A302: HV Transformer Assembly: includes C309, C310, T301, V303, and V304;
Stock No. 185A-llA; Mfr. 28480; Mfr. Part No. 185A-llA; TQ
C309, C310: Same description and @ Stock No.
L603: Same description and
T301: HV Transformer;
Figure 1-1. Model 185B Oscilloscope with Model 187B Dual-Trace Vertical Plug-In Unit
1-0
~
01255-1
Model
185B
Paragraphs
Section I
1-1
to
1-10
SECTION
GENERAL INFORMATION
1-1.
INTRODUCTION.
1-2.
This manual gives operation and maintenance
information for the
An operational check
operation
performance check
specifications during incoming inspection
paragraph
1-3.
1-4.
P2
phosphor with external graticule.
options, however, are available:
a.
b. Option
c. Option
d:
e.
1-5.
option
differences in crt persistence, color, and oscilloscope
photography techniques, instruments with or without
option
are, however,
(see section VI).
in this manual apply to all options unless otherwise indicated.
is
given in paragraph
5-67.
OPTIONS.
The
Model
Option
Option
Option
1.
P1
Internal graticule in cathode ray
3.
7.
P7 phosphor
11.
P11
31.
P31 phosphor
Option
31
1
is
available only with option
3
are operationally interchangeable. There
some differences in component parts
@
Model
to
185B
assure basic instrument
2-10.
that
may be used for verifying
185B
is
normally supplied with
phosphor
phosphor
is
not available with option
All
references to the Model
Oscilloscope.
A
complete
is
given in
The following
tube
3,
3.
Other than
and
185B
I
1-6.
INSTRUMENT IDENTIFICATION.
1-7.
Hewlett-Packard Company uses a two-section
eight-digit serial number (e.g.
first
three digits of the serial number onyour instru-
ment are not
with
this manual which define differences between
your instrument and
sheets are missing, your Hewlett-Packard engineering representative can supply you with the necessary
information.
144-,
change sheets have been supplied
this
a
1-8.
DCSCRl?TlON.
1-9.
GENERAL. The
provides
ena
will
below
spaced.
signal
operate reliably.
above
obtain jitter-free presentation.
1-10.
high speed characteristics
nique. Using this system the entire signal under examination
taken on different occurrences of the input at slightly
later points along the waveform.
a
with repetition rates to
visual display of very high-speed phenom-
present a steady display of pulse repetition rates
100
kc even when the signals are randomly
For frequencies above
is
divided down
100
kc, more uniform spacing
PRESENTATION. The Model
is
scanned, with each succeeding sample
@
Model
so
that the input circuitry will
As
the input frequency increases
by
000-00000).
manual. If
1000
mc. The instrument
100
185B
kc,
is
these
Oscilloscope
the
185B
using the sampling tech-
Each time such a
If the
change
incoming
required
obtains
to
it8
I
HORIZONTAL
Sweep Speeds:
10 ranges,
within
between ranges and increases fastestunmaanified
sweep speedto
sweep
the exception of time represented by
of
the unmagnified sweep.
Magnification:
7
calibrated ranges
and
X100.
speed to
sweep speed
Intensity and sample density are not affected by
magnification.
Delay Control:
Three-turn
when using magnified sweep. Permits any
tion of unmagnified trace
10
nsec/cm
Vernier gives continuous adjustment
&5%.
4
is
maintained at all magnifier settings with
Increases maximum calibrated sweep
0.1
nsec/cm; with vernier, maximum
is
further extended to
variable delay control
to
10
psec/cm, accuracy
nsec/cm. Accuracy of the basic
X1,
X2,
X5,
X10, X20, X50,
0.04
is
to
be
viewed on screen.
01255-1
Table
first
nsec/cm.
available
1/4
por-
1-1.
Specifications
cm
Minimum Delay (input trigger
Less than
faster. On slower sweep
lay increases to
5
psec on the lO-psec/cm range.
Sample
Continuously adjustable from approximately 70
samples per trace
Scanning Functions:
Internal
for normal viewing.
Record
approximately
Manual
knob.
