IF YOU CANNOT TURN OFF THE ELECTRICAL
POWER, PULL, PUSH, OR LIFT THE PERSON TO
SAFETY USING A WOODEN POLE OR A ROPE OR
SOME OTHER INSULATING MATERIAL
SEND FOR HELP AS SOON AS POSSIBLE
AFTER THE INJURED PERSON IS FREE OF
CONTACT WITH THE SOURCE OF ELECTRICAL
SHOCK, MOVE THE PERSON A SHORT DISTANCE
AWAY AND IMMEDIATELY START ARTIFICIAL
RESUSCITATION
SAFETY
This product has been designed and tested according to International Safety Requirements. To ensure
safe operation and to keep the product safe, the information, cautions, and warnings in this manual
must be heeded. Refer to Section I for general safety considerations applicable to this product.
This manual includes copyright material reproduced by permission of the HEWLETT-PACKARD Company.
TM 11-6625-3014-14
TECHNICAL MANUAL
NO. 11-6625-3014-14
DIRECT SUPPORT, AND GENERAL SUPPORT
DEPARTMENT OF THE ARMY
Washington, D.C.
OPERATOR’S,ORGANIZATIONAL,
MAINTENANCE MANUAL
MICROWAVE FREQUENCY COUNTER
TD-1225A(V)1/U
(NSN 6625-01-103-2958)
HEADQUARTERS
10 September 1981
REPORTING OF ERRORS
You can improve this manual by recommending improvements using DA Form 2028-2 located
in the back of the manual. Simply tear out the self-addressed form, fill it out as shown on the sample, fold it where shown, and drop it in the mail.
If there are no blank DA Forms 2028-2 in the back of your manual, use the standard DA Form
2028 (Recommended Changes to Publications and Blank Forms) and forward to Commander, US
Army Communications and Electronics Materiel Readiness Command, ATTN: DRSEL-ME-MQ,
Fort Monmouth, NJ 07703.
In either case a reply will be forwarded direct to you.
Destruction of army electronics materiel . . . . . 0-1
SERIAL PREFIX: 1840A
This manual applies to Serial Prefix 1840A, unless
accompanied by a Manual Change Sheet indicating
otherwise.
This manual is an authentiation of the manufacturer’s commercial literature which,through usage, has been found to cover the
data required to operate and maintain this equipment. Since the manual was not prepared in accordance with military specifications
and AR 310-3, the format has not been structured to consider levels of maintenance.
Equipment Supplied
Accessories Available
Recommended Test Equipment
HP-IB Interface Capability
5342A Bus Message Usage
Address Selection
Option 001 HP-IB Program Code Set
Operational Verification Record
Model 5342A Program
Model 9825A Program Description
Sample Printout
Performance Test Record
Adjustments
Exchange Assemblies
Abbreviations and Reference Designations
Replaceable Parts
Option 001 Replaceable Parts
Option 002 Replaceable Parts
Option 003 Replaceable Parts
Option 004 Replaceable Parts
Option 011 Replaceable Parts
Manufacturers Code List
Overall Troubleshooting
Assemblies Tested by Test Mode
Probable Failed Assemblies by Test Mode
Diagnostic Modes of the 5342A
A14 Microprocessor Troubleshooting
A19, A20, A21 Power Supply Troubleshooting
A1, A2 Keyboard/Display Troubleshooting
This manual describes Microwave Frequency Counter TD-1225A(V)l/U (fig. l-l)
and provides maintenance instructions.Throughout this manual, the TD-1225A(V)l/U
is refered to as the Hewlett-Packard (HP) Model 5342A.
0-2.
whether there are new
INDEXES OF PUBLICATIONS.
a.
DA Pam 310-4.
Refer to the latest issue of DA Pam 310-4 to determine
editions,
changes, or additional publications pertaining
to the equipment.
b.
DA Pam 310-7.
Refer to DA Pam 310-7 to determine whether there are
modification work orders (MWO’S) pertaining to the equipment.
0-3.
forms,
FORMS AND RECORDS.
a.
Reports of Maintenance and Unsatisfactory Equipment.Maintenance
records,
and reports which are used by maintenance personnel at all
levels of maintenance are listed in and prescribed by TM 38-750.
b.
Report of Packaging and Handling Deficiencies.
