Microfiche Part No. 60503-90010 Printed in USA: September, 1999
CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent
Technologies further certifies that its calibration measurements are traceable to the United States National Bureau of
Standards, to the extent allowed by the Bureau's calibration facility, and to the calibration facilities of other International
Standards Organization members.
WARRANTY
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three
years from date of delivery. Agilent Technologies software and firmware products, which are designated by Agilent
Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to
fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date
of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which
prove to be defective. Agilent Technologies does not warrant that the operation of the software, firmware, or hardware shall
be uninterrupted or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated
by Agilent Technologies. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products returned
to Agilent Technologies for warranty service. Except for products returned to Customer from another country, Agilent
Technologies shall pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies' product price, only if
Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva
Export price).
If Agilent Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the
Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer,
Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED
OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. AGILENT
TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Warranty options, extended support contracts, product
maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent
Technologies Sales and Service office for further information on Agilent Technologies' full line of Support Programs.
2
SAFETY SUMMARY
The following general safety precautions must be observed during all phases of operation, service and repair of this
instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the
customer's failure to comply with these requirements.
BEFORE APPLYING POWER.
Verify that the product is set to match the available line voltage and the correct fuse is installed.
GROUND THE INSTRUMENT.
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis
and cabinet must be connected to an electrical ground. The instrument must be connected to the ac power supply mains through a threeconductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. For instruments
designed to be hard-wired to the ac power lines (supply mains), connect the protective earth terminal to a protective conductor before any
other connection is made. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will
cause a potential shock hazard that could result in personal in jury. If the instrument is to be energi zed via an external autotransformer for
voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earth pole) of the ac power lines
(supply mains).
FUSES.
Only fuses with the required rated current, voltage and specified type (normal blow, time delay, etc.) should be used. Do not use repaired
fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the instrument in the presence of flammable gases or fumes.
KEEP AWAY FROM LIVE CIRCUITS.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified
service personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even
with the power cable removed. To avoid injuries, always disconn ect power, discharge circuits an d remove external voltage sources before
touching components.
DO NOT SERVICE OR ADJUST ALONE.
Do not attempt internal service or adjustment unless another person capable of rendering first aid and resuscitation, is present.
DO NOT EXCEED INPUT RATINGS.
This instrument may be equipped with a line filter to reduce electromagnetic interference and must be connected to a properly grounded
receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those stated on the line rating label
may cause leakage currents in excess of 5.0 mA peak.
SAFETY SYMBOLS.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.
Because of the danger of introducing additional hazards, do not install substitu te parts or perform any unauthorized modification to the
instrument. Return the instrument to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features
are maintained.
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the
instruction manual (refer to Table of Contents) .
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or th e like, which, if not correctly
performed or adhered to, could result in personal inj ury. Do not proceed beyond a WARNING sign until the
indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls at tention to an operating procedure, or the like, which, if not correctly
performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed
beyond a CAUTION sign until the indicated conditions are fully understood and met.
Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
3
Table of Contents
GENERAL INFORMATION5
About This Manual5
Troubleshooting Precautions5
Manual Revisions5
Module Initialization6
VERIFICATION7
Introduction7
Test Equipment Required7
CC Mode Test8
CV MODE TEST9
CR MODE TEST9
Transient Operation and Slew Circuit Test11
CC Mode PARD Test11
CC MODE POWER LIMIT12
This manual is designed to be used along with the Agilent 6050A/6051A Service Manual. It includes service information
that is specific to the 60503B Module. Troubleshooting information such as fault isolation, signature analysis, and
block-level troubleshooting is the same for all modules and is found in the Agilent 6050A/6051A Electronic Load
Mainframe Service Manual. The mainframe Service Manual also explains how to safely disassemble and connect the
module to the mainframe for troubleshooting.
Typically, you will need to refer to this manual when you are performing the verification routines, locating a test point on
the component/test point diagram, referring to the schematics for additional troubleshooting information, and initializing the
module after replacing the EEPROM.
You will also need to refer to this manual for identifying and locating replaceable parts. The parts list identifies all
replaceable parts in the module, and the component/test point diagram identifies the location of all electrical parts in your
module.
Troubleshooting Precautions
Use extreme caution when troubleshooting the module when it is connected to the mainframe. AC mains
voltage is present on the exposed pins on the top edge of the mainframe GPIB board and each module
whenever the units are turned on.
Observe all standard antistatic procedures when handling the module assemblies to avoid the possibility
of electrostatic damage (refer to mainframe Service Manual) .
