HP M1705B, M1772A, M1771A, M1770A, PageWriter 300 Service Manual

...
PageWriter 100, 200 and 300 Series Cardiographs M1770A, M1771A, and M1772A Mobile Cart M1705B
Service Manual
+
+33DUW1R0 3ULQWHGLQ86$2FWREHU
October 22, 1998 3:00 pm DRAFT
&RS\ULJKW +HZOHWW3DFNDUG&RPSDQ\ 3ULQWHGLQ86$
Notice
WARNING
0Notice
The information in this document is subj ect to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, includ ing, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in conn ection with the furnishing, performance, or use of this material.
This document contains or refers to proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Hewlett­Packard Company.
Responsibility of the Manufacturer
Hewlett-Packard only considers itself responsible for any effects on safety, reliability and performance of the equipment if all the following are true:
• Assembly operations, extensions, re-adjustments, modifications or repairs are done by persons authorized by Hewlett-Packard.
• The electrical installation of the relevant room complies with the IEC or national require­ments.
• The instrument is used according to the instructions for use presented in this manual.
As with all electronic equipment, radio frequency interference between this cardiograph and any existing RF transmitting or receiving equipment at the installation site, including electrosurgical equipment, should be evaluated carefully and any limitations noted before the equipment is placed in service. Monitoring during electrosurgery should not be attempted and monitoring electrodes should be removed from the patient to preclude the possibility of burns. Radio frequency generation from electrosurgical equipment and close proximity transmitters may seriously degrade cardiograph performance. Hewlett­Packard assumes no liability for failures resulting from RF interference between HP medical electronics and any radio frequency generating equipment at levels exceeding those established by applicable standards.
ii
Notice
CAUTION
CAUTION
Like all electronic devices, this cardiograph is susceptible to electrostatic discharge (ESD). Electrostatic discharge typically occurs when electrostatic energy is transferred to the patient, the electrodes, or the cardiograph. ESD may result in ECG artifact that may appear as narrow spikes on the cardiograph display or on the printed report. When ESD
occurs, the cardiograph’s ECG interpretation may be inconsistent with the physician’s interpretation.
ESD discharges to exposed metal on the rear of the cardiograph can occasionally cause an error message to appear on the cardiograph display. The cardiograph returns to normal operation after turning the power off, then on again.
The data transmission cable must have a suppression device attached to assure the cardiograph’s compliance with the European Radiated Emissions Standard found in CISPR 11. If your data transmission cable does not include a suppression device, compliance can be achieved by attaching one of the following suppression devices to the cable, near the cardiograph:
part number 0443164251 Fair-Rite Products Corporat ion
P. O. Box J One Commercial Row Wallkill, Ne w York 12589 telephone: (914) 895-2055 FAX: (914)895-2629
or Euro-Schaffner, S. A.
1 B Avenue de Suisse - BP 16 68311 Illzach Cedex, France telephone: 33-8-931-0400 FAX: 33-8-931-0401
part number 28B2025-0A0 Steward
East 36th Street P. O. Box 510 Chattanooga, TN 37401 telephone: (615) 867-4100 FAX: (615) 867-4102
or Steward EMC, S.A.
Rue Fritz-Couvoisier 40 Ch-2300 La Chaux-de-Fonds Switzerland telephone: 41-39-282-387 Fax: 41-39-280-277
This is to certify that this equipment is in accordance with the Radio Interference Requirements of the EMC Directive.
iii
Notice
Warranty
Hewlett-Packard warrants this medical product against defects in materials and workmanship for a period of three years in certain geographics, or one year with onsite support.
If Hewlett-Packard receives notice of such defects during the warranty period, Hewlett­Packard shall, at its option, either repair or replace hardware products which prove to be defective.
Hewlett-Packard software and firmware products that are designated by Hewlett-Packard for use with a hardware product, when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defect s during the warranty period. Hewlett-Packard shall repair or replace software media and firmware that do not execute their programming instructions due to such defects. Hewlett-Packard does not warrant that the operation of the software, firmware, or hardware shall be uninterrupted or error free.
If Hewlett-Packard is unable, within a reasonable time, to repair or replace any product to a condition as warranted, Buyer shall be entitled to a refund of the purchase upon return of the product to Hewlett-Packard.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from any of the following:
1. Improper or inadequate maintenance by Buyer.
2. Buyer-supplied software or interfacing.
3. Unauthorized modification or misuse.
4. Operation outside of the environmental specifications for the product.
5. Improper site preparation and maintenance.
THE WARRANTY SET FORTH ABOVE IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
iv
Notice
Printing History
October 1994 Edition 1 May 1996 Edition 2 October 1998 Edition 3
v
Safety Summary
0Safety Summary
Safety Symbols Marked on the Cardiograph
The following symbols are used on the cardiograph or the cart:
Caution - See operating instructions
Alternating current.
