IKA RC 5 basic User Manual

Page 1
20000017145a
RC 5 basic_062018
IKA RC 5 basic
Page 2
Device setup
1
2
3
4
3
Fig. 1
Item Designation 1 Filling opening lid 2 Operator panel and display 3 Handle 4 Venting grid 5 RS 232 port 6 USB port 7 External temperature sensor socket 8 Pump connector IN 9 Pump connector OUT 10 Power switch 11 Power socket 12 Backflow 13 Overflow
8 9
5 6
7
10
11 12 13
3
4
3
2
Page 3
Sourcing language: German
EN
Contents
Page
Device setup ............................................................................................................................................. 2
Declaration of Conformity ........................................................................................................................ 4
Explication of warning symbols ............................................................................................................... 4
Safety instructions ................................................................................................................................... 5
General information ......................................................................................................................................... 5
Disposal of the device ...................................................................................................................................... 5
Fluids ............................................................................................................................................................... 6
Refrigerant
Correct use ............................................................................................................................................... 6
Use .................................................................................................................................................................. 6
Range of use (indoor use only) ......................................................................................................................... 6
Unpacking ................................................................................................................................................ 7
Unpacking ....................................................................................................................................................... 7
Delivery scope .................................................................................................................................................. 7
Preparations ............................................................................................................................................. 7
Setting up ........................................................................................................................................................ 7
Connecting the tubings .................................................................................................................................... 7
Filling and draining ........................................................................................................................................... 8
Fluid (Standard information for IKA fluid) ......................................................................................................... 9
Moving the device ......................................................................................................................................... 10
Operator panel and display .................................................................................................................... 10
Commissioning and operating ................................................................................................................ 11
Menu settings ........................................................................................................................................ 12
Menu structure .............................................................................................................................................. 12
Basic guide to setting the menu options ......................................................................................................... 12
Operating mode (MODE)................................................................................................................................ 12
Maximum temperature (HI T) ......................................................................................................................... 12
Minimum temperature (LO T) ......................................................................................................................... 12
Maximum speed (HI R) ................................................................................................................................... 12
Minimum speed (LO R) ................................................................................................................................... 12
Fluid type (FLUI) ............................................................................................................................................. 12
Temperature control type (AUTO) ................................................................................................................... 13
Alarm and button tone (BEEP)
Calibration (CALI) ........................................................................................................................................... 13
Interface and output .............................................................................................................................. 14
Maintenance and cleaning ..................................................................................................................... 16
Error codes ............................................................................................................................................. 17
Accessories ............................................................................................................................................. 17
Technical data ........................................................................................................................................ 18
Warranty ................................................................................................................................................ 19
Pump performance curve ....................................................................................................................... 19
....................................................................................................................................................... 6
......................................................................................................................... 13
3
Page 4
Declaration of Conformity
We declare under our sole responsibility that this product corresponds to the regulations 2014/35/EU, 2006/42/EC, 2004/108/EC and 2011/65/EU and conforms with the standards or other normative documents: EN 61010-1, EN 61010-2-010, EN 61326-1, EN 60529, EN ISO 12100 and DIN 12876-1.
Explication of warning symbols
Indicates an (extremly) hazardous situation, which, if not avoided, will result in death,
DANGER
serious injury.
Indicates a hazardous situation, which, if not avoided, can result in death, serious injury.
WARNING
Indicates a potentially hazardous situation, which, if not avoided, can result in injury.
CAUTION
Indicates practices which, if not avoided, can result in equipment damage.
NOTICE
WARNING
WARNING
Indicates the risk of fire or explosion!
Indicates crushing risk to fingers/hand.
Warning: flammable material!
4
Page 5
Safety instructions
General information:
Read the operating instructions completely before
starting up and follow the safety instructions.
• Keep the operating instructions in a place where it can be accessed by everyone.
• Ensure that only trained staff work with the device.
• Follow the safety instructions, guidelines, occupational health, safety and accident prevention regulations.
• Set up the device in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface.
• The device has been constructed to the requirements of EU (European Union) and EFTA (European Free Trade As­sociation) countries.
• Do not operate the device in explosive atmospheres.
• Protect the device and accessories from bumping and im­pacting.
• Check the device and accessories for damage before each use. Do not use damaged components.
• Safe operation is only guaranteed with the accessories de­scribed in the ”Accessories” section.
