In all communications with INFICON, please specify the information on the product
nameplate. For convenient reference copy that information into the space provided
below.
INFICON AG, LI-9496 Balzers
Model:
PN:
SN:
V W
Validity
Intended Use
This document applies to products with the following part numbers:
BCG450 (without display)
353-550 (vacuum connection DN 25 ISO-KF)
353-551
(vacuum connection DN 40 CF-R)
353-561 (vacuum connection DN 25 ISO-KF, with baffle)
BCG450 (with display)
353-552 (vacuum connection DN 25 ISO-KF)
353-553
(vacuum connection DN 40 CF-R)
BCG450-SD (with DeviceNet interface and switching functions)
353-557 (vacuum connection DN 25 ISO-KF)
353-558
(vacuum connection DN 40 CF-R)
353-562 (vacuum connection DN 25 ISO-KF, with baffle)
BCG450-SP (with Profibus interface and switching functions)
353-554 (vacuum connection DN 25 ISO-KF)
353-556
(vacuum connection DN 40 CF-R)
The part number (PN) can be taken from the product nameplate.
If not indicated otherwise in the legends, the illustrations in this docu-
ment correspond to gauge with part number 353-552. They apply to the
other gauges by analogy.
All BCG450 versions are shipped with an instruction sheet (→ [8]). BCG450-SD
and BCG450-SP come with a supplementary instruction sheet describing the fieldbus interfaces and the switching functions (→ [9]).
We reserve the right to make technical changes without prior notice.
The BCG450, BCG450-SD and BCG450-SP gauges have been designed for
vacuum measurement of gases in the pressure range 5×10
-10
… 1500 mbar.
They must not be used for measuring flammable or combustible gases in mixtures
containing oxidants (e.g. atmospheric oxygen) within the explosion range.
The gauges can be operated in connection with the INFICON Vacuum Gauge
Controller VGC401, VGC402 or VGC403 or with other control devices.
2
tina40e1-b (2011-04)
Functional Principle
Trademarks
Due to the combination of three sensor technologies incorporated in the gauge
(Capacitance diaphragm sensor, Pirani sensor and hot cathode ionisation sensor
(BA)), a minimized gas type dependence is achieved.
Between 10 mbar and atmospheric pressure, the capacitance diaphragm sensor
operates without any gas type dependence. Below 1 mbar, the Pirani sensor and
the hot cathode ionisation sensor take over with only a small gas type dependence.
-3
Between 1 … 10 mbar and 5×10
… 2×10-2 mbar the gauges built in electronic
circuits take care of continuous and smooth crossovers between the ranges. Over
the whole measurement range, the measurement signal is output as a logarithm of
the pressure.
The hot cathode is switched on by the Pirani measurement system only below the
switching threshold of 2.4×10
off when the pressure exceeds 3.2×10
-2
mbar (to prevent filament burn-out). It is switched
-2
mbar.
Gauge adjustment is carried out automatically, no manual adjustment is required.
A user programmable atmospheric pressure switching function is incorporated.
DeviceNet™ Open DeviceNet Vendor Association, Inc.