Meets IEC type CF leakage current requirements and is defibrillator protected. (Isolated ECG input.)
Equipotential (identifies independent protective earth conductor to the cardiograph).
Fuse.
Indicates power control for cardiograph.
Hz Indicates operating frequency in cycles per second.
The maximum weight that the cart can hold
Displays the configuration menu on the PageWriter 200/200/300pi.
Please see Chapter 2, “Performance Verification and Maintenance,” for safety requirements that apply to the cardiograph.
vi
Conventions Used in This Manual
WARNING
CAUTION
NOTE
0Conventions Used in This Manual
Warning statements describe conditions or actions that can result in personal injury or loss of life.
Caution statements describe conditions or actions that can result in damage to the equipment or software.
Notes contain additional information on cardiograph usage.
TEXT represents the labels that appear on the display.
.H\
Softkey
represents keys on the key panel.
represents the temporary key labels that appear on the display.
vii
Preface
0Preface
This manual contains service information for the Hewlett-Packard M1 770A pageWriter 300pi, PageWriter 200i, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs. The information and procedures in this manual apply to all models unless otherwise specified.
This manual is organized as follows:
Chapter 1 — I n troduction. Contains a general description of the cardiographs, lists of technical specifications, and lists of accessories and options.
Chapter 2 — Performance Verification and Maintenance. Explains how to check the cardiograph’s performance using built-in self-tests, and lists maintenance procedures and safety requirements that apply to the cardiograph.
Chapter 3 — Theory of Operation. Provides an overview of how the cardiograph
works and describes the operation of the m a jor subassemblies.
Chapter 4 — Troubleshooting. Contains procedures and error codes to aid the service person in localizing faults to a replaceable subassembly.
Chapter 5 — Removal and Replacement. Contains procedures for removing and replacing each of the cardiograph’s major subassemblies.
Chapter 6 — Parts List. Lists part numbers for the cardiog raph’s replaceable parts, an d provides assembly drawings.
Appendix A — Connector Pin Assign ment s. Identifies and defines the signals
assigned to the subassembly interconnections.
Index.
viii
Contents
Notice. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Limitation of Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Safety Symbols Marked on the Cardiograph. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Conventions Used in This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vii
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
1. Introduction
PageWriter 100, 200, 200i, and 300pi Series Cardiographs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
M1705B Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Transmission and Storage, Option #A05 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Inquiries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Specification Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Sampling Characteristics of Cardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
PageWriter 100, 200, 200i, and 300pi Cardiographs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Country/Region Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
2. Performance Verification and Maintenance
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Visual Inspection and Power On Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Extended Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Internal Circuitry Testing of PageWriter 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Internal Circuitry Testing of PageWriter 200/200i/300pi . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Memory Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
How to Read the Extended Self-Test Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Display Test (PageWriter 200/200i/300pi only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Indicator Light Test (PageWriter 100 only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Printer Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Keyboard Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -13
PageWriter 200/200i/300pi Keyboard Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
PageWriter 100 Keyboard Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
ECG Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Changing the Default Operating Language (200/200i/300pi Only) . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Resetting the Cardiograph to the Factory Default State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Resetting the PageWriter 200/200i/300pi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Resetting the PageWriter 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Cleaning the Cardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Cleaning the Keyboard Overlay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -19
Cleaning the Digital Array Printhead and Paper Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
ix
Cleaning the Electrodes and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Caring for the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Storing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix . . . . . . . . . . . . . . . . 2-24
3. Theory of Operation
Operational Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
ECG Data Path. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Power-on and Power-off Sequences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Power-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Power-off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Circuit Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
The Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
CPU Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
CPU/System Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
System Gate Array/Real-Time Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
ECG Front End Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Interrupt Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
DMA for ECG Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
DMA for LCD Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
System Reset Circuitry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Watch-dog Circuitry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Real-Time Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
DRAM RAS and CAS Generation and Refresh . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
NVRAM Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Keyboard Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Display Control for PageWriter 20 and 300 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
NVRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
ECG Front End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
DRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
ROM-1 and ROM-2 Address Space . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Printer Gate Array/SRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Motor Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Analog/Digital Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Thermal Printhead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
System Expansion Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Voltage Selector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Primary Line Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Power Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Rectifier/Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Switch-Mode Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
VPRINT Boost Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Switched +6V Boost Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
5V Linear Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
ECG Front End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Isolation Power Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Front End Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Opto-Isolators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