• The device must only be operated with the original power cable.
• The socket for the power cable must be easily accessible.
• Socket must be earthed (protective ground contact).
• The voltage stated on the type plate must correspond to the power voltage.
• The device can only be disconnected from the power sup­ply by pulling out the power plug or the connector plug.
• Disconnect the power plug before attaching or changing any accessories.
• Disconnect the power plug before cleaning, maintenance and transportation of the unit.
• The device must only be opened by trained specialists,
even during repair. The device must be unplugged from the power supply before opening. Live parts inside the device may still be live for some time after unplugging from the power supply.
NOTICE
safe operation, for example to keep foreign objects and liquids, etc. from getting into the device.
Coverings or parts that can be re­moved from the device must be put back on the device again to ensure
CAUTION
can be easily accessed must be installed in the work area.
• A laboratory circulator refrigerates and circulates fluid ac­cording to specified parameters. This involves hazards due to high and low temperatures and general hazards due to the device of electrical energy. The user safety can not be ensured simply with design requirements on the part of the device. Further hazard sources may arise due to the type of tempering fluid, e.g. by exceeding or undercut­ting certain temperature thresholds or by the breakage of the container and reaction with the carrier fluid. It is not possible to consider all eventualities. They remain largely subject to the judgment and responsibility of the opera­tor. For this reason, it become necessary for user to take other precautionary safety measures.
• When device is used for external circulation, extra precau-
tion must be taken for cold fluid leakage due to damaged hose.
- Use suitable hoses for connection.
- Secure hoses and tubes against slippage and avoid kinks.
- Check hoses, tubes and bath at regular intervals for possible material fatigue (cracks/leaks).
WARNING
• After a power failure during operation, the device may start automatically (depending on operating mode).
WARNING
• Transport the device with care (see section “Moving the device“).
• Do not transport or empty the bath while it is still cold.
• Always empty the bath before moving the unit.
If the power switch is not within reach when device is operating, an EMERGENCY STOP switch that
Do not start up the device if:
- It is damaged or leaking
- Cable (not only supply cable) is damaged.
Because of the heavy weight of the device, at least two person are needed for carrying the device.
• The device may only be used as prescribed and as de­scribed in these operating instructions. This includes op­eration by instructed specialist personnel.
• When using critical or hazardous materials in your pro­cesses, IKA recommends to use additional appropriate measures to ensure safety in the experiment. For ex­ample, users can implement comprehensive monitoring equipment.
• Process pathogenic material only in closed vessels under a suitable fume hood. Please contact IKA application sup­port if you have any question.
NOTICE
Disposal of the device:
• The device must be disposed of in accordance with na­tional or local regulations
Always empty the bath if the unit is unused for long time.
5
Page 6
Fluids:
WARNING
the chemical hazards that may be associated with the bath fluid used. Observe all safety warning for the fluids.
Depending on the bath fluid used and the type of opera­tion or toxic can arise. Ensure suitable extraction.
• Do not use any fluid which may cause dangerous reac­tions during processing. Only use r
• non corr
ecommended bath fluid. Only use non-acid and
oding fluid.
Only use the fluids, which fulfill the requirements for safety device compatibility
, health and
. Be aware of
Refrigerant:
WARNING
WARNING
WARNING
The device is not allowed to be used in an ATEX (Atmosphere EX­losible) zone.
Flammable refrigerant used. Do not use mechanical devices or other means to accelerate the defrosting process, other than those recom­mended by the manufacturer.
Do not damage the refrigerant circuit.
NOTICE
• Continuous monitoring of the filling level of the bath fluid is r
equired.
• To ensure a sucient fluid circulation, the viscosity of the bath fluid must not exceed of 50mm erating temperature.
Untreated tap water is not recommended. It is recom­mended to use distilled water or high purity water (ion ex­changers) and add 0.1 g soda (sodium carbonate Na /liter
, to reduce corrosive properties.
Never operate the device without sucient fluid! Check the fluid level detection at a regular basis (see the section “Filling and draining”).
2
/s at the lowest op-
2CO3
Don’t use following fluids:
)
NOTICE
-
Untreated tap water Acids or bases
-
-
Solutions with halides: chlorides, fluorides, br iodides or sulfur
- Bleach (Sodium Hypochlorite)
-
Solution with chromates or chromium salts
- Glycerine
errous water.
