The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the furnishing, performance, or
use of this material.
Certification
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute’s calibration facility, and to the
calibration facilities of other International Standards Organization members.
Regulatory Information
The specifications and characteristics chapters and General Regulatory and Safety
Information chapter in this manual contain regulatory information.
Warranty
This Agilent Technologies instrument product is warranted against defects in material
and workmanship for a period of one year from date of shipment. During the warranty
period, Agilent Technologies will, at its option, either repair or replace products which
prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent
Technologies and Agilent Technologies shall pay shipping charges to return the product to
Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products
returned to Agilent Technologies from another country.
2
Agilent Technologies warrants that its software and firmware designated by Agilent
Technologies for use with an instrument will execute its programming instructions when
properly installed on that instrument. Agilent Technologies does not warrant that the
operation of the instrument, or software, or firmware will be uninterrupted or error-free.
IMITATIONOF WARRANTY
L
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
XCLUSIVE REMEDIES
E
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE
REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Cleaning
The instrument front and rear panels should be cleaned using a soft cloth with water or a
mild soap and water mixture.
3
General Safety Considerations
WAR NINGBefore this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to
a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or
outside the instrument, or disconnection of the protective earth
terminal can result in personal injury.
WAR NINGThere are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed
only by trained service personnel.
CAUTIONBefore this instrument is switched on, make sure its primary power circuitry
has been adapted to the voltage of the ac power source.
Failure to set the ac power input to the correct voltage could cause damage to
the instrument when the ac power cable is plugged in.
Differences between Agilent Technologies 8560
E-Series and EC-Series Spectrum Analyzers
1. EC-series analyzers use a color LCD display which is not adjustable. E-series
instruments use a monochrome CRT display which can be adjusted for focus and
intensity. E-series analyzers use a high-voltage module (HVM), located in the power
supply, to supply power to the CRT. EC-series analyzers do not use an HVM.
2. EC-series analyzers offer fast analog-to-digital-conversion (FADC) digitized sweep
times as a standard feature. EC-series FADC circuitry is integrated into the A2
controller board. E-series instruments offer FADC feature as an option (Option 007)
which is implemented through the use of the A16 FADC board. FADC enables you to
view displays with fast sweep times, between 30 ms and 50
3. EC-series instruments offer a rear-panel VGA connection as a standard feature.
E-series instruments do not have a rear-panel VGA connector. The VGA connection is
always active and does not require user activation.
In all other operational respects the 8560 EC-series and E-series are identical.
4
µs.
Contents
1. Using Operation Verification Software − Volume I
This automated test software is designed to give a high confidence level
of spectrum analyzer operation in a reasonable amount of time.
Instructions for using the software, brief descriptions of the tests, and a
list of the software error messages are included in this chapter. Refer to
Table 1-3, on page 20, for the names of manual performance tests that
are not included in the automated operation verification software.
What Is Operation Verification?
Operation Verification automates many of the performance tests. It is
recommended as a check of instrument operation for incoming
inspection or after a repair. It performs 80 to 85 percent of the manual
performance tests in less than 60 minutes (75 minutes for the Agilent
8564E/EC and Agilent 8565E/EC). It is designed to test an instrument
operating within a 20
set of test equipment. Refer to Table 1-1, on page 16 for a list of tests
performed.
°C to 30 °C temperature range using a minimum
If a test does not pass, the related manual performance test needs to be
run. Related manual performance tests are listed at the top of each
operation verification test description and the test procedures are in
“Using Performance Tests” Chapter 10, of Volume II of the calibration
guide.
Operation Verification versus Performance Tests
The performance tests verify that the analyzer performance meets all
specifications. Performance verification consists of executing all of the
manual performance tests. It is time-consuming and requires extensive
test equipment.
12Chapter 1
Using Operation Verification Software − Volum e I
Getting Started
Getting Started
First, make sure you have a compatible controller (computer), the
proper test equipment, and a printer for recording test results. The
following paragraphs describe requirements for controllers, test
equipment, and printers. Once the proper equipment is identified,
proceed to “Equipment Connections.”