x
Integrated Front End Circuits/Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Defibrillator Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Storage and Transmission (Option #A05 0nly) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
M177XA Expansion Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Gate Array . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
ROM-X1, ROM-X2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Flash1, Flash2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
VPROG Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Dual UART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
18.432 MHz Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 -15
RS232 Drivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Serial Port Connectors 1 and 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
4. Troubleshooting
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Maintenance Philosophy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Test Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Patient Cable Test Tool (M1770-87908) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Front End and RS-232 Port Test Tool (M1770-87909, Rev. B) . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Testing the Instrument Signal Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Testing the Cardiograph’s RS-232 Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
The Error and Event Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Using Extended Self-test in Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
How the PageWriter 100 Communicates Error Codes and Messages . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
PageWriter 100 Error Code Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
PageWriter 100 Error Message Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Troubleshooting Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Testing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
5. Removal and Replacement
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Tool Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
The Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Removing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
The Keyboard Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Removing the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Removing the Keyboard Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Removing the Keyboard Flexible Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Removing the Keyboard Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
Replacing the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
The Top Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Opening and Removing the Top Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Replacing the Top Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
The Printer Door Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Removing the Printer Door Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Replacing the Printer Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
The Printer Drive Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Removing the Printer Drive Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Replacing the Printer Drive Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
xi
The Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Removing the Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Removing the Optical Paper Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Replacing the Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
The Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Removing the Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Replacing the Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
The Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Removing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Replacing the Power Supply Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
The Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
The Internal Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
Replacing the Power Supply Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
Replacing the Capacitor Board Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
Replacing the Printer Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
The Storage and Transmission Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
Removing the Storage and Transmission Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
6. Parts Lists
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Calling for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
7. Connector Pin Assignments
CPU Assembly Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
xii
Introduction
WARNING
NOTE
This chapter introduces you to the M1770A PageWriter 200i/300pi, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs and lists their technical specifications.
These cardiographs are only to be serviced by HP qualified personnel.
Safe and effective use of medical instrumentatio n requires periodic inspection and preventive maintenance. Perform the preventive maintenance procedures in Chapter 2 of
this manual at required intervals to ensure satisfactory instrument performance.
The cardiographs use a thermal printhead to record waveforms and label the ECG report. The paper supplied with the cardiographs is a thermal paper designed to work with this printhead and with the photo detector used to advancing the paper.
Hewlett-Packard guarantees the performance of the cardiographs only when used with Hewlett-Packard supplies, accessories, and paper that meet or exceed Hewlett-Packard specifications.
0PageWriter 100, 200, 200i, and 300pi Series Cardiographs
The M1770A PageWriter 200i/300pi is Hewlett-Packard’s economical, interpretive
cardiograph. The M1771A PageWriter 200 is the economical, non-interpretive cardiograph. The M1772A PageWriter 100 is the most economical, non-interpretive cardiograph. The cardiogr aph con tains the user controls, the p rinter, and all the pr ocessing circuitry. All models use the same enclosure. The M1705B cart is designed for these cardiographs. Figure 1-1 shows the M1770A mounted on the cart. Option #A05 provides transmission and storage capability for the M1770A and M1771A. The StressWriter option provides capability for communication between a PageWriter 100/200/200i/300pi cardiograph and a StressWrit er system.
1-1
PageWriter 100, 200, 200i, and 300pi Series Cardiographs
Figure 1-1 M1770A/M1771A PageWriter 300pi/200i/200 Cardiograph.
The PageWriter 100, 200, and 300 family of cardiographs print ECGs in Auto and Manual formats. All PageWriter cardiographs use a continuous feed, high-resolution, thermal array printer. This produces electrocardiograms on standard- size paper that can be included in patient records without cutting or mounting. ECG reports clearly show the ECG waveforms. The PageWriter 200 and 300 series cardiographs also annotate records with patient information and ECG measurements. The Pa geWriter 200i and 300pi also includes ECG interpretation for both adult and pediatric patients.
From a service standpoint, the PageWriter 100, 200, and 300 family of cardiographs are similar except for two major hardware differences:
• The M1770A has additional firmware for ECG measurements and interpretation.
• The M1771A has additional firmware for ECG measurements.
• The M1772A keypanel does not have an LCD preview dis play and uses operation al keys instead of alphanumeric keys for user input.
All other model differences are found in the cardiograph’s software.
The PageWriter 100, 200, and 300 family of cardiographs is designed for low-cost, long­term reliability. The modular design makes extensive use of VLSI and gate array technology, resulting in a minimum number of subassemblies. The modular approach means less down-time for the user, since replacing subassemblies allows faster field repairs. The internal self-test efficiently identifies faulty subassemblies , further speeding the repair process.
