HP 8648A, 8648B, 8648C, 8648D User Manual

Operation and Service Guide
HP
8648A/B/C/D
Signal Generator
SERIAL NUMBERS
This manual applies directly to the following HP 8648 model/serial prefix combinations and below:
8648A
3847A 3847U 3847A 3847U 3847A 3847U 3847A 3847U
8648B BP 8648C
8648D
HP Part No. 08648-90048 Printed in USA April 1999
Supersedes: November 1998
Notice. The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
@
Copyright Hewlett-Packard Company 1996, 1998 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
FLEX and Motorola are trademarks of Motorola, Inc. Windows@ and MS Windows@ are U.S. registered trademarks of Microsoft Corporation. Windows NT@ is a U.S. registered trademark of Microsoft
Corporation.

Certification

Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Hewlett-
Information

Warranty

Regulatory
The regulatory information is in Chapter 4, “Specifications.”
This Hewlett-Packard instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Hewlett-Packard Company will, at its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Hewlett-Packard. Buyer shall prepay shipping charges to Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and firmware designated by Hewlett-Packard for use with an instrument will execute its programming instructions when properly installed on that instrument. Hewlett- Packard does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
. . .
III

Assistance

Product
agreements assistance, contact Ome.
wm.
maintenance
are available for
Refer to
the
agreements
gour
list of
and
other
customer
Hewlett-Rzckard products.
nearest Hewlett-Rxckard Sales
Sales
and
Service Omes on the
assistance
$br
any
and
Service
following
‘able
O-l. Hewlett-Packard Sales and Service
Offices
UNITED
Instrument
Hewlett-Packard Company (800) 403-0801
Headquarters
Hewlett-Packard S.A. Hewlett-Packard France 150, Route du Nant-d’Avril 1 Avenue Du Canada 12 17 Meyrin O/Geneva Switzerland F-91947 Les
(41 22) 780.8111 France
Great Britain
Hewlett-Packard Ltd.
,
~~~~~~~e~;~~~;~i~g
England
(44 734) 696622
Headquarters
Hewlett-Packard Company 3495 Deer Creek Road Palo Alto, California, USA 94304-1316 (415) 857-5027
Support
Center
EUROPEAN FIELD
France
Zone D’Activite De Courtaboeuf 61352 Bad Homburg v.d.H
(33 1) 69 82 60 60
INTERCON FIELD
Australia
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 (61 3) 895-2895 Kirkland, Quebec
Ulis
STATES
Cedex
OPERATIONS
Germany
Hewlett-Packard Hewlett-Packard Strasse
Germany (49 6172) 16-O
OPERATIONS
Canada
Hewlett-Packard (Canada) Ltd.
17500 South Service Road
Trans-Canada Highway
Canada (514) 697-4232
GmbH
H9J
2X8
China
China Hewlett-Packard Company Hewlett-Packard Japan, Ltd. 38
Bei
San Huan Xl Road Shuang Vu Shu Tokyo 192, Japan
Hai
Dian District (81 426) 60-2111 Beijing, China (86 1) 256-6888
lbiwan
Hewlett-Packard Taiwan 8th Floor, H-P Building 337 Fu Hsing North Road Taipei, Taiwan (886 2) 712-0404
Japan
9-l
‘lakakura-Cho,
Hachioji
Singapore
Hewlett-Packard Singapore (Pte.) Ltd.
150 Beach Road #29-00 Gateway West Singapore 0718 (65) 291-9088
V

Safety Notes

The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument.
Caution
Warning

Instrument Markings

Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, would result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
The following markings and caution and warning labels are used on the instrument. Be sure to observe all cautions and warnings.
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the manual.
The CE93 mark shows compliance with the European Community 1993 standards.
A
CE
!
Warning
Warning Warning
Caution
The CSA mark is the Canadian Standards Association safety mark.
“ISMl-A” This is a symbol of an Industrial Scientific and Medical
Group 1 Class A product.
Hazardous voltage always present in this area with instrument
power cord connected to ac line.
Do not remove this screw when removing cover from instrument.
Hazardous Voltage
Hazardous electrical shock. Heat sink is live. Disconnect power supply before servicing.
vi

General Safety Considerations

Warning
Warning
Warning
Caution
No operator serviceable parts inside. Refer servicing to qualified personnel. lb prevent electrical shock, do not remove covers.
If this instrument is used in a manner not specified by Hewlett-Packard Co., the protection provided by the instrument may be impaired.
For continued protection against fire hazard replace line fuse only with same type and rating (3 A 250 V type F). The use of other fuses or material is prohibited.
Always use the three-prong ac power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
vii

How to Use This Guide

This guide uses the
following convention:
Documentation
Description
CFRONT-PANEL
KEY] This represents a key physically located on
the instrument.
Display
This font is used to represent text which appears on the instrument display.
This guide contains the information required to operate, calibrate, and repair the signal generator to the assembly level. Included are the following:
n a quick overview of the signal generator n examples of typical operation n a reference section that describes all operation features n explanations of error messages displayed on the signal generator n installation instructions
w
tables of specifications
n tables of post-repair information and recommended equipment
required
n theory of operation of the signal generator n troubleshooting procedures to identify failed assemblies n disassembly procedures for removal and replacement of assemblies n replaceable part numbers n adjustments required after repair or performance test failure n performance tests to test the instrument to specifications n supplemental verification tests to test some unspecified parameters
of the instrument
VIII
. . .
Contents
1. Operation
Quick Overview ...................
1. Power Key ...................
2. Display .....................
3. Function and Data Keys
4. Increment Set Keys
5. Knobs. .....................
6. MEMORY
7.
Modulation Source
la. Operation Examples
Getting Started ....................
Operation Examples
Setting the RF Output Signal
Setting the Frequency Setting the Amplitude Turn on the RF Output Setting the Modulation
Incrementing or Decrementing the RF Output Signal
Preliminary Steps ..................
Using the Knob ...................
Using the Increment keys
Using the Memory Registers
Saving Instrument Settings in Register Sequences
Selecting the Sequence Saving Settings in Registers Checking the Sequence Checking a Different Sequence
Deleting a Register from the Sequence
Selecting the Sequence Deleting a Register
Renumbering the Registers in a Sequence
Decreasing the Register Number Checking the Sequence
Inserting a Register in a Sequence
Saving a New Register
Offsetting the RF Output from a Reference
Setting the Reference Value Offsetting the RF Output
Turning the Reference Mode Off or On Setting a New Reference Value
Holding the Output Attenuator Range
Set the Amplitude Level Holding the Attenuator Adjusting the Amplitude
....................
................
................
................
.............
...............
.............
...............
...............
...............
...............
..............
..............
..............
............
..............
..........
........
..............
......
..........
..............
..........
..............
.......
.............
..............
.......
..........
.........
..............
...............
..............
. .
...
l-2 l-2 l-2 l-3 l-4 l-4 l-4 l-5
la-l la-2 la-3 la-3 la-3 la-4 la-4 la-5 la-5 la-5 la-6 la-7 la-8
la-8
la-8 la-10 la-10 la-12 la-12 la-12
la-14
la-14
la-15 la-16
la-17
la-18 la-18 la-19 la-19 la-20 la-21 la-21 la-21 la-22
Contents-l
Setting a User Selectable Modulated Frequency and
Waveform (Option
lE2
or
1EP
Only) ........
Setting the Modulation Level ............
Setting the Modulated Waveform ..........
Setting the Modulated Frequency ..........
Signaling a Numeric-Type FLEX Pager (Option
Setting Up Pager Encoding .............
Entering Pager Encoding Settings ..........
Selecting the Format Settings ............
Selecting the Data Rate and Pager Type Settings ...
Selecting the Message Settings ............
Selecting Transmission Repetitions and amplitude . . Selecting the Pager
Capcode
(Address) ........
Selecting the Protocol Settings ...........
Selecting the Roaming Mode Settings ........
Encoding ......................
lb. Operation Reference
Frequency and Amplitude ...............
1. Knob ......................
2. Digit-Select Arrow Keys .............
3. REFSET ....................
Units
4. REF
OkCFF’
: : : : : : : : : : : : : : : : : :
Function .......................
1. FREQUENCY ..................
2. AMPLITUDE ..................
3.FMAM4M
...................
Setting Up the Pager Encoder .............
1. ENCODER. ...................
Setting the Format .................
FLEX/FLEX-TD
..................
Setting the Data Rate and Pager Type .......
Setting the Message ................
Setting the Encoding Mode ............
Entering the Pager
Capcode
(Address) .......
Setting the Protocol ...............
Setting the Roaming Mode ............
Message During Encoding .............
Signaling Examples ................
SSID/NID
POCSAG
Roaming Example ............
......................
Setting the Data Rate and Pager Type .......
Setting the Message ................
Setting the Encoding Mode ............
Entering the Pager
Capcode
(Address) .......
Message During Encoding .............
RESYNC.. ....................
Setting the Encoding Mode ............
Message During
PN15
.......................
Resynchronizing
..........
Setting the Data Rate ...............
Message During Encoding .............
SERVICE
......................
Setting the Data Rate . . . . . . . . . . . . . . . lb-35
1EP
Only)
la-23 la-23 la-24 la-24 la-25
la-26 la-27 la-27 la-28 la-29 la-29
la-30 la-31 1 a-32 la-32
lb-2 lb-2 lb-2 lb-3 lb-3 lb-3 lb-4 lb-4 lb-5 lb-5 lb-6 lb-7 lb-8 lb-9
lb-9 lb-12 lb-14
lb-16 lb-18 lb-19 lb-22 lb-23
lb-25 lb-27 lb-27 lb-29 lb-31 lb-32 lb-32 lb-33 lb-33 lb-33 lb-34 lb-34 lb-34 lb-35
Contents2
Message During Servicing
.............
Pulse Modulation ...................
1. PULSE .....................
Increment Set ....................
1.
INCRSET
START/STOP Encoding PREV and NEXT
....................
..............
.................
Data .........................
1. MHz/dBm kHz/mV Units Conversion ENTER SHIFT
.....................
......................
%/pV rad/dBpV
.................
........
2. Backspace ....................
3. emf .......................
4.f ........................
Instrument Preset ..................
pGim)@
.....................
p?Fmg(DEL)
.....................
HP-IB
........................
1. ADRS ......................
2. LOCAL .....................
Memory .......................
l.SAV .......................
2. REG ......................
3. Register Recall Arrows
4.SEQ..
.....................
..............
5. DEL ......................
Renumbering the Registers
............
Modulation Source ..................
1. MOD ON/OFF ..................
2.
INT400HzINTl
kHz
3. (FREQUENCY) 4. (FREQUENCY/ WAVEFORM)
5. EXTACEXTDC
.................
6.1kHz+EXTDC
Setting the Modulation Level
7. MOD INPUT/OUTPUT
..............
...
................
...........
..............
RF Output ......................
1. RF ON/OFF ...................
2.A’I”l’NHOLD ...................
Vernier Ranges ..................
3. RF OUTPUT ...................
Rear Panel ......................
1. 10 MHz REF INPUT and OUTPUT
2. DISPLAY CONTRAST
3. AUXILIARY INTERFACE
4. Line Voltage Connector
5. HP-IB Connector
6.
TIMEBASE
ADJ and Language Switches
7. External Pulse Input
Remote Interface (Accessory)
...............
.............
.............
.................
...............
.............
.........
......
1. MOD ON/OFF ..................
2. RF ON/OFF ...................
3. Sequence Selection Arrows
4. Register Recall Arrows
Memory Interface (Accessory)
............
..............
.............
lb-35
lb-36
lb-36
lb-38
lb-38
lb-38
lb-38
lb-39
lb-39 lb-39 lb-39 lb-39 lb-40 lb-40 lb-40 lb-41 lb-41 lb-41 lb-45 1 b-45 lb-45 lb-46 lb-47 lb-47 lb-48 lb-48
lb-49 lb-49 lb-50 lb-51
lb-51
lb-51 lb-52
lb-52 lb-53
lb-53
lb-54
lb-54
lb-54
lb-54
lb-55
lb-56
lb-56
lb-56
lb-56
lb-57
lb-57
lb-57
lb-57
lb-58
lb-58
lb-58
lb-58
lb-59 1 b-60
Contents-3
1. POWER
2. Copy Arrow Keys
.....................
................
Making a Copy ..................
3.
BUSY
......................
lc. Operation Messages
Front Panel Operation Messages ............
HP-IB
. Command Errors ...............
HP-IB Execution Errors ................
HP-IB Device-Specific Errors .............
HP-IB Query Errors ..................
Service Messages ...................
2. HP-IB Programming
Background .....................
Programming Guidelines ...............
HP-IB Definition ..................
What is Programmable ...............
HP-IB Address ...................
Error Messages ...................
Programming Language ...............
Query .......................
Advanced Programming ...............
Programming Examples ................
Programming RF Frequency .............
Programming RF Frequency and FM Modulation ...
Querying RF Frequency ...............
Programming RF Amplitude .............
Programming Pulse Modulation (Option Programming Pager Encoder (Option
lE6)
1EP)
......
HP-IB Status Reporting ................
External Modulation Input Level Status .......
Example: Check the Condition of Modulation Input
(High or Low) .................
Example: Generate a Service Request for External