Fill out and forward
SF 364 (Report of Discrepancy (ROD))as prescribed in AR 735-11-2/DLAR
4140.55/NAVSUPINST 4610.33B/AFR 75-18/MCO p4610.19C and DLAR 4500.15.
c. Discrepancy in Shipment Report (DISREP) (SF 361). Fill out and
forward Discrepancy in Shipment Report (DISREP) (SF 361) as prescribed in
AR 55-38/NAVSUPINST 4610.33B/AFR 75-18/MCO P4610.19C and DLAR 4500.15.
0-4.
REPORTING OF EQUIPMENT IMPROVEMENT RECOMMENDATIONS (EIR).
EIR’s will be prepared using DA Form 2407, Maintenance Request.
Instructions for preparing EIR’s are provided in TM 38-750, The Army
Maintenance Management System.
EIR’s should be mailed directly to
Commander, US Army Communications and Electronics Materiel Readiness Command,
ATTN:
DRSEL-ME-MQ, Fort Monmouth, New Jersey 07703.
A reply will be
furnished directly to you.
0-5.
ADMINISTRATIVE STORAGE.
Administrative storage of equipment issued to and used by Army activities
shall be in accordance with TM 740-90-1.
0-6.
DESTRUCTION OF ARMY ELECTRONICS MATERIEL.
Destruction of Army electronics materiel to prevent enemy use shall be in
accordance with TM 750-244-2.
0-1
Model 5342A
General Information
SECTION I
GENERAL INFORMATION
1-1. INTRODUCTION
1-1. This manual provides operating and service information for the Hewlett-Packard Model
5342A Microwave Frequency Counter, shown in Figure 1-1.
1-3. SPECIFICATIONS
1-4. Specifications of the 5342A are listed in Table 1-1.
the minimum signal level is not less than the
sensitivity specification.
Automatic amplitude discrimination: Automati-
cally measures the largest of all signals present, providing that signal is 6 dB above any
signal within 500 MHz; 20dB above any signal,
500 MHz—18 GHz.
Modes of operation:
Automatic:
and displays highest level signal within sensitivity range.
Manual:
±50 MHz to true value.
Acquisition time:
Automatic mode: Normal FM 530 ms worst
case; wide FM 2.4 s worst case.
Manual mode:
INPUT 2:
Frequency range:
Count.
Sensitivity:
50 10 Hz to 520 MHz 25 mV rms,
1 M 10 Hz to 25 MHz 50 mV rms.
Impedance:
nominal.
Coupling: ac
Connector: Type BNC female.
Maximum input:
dc fuse protected; 1 MW 200V dc +5.0V rms.
Crystal frequency: 10 MHz
Stability:
Aging rate: <1 X 10-
Short term:
50°C.
Line variation:
nominal.
(TTL compatible); 1.5V peak-to-peak into
50W available from rear panel BNC.
Table 1-1. Model 5342A Specifications
INPUT CHARACTERISTICS
-20 dBm
25 dB
Precision Type N female
dc to load, ac to instrument.
Switch selectable (rear panel)
50 MHz p-p worst case.
Any modulation index provided
Counter automatically acquires
Center frequency entered to within
80
ms after frequency entered.
10 Hz to 520 MHz Direct
Selectable: 1 MW, <50 pF or 50W
50Ω
3.5V rms (+24 dBm) or 5V
TIME BASE
7
per month.
External time base:
Requires 10 MHz, 2.0V peakto-peak sine wave or square wave into 1 KW
via rear panel BNC connector.
Switch
selects either internal or external time base.
OPTIONAL TIME BASE
(OPTION 001)
Option 001 provides an oven-controlled crystal
oscillator time base, 10544A (see separate data
sheet), that results in better accuracy and longer
periods between calibration.
Crystal frequency: 10 MHz
Stability:
Aging rate:
up.
Temperature:
50°C.
Short term:
Line variation:
nominal.
Warm-up:
after turn-on, at 25°C.
AMPLITUDE MEASUREMENT
(OPTION 002)
Option 002 provides the capability of measuring
the amplitude of the incoming sine wave signal,
and simultaneously displaying its frequency (MHz)
and level (dBm). The maximum operating level
and the top end of dynamic range are increased to
+20 dBm. Amplitude offset to 0.1 dB resolution
may be selected from front panel pushbuttons.
INPUT 1:
Frequency range: 500 MHz—18 GHz,
Dynamic range (frequency and level):
-22 dBm to +20 dBm 500 MHz to 12.4 GHz
-15 dBm to +20 dBm 12.4 GHz to 18 GHz
Maximum operating level: +20 dBm
Damage level: +25 dBm, peak
Resolution: 0.1 dB
Accuracy:
Bail Handle Kit5061-2002
Rack Mounting Adapter Kit (Option 908)5061-0057
Rack Mounting Adapter Kit with slot for access
to front connectors from rear.