1-2
Inquiries
M1705B Cart
The optional M1705B Cart provides mobility for the PageWriter 100, 200, and 300 family of cardiographs. The large wheels make the cart easy to move, yet steady. Slots in the
cardiograph’s feet and a thumbscrew secure the cardiograph to the cart rails. The tray directly below the cardiograph provides stor age for the user’s gu ide. Storage for the power cable is built in. Other built-in cable retainers hold the patient data cable ou t of the way. Two compartmented trays provide storage for spare patient electrodes and patient cables, additional thermal paper, and consumable supplies.
Transmission and Storage, Option #A05
Transmission and storage features are available for the M1770A and M1771A with purchase of Option #A05. Up to thirty Auto ECGs can be stored in the cardiograph’s internal flash memory. Stored ECGs can be recalled later for editing, re-analyzing, printing, or transmission to another PageWriter 200/200i/300pi equip ped with Option #A05, a PageWriter XLi cardiograph, a TraceMaster ECG Management System, or facsimile machine. Manual ECGs cannot be transmitted or stored.
1Inquiries
Refer any questions or comments regarding these instruments to the nearest Hewlett­Packard Sales/Service Office or to one of Hewlett-Packard’s Service Dispatch Centers. Always identify the instrument by model number and serial number in all correspond ence. Telephone numbers for Service Dispatch Centers and Sales/Service Offices are listed in Chapter 6, “Parts List.”
1Specification Data
The following tables list the technical specifications for the cardiographs and the mobile cart. Specifications are the same for all models except as noted
Table 1-1 Physical Specifications
Parameter Specification
Dimensions (h × w × l)
× 15 in × 17 in (11. 5 cm × 39.2 cm × 43.7 cm)
cardiograph cart
Weight cardiograph
cart
4 in
× 17 in × 33 in (91.4 cm × 43.5 cm × 84.8 cm)
36 in
18.7 lbs. (8.5 kg) or l es s (includes battery , 200 sheets of paper, pa ti ent cable, and power cord)
38 lbs. (17.24 kg)
Chemical resistance, cleaners
Withstands the following: isopro py l alcohol (except patient cabl e ), mild soap and water, chlorine bleach and wate r (30 ml/l of water).
1-3
Specification Data
Table 1-2 Electrical Specifications
Parameter Specification
Resolution ECG (internal) display
5 µV 128 row pixels by 24 0 column pixels
display sweep sp eed Input impedance greater than 2.5 M@ 10 Hz, typically greater than 100MΩ @ DC
Gain accuracy
Input bias less than 40 nA for input leadwire, les s t han 500 nA for Right L eg output
Common mode rejection
Defibrillator reco ve ry
Crosstalk rejection
Sample rate 4.096M samples per second per input le adwire (unfiltere d) DC offset tolerance
Noise
Standardizing voltage
50/60 Hz notch filter 50/60 Hz AC line rejectio n filters are always active. Pacemaker pulse display > 0.2 mV indication for pa ce pulses between 0. 5 and 2 ms dur ation at amplit ude
1
Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devi ces.
2
Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Tested according to AAMI
23.1 mm/sec
±5% of input signal or ±40 µV, whichever is greater
leadwire
110 dB or greater with A A MI test circuit
System recovery 8 seconds after 360 joule discharge, no damage
Less than 2% chan nel crosstalk
≥ ±300 mVdc with less tha n 5% gain change 30 µV peak-to-peak RTI
1 mV
2 to 250 mV
± 1%
1
1
1
± 5% for 100 mSec ± 5 mSec
1
1
2
1
1
1
1
EC11-1991 test method 4.2.7.1.