Spectrum Analyzers
The 8560 E-Series and EC-Series Operation and Verification software
tests the following spectrum analyzers:
Spectrum Analyzer Software Revision
8560E/ECall
Agilent 8561E/EC
Agilent 8562E/EC
Agilent 8563E/EC
Agilent 8564E/EC
Agilent 8565E/EC
all
D.00.00 and later
all
C.00.00 and later
C.00.00 and later
Errors will occur if this software is used to test spectrum analyzers
other than those listed above. Other versions of the operation
verification software are available for testing other 8560 Series
spectrum analyzers.
Controller (Computer)
Operation Verification software requires using any combination of one
of the following controllers and the HP BASIC operating system:
Controller
9000 model 216
9000 model 236
9000 model 310
Chapter 113
Using Operation Verification Software − Volu me I
Getting Started
Operating System
HP BASIC 2.0 with extensions 2.1
HP BASIC 3.0 and required BIN files
HP BASIC 4.0 and required BIN files
Other 9000 Series 300 controllers are compatible with the Operation
Verification software; however, the graph printouts might not be full
width. This is especially true with medium- and high-resolution
displays.
The Operation Verification program requires at least 500K of free
memory. The computer can have either single or dual GPIB ports. Refer
to “Dual-Bus Operation,” on page 28, under “Using Operation
Verification,” on page 22, for information on using the program with
dual GPIB ports.
Test Equipment
Table 1-1, on page 16, lists the operation verification tests and the test
equipment required for each test. You do not need all the test
equipment connected to perform operation verification. You need only
connect the equipment specified in each test to run that test.
NOTEThe Agilent 3335A Synthesizer Level Generator, which is used in
several operation verification tests, has been obsoleted. If an Agilent
3335A is not available, alternate test equipment can be used to perform
comparable tests manually. Chapter 8 in this volume contains the
information on alternate test equipment and manuals test procedures
you will need to perform these tests.
Table 1-2, on page 19, summarizes the equipment required to run the
Operation Verification tests. Some tests, like 10 MHz Reference
Accuracy, can use various model numbers of a particular equipment
type. Information about selecting the equipment model number you
want to use is provided in “Setting GPIB Addresses,” on page 25, in this
chapter under “Using Operation Verification,” on page 22.
NOTEThe validity of Operation Verification program measurements depends
in part on required test equipment measurement accuracy. Verify
proper calibration of test equipment before testing the analyzer with
the software.
14Chapter 1
Using Operation Verification Software − Volum e I
Getting Started
Printers
All test results are sent to a GPIB printer. The program does not run
without being connected to a GPIB printer. Virtually any GPIB
graphics workstation printer can work. These tests have been run using
the HP ThinkJet, 2671G, 82906A, and 9876G printers.
Warmup Time
Test Equipment Warmup Allow sufficient warmup time for test
equipment. Refer to their individual operating and service manuals for
warmup specifications.
Spectrum Analyzer Warmup Warm the spectrum analyzer up for at
least five minutes before performing the first test.
NOTESince the preselector in the Agilent 8561E/EC, Agilent 8562E/EC,
Agilent 8563E/EC, Agilent 8564E/EC and Agilent 8565E/EC spectrum
analyzers is not peaked during the Frequency Response test, allow a 30
minute warmup at 20
°C to 30 °C before beginning this test. If you start
running All Tests after five-minute warmup, the spectrum analyzer will
have warmed up for 30 minutes by the time the Frequency Response
test begins.