External
approximately
input impedance greater than
120
Density:
-
X
-
X
-
X
-
nsec at
a
maximum of approximately
to
axis
driven
axis driven by internal slow ramp;
60
seconds for one trace.
axis driven by manual scan control
X
axis
driven
12
volts for 10-cm deflection,
to
start
100
nsec/cm sweep and
speeds,
1000
by
by
minimum de-
samples per trace.
internal staircase
external voltage;
25,000
Specifications cont'd
of trace):
ohms.
1-1
Section
Table
I
1-1
(cont‘d)
1-1.
Table
“RIGGER FUNCTIONS
Normal-External Trigger
to
Amplitude: 450 mv
volts rms or 100 volts peak will not damage
input circuit.
Width: 5 nsec at minimum amplitude.
Rate:
50 cps to 1 mc
speed setting. Maximum
100 mc on the 200 nsec/cm
Jitter:
Input Impedance: 50 ohms nominal
Sensitive-External Trigger
Amplitude: *15
Width:
Rate: Same as normal
Jitter:
Input Impedance: 50
High Frequency
Input Frequency: 50 to
Sensitivity: 200 mvpeak-to-peak. Operates from
Jitter: 4% of cycle from 50 to 400 mc;
Signal at Input Connector: Less than 15 mvpeak-
Input Impedance: 50 ohms nominal, ac coupled.
SYNC PROBE
The 185B-21A (use with any trigger function) in-
creases input impedance to more than 7500
ac coupled; reduces sensitivity by approximately
4:l at 10 mc and higher and by approximately 20:l
at
low frequencies.
SYNC PULSE OUTPUT
Amplitude: Positive; at least 1.5 v into 50 ohms
Rise Time: Less than 2 nsec
Width: Approximately 5 psec
Recurrence: One pulse per sample
CALIBRATOR
Voltage:
20 mv, 100 mv,
Time:
Approximately 5 psec burst of 50 mc sinewave.
Frequency accuracy
Less than 0.03 nsec or 0.02% of the time
represented
ever
is
approximately 5:l in the “smoothed” response
position.
Reflection from step of
by
greater (fast rise Signals). Reduced
to
rms or 10 volts peak
circuit.
5 nsec at minimum amplitude
Same as normal
of 200 nsec/cm
smaller signals at some increase in jitter. Up
to 5 volts rms or 15 volts peak will not damage
input circuit.
from 400 to 1000 mc
to-peak, approximately 10 mc
Reflection from step of
less than 8%.
i2 volts peak. Up
on
the lO-Fsec/cm sweep
rate
increases
and
faster
the
unmagnified speed, which-
1/2
nsec
is
*200 mv peak. Up to 5 volts
will
not damage input
ohms
nominal, dc coupled
1000
mc forsweepspeeds
and
faster
1/2
nsec rise time
200
mv, and 1000 mv;
dc
less
8%
rangee.
coupled.
than
of Cycle
*2%.
Specifications (cont’d)
X-Y
RECORDER OUTPUT
X-
and
to
5
to
8%.
is
ohms,
&
minals in all positions of the scanning control. In
the
187A-76A BNC Adapter
187A-76B Type
187B-76C 10:
187A-76D Blocking Capacitor
187B-76E 50-ohm T Connector
Model
Model 908A 50-ohm Coaxial Termination
187B-76F Adapters
187B-76G Probe Socket
Model 213A Pulse Generator,
Model 185B
Y-axis
used to make pen recordings with a con-
signals are available at rear
X-Y
recorder.
to
+13 volts at end of sweep (1.2
Source
impedance approx. 20,000 ohms.
-1
volt at bottom (0.2 v/cm), Source
0
volt
+1
volt at top of
at
ter-
start
Tube:
P1,
P7, and
10 cm, marked in centimeter squares. Major
300
250 ohms
sion, 500 ohms
sion, 2500 ohms
sion, 5000 ohms
llOOA
nsec
rise
time k350 mv amplitude
P11
watts
N
Adapter
1
Divider
Delay Line
phosphors available.