Sampling Characteristics of Cardiograph
Physiological factors such as breathing can cause variation in amplitudes of heart beats, independent of the heart rate. In addition, sampled systems (as opposed to continuous systems) may show changes in the apparent height of the R-wave when sampling of the R­wave occurs slightly off-peak. Because the sampling rate and the heart rate are asynchronous, the time between the peak and an adjacent sample can vary from one QRS complex to the next. This results in a slight variation o f dis playe d QRS amplitude. This effect is more pronounced wi th narrower si gnals, more commonly found in some pediat ric electrocardiograms and in pacemaker pulses. The M177XA family of cardiographs minimizes this effect by:
• using an integrating-type A/D converter
• sampling all leadwires simultaneously
• oversampling the signals
• processing the oversampled si gnals with appropriate digital si gnal process ing techniques
1-4
Specification Data
Table 1-3 ECG to Paper Specifications
Parameter Specification
Frequency response of PageWriter
High Pass Low Pass 200/200i/300pi (-3dB bandwidth)
Auto ECG
Manual ECG
0.05 Hz
0.15 Hz
0.5 Hz
0.05 Hz
0.15 Hz
0.5 Hz
1 1
3
150 Hz
100 Hz
40 Hz
150 Hz
100 Hz
40 Hz
1 2
3
2
Frequency response of PageWriter 100 Auto ECG Manual ECG
1
3
<3 dB down at (u ser selectable):
3 dB down at (user selectable):
<
0.15-150 Hz or 0.15-40 Hz
0.05-150 Hz or 0.05-40 Hz
ECG resolution on paper voltage axis time axis
Approximately 8 dots/mm (200 dots/in) Approximately 20 dots/mm (500 dots/in) at 25mm/sec and slower Approximately 10 dots/mm (250 dots/in) at 50mm/sec
Recorder speed 5, 10, 25, and 50 mm/sec +
1.5%
4
ECG visibility with pace pulse Meets or exceed AMMI EC11-1991 standard for Diagnostic Electrocardio-
graphic Devices
1
Bold entries mean meets or exceeds AAMI E C 11-1991 standard for Diagnostic Electrocardiographic Devices. Meets
frequency response standard 3.2.7.2 using methods A, D, and E with auto report filter settings at 0.05-150 Hz or
0.15-150 Hz.
2
-4 dB @ 100 Hz
3
Bold entries mean meets or exceeds AAMI E C 11-1991 standard for Diagnostic Electrocardiographic Devices. Meets
frequency response standard 3.2.7.2 using methods A, D, and E with manual report filter settings at 0.05-150 Hz.
4
Bold entries mean meets or exceeds AAMI E C 11-1991 standard for Diagnostic Electrocardiographic Devices.
1-5
Specification Data
Table 1-4 Power Supply and Battery Specifications
Parameter Specification
AC line frequenc y 115 V setting 230 V setting AC input power rating
Battery voltage 6 V DC, 6.5 Ah Battery recharge time with unit off
(battery full y discharged) to 90% capacity to full capacity
Battery capacity Auto continuous rhythm (Manual mode)
Battery life AC and battery operation battery only operation
Low battery warning At least 2 Auto ECGs or at least 2 minutes of Man-
Table 1-5 Safety Specifications
Parameter Specification ECG leads source curr ent to ground Less than 10 µA RMS
50 and 60 Hz (nominal)
90–132 Vac 198–264 Vac 50 VA
7 hours typical 16 hours typical
40 ECGs 40 minutes
Typically 24 month s Typically 14–16 months
ual ECG is allowed after the Low Battery indica­tion is given.
1
ECG leads sink current 120 Vac, 60 Hz
240 Vac, 50 Hz
1 2
Ground wire integrity
Shunting of defibrillator ene rg y
Operator safety during defibrillation
1
Meets or exceeds AAMI EC11-1991 st andard for Diagnostic Electrocardiographic
Less than 20 µA RMS with patient cable Less than 50 µA RMS with patient cable
Less than 100 mΩ @ 25 A AC
1
< 10%
< 100 µC
1
1
Devices.
2
Meets or exceeds IEC 601-2-25 International Standard for Medical Electrical Equipment,
Part 2 for Electrocardiographs.
1-6
Specification Data
Table 1-6 Environmental Specifications
Parameter Specification Temperature
operating storage
Humidity operating storage
Pressure (altitude) operating storage
Table 1-7 Miscellaneous Specifications
Parameter Specification Real time clock accuracy
10 to 40
° C
50
° to 104° F (10° to 40° C) ° to 122° F (0° to 50° C)
32
15 to 80% RH, non-condensing 15 to 90% RH, non-condensing
15,000 ft. (4600 m) for 2 hours 15,000 ft. (4600 m)
Less than 3 minutes deviation per month.
Printhead life Typically 100,000 pa ges
Table 1-8 Storage and Transmission Specifications (Option #A05 only)
Parameter Specification
Auto ECG storage capacity 30 ECGs Transmission Protocols
DT
SCP Modem command interfaces
Data modem FAX modem
Modem protocols Modulation Error correction Compression FAX modulation
1
Described in the European Committee for Standardization, Standard Communications
Protocol - Computer - Assisted Electrocardiography. (CEN/TC 251).
Data transmission standard used betwee n HP PageWriter cardio­graphs and HP TraceMaster ECG Management System s
Standard Communications Protocol
Hayes standard AT command set EIA/TIA-578 Serv i ce Cl ass 1
V.34 V.42 V.42 bis V.17
1
1-7
Options and Accessories
1Options and Accessories
PageWriter 100, 200, 200i, and 300pi Cardiographs
These tables list the options and accessories available for the PageWriter 100, 200, and 300 family of cardiographs.