Chapter 115
Using Operation Verification Software − Volu me I
Getting Started
Table 1-1 Equipment Required for Each Operation Verification Test
Test Name Equipment Required
10 MHz
Reference
Accuracy
1ST LO
OUTPUT
Amplitude
Accuracy
Calibrator
Amplitude
Accuracy
Displayed
Average
Noise Level
Fast Sweep
Time
Accuracy
Agilent
8560E/EC
5342A,
5343A, or
5350B*
8902A or
436A or
438A
8485A or
8481A
8902A,
436A or
438A
8481A, or
8482A
909D909D909D909D85138B85138B
5342A,
5343A, or
5350B**
Agilent
8561E/EC
5342A,
5343A, or
5350B*
8902A or
436A or
438A
8485A or
8481A
8902A,
436A or
438A
8481A, or
8482A
5342A,
5343A, or
5350B**
Agilent
8562E/EC
5342A,
5343A, or
5350B*
8902A or
436A or
438A
8485A or
8481A
8902A,
436A or
438A
8481A, or
8482A
5342A,
5343A, or
5350B**
Agilent
8563E/EC
5342A,
5343A, or
5350B*
8902A or
436A or
438A
8485A or
8481A
8902A,
436A or
438A
8481A, or
8482A
5342A,
5343A, or
5350B**
Agilent
8564E/EC
5342A,
5343A, or
5350B*
8902A or
436A or
438A
8485A or
8481A
8902A,
436A or
438A
8481A, or
8482A
5342A,
5343A, or
5350B**
Agilent
8565E/EC
5342A,
5343A, or
5350B*
8902A or
436A or
438A
8485A or
8481A
8902A,
436A or
438A
8481A, or
8482A
5342A,
5343A, or
5350B**
Frequency
Readout/
Counter
Accuracy
Frequency
Span
Accuracy
8340A/B or
83640A or
83650A
8120-4921
8340A/B or
83640A or
83650A
11667A/B
8120-4921
8340A/B or
83640A or
83650A
8120-4921
8340A/B or
83640A or
83650A
11667A/B
8120-4921
8340A/B or
83640A or
83650A
8120-4921
8340A/B or
83640A or
83650A
11667A/B
8120-4921
8340A/B or
83640A or
83650A
8120-4921
8340A/B or
83640A or
83650A
11667B
8120-4921
83640A or
83650A
8120-6164
83640A or
83650A
11667C
8120-6164
83650A
8120-6164
83650A
11667C
8120-6164
16Chapter 1
Using Operation Verification Software − Volum e I
Table 1-1 Equipment Required for Each Operation Verification Test
Test Name Equipment Required
Getting Started
Frequency*
Response
IF Gain*
Uncertainty
Image and
Multiple
Responses
Agilent
8560E/EC
8340A/B or
83640A or
83650A
3335A
8902A or
436A or
438A
8482A
11667A/B
8120-4921
3335A3335A3335A3335A3335A3335A
8340A/B or
83640A or
83650A
Agilent
8561E/EC
8340A/B or
83640A or
83650A
3335A
8902A or
436A or
438A
8481A
11667A/B
8120-4921
8340A/B or
83640A or
83650A
Agilent
8562E/EC
8340A/B or
83640A or
83650A
3335A
8902A or
436A or
438A
8481A
11667A/B
8120-4921
8340A/B or
83640A or
83650A
Agilent
8563E/EC
8340A/B or
83640A or
83650A
3335A
8902A or
436A or
438A
8485A
11667B
8120-4921
8340A/B or
83640A or
83650A
Agilent
8564E/EC
83640A or
83650A
3335A
8902A or
436A or
438A
8487A
11667C
8120-6164
83640A or
83650A
Agilent
8565E/EC
83650A
3335A
8902A or
436A or
438A
8487A
11667C
8120-6164
83650A
Input*
Attenuator
Switching
Uncertainty
Noise
Sidebands
RES BW*
Accuracy/
Selectivity
8902A or
436A or
438A
8481A
11667A/B
8120-4921
3335A3335A3335A3335A3335A3335A
8662A or
8663A or
CAL
OUTPUT
signal
3335A3335A3335A3335A3335A3335A
8902A or
436A or
438A
8481A
11667A/B
8120-4921
8662A or
8663A or
CAL
OUTPUT
signal
8902A or
436A or
438A
8481A
11667A/B
8120-4921
8662A or
8663A or
CAL
OUTPUT
signal
8902A or
436A or
438A
8485A
11667B
8120-4921
8662A or
8663A or
CAL
OUTPUT
signal
8902A or
436A or
438A
8487A
11667C
8120-6164
8662A or
8663A or
CAL
OUTPUT
signal
8902A or
436A or
438A
8487A
11667C
8120-6164
8662A or
8663A or
CAL
OUTPUT
signal
Chapter 117
Using Operation Verification Software − Volu me I
Getting Started
Table 1-1 Equipment Required for Each Operation Verification Test
Test Name Equipment Required
RES BW*
Switching
Uncertainty
Residual
FM
Scale*
Fidelity
Second
Harmonic
Distortion
Agilent
8560E/EC
Agilent
8561E/EC
Agilent
8562E/EC
Agilent
8563E/EC
Agilent
8564E/EC
Agilent
