pressure switching function is incorporated.
DeviceNet™ Open DeviceNet Vendor Association, Inc.
TripleGauge™ INFICON AG, Balzers
tina40e1-b (2011-04) 3
Contents
Product Identification 2 Validity 2 Intended Use 2 Functional Principle 3 Trademarks 3
1
Safety 6
1.1 Symbols Used 6
1.2 Personnel Qualifications 6
1.3 General Safety Instructions 7
1.4 Liability and Warranty 7
2 Technical Data 8 3 Installation 13
3.1 Vacuum Connection 13
3.1.1 Removing and Installing the Electronics Unit 15
3.1.2 Using the Optional Baffle 16
3.2 Power Connection 17
3.2.1 Use With INFICON VGC40x Vacuum Gauge Controller 17
3.2.2 Use With Other Controllers 18
3.2.2.1 Making an Individual Sensor Cable 18
3.2.2.2 Making a DeviceNet Interface Cable (BCG450-SD) 21
3.2.2.3 Making a Profibus Interface Cable (BCG450-SP) 22
3.2.3 Using the Optional Power Supply (With RS232C Line) 23
4 Operation 25
4.1 Measuring Principle, Measuring Behavior 25
4.2 Operational Principle of the Gauge 27
4.3 Putting the Gauge Into Operation 28
4.4 Degas 28
4.5 Emission Control Mode 29
4.6 Atmosphere Switching Function 30
4.6.1 Functional Principle 30
4.6.2 Programming the Atmospheric Pressure Threshold 31
4.6.3 Wiring the relay "Atmospheric Pressure Reached" (BCG450) 32
4.7 Display (BCG450) 32
4.8 RS232C Interface 33
4.8.1 Description of the Functions 34
4.8.1.1 Output String (Transmit) 34
4.8.1.2 Input String (Receive) 36
4.9 DeviceNet Interface (BCG450-SD) 38
4.9.1 Description of the Functions 38
4.9.2 Operating Parameters 38
4.9.2.1 Operating Software 38
4.9.2.2 Node Address Setting 38
4.9.2.3 Data Rate Setting 39
4.9.3 Status Lights 39
4.10 Profibus Interface (BCG450-SP) 40
4.10.1 Description of the Functions 40
4.10.2 Operating Parameters 40
4.10.2.1 Operating Software 40
4.10.2.2 Node Address Setting 40
4.11 Switching Functions (BCG450-SD, -SP) 41
4.11.1 Setting the Switching Functions 41
5 Deinstallation 43
4
tina40e1-b (2011-04)
6 Maintenance, Repair 45
6.1 Maintenance 45
6.1.1 Cleaning the Gauge 45
6.2 Adjusting the Gauge 45
6.3 Adjusting the Atmosphere Sensor 45
6.4 What to Do in Case of Problems 47
6.5 Replacing the Sensor 49
7 Options 50 8 Spare Parts 50 9 Storage 50 10 Returning the Product 51 11 Disposal 51
Appendix 52
A: Relationship Measuring Signal – Pressure 52 B: Gas Type Dependence 53 C: Literature 55
Declaration of Contamination 56
For cross-references within this document, the symbol ( XY) is used, for cross- references to further documents and data sources, the symbol ( [Z]).
tina40e1-b (2011-04) 5
1 Safety
1.1 Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and environmental damage.
Caution
Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage.
Notice
1.2 Personnel Qualifications
Hint, recommendation
The result is O.K.
The result is not as expected.
Optical inspection
Waiting time, reaction time
Skilled personnel
All work described in this document may only be carried out by persons who have suitable technical training and the necessary experience or who have been instructed by the end-user of the product.
6
tina40e1-b (2011-04)
1.3 General Safety Instructions
1.4 Liability and Warranty
Adhere to the applicable regulations and take the necessary precautions for the process media used.
Consider possible reactions between the materials ( 11) and the process media.
Consider possible reactions of the process media (e.g. explosion) due to the heat generated by the product.
Adhere to the applicable regulations and take the necessary precautions for all work you are going to do and consider the safety instructions in this document.
Before beginning to work, find out whether any vacuum components are con­taminated. Adhere to the relevant regulations and take the necessary precau­tions when handling contaminated parts.
Communicate the safety instructions to all other users.
INFICON assumes no liability and the warranty becomes null and void if the end­user or third parties
disregard the information in this document
use the product in a non-conforming manner
make any kind of interventions (modifications, alterations etc.) on the product
use the product with accessories not listed in the corresponding product docu-
mentation.
The end-user assumes the responsibility in conjunction with the process media used.