3.1.1 Removing and Installing the Electronics Unit 15
3.1.2 Using the Optional Baffle 16
3.2 Power Connection 17
3.2.1 Use With INFICON VGC40x Vacuum Gauge Controller 17
3.2.2 Use With Other Controllers 18
3.2.2.1 Making an Individual Sensor Cable 18
3.2.2.2 Making a DeviceNet Interface Cable (BCG450-SD) 21
3.2.2.3 Making a Profibus Interface Cable (BCG450-SP) 22
3.2.3 Using the Optional Power Supply (With RS232C Line) 23
4 Operation 25
4.1 Measuring Principle, Measuring Behavior 25
4.2 Operational Principle of the Gauge 27
4.3 Putting the Gauge Into Operation 28
4.4 Degas 28
4.5 Emission Control Mode 29
4.6 Atmosphere Switching Function 30
4.6.1 Functional Principle 30
4.6.2 Programming the Atmospheric Pressure Threshold 31
4.6.3 Wiring the relay "Atmospheric Pressure Reached" (BCG450) 32
4.7 Display (BCG450) 32
4.8 RS232C Interface 33
4.8.1 Description of the Functions 34
4.8.1.1 Output String (Transmit) 34
4.8.1.2 Input String (Receive) 36
4.9 DeviceNet Interface (BCG450-SD) 38
4.9.1 Description of the Functions 38
4.9.2 Operating Parameters 38
4.9.2.1 Operating Software 38
4.9.2.2 Node Address Setting 38
4.9.2.3 Data Rate Setting 39
4.9.3 Status Lights 39
4.10 Profibus Interface (BCG450-SP) 40
4.10.1 Description of the Functions 40
4.10.2 Operating Parameters 40
4.10.2.1 Operating Software 40
4.10.2.2 Node Address Setting 40
4.11 Switching Functions (BCG450-SD, -SP) 41
4.11.1 Setting the Switching Functions 41
5 Deinstallation 43
4
tina40e1-b (2011-04)
6 Maintenance, Repair 45
6.1 Maintenance 45
6.1.1 Cleaning the Gauge 45
6.2 Adjusting the Gauge 45
6.3 Adjusting the Atmosphere Sensor 45
6.4 What to Do in Case of Problems 47
6.5 Replacing the Sensor 49
7 Options 50
8 Spare Parts 50
9 Storage 50
10 Returning the Product 51
11 Disposal 51
Appendix 52
A: Relationship Measuring Signal – Pressure 52
B: Gas Type Dependence 53
C: Literature 55
Declaration of Contamination 56
For cross-references within this document, the symbol (→ XY) is used, for cross-
references to further documents and data sources, the symbol (→ [Z]).
tina40e1-b (2011-04)5
1 Safety
1.1 Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and environmental damage.
Caution
Information on correct handling or use. Disregard can lead to malfunctions or
minor equipment damage.
Notice
1.2 Personnel Qualifications
Hint, recommendation
The result is O.K.
The result is not as expected.
Optical inspection
Waiting time, reaction time
Skilled personnel
All work described in this document may only be carried out by persons who
have suitable technical training and the necessary experience or who have been
instructed by the end-user of the product.
6
tina40e1-b (2011-04)
1.3 General Safety
Instructions
1.4 Liability and Warranty
• Adhere to the applicable regulations and take the necessary precautions for the
process media used.
Consider possible reactions between the materials (→ 11) and the process
media.
Consider possible reactions of the process media (e.g. explosion) due to the
heat generated by the product.
• Adhere to the applicable regulations and take the necessary precautions for all
work you are going to do and consider the safety instructions in this document.
• Before beginning to work, find out whether any vacuum components are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.
Communicate the safety instructions to all other users.
INFICON assumes no liability and the warranty becomes null and void if the enduser or third parties
• disregard the information in this document
• use the product in a non-conforming manner
• make any kind of interventions (modifications, alterations etc.) on the product
• use the product with accessories not listed in the corresponding product docu-
mentation.
The end-user assumes the responsibility in conjunction with the process media
used.
Gauge failures due to contamination or wear and tear, as well as expendable parts
(e.g. filament), are not covered by the warranty.
tina40e1-b (2011-04)7
2 Technical Data
Measurement principle
Measuring range
Emission
Degas
Output signal
Pressure range
10 … 1500 mbar capacitance diaphragm sensor
1 … 10 mbar crossover range
-2
2×10
5×10
5×10
… 1 mbar Pirani sensor
-3
…2×10-2 mbar crossover range
-10
…5×10-3 mbar hot cathode ionisation (BA)
Range (air, O2, CO, N2) 5×10
-10
… 1500 mbar, continuous
Accuracy
-8
1×10
… 50 mbar ±15% of reading
50 … 950 mbar ±5% of reading
950 … 1050 mbar ±2.5% of reading
(after 10 min. stabilisation)
-8
Repeatability
Gas type dependence
5% of reading, 10
(after 10 min. stabilisation)
→ Appendix B
… 10-2 mbar
Switching on threshold
Switching off threshold
Emission current
p ≤7.2×10
7.2×10
-6
mbar
-6
mbar < p < 3.2×10-2 mbar
Emission current switching
25 µA ⇒ 5 mA
5 mA ⇒ 25 µA
2.4×10-2 mbar
-2
3.2×10
mbar
5 mA
25 µA
-6
7.2×10
3.0×10
mbar
-5
mbar
Degas emission current
(p <7.2×10
-6
mbar)
≈20 mA (P
degas
≈4 W)
Control input signal 0 V/+24 VDC, active high
(control via RS232 → 33)
Duration max. 3 min, followed by automatic stop
A new degas cycle can only be started
after a waiting time of 30 minutes.