- F
omides,
• Insucient ventilation may result in the formation of ex­plosive mixtur
eas and keep all air vents open.
ar
• Flammable refrigerant used. Only the manufacturer can
open the device and system (substitute device).
The minimum room size is 12m³, so that in the event of
an err
or (exit of the complete refriger no flammable r max. 8 g/m³ ambient air).
es. Only use the device in well ventilated
repair the device, closed refrigeration
ant filling quantity)
efrigerant-air mixtur
e can arise (R 290
Correct use
• Use:
Use RC (Refrigerated Circulators) for cooling and circulat­ing fluids.
• Range of use (indoor use only):
- Laboratories -
- Pharmacies -
Schools Universities
This instrument is suitable for use in all areas except:
- Residential areas
-
Areas that are connected directly to a low-voltage supply
network that also supplies r
The safety of the user cannot be guaranteed:
-
If the device is operated with accessories that are not sup­plied or recommended by IKA .
-
If the device is operated improperly or in contr
IKA specifications.
-
If the device or the printed circuit board ar thir
d parties.
esidential areas.
ary to the
e modified by
6
Page 7
Unpacking
• Unpacking:
- Please unpack the device carefully
- Any damage should be notified immediately to the shipping agent (post oce, railway network or logistics company).
Preparations
• Setting up:
- Place the unit on an even, stable, clean, nonslip, dry and fireproof surface.
- Keep at least 20 cm of open space on the front and rear side.
- The place for installation should be large enough and provide sucient air ventilation to ensure the room does not warm up excessively because of the heat from device radiates to the environment.
- Do not set up the device in the immediate vicinity of heat sources and do not expose to sun light.
- Cooling machine, pump motor and electronics produce intrinsic heat that is dissipated via the venting grids! Never cover these venting grids!
Note: After setting up the device, wait at least one hour before starting the operation to avoid the damage to the cooling system.
• Connecting the tubings:
- Unscrew the union nuts and stoppers using a wrench (SW 19)from the pump connection IN (8) and OUT (9).
Stopper
• Delivery scope:
- RC 5 basic
- Power cables
-
Hose olive NW 8 (2 pieces)
-
Hose olive NW 12 (2 pieces)
- Plastic cap (for “Overflow” connector)
- Plastic cap (for “Backflow” connector)
- USB cable 2.0 A - micro B
-
User guide
- Warranty card.
- Via a hose to the “Overflow” connection, overflowing flu­id can be directed into a suitable vessel. The vessel should be positioned lower than the “Overflow” connection. In case of closed external system, if there is in any case no risk of exceeding the maximum fluid level, you can close the “Overflow” connector with the included overflow cap. This minimizes the energy loss and pro­tects, in case of long-term applications with low tem­peratures and high humidity, against unwanted water input and the associated volume increase. Check fluid heat expansion!
- Connect the “Backflow” connector to IKA with a suitable hose. When IKA calorimeter is not connected, please close the “Backflow” connector with included backflow cap.
Olives
Union nut
Fig. 3
calorimeter
Union nut
Fig. 2
- Connect the hoses for circulating the external system to the pump connectors M 16 x 1 for IN and OUT directly or with the olives.
- Screw the hose olive to the pump connection IN and OUT with union nuts. Slide the hoses (NW 12) onto the olives. The hoses must be secured with suitable clamps.
External
system
IKA®
Calorimeter
Fig. 4
Note: When the external system is not necessary, please
seal the pump connection IN and OUT with the existing union nuts and stoppers.
7
Page 8
• Filling and draining:
- Before filling the fluid into the bath, open the venting grid as indicated in following figure.
Low Level
High Level
Fig. 5
- Check and make sure that the drain valve is closed (ro­tate clockwise to the stop position, see Fig. 6).
Note: Please note information in section “Commissioning”.
- Connect the power plug and turn on the device with power switch (10).
- The low level warning message appear on the display. Meanwhile, the bottom LED segment of the lights up in red color to indicate the low fluid level
- Open the filling opening lid (1) and remove the stopper from the filling opening. Then, fill fluid to the bath.
(see Fig. 7)
Fig. 6
.
Red (High Level)
Red (Low Level)
Fig. 7
- To drain the fluid from the bath, connect a hose to the drain port and turn the drain valve in counter clock­wise direction with a straight screwdriver.