Gauge failures due to contamination or wear and tear, as well as expendable parts
(e.g. filament), are not covered by the warranty.
tina40e1-b (2011-04) 7
2 Technical Data
Measurement principle
Measuring range
Emission
Degas
Output signal
Pressure range
10 … 1500 mbar capacitance diaphragm sensor
1 … 10 mbar crossover range
-2
2×10
5×10
5×10
… 1 mbar Pirani sensor
-3
…2×10-2 mbar crossover range
-10
…5×10-3 mbar hot cathode ionisation (BA)
Range (air, O2, CO, N2) 5×10
-10
… 1500 mbar, continuous
Accuracy
-8
1×10
… 50 mbar ±15% of reading
50 … 950 mbar ±5% of reading
950 … 1050 mbar ±2.5% of reading
(after 10 min. stabilisation)
-8
Repeatability
Gas type dependence
5% of reading, 10 (after 10 min. stabilisation)
Appendix B
… 10-2 mbar
Switching on threshold Switching off threshold
Emission current
p 7.2×10
7.2×10
-6
mbar
-6
mbar < p < 3.2×10-2 mbar
Emission current switching
25 µA 5 mA 5 mA 25 µA
2.4×10-2 mbar
-2
3.2×10
mbar
5 mA 25 µA
-6
7.2×10
3.0×10
mbar
-5
mbar
Degas emission current (p <7.2×10
-6
mbar)
20 mA (P
degas
4 W)
Control input signal 0 V/+24 VDC, active high
(control via RS232 33)
Duration max. 3 min, followed by automatic stop
A new degas cycle can only be started after a waiting time of 30 minutes.
In degas mode, BCG450 gauges keep supplying measurement values, however their tolerances may be higher than during normal operation.
Output signal (measuring signal) 0 … +10.13 V
Measuring range 0.774 … +10.13 V
(5×10
-10
mbar … 1500 mbar
Relationship voltage-pressure logarithmic, 0.75 V/decade
Appendix A)
(
Error signal +0.1 V Diaphragm sensor or
EEPROM error +0.3 V BA sensor error +0.5 V Pirani sensor error ( 47)
Minimum load impedance
10 kΩ
8
tina40e1-b (2011-04)
Display (BCG450)
Power supply
Display panel
Background illumination
Dimensions
Pressure units (pressure p)
LCD matrix, 32×16 pixels
two colors red/green
16.0 mm × 11.2 mm
mbar (default), Torr, Pa (selecting the pressure unit 33)
DANGER
The gauge may only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extra­low voltage (SELV). The connection to the gauge has to be fused
Operating voltage at the gauge +24 VDC (+20 … +28 VDC) 1)
Power consumption
Standard Degas Emission start (<200 ms)
Power consumption
BCG450 BCG450-SD BCG450-SP
Fuse necessary 1.25 AT
(INFICON controllers fulfill these requirements).
ripple max. 2 V
pp
0.5 A 0.9 A 1.4 A
18 W 20 W 20 W
Sensor cable connection
BCG450-SD requires an additional, separate power supply for the
DeviceNet interface ( 21).
Supply voltage at the DeviceNet con­nector, (Pin 2 and Pin 3)
+24 VDC (+11 … +25 VDC)
Power consumption <2 W
The gauge is protected against reversed polarity of the supply voltage.
For reasons of compatibility, the expression "sensor cable" is used for all
BCG450 versions in this document, although the pressure reading of the gauges with fieldbus interface (BCG450-SD and BCG450-SP) is nor-
Electrical connector
mally transmitted via the corresponding bus.
BCG450 BCG450-SD, -SP
D-Sub,15-pin, male
19 → 20
Measuring cable shielded, number of conductors de-
pending on the functions used (max. 15 conductors plus shielding)
1)
Cable length (supply voltage 24 V
Analog and fieldbus operation
RS232C operation
Gauge identification
)
35 m, conductor cross-section 0.25 mm²50 m, conductor cross-section 0.34 mm²100 m, conductor cross-section 1.0 mm²
30 m
42 kΩ resistor between Pin 10 and Pin 5 (sensor cable)
1)
Measured at sensor cable connector (consider the voltage drop as function of
the sensor cable length).
tina40e1-b (2011-04) 9
RS232C interface
DeviceNet interface (BCG450-SD)
Switching functions
BCG450
BCG450-SD, -SP
Adjustment range
Atmosphere switching function
2 (setpoints A and B)
-9
mbar … 100 mbar
1×10
Setpoints adjustable via potentiometers, one floating, normally open relay contact per setpoint ( 20, 41)
(Adjusting the setpoints via field bus is described in the corresponding bus sec­tions.)