In degas mode, BCG450 gauges keep supplying measurement values, however
their tolerances may be higher than during normal operation.
EEPROM error
+0.3 V BA sensor error
+0.5 V Pirani sensor error
(→ 47)
Minimum load impedance
10 kΩ
8
tina40e1-b (2011-04)
Display (BCG450)
Power supply
Display panel
Background illumination
Dimensions
Pressure units (pressure p)
LCD matrix, 32×16 pixels
two colors red/green
16.0 mm × 11.2 mm
mbar (default), Torr, Pa
(selecting the pressure unit → 33)
DANGER
The gauge may only be connected to power supplies, instruments or
control devices that conform to the requirements of a grounded extralow voltage (SELV). The connection to the gauge has to be fused
Operating voltage at the gauge +24 VDC (+20 … +28 VDC) 1)
Power consumption
Standard
Degas
Emission start (<200 ms)
Power consumption
BCG450
BCG450-SD
BCG450-SP
Fuse necessary 1.25 AT
(INFICON controllers fulfill these requirements).
ripple max. 2 V
pp
≤0.5 A
≤0.9 A
≤1.4 A
≤18 W
≤20 W
≤20 W
Sensor cable connection
BCG450-SD requires an additional, separate power supply for the
DeviceNet interface (→ 21).
Supply voltage at the DeviceNet connector, (Pin 2 and Pin 3)
+24 VDC (+11 … +25 VDC)
Power consumption <2 W
The gauge is protected against reversed polarity of the supply voltage.
For reasons of compatibility, the expression "sensor cable" is used for all
BCG450 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BCG450-SD and BCG450-SP) is nor-
Electrical connector
mally transmitted via the corresponding bus.
BCG450
BCG450-SD, -SP
D-Sub,15-pin, male
→ 19
→ 20
Measuring cable shielded, number of conductors de-
pending on the functions used
(max. 15 conductors plus shielding)
1)
Cable length (supply voltage 24 V
Analog and fieldbus operation
RS232C operation
Gauge identification
)
≤35 m, conductor cross-section 0.25 mm²
≤50 m, conductor cross-section 0.34 mm²
≤100 m, conductor cross-section 1.0 mm²
≤30 m
42 kΩ resistor between Pin 10 and Pin 5
(sensor cable)
1)
Measured at sensor cable connector (consider the voltage drop as function of
the sensor cable length).
tina40e1-b (2011-04)9
RS232C interface
DeviceNet interface
(BCG450-SD)
Switching functions
BCG450
BCG450-SD, -SP
Adjustment range
→ Atmosphere switching function
2 (setpoints A and B)
-9
mbar … 100 mbar
1×10
Setpoints adjustable via potentiometers,
one floating, normally open relay contact
per setpoint (→ 20, 41)
(Adjusting the setpoints via field bus is
described in the corresponding bus sections.)
Injury caused by released parts and harm caused by escaping process
gases can result if clamps are opened while the vacuum system is
pressurized.
Do not open any clamps while the vacuum system is pressurized. Use
the type of clamps which are suited to overpressure.
DANGER: overpressure in the vacuum system >2.5 bar
KF flange connections with elastomer seals (e.g. O-rings) cannot
withstand such pressures. Process media can thus leak and possibly
damage your health.
Use O-rings provided with an outer centering ring.
DANGER
DANGER
The gauge must be electrically connected to the grounded vacuum
chamber. This connection must conform to the requirements of a protective connection according to EN 61010:
• CF connections fulfill this requirement
For gauges with a KF vacuum connection, use a conductive metallic
clamping ring.
Caution: vacuum component
Dirt and damages impair the function of the vacuum component.
When handling vacuum components, take appropriate measures to
ensure cleanliness and prevent damages.
Caution: dirt sensitive area
Touching the product or parts thereof with bare hands increases the
desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
DANGER
Caution
Caution
The gauge may be mounted in any orientation. To keep condensates
and particles from getting into the measuring chamber, preferably
choose a horizontal to upright position. See dimensional drawing for
space requirements (→ 12).