8565E/EC
3335A3335A3335A3335A3335A3335A
8662A or
8663A
8662A or
8663A
8662A or
8663A
8662A or
8663A
8662A or
8663A
8662A or
8663A
3335A3335A3335A3335A3335A3335A
8340A/B or
83640A or
83650A
8902A or
436A or
438A
8485A or
8481A
11667A/B
0955-0306
8120-4921
8340A/B or
83640A or
83650A
8902A or
436A or
438A
8485A or
8481A
11667A/B
11689A (2
required)
0955-0306
8120-4921
8340A/B or
83640A or
83650A
8902A or
436A or
438A
8485A or
8481A
11667A/B
11689A (2
required)
0955-0306
8120-4921
8340A/B or
83640A or
83650A
8902A or
436A or
438A
8485A or
8481A
11667B
11689A (2
required)
0955-0306
8120-4921
83640A or
83650A
8902A or
436A or
438A
8485A or
8481A
11667C
11689A (2
required)
0955-0306
8120-6164
83650A
8902A or
436A or
438A
8485A or
8481A
11667C
11689A (2
required)
0955-0306
8120-6164
* The 3335A Synthesizer Level Generator, which is used in several operation verification tests, has been
obsoleted. If an 3335A is not available, alternate test equipment can be used to perform comparable tests
manually. Chapter 8 in this volume contains the information on alternate test equipment and manuals test
procedures you will need to perform these tests.
* *The Agilent 5350B is supported only in software revision D.00.00 and later.
18Chapter 1
Using Operation Verification Software − Volum e I
Table 1-2 Required Test Equipment Summary
Type of Equipment Model Number
Controller*Series 200 9000 model 216 ( 9816)
9000 model 236 ( 9836)
or 9000 model 310
Synthesizer/level generator3335A
Getting Started
Synthesized sweeper (10 MHz to 26.5 GHz)
(10 MHz to 40 GHz)
(10 MHz to 50 GHz)
Synthesized signal generator8662A/8663A
Measuring receiver8902A
Power meter436A or 438A (alternate)
Microwave frequency counter5343A
Microwave frequency counter5342A (alternate)
Microwave frequency counter5350B (alternate) †
Power sensor (100 kHz to 4.2 GHz)8482A
Power sensor (50 MHz to 26.5 GHz)
(50 MHz to 50 GHz)
Power sensor (10 MHz to 18 GHz)8481A (alternate for 8560E/EC, 8561E/EC, and
Power splitter (dc to 18 GHz)
(dc to 26.5 GHz)
(dc to 50 GHz)
8340A/B ( 8560E/EC, 8561E/EC, 8562E/EC, and
8563E/EC)
85640A ( 8564E/EC)
85650A ( 8565E/EC)
8485A ( 8560E/EC, 8561E/EC, 8562E/EC, and
8563E/EC)
8487A ( 8564E/EC and 8565E/EC)
8562E/EC)
11667A ( 8560E/EC, 8561E/EC, and 8562E/EC)
11667B ( 8560E/EC, 8561E/EC, 8562E/EC, and
8563E/EC)
Miscellaneous cables and adaptersAs per test setup
GPIB printerSee “Printers”
*
500 kilobytes of free memory is required for the test program.
†
Th 5350B is supported only in software revision D.00.00 and later.
909D ( 8560E/EC, 8561E/EC, 8562E/EC, and 8563E/EC)
85138B ( 8564E/EC and 8565E/EC)
Chapter 119
Using Operation Verification Software − Volu me I
Getting Started
Table 1-3 Manual Performance Tests That Are Not Automated
Pulse Digitization Uncertainty
Third Order Intermodulation Distortion
Gain Compression
Sweep Time Accuracy
Residual Responses
IF Input Amplitude Accuracy
Gate Delay Accuracy and Gate Length Accuracy
Delayed Sweep Accuracy
These tests apply only to 8560E/EC Option 002:
Tracking Generator Level Flatness
Absolute Amplitude and Vernier Accuracy
Maximum Leveled Output Power
Power Sweep Range
RF-Power-Off Residuals
Harmonic Outputs
Non-Harmonic Outputs
Tracking Generator Feedthrough
LO Feedthrough Amplitude
Frequency Tracking Range
Tracking Generator Frequency Accuracy
Equipment Connections
Computer (Controller) Setup
For 9000 model 216 or model 236 computers, setup instructions are
provided in Chapter 1, “Computer Installation,” of the BASIC Operating Manual. For 9000 model 310 computers, setup information
is provided in Configuration Reference Manual for Series 300
computers.