Relay contact rating
Atmosphere switching function
BCG450
Relay contact rating
BCG450-SD/SP
Relay contact rating
Data rate
Data format
Connections (sensor cable connector) TxD (Transmit Data) RxD (Receive Data) Supply common (GND)
60 VDC, 0.5 ADC
Atmospheric pressure threshold pro­grammable via serial interfaces ( 30)
Atmospheric pressure threshold pro­grammable via RS232
Floating, normally open relay contact "atmosphere pressure reached" avail­able at the sensor cable connector
30 VAC/DC, 0.3 AC/DC
Atmospheric pressure threshold and re­lay function "atmosphere pressure reached" programmable via fieldbus in­terfaces ( [1] or [2]).
60 VDC, 0.5 ADC (same as SP A/B)
9600 Baud
binary 8 data bits one stop bit no parity bit no handshake
Pin 13 Pin 14 Pin 5
Function and communication protocol of the RS232C interface 33
Fieldbus name DeviceNet
Standard applied
Communication protocol, data format
Interface, physical CAN bus
Data rate (adjustable via "RATE"
switch)
[6]
[1], [4]
125 kBaud 250 kBaud 500 kBaud (default) "P" (125 kBaud, 250 kBaud, 500 kBaud programmable via DeviceNet ( [1])
Node address (MAC ID)
(Adjustable via "ADDRESS", "MSD", "LSD" switches)
0 … 63
(default = 63
dec
"P" (0 … 63 programmable via DeviceNet, [1])
dec
)
DeviceNet connector Micro-Style, 5-pin, male
Cable Shielded, special DeviceNet cable,
5 conductors ( 21 and [4])
Cable length, system wiring According to DeviceNet specifications
( [6], [4])
10
tina40e1-b (2011-04)
Profibus interface (BCG450-SP)
Materials used
Ambiance
Fieldbus name Profibus
Standard applied
Communication protocol data format
Interface, physical RS485
Data rate
Node address
Local (Adjustable via hexadecimal "ADDRESS", "MSD", "LSD" switches)
Default setting
Via Profibus (hexadecimal "ADDRESS" switches set to >7d
hex
(>125
dec
))
[7]
[9], [7]
12 MBaud ( [2])
00 … 7D
5C
hex
(0 … 125
hex
00 … 7D
(0 … 125
hex
dec
dec
)
)
Profibus connection D-Sub, 9-pin, female
Cable Shielded, special Profibus cable
( 22 and [5])
Cable length, system wiring According to Profibus specifications
( [7], [5])
Materials exposed to vacuum
Housing, supports, screens Feedthroughs Insulator Cathode Cathode holder Pirani element Sensor diaphragm Sensor contacts
Internal volume
DN 25 ISO-KF DN 40 CF-R
stainless steel NiFe, nickel plated glass iridium, yttrium oxide (Y2O3) molybdenum tungsten, copper ceramic (Al
2O3
)
SnAg
3
24 cm34 cm3
Pressure max. 5 bar (absolute)
Admissible temperatures
Storage
Operation
Bakeout
-20 … 70 °C
0 … 50 °C
+150 °C (at vacuum connection, without electronics unit, horizontally mounted
Relative humidity (year's mean / during 60 days)
65 / 85% (no condensation)
Use indoors only
altitude up to 2000 m NN
Mounting orientation any
Degree of protection IP 30
tina40e1-b (2011-04) 11
Dimensions [mm]
4-40UNC 2B
4-40UNC 2B
DN 25 ISO-KF
DN 40 CF-R
Weight
Gauges with DeviceNet connector are 14 mm longer.
353-550, 353-552, 353-561
353-551, 353-553 353-554, 353-557, 353-562 353-556, 353-558
305 g 565 g 445 g 710 g
12
tina40e1-b (2011-04)
3 Installation
3.1 Vacuum Connection
DANGER: overpressure in the vacuum system >1 bar
Injury caused by released parts and harm caused by escaping process gases can result if clamps are opened while the vacuum system is pressurized.
Do not open any clamps while the vacuum system is pressurized. Use the type of clamps which are suited to overpressure.
DANGER: overpressure in the vacuum system >2.5 bar
KF flange connections with elastomer seals (e.g. O-rings) cannot withstand such pressures. Process media can thus leak and possibly damage your health.
Use O-rings provided with an outer centering ring.
DANGER
DANGER
The gauge must be electrically connected to the grounded vacuum chamber. This connection must conform to the requirements of a pro­tective connection according to EN 61010:
CF connections fulfill this requirement
For gauges with a KF vacuum connection, use a conductive metallic clamping ring.
Caution: vacuum component
Dirt and damages impair the function of the vacuum component.