Country/Re gion Options
Each country/region option includes the appropriate power cord, voltage, printer, patient cable, and language. See Table 1-9, which shows th e configuration o f each coun try/region option.
1-8
Options and Accessories
NOTE
Table 1-9 Country/Region Option Configurations
Country/
Option
ABA North America English English A AHA 120 903–US Disposable ABB Europe En gl ish English English A4 IEC 220 902–Europe None ABC French Ca nada French French A AHA 1 20 903–US Disposable ABD Germany German German A4 IEC 220 902–Europe None ABE Spain Spanish Spanish A4 IEC 220 902–Euro pe None
ABF France French French A4 IEC 220 902–Europe None ABG Australia English English A4 AHA 240 901–Australia Welsh ABH Netherlands Dutch Dutch A4 IEC 220 9 02–Europe None ABK Intercon
ABM Latin America Spanish Spanish A AHA 120 903–US Welsh ABN Norwegian Norwe-
ABS Sweden Swedish Swedish A4 IEC 220 902-Europe None ABU United Kingdom English English A4 IEC 240 900 –U K None
Region Labels
English English A AHA 120 903–US Welsh
English, Taiwan
gian
User Manuals Printer
Norwe-
gian
A4 IEC 230 902-Europe None
AHA/
1
IEC Volts Power Cord
2
Electrodes
ABX Finland Finnish Finnish A4 IEC 220 902-Eu rope None ABZ Italy Italian Italian A4 IEC 220 902–Europe None
AB2 China Chinese Chinese A4 IEC 220 922–Chi na Welsh
AB4 Singapore English Englis h A4 AHA 240 900–UK Welsh
AC6 S. Korea English English A AHA 220 902-Europe Welsh
ACJ India English English A4 IEC 240 917-SA Welsh
AKM China Chinese English A4 IEC 220 922-China Welsh
ACQ South Africa English English A4 IEC 220 9 17–SA None AKV South America Spanis h Spanish A A H A 220 902–Europe Welsh
1
In Table 1-9 an “A” in the Printer column refers to 8.5 x 11-in. paper; “A4” refers to 210 x 29 7-mm paper
2
See Table 1-11 for HP part numbers
Switzerland should order the desired language (German, French, Italian, or English) and the Swiss power cord (HP 8120-2296).
1-9
Options and Accessories
Table 1-10 PageWriter 300pi, 200/200i and 100 Series Documentation Part Numbers
Document Title Language Part Number
PageWriter 100 Cardiograph User’s Gui de (M1772A)
PageWriter 200/20 0i Car diograph Users’s Guide (M1770A, M1771A with Serial # Prefix CNA, CNB, CNC, 3807A or earlier version)
English M1772-91900
French M1772-91901 German M1772-91902 Dutch M1772-91903 Spanish M17 72-91904 Italian M1772-91905 Finnish M1772-91910 Chinese M1772-91908 English M1772-91900
French M1770-91901 German M1770-91902 Dutch M1770-91903 Spanish M1770-91904 Italian M1770-91905
PageWriter 300p i Ca rdiograph Users’ s G uide (M1770A with Serial # Prefix CND)
Swedish M1770-91906 Finnish M1770-91910 English M1770-91930
French M1770-91931 German M1770-91932 Dutch M1770-91933 Spanish M1770-91934 Italian M1770-91935 Swedish M1770-91936 Norwegian M1770-91938 Finnish M1770-91939
1-10
Options and Accessories
Table 1-10 PageWriter 300pi, 200/200i and 100 Series Documentation Part Numbers
PageWriter 300p i Pred ictive I nstr uments Quic k Reference Guide (M1770A with Seria l # Pre fix CND)
Interpretive Ph ysician’s Guide (M1770A) English M1700-92 908
English M1770-92800
French M1770-92801 German M1770-92802 Dutch M1770-92803 Spanish M1770-92804 Italian M1770-92805 Swedish M1770-92806 Norwegian M1770-92808 Finnish M1770-92809
French M1700-92918 German M1700-92928 Dutch M1700-92938 Spanish M17 00- 92948 Italian M1700-92958
Interpretive Physician’s Guide - Addendum (M1770A)
Predictiv e Instruments Physic ian’s Guid e (M1770A with Serial # Prefix CND)
English M1700-90700
French M1700-90701 German M1700-90702 Dutch M1700-90703 Spanish M1700-90704 Italian M1700-90705 English M1792-93900
French M1792-93901 German M1792-93902 Dutch M1792-93903 Spanish M1792-93904 Italian M1792-93905
1-11
Options and Accessories
Table 1-11 Power Cord Part Numbers
Power Cord Key HP Part Number
900 8120-1703 901 8120-4464 902 8120-1692 903 8120-5213 917 8120-4600 922 8120-8377
Table 1-12 Options
Option Description
A05 Adds storage and transmission capability M2488A-#A70, A71, A72 StressWriter system includes Pa geWriter cardiograph M2488A-#C70 StressWriter system includes interface card for PageWriter
cardiograph
Standard Accessories
Accessories included are based on model number and localization.