• The gauge is supplied with a built-in grid. For potentially contaminating applications and to protect the electrodes against light and fast charged particles, installation (→ 16) of the optional baffle is recommended (→ 50).
• The sensor can be baked at up to 80 °C (at vacuum connection, horizontally
mounted). At temperatures exceeding 50 °C, the electronics unit has to be removed (→ 15).
tina40e1-b (2011-04)13
Procedure
Remove the protective lid.
The protective lid will be needed for maintenance..
Make the flange connection to the vacuum system
INFICON recommends to install the gauge without applying vacuum
grease.
Seal with centering ring
or
When installing the gauge, make sure that the area around the con-
nector is accessible for the tools required for adjustment while the gauge
is mounted (→ 41, 45).
When installing the gauge, allow for installing/deinstalling the connectors
and accommodation of cable loops.
If you are using a gauge with display, make sure easy reading of the
display is possible.
14
Seal with centering ring
and baffle (Option)
tina40e1-b (2011-04)
3.1.1 Removing and Installing
the Electronics Unit
Required tools/material
Removing the electronics unit
• Allen key, AF 2.5
Unscrew the hexagon socket set screw (1) on the side of the electronics
unit (2).
2
1
Installing the electronics unit
Remove the electronics unit without twisting it.
Place the electronics unit (2) on the sensor (3) (be careful to correctly align
the pins and notch (4)).
4
3
2
Slide the electronics unit in to the mechanical stop and lock it with the hexa-
gon socket set screw.
tina40e1-b (2011-04)15
3.1.2 Using the Optional Baffle
Requirement
Required tools / material
Installation
In severely contaminating processes and to protect measurement electrodes optically against light and fast charged particles, replacement of the built-in grid by the
optional baffle (→ 50) is recommended.
The gauge is deinstalled (deinstallation gauge → 43).
• Baffle (→ 50)
• Pointed tweezers
• Pin (e.g. pencil)
• Screwdriver No 1
Carefully remove the grid with tweezers.
Carefully place the baffle onto the sensor opening.
16
tina40e1-b (2011-04)
Using a pin, press the baffle down in the center until it catches.
Deinstallation
3.2 Power Connection
3.2.1 Use With INFICON
VGC40x Vacuum Gauge
Controller
Carefully remove the baffle with the screwdriver.
If the gauge is used with an INFICON VGC40x controller, a corresponding sensor
cable is required (→ [3]). The sensor cable permits supplying the gauge with
power, transmitting measurement values and gauge statuses, and making parameter settings.
Caution
Caution: data transmission errors
The attempt to operate a fieldbus gauge (BCG450-SD / -SP) with the
VGC40x Vacuum Gauge Controller (RS232C) causes data transmission errors.
Fieldbus gauges must not be operated with an INFICON VGC40x
Required material
• Sensor cable (→ [3], INFICON sales literature)
tina40e1-b (2011-04)17
controller.
Procedure
Plug the sensor connector into the gauge and secure it with the locking
screws.
Connect the other end of the sensor cable to the INFICON controller and
secure it.
3.2.2 Use With Other
Controllers
3.2.2.1 Making an Individual
Sensor Cable
Cable type
Procedure
The gauge can now be operated with the VGC40x controller.
The gauge can also be operated with other controllers.
Especially the fieldbus versions BCG450-SD (DeviceNet) and BCG450-SP
(Profibus) are usually operated as part of a network, controlled by a master or bus
controller. In such cases, the control system has to be operated with the appropriate software and communication protocol.
For reasons of compatibility, the expression "sensor cable" is used for all
BCG450 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BCG450-SD or BCG450-SP) is normally
transmitted via DeviceNet or Profibus.
The sensor cable is required for supplying all BCG450 types with power.
It also permits access to the relay contacts of the switching functions
(→ 20).
The application and length of the sensor cable have to be considered when determining the number and cross sections of the conductors (→ 9).
Open the cable connector (D-Sub, 15-pin, female).
Prepare the cable and solder/crimp it to the connector as indicated in the
diagram of the gauge used:
18
tina40e1-b (2011-04)
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.