To reduce the risk of electrical shock when troubleshooting a defective module, make sure the GPIB board is installed in
the mainframe. Also, to make it easier to troubleshoot the module, connect the module to the GPIB board using an extender
service cable (P/N 06050-60030). This cable is included with the Service kit (P/N 06050-60004), which must be ordered
separately.
Manual Revisions
Agilent Technologies instruments are identified by a ten-character serial number such as US37250101. This manual was
written for Electronic Load Modules with serial numbers equal to and higher than those shown on the title page. If the serial
number of your module is higher than the one shown on the title page, then the module may have hardware or firmware
differences that are not covered in this manual. If there are such differences, they are documented in one or more Manual
Change sheets sent with this manual.
5
Module Initialization
EEPROM chip U342 on the Control Board stores the module's GPIB address and model number as well as other constants.
The EEPROM was programmed with the proper constants at the factory. If the Control board or the EEPROM chip (U342)
is replaced, the module must be reinitialized with the proper constants by programming the following commands in the
order indicated. After it has been initialized, the module must also be recalibrated as described in the Operating Manual.
60503B Initialization
"CAL 1"! turn calibration mode on
"CAL: INIT 240 , 10 ! initialize default calibration parameters
"CAL:SAVE"! store calibration constants in EEPROM
"DIAG:CAL:SEC -5033"! model number
"DIAG: CAL: SEC 1 , 16896"! model number suffix
"DIAG: CAL: SEC 26 , 1"! module width
"DIAG: CAL: SEC 27 , 1" ! module type
"DIAG: CAL: SEC 28 , 5000"! voltage for soft over power
"DIAG:CAL:SEC 29 , 5200"! current for soft over power
"*RST"! reset factory default state
"CURR : SLEW 0.167E6"! turn on slew rate
"*SAV 0" ! to location 0
"CAL 0" ! turn calibration mode off
6
2
Verification
Introduction
This chapter contains test procedures that check the operation and calibration of an Agilent 60503A Electronic Load
Module. The tests are performed from the front panel of an Agilent 6050A/6051A Electronic Load Mainframe with the
module installed in slot 1. The tests can also be used to determine which circuits are faulty when troubleshooting. There are
some transient, trigger, and pulse functions that require an GPIB controller and will not be verified with manual testing
from the front panel. The following tests will verify, with a high level of confidence, that the module is operating properly
without testing all of its capabilities.
At the end of this chapter are performance record tables where actual measured values can be recorded.
Test Equipment Required
Table 2-1 lists the test equipment required to perform the tests in this chapter. Test setups for the tests are shown in Figures
2-1 through 2-3. Make sure the sense switch on the rear of the module is se t to the LCL position since local sensing is used
in all of the test setups. Use adequate wire gauge when making connections (see Chapter 3 in the Operating Manual).
NoteThe Electronic Load must pass the selftest at power turn-on before the following tests can be performed.
If the unit fails selftest, refer to the overall troubleshooting procedures in the mainframe Service Manual.
Table 2-1. Test Equipment Required for Verification
TypeRequired CharacteristicsRecommended Model
60V/60A Source0 to 500V/0 to 5A
0 to 60V/0 to 50A
Current Monitor
Resistor
Digital Voltmeterdc accuracy of 0.01%
Current Probe with
Amplifier and Power
Supply
OscilloscopeSensitivity: 1mV
0.100 ohms @ 15A9230/15
6 digit readout
Sensitivity of 1mA/10 mV to 50MHz with
less than 300µA of noise to 5MHz.
Bandwidth: 20MHz
Agilent 6035A or equivalent
Agilent 6032A or equivalent
(Guildline Instruments)
Agilent 3455A, 3456A, or 3458A
Tektronix A6302 probe, AM503 probe amplifier,
and TM501 probe power supply.
Agilent 54504
7
CC Mode Test
This test verifies that the module operates in the CC Mode and that the current programming and readback to the front panel
display are within specifications. For each DMM reading, the front panel display should be equal to:
DMM reading in amps ± ((DMM reading in amps x 0.0012) + 0.010).