Transit Case9211-2682
Service
Accessory Kit (refer to paragraph
Microwave Attenuators
Signature Analyzer
1-16)
HP PART NUMBER
K70-59992A
Model 10842A
Model 8491B, 8494/5/6H
Model 5004A
1-11. DESCRIPTION
1-12. The 5342A Microwave Frequency Counter measures the frequency of signals in the range
of 10 Hz to 18 GHz, with a basic sensitivity of -25 dBm. Signals in the frequency range of 10 Hz to
500 MHz are measured by the direct count method. Signals in the frequency range of 500 MHz to
18 GHz are down-converted to an IF by a heterodyne conversion technique for application to the
counter circuits. The unique conversion technique employed results in high sensitivity and FM
tolerance in addition to automatic amplitude discrimination. The counted IF is added to the local
oscillator frequency to determine the unknown frequency for display.
1-13. OPTIONS
1-14. Options available with the 5342A are described in Table 1-1 and paragraph 3-57. If an
option is included in the initial order, it will be installed at the factory and ready for operation
upon receipt. If an option is ordered
for field installation it will be supplied as a retrofit kit. Refer
to Section II for kit part numbers and installation instructions.
1-15. SERVICE EQUIPMENT AVAILABLE
1-16. Extender boards are available for servicing printed circuit assemblies while extended from
the instrument. The extender boards allow assemblies to be extended from their plug-in con-
nectors for monitoring with appropriate test equipment. Extender boards for each assembly are
supplied in Service Accessory Kit 10842A as described in paragraph 8-46.
1-17. RECOMMENDED TEST EQUIPMENT
1-18. The test equipment listed in Table 1-4 is recommended for use during performance tests,
adjustments, and troubleshooting. Substitute test equipment may be used if it meets the required
characteristics listed in the table.
1-4
Table 1-4 Recommended Test Equipment
Model 5342A
General Information
INSTRUMENT
Oscilloscope
Signal Generator
Spectrum AnalyzerRF inputs from 1 MHz—500 MHzT,A,P
DC Voltmeter20V Range, 0.05V ResolutionT,A
AC Voltmeter10 MHz-350 MHzT,A
AC Voltmeter100 kHz, 1% accuracy
Logic State AnalyzerHP 1740A compatibility
Signature Analyzer5342A compatibilityT
Power SplitterDC—18 GHz
Logic Pulser
Current Tracer
Logic Probe
Step Attenuator
AP Clips (4)Clip for 14 pin/16 pin IC’s
Isolation Transformer120V IN — Isolated 120V OUT
Extender Boards2 X 10 pin
REQUIRED
CHARACTERISTICS
100 MHz bandwidth
10 Hz—10 MHz
10 MHz—2.4 GHz
2 GHz—18 GHz
USE*
T,A,OV,P
T,A,OV,PHP 651B
A (Opt. 002)
OV,P
TTL compatibility
1 mA—1 A range
TTL compatibility
DC—18 GHz 10 dB steps
2 X 12 pin
2 X 15 pinHP P/N 05342-60032
2 X 18 pin (2)
OV,P
2 X 22 pin (2)
X 24 pin
2
A 14 Extender
A15 Extender
Power Meter
Power Sensor
50Ω Termination
Microwave Amplifier1 GHz, >+20 dBm Output
Signal Generator
Signal Generator
HP P/N 05342-60034
HP P/N 05342-60035
HP P/N 05342-60036
HP
P/N 05342-60039
HP 909A
(Option 012)
002)HP 489A
HP 3312A
P
P
P
P
P
P
P
HP 8755B
HP 11665B
HP 11664A
HP 182T
HP 11692D
HP 778D
HP 8620C Mainframe
1-5
Model 5342A
Installation
SECTION II
INSTALLATION
2-1. INTRODUCTION
2-2. This section contains information for unpacking, inspection, storage, and installation.
2-3. UNPACKING AND INSPECTION
2-4. If the shipping carton is damaged, inspect the instrument for visible damage (scratches,
dents, etc.). If the instrument is damaged, notify the carrier and the nearest Hewlett-Packard
Sales and Service Office immediately (offices are listed at the back of this manual). Keep the
shipping carton and packing material for the carrier’s inspection. The Hewlett-Packard Sales
and Service Office will arrange for repair or replacement of your instrument without waiting for
the claim against the carrier to be settled.
