IKA Roller 6 basic, Roller 6 digital, Roller 10 basic, Roller 10 digital User Manual

Page 1
20000005506
Roller 6/10 bd_102016
IKA® Roller 6 basic
IKA
IKA® Roller 10 digitalIKA® Roller 10 basic
®
Roller 6 digital
Page 2
Source language: German
EN
Contents
Page
Declaration of conformity 10 Explication of warning symbols 10 Warranty 10 Safety instructions 11 Unpacking 11 Correct use 12 Useful information 12 Commissioning 12 Accessories 14 Maintenance and cleaning 15 Error codes 15 Technical data 16
Declaration of conformity
We declare under our sole responsibility that this product corresponds to the regulations 2006/42/EC, 2011/65/EU, 2014/30/EU and 2014/35/EU and conforms with the standards or standardized documents: EN 61010-1, EN 61010-2-051, EN 61326-1, EN 60529 (A1+A2) and EN ISO 12100.
Explication of warning symbols
General hazard
Crush hazard!
Warranty
In accordance with IKA® warranty conditions, the warranty pe­riod is 24 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or mainte­nance not carried out in accordance with the instructions in this operating manual.
Page 3
Safety instructions
For your protection
Read the operating instructions in full before starting up and follow the safety instruc­tions.
Keep the operating instructions in a place where they can be accessed by everyone.
Ensure that only trained staff work with the appliance.
• Follow the safety instructions, guidelines, occupational health and safety and accident prevention regulations.
• Wear your personal protective equipment in accordance with the hazard category of the medium to be processed. Otherwise there is a risk of:
- splashing liquids
- projectile parts
- body parts, hair, clothing and jewellery getting caught.
• Set up the device in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface.
The feet of the appliance must be clean and undamaged.
Check the appliance and accessories beforehand for dama­ge each time you use them. Do not use damaged compo­nents.
Caution! There is a crush hazard between the moving parts and the casing.
Place the test tubes safely between the rollers.
• Ensure accessory parts are secured carefully as test tubes can be damaged or ejected otherwise.
Always load the samples evenly on the device.
The weight of the load must not exceed 2 kg. Do not ex­ceed the maximum permissible shaking weights.
Do not move or transport the device when it is operating or connected to the power supply.
• Reduce the speed if
- medium sprays out of test tube due to excessive speed,
- the appliance is not running smoothly,
- dynamic forces start to cause the device and/or test tubes to move around.
The appliance is not suitable for manual operation.
Beware of the risk of
- flammable materials
- glass breakage as a result of mechanical shaking power
The appliance may heat up when in use.
• Only process media that will not react dangerously to the extra energy produced through processing. This also applies to any extra energy produced in other ways, e.g. through light irradiation.
• Process pathogenic materials only in closed vessels under a suitable extractor hood. Please contact IKA
®
if you have
any questions.
• Do not operate the appliance in explosive atmospheres, with hazardous substances or under water.
• Safe operation is only guaranteed with the accessories de­scribed in the ”Accessories” chapter.
Always disconnect the plug before fitting accessories.
• Disconnect the device from the power supply before you open the cover or adjust the inclination angle of the device shaking table.
Do not operate the device with the cover open.
• The appliance can only be disconnected from the mains supply by pulling out the mains plug or the connector plug.
The socket for the mains cord must be easily accessible.
• The IKA when power is restored after a power outage. If the IKA
®
Roller 6/10 basic does not restart automatically
®
Roller 6/10 digital is running in mode A it does not restart
automatically when power is restored after a power outage.
For protection of the equipment
The appliance may only be opened by experts.
The voltage stated on the type plate must correspond to the mains voltage.
• Removable parts must be refitted to the appliance to pre­vent the infiltration of foreign objects, liquids etc.
Protect the appliance and accessories from bumps and im­pacts.
The device must only be operated with the original plug-in power supply unit.
Unpacking
Unpacking
Please unpack the device carefully.
In the case of any damage a detailed report must be sent immediately (post, rail or forwarder).
Delivery scope
®
• IKA
AC adapter, see Fig. 1
• Plate SH
• Operating instructions
• Warranty card
Roller 6/10 basic/digital, depending on the type that was ordered
Fig. 1
Page 4
Correct use
Use
For mixing and agitating liquids.
®
IKA
Roller 6/10 basic
Continuous mode for single or multiple test tubes.