*IO%,
50 to
less
60
1O:l
divi-
1OO:l
divi-
than 0.5
cp8,
A.
1-2
01255-1
Model
185B
Paragraphs
Section I
1-11
to
1-15
sample
is
taken, the
“spot"
on the
crtis
moved hori-
zontally along the waveform. Thus, a complete pic-
ture of
a
repetitive high speed signal
is
synthesized
by a buildup of image-retaining “dots” on the oscilloscope face
1-11.
The Model
ranging from
nanoseconds per centimeter depending on the
of the TIME SCALE switch. Any part
time scale can
by adjusting the TIME SCALE MAGNIFIER
as
a graph
185B
10
microseconds per centimeter
be
expanded without loss
is
provides
plotted point by point,
10
basic
time
scales
setting
of
this
of
calibration,
switch.
to
10
basic
Built-in time and amplitude calibrators provide a
convenient means
vertical calibration. Intensity of
pendent of duty cycle, and vertical deflection may
adjusted up
185B
to
provides output signals for
of
checking
10
centimeters. In addition, the Model
both
horizontal and
the
trace
X-Y
recorders and
is
inde-
be
provides means for controlling the display either manually or externally.
1-12.
VERTICAL AMPLIFIER. The vertical system
of the Model
such as the Model
185B
includes aplug-in vertical amplifier
187B.
It
is
this amplifier which
determines vertical characteristics such as bandpass,
sensitivity, etc. The vertical plug-in unit
of
the
basic
1-13.
F’igure
185B-21A
Model
Sync Probe shown in
185B
Oscilloscope.
1-1
illustrates the Model
the
figure
with the oscilloscope. The Model
Vertical Amplifier
plug-in unit
Table
1-1
1-14.
CATWODLRAY
1-15.
The cathode-ray tube supplied with the Model
185B
is
is
lists
guaranteed against electrical failure for one
year from the date
Company. The cathode-ray
trated in figure
in the appendix of
is
shown installed, although the
not part of the basic oscilloscope.
the specifications for the Model
TUBE
WARRANTY.
of
sale
by
the
tube
warranty
1-2.
A
sheet
for your use
this
manual.
is
not part
187B
185B.
is
Dual Trace
The
supplied
185B.
Hewlett-Packard
is
illus-
is
included
IEWLETT-PACKARD
01255-1
CO.
PAOC
MILL
ROAD.
PAL0
Figure
ALTO. CALIF.
1-2.
U.S.A.
Cathode-Ray
Tube
Warranty
1-3
se~tian
Figure
n
2-1
Model
185B
MODEL
DUAL TRACE
I
CAL AMPL I F I ER
VERT
1878
MODEL 196A/B
OSCILLOSCOPE
CAMERA
MODEL IlOOA
DELAY LINE
MODEL
DELAY
2-0
I
I
I
00A-76
LINE LOAD
A
Figure
2-1.
CONNECTING
Aellociated
@
Equipment
Available
MODEL l85A-76A
SYNC TAKE-OFF
UP
-Y-93
01255-1
Model 185B
Section
Paragraphs 2-1
to
II
2-19
SECTION
PREPARATION
2-1.
INCOMING INSPLCTION.
2-2. MECHANICAL INSPECTION.
2-3. Upon receipt of your Model 185B, check the
contents against the packing
strument for any obvious damage received in transit.
If
damage
(Refer
ditional information.)
all re-usable packing material until an operational
check
2-4. POWER REQUIREMENTS.
2-5. The Model 185B requires
115 or 230 volts
which can deliver approximately
2-6. 230-VOLT OPERATION.
2-7.
operated switch
ment. The existing fuse should
2-ampere slow-blow fuse.
Be
erly for the line voltage
power supplies may
switch
2-8. THREE-CONDUCTOR POWER CABLE,
2-9. To protect operating personnel, the National
Electrical Manufacturers' Association (NEMA)
ommends
grounded. This instrument
conductor power cable which, when plugged into an
appropriate receptacle, grounds the instrument. The
offset pin on the power cable three-prong connector
is
the ground pin. To preserve the protection feature
when operating the instrument from
outlet, use
connect the green pigtail to ground.