2-5. INSTALLATION REQUIREMENTS
CAUTION
Before connecting the instrument to ac power lines,
be sure that the voltage selector is properly
positioned as described below.
2-6.
LINE VOLTAGE REQUIREMENTS. The 5342A is equipped with a power module that con-
tains a printed-circuit line voltage selector to select 100- 120-, 220-, or 240-volt ac operation.
Before applying power, the pc selector must be set to the correct position and the correct fuse
must be installed as described
2-7. Power line connections are selected by the position of the plug-in circuit card in the
module. When the card is plugged into the module, the only visible markings on the card indicate the line voltage to be used. The correct value of line fuse, with a 250-voIt rating, must be
installed after the card is inserted. This instrument uses a 0.75A fuse (HP Part No. 2110-0360) for
100/120-volt operation; a 0.375A fuse (HP Part No. 2110-0421) for 220/240-volt operation.
2-8. To convert from one line voltage to another, the power cord must be disconnected from
the power module before the sliding window covering the fuse and card compartment can be
moved to expose the fuse and circuit card. See
below.
Figure 2-1.
Figure 2-1. Line Voltage Selection
2-1
Model 5342A
Installation
2-9. Power Cable
2-10. The 5342A is shipped with a three-wire power cable. When the cable is connected to an
appropriate ac power source, this cable connects the chassis to earth ground. The type of power
cable plug shipped with each instrument depends on the country of destination. Refer to
Figure 2-2 for the part numbers of the power cable and plug configurations available.
2-2
Figure 2-2. Power Cable HP Part Numbers versus Mains Plugs Available
WARNING
BEFORE SWITCHING ON THIS INSTRUMENT, THE
PROTECTIVE EARTH TERMINALS OF THIS INSTRU-
MENT MUST BE CONNECTED TO THE PROTECTIVE
CONDUCTOR OF THE (MAINS) POWER CORD. THE
MAINS PLUG SHALL ONLY BE INSERTED IN A
SOCKET OUTLET PROVIDED WITH A PROTECTIVE
EARTH CONTACT, THE PROTECTIVE ACTION MUST
NOT BE NEGATED BY THE USE OF AN EXTENSION
CORD (POWER CABLE) WITHOUT A PROTECTIVE
CONDUCTOR (GROUNDING).
Model 5342A
Installation
2-11. Operating Environment
2-12. TEMPERATURE. The 5342A may be operated in temperatures from 0°C to +55°C.
2-13. HUMIDITY. The 5342A may be operated in environments with humidity up to 95%. However, it should be protected from temperature extremes which cause condensation in the
instrument.
2-14. ALTITUDE. The 5342A may be operated at altitudes
2-15. STORAGE AND SHIPMENT
2-16. Environment
2-17. The instrument may be stored or shipped in environments within the following limits:
2-18. The instrument should also be protected from temperature extremes which cause
condensation within the instrument.
2-19. Packaging
2-20. ORIGINAL PACKAGING. Containers and materials identical to those used in factory
packaging are available through Hewlett-Packard offices. If the instrument is being returned to
Hewlett-Packard for servicing, attach a tag indicating the type of service required, return address,
model number, and full serial number. Also, mark the container FRAGILE to ensure careful
handling. In any correspondence, refer to the instrument by model number and full serial
number.
2-21. OTHER PACKAGING. The following general instructions should be used for repacking
with commercially available materials:
7,620 metres (25,000 feet)
up to 4,600 metres (15,000 feet).
Up to 95%
Wrap instrument in heavy paper or plastic. (If shipping to Hewlett-Packard office or
a.
service center, attach tag indicating type of service required, return address, model
number, and full serial number.)
b. Use strong shipping container. A double-wall carton made of 350-pound test material is
adequate.
c.
Use a layer of shock-absorbing material 70 to 100 mm (3-to 4-inch) thick around all sides
of the instrument to provide firm cushioning and prevent movement inside container.
Protect control panel with cardboard.
d. Seal shipping container securely.
Mark shipping container FRAGILE to ensure careful handling.
e.
f.In any correspondence, refer to instrument by model number
2-22. FIELD INSTALLATION OF
2-23. Procedures for field installation of Options 001,002,003,004, and 011 are described in the
OPTIONS
and full serial number.
following paragraphs.