When handling vacuum components, take appropriate measures to ensure cleanliness and prevent damages.
Caution: dirt sensitive area
Touching the product or parts thereof with bare hands increases the desorption rate.
Always wear clean, lint-free gloves and use clean tools when working in this area.
DANGER
Caution
Caution
The gauge may be mounted in any orientation. To keep condensates
and particles from getting into the measuring chamber, preferably choose a horizontal to upright position. See dimensional drawing for space requirements ( 12).
The gauge is supplied with a built-in grid. For potentially contaminating appli­cations and to protect the electrodes against light and fast charged particles, in­stallation ( 16) of the optional baffle is recommended (→ 50).
The sensor can be baked at up to 80 °C (at vacuum connection, horizontally mounted). At temperatures exceeding 50 °C, the electronics unit has to be re­moved ( 15).
tina40e1-b (2011-04) 13
Procedure
Remove the protective lid.
The protective lid will be needed for maintenance..
Make the flange connection to the vacuum system
INFICON recommends to install the gauge without applying vacuum
grease.
Seal with centering ring
or
When installing the gauge, make sure that the area around the con-
nector is accessible for the tools required for adjustment while the gauge is mounted ( 41, 45).
When installing the gauge, allow for installing/deinstalling the connectors and accommodation of cable loops.
If you are using a gauge with display, make sure easy reading of the display is possible.
14
Seal with centering ring and baffle (Option)
tina40e1-b (2011-04)
3.1.1 Removing and Installing the Electronics Unit
Required tools/material
Removing the electronics unit
Allen key, AF 2.5
Unscrew the hexagon socket set screw (1) on the side of the electronics
unit (2).
2
1
Installing the electronics unit
Remove the electronics unit without twisting it.
Place the electronics unit (2) on the sensor (3) (be careful to correctly align
the pins and notch (4)).
4
3
2
Slide the electronics unit in to the mechanical stop and lock it with the hexa-
gon socket set screw.
tina40e1-b (2011-04) 15
3.1.2 Using the Optional Baffle
Requirement
Required tools / material
Installation
In severely contaminating processes and to protect measurement electrodes opti­cally against light and fast charged particles, replacement of the built-in grid by the optional baffle ( 50) is recommended.
The gauge is deinstalled (deinstallation gauge 43).
Baffle (→ 50)
Pointed tweezers
Pin (e.g. pencil)
Screwdriver No 1
Carefully remove the grid with tweezers.
Carefully place the baffle onto the sensor opening.
16
tina40e1-b (2011-04)
Using a pin, press the baffle down in the center until it catches.
Deinstallation
3.2 Power Connection
3.2.1 Use With INFICON VGC40x Vacuum Gauge Controller
Carefully remove the baffle with the screwdriver.
If the gauge is used with an INFICON VGC40x controller, a corresponding sensor cable is required ( [3]). The sensor cable permits supplying the gauge with power, transmitting measurement values and gauge statuses, and making pa­rameter settings.
Caution
Caution: data transmission errors
The attempt to operate a fieldbus gauge (BCG450-SD / -SP) with the VGC40x Vacuum Gauge Controller (RS232C) causes data transmis­sion errors.
Fieldbus gauges must not be operated with an INFICON VGC40x
Required material
Sensor cable ( [3], INFICON sales literature)
tina40e1-b (2011-04) 17
controller.
Procedure
Plug the sensor connector into the gauge and secure it with the locking
screws.
Connect the other end of the sensor cable to the INFICON controller and
secure it.
3.2.2 Use With Other Controllers
3.2.2.1 Making an Individual
Sensor Cable
Cable type
Procedure
The gauge can now be operated with the VGC40x controller.
The gauge can also be operated with other controllers.
Especially the fieldbus versions BCG450-SD (DeviceNet) and BCG450-SP (Profibus) are usually operated as part of a network, controlled by a master or bus controller. In such cases, the control system has to be operated with the appropri­ate software and communication protocol.
For reasons of compatibility, the expression "sensor cable" is used for all
BCG450 versions in this document, although the pressure reading of the gauges with fieldbus interface (BCG450-SD or BCG450-SP) is normally transmitted via DeviceNet or Profibus.
The sensor cable is required for supplying all BCG450 types with power. It also permits access to the relay contacts of the switching functions ( 20).
The application and length of the sensor cable have to be considered when deter­mining the number and cross sections of the conductors ( 9).
Open the cable connector (D-Sub, 15-pin, female). Prepare the cable and solder/crimp it to the connector as indicated in the
diagram of the gauge used:
18
tina40e1-b (2011-04)
Loading...
+ 42 hidden pages