®
IKA
Roller 6/10 digital
Continuous mode for single or multiple test tubes.
Continuous mode for single or multiple test tubes.
Useful information
Individual rollers can be removed to allow tubes of different sizes to be processed. In this case the rollers automatically take up symmetrical positions on the shaking table, so that the rollers can be removed and inserted as desired.
Commissioning
• Check that the power supply matches the data on the ra­ting plate.
• Plug the AC adapter into a standard mains socket in order to supply the device with power. See the section "Technical data" for the connection values.
• Also pay attention to the ambient conditions listed in the “Technical data”.
Range of use
- Laboratories - Schools
- Pharmacies - Universities
The unit is suitable for use in all areas apart from domestic areas and areas directly connected to a low-voltage supply which also serves domestic areas.
The safety of the user cannot be guaranteed if the appliance is operated with accessories that are not supplied or recommended by the manufacturer or if the appliance is operated improperly contrary to the manufacturer’s specifications or if the appliance or the printed circuit board are modified by third parties.
The integral tube stopper prevents the tubes falling out. For the safety of the operator, the device switches off automati­cally if the rollers jam and/or can no longer rotate freely.
®
• The IKA when power is restored after a power outage. If the IKA Roller 6/10 digital is running in mode A it does not restart automatically when power is restored after a power outage.
Roller 6/10 basic does not restart automatically
If these conditions are met the unit is ready for service when the mains plug has been plugged in.
®
IKA® Roller 6/10 basic
The AC adapter is suitable for 100 V upto 240 V. Insert an appro­priate adapter corresponding to the country, before plugging the AC adapter into the mains socket.
Switching the device on
Press the "Power" button (L) to switch the device on. Î The LED (M) lights up green.
Starting the device
Press the "Start/Stop" button (F) to start the shaking process. The device starts at the preset speed of 30 rpm.
Stopping the device
Press the "Start/Stop" button (F) to stop the shaking process.
Switching off the device
Press the "Power" button (L) to switch the device off. Î The LED (M) goes out.
Page 5
IKA® Roller 6/10 digital
The AC adapter is suitable for 100 V upto 240 V. Insert an appro­priate adapter corresponding to the country, before plugging the AC adapter into the mains socket.
Switching the device on
Press the "Power" button (L) to switch the device on.
The following data appear on the display in the sequence listed:
the test image
the software version
the operating mode
the values and settings last used
Speed setting
Use the "Speed setting" button (K) to set the speed to a value between 5 and 80 rpm in units of 1 rpm.
Timer/Counter
Select "Counter" or "Timer" (I).
Timer When "Timer" is selected, the set value counts down. Once the timed shaking cycle has been completed the device switches off automatically.
An acoustic signal beeps three times and the "Timer" LED (H)
Î flashes red.
Counter The counter measures the time from the start of shaking to the end. Î The "Counter" LED (G) flashes red. If the "Start/Stop" button (F) is pressed the shaking process stops and the elapsed time is displayed. If the counter is restarted the measurement restarts afresh.
To switch from counter to timer or vice versa, press the "Timer" button (I).
Setting Set the timer from 1 sec to 99 h 59 min using the "Plus/Minus" timer setting buttons (J). Press the "Start/Stop" button (F) to start the shaker. The last timer value setting is saved and is loaded for the next shaking process.
Resetting to 0 To reset the timer, press the "Timer Minus" and "Timer Plus" but­tons (J) simultaneously.
Starting the device
Press the "Start/Stop" button (F) to start the shaking process. The device starts with the desired settings.
Stopping the device
Press the "Start/Stop" button (F) to pause operation.
Switching off the device
Press the "Power" button (L) to switch the device off.
Page 6
Operating modes A and B
Operating mode A
The device does not restart automatically when power is resto­red after a power outage.
Operating mode B
The device restarts automatically when power is restored after a power outage, using the set values, and displays the (Pr) symbol to indicate that there has been a power outage within a certain period of time.
Switching between operating modes
To switch between operating modes, switch the device off, then keep the "Start/Stop" button (F) depressed and press the "Pow­er" button (L).
Caution! Do not change the tube fences or rollers whilst the device is in operation! Stop the device by pressing the "Start/Stop" button (F). Do not use your hand to stop the device. Press the "Power" button (L) to switch the device off.
Removal / fitting tube fences
The front and back tube fences must be in position when the device is in operation. The side tube fences are optional; for certain sizes of tubes the side tube fences offer additional safety.