2-3
Model 5342A
Installation
2-24. Part Numbers for Ordering Option Kits
2-25. To obtain the necessary parts for installation of an option, order by part number as listed
below (refer to Section VI for ordering information):
Option
001
002
003
*004
001
If the instrument in which Option 004 is to be installed
has a series number 1812 or lower, the U7 ROM on
A14 Microprocessor will have to be replaced. Order
U7 ROM Part Number 1818-0706 to replace the old
U7 ROM (1818-0331).
2-26. Installation of 10 MHz Oscillator Option 001
2-27. Option 001 consists of oven-controlled crystal oscillator time base 10544A, which has a pc
card connector. Option 001 is installed in the same connector on the motherboard as the standard oscillator (A24). See Figure 8-44. To install Option 001, proceed as follows:
High Stability Time BaseHP Model 10544A
Amplitude Measurement
Extended Dynamic Range05342-60201 (Kit)
Digital-to-Analog Converter
HP-IB I/O05342-60019 (HP-IB Assy.)
NamePart Number
05342-60200 (Kit)
05342-60202 (Kit)
05342-60029 (HP-IB Input Assy.)
*NOTE
a.
Remove the standard oscillator from A24 connector.
b. Install Option 001 oscillator into A24 connector.
Attach Option 001 oscillator to the motherboard by means of two 6/32X5/16 pan head
c.
screws. Install the screws from the bottom of the motherboard using star washers.
d. Perform Option 001 oscillator adjustment as described in paragraph 5-32.
2-28. Installation of Amplitude Measurement Option 002
2-29. Option 002 consists of U2 High Frequency Amplitude assembly and A27 Low Frequency
Amplitude Assembly modules and the A16 Amplitude Assembly pc board. U2 is connected to the
high frequency input of the 5342A, A27 is connected to the low frequency input and both of the
modules are connected to the A16 board by the coax wires supplied. See photo of installed
option, Figure 8-22, and schematic diagram, Figure 8-39. To install the components proceed as
follows:
NOTE
The parts that comprise this option are listed in Table 6-5.
a.
Remove the top and bottom covers and top plate from instrument.
b. Place instrument top down.
c.
At inside front panel, disconnect cables from A1J1,J1J3,J25J1 (IF OUT lNT), and A25J2 (IF
OUT EXT).
d. Solder one end of the white/red/green 14-inch wire (8120-0483) to AT1 feedthrough
capacitor terminal on A25 Preamplifier assembly.
2-4
Install coax assembly 8120-2268 through A22 motherboard from top of instrument at A16
e .
slot. Place the wires through the holes as shown below:
NOTE
Model 5342A
Installation
Prior to installing A27 Low Frequency
Amplitude Assembly, connect the wires as
described below.
f.
Solder one
capacitor terminal on A27.
Place heat shrinkable tubing (0890-0983) over connection at C7,
g.
h.
Place heat shrinkable tubing (0890-0983) over three of the coax wires (red, blue, and
green) that were installed in step e. and solder
Coax
Red
Blue
Green
Apply heat to shrink the tubing at the connections made in step g and h.
i.
Remove attaching nut from front panel N-type input connector and disconnect rigid
j.
coax W1 from J1 on U1 Sampler. Remove W1 from instrument.
Mount A27 Low Frequency Amplitude Assembly in the recessed angle of the casting
k.
behind front frame, see Figure 8-22. Attach A27 to casting with two pan head screws
supplied. Place a star washer under the other screw.
The wire previously soldered to A27C10 has a black ground wire attached. Solder the end
l.
of this black wire to the ground lug installed in preceding step.
end of the black/white/blue 14-inch wire (8120-0471) to C7 feedthrough
these wires to the terminals listed below:
Terminal
A27C10
A27C9
A27C8
Solder the free end of white/red/green wire (other end connected to A25AT1 in step d)
m.
to A22 motherboard at XA16B, pin 3 (ATT).
NOTE
Prior to installing U2 High Frequency Amplitude
Assembly, connect the color-coded wires as shown
below. Place heat shrinkable tubing (0890-0983 for
coax and 0890-0706 for single wires) over all connections to U2.
2-5
Model 5342A
Installation
n.
o.
p.
q.
r.
s.
Connect rigid coax (8120-2516) from U2 High Frequency Amplitude Assembly to J1 on
Sampler U1. Install U2 input connector through front panel. Fasten with attaching nut.
Solder white/black/red wire (from U2) to A22 motherboard XA16B, pin
Solder white/brown/red wire (from U2) to A22 motherboard XA16B, pin
Harness the coax cables and wires with tie wraps supplied.