Drain valve
Drain port
Note: Pay attention to the fluid level information:
Fig. 8
Warning: Do not empty the bath while the fluid is still
cold, there is a risk of freezing.
8
Page 9
• Fluids (Standard information for IKA fluid):
No.
IKA
Designation
CF.EG28.N10.80.8
0
CF.EG39.N20.80.16
1
CF.EG44.N25.80.19
2
CF.EG48.N30.80.22
3
UF.Si.N30.150.10LV
4
HF.Si.20.200.50
5
HF.Si.20.250.50A
6
7
Water o
8
Customized
Operating temperature
range for open bath
application
(°C)
-10 ... 80 -10 … 80 90 115
-20 … 80 -20 … 80 90 115
-25 … 80 -25 … 80 90 115
-30 … 80 -30 … 80 90 115
-30 … 130 -30 … 150 20 … 200 20 … 200 255 >280 20 … 200 20 … 250 255 >280
5 … 95 5 … 95 - -
p
Operating temperature
range for closed bath
applications
(°C)
Safety temperature
(°C)
145 n
The listed fluid temperature range may deviate from the device operating temperature ranges and settings. Nomenclaturefor IKA fluids:
CF.EG28.N10.80.8 --
(5) Addition information
mm²/s
(4) Viscosity (8
) (3) Temperature range (-10 – 80 °C) (2) Chemical composition (Ethylene Glycol 28%) (1) Classification (Cooling fluid)
(1) Classification:
HF: Heating Fluid CF: Cooling Fluid UF: Universal Fluid
Flash point
(°C)
>170
(2) Chemical composition:
Si: Silicone oil EG: Ethylene Glycol
(3) Temperature range: (
Minimum temperature. Maximum temperature)
N: Negative Temperature
(4) Viscosiy:
Viscosiy at 25 °C for Heating Fluid (HF) Viscosity at -20 °C for Cooling Fluid (CF) Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa•s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the fluid divided by 1000.
(5) Additional information:
A: Oil Additives LV: Low Viscosity
n
Note: for open bath application!
o
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to cor­rosive properties of these media. High purity water and distillates are suitable as a medium after adding 0.1 g soda (Na
p Note: The temperature limit values are adjustable in accordance with the fluid used.
, sodium carbonate) per liter of water.
2CO3
9
Page 10
• Moving the device:
Empty all fluid in the bath before moving device from one place to other place. The device must be lifted by two persons with the upper handles. It can also be moved on flat surface by lifting and push­ing the front of the device. It is easy to move the device with the help of the wheels. The angle of inclination should never be more than 10 º at any direction when move the device!
Note: The device must not be moved during operation. After moving the device, you must wait at least one hour before restarting the unit.
Operator panel and display
A
O
Fig. 10
Fig. 9
EDGBC
LJK
max. 10˚
F
H
IMN
Item Designation Function A On/Off button: Switch on/off the circulator.
B LED, set: The LED lights up at the same time as the set value is displayed. C LED, temperature sensor: Indicate that an external temperature sensor is connected. D LED, cooling: The light LED indicate the refrigerating function is activated. E LED display: Display the settings and actual values. F “Menu” button: Press it once: menu option is display.
Press it once again: back to working screen.
G LED, lock: Indicate the buttons are deactivated. H Lock button: Lock/unlock buttons. I “OK/Pump” button: Start/stop the pump function.
Confirm the menu options. J “Pump (+)” button: Increase the pump speed setting. Change the menu value setting. K “Pump (-)” button: Decrease the pump speed setting. Change the menu value setting.
L “Temp” button: Start/stop the refrigerating function. M “Temp (+)” button: Increase the temperature setting.
Navigation, selecting the settings in the menu. N “Temp (-)” button: Decrease the temperature setting. Navigation, selecting the settings in the menu. O Fluid level indicator: Indicate the fluid level (when the fluid level is too low or too high, the bottom
or the top LED segment change into red color).
10
Page 11
Commissioning and operation
Note: Before commissioning, make sure that the device has not been moved for one hour.
Check whether the voltage specified on the type plate matches the power voltage available.
The power socket used must be earthed (protective earthing).
If above conditions are met, the instrument is ready for operation after plugging in the power plug. If these conditions are not met, safe operation is not guar­anteed and the device could be damaged.
Observe the ambient conditions (temperature, humidity, etc.) listed underTechnical Data”.