Note that if the test readings significantly disagree with the specified values or no readings can be recorded, perform the CC
MODE TEST troubleshooting procedures in Figure 3-1 (sheet 2 of 3) in Chapter 3 of the mainframe Service Manual. If the
readings are out of tolerance, calibrate the applicable current range (see Chapter 6 in the Operating Manual).
a. Connect the Electronic Load, power supply (Agilent 6032A or equivalent), DMM, and current monitor resistor (0.100
ohm) as shown in Figure 2-1.
b. Turn on the Electronic Load.
c. Check the high amp current range as follows:
1. Press
2. Turn on the power supply and set for 5V and greater than 10 amps.
3. Wait 30 seconds and then record the DMM and front panel display readings. DMM reading should be between
0.997V (9.975A) and 1.0025V (10.025A). Note that the Electronic Load's CC annunciator is on.
4. Press
5. Wait 10 seconds then record the DMM and front panel display readings. DMM reading should be between
98.85mV (0.988A) and 101.15mV (1.0115A).
, then .
.
d. Check the low current range as follows:
1. Press
2. Wait 10 seconds then record the DMM and front panel display readings. DMM reading should be between
98.85mV (0.998A) and 101.15mV (1.011A.)
3. Press
4. Wait 10 seconds and record the DMM and front panel display readings. DMM reading should be between
8.985mV (89.85mA) and 11.015mV (110.15mA).
then
.
Figure 2-1. Test Setup A
.
8
CV MODE TEST
This test verifies that the module operates in the CV Mode and that the voltage programming and readback to the front
panel display are within specifications. For each DMM reading, the corresponding front panel display should be equal to:
DMM reading +((DMM reading x 0.0010) + 0.150).
Note that if the test readings significantly disagree with the specified values or no readings can be recorded, perform the CV
MODE TEST troubleshooting procedures in Figure 3-1 (sheet 2 of 3), in Chapter 3 of the mainframe Service Manual. If the
readings are out of tolerance, calibrate the voltage range (see Chapter 6 in the Operating Manual).
a. Connect the Electronic Load, power supply (Agilent 6030A in series with Agilent 6032A or equivalent), and DMM as
shown in Figure 2-2. Take care in making connections so that contact resistance voltage drop will not affect the
readings.
b. Press
c. Turn on and set power supply for 250V and 1A.
d. Record the DMM and front panel display readings. DMM reading should be between 239.59V and 240.408V. Note
that the Elect ronic Load's CV annunciator is o n.
e. Press
f. Record the DMM and front panel display readings. DMM reading should be between 2.876V and 3.123V.
, then .
.
Figure 2-2. Test Setup B
CR MODE TEST
This test verifies that the module operates in the CR Mode and that the resistance programming is within specifications. The
programmed resistance values are checked by recording the voltage across the current monitor resistor and the input voltage
(voltage across the module's input terminals), and then calculating the resistance value as follows:
Load resistance = Input voltage/(voltage across resistor/resistor value)
Note if the calculation significantly disagrees with the specified range of values, perform the CR MODE TEST
troubleshooting procedures in Figure 3-1 (sheet 3 of 3) in Chapter 3 of the mainframe Service Manual. If the calculation is
out of tolerance, calibrate the applicable resistance range (see Chapter 6 in the Operating Manual).
9
a. Connect the Electronic Load, power supply (Agilent 6035A or equivalent), and current monitor resistor (0.100 ohm) as
shown in Figure 2-1. Use the DMM to measure the voltage across the monitor resistor and across the module's input
terminals.
b. Check the low ohm range as follows:
1. Press
2. Turn on power source and set for 60V and 1.82A. For the low ohm range test, the power supply will operate in the
current limit mode.
3. Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the
Electronic Load resistance. The result should be between 23.6 and 24.4 ohms. Note that the Electronic Load's CR
annunciator is on.
4. Press:
5. Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the
Electronic Load resistance. The result should be between 0.792 and 1.208 ohms.
c. Check the middle ohms range as follows:
, then
.
, then .
1. Press
2. Set power supply for 44V and 4A.
3. Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the
4. Press
5. Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the
d. Check the high ohms range as follows:
1. Press:
2. Set power source for 240V and 2A.
3. Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the
4. Press:
5. Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the
Electronic Load resistance. The result should be between 433 and 590 ohms.
Electronic Load resistance. The result should be between 23.75 and 24.24 ohms.
Electronic Load resistance. Calculation should be between 1247 and 5037ohms.
Electronic Load resistance. The result should be between 223 and 259 ohms.
, then
.
, then .
.
.