Modulation ..................
Reverse Power Protection Status
Example: Check the condition of’the
RPP : : : : :
Unspecified Power (Amplitude) Entry Status .....
Example: Check the Condition of Unspecified Power
Entry .....................
Pager Encoding Status (Option
1EP
Only) ......
Example: Check the end of message encoding ...
Example: Check the start of each frame ......
SCPI Command Reference ...............
ABORt
Subsystem (Option
1EP
Only) ........
AM Subsystem ...................
CAL Subsystem ...................
DM Subsystem (Option 1EP Only) ..........
FM Subsystem ...................
FREQuency Subsystem ...............
INITiate Subsystem (Option
1EP
Only) ........
OUTPut Subsystem .................
PAGing
Subsystem (Option
1EP
Only) ........
PM Subsystem . . . . . . . . . . . . . . . . . . .
.....
lb-60 lb-60 lb-60 lb-61
lc-1
lc-5 lc-7 lc-7 lc-7 lc-8
2-l
2-2 2-2 2-2
2-2
2-2 2-2 2-2
2-2 2-3 2-3 2-3 2-4 2-4 2-4 2-5
2-12
2-14
2-14
2-15 2-16 2-16 2-17
2-17 2-18 2-18
2-19
2-20 2-21
2-21
2-22 2-22
2-23
2-24
2-24
2-24 2-25
2-37
Contents-4
POWer
Subsystem PULM Subsystem
STATUS SYSTem
Subsystem
Subsystem
TRIGger Subsystem (Option
Changing Parameters While Encoding (Option
.................
..................
.................
.................
1EP
Only)
........
1EP
Using the Buffer Memory for the Arbitrary Messages . . HP-IB Capabilities HP-IB Connector Information HP 8656/57 Compatible Language
Program Code Implementation Receiving the Clear Message Additional Programming Information
..................
.............
...........
...........
.............
........
3. Installation
Unpacking Your Signal Generator Connecting AC Power
Power Requirements
Replacing the Fuse
.................
................
.................
Turning On the Signal Generator Connecting to Other Instruments Storing the Signal Generator Shipping the Signal Generator
...........
...........
...........
.............
.............
4. Specifications
Options Frequency Specifications Internal Reference Oscillator output Spectral Purity Frequency Modulation Phase Modulation Amplitude Modulation Modulation Source Remote Programming
Environmental General Modulation Generator Option 1 E2 Pulse Modulation Option
Pager Encoder/Signaling Option 1 EP
Frequency Frequency Modulation Pager Signaling Modulation Source General
Regulatory Information
IS0 9002 Compliant
Statement of Compliance
Noise Declaration
.......................
...............
.............
........................
....................
................
...................
................
..................
.................
....................
.......................
...........
.............
lE6
..........
.....................
...............
...................
.................
......................
................
.................
..............
..................
only)
2-38 2-38 2-39 2-40 2-40 2-41 2-41 2-43 2-44 2-45 2-46 2-48 2-48
3-l 3-2 3-2 3-3 3-5 3-5 3-5 3-6
4-1 4-2 4-2 4-3 4-4 4-5 4-6 4-7 4-8
4-8 4-9 4-9
4-10 4-11
4-11 4-11
4-12
4-12 4-12 4-12 4-13 4-13 4-13 4-13
Contents-5
5. Service
Shipping Your Instrument Back to Hewlett-Packard . . .
Recommended Test Equipment . . . . . . . . . . . .
Post-Repair . . . . . . . . . . . . . . . . . . . . . .
Safety Notes . . . . . . . . . . . . . . . . . . . . .
5a. Theory of Operation
Introduction
Overview Al Front Panel A2 Power Supply
A3 Motherboard A4 Reference A5 Sig Gen Synth A6 Output (HP 8648A) A6 Output (HP A7 Attenuator (HP 8648A) A10 Frequency Extension (HP Al 1 Attenuator (HP Al2 Reverse Power Protection (HP Al3 Pulse Modulator (HP Al4 Modulation Generator (Option A30 Pager Encoder (HP 8648A Option
.....................
.....................
...................
..................
..................
...................
..................
...............
8648B/C/D)
.............
.............
8648B/C/D)
8648B/C/D)
...........
8648B/C/D)
8648B/C/D
Option
lE2)
.......
1EP)
5b. Troubleshooting Information
Introduction Troubleshooting Checklist
.....................
...............
AC Mains (line) Fuse Removal .............
lb
Remove the Fuse
................
Modulation Test Points and Power Supply LEDs
Power Supply Distribution ..............
Block Diagrams
....................
......
....
lE6)
...
.....
.....
5-l 5-2
5-5
5-8
5a-1 5a-3 5a-3 5a-5 5a-5 5a-6 5a-6 5a-7 5a-7 5a-7 5a-8 5a-8 5a-8 5a-9
5a-9
5a-9
5b-1 5b-2 5b-3 5b-3 5b-4 5b-5
5b-7
Contents-6
5~.
Service Error Messages
6. Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . .
Assembly Replacements . . . . . . . . . . . . . . .
7. Adjustments
Test Equipment ....................
Equipment Setup for Automated Tests Test Point Extender
.................
........ 7-l
Manual Adjustments .................
Internal Reference Oscillator Adjustment
.......
Pager Encoder Timebase Frequency Adjustment
Automated Adjustments ................
AM Level and Distortion
..............
AM Level .....................
Detector Offset Output Level AM Level: FE Predistortion and Detector Offset Prelevel
......................
...................
....................
...................
..........
Output Level: Frequency Extension Calibration
...
...
6-l 6-l
7-l
7-2 7-3 7-4
7-6 7-8 7-9
7-11 7-13 7-14 7-15 7-17 7-18 7-19
AM Modulator Time Base DAC Motherboard Audio Path
DCFM
.......................
Audio Generator HF Power Level Accuracy LF Output Level LF Power Level Accuracy FSK Deviation Filter Path
Service Support Software
Required Test Equipment Installing the Software Running the Service Support Software
Starting the Software Identifying the DUT
...................
...................
..............
..................
..............
..................
..............
...................
.....................
...............
..............
...............
........
...............
...............
Selecting the Performance Test or Adjustments and
the Test. ...................
Defining the Location where the Test Results are
Saved ....................
Running the Tests and Adjustments
........
Reviewing the Test and Adjustment Results Printing the Test and Adjustment Results Exiting the Software
Support Software Administration
Software Configuration
The User Configuration
The Administration Configuration Adding Test Equipment Removing Test Equipment Editing Test Equipment
Adding Device Drivers Removing Device Drivers Adding Test Drivers Removing Test Drivers Adding Datapacks Removing Datapacks
Motherboard Repair Utility
...............
..........
..............
.............
........
..............
.............
..............
..............
.............
................
..............
................
...............
.............
Reading Information from the A3 Motherboard Storing Information in the A3 Motherboard
....
......
...
....
7-20 7-22 7-23 7-25 7-26 7-27 7-30 7-31 7-33 7-34 7-35 7-35
7-36
7-42
7-42
7-43
7-44
7-46 7-47 7-47 7-49 7-49 7-50 7-50 7-50 7-50
7-51 7-54 7-55
7-56
7-59
7-60
7-62
7-63
7-65
7-66 7-68 7-70
8. Performance Tests
Calibration Cycle Required Test Equipment Performance Test Descriptions
FM Accuracy Performance Test
...................
...............
.............
...........
FM Accuracy Performance Test (Option FM Distortion Performance Test AM Accuracy Performance Test
...........
...........
AM Accuracy Performance Test (Option lE2 only) AM Distortion Performance Test
...........
Phase Modulation Distortion Performance Test Residual FM Performance Test Harmonics Performance Test Spurious Performance Test
............
............
.............
lE2
Only)
. .
. .
....
Contents-7
8-l 8-2
8-5 8-6 8-8
8-10
8-12 8-13 8-14
8-15
8-17 8-19 8-20
DC FM Frequency Error Performance Test . . . . . .
RF Level Accuracy Performance Test . . . . . . . . .
Pulse Modulation On/Off Ratio Performance Test
(Option
lE6
Only) . . . . . . . . . . . . . . . .
Pulse Modulation Rise Time Performance Test (Option
lE6
Only) . . . . . . . . . . . . . . . . . . . .
Pager Encoder Timebase Accuracy Performance Test
(Option 1EP Only) . . . . . . . . . . . . . . . .
FSK Deviation Accuracy Performance Test (Option
Only) . . . . . . . . . . . . . . . . . . . . . .
Internal Timebase: Aging Rate Performance Test
(Option
lE5
Only) . . . . . . . . . . . . . . . .
Power Level Accuracy Performance Test (Automated) .
HP 8648A Test Record . . . . . . . . . . . . . . .
HP
8648B
Test Record . . . . . . . . . . . . . . .
HP 8648C Test Record . . . . . . . . . . . . . . .
HP 8648D Test Record . . . . . . . . . . . . . . .
9. Supplemental Verification Tests
Required Test Equipment List . . . . . . . . . . . . .
CW Frequency Accuracy Supplemental Verification Test 9 kHz RF Level Accuracy Supplemental Verification Test
Index
8-21
8-22
8-26
8-27
8-29
1EP
8-30
8-35 8-38 8-41 8-61
8-85
8-109
9-2 9-3 9-8
Contents-l
Figures
2-l. HP 8648 Status Register Model 2-2. Paging Encoding Status 3-l. Replacing the Fuse
...............
.................
3-2. Power Cable and Mains Plug 4-l. Typical Output Power with Option
5a-1.
HP 8648A Simplified Block Diagram
5a-2.
HP
8648B/C/D
5b-1.
Fuse Removal
5b-2.
Location Diagram
5b-3.
Bottom View of Motherboard with Cover Removed . .
5b-4.
HP 8648A Block Diagram
5b-5.
HP
8648A
Simplified Block Diagram
...................
..................
..............
Option
1EP
A30 Pager Encoder Block
...........
............
1EA
.......
.........
.......
Diagram ....................
5b-6.
HP 8648A Option
lE2
Al4 Modulation Generator Block
Diagram ....................
5b-7.
HP
8648B/C/D
Block Diagram (1 of 2) 6-l. HP 8648A Replaceable Parts 6-2. HP 6-3. HP
8648B/C/D
8648B/C/D
Replaceable Parts
Replaceable Parts -
.............
...........
........
All/A12/A13/A14
Detailed View ..................
7-l.
531
Test Point Extender 7-2. Timebase Adjust Switch Location 7-3. Internal Reference Oscillator Adjustment Setup
...............
..........
....
7-4. Pager Encoder Timebase Frequency Adjustment Setup 7-5. Variable Capacitor Location 7-6. AM Level and Distortion Test Setup 1 7-7. AM Level and Distortion Test Setup 2 7-8. Location of
530,
531, and
7-9. AM Level Test Setup 1
7-10. AM Level Test Setup 2
7-11. Location of
531
and
532
7-12. Detector Offset Test Setup 7-13. Output Level Test Setup 7-14. AM Level: FE Test Setup 1 7-15. AM Level: FE Test Setup 2 7-16. Location of
531
and
532
7-17. Predistortion and Detector Offset Test Setup 7-18. Prelevel Test Setup
.................
.............
........
........
532
on the Motherboard
...............
...............
on the Motherboard
.....
..............
..............
.............
.............
on the Motherboard
.....
.....
7-19. Output Level: Frequency Extension Calibration Test
Setup ......................
7-20. AM Modulator Test Setup 7-21. Location of
531
on the Motherboard 7-22. Time Base DAC Test Setup 7-23. Motherboard Audio Path Test Setup 7-24. DCFM Test Setup
..................
7-25. Audio Generator Test Setup
..............
.........
.............
.........
.............
. .
2-13 2-18
3-3 3-4 4-3
5a-1
5a-2 5b-3 5b-4 5b-5 5b-7
5b-9
5b-9
5b-11
6-3 6-9
B-11
7-3 7-4 7-4 7-6 7-7 7-9
7-9 7-10 7-11 7-11 7-12 7-13 7-14 7-15 7-15 7-16 7-17 7-18
7-19
7-20 7-21 7-22 7-24 7-25 7-26
Contents-9
7-26. HF Power Level Accuracy Test Setup for Power Levels
>-10dBm
...................
7-27. HF Power Level Accuracy Test Setup for Power Levels
of-lOto-70dBm
...............
7-28. HF Power Level Accuracy Test Setup for Power Levels
< -70 dBm and < 1300 MHz ...........
7-29. HF Power Level Accuracy Test Setup for Power Levels
< -70 dBm and > 1300 MHz ...........
7-30. LF Output Level Test Setup .............
7-31. LF Power Level Accuracy Test Setup for Power Levels
of?-40dBm
.................
7-32. LF Power Level Accuracy Test Setup for Power Levels
of<-40dBm .................
7-33. FSK Deviation Test Setup ..............
7-34. Filter Path Test Setup ................
7-35. Welcome Screen ..................
7-36. Important Information Screen ............
7-37. Choose Destination Location Screen .........
7-38. Select Program Folder Screen ............
7-39. Start Copying Files Screen .............
7-40. Installation Status Gauge ..............
7-41. Setup Complete Screen ...............
7-42. Setup Complete Screen ...............
7-43. HP Service Support Program Group .........
7-44. HP Service Software for PC’s Selections .......
7-45. User Information Dialog Box ............
7-46. DUT Selection Dialog Box ..............
7-47. Select Test Equipment and Tests Window .......
7-48. Save As Dialog Box .................
7-49. HP Service Support Software Window ........
7-50. HP Service Support Software Window Displaying Test
Results .....................
7-51. Print the Log File? Dialog Box ...........
7-52. The User Information Window ............
7-53. Test Equipment Drivers in the File Drop-Down Menu . 7-54. Adding Test Equipment Using the Test Equipment Menu 7-55. Adding the Equipment Information Using the New Test
Equipment Window ...............
7-56. Removing and Editing Test Equipment Using the Test
Equipment Window ...............
7-57. The Select Test Equipment and Tests Window ....
7-58. Test Equipment Drivers in the File Drop-Down Menu . 7-59. Adding a Device Driver Using the Test Equipment
Drivers Window .................
7-60. Using the Open Dialog box to Search for a Device
Driver File to Add ................
7-61. Removing a Device Driver Using the Test Equipment
Drivers Window .................
7-62. Test Drivers in the File Drop-Down Menu ......
7-63. Adding a Test Driver Using the Test Drivers Window 7-64. Using the Open Dialog Box to Search for a Test Driver
File to Add ...................