Connect cable 05342-60119 from A27J1 to A1J3.
Connect cable A1J3/A27J2 to A27J2.
Reconnect A1J1,J1 (IF OUT INT) and J2 (IF OUT EXT) and harness with tie wrap.
t.
Harness the white cables with tie wraps supplied.
u.
NOTE
The ROM and U2 High Frequency Amplitude
Assembly are supplied as a matched pair and are
included
(05342-80005).
v.
Install the ROM (supplied with option) into U3 socket on A16 (05342-60038) board.
Replace resistor R2 on A16 board with a resistor of the value labeled on U2 assembly.
w.
Insert the plug of 8120-2268 cable into mating socket on A16 board (05342-60038) and
x.
install A16 into connector XA16.
Perform the Option 002 adjustments listed under paragraph 5-33 through 5-39 of this
y.
manual.
Perform the operational verification procedures in paragraphs 4-14, 4-15, and 4-17
z.
of this manual.
under one replaceable part number
NOTE
If the instrument does not meet the specified
accuracy of ±1.5 dB as described in paragraph 4-14,
perform the following procedures.
Replace resistor R6 from the A27 Low Frequency Amplitude Assembly and replace with a
resistor of a higher or lower value as shown below. For lower power readings increase the
value and for higher power readings decrease the value of resistor R6 as follows:
2-6
dB Change
0.2
0.4
0.6
0.8
1.0
R6 Changes
(ohms)
10
20
30
40
50
Model 5342A
Installation
2-30.
Installation of Extended Dynamic Range Option 003
2-31. Option 003 consists of A16 Extended Dynamic Range Assembly (05342-60037) and U2
Attenuator Assembly (5088-7038). See Figure 8-22 for location of U2 (Option 002 or 003).
NOTE
The parts that comprise this option are listed at the
end of Table 6-6.
Remove the top and bottom covers and top plate from instrument.
a.
Place instrument top down.
b.
At inside front panel, disconnect cable from A1J1,A1J3,A25J1 (IF OUT INT), and A25J2
c.
(IF OUT EXT).
Solder one end of the white/red/green 14-inch wire (8120-0483) to AT1 feedthrough
d.
capacitor terminal on A25 Preamplifier Assembly.
Solder free end of white/red/green wire (other end connected to A25AT1 in step d) to
e.
A22 Motherboard at XA16B, pin 3 (ATT).
Solder white/bIack/red wire (from U2) to A22 Motherboard XA16B, pin
f.
Solder white/brown/red wire (from U2) to A22 Motherboard XA16B, pin
g.
h.
Remove the N-type input connector from front panel and replace with U2 (5088-7038).
Connect rigid coax (supplied) from U2 to J1 on Sampler U1.
i.
Install A16 board (05342-60037) into XA16 connector.
j.
Perform the operational verification procedures in paragraphs 4-13 and 4-16
k.
of this manual.
2-32.
Installation of Digital-to-Analog Conversion (DAC) Option 004
2-33. Option 004 consists of an A2 Display Driver Assembly (05342-60028) that contains DAC
circuitry added to the standard A2 circuit. Interconnecting wires are included with the Option
004 retrofit kit (05342-60202). Procedures for installation of Option 004 are as follows:
2-7
Model 5342A
Installation
a.
b.
Remove top and bottom covers, front frame and A1-A2 assemblies. Refer to disassembly
procedures, paragraph 8-22.
Replace the original A2 board (05342-60002) with Option 004 A2 board (05342-60028) and
reassemble unit.
If the series number of the instrument is
c.
1812 or lower, the U7 ROM, 1818-0331 on the
A14 Microprocessor board will have to be replaced with U7 ROM, 1818-0706 as described
in step d. If instrument has the
1818-0706 ROM, proceed to step e.
CAUTION
ROM U7 is a large-scale MOS IC. Its inputs are
susceptible to damage by high voltage and by static
charges. Particular care should be exercised when
servicing this IC or handling it under conditions
where static charges can build up.
Remove top plate from 5342A. Remove A14 Microprocessor and replace ROM U7 part
d.
number 1818-0331 with part number 1818-0706. Install A14.
At bottom of 5342A connect coax cable to the connector at the bottom rear of A2 board
e.
labeled D/A OUTP. Solder the other end of this cable to the DAC OUT connector on the
rear paneI.
f.
Connect the white/gray wire to the pin (push-on) labeled LDA at bottom rear of A2
Display Driver board. Solder other end of wire to LDA terminal on A22 Motherboard as
shown in figure below.