10
Transient Operation and Slew Circuit Test
This test verifies transient and slew circuit operation. The slew circuits cannot be calibrated. If slew rise time and/or fall
time are not within specifications or the slew circuits are inoperative, perform either the "Transient Generator
Troubleshooting", or the "Slew Circuit Troubleshooting" in Chapter 3 of the mainframe Service Manual.
a. Use the test setup of Figure 2-1 except connect an oscilloscope across the 0.100 current monitor resistor in place of the
DMM. Set power supply for 10V and 10A.
b. Recall the factory default values by pressing
c. Select the current range by pressing
d. Set up transient operation by pressing
Set the slew rate to 0.083A/µs (83A/ms) by pressing
e.
, then
f. Adjust the oscilloscope for a single rise or fall time display. Use delayed sweep. The rise time when measures from
10% to 90% or the fall time when measured from 90% to 10% should be between 60 and 100µs as shown below. Note
that the Electronic Load's Tran annunciator is on.
9.3A
(blue shift key), then
.
.
.
, then .
1.0A
20µµµµS/DIV
SLEW 0.083
g. Set the slew rate to 0.0042A/µs (4.2A/ms) by pressing (blue shift key), then
.
h. Adjust the oscilloscope for a single rise or fall time display. Use delayed sweep. The rise time when measures from
10% to 90% or the fall time when measured from 90% to 10% should be between 1.2 and 2.0ms.
, then
, then
CC Mode PARD Test
CC mode PARD (periodic and random deviations) is specified as the rms input current in a frequency range 20Hz to 10Mhz.
This test checks CC Mode PARD.
a. Connect the Electronic Load, power supply (Agilent 6032A or equivalent, DMM, and current probe as shown in
Figure 2-3. Set power supply for 10V and greater than 10A.
b. Turn the load's ac power off then on.
11
c. Press
d. DMM reading should be less than 1mA rms.
.
Figure 2-3. Test Setup C
CC MODE POWER LIMIT
This test verifies that the module's power limit circuit is operating properly. If the results specified in steps d through i are
not obtained, troubleshoot the circuits as described in "Overpower Circuits Troubleshooting" in Chapter 3 of the mainframe
Service Manual
If the overpower circuit does not turn the load off within three minutes after performing step d, stop the
tests and troubleshoot the overpower circuits.
a. Connect the Electronic Load and the power source as shown in Figure 2-2.
b. Turn on the Electronic Load and run for approximately five minutes with no power being dissipated (no input power).
c. Then press
d. Turn on and set the power supply for 45 volts and 13 amps.
The Electronic Load's front panel should indicate approximately 45 volts and between 7.2 to 10.9 amps. The front panel
Prot annunciator should also be on.
e. Press
limit.
f. Let the Elect ronic Load continue running. Within three minutes the Electro nic Load should turn its input off, and the
display should show ''PS − OP'' indicating protection shutdown. IF THE OVERPOWER CIRCUIT DOES NOT TURN
THE LOAD OFF WITHIN THREE MINUTES, STOP THE TESTS AND TROUBLESHOOT THE OVERPOWER
CIRCUITS.
g. Immediately press
protection shutdown is latched.
h. Wait approximately one minute and press
displayed.
i. Reduce the power source output to 35 volts. The display should change to "−−−" indicating that the protection
shutdown and overpower conditions are cleared.
to display ''−−OP", indicating that an overpower condition exists and the Electronic Load is in power
, then .