7-65. Removing a Test Driver Using the Test Drivers Window
7-66. Datapacks in the File Drop-Down Menu .......
7-67. Adding a
Datapack
Using the Datapacks Window . . .
7-27 7-28
7-28 7-29
7-30
7-31
7-32 7-33 7-34 7-36 7-37 7-38 7-38 7-39 7-39 7-40 7-40
7-42 7-42 7-43 7-43 7-44 7-46 7-47
7-47
7-49
7-50
7-51 7-52
7-53
7-54
7-56 7-57
7-57
7-58
7-59
7-60
7-61
.
7-61 7-62
7-63
7-64
Contents-l
0
7-68. Using the Open Dialog Box to Search for a
File to Add 7-69. Removing a 7-70. 110 Port Error Message
...................
Datapack
Using the Datapacks Window
...............
7-71. HP Service Support Program Group
Datapack
.........
7-72. HP Service Software for PC’s Selections .......
7-73. Password Requested by User Information Dialog Box .
7-74. HP8648 Motherboard Repair Utility Window .....
7-75. HP Service Support Program Group
.........
7-76. HP Service Software for PC’s Selections .......
7-77. Password Requested by User Information Dialog Box .
7-78. HP8648 Motherboard Repair Utility Window .....
8-l. FM Accuracy Equipment Setup 8-2. FM Accuracy Equipment Setup for HP 8-3. FM Accuracy Equipment Setup for Option 8-4. FM Accuracy Equipment Setup for HP
Option
lE2
...................
8-5. FM Distortion Equipment Setup 8-6. FM Distortion Equipment Setup for HP 8-7. AM Accuracy Equipment Setup 8-8. AM Accuracy Equipment Setup for Option 8-9. AM Distortion Equipment Setup
...........
8648B/C/D
lE2
. .
....
8648B/C/D
...........
8648B/C/D
. .
...........
lE2
....
...........
8-10. Phase Modulation Distortion Equipment Setup ....
8-11. Phase Modulation Distortion Equipment Setup for HP
8648B/C/D
8-12. Residual FM Equipment Setup
8-13. Harmonics Equipment Setup
8-14. Spurious Equipment Setup
...................
............
............
.............
8-15. DC FM Frequency Error Equipment Setup ......
8-16. Equipment Setup for the HP 8648A and HP
5 1300MHz.
8-17. Equipment Setup for the HP
..................
8648B/C/D
8648B/C/D
> 1300 MHz .
8-18. Pulse Modulation On/Off Ratio Equipment Setup. ...
8-19. Pulse Modulation On/Off
Risetime
Equipment Setup.
. 8-20. Pager Encoder Timebase Accuracy Equipment Setup .
8-21. FSK Deviation Accuracy Equipment Setup ......
8-22. Internal Timebase: Aging Rate Test Setup ......
8-23. HF Power Level Accuracy Test Setup Setup for Power
Levels > -10 dBm
...............
8-24. HF Power Level Accuracy Test Setup for Power Levels
of-lOto-70dBm
...............
8-25. HF Power Level Accuracy Test Setup for Power Levels
< -70 dBm and 5 1300 MHz
...........
8-26. HF Power Level Accuracy Test Setup for Power Levels
< -70 dBm and > 1300 MHz
9-
1. Frequency Accuracy Equipment Setup
...........
........
9-2. 9 kHz RF Level Accuracy Equipment Setup .....
.
7-64 7-65
7-67
7-68 7-68 7-69 7-69
7-70 7-70 7-71 7-72
8-6 8-7
8-8
8-9
8-10 8-11 8-12 8-13 8-14
8-15
8-16
8-17
8-19 8-20 8-21
8-22
8-23 8-26 8-27
8-29 8-30 8-36
8-38
8-39
8-39
8-40
9-3 9-8
Contents-l 1
Ihbles
O-l. Hewlett-Packard Sales and Service Offices ......
2-1. Programming Command Statements and Descriptions .
2-2. Dictionary of Terms .................
2-3. IEEE 488.2 Capabilities ...............
5
1. Recommended Test Equipment ...........
5-2. Adjustments and Performance Tests Required after
Repair or Replacement of an HP 8648A Assembly .
5-3. Adjustments and Performance Tests Required after
Repair or Replacement of an HP
Assembly ....................
5a-1.
Al Front Panel (Keyboard) .............
6-l. HP 8648A Replaceable Parts .............
6-2. HP
7-l. 7-2. An Example of Calibration Data for Power Sensors . .
8-l. High Power Level Accuracy Work Table .......
8-2. FSK Deviation Accuracy Work
8-3. HP 8648A Test Record ...............
8-4. HP 8648A Test Record ...............
8-5. FM Accuracy Performance Test ...........
8-6. FM Accuracy Performance Test Option
8-7. FM Distortion Performance Test ...........
8-8. AM Accuracy Performance Test ...........
8-9. AM Accuracy Performance Test Option
8-10. AM Distortion Performance Test ...........
8-l 1. Phase Modulation Distortion Performance Test ....
8-12. Residual FM Performance Test ............
8-13. Harmonics Performance Test .............
8-14. Spurious Performance Test .............
8-15. DC FM Frequency Error Performance Test ......
8-16. RF Level Accuracy Performance Test .........
8-17. Pager Encoder Timebase Accuracy Performance Test
8-18. FSK Deviation Accuracy Performance Test (Option
8-19. Internal Timebase: Aging Rate Performance Test
8-20. HP 8648B Test Record ...............
8-21. HP 8648B Test Record ...............
8-22. FM Accuracy Performance Test - Part 1 .......
8-23. FM Accuracy Performance Test - Part 2 .......
8-24. FM Accuracy Performance Test Option 8-25. FM Accuracy Performance Test Option
8-26. FM Distortion Performance Test - Part 1 .......
8-27. FM Distortion Performance Test - Part 2 .......
8-28. AM Accuracy Performance Test ...........
8648B/C/D
531
Test Point Extender Parts List ..........
(Option
Only) ......................
(Option
Replaceable Parts ...........
1EP
Only) ................
lE5
Only) ................
Table
8648B/C/D
.........
lE2
......
lE2
lE2 -
lE2 -
.....
1EP
Part 1 . .
Part 2 . .
2-i
2-20
2-43
5-2
5-5
5-6
5a-4
6-5
6-13
7-2 7-53 8-25 8-33 8-41 8-42 8-43 8-44 8-45 8-46 8-48 8-50
8-51 8-51
8-52 8-53 8-54 8-55
8-60
8-60
8-60
8-61 8-62
8-63 8-63 8-64 8-64 8-65 8-65 8-66
Contents-l 2
8-29. AM Accuracy Performance Test Option
8-30. AM Distortion Performance Test . . . . . . . . . . .
8-31. Phase Modulation Distortion Performance Test - Part 1 8-32. Phase Modulation Distortion Performance Test - Part 2
8-33. Residual FM Performance Test . . . . . . . . . . . .
8-34. Harmonics Performance Test . . . . . . . . . . . .
8-35. Spurious Performance Test . . . . . . . . . . . . .
8-36. DC FM Frequency Error Performance Test . . . . . .
8-37. RF Level Accuracy Performance Test - Part 1 . . . . .
8-38. RF Level Accuracy Performance Test - Part 2 . . . . .
8-39. RF Level Accuracy Performance Test - Part 3 . . . . .
8-40. RF Level Accuracy Performance Test with Options
andlE6-Part3.
8-41. RF Level Accuracy Performance Test - Part 4 . . . . .
8-42. Pulse Modulation On/Off Ratio Performance Test
(Option
8-43. Pulse Modulation Rise Time Performance Test (Option
lE6
8-44. Internal Timebase: Aging Rate Performance Test
(Option
8-45. HP 8648C Test Record . . . . . . . . . . . . . . .
8-46. HP 8648C Test Record 8-47. FM Accuracy Performance 8-48. FM Accuracy Performance Test - Part 2 8-49. FM Accuracy Performance Test Option
8-50. FM Accuracy Performance Test Option
8-51. FM Distortion Performance Test - Part 1 . . . . . . .
8-52. FM Distortion Performance Test - Part 2 . . . . . . .
8-53. AM Accuracy Performance Test . . . . . . . . . . .
8-54. AM Accuracy Performance Test Option lE2
8-55. AM Distortion Performance Test . . . . . . . . . . .
8-56. Phase Modulation Distortion Performance Test - Part 1 8-57. Phase Modulation Distortion Performance Test - Part 2
8-58. Residual FM Performance Test . . . . . . . . . . . .
8-59. Harmonics Performance Test . . . . . . . . . . . .
8-60. Spurious Performance Test . . . . . . . . . . . . .
8-61. DC FM Frequency Error Performance Test . . . . . .
8-62. RF Level Accuracy Performance Test - Part 1 . . . . .
8-63. RF Level Accuracy Performance Test - Part 2 . . . . .
8-64. RF Level Accuracy Performance Test - Part 3 . . . . .
8-65. RF Level Accuracy Performance Test with Options
andlE6-Part3.
8-66. RF Level Accuracy Performance Test - Part 4 . . . . .
8-67. Pulse Modulation On/Off Ratio Performance Test
(Option
8-68. Pulse Modulation Rise Time Performance Test
(Option 8-69. Internal Timebase: Aging Rate Performance Test
(Option lE5 Only) . . . . . . . . . . . . . . . .
8-70. HP 8648D Test Record . . . . . . . . . . . . . . .
8-71. HP 8648D Test Record . . . . . . . . . . . . . . .
8-72. FM Accuracy Performance Test - Part 1 8-73. FM Accuracy Performance Test - Part 2 8-74. FM Accuracy Performance Test Option 8-75. FM Accuracy Performance Test Option
lE6
Only) . . . . . . . . . . . . . . . . . . . .
lE5
lE6
lE6
. . . . . . . . . . . . . . . .
Only) . . . . . . . . . . . . . . . .
Only) . . . . . . . . . . . . . . . .
Test’- Part
. . . . . . . . . . . . . . . .
Only) . . . . . . . . . . . . . . . .
Only) . . . . . . . . . . . . . . . .
lE2
‘1 ’ : 1 : : : :
lE2 ­lE2 -
lE2 -
lE2 -
. . . . .
1EA
. . . . . . .
Part 1 . .
Part 2 . .
. . . . .
1EA
. . . . . . .
. . . . . . .
Part 1 . . Part 2 . .
:
8-68 8-70 8-71 8-71 8-71
8-72
8-73 8-74 8-75 8-81 8-81
8-82 8-82
8-82
8-83
8-83
8-85
8-86 8-87 8-87 8-88 8-88 8-89 8-89 8-90 8-92 8-94 8-95
8-95
8-95 8-96 8-97 8-98 8-99
8-105 8-106
8-106 8-107
8-107
8-108
8-108 8-109 8-110 8-111 8-111 8-112 8-112
Contents-13
8-76. FM Distortion Performance Test - Part 1 8-77. FM Distortion Performance Test - Part 2 8-78. AM Accuracy Performance Test
...........
8-79. AM Accuracy Performance Test Option 8-80. AM Distortion Performance Test
...........
.......
.......
lE2
.....
8-81. Phase Modulation Distortion Performance Test - Part 1 8-82. Phase Modulation Distortion Performance Test - Part 2 8-83. Residual FM Performance Test 8-84. Harmonics Performance Test 8-85. Spurious Performance Test 8-86. DC FM Frequency Error Performance Test 8-87. RF Level Accuracy Performance Test - Part 1 8-88. RF Level Accuracy Performance Test - Part 2 8-89. RF Level Accuracy Performance Test - Part 3 8-90. RF Level Accuracy Performance Test with Options
andlE6-Part3.
................
8-91. RF Level Accuracy Performance Test - Part 4
............
............
.............
......
.....
.....
.....
1EA
.....
8-92. Pulse Modulation On/Off Ratio Performance Test
(Option
lE6
Only) ................
8-93. Pulse Modulation Rise Time Performance Test
(Option
lE6
Only) ................
8-94. Internal Timebase: Aging Rate Performance Test
(Option
9-
1. HP 8648A Frequency Accuracy Supplemental
lE5
Only) ................
Verification Test .................
9-2. HP 8648B Frequency Accuracy Supplemental
Verification Test .................
9-3. HP 8648C Frequency Accuracy Supplemental
Verification Test .................
9-4. HP
86481)
Frequency Accuracy Supplemental
Verification Test .................
9-5. HP 8648A Option
Supplemental Verification Test
9-6. HP 8648B Option
Supplemental Verification Test
9-7. HP 8648C Option
Supplemental Verification Test
lE5
Frequency Accuracy
lE5
Frequency Accuracy
lE5
Frequency Accuracy
..........
..........
..........
9-8. HP 8648D Option lE5 Frequency Accuracy
Supplemental Verification Test
..........
9-9. HP 8648B 9 kHz RF Level Accuracy Supplemental
Verification Test .................
9-10.
HP 8648C 9 kHz RF Level Accuracy Supplemental
Verification Test .................
9-l 1. HP 8648D 9 kHz RF Level Accuracy Supplemental
Verification Test .................
8-113 8-114 8-115 8-116
8-117 8-117 8-118 8-118
8-119
8-120
8-121
8-122
8-128
8-129
8-129
8-130
8-131
8-131
8-131
9-4
9-4
9-5
9-5
9-6 9-6
9-7
9-7 9-9
9-9
9-10
Contents-14

Operation

“Operation” contains the following information:
1 Operation Provides a quick overview of the instrument’s
operation.
la Operation
Examples
lb Operation
Reference the instrument’s functions.
lc Operation
Messages HP-IB remote operation messages.
Provides examples to help you learn how to operate the instrument.
Provides quick access to information about each of
Provides information about both front-panel and
1
Note
For information about service messages numbered 500 and above, refer to Chapter
5c,
“Service Error Messages.”
Operation l-l

Quick Overview

1. Power Key

2. Display

234
Press
s
HP
8648 Signal Generator
(i%iK)
to power up the instrument. The instrument powers up
7
to the same state it was in when power was turned off, except that the RF output will be turned off and the digit-select arrow keys
(@
and
a)
will be reset to the least significant digit.
The display can be one of two displays depending on the serial number prefix of your instrument as illustrated below.
4
,\FRfgUENCY -
MCOUAIION -
AMPUI~E
I
A
/
B
4
FREQUENCY -
MCOUATION -
AMPUIUDE -
/
,
1-2 Operation
ot77a
The following table describes the prefixes that apply to the various HP 8648 models equipped with an LCD.
A. Liquid Crystal Display (LCD)
(labels located above the
display)
BP 8648A BP 8648B
Prefix Preflx
3636A and below 3623A and below 3623A and below 3613A and below 3643U and below 3642U and below 3642U and below 3642U and below
The display contrast of the LCD can be achieved using the adjustment that is located on the rear panel of these instruments. Note that this adjustment is only available for instruments equipped with an LCD. It allows you to adjust the contrast of the LCD. Turn the adjustment to optimize the display for viewing from most angles. If the display is blank, first attempt to adjust the display contrast before returning the instrument for service.
The following table describes the prefixes that apply to the various HP 8648 models equipped with a VFD.
B. Vacuum Fluorescent Display (VFD)
(labels located below the
BP 8648A BP 8648B BP 8648C
Prefix Prefix Prefix Prefix
3836A and above 3836A and above 3836A and above 3836A and above
BP 8648C
Prefix
display)
BP 8648D
Prefix
ElP
8648D

3. Function and Data Keys

3836U and above 3836U and above 3836U and above 3836U and above
The VFD is a 2x40 display. The intensity of the this display is at 100
% and cannot be adjusted.
The keys in the FUNCTION and DATA blocks allow you to enter values for setting the frequency, amplitude, and modulation level of the RF output signal.
If Option pager encoder (ENCODER) mode and FM mode.
If Option mode, the input alphabetical characters using the DATA and MODULATION SOURCE blocks when you are in pager encoder mode.