• 200 sheets of z-fold paper
-English paper p/n M2481A
-Metric paper p/n M2483A
PageWriter 300pi HP M1770A Cardiograph User’s Guide (PageWriter 300pi)
PageWriter 200/200i HP M1771A/M1770A Cardiograph User’s Guide (PageWriter
200/200i)
Hewlett Packard Interpretive Cardio graph Physician’s Guide (PageWriter 200i/300pi only)
PageWriter 100 HP M1772A Cardiograph User’s Guide (PageWriter 100 only)
Using the HP PageWriter 200/200i Cardiograph Operator Training Video (PageWriter 200/200i, also for use with PageWriter 300pi)
Using the HP PageWriter 100 Cardiograph Operator Training Video (PageWriter 100 only)
1-12
Performance Verification and Maintenance
NOTE
1Introduction
This chapter describes how to verify the cardiograph’s performance, explains preventive maintenance, presents patient safety informatio n, and provides a performance verification checklist.
1Performance Verification
Several procedures make up performance verification: visual inspection of the cardiograph exterior, execution of Extended Self-test, visual examination of an ECG recorded from an ECG patient simulator, and system safety tests using a safety analyzer.
A Performance Verification matrix which appears at the end of this section specifies the tests and inspections which must be performed follo wing servicing of the PageWriter cardiographs. The Performance Verification test results must be recorded on
Customer Service order records
Make copies of the matrix. Fill out a copy each time the cardiograph is tested. Attach the printer test output, and simulator ECG trace to the completed matrix and file with the cardiograph’s permanent maintenance record.
If any of the performance verification tests fail, refer to Chapter 4, Troubleshooting.
2-1
Performance Verification
Visual Inspection and Power On Self Test
Before beginning the inspectio n, press the key to put the cardiograph in
2Q6WDQGE\
Standby mode, and unplug the power cord from the wall outlet. Inspect the cardiograph for the following:
• Worn or damaged power cord
• Loose or missing hardware
• Mechanical damage
• Evidence of liquid spill
• Worn printer drive gear
• Worn printer roller
• Corroded or damaged reusable electrodes, if present
• Damaged patient cable
• Dirt/paper residue on the thermal printhead
• Frayed or damaged wiring
Replace any damaged or missing items, and clean the printhead and patient electrodes as necessary. Cleaning instructions are listed under “Preventive Maintenance” later in this chapter. Connect the cardiograph to AC power. Be sure that the AC indicator comes on. Turn the cardiograph On and observe that the display turns on (PageWriter 200/200i/300pi) or that the LED’s flash on (PageWriter 100).
Extended Self-test
The cardiograph’s extended self-test checks each major subassembly. The extended self­test consists of two sections: the testing of internal circu itry and the testing of printer, display, keyboard, and modem capabilities.
To begin the extended self-test, press and hold both the and keys while turning on the cardiograph with the key.
2Q6WDQGE\
To stop or exit extended self-test, you must press to place the cardiograph
$XWR 0DQXDO
2Q6WDQGE\
in Standby.
Internal Circuitry Testing of PageWriter 100
These tests are performed on the internal circuitry:
•Memory
• Gate Array
• Front-End Interface
• Serial Interface
Each of these tests is described in detail later in this chapter.
When the internal circuitry test of the PageWriter 100 is performed, the LEDs show which test is being performed. The following table shows which LED is associated with each test.
2-2
Performance Verification
Table 2-1 PageWriter 100 Test LED’s
Test Number Test Name LED(s) lit on Page-
Writer 100
1 ROM-1 5 mm/mV OK 2 ROM-2 10 mm/mV Normally fails for 100. 3 ROM-X1 20 mm/mV OK for A#10. 4 ROM-X2 V Leads OK for A#10. 5 RAM Filter OK 6 NVRAM Auto OK 7 GA I II III Lead Group Will also flash all LEDs. OK 8 ECG FE aVR aVL aVF Lead Group OK
9 SIO II aVF V2 Lead Group OK for A#10. A REMOTE V Leads For future use - normally fail s. B MODEM 5 mm/sec OK for A#10, with modem or test tool
C FLASH1 10 mm/sec OK for A#10. D FLASH2 25 mm/sec OK for A#10.