2-10. OPERATIONAL CHECK.
2-11. The following procedure
for checking basic operation
controls mentioned in the following procedure
shown in figure 3-1. A complete check-out procedure
to
verify specifications
a.
Install the vertical plug-in unit.
gain
to
b.
Turn INTENSITY
c.
Set
100 NSEC/CM,
is
evident, file claim with the carrier.
to
the
warranty sheet in
To
has
been successfully completed.
*lo%,
If
.230-volt operation
is
provided on the rear of the instru-
CAUTION
sure
to
set
the 115-230 volt switch prop-
is
set
to the wrong position.
that
the instrument panel and cabinet
a
three-prong to two-prong adapter and
approximately 50 mv/cm.
full
SCANNING to INTERNAL, TIME SCALE to
DENSITY full clockwise, MODE
list
and inspect the in-
this
facilitate reshipment, keep
single phase,
is
desired, a screwdriver-
be
is
equipped with a three-
of
is
given in paragraph 5-67.
counterclockwise.
manual for ad-
a
power
300
be
to
be
damaged
is
given
the Model 185B. All
source
50
to
watts.
replaced with
used. The
a
60 cps,
if
this
two-contact
as
a
means
Set
vertical
of
rec-
be
are
to
II
FOR
FREE RUN, TIME SCALE MAGNIFIER and CALIBRATOR AND SYNC PULSE OUTPUT switches full
counterclockwise.
d. Center HORIZONTAL POSITION and VERTICAL
POSITION controls.
e.
minutes
vertical plug-in input
f.
pears.
and readjust HORIZONTAL POSITION and VERTICAL
POSITION controls as necessary.
g.
h
mately five cycles of
2-12.
a
2-13. COOLING.
2-14. The Model 185B uses
tem
within the cabinet. The air intake and filter
cated on
ment, choose
of clearance around rear and sides of cabinet..
2-15.
the
be
coated with a
No.
filter
cooling. Refer to paragraph
procedures.
2-16. CABINET MOUNT.
2-17. The Model 185B cabinet mount
instrument.
ated with
plane.
ment to a better viewing angle. Be sure to maintain
clearance required for proper cooling mentioned in
paragraph 2-14.
2-18. RACK MOUNT.
2-19. The standard rack model
rear-support pins and bushings. The bushings
installed on rear flanges. The pins are for use in
mounting the instrument
tended to mate with bushings when instrument
stalled. Do not mount the instrument with only frontpanel screws, particularly
to
any
support pins
securing front panel to rack are not supplied. Install
instrument
USE
Turn the Model 185B on and allow about
for
the instrument
to
Rotate INTENSITY clockwise until a
If
the crt remains
Adjust FOCUS for a
resultant presentation should
The
a
INSTALLATION.
to
maintain tolerable operating temperatures
rear
of instrument. When mounting instru-
a
site
that provides
AIR
air
FILTER. Before operating the Model 185B,
filter,
3
from Research Products Co. In addition, the
should
A
vibration or shock.
located at rear of instrument, should
filter
be
cleaned periodically to insure proper
The instrument
its
front panel in a verticalor near-vertical
bail
is
provided for raising front of instru-
are
shown in figure 2-2. Screws for
as
follows:
to
the 50-mc connector.
50-mc sine wave.
adhesive such
stabilize. Connect the
blank,
press BEAM FINDER
thin,
well-defined
a
forced-air cooling
at
least three inches
as
5-6
for proper cleaning
be
is
is
intended
is
at
installation and are in-
if
installation
Location dimensions for
to
supplied with
trace
approxi-
Filter
a
portable
be
is
two
ap-
trace.