If Option mode, the key must be used to store any numeric or alphabetic characters entered by way of the DATA and MODULATION source blocks.
I
1EP
is present, the [FM) (ENCODER) key will toggle between
1EP
is present and the signal generator is in the ENCODER
L-j
1EP
(jjdB(ml)
key functions as a SHIFT key. This key lets you
is present and the signal generator is in the ENCODER
key functions as an ENTER key. The ENTER
I
I
Operation 1-3

4. Increment Set Keys

When you press a FUNCTION key, that function becomes the active function. Press the active function. Press @) or @J at any time to change the active function setting by the increment value. (If Option and the signal generator is in the ENCODER mode, these keys have alternate functions.)
(‘NCR]
to view or change the increment value for
1EP
is present

5. Knobs

6. MEMORY

If Option mode, the starts or stops any pager encoding activity. In addition, in this mode, the move the blinking cursor between each parameter when you are entering the pager encoding settings.
The knobs are always active when the instrument is in local (front panel) control. Turn them to increase or decrease the frequency or amplitude of the RF output. Press a or adjust the knob’s resolution.
Press reference value and turn on the reference mode. Press (REF ON/OFF) to turn on and off the reference mode without changing the reference value. When the reference mode is on, the displayed value indicates the offset between the reference value and the RF output signal.
If Option 1EP is present and the signal generator is in the ENCODER mode, the AMPLITUDE/ENCODER knob is used to enter a setting for a pager encoding parameter.
Memory registers allow you to save instrument set-ups and recall them whenever you wish. Press
number to save the instrument’s current settings.
settings, press allow you to recall registers in numerical sequence. You can arrange your registers in up to ten different sequences.
(jjj
and
@GZQ,
1EP
is present and the signal generator is in the ENCODER
(jj]
(7J-J
key functions as a START/STOP key. This key
function as PREV and NEXT keys. These keys let you
(ZJ
next to each knob, to
next to each knob, to set the displayed value as the
a
and enter a two-digit register
‘Ib
recall the
(REG)
and enter the register number. The arrow keys
1-4 Operation
The number of the currently selected sequence and the last register selected are always displayed in the lower-left corner of the display to help you keep track of where you are in your testing process. (If Option on any pager encoding menu.) The memory register examples provided in Chapter la, “Operation Examples,” show you how to create a sequence and how to delete or add registers in your sequence.
1EP
is present, the sequence and register are not displayed

7. Modulation Source

Press
CMOD
ON/OFF) to turn on or off the modulation source. Press
@KiZiG@
or
@i7iiKJ
to select one of the internal source tones
for modulating the RF output signal. These tones are also available
as an output signal at the MOD INPUT/OUTPUT port when they are
selected. Press
C-1
or
[EXTDC)
to ac- or dc-couple an external
audio source via the MOD INPUT/OUTPUT port. Press
(1kt-b
+ EXT
DC)
to frequency modulate the RF signal with
the internal 1 kHz tone and an external source at the same time. (Additional internal plus external modulation capabilities are available for HP-IB operation.)
ClkHz
+ EXT
DC)
will also amplitude or phase
modulate the RF signal with the internal 1 kHz tone but it will not be dc-coupled.
If Option 1EP is present, the Option
lE2
is present, the
[@7iiK]
(NTIkHz)
(FREQUENCY) key, or if
(FREQUENCY/WAVEFORM) key scrolls between five states: a fixed 1 kHz internal source and a variable-frequency internal source with four different waveform selections. The four modulation waveforms are sine, triangle, square, and sawtooth (or ramp).
Operation
l-5

Operation Examples

This section contains operating examples to help you learn how to operate the signal generator. These examples can be performed without any additional equipment. The pager testing example can only be performed if Option 1EP is present.
la

Getting Started

If this is the first time you have operated this instrument, perform each of the following examples for a quick introduction to general operation. After you have completed the examples, try operating the instrument’s remaining functions on your own. If you have trouble or want additional information on a function, refer to Chapter lb,
“Operation Reference.” If a message is displayed that you do not
understand, refer to Chapter lc, “Operation Messages.”
Operation Examples la-l

Operation Examples

This section provides the following examples of signal generator
operation. The item numbers of the operation examples correspond to
the numbers called out on drawing of the instrument front panel.
1. Setting the RF Output Signal
2. Incrementing or Decrementing the RF Output Signal
3. Using the Memory Registers
4. Offsetting the RF Output from a Reference
5. Holding the Output Attenuator Range
6. Setting a User Selectable Modulated (Option
lE2
or
1EP
Only)
Freauency
7. Signaling a Numeric-Type FLEX Pager (Option
YcnnJLATIoN
SOURCE
RF
OUTPUT
A
and Waveform
1EP
Only)
“*Lu”
\
la-2 Operation Examples
an622a2d

Setting the RF Output Signal

In this example, you will set the frequency, amplitude, and modulation level of the RF output signal.

Setting the Frequency

/

Setting the Amplitude

1. Set the frequency to 100 MHz using the keys shown below the instrument diagram.
If you make a mistake while entering a value, press a to correct it.
J
\
FREQUENCY
100.00000 MHz
2. Set the amplitude to -100 dBm.
-
MODUfATlON -
AMP-
fund1
.drw
4
J
funcfZ.dw
Operation Examples la-3

Turn on the RF Output

3. Press
C-1
to turn on the RF output.
RF OFF is displayed below the amplitude setting when the RF
output is turned off.
Setting the
Modulation
II ” ~“““’ ;I
4. Set the FM deviation to 3 kHz. The modulation rate is displayed below the deviation setting. Use
the MODULATION SOURCE keys to select a modulation source and turn modulation on or off.
3
FREQUENCY
100.00000 MHz
\
-
MODULATION
FM 3.00
kHz
1kHz
-
AMPLITUDE
-100.0 dBm
funct4.drw
,
/
la-4 Operation Examples

Incrementing or Decrementing the RF Output Signal

In this example, you will increment the amplitude and frequency of the RF output signal.

Preliminary Steps

[ FwJ,‘,,
]
[+] [o] 10 [z]

Using the Knob

1. If they are not already set, set the frequency to 100 MHz, and the amplitude to -100 dBm.
FREQUENCY
100.00000 MHz
2. Increment the amplitude using the knob. Press @ or
(TJ
-
when you wish to adjust the increment resolution.
MODULATION
MODUl.ATlON
-
AMPLITUDE
-100.0
dBm
lncrl
.dnv
incr2.dw
Operation Examples la-5
Using the Increment
keys
3. Enter a frequency increment of 25 kHz. The $ symbol is displayed when you press
that the displayed value is the increment set value.
2500000:kHz
L
4. Increment the RF output frequency in 25 kHz steps. The increment keys affect the last FUNCTION selected
(FREQUENCY, AMPLITUDE, FM, AM or
@iiZWi]
-101.0
dM).
to indicate
dBm
/
000
OOn
0
o--
0 00
q c
0 not
0
0 q OCI
OrI
II
$
FREQUENCY
100.02500 MHz
-
MODULATION
,-
AMP”>
-101 .O dBm
h
fncr4.dw
la-6 Operation Examples

Using the Memory Registers

The memory register examples show you how to create a sequence of
registers, delete a register from that sequence, renumber the registers in the sequence, and insert a new register in the sequence.
Up to 10 register sequences can be defined (0 through 9). A sequence can contain up to 100 registers (00 through 99). There are a total of 300 registers available in the instrument. The registers can be used
in the sequences in any combination (such as 10 sequences of 30
registers each, or 3 sequences of 100 registers each) as long as the total does not exceed 300 registers. It is not possible to have all 10 sequences each contain 100 registers as that would be 1000 registers.
(If Option
1oo.ooOOoMHz
1EP
is present, there are a total of 70 registers available.)
FREWENCY
B
EGOI
\
MODuATlON
FM
3.OOKz 1lcHzoFF
m .uw”nmE
-I
36.0dBm RFOFF
z
SEQ 0
SEQ
9
REG 00
REG
99
Operation Examples la-7
Saving Instrument
Settings in Register
Sequences
SE0
0
b3
In this ten step example, you will use the memory keys to create
a sequence containing three registers. Each register will contain a different frequency setting.
Selecting the Sequence
1. Select sequence 0. If there are registers saved in sequence 0, the message shown
in the display below will not appear. Note that the steps in this example will cause the settings in registers 00, 01, and 02 of
sequence 0 to be changed.
regseql
Saving Settings in Registers
.dnv
2. Set the frequency to 10 MHz.
3. Save the instrument settings in register 00.
J
FREQUENCY
10.00000 MHz
SEQ 0 REG 00
-
MODUIATION
-
AMPUTLIDE
regseqZ.dnv
1
,
I
la-8 Operation Examples
4. Set the frequency setting to 11 MHz.
FFIEQUENCY ,-
MODULATION
-
5. Save the instrument settings in register 01.
6. Set the frequency to 12 MHz.
AMPLITU
regseqldrw
regseq&drw
Operation Examples la-9
7. Save the instrument settings in register 02.
regseq9.dw
FREQUENCY
-
MODULATION
-
AMPLITUDE
12.00000 MHz
SEQ 0 REG 02
Checking the Sequence
8. Recall the registers in sequence 0. The @) and @J keys recall registers or sequences depending on
which key was pressed last
FREQUENCY
SEQ 0 REG 01
-
12.00000 MHz SEQ 0 REG 02
10.00000 MHz
SEQ 0 REG 00
(m
or
(SEQ).
MODULATION
-
AMPLITUDE
regseq7.drw
la-l 0
Operation Examples
Checking a
Different
9. Select sequence 1.
16
\
SEQ
Sequence
FREQUENCY
1
-
MODULATION
--,
AMPUTUDE
/
rea9.dw
10. Step through the registers in sequence 1 if there are registers saved in it.
Note
Sequence 1 does not contain the settings you saved in sequence
0. The instrument enables you to save different settings in each sequence to create up to ten different sequences for your testing. Remember when you save or recall a register, be sure that the correct sequence is also selected.
reqO.dw
Operation Examples
la-l 1
Deleting a Register
from the Sequence
In this example, you will delete a register from the sequence you created in the preceding example.
Selecting the Sequence
1. Select sequence 0.
I
1.i..
SE0
flc5
0
Note
FREQUENCY
\
SEQ 0
-
MODULATION
-
AMPLITUDE
delregO.drw
Deleting a Register
2. Delete register 01 from sequence 0. The contents of the register are recalled when it is deleted. This
allows you to resave the contents if you need to.
4
FREQUENCY
11 .OOOOO MHz SEQ 0 REG --
-
MODULATION
-,
AMPLITUDE
,
la-12 Operation Examples
delregl
.drw
Step through the remaining registers in sequence 0.
3. The deleted register number has been removed from the sequence.
Note that the instrument does not renumber the registers when one is deleted.
FREOUENCY -
12.00000 MHz
SEQ 0 REG 02
10.00000 MHz
SEO
0 REG 00
MODULATION
-
AMPUTUDE
delreg2.drw
Operation Examples la-13
Renumbering the
Registers in a
Sequence
In this example, you will eliminate the skip from register 00 to register 02 in sequence 0 caused when you deleted register 01 in the previous example.
Decreasing the Register Number
1. Delete register 02. The settings saved in register 02 are recalled when it is deleted.
2. Save the settings from register 02 into register 01.
12.00000 MHz
\
SEQ 0 REG 01
nemwta2.dnv
/
la-14 Operation Examples
Checking the Sequence
3. Step through the register sequence.
I
I
12.00000
LSEQOREGO~
I
10.00000MHz
SEOOREGOO
MHz
I
r
numseq3.drw
Note
In this example, you renumbered one register. When you need to renumber two or more registers, use
each register until you get to the last register in the sequence, then
use (DEL).
[REG)
instead of (DEL) to recall
Operation Examples la-15
Inserting a Register in
a Sequence
In this example, you will insert a register into the sequence you created in the previous example. The process involves incrementing each register number that comes after the point in the sequence where you wish to insert a register.
1. Recall the last register in sequence 0.
mu
L
i ‘I””
000 Em
0 00 00 0
0 or
0
PJ
!a GCEI
4
I \
-J
FREQUENCY
12.00000 MHz
\
SEQ 0 REG 01
-
MODULATION
2. Save the recalled settings into register 02.
-
AMPUTUDE
/
insseql .dw
insseq2.dnv
Operation Examples
la-l 6
3. Recall register 00. Register 01 can now be used to save the settings that are saved in
register 00.
FREQUENCY
-
MODULATION
-
AMP>
Insseq3.dw
,
4. Save the recalled settings into register 01. Register 00 can now be used to save the new settings.
9
FREQUENCY
-
10.00000 MHz
SEQ 0 REG 01
Saving a New Register
5. Set the frequency to 8 MHz.
FREQUENCY
8.00000 MHz
I
6. Save the settings in register 00.
-
MODULATION
MODULATION
-
AMPUT-
,
/
inswq4.drw
-
AMPUTUDE
I
insseq5.drw
Press
0-J
to check the new sequence.
8.00000 MHz
insseq6.drw
Operation Examples
la-l 7

Offsetting the RF Output from a Reference

In this example, you will enter an RF output frequency, set it as the reference value, and then offset the RF output frequency 10 MHz below the reference value.
Setting the
Reference
Value
1. Set the frequency to 500 MHz.
500.00000 MHz
2. Set 500 MHz as the reference frequency. The A symbol appears in the display to indicate that the reference
mode is selected. The output frequency is still 500 MHz.
refl
.drw
la-l 6
refZ.dn%
Operation Examples
Offsetting the RF
output
3. Offset the output frequency 10 MHz below the reference frequency.
You can enter in the offset value directly, or use the knob or and
(IJ
keys.
@J
i
i ‘I’
EE
00
00
n
000
000
r--l
[pfjry
Attention!
[o] [Z]
FREQUENCY
-10.00000
-
nMHz
MODULATION
-
AMPLITUDE
L
ref3,drw
In the reference mode, the output frequency equals the reference