20Chapter 1
Using Operation Verification Software − Volum e I
Getting Started
GPIB Cables
All test equipment controlled by GPIB should be connected to the
internal GPIB of the controller (select code 7). If the controller has only
one GPIB connector, connect the spectrum analyzer to it as well. If the
controller has dual GPIB connectors, connect the spectrum analyzer
under test to the second GPIB (typically, select code 8).
10 MHz Reference
The 10 MHz REF IN/OUT on the spectrum analyzer under test should
be connected to the synthesized sweeper external frequency reference.
That is, it should be connected to the FREQUENCY STANDARD EXT
of the 8340A/B or the 10MHz REF INPUT of the 83640A or 83650A.
Connect the synthesized sweeper 10 MHz REF OUTPUT to the
40/N MHz REF INPUT of the 3335A. Do not connect the spectrum
analyzer 10 MHz REF IN/OUT to the external frequency reference
input of the 8663A; doing so invalidates the Noise Sidebands test
results.
NOTETerminate the 3335A 10 MHz REF OVEN OUTPUT in 50 Ω. Do not
connect the 10 MHz output to the external frequency reference input of
any other test equipment.
Test Setups
Test setups for each test are included with the test. These are in the
“Test Descriptions” section of this chapter. The program prompts the
operator to make appropriate equipment connections if the correct
equipment setup is not detected.
Chapter 121
Using Operation Verification Software − Volu me I
Using Operation Verification
Using Operation Verification
Loading the Program
Load BASIC into the computer. BASIC choices are:
• BASIC 2.0 and extensions 2.1
• BASIC 3.0 or 4.0, which must include the following binaries:
MAT
IO
GRAPH
GRAPHX
PDEV
IB
MS
CLOCK
CS80
DISC
KBD
For configuration instructions, refer to the BASIC Operating Manual.
Next, insert the Operation Verification software disk into the disk
drive, then type:
LOAD “VERIFY_6XE”,1
NOTEA double-sided disk drive must be used. The Operation Verification
software will not fit on a single-sided formatted disk.
Press
Series 300 computers, to load the software and start the program
running.
EXECUTE on 9000 Series 200 computers, or RETURN on 9000
22Chapter 1
Using Operation Verification Software − Volum e I
Using Operation Verification
Program Operation
Operation Verification consists of three menus. They are the Conditions
Menu, the Test Menu, and the Sensor Utilities Menu, which are
accessed from the Conditions Menu. Program operation is controlled
through a combination of softkeys and user prompts. Some prompts,
primarily in the Conditions and Sensor Utilities Menus, require
computer keyboard entries. Terminate keyboard entries with the
RETURN or ENTER key. Most prompts, however, tell the user what to do
next or provide informational messages.
If the message (any key) follows a prompt, pressing any key on the
keyboard continues the program. If the message (any keyor ‘Q’ to quit) follows a prompt, pressing any key except Q
continues the program. Pressing Q terminates the current procedure at
the next, most logical point in the program.
Conditions Menu
The first menu screen displayed is the Conditions Menu. The pointer
displayed along the left edge of the screen may be moved with the knob
(if one is present) or the up (
menu has two pages. Moving the pointer below the last entry on the
page brings up the next page. Similarly, moving the pointer above the
first entry on a subsequent page brings up the preceding page. The two
pages of the Conditions Menu have a four-line overlap. The last four
lines of page 1 appear as the first four lines on page 2.
⇑) and down (⇓) arrow keys. Notice that the
Chapter 123
Using Operation Verification Software − Volu me I
Using Operation Verification
Test Record Header Information
The information in the first six entries of this menu is printed out as
part of the operation verification test record. The spectrum analyzer
model number and serial number are stored in the analyzer memory.