. The ''PS'' display should blink and the input will remain shut down, indicating that
again. This time the load should turn on with only ''OP"
12
PERFORMANCE TEST RECORD - Agilent 60503B LOAD MODULE (Page 1 of 2)
Model Agilent 60503B Ambient temperature __________________________°C
Serial No._________________________________________ Relative humidity_____________________________%
Options___________________________________________ Line frequency______________________Hz (nominal)
Firmware Rev._____________________________________
PERFORMANCE TEST RECORD - Agilent 60503B LOAD MODULE (Page 2 of 2)
Model Agilent 60503B Report No.____________________ Date______________________________
Test DescriptionMinimum
Specification
ResultsMaximum
Specification
Measurement
Uncertainty
CONSTANT CURRENT MODE TESTS
10 Ampere Range Programming
and Readback
High Current (l0A)9.975____________A10.0254 mA
Front Panel DisplayA
Low Current (1A)0.9885____________A1.0115
Front Panel DisplayA
-0.022____________AA
OUT
-0.011____________AA
OUT
+0.0224 mA
OUT
+0.011
OUT
427µA
427µA
1 Ampere Range Programming
and Readback
High Current (1A)0.9885____________A1.0115
Front Panel DisplayA
-0.011____________AA
OUT
OUT
+0.011
Low Current (0.1A)0.0899____________A1.101
Front Panel DisplayA
-0.010____________AA
OUT
OUT
+0.010
427µA
427µA
56µA
56µA
CONSTANT VOLTAGE MODE TESTS
Voltage Programming and
Readback
High Voltage (240V)239.59____________V240.4083mV
Front Panel DisplayV
Low Voltage (3V)2.876____________V3.1236
Front Panel DisplayV
-0.390____________VV
OUT
-0.153____________VV
OUT
+0..3903mV
OUT
+0.153
OUT
35µV
35µV
CONSTANT RESISTANCE MODE TESTS
Low Resistance Range
Resistance (24 Ω)
Resistance (1Ω)
23.6
0.792
____________Ω
____________Ω
24.4
1.208
Middle Resistance Range
Resistance (500Ω)
Resistance (24 Ω)
433
23.75
____________Ω
____________Ω
590
24.25
High Resistance Range
Resistance (2000Ω)
Resistance (240Ω)
1247
223.3
____________Ω
____________Ω
5037
259.5
TRANSIENT SLEW TEST
Fast Slew Transient
Slew Rate 0.083 A/µs
Slew Rate 0.0042 A/µs
75
____________µs
1.5____________ms2.5
125
CONSTANT CURRENT PARD TEST
Current (10A)0__________mA1mA RMS
14
Replaceable Parts
Introduction
Tables 3-3 and 3-4 list the electrical components and Table 3-5 lists the mechanical components for the Agilent 60503B
Electronic Load Modules. These tables provide the following information.
• Reference designation (see Table 3-1).
• Agilent part number.
• Description of part (see Table 3-2).
Refer to Figures 4-2 and 4-4 for component locations.
You can order parts from your local Agilent Technologies sales office (refer to the list at the end of this manual for the
office nearest you). When ordering parts, include the following information:
• Agilent part number
• Description of the part
• Quantity desired
• Electronic Load model number (Agilent 60503B)
Table 3-2. Part Description Abbreviations
ALAluminumPEPolyester
CCCarbon CompositionPDPower Dissipation
CERCeramicPPPolypropylene
DIPDual In-Line PackagePWRPower
DPDTDouble Pole Double ThrowRECTRectifier
FFFlip FlopSIPSingle In-Line Package
FXDFixedTATantalum
GEN-PURPGeneral PurposeTCTemperature Coefficient
ICIntegrated CircuitTFT hi n Film
MACHMachineW/With
MOMeta1 Oxide
15
Table 3-3. Agilent 60503B Control Board - Electrical Parts
9211-61961CARTON, SHIPPING
60503-900071MANUAL, OPERATING
60503-900091MANUAL, SERVICE
24
4
Diagrams
Schematic Diagrams
Schematic diagrams and component location diagrams are provided for the Control board as well as the Power board on two
foldout pages. Two schematic diagrams of the Power board are included in Figure 4-2 on the first foldout; three schematic
diagrams of the Control board are included in Figure 4-4 on the second foldout. Table 4-1 lists the notes that apply to both
the Control board and Power board schematic diagrams.
Table 4-2 lists, alphabetically, all of the signal names that appear on the schematic, along with a brief description of the
signal's function. To help you locate where signals come from and go to, Table 4-1 lists the coordinates for each appearance
of a signal on each sheet of the schematic. Coordinates printed in BOLD indicate the signal origin. On the Control board for
example, when ACLR* is active, it clears the analog circuits. ACLR* originates in area 4D of sheet 1 and also appears in
area 2B of sheet 3.
Table 4-1. Schematic Diagram Notes
1. All resistors are in ohms ± 1%, 1/8 W unless otherwise specified.
2. All capacitors are in microfarads unless otherwise specified.
3. All unmarked capacitors are 0.047µF.
4. An asterisk negates a signal name. For example,
5. Signal lines that are terminated by flags continue on other sheets, and may also go to other locations on the same sheet.
Table 4-2 documents all signal terminations on the schematics. Note that flags do not indicate signal flow direction.
appears on the schematic as CS2*.
CS2
Example:
6. Unterminated signal lines go to at least one other location of the same schematic sheet.
Example:SPCLR*
7. Heavy signal lines represent multiple-wire data buses.
8. Unless otherwise noted, bias connections to IC packages are as follows:
Each foldout includes a component locations diagram with a look-up table for locating electrical components. Figure 4-1
gives the component locations for the Control board and Figure 4-3 gives the component locations for the Power board. The
coordinates in the look-up table under each diagram reference the grid on the diagram and give the location of some point
on each component to within 1/10 of a unit. The mainframe Service Manual gives several examples of how to use the table.