sys-
are
lo-
Coat
oper-
two
are
is
in-
subject
01255-1
2-
1
Section
Paragraphs
II
2-20
to
Model 185B
2-23
(9-1
196A/B
Oscilloscope Camera
187B
Dual
Trace Amplifier
1
lOOA
Delay
Line
Sync Take-Off
11OOA-76A
Delay Line
Resistive Dividers
185A-21C
185A-21D
185A-21E
185A-21F
AC-16W
Cable
Load
Use
A
High quality camera for use in
permanently recording oscilloscope
presentations
~~ ~~
A dual-channel vertical amplifier
(plug-in) for Model 185B (shown
installed in figure
Overcomes inherent oscilloscope delay,
allowing
rate
Inserted
line
sumdied
~~
Termination for
(normally suppliedwith Model
system)
~
These dividers provide a means for
obtaining
capacitance input
Delay
(All
cable terminated in 50 ohms)
Connecting
Dividers
rise
pulses
to
to
between
permit sync
with
a
high-impedance, low-
Line
or other 50-Ohm systems.
dividers must
cable
1-1)
times of slow repetition
be
viewed
signal
Model
llOOA
for 185A-21C/F
line
take-off
llOOA
Delay Line
to
Model llOOA
be
used with
and delay
(normally
system)
llOOA
Features
Image
to
object ratio:
(1:
1
available)
Model 196B allows oscilloscope
photograph on either internal
external
I
Sensitivity 4 mv/cm
Badwidth:
1000 mc
Input
2 pf nominal
Rise Time: 0.25
Delay:
Insertion
6
db for both
Termination Resistance: 50
graticule oscilloscopes
DC
to
Impedance:
120
ne
Loss:
channels
l:O.
9
to
200 mv/cm
800mc ueable
lOOK shunted
M
ohma
or
1
to
by
c
3
Division
Input Res.
250 5:
a
500
2500 50:
5000
Ratio
10:
100:
Max
Inpl
VRMS
1
1
1
1
10
15
35
50
J
186A
Switching Time Tester
I
a.
Fabricate a bracket for rear support pins and
fasten pins
b.
c.
port
2-20.
2-21.
is
available from Hewlett-Packard Company
crease the usefulness of your Model 185B. Additional
equipment
in unit. Refer to the manual for operating information
regarding your particular plug-in or contact your
nearest
this
information.
2-2
in
place.
Install
Lift instrument into place, engaging rear-sup-
pins, and secure front panel firmly to rack.
bracket at
ASSOCIATED EQUIPMENT
Figure
2-1 and
is
available for use with the vertical plug-
@
representative and he
rear
table
Measures switching time of transistors,
diodes and tunnel diodes. Teste pulse
response of active and passive net-
works. Triggers Model 185B in
vance of pulse output. Accessory
universal adapter available.
18. HIGH FREQUENCY STABILITY.
ger stability on HIGH FREQ. position of TRIGGERING.
19.Model 187B
plug-in unit. Not part of
20.SYNC. PULSE OUTPUT. Provides
sync pulse out for triggering
to use as a
21.50 MC. Provides a pulsed 50-mc output.
22. AMPLITUDE DC (OPEN CLRCUIT). Provides
output of four calibrated dc voltages.
23,CALIBRATOR AND SYNC PULSE. Selects
calibrated
to appropriate connectors.
Dual
test
dc
this
switch according
of
trigger signal.
Adjusts
Channel Vertical Amplifier
basic
oscilloscope.
a
test
circuits, or
pulse.
voltages or sync pulse outputs
be
to
trig-
delayed
~~
Figure
3-0 01255-1
3-1.
Operating Controls and Connectors
Model 185B
Paragraphs 3-1
Section
to
111
3-5
SECTION
0
P
ERATI
3-1.
INTRODUCTION.
3-2.
This
tion of all controls in the Model 185B Oscilloscope.
If
more theoretical information
Section
3-3. The vertical amplifier system for the Model
185B includes
for operating the plug-in unit are contained in the in-
struction manual for
3-4.