When the tests are completed, the results are also printed out in a pass/fail format on the left-hand side of the first page of the test repor t.
Comments/Expected Results
attached
Table 2-2 Test Results Format
Test Passes Test Fails
Printed Result OK **
In some cases, a test failure halts the test/test printout and an error code indicating the
failure is flashed on the LED’s. See Chapter 4, “Troubleshooting,” for a list of self-test failure codes and failure symptoms. The section “How the PageWriter 100 Communicates Error Codes and Messages” in Chapter 4 describes how to read the LED’s.
2-3
Performance Verification
Internal Circuitry Testing of PageWriter 200/200i/300pi
These tests are performed on the internal circuitry:
•Memory
• Gate Array
• Front-End Interface
• Serial Interface
Each test is described detail later in this chapter.
When testing the internal circuitry of the PageWriter 200/200i/300pi, the numbers 1 through 9 and the characters A through D are displayed one at a time beginning with number 1. Each number represents a particular test. If the test fails, the character X is placed after the test number, for example, 3X. The test descriptions below are listed in the order the tests are performed.
Table 2-3 PageWriter 200/200i/300pi Test Numbers
Test Number Test Name Character Displayed on Page-
Writer 200/200i/300pi
1ROM-1 1 OK 2 RO M-2 2 Passes for 300pi, 200i, and 200 with
3 ROM-X1 3 Passes for Options #A05 an d #A10
4 ROM-X2 4 Passes for Options #A05 an d #A10
5RAM 5 OK 6 NVRAM 6 OK 7GA 7 OK 8ECG FE 8 OK 9 SIO 9 Passes for Options #A05 and #A10
A REMOTE A For future use - norm ally fails B MODEM B Passes for Options #A05 and #A10
Comments
software revision A.05.0 6 or later.
only.
only.
only.
only, with modem or test tool
attached
C FLASH1 C Passes for Options #A05 and #A10
only
D FLASH2 D For future use - normal ly fails
2-4
Performance Verification
CAUTION
If your cardiograph is equipp ed with Option s #A05 or #A10, do not tu rn it off duri ng Test C or D. If you turn off the cardiograph while Test C or D are in process, you could lose stored ECGs, the Log of ECGs Stored, or the Log of ECGs Taken.
When the tests are completed, the results are printed out in a pass/fail format on the left­hand side of the first page of the test report.
Table 2-4 Test Results Format
Test Passes Test Fails
Printed Result OK **
In some cases, a test failure halts the test/test printout and an error code indicating the
failure appears on the display. See Chapter 4, “Troubleshooting,” for a list of self-test failure codes and failure symptoms.
Memory Test
This test looks for failures in the following memory subsystems:
• Read-Only Memory (ROM)
• Random Access Memory (RAM)
• Non-Volatile Random Access Memory (NVRAM)
Read-Only M emory (tests 1-4) . The cardiograph is designed to have up to four ROMs installed, two on the main control board and two on an option board. The Read-Only Memory test reads the contents and performs a 32-bit cyclic redundancy check (CRC) of each of the four ROM address spaces. The test of a ROM address space will fail if the ROM module is faulty or is not present. The four ROM address spaces are identified on the report as
•ROM-1
•ROM-2
•ROM-X1
•ROM-X2
Depending on the serial number and model of the unit you are servicing, the corresponding ROM part numbers vary. Refer to tables 2-5, 2-6, 2-7, and 2-8 for the correct ROM replacement part numbers.
2-5
Performance Verification
Table 2-5 ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial Number
Prefix CNA, CNB, or CNC,3807A or earlier
Test ROM Part Number Language
ROM 1
ROM 2 U206 - Main CPU Board M1770-89521
ROM X1 U9 - Interface Board M2488-17901 ROM X2 U10 - Inte rf ace Board M 2488-17902
U205 - Main CPU Board M1770-89505
M1770-89521 M1770-89522 M1770-89523 M1770-89524 M1770-89525
English French German Dutch Spanish Italian
Table 2-6 ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial Number
Prefix CND
Test ROM Part Number Language
ROM 1 ROM 2
U205 - Main CPU Board M1770-89550
M1770-89551 M1770-89552 M1770-89553 M1770-89554 M1770-89555 M1770-89556 M1770-89558 M1770-89559
English French German Dutch Spanish Italian Swedish Norwegian Finnish
ROM X1 U9 - Interface Board M2488-17901 ROM X2 U10 - Inte rf ace Board M 2488-17902
2-6
Loading...
+ 134 hidden pages