/
frequency f the displayed offset frequency.
Turning
the Reference Mode Off or On
4. Turn off the reference mode to display the actual output frequency.
FREQUENCY
490.00000
-
MHz
MODULATION
-
AMPLITUDE
5. Turn on the reference mode without changing the reference frequency.
.
----=A
-
011
\I
y-1(1.00000~MH~
i-----~
;
I
i
/I
11 “I
-
-------
11
110 ij$!~jE&i~~~~~
‘h?fEEk3~iA
1 0 ~J~Kqp+i
FIEF
ON/OFF
i-
0
JI;; 0 Jj
r^L^. ._..^.,
riit""eNLT
p.
L
..^ -... __.^..
M"U"lAl I"N -
rel4.dnv
-TIITl,n~~
AM, Y YYL
1
ref5.dw
Operation Examples la-19
6. Change the displayed units to
kHz.
Note that for amplitude, reference settings are displayed in
dB
units only.
FREQUENCY
-
MODULATION
-
AMPUTUDE
-1OOOO.OOakHz
Setting a New Reference Value
7. Set the current output frequency as the new reference frequency at any time.
ref7.dr-a
la-20 Operation Examples

Holding the Output Attenuator Range

In this example, you will hold the output attenuator so it does not
change ranges when you change the amplitude setting. This will
prevent attenuator range changes from affecting the output signal.
Set the Amplitude
Level
[=I
I+/-]
[s]
[T--y
Holding the
Attenuator
a&7?2iJjm:I,I’
1. Set the amplitude level to -82 dBm.
[Z]
2. Hold the attenuator at this setting.
I
\
J ,
FREGUENCY -
MODULATION
-82.0
dBm
-
AMPLITUDE
-82.0
dBm
HOLD
/
Operation Examples la-21
Adjusting the
Amplitude
3. Adjust the amplitude setting. Now amplitude changes do not cause the attenuator to change its
range setting. Consequently, amplitude changes are limited to the range provided by the instrument’s vernier. For information about the instrument’s vernier ranges, refer to Chapter lb, “Operation Reference.
la-22 Operation Examples
Setting a User Selectable
Modulated
Frequency and
Waveform
(Option lE2 or
Only)
1EP
Note
Setting the
Modulation Level
This modulation example can only be performed if Option is present.
In this example, you will select the modulation level and the modulated frequency and waveform of the RF signal output.
1. Select FM modulation with a deviation of 25 kHz. Either LAM) or
The modulation type (FM, AM, or (deviation or depth) is displayed on the top line of the front-panel display as shown.
(jKJ
modulation may be used instead of (FM.
4M)
and the modulation level
pRE9uENCY
-
MOD”LATlON
FM25.0kHz
-
AMPUTUDE
lE2
or
1EP
c
Operation Examples la-23
Setting the Modulated
Waveform
(Frequency/
Waveform)
2. Press the
@KiiZ’
(FREQUENCY/WAVEFORM) key until
selected. Repetitively pressing the
(INT’
(FREQUENCY/WAVEFORM) key
selects one of five states:
n A fixed 1 kHz sine-wave internal source. n A variable-frequency sine-wave source
(indicated by SIN preceded by the frequency value).
n A variable-frequency triangle source (TRI). n A variable-frequency square-wave source n A variable-frequency sawtooth (or ramp) source (SAW).
(SqU).
SqU
is
one1 md
Setting the Modulated
Frequency
3. Set the modulated frequency to 1.5 kHz. The
IkHz)
key is the only accepted units key.
1.5OkHz S&R
la-24 Operation Examples
Signaling a Numeric-Type FLEX Pager
(Option
1EP
Only)
In this example, you will set up the pager encoder to send a signal appropriate to test a numeric-type FLEX pager.
[rhixcz
Note
8LeA mMi*-10WMHz
SIGNAL GENER.4TOR
This pager encoding example can only be performed if Option present.
The front panel of the Option 1EP instrument is different from the standard instrument’s front panel. The green and blue labels are incorporated for the pager encoding mode only. Therefore, these keys have multiple functions on instruments with Option
A-
FREWENCY
FORMAT FLEX POLARITY NORMAL
-
MODUtATlON
~AMPLITUDE-
FILTER ON
!
1EP.
I:7
gWLMWENCO~ER
El
h-1
gEJj
I
1EP
is
HP 8648A Option
1EP
Signal Generator
Operation Examples la-25
Setting Up Pager
Encoding
The following steps are required to set up pager encoding on the
signal generator. Details of setting each parameter are provided following this overview.
1. Before entering the signal generator pager encoder mode, set the correct carrier frequency and FM deviation for the pager-under-test. (The FM deviation is 4.8 kHz for FLEX and is
4.5
kHz
for POCSAG.)
2. Press
3. Enter the pager encoder mode by pressing m (ENCODER). If FM
4. Set up the pager encoder parameters to meet the test requirements
5. Start signaling the pager after all of the encoder parameters are set
[-ON/OFF_)
below the amplitude setting when the RF output is turned off.)
wasn’t the last active function, press (FM) (ENCODER) twice.
of your specific pager.
Use the encoder parameters. The cursor will blink around the first letter of the active parameter. The @J (NEXT) key moves forward sequentially through each pager encoder parameter and the
@‘j
(PREV) ke
To move between settings for each parameter, rotate the AMPLITUDE/ENCODER knob until the desired setting is displayed. When characters are entered by way of the keypad (such as the capcode), terminate the entry with the (ENTER) key.
by pressing the
to turn on the RF output. (RF OFF is displayed
(IJ
(NEXT) and @) (PREV) keys to scroll through the
y moves backward sequentially.
(jjdB(mL)
(jj]
(START/STOP) key.
la-26 Operation Examples
Entering Pager
Encoding Settings
1. Display the first pager encoder menu. After setting your pager’s carrier frequency and an FM deviation
of 4.8 kHz (for a FLEX pager) on the signal generator, press the
a
(ENCODER) key once again to display the first pager encoder
menu (FORMAT). The
(KiJ
(ENCODER) key toggles between the
frequency modulation menu and the pager encoder menu.
Selecting the Format
Settings
FREQuENcl B
FORMAT
POLARITY
MODmION
FLEX
NORMAL
FILTER ON
Encodel.drw
2. Set FORMAT to FLEX, POLARITY to NORMAL, and FILTER to ON. a. With the blinking cursor on the “F” of FORMAT, set the FLEX
format by rotating the AMPLITUDE/ENCODER knob until FLEX is displayed.
b. Press the
(JJ
(NEXT) key to move the blinking cursor to POLARITY, then use the AMPLITUDE/ENCODER knob to set the polarity to NORMAL.
c. Using the
(7JJ
(NEXT) key and the AMPLITUDE/ENCODER
knob, set FILTER to ON.
v
FREpuENol B
FORMAT POLARITY
MODuLA!mON
FLEX NORMAL
~AMPLIlvDE
FILTER ON
Press the
(JJ)
(NEXT) key to move to the next page.
Encodel.drs
Operation Examples la-27
Selecting the Data
Rate and Pager Type
Settings
3. Set DATA RATE to AMPLITUDE/ENCODER knob and the
3200/2
and PAGER TYPE to NUMERIC, using the
(J-j
(NEXT) key.
You may choose to set the data rate to one of the other settings;
your FLEX pager should automatically adjust.
1
FREQUPNCY -
DATA RATE PAGER TYPE NUMERIC
32oOL2
MOD-ION -X,MLZL~,DE
EncodeZ.drv
Press the @J (NEXT) key to move to the next page.
4. Set VECTOR TYPE to STANDARD using the AMPLITUDE/ENCODER knob.
B-QMNCY -
VEmOR TYPE
-DUIA!CION -TU.SFT,ITUDE
STANDARD
-77
la-28 Operation Examples
Press the
(J’Q
(NEXT) key to move to the next page.
Encode3,drr
Selecting the Message
Settings
5. You may choose one of the five fixed messages (only numbers one and five are useful for numeric pagers) or you may define your own message. For this example, use your own phone number as the user-defined message:
a. Set MESSAGE NO. to 6 and MESSAGE LENGTH to 40 using the
AMPLITUDE/ENCODER knob and the @) (NEXT) key.
b. Press the @ (NEXT) key again to select the FREE MESSAGE
parameter.
c. Enter your phone number with the numeric keys and
terminate your entry with the
ljjdB(mL)
(ENTER) key.
Selecting
Transmission
Repetitions and
amplitude
v
Press the
MESSAGE NO.6
FREEMESSAGE
(IJ
(NEXT) key to move to the next page.
6. Set MODE to SINGLE and AMPLITUDE to 0. AMPLITUDE/ENCODER knob and the
MESSAGELENGTH 40
OdBm
(7J-J
(NEXT) key. Enter the numeric values using the numeric keys and the (ENTER) key.
FREQUENCY-
MODULATION ~-
using the
C-j-
AMPLITLID
encode5 dm
Press the NEXT
(a)
key to move to the next page.
Operation Examples la-29
7. Set IMMEDIATE STOP to OFF, HEADER to ON, and TERMINATOR to ON using AMPLITUDE/ENCODER knob and the @J (NEXT) key.
These are default settings that normally would not be adjusted.
Selecting the Pager
Capcode
(Address)
Note
FREQUENCY
IMMEDIATE STOP OFF HEADER ON
Press the
8. Enter your pager’s menu. The alphabetical characters. For example, to enter the following
capcode:
c-j-
Th
AMPLITUDE/ENCODER knob, set DUMMY CALL to OFF.
The pager’s parameters of the last two menus.
I
1
I
(JJJ
(NEXT) key to move to the next page.
(-1
A0012477, press
(ENTER).
capcode
contains information that automatically sets the
MODULATION
capcode
(SHIFT) key is required to enter
-
(address) in the pager code
c-1
en, using the @J (NEXT) key and the
(SHIFT), 0 (A), @i?ZV),
TERMINATOR ON
AMPLITU
/
p=&qihJ~~
la-30 Operation Examples
Press the
PAGER CODE A0012477
(lJJ
(NEXT) key to move to the next page.
q
encoddO.cdr
9. ADDRESS TYPE and ADDRESS1 are set automatically when the
capcode
is entered in the previous menu. If A0012477 was entered, SHORT and 0045245 would be displayed respectively. Generally, you would not change these settings.
cncodc41 .cdr
Selecting the Protocol
Settings
Note
Press the
(JJ)
(NEXT) key to move to the next page.
10. FRAME, PHASE, and COLLAPSE CYCLE are also set automatically when you enter the pager’s
capcode.
If you entered A0012477
previously, 0 11, D, and 4 would be displayed.
If MODE is set to BURST or CONT instead of SINGLE, you may choose to change the collapse cycle to 0 to 7. With a collapse cycle of 4, the pager will receive the message once every 16 frames
(24).
If the collapse is 0 or 1, the pager will receive the message either every frame (2’), or every other frame (2l).
Press the
(iJJ
(NEXT) key to move to the next page.
encode42
..C&
Operation Examples
la-3 1
Selecting the Roaming
Mode Settings
11. Set ROAMING MODE to NONE using the AMPLITUDE/ENCODER knob.
I I
&@j=@<_-
0

Encoding

FREQUENCY --
CYCLE 00 COLLAPSE CYCLE 4
FRAME 011
-AMpLrmDE
PHASE D
cncodc42.cdr
Press the @j (NEXT) key to move to the next menu.
12. To start encoding after selecting all pager encoder parameters, press the
(j-SET_)
MESSAGE NO. CYCLE 00
(START/STOP) key.
6
FRAME 011
AMPLITUDE 0.0 dBm
PHASE D
cncodc43.cdr
la-32 Operation Examples

Operation Reference

This chapter describes each of the instrument’s functions including all of the front panel keys, the rear panel connectors, and the optional remote interface and memory interface. This information is presented in the same functional groups as the front panel key functional groupings.
lb
Note
Option Consequently the front panel of the Option different from the standard instrument’s front panel. The green and
blue labels are incorporated for the pager encoding settings only.
Therefore these keys have multiple functions on instruments with Option
1EP
adds a pager encoder capability to the standard functions.
1EP
instrument is
lEl?
Operation Reference
1 b-l
Frequency/Amplitude

Frequency and Amplitude

1. Knob

2. Digit-Select Arrow Keys

Note
'I
The knob and reference set keys work similarly for both frequency and amplitude settings.
Turn the knobs to increment or decrement the frequency and amplitude settings. The knobs are always active when the instrument is in local operation.
If Option 1EP is present and the signal generator is in the ENCODER mode, the knob in the AMPLITUDE/ENCODER block is used to select the desired setting for each pager encoder parameter.
Press these digit-select arrow keys changed with the knob.
The knobs increment the selected digit only. For information about incrementing by an arbitrary value using the increment set keys, see
“Increment Set” in this chapter.
2
3
4
(@ a)
to select the digit to be
ireqkey&dnv
Operation Reference
1 b-2

3. REF SET

Frequency/Amplitude
Press
[REF)
RF output setting as the reference value. The reference value is stored in non-volatile memory until you replace it by pressing
[REF)
to turn on the reference mode and to set the current
again.

4. REF ON/OFF

Output Power
Trouble?
When you press value and the units. When A appears, the displayed value indicates the offset between the reference value and the RF output signal.
The RF output signal is not changed when you press this key.
@EKE),
the A symbol is displayed between the
Units
When you press or kHz. For amplitude, values can be entered in any of the amplitude units provided, but they are displayed in dB only.
Press (REF ON/OFF] to turn off the reference mode if it is on, or to turn on the reference mode without changing the reference value.
When you turn on the reference mode, the displayed value indicates the offset between the reference value and the current RF output setting.
The RF output signal is not changed when you press this key. If the RF output power seems too low, look for A in the display
between the power level value and the dB indicator. The A tells you that reference mode is turned on. The displayed value is not the
output power level; it is the offset between the reference value and
the output power. To exit the reference mode, press can then reset the output power to the desired level.
[-SET)
for frequency, values can be entered in MHz
[REF
ON/OFF].
You
Operation Reference
1 b-3

Function

Function
hmckeys.drw

1. FREQUENCY

The RF output frequency range for each model is shown in the following table. When making frequency changes, the instrument does not turn off the RF output. However, the power does blank for a few milliseconds when crossing the frequencies at 249, 501, 1001, 1260,
1600, 2001, 2520, and 3200 MHz.
Frequency Ranges
Model
HP 8648A HP 8648B
HP 8648C9 kHzto
HP
8648D9 kHzto
Frequency Range
100kHzto
9 kHzto2000 MHz
1000 MHz
3200 MHz 4000 MHz
Frequency switching typically takes less than 75 ms at frequencies lower than 1001 MHz, and less than 100 ms at higher frequencies.