Software revisions before C.00.00 will not display the complete list of
options if there are more than two. The program queries these numbers
via GPIB and displays them. If the spectrum analyzer under test does
not respond at the address listed under GPIB Addresses, or no address
is listed, a message appears where the model and serial numbers are
normally displayed.
The program also queries the time and date in the computer. If an 9000
Series 200 computer is used, it might be necessary to reset the time and
date; 9000 Series 300 computers have built-in real-time clocks.
Entries for Operator, Test Conditions, and Other Comments are
optional. Blank spaces are provided on the test record if no entry is
made. To make or to change an entry, move the pointer to the line
where the entry is to be made or changed. Press
Change Entry and type
in your new entry. Entries for Operator, Test Conditions, and Other
Comments can be up to 37 characters long, but only the first 25
characters of the Operator entry are printed on the test record.
System Mass Storage File Location
Calibration factor data for different power sensors and a customized set
of conditions may be stored on disk. The mass storage unit specifier
(msus) for the disk containing this information should be entered as the
system mass storage file location. Refer to the BASIC Operating Techniques Manual for information on the syntax of the msus. Software
revision C.00.00 and later allows a system mass storage file location
with more than 37 characters though only the first and last 17
characters will be displayed.
The Operation Verification program disk comes write-protected from
the factory. If you want to use this disk for storing your power sensor
and conditions data files, it is necessary to disable the write-protect
mechanism.
NOTEA double-sided disk drive must be used. The Operation Verification
software will not fit on a single-sided formatted disk.
24Chapter 1
Using Operation Verification Software − Volum e I
Using Operation Verification
Pow er Sensors
The Operation Verification program supports four models of power
sensors, but only two models are necessary to run all the tests. The
Agilent 8481A may be substituted for the Agilent 8482A. The
Agilent 8485A is required for the Agilent 8562E/EC and
Agilent 8563E/EC, and the Agilent 8487A is required for the
Agilent 8564E/EC and Agilent 8565E/EC. Refer to “Sensor Utilities
Menu,” on page 28, for more information regarding storing, viewing,
editing, and purging cal factor data for power sensors.
To select a particular sensor of a certain model number, move the
pointer to the desired model number and press
Change Entry. Enter the
last five digits of the power sensor serial number (that is, the serial
number suffix). The program checks to see that a data file containing
the cal factor data for that particular sensor exists.
To create, edit, view, or purge power sensor cal factor data files, press
Sensor Utils to bring up the Sensor Utilities Menu. Refer to “Sensor
Utilities Menu,” on page 28, in this chapter for more information.
A WARNING message appears if the program does not find a data file for
the sensor. If this occurs, check that the system mass storage file
location specifies the disk where the power sensor data resides. If the
system mass storage file location is correct, the cal factor data for that
particular sensor has not been stored.
NOTEPower sensor data files created using the VERIFY_62 Operation
Verification software (for 8560A/61A/61B/62A/62B/63A spectrum
analyzers) are not compatible with the VERIFY_6XE Operation
Verification software.
Refer to “Sensor Utilities Menu,” on page 28, in this chapter for
additional information.
Setting GPIB Addresses
The last 11 lines (last 12 lines, for revision D.00.00 and later) of the
Conditions Menu are for selecting the GPIB addresses of test
equipment used for the Operation Verification program. It is not
necessary to use all the test equipment listed. Some model numbers
listed are “alternates.” Table 1-1, on page 16, lists the test equipment
required for each test and Table 1-2, on page 19, lists model numbers
allowed for a particular type of test equipment.
NOTEFor software revisions B.01.00 and earlier, the Conditions Menu has
two entries for Agilent 8340A/B synthesized sweepers. Only
Agilent 8340A/B #1 is used.
Software revisions D.00.00 and later have an entry for an Agilent 5350
Series microwave frequency counter.
Chapter 125
Using Operation Verification Software − Volu me I
Using Operation Verification
Entering zero as the test equipment address results in that model
number being unavailable in the program (NA is displayed in the
address field). To minimize possible confusion later, enter a zero for the
address of each piece of test equipment that is not available.
Enter the address for each piece of test equipment that is available,
including the spectrum analyzer under test. Addresses must contain
the select code of the bus to which the equipment is connected followed
by the equipment address on that bus. For example, if the 8902A is at
address 14 on a bus with a select code of 7, enter an address of 714. If
the 8902A were on a bus with a select code of 12, you should enter an
address of 1214.