The component locations diagram also indicates specific troubleshooting test point locations. The test points are described
in Chapter 3 of the mainframe Service Manual and are used in various troubleshooting procedures described in that chapter.
25
Table 4-2. Signal-Name Descriptions
MnemonicFunctionSheet 1Sheet 2Sheet 3
CONTROL BOARD
ACLR*Analog circuits clear (D)
BO*Brown out input disable (D)3D2A
CC_EN*CC mode enable (D)
CC_PROGProgramming voltage for CC mode (A)
CCVTSTComparator output, main DAC self-test (D)6D
CG*CR-mode middle-and-high-range select (D)
CLRInitialize transient generator (D)
CR*CR-mode low-range select (D)
CS0*Main DAC chip select (D)
CS1*Transient DAC chip select (D)
CS2*Slew range chip select (D)
CS3*Readback DAC chip select (D)
CS4*Secondary-data-bus buffer chip select (D)
CS5*Control-signals latch chip select (D)
CS6*Control-signals latch chip select (D)
CS7*Non-volatile memory R/W chip select (D)
CV_EN*CV mode enable (D)
CV_PROGProgramming voltage for CV mode (A)
DAC_REF*Main DAC reference enable, CV/CC modes (D)
E
EEPON*EEPROM power-on disable (D)2C
EPC_EN*Extended power capability enable/disable (D)
EXT_PROGExt programming input (A) from rear-panel A4
FLTVoltage-fault signal (D) to rear-panel A6
FSEL0,FSEL1,FSEL2Transient-generator frequency-select bits (D)
HIGH*Transient-DAC output enable (D)
H/L*Main-DAC transfer control (D)
H/L-A/DReadback-DAC transfer control (D)
IMONCurrent-monitor output (A) to rear-panel A1
IMON*Input-current monitor signal (A)3D, 6A2C
IMONRCurrent-monitor-comparator output (D)6D
LCLR*Clear status latch (D)
MODULE_INSTALLED*Indicate how many modules are installed (D)
OP*Overpower status (D)3D2A
OV*Overvoltage status (D)3D2A
PCLR1*Power-on clear signal from mainframe (D)
PORTPORT output (D) to rear-panel A7
PRX
PTX
P_TRIGContinuous mod e pulse trigger (D)6A, 5D
PRI_TRIGTrigger signal from mainframe (D)
PULSE_ENP ulse-mode enable (D)
RCK_HI*Loads transient-level counter (D)
RCK_LOW*Loads main-level counter (D)
RNGCC-mode range select (D)
(A) = analog signal(D) = digital signal 8C = signal origin
- SRemote - sense (A) from rear-panel - S
SA_EN*
SD
SDB
SKPSkip self-test (D) from TP301-5
SLEWProgramming-voltage input to slew circuit (A)5B
SLW1,SLW2,SLW3,SLW4Slew-rate control signals (D)
SPROG(not used)2D2C
SRX
START
STAT_ENEnable status latc h (D)
STB*Enable chip-select decoder (D)
STOP
STX
S0, S1, S2Chip-select-decoder input (D)
TEMP1Power-board temperature (A)5A2C
TMONRTemperature-monitor-comparator output (D)6D4A
TOGGLE*Toggle mode select (D)6B4A
TRANS_ENTransient-generator clear and enable (D)
TRIGTrigger (D)6A8C
TRIG_EN*Main and transient DACs trigger enable (D)
UNREG*Unregul ated-input status (D)3D2A
UXFERMain and transient DACs transfer control (D)
VMONVoltage-monitor output (A) to rear-panel A2
VMON*Input-voltage monitor signal (A)3C, 6A2B
VMONRVoltage-monitor-comparator output (D)6D
VREFVoltage reference for DACs (A)
X(not used)3D2 A
-10V_REFDAC reference for CV and CC modes (A)
+15VTo power board (A)
-15VTo power board (A)
+ 12V_REFRef for readback DAC & main DAC self-test (A)
(A) = analog signal(D) = digital signal 8C = signal origin
Secondary µP SA enable (D) from TP301-1
Secondary µP data bus (D)
Secondary µP data bus, buffered (D)
Secondary-µP receiver (D)
Secondary-µP SA start signal (D)
Secondary-µP SA stop signal (D)
Secondary-µP transmitter (D)