1 b-4
Operation Reference
Function

2. AMPLITUDE

3. FM AM
oM
The RF output amplitude range is shown in the following table. When making amplitude changes, the instrument does not turn off the RF output. The period of any over- or under-ranging that may occur during level transitions is typically less than 30 ms.
Output Amplitude Ranges
Amplitude
+10to-
+13to-136 +10to-136 +17to­+20to-136 +19to­+17to­+15to­+13to-
136
136
136 136 136 136
dBm dBm dBm dBm dBm dBm dBm dBm dBm
Press
Model
HP 8648A
HP 8648BICID
HP
8648BIClD
with Option
IFM)
to set the peak deviation for frequency modulation. Then
Frequency Range
100
1EA
kHz
to 1000 MHz
5 2500
MHz
> 2500 MHz
< 100
kHz
5
1000 MHz
5
1500 MHz
5
2100 MHz
5
2500 MHz
>
2500
MHz
use the data entry keys to enter the desired value of deviation. The
values allowed depend on the RF frequency selected. See Chapter 4,
“Specifications” for peak deviation specifications.
Press (AM) to set the amplitude modulation range. Then use the data entry keys to set the desired value of range. Values from 0 through
100% are allowed.
Press
Idrvl]
to set the peak deviation for phase modulation. Then use
the data entry keys to enter the desired value of deviation. The
values allowed depend on the RF frequency selected. See Chapter 4,
“Specifications” for peak deviation specifications.
Operation Reference
1 b-5
Function
Setting Up the
Pager Encoder
m=
Note
*MBA m0wlz-10mHz
SGNN GENERATOR
The pager encoder mode is available only on instruments with Option
Option functions. Consequently the front panel of the Option is different from the standard instrument’s front panel. The green and blue labels are incorporated for the pager encoding settings only. Therefore these keys have multiple functions on instruments with Option
c
FORMAT FLEX POLARITY NORMAL FILTER ON
1EP.
1EP
adds pager encoding/signaling capabilities to the standard
1EP.
FREOUENCY
D
MODWATION
~AwLlTLJDE-
1EP
instrument
\
\
I
Operation Reference
1 b-6
HP 8648A Option
For numeric messages, the following keys are assigned for the special characters in addition to the standard numeric keys (0 to 9).
1EP
Signal Generator
Front
Panel Key
S
S
UU UU
Space
Space SpaceSpace
1
[ [[
Character B3 B2 Bl BO
-
(Hyphen)(Hyphen)
SpareSpare
]]
1 0 1 01 0 1 0 1 0 1 11 0 1 1 1 1 0 01 1 0 0 1 1 0 11 1 0 1 11101110 11111111
1'

1. ENCODER

00000
00000
00000
00000
D pLY=cJ
Function
Note
Before selecting the pager encoder mode check that the correct pager carrier frequency and FM deviation have been set.
To select the pager encoder mode, press m (ENCODER) twice:
once to set FM deviation, then again to toggle between the standard functions and the pager encoder mode. The display will look like the following:
FORMAT
POLARITY
This format menu is always displayed first when the pager encoder
mode is selected. The @) (PREV) and
cursor between each parameter.
($J
(NEXT) keys are used to move the blinking
FLEX NORMAL FILTER ON
Enccdel.drw
7
The knob in the AMPLITUDE/ENCODER block is used to select the desired setting for each pager encoder parameter.
To start or stop encoding, press
@iGZ=iSET)
(START/STOP).
Operation Reference
1 b-7
Function

Setting the Format

Note
The format menu, shown in the previous figure, has the following
settings:
n FORMAT selects the pager protocol being tested.
FLEX: FLEX-TD: POCSAG: PN15: RESYNC:
FLEX format FLEX-TD format
POCSAG format Pseudorandom Sequence (PN 15) Re-synchronization function (for FLEX/FLEX-TD pagers only)
SERVICE:
Servicing the HP 8648A Option
1EP
Depending on the format selected, the parameter menus will be different. Messages displayed while sending encoding signals are also unique to each selection.
n POLARITY selects the data polarity.
NORMAL:
POCSAG FLEX/FLEX-TD
“1”: Carrier - Deviation
“0”: Carrier + Deviation
Normal polarity
FLEX/FLEX-TD
(a-Level
RESYNC, PN15
“1”: Carrier + Deviation “10”: Carrier + Deviation
“0”:
Carrier - Deviation “00”: Carrier - Deviation
FSK),
“11”: Carrier + (Deviation/J) “01”: Carrier -
(4-Level FSK)
(Deviatiod3)
INVERSE:
POCSAG
“1”: Carrier + Deviation “1”: Carrier - Deviation “10”: Carrier - Deviation
“0”: Carrier - Deviation
n FILTER selects whether or not the baseband modulation filter
(lOth-order
Inverse polarity
FLEX/FLEX-TD FLEX/FLEX-TD
(S-Level FSK),
RESYNC, PNlS
“11”: Carrier - (Deviation/J) “01”: Carrier + (Deviation/S)
“0”:
Carrier + Deviation
“00”: Carrier + Deviation
Bessel low-pass filter, -3 dB at 3.9
(4-Level FSK)
kHz)
is used.
Normally the filter is on. ON:
OFF:
Uses the low-pass filter Does not use the low-pass filter
lb-6 Operation Reference

FLEX/FLEX-TD

Function
Setting the Data Rate and Pager Type
The parameter menu for the data rate and the pager type looks like
the following:
F=QuENcl-
DATA RATE
PAGER TYPE NUMERIC
n DATA RATE specifies how fast the data stream is output in bits per
3200/2
MODULATION
-AMPLITUDE
Encode2.dra
second and whether it is 2-level or 4-level FSK.
1600/2: 3200/2: 3200/4: 6400/4:
n PAGER TYPE specifies the type of message a pager under test can
1600 bps, 2-level FSK 3200 bps, 2-level FSK 3200 bps, 4-level FSK 6400 bps, 4-level FSK
respond to and is selectable from the following: TONE ONLY:
NUMERIC: ALPHANUMERIC: HEX/BIN:
Short Message/Tone only type Numeric type Alphanumeric type HEX/Binary type
Operation Reference lb-9
Function
q If NUMERIC is selected, the next setting to be selected is VECTOR
TYPE as shown:
VECFOR
TYPE
n VECTOR TYPE identifies the type of message as follows:
STANDARD
EncodeS.drr
STANDARD:
The pager-under-test displays the numeric
message it receives.
SPECIAL:
The pager-under-test converts the numeric message to a pre-defined message which is displayed.
NUMBERED:
The message numbers are assigned for each paging address separately starting at 0 and progressing up to a maximum of 63 in consecutive order.
If NUMBERED is selected, the next parameter to be selected is
NUMBER as follows:
encode21
.dtw
NUMBER is for setting the initial number of the numbered numeric message that will be transmitted. The allowable range is 0 to 63.
lb-10 Operation Reference
q
If HEX/BIN
LENGTH:
is selected, the next setting to be selected is
BLOCKING
Function
Note
FREQLENCY~.
BLOCKING LENGTH
I
n
BLOCKING LENGTH
1BIT:
7BITS: 8BITS:
14BITS:
16BITS:
1 bit per character 7 bits per character 8 bits per character
14 bits per character
16 bits per character
MODULAllON -
1BIT
indicates the bits per character as follows:
Enter the message in the hexadecimal format to the MESSAGE NO. 6 in the message menu.
When
7 BITS
or
14 BITS
are selected, the total number of hexadecimal character of the message must be a multiple of 7 characters (28 bits).
I
I
lb-l
Operation Reference
1
Function
Setting the Message
The message menu looks like the following:
MESSAGE NO.1
w
MESSAGE NO. is defined as the following where “X” represents the
MESSAGE LENGTH 40
phase setting: 0:
Arbitrary message set from the external controller using the HP-IB capability. The data must be entered in units of words. Refer to
:ARBitrary :DEFine <NRl>, <NRl>,
“[SOURceI
:PAGing
[:FORMat] :(POCSaglFLEXIFTD)
. . . ,
<NRl>"
and “Using
the Buffer Memory for the Arbitrary Messages” in Chapter
START
FRAME and STOP FRAME are defined in the following
menu:
FREQLENCY-
START FRAME
000
MODULATION -
AMPLII’UDE
STOP FRAME 127
2.
lb-12 Operation Reference
encode22.drw
The frames between the START FRAME and STOP FRAME will be transmitted. The allowable ranges for both are 000 to 127.
1: Displays: 0123456 [Xl
where “X” =
“3” for phase D.
2: Displays: ABCDEFG
where “X” = phase “A”, “B”, “C”, “D”.
3: Displays: TEST PAGING :XXXXXXX
where “XXXXXXX” = “PHASE A”, “PHASE B”, “PHASE C”, or
“PHASE D”.
4: Displays: ALPHANUMERIC 40CHARS TEST PAGING :XXXXXXX
where “XXXXXXX” = “PHASE A”, “PHASE B”, “PHASE C”, or
“PHASE D” .
5: Displays: 8888888888888888888888888888888888888888
“0”
for phase A, “1” for phase B, “2” for phase C,
[XI
Function
for LCD test with 40 characters of “8”.
6:
User-definable message, up to 40 characters long, set from the front panel. Enter numeric messages with the keypad. Alphanumeric messages require the
[Z7iQY]
(SHIFT) key to access the alphabetical characters (blue letters) above the keypad. If required alphabetical characters are not available on the front panel, send a message from an external controller over HP-IB. Refer to “
:
(POCSag I
n MESSAGE LENGTH specifies the maximum number of characters that
FLEX I
FTD3 :MESS :DEFine ’
[SOURceI
:
PAGing
[
:
FORMat]
string”’ in Chapter 2.
will be sent out and displayed in the message field. The allowable range is 1 to 40. After specifying the number of characters, press
[vdB(mL)
(ENTER) to save the new value.
When the value of this parameter is equal to or greater than the length of the message, the whole message will be sent out.
When the value of this parameter is less than the length of the message, only the characters specified by this parameter are output. (For example, if the MESSAGE NO. is 1 (0123456[Xj) and the MESSAGE LENGTH is 5, the message which is displayed and will be sent out is actually 01234.)
Operation Reference lb-13
Function
Setting the Encoding Mode
The number of times a message is transmitted and the level of that transmission are determined by the following menu:
EmEQ”ENCY
MODE BURST BURSTS 003
MODE determines whether the message will be output once or multiple times when
SINGLE: BURST:
CONT:
AMPLITUDE sets the output signal level of the carrier.
BURSTS sets the number of times the message will be transmitted. This parameter is displayed only when the MODE parameter is set to
BURST. The allowable range is 1 to 256.
Outputs the message once. Outputs the message the number of times specified in the BURSTS field (3 times in this example). Outputs the message continuously until the
(lNCR]
time. (Note: The collapse cycle described later on in this section will determine how often the message will be transmitted per cycle.)
-
@iK?‘K]
(START/STOP) key is pressed a second
MOD”LATlcm
(START/STOP) is pressed.
-
AMFUTUDE O.OdBm
AMP‘lTUDE
lb-14 Operation Reference
RF OFF is displayed below the amplitude setting when the RF output is turned off. To toggle the RF signal output on and off, press
(EKGipFON/OFF).
HOLD is displayed when the step attenuator is held at its current range setting. To toggle the attenuator hold function on and off, press
[ATTNJ
Function
Another menu is displayed as follows for completing the Encoding
Mode:
IMMEDIATE STOP OFF
HEADER ON
TERMINATOR ON
IMMEDIATE STOP terminates pager signaling as follows when a stop event occurs.
/
Note
ON: OFF:
Terminates pager signaling immediately.
Default setting. Terminates pager signaling just before the next instance the pager is on. (If TERMINATOR is set to ON, the pager encoder generates the re-synchronization pattern in the frame that was supposed to contain the message.)
HEADER sets whether the idle frame signal is output in the frame before the first message.
ON: OFF:
Default setting. Outputs the idle frame signal. Does not output the idle frame signal.
TERMINATOR sets whether the re-synchronization pattern is output after the last message. The collapse cycle will determine in which frame the re-synchronization pattern appears after the
@iKZQ
(START/STOP) key is pressed to stop or at the end of a burst
sequence (refer to “FLEX/FLEX-TD Signaling Examples”). ON:
Default setting. Outputs the re-synchronization pattern.
OFF:
Does not output the re-synchronization pattern.
If TERMINATOR is set to OFF, the pager under test is not asynchronized. Therefore, the pager can not be synchronized with the future page until the re-synchronization pattern is sent. To send the re-synchronization pattern, set TERMINATOR to ON in this menu or select RESYNC in the format menu.
Operation Reference lb-15
Function
Note
Entering the Pager
Capcode
(Address)
The pager code menu looks like the following:
PAGER CODE A0012477
cncode40.cdr
n PAGER CODE accepts alphanumeric characters up to 16 digits.
Use the numeric keys to enter the alphabetical character, press
(j-1
capcode.
To select an
(SHIFT) before the numeric key associated with the desired alphabetic character. Terminate your entry with the
A FLEX/FLEX-TD pager’s information. Entering the
[MHz/dB(mZ)
(ENTER) key.
capcode
capcode
in the PAGER CODE data field
contains address and protocol
automatically sets up the ADDRESS TYPE, ADDRESSI, and ADDRESS2 in the address menu, and the FRAME, PHASE, and COLLAPSE CYCLE in the protocol menu as well.
The following table shows an example of the conversion from
capcodes
to the parameters. Note that the values of COLLAPSE CYCLE
and FRAME are changed.
Capcode
A0012477
2A0012477
1272A0012477
n DUMMY CALL inserts a message with all 5s in all non-call phases. This
ADDRESS
TYPE
SHORT SHORT SHORT
ADDRESS1 ADDRESS2 FRAME PHASE
0045245 0045245 0045245
None None None
011 011
127
D D D
COLLAPSE
CYCLE
4 2 2
is required by the FLEX-TD standard (RCR STD-43A) for sensitivity testing.
ON:
Default setting. Activates the dummy call function for equalizing the FSK deviation. An address for the non-call phases must be set.
OFF:
Does not activates the dummy call function. Idle
frames are set in the non-call phases.
lb-16 Operation Reference
The address menu will display the following information:
n ADDRESS TYPE selects the address length.
Function
encode41 .Cdr
SHORT: LONG:
w
ADDRESS1 is the 7-digit short address or the primary address of the
Short address Long address
long address.
n ADDRESS2 is the 7-digit secondary address of the long address. (This
is displayed only when ADDRESS TYPE is set to LONG.)
If DUMMY CALL is ON, the following address menu is displayed.
FREQUENC,’ MODULATION
ADDRESS TYPE