After switching on the power switch (10) at the back of the device or pressing the On/Off button (A), all LED seg­ments light up during the self test.
c
Display field
d
Software version (ver)
e
Operating mode (mode)
f
Then the device enters standby status and is ready for operation.
Change the temperature setting with “Temp (+)“ button (M) or “Temp (-)“ button (N). Change the pump speed setting with “Pump (+)“ button (J) or “Pump (-)“ button (K).
In standby status, activate the refrigerating function by pressing the “Temp“ button (L). The pump and fan starts to run at once. The refrigerating function may delay to start about 2 minutes.
In working status, press the “OK/Pump“ button (I) to stop the pump and refrigerating function.
Note: In running status, the pump speed adjusts auto­matically according to the load. When the actual speed doesn’t reach the setting speed, the blinking set speed and the actual speed appears alternately.
In standby status, press the “OK/Pump“ button (I) to start the pump function. The refrigerating function will not be activated. In working status, press the “Temp“ button (L) to stop the refrigerating function, the pump keep running.
Pressing the lock button (H) for about 2 seconds, the de­vice control elements can be locked to prevent accidental changes that can be done during operation (LED (G) will light up). To unlock the control elements, press the lock button (H) again for about 2 seconds (LED (G) goes out).
Working setting
After that, following two screens appear alternately.
g
h
11
Page 12
• Menu structure:
MODE
Menu settings
A B C
Default settings
Activated
-
-
Maximum temperature (HI T) Minimum temperature (LO T) Maximum speed (HI R) Minimum speed (LO R) Fluid type (FLUI)
Menu
The type of temperature PID control arithmetic (AUTO)
Alarm and button tone (BEEP)
Calibration (CALI)
Reset to factory settings (RST)
Basic guide to setting the menu options:
) Enter the menu by pressing ”Menu” button (F). ) Press the
ton
)
Press the (K)
) Confirm the menu settings by pressing the
button (I)
Operating mode (MODE):
Temp (+)“ button (M) or “Temp (-)“ but-
(N)
to change the menu options.
Pump (+)“ button (J) or “Pump (-)“ button
to change the menu value settings.
OK/Pump“
.
AUTO 1 AUTO 0
BEEP 0 BEEP 1
CALI 0 CALI 2 CALI 3
No Yes
Minimum temperature (LO T):
The minimum adjustable value: -30 °C.
Maximum speed (HI R):
Proportional coecient of PID (Kp 1) Integrate time of PID (Ti 1) Differential time of PID (Td 1)
The maximum adjustable value: 4600 rpm.
40 °C
-30 °C 4600 rpm 1000 rpm 8
Activated
80.0
10.0
2.0
­Activated
Activated
-
-
Activated
-
Mode A: After power on/power failure no automatic restart of functions.
Mode B: After power on/power failure automatic re­start of functions, depending on previous settings.
Mode
C:
Set values (set in A or B) cannot be changed. After power on/power failure automatic restart of func­tions, depending on previous settings.
Maximum temperature (HIT):
The maximum adjustable value: 40 °C.
Minimum speed (LO R):
The minimum adjustable value: 1000 rpm.
Fluid type (FLUI):
The chosen fluid (No.) limits the maximum and mini­mum temperature. Customized fluid (No.: 8) enables the whole working temperature range. See Section “Fluids (Standard information for IKA fluid)”.
12
Page 13
Temperature control type (AUTO):
Proportional coecient of PID (Kp 1):
AUTO 1: AUTO 1 is the default setting. The default PID settings are used automatically. Cooling down curve in auto-turning control (AUTO 1):
°C
t
Fig. 11
AUTO 0: For special requirements, AUTO 0 can be used with manual adjustment of the control parameters.
If the PID control (AUTO 0) option is selected, the fol­lowing items including Kp 1, Ti 1 and Td 1 could be set. Otherwise they are not displayed in the menu list.
Note: Inappropriate settings may produce the follow­ing cooling down curves:
°C
Kp too high
°C
t
t
The proportional coecient Kp is the controller ampli­fication and determines how strongly the control devia­tion (the difference between the target temperature and actual temperature) directly affects the control variable (cooling down time). Kp values that are too large can lead to the controller overshooting.