3-5. Figure 3-1 shows
trols and gives a short description of theiruse. Num-
bers
ure
3-1
dicate operational procedure. More detailed information related to
a. BEAM FINDER. Pressing the BEAM FINDER
reduces gain of horizontal and vertical amplifiers to
a point where a “lost” presentation may
adjusted
HORIZONTAL POSITION and VERTICAL POSITION
controls.
b. SCALE. The SCALE control adjusts intensity
of graticule on those instruments provided withanexternal graticule. On internal graticule instruments,
SCALE aligns trace with graticule.
c. INPUT. This connector
chronizing circuits. These circuits insure an exact
time relationship between input signal and moment of
sampling. To operate properly, the sampling oscilloscope
nal that
supply
step
i).
d. TRIGGERING. The TRIGGERING switch may
be
set to SENSITIVE, NORMAL or HIGH FREQ.
pending on amplitude and frequency of trigger signal.
SENSITIVE position
200 millivolts with a frequency below 100 mc. Do not
exceed 10 volts peak input on SENSITIVE position.
NORMAL position
nals (200 mv to
position should be used for trigger signals of
plitudes above 100 mc. The input circuits on HIGH
FREQ. position
1000 mc
circuits.
Do
peak to INPUT when TRIGGERING
SITIVE. Doing
section contains information on the func-
is
desired, refer
IV,
Principles of Operation.
a
plug-in unit. Detailed instructions
that
particular unit.
FRONT PANEL CONTROLS AND
CONNECTORS.
all
front panel operatingcon-
in figure 3-1 are given
to the photograph and do not necessarily in-
these
to
center of cathode-ray tube using the
either
must
be
is
time-related to the input signal, or must
a
sync pulse to trigger circuit under
is
is
2
volts) below 100 mc. HIGH FREQ.
will
so
they may
not connect any voltage exceeding 10 volts
divide down signals
be
CAUTION
so
may damage inputcircuits.
to
relate the
controls
is
accurately triggered by a sig-
used for trigger signals below
used for higher amplitude sig-
used
to
is
listed
be
the input
operate synchronizing
is
on SEN-
N G I
text
in fig-
below:
found and
to
the syn-
test
all
as
high as
to
(see
de-
am-
111
NSTRU
adjusts frequency of count-down oscillator, enabling
it
either positive-going or negative-going slope
signal.
of input circuits
same point on input signal.
full
sync circuits will
at a 100-kc rate.
quency above
near 100
of the input signal.
triggering stability for repetition
by
PUT connector provides
is
With sync pulse from SYNC PULSE OUTPUT connected
in Model 185B
delaying input signal for the purpose of examining
fast
wave output, synchronized with pulse from SYNC
PULSE OUTPUT. Frequency of sine wave
cycles. This signal
reference on the cathode ray tube, or for checking
time
nector provides dc voltages selected by CALIBRATOR
AND SYNC PULSE switch for calibration of vertical
amplifier in oscilloscope. Voltages selected are accurate to within i3$~ of selected value.
when in
sync pulse output and pulsed 50-mc output to appropriate jacks. On other positions of this switch,
voltage
CUIT) connector
mines time
time calibration of
fications when VERNIER control
clockwise.
fine adjustment between any
TIME SCALE switch, resulting in continuous variation
of time scale from
CTI
0
NS
e.
HIGH FREQUENCY STABILITY. This control
to
lock
in
at a submultiple of input trigger frequency.
f.
TRIGGER SLOPE. The TRIGGER SLOPE switch
may
be
used
to
synchronize Model 185B circuits
of
g. MODE. The MODE control adjusts sensitivity
clockwise
kc
h. STABILITY. The STABILITY control adjusts
varying hold-off time in triggering circuits.
i.
SYNC
PULSE OUTPUT. The SYNC PULSE OUT-
delayed 130 nanoseconds from trigger initiation.
to
circuit under
rise
time.
MC. This connector provides
j.