A question mark (?) next to a GPIB address indicates the address has
not been checked to verify a response. An asterisk (*) next to a GPIB
address indicates the address was checked and that an instrument
responds at that address. If there is neither an asterisk nor a question
mark next to an address, the address has been checked and no response
was detected.
Storing and Loading the Conditions File
The information in the Conditions Menu may be stored for future use
by pressing
Store Conds. A file named CONDITIONS is created on the
disk specified by the system mass storage file location.
NOTECONDITIONS files for software versions B.01.00 and earlier are not
compatible with software versions C.00.00 and C.01.00. Versions
C.00.00 and later CONDITIONS files include entries for the 83640A
and 83650A sweepers and the 8487A power sensor.
CONDITIONS files for software versions C.01.00 and earlier are not
compatible with software versions D.00.00 and later. Versions D.00.00
and later CONDITIONS files include entries for 5350 Series
microwave frequency counters.
When running the Operation Verification program in the future, set the
system mass storage file location to read the disk where the
CONDITIONS file is located and press
Load Conds. If the
CONDITIONS file resides on the default system mass storage file
location, the CONDITIONS file is loaded automatically the next time
the program is run.
The default system mass storage file location is :,700,1. For the 9000
Model 236 ( 9836), it is necessary to use an external disk drive. The disk
drive must support double-sided format.
26Chapter 1
Using Operation Verification Software − Volum e I
Using Operation Verification
Getting to the Test Menu
Once all necessary items in the Conditions Menu are selected, you can
run a test by pressing
Test Menu. Before the Test Menu is displayed,
these things occur:
• Appropriate power sensor data files are loaded.
• GPIB is checked for a response at each address.
• Serial and model number of the spectrum analyzer under test are
queried.
• Reference level calibration is performed.
NOTEPower sensor data files created using the VERIFY_62 Operation
Verification software (for 8560A/61A/61B/62A/62B/63A spectrum
analyzers) are not compatible with the VERIFY_6XE Operation
Verification software.
Refer to “Test Menu,” on page 30, “Test Menu” in this chapter for more
details on running the tests. If a printer is unavailable, the Conditions
Menu is displayed again rather than the Test Menu. All test results
must be sent to the printer.
NOTEWithout a printer, Operation Verification tests do not run.
Verifying the GPIB
To see which test equipment responds on GPIB, press
Verify Bus. This
check only verifies that there is a response at the address listed; it
cannot tell that a particular piece of equipment is at a particular
address. This is useful for verifying GPIB connections without entering
the Test Menu.
Querying the Spectrum Analyzer Serial Number
The Operation Verification program automatically queries the
spectrum analyzer serial and model number on three occasions: at
program initiation, when loading the CONDITIONS file, and when
entering the Test Menu. To query the analyzer serial and model
numbers at any other time, press
Query DUT S/N. This is helpful for
testing multiple spectrum analyzers; you do not have to reload the
CONDITIONS file or restart the program.
Exiting Operation Verification
Press
Exit Program to exit the Operation Verification program.
Chapter 127
Using Operation Verification Software − Volu me I
Using Operation Verification
Dual-Bus Operation
The Operation Verification program may be used on dual GPIB
systems, such as the microwave test set. In these systems, all the test
equipment is connected to GPIB at select code 7, and the device under
test (for example, the spectrum analyzer) is connected to GPIB at select
code 8.
To run this program in a dual-bus configuration, enter equipment
addresses as described in “Setting GPIB Addresses,” on page 25,
making sure that each address properly identifies the bus select code to
which it is connected. Program operation is the same for dual-bus and
single-bus configuration.
Sensor Utilities Menu
Operation Verification needs to know the cal factors of each power
sensor being used. Create, edit, view, and delete data files containing
cal factors for each power sensor in the Sensor Utilities Menu. Power
sensor data filenames include the last five digits of the power sensor
serial number. For example,
for 8481A power sensors the filename is SEN81NNNNN
for 8482A power sensors the filename is SEN82NNNNN
for 8485A power sensors the filename is SEN85NNNNN
for 8487A power sensors the filename is SEN87NNNNN
where NNNNN represents the last five digits of the power sensor serial
number (the serial number suffix). Note that the first two digits in the
filename correspond to the last two digits of the power sensor model
number.