ADDRESS1 0032769
w
ADDRESS TYPE (DUMMY CALL) displays the address type used in the
(DUMMYCALL)
-
SHORT
AMPLITUDE
encode23.dtw
non-call phases (automatically set). SHORT:
LONG:
Short address Long address
Note
n ADDRESS1 is the 7-digit short address or the primary address of the
long address.
n ADDRESS2 is the 7-digit secondary address of the long address. (This
is displayed only when ADDRESS TYPE (DUMMY CALL) is LONG.)
To equalize the FSK deviation, the addresses for DUMMY CALL should
contain an equal number of “1” and “0”.
Operation Reference lb-17
Function
Setting the Protocol
The protocol menu displays the following information when
“A0000001”
is the previously entered
capcode.
I
1’gMRl
I-
/
/
0 non 000
I
mEPU=NCY B
CYCLE 00
COLLAPSE
CYCLE 0
MODUI.A~ION -
FFL4ME
000
AMPL,,,,~
PHASE A REPEAT 0
EO
These parameters, except for CYCLE, are all automatically set when the
capcode
n CYCLE: Defines the start cycle. The allowable range is 0 to 14.
w
FRAME: Defines the start frame where the message is located. The
allowable range is 0 to 127.
n PHASE: Defines the phase for the pager. The allowable range is A, B,
C, or D.
w
COLLAPSE CYCLE: Determines how many times a message repeats
within a cycle when BURST or CONT is active in the encoding MODE setting. The default setting of 4 will repeat the message once every
16 frames
n REPEAT: Displayed only if FLEX-TD is selected in the format menu.
Defines the repetition transmission value. Determines how many cycles the message will be repeated. The allowable range is 0 to 3.
is entered in the PAGER CODE data field:
(24).
The allowable range is 0 through 7.
an613abc
1 b-l
8 Operation Reference
Setting the Roaming Mode
Function
FREQUENCY-
ROAMING MODE NONE
n ROAMING MODE determines the mode in which a pager is tested.
NONE: SSID:
Tests a pager without the roaming mode. Tests a pager in the SSID mode which responds to
MODULATION -
AMPLmUDE
encode24.drw
a simulcast transmission system. SSID consists of three words as follows:
q
SSIDl (BIWOOO):
Local Channel ID Coverage Zone
•I
SSIDB (BIWlll):
Country Code Traffic Management Flag
q Frame Offset (BIWlOl)
SSID, NID:
Tests a pager in the SSID and NID modes. The NID mode, in addition to the SSID mode, is a roaming network for covering wider multiple areas. NID consists of a network address and a short message vector as follows:
•I SSIDl (BIWOOO): Local (Channel) ID
Coverage Zone
q
SSID2
(BIWlll):
Country Code SSID Traffic Management Flag (TMF)
q Frame Offset (BIWlOl)
q
NID
Network Address: 21 bits Short Message Vector:
RF Channel (M) Multiplier (Service) Area NID Traffic Management Flag (TMF)
Operation Reference
1 b-l 9
Function
Note
When ALPHANUMERIC is selected in the PAGER TYPE field and SSID, NID is selected in the ROAMING MODE field, the maximum allowable range of the MESSAGE LENGTH field is limited to 35.
The following menu is displayed for setting the SSID mode for a pager to be tested:
encode25.drw
n LOCAL ID: Defines the local channel ID of a pager under test. The
allowable range is 0 to 511.
n COVERAGE ZONE: Defines the coverage zone of a pager under test.
The allowable range is 0 to 3 1.
n COUNTRY CODE: Defines the country code. The allowable range is
0 to 1023.
n SSID TMF: Defines the SSID traffic management flag. This value is
automatically set when the
capcode
(address) is entered in the
PAGER CODE data field. The allowable range is 0 to 15.
After completing the previous menu, set FRAME OFFSET in the following menu:
n FRAME OFFSET: Defines the number of frames to be offset by
BIWlOl from the signaling frame at the home area. The allowable
range is 0 to 63.
lb-20 Operation Reference
Function
The next menu is displayed for setting the NID mode after completing
the SSID mode settings only if the SSID and NID modes are selected:
encode27,drw
n ADDRESS: Defines the network address which appears on the RF
channel within the address field in predicted frames. The allowable range is 2058240 to 2062335.
w
RF CH: Represents each RF channel obtained by the formula as
follows: RF CH = Modulo 8 of Integer [(RF Freq. kHz)/(channel space
kHz)]
The allowable range is 0 to 7.
n AREA: Defines the service area identification. The allowable range is
0 to 31.
n MULTIPLIER: Defines how much the network address is extended.
The allowable range is 0 to 7.
n NID TMF: Defines NID traffic management flag. This value is
automatically set when the
capcode
(address) is entered in the
PAGER CODE data field. The allowable range is 0 to 15.
Operation Reference lb-21
Function
Message During Encoding
To start and stop encoding after selecting all pager encoder parameters, press the
display is shown during encoding a pager under test:
@Ki7ZKSET_)
(START/STOP) key. The following
encode28.drw
Note
Note
REPEAT is displayed additionally only if FLEX-TD is selected in the
format menu.
If MESSAGE NO. is set to 0, the display is changed as follows:
encode29.drw
If the RF output is turned off, AMPLITUDE x .
xdBm
will be substituted for RF OFF. While encoding, the blinking cursor is on the first character A of AMPLITUDE or R of RF OFF.
1 b-22
Operation Reference
Function
Signaling Examples
The following examples illustrate the frame structure for the three transmission repetition settings. The HEADER and TERMINATOR parameters are set to ON (default), and the COLLAPSE CYCLE (CC) is set to 1 and 2 (messages will be repeated every 2 frames (2l) and
4 frames
n SINGLE Mode Example
(22)
respectively).
Start Cycle: 10 Start Frame:
127
Collapse Cycle: Has no effect
q Since HEADER is set to ON, the idle signal is output at Frame 126. q With TERMINATOR set to ON, the asynchronous state signal is
output after the message.
F
CC= 1 idle message asynchronous
ICC=2 Iidle)messageIasynchronousI
n BURST Mode Example
Start Cycle: 10 Start Frame: Collapse Cycle: 1 (2l) and 2 Burst:
q Since HEADER is set to ON, the idle signal is output at Frame 126. q With TERMINATOR set to ON, the asynchronous state signal is
output in the frame where the next message was supposed to occur (as determined by the collapse cycle).
Cycle 10 Frame 126
H-t
CC= 1 idle message idle message idle message idle
CC=2 idle message idle idle idle message idle
10
127 0
11
11
1
11
2
11
3
127
3
11
4
11
5
asyn-
chronous
idle
(22)
11 11
6 7
idle message
11 11
asyn-
chronous
Operation Reference
1 b-23
Function
n CONT Mode Example
Start Cycle: 10 Start Frame:
127
Collapse Cycle: 1 (2l) and 2
(29
stop: Selected in frame f+ 1
0
Since HEADER is set to ON, the idle signal is output at Frame 126.
cl
With TERMINATOR set to ON and IMMEDIATE STOP set to OFF, selecting stop will replace the next message with the asynchronous signal.
c
Cycle 10 10 11 11 11
0
1
2
message idle message idle
idle
CC=2 idle message
I I
idle
idle
idle . .
message idle
C C C C
f+l
f
f+2
asyn-
chronous
idle idle asyn-
f+3 f+4
chronous
1 b-24
Operation Reference
Function
SSID/NID
Roaming Example
When the roaming function with SSID and NID is set to on with the following conditions, the contents in each phase are shown in the following table:
n 6400 bps, 4-level FSK n Call in Phase: A n Address Type: Long Address
n Roaming Mode: SSID, NID
(for Network Address)
IB
Idle
1 BIW 101 System Messages: Time Instruction,
Seconds Adjust: 0, Daylight Savings: No (Standard Time), Local Time Zone: Greenwich Meridian
2 Vector Type: Short Message/Tone Only,
Message Type: 3 Numeric Characters (000) with Short Address
Idle Idle
where, BI:
AF: VF: MF: IB:
Block Information Address Field Vector Field Message Field Idle Block
Operation Reference
1 b-25
Function
If the firmware revision of the HP 8648A Option
1EP
is B.04.08 or
below, the contents in each phase are shown in the following table:
n 6400 bps, 4-level FSK
n Call in Phase: A n Address Type: Long Address
w
Roaming Mode: SSID, NID
Dummy Address 2
block 2
where,
BI: AF:
VF: MF:
IB:
IB
Idle
Block Information Address Field Vector Field Message Field Idle Block
Dummy Message
Dummy Message
Idle
Dummy Message
Dummy Message
Idle
1 b-26
Operation Reference

POCSAG

Function
Setting the Data Rate and Pager Type
The parameter menu for the data rate and the pager type looks like
the following:
==-7 m "-
n
DATA RATE
512:
1200:
2400:
n
PAGER TYPE
NUMERIC: TONE ONLY: ALPHANUM 7BIT:
ALPHANUM 8BIT:
ALPHANUMERIC:
FREQUENCY-,
DATA RATE PAGER TYPE
5 12
512 bps 1200 bps 2400 bps
Numeric type Tone only type 7-bit coded alphanumeric type entered by hex-binary code
8-bit coded alphanumeric type entered by
hex-binary code 7-bit coded alphanumeric type
NUMERIC
MODULATION -
AMPLITUDE
FUNCTION 00
encode32,tif
Operation Reference lb-27
Function
Note
The ALPHANUM 7BIT and ALPHANUM 8BIT types can be used for displaying two-byte special characters, for example Chinese characters, through a conversion table.
The following examples show how the actual bit streams are transmitted if
q
For ALPHANUM 8BIT
“BOAl”
is set to the message (MESSAGE NO. 6):
First Byte Second Byte
Bit number Bit streams
Characters
q
For ALPHANUM
MSB
b8 b7 b6 b5 b4 b3 b2 bl b8 b7 b6 b5 b4 b3 b2 bl
1011000010100001
B 0
7BIT
LSB MSB
A
1
The eighth bit (b8) is ignored as follows:
First Byte
MSB
LSB MSB
Second Byte
LSB
LSB
1
Bit number
b7 b6 b5 b4 b3 b2 bl I b7 b6 b5 b4 b3 b2 bl Bit streams 01100000100001 Characters
B
(3)
0
A
(2)
1
When converting the binary data stream to 2-level FSK, the first bit out becomes the least significant bit (LSB) of the 2-level symbol with the following bit the most significant bit (MSB) of the symbol.
n FUNCTION determines the pager’s response depending on the pager
type. The allowable range is 00 to 11.
1
1 b-28
Operation Reference
Setting the Message
The message menu looks like the following:
Function
MESSAGE NO.1
0123466[0]
n MESSAGE NO. is defined as the following where “X” represents the
MESSAGE LENGTH 40
phase setting:
0:
Arbitrary message set from the external controller using the HP-IB capability. The data must be entered in units of words. Refer to “
[SOURce]: PAGing
:ARBitrary :DEFine
<NRl>,
[
:
FORMat]
CNRl>,
. . . ,
CPOCSag I
<NRl>”
FLEX 1
FTD)
and “Using
:
the Buffer Memory for the Arbitrary Messages” in Chapter 2. This arbitrary message buffer is divided into multiple segments
by assigning START BATCH and STOP BATCH in the following menu:
r
I
FREQUENCY
START BATCH 000
I
B
~~~
-
STOP BATCH 127
AMlJlUDE
I
ellcodc44.cdr
I
The allowable ranges for both are 000 to 127. Displays: 0123456
1:
C-1
2: Displays: ABCDEFG [PI 3: Displays: TEST PAGING: POCSAG 4: Displays: ALPHANUMERIC 40CHARS TEST PAGING : POCSAG 5: Displays: 8888888888888888888888888888888888888888
for LCD test with 40 characters of “8”.
6:
User-definable message, up to 40 characters long, set from the front panel. Enter numeric messages with the keypad. Alphanumeric messages require the
(-1
(SHIFT) key to access the alphabetical characters (blue letters) above the keypad. If required alphabetical characters are not available on the front panel, send a message from an external
Operation Reference
1 b-29
Function
controller over HP-IB. Refer to “
:
CPOCSag
n MESSAGE LENGTH specifies the maximum number of characters that
I FLEX I
FTD) :MESS
[SOURceI
:
PAGing C
:
FORMat]
: DEFine ‘string”’ in Chapter 2.
will be sent out and displayed in the message field. The allowable range is 1 to 40. After specifying the number of characters, press
ljdB(mL)
(ENTER) to save the new value.
When the value of this parameter is equal to or greater than the length of the message, the whole message will be sent out.
When the value of this parameter is less than the length of the message, only the characters specified by this parameter are output. (For example, if the MESSAGE NO. is 1(0123456[-1) and the MESSAGE LENGTH is 5, the message which is displayed and will be sent out is actually 01234.)
lb-30 Operation Reference
Setting the Encoding Mode
Function
I
pijti
The number of times a message transmission are determined by
I 1
I
FREQUENCY
MODE BURST BURSTS003
MODE determines whether the message will be output once or multiple times when
SINGLE: BURST:
Outputs the message once. Outputs the message the number of times specified in the BURSTS field (3 times in this example).
CONT:
Outputs the message continuously until the
:yJ
AMPLITUDE sets the signal output level of the carrier.
is transmitted and the level of that the following menu:
1-1
[SET)
MOD”LATlON
AMPLITUDEO.OdE!m
(START/STOP) is pressed.
(START/STOP) key is pressed a second
BURSTS sets the number of times the message will be transmitted.
This parameter is displayed only when the MODE parameter is set to BURST. The allowable range is 1 to 255.
RF OFF is displayed below the amplitude setting when the RF output is turned off. To toggle the RF signal output on and off, press
(jj-.
HOLD is displayed when the step attenuator is held at its current range setting. To toggle the attenuator hold function on and off, press
CATTNj.
Operation Reference lb-31
Function
Entering the Pager
The pager code menu looks like the following:
\
mmiii\
n PAGER CODE accepts only numeric characters up to 7 digits.
Use the numeric keys to enter the with the
Message During Encoding
To start and stop encoding after selecting all pager encoder
parameters, press the display is shown during encoding a pager under test.
fjj]
Capcode
FREQUENCY -
PAGER CODE
(ENTER) key.
[iKZKK]
(Address)
MODULATION -
0000000
capcode.
(START/STOP) key. The following
Terminate your entry
AMPLKUDE
Note
$
If the RF output is turned off, AMPLITUDE x . for RF OFF. While encoding, the blinking cursor is on the first character A of AMPLITUDE or R of RF OFF.
FREQ-CY
MESSAGE NO. 6
MODULATION
AMPLl’WDE
AMPLITUDE O.OdBm
xdBm
will be substituted
encode34.dw
1 b-32
Operation Reference
1 n