Integral time of PID (Ti 1):
The integral time Ti (s) is the correction time and deter­mines how strongly the duration of the control devia­tion affects the control variable. Ti compensates an ex­isting control deviation. A high Ti means a smaller and slower effect on the control variable. Ti values that are too small can lead to instability of the controller.
Differential time of PID (Td 1):
The differential time Td (s) is the derivative time and de­termines how strongly the rate of change of the control deviation affects the control variable. Td compensates for rapid control deviations. A high Td means a smaller and slower effect on the control variable. Td values that are too small can lead to instability of the controller.
Alarm and button tone (BEEP):
°C
°C
Fig. 12
Ti too low
t
Kp too low Ti too high
t
Td too high
BEEP 0: no alarm and button tone BEEP 1: alarm and button tone is activated
Calibration and adjustment (CALI):
CALI 0: Reset calibration CALI 2: 2-point calibration CALI 3: 3-point calibration
13
Page 14
Example: 2-point calibration:
Dip the temperature sensor of the reference measuring instrument into the bath fluid.
Select 2-point calibration in the menu. Press the “OK/ Pump“ button (I) to start the 2-point calibration.
Set the first point temperature (step 1, e.g. 10 °C) with “Pump (+)” button (J) or “Pump (-)” button (K). Con­firm the setting by pressing “OK/Pump“ button (I).
Left area of the display appears the set value (10 °C) and the right area of the display appears the actual tempera­ture value measured. The “Set” LED (B) flashes briefly.
The unit now starts and controls to the set value. When the temperature has reached the set value and has be­come constant, the “Set” LED (B) no longer flashes and the following display appears. The temperature mea­sured by the unit appears on the left hand side.
Input the calibration value from the reference measuring instrument (e.g. 10.1 °C) with “Pump (+)” button (J) or “Pump (-)” button (K).
Confirm the value by pressing the “OK/Pump“ button (I). Back to previous screen for input a new value by press­ing “Temp” button (L).
Back OK
The first point calibration is finished now. Calibration of the other points is performed in the same way.
Reset to factory settings (RST):
Reset (No): Without resetting to factory settings.
Reset (Yes): Resetting to factory settings.
Interface and output
The device can be operated in “Remote” mode via the RS 232 port (5) or the USB port (6) connected to a PC and with the laboratory software Labworldsoft
Note: Please observe the system requirements together with the operating instructions and help section included with the software.
USB interface:
The Universal Serial Bus (USB) is a serial bus for connect­ing the device to the PC. Equipped with USB devices can be connected to a PC during operation (hot plugging). Connected devices and their properties are automatically recognized. The USB interface can also be used to update firmware.
USB device drivers:
First, download the latest driver for IKA devices with USB interface from: http://www.ika.com/ika/lws/download/usb-driver.zip. Install the driver by running the setup file. Then connect the IKA device through the USB data cable to the PC. The data communication is via a virtual COM port. Con­figuration, command syntax and commands of the virtual COM ports are as described in RS 232 interface.
®
.
Device software update: For device software update, visit IKA website www.ika. com and enter the “Service” menu. Download and run
the Firmware Update Tool. Find and click the Firmware Update Tool in your PC after installation. Register your E-mail and password. Connect the device to your PC via the USB cable. Then, you can update the device software according to instruc­tions of the Firmware Update Tool.
RS 232 interface:
Configuration:
- The functions of the interface connections between the stirrer machine and the automation system are chosen from the signals specified in EIA standard RS 232 in ac­cordance with DIN 66 020 Part 1.
-
For the electrical characteristics of the interface and the al­location of signal status, standard RS 232 applies in accor­dance with DIN 66 259 Part 1.
- Transmission procedure: asynchronous character trans­mission in start-stop mode.
- Type of transmission: full duplex.
14
Page 15
- Character format: character representation in accordance with data format in DIN 66 022 for start-stop mode. 1 start bit; 7 character bits; 1 parity bit (even); 1 stop bit.
- Transmission speed: 9600 bit/s.
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to the computer takes place only at the computer’s request.
Command syntax and format:
The following applies to the command set:
- Commands are generally sent from the computer (Master) to the device (Slave).
- The device sends only at the computer’s request. Even fault indications cannot be sent spontaneously from the device to the computer (automation system).
- Commands are transmitted in capital letters.
- Commands and parameters including successive parame­ters are separated by at least one space (Code: hex 0x20).