All power sensor data files available on the system mass storage
location file are listed upon entering the Sensor Utilities Menu.
NOTEPower sensor data files created using the VERIFY_62 Operation
Verification software (for 8560A/61A/61B/62A/62B/63A Spectrum
Analyzers) are not compatible with the VERIFY_6XE Operation
Verification software.
28Chapter 1
Adding a Power Sensor Data File
Using Operation Verification Software − Volum e I
Using Operation Verification
To add a new power sensor data file, press
Add File, and enter the power
sensor model number as requested. An error message is displayed if a
disk is not found at the current system mass storage file location.
When prompted for the power sensor serial number, enter only the last
five digits (the serial number suffix). You are then prompted for a cal
factor frequency and for the cal factor. These frequency/cal-factor pairs
need not be entered in order of increasing frequency; the program
inserts the pairs in their proper place. All frequencies should be entered
in MHz.
A 50 MHz Cal-Factor must be entered in order to calibrate the power
sensor. Some power sensors do not include a 50 MHz Cal-Factor on
their chart or calibration record; it is listed as part of the Calibration
Procedure on the case of the power sensor.
If a mistake is made entering a cal factor, enter the frequency of the
erroneous cal factor at the next frequency prompt. Enter the correct cal
factor at the next prompt. If an error was made entering the frequency
value, enter the erroneous frequency at the next frequency prompt and
a zero for the cal factor to delete that frequency point.
Once all cal factor data for a power sensor is entered, enter an S at the
next frequency prompt. The power sensor data is then stored on disk.
Viewing and Editing a Power Sensor Data File
Press
View/Edit to view or edit a power sensor data file. Only data files
listed on the screen can be viewed or edited. If a file is created but data
is not stored, the filename is listed, but no data is viewed and it cannot
be edited.
To change the cal factor at a particular frequency, enter that frequency
at the frequency prompt, then enter the new cal factor.
To delete a frequency/cal factor pair, enter the frequency of the pair to
be deleted and a cal factor of zero. Add a frequency/cal factor pair by
entering the new frequency at the frequency prompt and the new cal
factor.
Deleting a File
Press
Delete File to remove a listed file. At the prompt, enter the
filename exactly as it appears on-screen. You are asked for confirmation
to delete the file.
Chapter 129
Using Operation Verification Software − Volu me I
Using Operation Verification
Changing the System Mass Storage File Location
To add, edit, or view power sensor data on a disk other than the one
currently specified by the system mass storage file location, press
System File. Enter the msus of the new system mass storage file
location. All power sensor files residing on that disk are listed. Upon
returning to the Conditions Menu, the system mass storage file location
is the one determined in the Sensor Utilities Menu.
Listing Available Power Sensor Data Files
Press
List Files to list all power sensor data files on the currently
specified system mass storage location file.
Returning to the Conditions Menu
Press
Cond Menu to return to the Conditions Menu.
Test Menu
The Test Menu displays all tests that can be performed by the
Operation Verification program. Notice that the menu has two pages.
Tests may be run in any of five modes. These are listed below.
• All Tests runs all 17 tests in the sequence shown on screen.
• Single Sequence runs a user-defined sequence of tests once.
• Single Test runs one test once.
• Repeat Sequence runs a user-defined sequence of tests until testing
is aborted.
• Repeat Test runs a single test until the testing is aborted.
If GPIB controlled equipment for a given test does not respond over
GPIB, that test is flagged MISSING ETE (missing electronic test
equipment). These tests cannot be run and, if they are included as part
of a sequence (All Tests, Single Sequence, or Repeat Sequence), they are
ignored. See “List Equipment,” below.
If a test does not apply to the spectrum analyzer, the test will be flagged
TEST NOT APPLICABLE. For example, if the spectrum analyzer is an
E-Series instrument which does not have Option 007, the Fast Sweep
Time Accuracy test will have the message TEST NOT APPLICABLE next
to it.
Equipment connection prompts are displayed on the computer screen.
Most tests check equipment connections and only prompt the operator
if a misconnection is detected.
30Chapter 1
Loading...
+ 1058 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.