RESYNC

mm/r, 6
tir,‘n\--F=Q-C”-
Function
Setting the Encoding Mode
The following parameters must be set to activate the resynchronization function for a FLEX/FLEX-TD pager under test:
.
M=‘D”LJJ”ON-
BURSTS 001 AMPLITUDE
Ah4PLrr”DE
O.odBm
encode38.drw
I
I
I
Note
Note
n BURSTS defines the number of data to make a pager
resynchronized
with this pager encoding signal. One burst is 1.92 s. The allowable range is 000 to 256. 0 represents the continuous mode.
This setting is required only when RESYNC is selected in the format menu. The @) (PREV) and
($‘J
(NEXT) keys toggle the format menu
and this BURSTS menu. No other menus are displayed.
Message During
Resynchronizing
To start and stop encoding after selecting all pager encoder parameters, press the
RESYNC SENDING
I
I
If the RF output is turned off, AMPLITUDE x .
[‘NCR)
FREQUENCY-
(START/STOP) key.
MODULATION -
xdBm
AMPLITUDE-
I
encode37.dr.v
will be substituted
for RF OFF. While encoding, the blinking cursor is on the first character A of AMPLITUDE or R of RF OFF.
I
Operation Reference
1 b-33
Function

PN15

Setting the Data Rate
If PNl5 modulation is selected in the format menu, the following parameters must be set to complete the pager encoder setup process.
encode30,drw
H
DATA RATE
> ~~~g~l;~;;~ar;
\
(
‘1
Note
512/2:
1200/2:
1600/2: 2400/2: 3200/2: 3200/4: 6400/4:
512 bps, 2-level FSK
1200 bps, 2-level FSK
1600 bps, 2-level FSK 2400 bps, 2-level FSK 3200 bps, 2-level FSK 3200 bps, 4-level FSK 6400 bps, 4-level FSK
This setting is required only when PNl5 is selected in the format menu. The m (PREV) and
(IJJ
(NEXT) keys toggle the format menu
and this PN15 FSK data rate menu. No other menus are displayed.
Message During Encoding
To start and stop encoding, press the
/
-
IIY
FREQUENCY-
I
PN15 SENDING
@ZCZY]
MODULATlON -
(START/STOP) key.
AMP
AMPLITUDE
O.OdBm
I
L
encode31
J
Aif
1 b-34
Note
Operation Reference
If the RF output is turned off, AMPLITUDE x .
xdBm
will be substituted for RF OFF. While encoding, the blinking cursor is on the first character A of AMPLITUDE or R of RF OFF.
Function

SERVICE

This function is used for servicing the HP 8648A Option lEP, and provides a continuous FSK signal.
Setting the Data Rate
The following parameter must be set to activate the service function:
FORMAT SERVICE
encodc45.cdr
n CALIBRATION FSK is one of the data rates as follows:
512/2:
1200/2: 1600/2:
2400/2:
3200/2: 3200/4: 6400/4:
512 bps, 2-level FSK 1200 bps, 2-level FSK 1600 bps, 2-level FSK 2400 bps, 2-level FSK 3200 bps, 2-level FSK 3200 bps, 4-level FSK 6400 bps, 4-level FSK
Note
Note
This setting is required only when SERVICE is selected in the format menu. The @ (PREV) and
(iJJ
(NEXT) keys toggle the blinking cursor between the format parameter and this CALIBRATION FSK parameter. No other menus are displayed.
Message During Servicing
To start and stop encoding, press the
If the RF output is turned off, AMPLITUDE x. for RF OFF. While encoding, the blinking cursor is on the first character A of AMPLITUDE or R of RF OFF.
c-1
(START/STOP) key.
xdBm
will be substituted
Operation Reference
1 b-35

Pulse Modulation

Pulse Modulation
Note

1. PULSE

Pulse modulation is valid only for instruments with Option
~FUNCTION
[FREOUENCI]
[Z)
Eel
BIcl
I
funckeys.drw
-1
4
lE6.
To select pulse modulation, press (AM) (PULSE) twice. If RF is on, the
display will look like the following representation. The
CMOD
ON/OFF]
key enables or disables the pulse modulator.
Note
When the instrument is in the pulse modulation mode, pulse modulation should be turned off using
CMOD
ON/OFF) before switching
off the instrument. If the instrument is switched off while the pulse modulation mode is turned on, the instrument will power up with no RF output.
PULSE MOD
PULSE MOD
REARIN
OF’F
The word PULSE is shown in the AMPLITUDE portion of the display
when pulse modulation is enabled ([MOD ON/OFF]) and the RF is turned on
(c-1).
In this case, no RF will be present at the RF output unless a TTL-high signal is applied to the rear-panel PULSE IN connector.
1 b-36
Operation Reference
Pulse Modulation
The following keys are invalid when the pulse modulation screen is visible:
n
@iTxiFq
n
@iFiiq
n @EC--
D
pm?--
n
[lkHz+EXT DC)
Operation Reference lb-37

Increment Set

Increment Set

1. INCR SET

Press
[SET_)
to view or change the increment set value for the
currently active function (the active function is the last function key
you press;
press
(FREQUENCY],@~~~TEF],(FMI,(AM),
[SET),
0 is displayed between the value and the units. The
or@iiJ When YOU
indicates that the displayed value is the increment set value.
Q
Note
Note
An increment value cannot be set for the knobs or the memory recall arrow keys.
Incr 2ment
Function
Set Ranges
Range
1 Hz to 999.75 MHz
> 0.0 to 149.0 dB (HP 8648A)
FM Deviation AM Denth
> 0.0 to 150.5 dB (HP > 0.0 to 100 kHz > 0.0 to 100%
8648B/C/D)
> 0.0 to 10.0 Radians
0.01 to 1.0 kHz
It is possible to set an increment value of greater resolution than can be displayed or than the hardware can respond to. However, the instrument records each arrow key press and will respond after the
appropriate number of presses.
START/STOP Encoding
1 b-38
Operation Reference
If Option mode, the
1EP
is present and the signal generator is in the ENCODER
(jj)
key functions as a START/STOP key to start or
stop the pager encoding when in the pager encoder mode.
PREV and NEXT
If Option
1EP
is present and the signal generator is in the ENCODER mode, the @) and @J keys function as the PREV and NEXT keys to change the blinking first character between each parameter when in the pager encoder mode.

Data

Data
datakeys.drw
1. MHz/dBm
kHz/mV
%/pV
rad/dBpV
Note
\
\
4
Press a units key after you enter a value. This terminates the entry.
Note that the units keys in the left column are each labeled with an amplitude unit on the bottom and a frequency or modulation unit on the top. The instrument applies the appropriate unit for the function value you are entering. The bottom key in the row for instance, terminates a $M entry in radians or an amplitude entry in
Memory register selections, sequence selections, and HP-IB address entries do not require a units key to terminate the entry. These entries are automatically terminated after the last digit is entered.
dBpV.
Units Conversion
You can change the units of the displayed frequency or amplitude value by selecting the FUNCTION (frequency or amplitude) and then pressing a units key. The instrument will convert the displayed value to the equivalent value for the units key you pressed.
ENTER
If Option 1EP is present and the signal generator is in the ENCODER mode, the key must be used to store any numeric or alphabetic characters entered by way of the DATA and MODULATION source blocks.
If Option 1EP is present and the signal generator is in the ENCODER mode, the input alphabetical characters using the DATA and MODULATION SOURCE blocks when in the pager encoder mode.
C-0
[GZiZ7]
key functions as an ENTER key. The ENTER
key functions as a SHIFT key. This key lets you
Operation Reference lb-39

2. Backspace

Press
1--)
when entering a numeric value to backspace and remove
the last digit entered.

3. emf

4.
f
Press these keys to display the amplitude value indicated on the key label in electromotive force units. Emf is the RF output voltage with no load. It is twice the output voltage with a 50 ohm load.
Press this key at any time while you are entering an amplitude or reference offset value to change the sign of the value.
1 b-40
Operation Reference

Instrument Preset

Instrument Preset
CPOWER)@
(POWER)
ET)
Note
Turn the instrument on while pressing the backspace key perform an instrument preset. The instrument will power up to factory-defined settings shown in the following table. Save and recall registers are not affected by this operation.
Turn the instrument on while pressing the memory perform a clear memory. This function erases all save/recall registers, sets the HP-IB address to 19, and performs an instrument preset where the instrument powers up to factory-defined settings shown in the following table.
This will cause an error message to appear on the display: 627 Battery RAM failure
: memory lost.
This is normal.
IDEL)
(@)
key to
to
Operation Reference lb-41
Loading...