Fitting
Plug in the tube fences (1).
Removal
Take the tube fences off (1).
Fitting / removing the rollers
Removal
Take the tube fences off (1).
Release the roller and pull it out (2).
Remove the roller completely (3).
Fitting
Fitting is the reverse of removal. A "click" confirms achievement of the correct position.
1
2
1
3
Accessories
Designation Description
Plate For certain sizes of tubes the side tube fences offer SH additional safety, so that the tubes cannot fall off the shaker.
Page 7
Maintenance and cleaning
The device operates maintenance-free.
For cleaning disconnect the main plug.
Wear protective gloves when cleaning the devices.
• Do not place electrical appliances into the cleaning agents for cleaning purposes.
• Do not allow moisture to get into the appliance when cleaning.
• Please consult with IKA
®
before using any cleaning or de­contamination methods, other than those recommended here.
Only clean IKA
®
appliances using these IKA® approved cleaning
agents:
Dirt Cleaning agent
Dyes isopropanol Building materials water containing detergent, isopropanol Cosmetics water containing detergent, isopropanol Food water containing detergent Fuels water containing detergent Other materials please consult IKA
®
Error codes
Spare parts order When ordering spare parts, please give:
• Machine type
Manufacturing number, see type plate
Item and designation of the spare part, see www.ika.com, spare parts diagram and spare parts list
Repair
In case of repair the device has to be cleaned and free from any materials which may constitute a health ha­zard.
For this you should use the downloaded print of the "Certificate of compliance" form from the IKA
®
website www.ika.com.
If you require servicing, return the appliance in its original pa­ckaging. Storage packaging is not sucient. Please also use sui­table transport packaging.
IKA® Roller 6/10 basic
In case of a fault, a beep is heard and the “Power” LED flashes.
Proceed as follows in such cases:
1. Switch the device off.
2. Perform corrective measures as described below.
3. Perform a restart.
Error Cause Effect Solution
Motor blockage or overload Motor off Switch the device off.
Reduce the load on the motor and check all moving parts for freedom of movement.
Then perform a restart.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
• Contact the IKA
®
service department,
Send the device for repair, including a short description of the fault.
®
Roller 6/10 digital
IKA
The fault is shown by an error code in the display as following if the error occurs, e.g. Er 4.
Proceed as follows in such cases:
1. Switch the device off.
2. Perform corrective measures as described below.
3. Perform a restart.
Error Cause Effect Solution
Er 3 Temperature inside device is too
high
Motor off Switch the device off.
Allow the device to cool down.
Er 4 Motor blockage or overload Motor off Switch the device off.
Reduce the load on the motor and check all moving parts for freedom of movement.
Then perform a restart.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
• Contact the IKA
®
service department,
Send the device for repair, including a short description of the fault.
Page 8
Technical data
Power supply
Input V 100 - 240
A 0,8
Hz 50-60
Output VDC 24
24W
.LPS
(Limited power source)
Protection class
Shaker: IKA® Roller 6 basic 10 basic 6 digital 10 digital
DC V 24
Max. device power consumption W 24
Motor power consumption W 16
Motor output power W 9
Type of movement rolling
Permissible shaken weight (incl. fixture)
Inclination angle of the shaking table ° 3
Speed min. rpm -5
Speed preset definitely rpm 30 -
Speed max. rpm -80
Speed variation % 10 5
Adjustment
Speed display - 7-segment LED
Timer - yes
Timer display - 7-segment LED
Operating mode continuous timer and continu-operation
Adjustment
Adjustment
Abmessungen (B x H x T) mm 240 x 115 x 545 380 x 115 x 545
Dimensions (W x D x H) kg 4,6 6,8 4,6 6,8
Number of rollers 6 10 6 10
Roller diameter mm 32 32
Length of the working area mm 340 340
Max. travel at the end of the rollers mm 16 16
Perm. duration of operation % 100
Perm. ambient temperature °C
P
erm. rel. humidity % 80
Protection acc. DIN EN 60529 IP 21
Overvoltage category II
Protection class II
Contamination level 2
Operation at a terrestrial altitude m max. 2000 above sea level
speed - 1 rpm steps
timer min. sec -1
timer max. h:min - 99 h 59 min
2 (double insulated)
kg
2
+ 4 bis + 50
Subject to technical changes!
Loading...