- Each individual command (incl. parameters and data) and each response are terminated with CR LF (Code: hex 0x0d hex 0x0A) and have a maximum length of 80 characters.
- The decimal separator in a number is a dot (Code: hex 0x2E).
The above details correspond as far as possible to the recommendations of the NAMUR working party (NAMUR recom­mendations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment, rev. 1.1). The NAMUR commands and the additional specific IKA co mmands serve only as low level commands for communication between the device and the PC. With a suitable terminal or communications programme these commands can be transmit-
®
ted directly to the circulator equipment. The IKA software package, Labworldsoft
, provides a convenient tool for control­ling circulating equipment and collecting data under MS Windows, and includes graphical entry features, for pump motor speed ramps for example.
Commands:
Commands
Function
IN_PV_1 Read the external actual temperature IN_PV_2 Read the internal actual temperature IN_PV_4 Read the pump actual speed IN_SP_1 Read the internal setting temperature (if ET=0)
Read the external setting temperature (if ET=1) IN_SP_4 Read the pump setting speed IN_TMODE Read temperature control
0: internal regulation
1: external regulation OUT_SP_1 xxx Set the internal setting temperature XXX (if ET=0)
Set the external setting temperature XXX (if ET=1) OUT_SP_4 xxx Set the pump speed XXX OUT_SP_42@n Set the WD-safety speed with echo of the set (defined) value. OUT_TMODE_0 Set to Internal temperature control OUT_TMODE_1 Set to External temperature control OUT_WD1@n Start the watchdog mode 1 and set the time for the watchdog to n (20...1500) seconds.
Echo of the Watchdog time.
During a WD1-event, the refrigerating and pump functions are switched off and message PC 1 is
displayed.
This command needs to be send within the watchdog time. OUT_WD2@n Start the watchdog mode 2 and set the watchdog time to n (20...1500) seconds.
Echo of the watchdog time.
During a WD2-event, the set temperature is changed to the WD safety temperature and the pump
set speed is set to the WD safety speed. The PC 2 warning is displayed.
This command needs to be send within the watchdog time. RESET Reset the PC control and stop the device functions. START_1 Start the tempering function START_4 Start the pump function STOP_1 Stop the tempering function STOP_4 Stop the pump function
15
Page 16
PC 1.1 Cable: This cable is required to connect RS 232 port (5) to a PC.
6 7 8 9
1 2 3 4 5
1
RxD 2
TxD 3
4
GND 5
6
RTS 7 CTS 8
9
Fig. 13
USB 2.0 cable A–micro B: This cable is required to connect USB port (6) to a PC.
USB micro B
Fig. 14
Maintenance and cleaning
To avoid contamination, it is essential to check and change the bath fluid from regularly. If water is used for bath fluid, we recommend to add Wa­ter bath protective media. The protective media stops the growth of algae, bacteria and other microorganisms.
To keep the full cooling performance, the dust filter of the cooler must be checked regularly and cleaned if necessary.
• Switch off the device and disconnect power cable.
• Open the front venting grid.
• Draw out the dust filter.
• Clean condenser dust filter with a vacuum cleaner or wash
the filter in the water and dry the it before assembly.
Note: Don’t touch the condenser surface with hard parts.
Cleaning:
For cleaning disconnect the main plug!
Use only cleaning agents which have been approved by
IKA to clean the device.
Dirt Cleaning agent
Dye Isopropyl alcohol
Construction material Water containing tenside/isopropyl alcohol
Cosmetics Water containing tenside/isopropyl alcohol
Foodstuff Water containing tenside
Fuel Water containing tenside
5 4 3 2 1
1 2 RxD
3 TxD 4 5 GND 6 7 RTS 8 CTS 9
9 8 7 6
PC
A
USB A
For materials which are not listed, please request information from IKA application support. Wear protective gloves while cleaning the instruments. Electrical instruments may not be placed in the cleansing agent for the purpose of cleaning. Do not allow moisture to get into the instrument when cleaning. Before using another than the recommended method for cleaning or decontamination, the user must ascertain with
IKA that this method does not damage the instrument.
Spare parts order:
When ordering spare parts, please give:
- machine type
- serial number, see type plate
-
item and designation of the spare part, see www.ika.com, spare parts diagram and spare parts list
- Software version.