Connect red wire (+15V) and violet wire (-15V) to the proper terminals (push-on pins) on
g.
A2 Display Driver board (see
Figure 8-25, component locator for location). Connect
other end of these wires to terminals on A22 Motherboard as shown in figure below.
2-8
h.
Reassemble instrument and perform operational verification procedures in para-
graph 4-27 of this manual.
2-34.
2-35.
Installation of HP-IB Option 011
Option 011 consist of printed-circuit assembly A15 and interconnection board A29. The
interconnection board mounts inside the 5342A rear panel and is connected to A22 Motherboard
via a cable strap. Procedures for installation of Option 011 are as follows (see photo of installed
option, Figure 8-22):
Remove top and bottom covers and top panel from the 5342A.
a.
b. Insert A15 assembly into A15 slot. See Figure 8-21 for location.
c.
If 5342A is equipped with Option 001 Oscillator, remove oscillator assembly by removing
two attaching screws from A22 Motherboard.
NOTE
Model 5342A
Installation
In the following step, make sure
switch (A29S1) is located as shown
Insert the A29 Interconnection board (05342-60019) into the rear panel slots provided
d.
(from inside). Screw the two mounting studs (0380-0644) and washers (2100-3171) into the
HP-IB connector to attach the board to the rear panel,
Connect the plug of the cable strap from A29 to J2 on A22 Motherboard with arrow on
e.
installed plug pointing toward front panel.
f.
Perform the Option 011 HP-IB Verification in paragraph 4-19 of this manual.
Refer to paragraph 2-36 for HP-IB interconnection data and to paragraph 3-69 for
g.
programming information.
2-36. HP-IB Interconnections
2-37. HEWLETT-PACKARD INTERFACE BUS. Interconnection data concerning the rear panel
HP-IB connector is provided in Figure 2-3. This connector is compatible with the HP 10631A/
B/C/D HP-IB cables. The HP-IB system allows interconnection of up to 15 (including the controller) HP-IB compatible instruments. The HP-IB cables have identical “piggy back” connectors
on both ends so that several cables can be connected to a single source without special adapters
or switch boxes. System components and devices may be connected in virtually any configu-
ration desired. There must, of course, be a path from the calculator (or other controller) to every
device operating on the bus. As a practical matter, avoid stacking more than three or four cables
on any one connector. If the stack gets too large, the force on the stack produces great leverage
which can damage the connector mounting. Be sure each connector is firmly (finger tight)
screwed in place to keep it from working loose during use.
that the address
in Figure 8-20.
2-38. CABLE LENGTH RESTRICTIONS. To achieve design performance with the HP-IB, proper
voltage levels and timing relationship must be maintained. If the system cable is too long, the
lines cannot be driven properly and the system will fail to perform properly. Therefore, when
interconnecting an HP-IB system, it is important to observe the following rules:
The total cable length for the system must be less than or equal to 20 metres (65 feet).
a.
b. The total cable length for the system must be equal to or Iess than 2 metres (6.6 feet) times
the total number of devices connected to the bus.
The total number of instruments connected to the bus must not exceed 15.
c.
2-39. 5342A Listen Address
2-40. The 5342A contains a rear panel HP-IB Instrument address selection switch. There are five
setting and changing the listen address are provided in Section III of this manual along with
programming codes.
2-41. HP-IB Descriptions
2-42. A description of the HP-IB is provided in Section Ill of this manual, A study of this infor-
mation is necessary if the user is not familiar with the HP-IB concept. Additional information
concerning the design criteria and operation of the bus is available in IEEE Standard 488-1975,
titled “IEEE Standard Digital Interface for Programmable Instrumentation”.
2-9
Model 5342A
Installation
2-10
Figure 2-3. Hewlett-Packard Interface Bus Connection
Model 5342A
Operation
SECTION Ill
OPERATION
3-1. INTRODUCTION
3-2. This section contains operating information including operating characteristics, descrip-
tions of controls and indicators, and operating procedures.
3-3. OPERATING CHARACTERISTICS
3-4. The following paragraphs describe the operating ranges and modes, resolution, sample
rate, AM and FM characteristics, and auto-amplitude discrimination. Front panel controls and
indicators are described in Figure 3-1, rear paneI controls and connectors are described in Fig-ure 3-2. Operating procedures are explained in Figure 3-3. Amplitude measurements (Option
002) are described in Figure 3-4. DAC operation (Option 004) is described in Figure 3-5.
3-5. Operating Ranges
3-6. There are two basic operating ranges: 10 Hz to 500 MHz and 500 MHz to 18 GHz.
Frequencies in the lower range are measured directly while measurements in the 500 MHz to
18 GHz range are made indirectly by a harmonic heterodyne down-conversion technique.
Provision is made to select either range by a front-panel slide switch. A separate input connector
is provided for each range. When the range switch is in the 10 Hz—500MHz position, the signal at
the BNC connector is routed to the direct count circuits of the 5342A. In this range, input impedance is selectable via the
range, the input signal is applied via the front-panel type N connector to the down-conversion
circuits of the 5342A.
50Ω−1 ΜΩ switch. When the range switch is in the 500 MHz—18 GHz
3-7. Resolution Keys
3-8. The best case resolution is the value represented by the least significant digit (LSD) in the
display. In the 5342A, a maximum resolution of 1 Hz can be selected (by the pushbutton keys on
the front panel labeled in blue, preceded by the blue key being pressed). The display is divided
into four sections for ease of determining GHz,MHz,kHz,and Hz resolution. Half-sized
used as space fillers within a section to improve interpretation of the display. For example, a
signal measured to 100 kHz resolution will be displayed thus:
3-9. The pushbutton keys on the front panel under the RESOLUTION label are used for other
purposes when the blue key is not in effect (has not been pressed). When the blue key has not
been pressed, the keys are defined by the black number on the keys and are used to enter fre-
quency offsets, manual center frequencies, and amplitude offsets as described in Figure 3-1.
3-1
Model 5342A
Operation
3-10. CHECK, DAC, and ENTER keys
3-11. The CHECK, DAC, and ENTER keys are used as described in Figure 3-1.
3-12. FREQ Keys
3-13. Two of the pushbutton keys on the front paneI under the FREQ label are used to select the
automatic or manual mode of operation. The other keys in this section of the keyboard control
the use of the RESOLUTON keys. Use of these keys is described in detail in Figure 3-1.
3-14. Automatic Mode
3-15. The automatic mode of operation is selected by pressing the AUTO key. Input signals in
the 500 MHz—18 GHz range are acquired, measured, and displayed automatically. When power
is initially turned on, the 5342A goes into this mode automatically.
3-16. Manual Mode
3-17. The manual mode of operation is selected by pressing the MAN (MHz) key. To operate in
this mode, input signals in the 500 MHz-18 GHz range must be known to within 50 MHz and this
frequency (called the manual center frequency) must be entered into the display prior to the
measurement. Use of the manual mode is described in detail in Figure 3-3.
3-18. Offset Frequencies
3-19. It is sometimes desirable to add or subtract a constant to\from a frequency measurement.
For example, by measuring a radio IF and knowing the LO, the counter can display the RF input
when the LO frequency is entered as a positive offset. It may be easier to tune an oscillator to a
specific frequency if the desired frequency is entered as a negative offset and the oscillator tuned
until the counter reads zero. Frequency offsets are described in Figure 3-3.
3-20. Amplitude and Offset Measurements
3-21. When Amplitude Option 002 is installed, the amplitude is displayed in addition to the fre-
quency of the input signal. The frequency is displayed to 1 MHz resolution in the five leftmost
digits and the amplitude is displayed to 0.1 dB resolution in the four rightmost digits of the display. An arbitrary value can be selected as an amplitude offset and can be added to or subtracted
from the measured value as described in Figure 3-4.
3-22. Digital-to-Analog Converter (DAC) Operation
3-23. When DAC Option 004 is installed, any three consecutive digits of the display can be
selected and converted to a corresponding analog voltage output. The voltage is available at the
BNC connector on the rear panel (labeled DAC OUT) and is between
example, if the selected digits are
output is 9.99 volts dc. Operating procedures are listed in Figure 3-5.
3-24. SET, RESET, RECALL, and CHS Keys
∅∅∅ the output is ∅ volts and if the selected digits are 999 the
∅ and 9.99 volts dc. For
3-2
3-25. The SET, RESET, RECALL, and CHS keys allow offsets and center frequencies to be entered,
reset the measurement process, recall previous values, and change the sign of offsets as
described in Figure 3-3.
3-26. SAMPLE RATE, GATE, and REMOTE
3-27. The SAMPLE RATE control adjusts the deadtime between the end of one measurement
and the start of the next measurement. The duration of the measurement is determined by the
Loading...
+ 334 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.