Repair: Please send in instrument for repair only after it has been cleaned and is free from any materials which may constitute a health hazard. For repair, please request the “Decontamination Certifi­cate” from IKA, or download printout of it from the IKA
website www.ika.com. If you require servicing, return the instrument in its original packaging. Storage packaging is not sucient. Please also use suitable transport packaging.
16
Page 17
Error codes
Any malfunctions during operation will be identified by an error message on the display. Proceed as follows in such cases:
)
Switch off device using the main switch at the back of the device
)
Carry out corrective measures
)
Restart device
Error code Err 01 Pump off
Err 02 Pump off
Err 04 Pump off
Err 06 Pump off
Err 15 Pump off
Err 16 Pump off
Err 17 Pump off
Err 19 Pump off
Err 20 Pump off
Effect Cause Solution
Refrigerating off
Refrigerating off
Refrigerating off
Refrigerating off
Refrigerating off
Refrigerating off
Refrigerating off
Refrigerating off
Refrigerating off
No external
Motor over current (rate current) - Reduce pump motor speed
Motor hall signal
Too low liquid level - Check the fluid level and buoyage
Compressor power supply disconnected
Suction temperature error
Discharge temperature error - Restart the device
Fluid temperature too high - Switch the device off and let the fluid cool down
Condenser fan error - Restart the device
temperature sensor - Check the temperature sensor
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
missing - Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
- Restart the device
- Restart the device
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Accessories
Tubing and hoses: LT 5.20 Metal hose (isolated M16 x 1) LT 5.21 PTFE hose (isolated M16 x 1) H.PVC.8 PVC tube (nominal width 8) H.PVC.12 PVC tube (nominal width 12) H.SI.8 Silicone tube (nominal width 8) H.SI.12 Silicone tube (nominal width 12) H.FKM.8 FKM tube (nominal width 8) H.FKM.12 FKM tube (nominal width 12) H.PUR.8 PUR tube (nominal width 8) H.PUR.12 PUR tube (nominal width 12)
See more accessories on www.ika.com.
Tubing insulations: ISO. 8 Insulation (8 mm) ISO.12 Insulation (12 mm)
Additional accessories: PC 1.1 Cable (RS 232) Labworldsoft
®
17
Page 18
Technical data
Operating voltage VAC 230 ± 10 %
100–115 ± 10 % Frequency Hz 50 / 60 Max. input power W 1100 Working temperature range °C - 30 ... RT Operating temperature range (with external heating) Temperature stability – internal temperature control
(according to DIN 12876) Temperature control PID Temperature measurement, absolute accuracy K ± 0.5 Temperature setting Buttons Temperature setting resolution K 0.1 Temperature display LED Temperature display resolution K 0.1 Cooling capacity according to DIN 12876 (at 4600 rpm):
+ 20 °C + 10 °C 0 °C
- 10 °C
- 20 °C
- 30 °C Refrigerant R 290 Refrigerant quantity g 90 Max. refrigerating system pressure bar 21 Pump speed (adjustable) rpm 1000 ... 4600 Max. pump pressure/suction bar 0.61 / 0.45 Max. flow rate (at 0 bar) l/min 31 Bath volume I 5–7 Fluid maximum viscosity mm Low level protection Yes Interface USB, RS 232 Permitted on-time % 100 IP code according to EN 60 529 Protection class Excess voltage category II Contamination level 2 Permitted ambient temperature °C + 5 ... + 32 Permitted ambient humidity % 80 Dimension (W x D x H) mm 310 x 490 x 546 Weight kg 37.5 Operation at a terrestrial altitude m max. 2000
°C - 30 ... + 80 K ± 0.2
W 1400
1200
950
650
450
200
n
2
/s 50
IP 21
I
n Note: Refrigerant must be disposed of in accordance with local and national regulations.
Subject to technical changes!
18
Page 19
Warranty
In accordance with IKA warranty conditions, the warranty period is 24 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our factory, enclosing the delivery invoice and giv-
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or maintenance not carried out in accordance with the instructions in this operating manual.
ing reasons for the claim. You will be liable for freight costs.
Pump performance curve
Pump performance curve measured with water:
(Measurements done according DIN 12876-2 with water at 20 °C; pump in a closed-loop circuit).
700
500
300
100
ΔP [mbar]
-100
-300
-500
Fig. 15
0 5 10 15 20
Q [l/min]
25 30 35
4600 rpm
3200 rpm
2000 rpm
19
Loading...