50
scales
k. AMPLITUDE (DC OPEN CIRCUIT). This con-
m. CALIBRATOR AND SYNC PULSE. This switch,
full
is
supplied to AMPLITUDE (DC OPEN CIR-
n. TIME SCALE. The TIME SCALE switch deter-
p. VERNIER. The VERNIER control provides
so
they
will
trigger reliably
If
MODE control
to
FREE RUN position, Model 1858
free
run,
i.e.
sample automatically
If
they
are
triggered withafre-
100
kc,
they
will
but
will
synchronize with a submultiple
test,
is
overcome. The effect
is
valuable for setting up a time
on oscilloscope.
counterclockwise position, connects both
as
explained in step
scale
in nsec or p sec per centimeter. The
this
switch
operate
a
fast
inherent delay (0.1 psec)
is
two
IO
psec/cm to 3 nsec/cm
at
a
frequency
rates
above 100
rise
time pulse
is
similar
a
pulsed sine
is
50 mega-
k.
correct within speci-
is
full Counter-
steps selected
on
trigger
at
the
is
set
kc
that
to
a
dc
by
a
01255-1
3-1
Section
Figure
III
3-2
Model
185B
1.
Regulated dc output connector
instruments.
2.
Channel A vertical output voltage
deflection present on
187B). Sensitivity 0.2 volt/cm, Graticule
Center:
3.
Channel B vertical
as
item
0
2.
volt.
For
output.
this
use
for
use with future
for
connector (with Model
with
X-Y
Same
characteristics
vertical
recorder.
-
4. OUTPUT TIME BASE SCAN. Horizontaldrive
voltage
with
Sweep
5. INPUT EXT SCAN. Connect
drive
volts/cm, Input Range:
6.
115-230 Volt Switch.
used.
volt operation, 2-ampere slow-blow fuse
230-volt operation (see section
is
available at
X-Y
recorder. Sensitivity:
Start:
0
volt.
signal
to
this
connector. Sensitivity:
0
Use
4-ampere slow-blow fuse
Set
this
connector.
external
to
12
volts.
to
line voltagebeing
VI).
1.2
volb/cm,
horizontal
for
POWER
CORD
For
use
1.2
115-
for
3-
Figure
2
3-2.
Rear-Panel Controls
01255-1
Model
185B
Paragraphs
Section
3-6
to
111
3-7
.
q.
TIME SCALE MAGNIFIER. The TIME SCALE
MAGNIFIER switch
selected by TIME SCALE switch
by TIME SCALE MAGNIFIER switch.
r.
DELAY. The DELAY control selects any part
of unmagnified presentation for magnification;
with delay control you effectively slide cathode-ray
tube along expanded presentation
divides
the value
by
number selected
to
view portion you
that
has
been
i.e.,
want
8.
SCANNING. The SCANNING switch determines
type
of horizontal deflection
point
it
should
beam across the face of the cathode-ray tube
relation
The beam may
of the tube and yet the time
ample,
185B
deflection; they
to
50
provides a choice
(1)
INTERNAL. The time
setting of TIME SCALE and TIME SCALE
MAGNIFIER switches. On INTERNAL, the
beam
of the tube
DENSITY control and sampling frequency.
(2)
MANUAL. On MANUAL position, scanning
accomplished
DENSITY (INTERNAL) control. Manual operation may
sweep
the only visible portion of trace
illuminated by setting of MANUAL SCAN control. The time scale in seconds per centimeter
is
determined again
MAGNIF'IER and TIME SCALE switches. Man-
ual scan
it
is
e.g., when
RECORD. On RECORD position, operation
exactly the same as on MANUAL except
in
very slowly across
time scale determined by setting
SCALE MAGNIFIER and TIME SCALE switches.
RESET OR EXTERNAL. The RESET OR EXTERNAL position
vides a means for resetting scan when oper-
ating on RECORD position,
a
means of scanning with an external signal.
An input connector, INPUT EXTERNAL SCAN,
is
horizontal input,
t.
SCAN (MANUAL) - DENSITY (INTERNAL), The
function of this control depends on setting of SCANNING switch.
NAL, the SCAN (MANUAL)
control adjusts number of samples per centimeter
and hence the density of sample dots
viewer. Reducing scan density
ing up the physical speed of horizontal sweep,
be
emphasized
its
time scale in seconds per centimeter.
take
10
nanoseconds per centimeter. The
are
as
is
automatically swept across the face
at
an actual speed determined
by
be
thought of as a condition where
is
always present on oscilloscope, but
is
useful in
necessary
fast
spikes are present on waveform.
this
case the beam
provided on rear of instrument for external
With
SCANNING switch
that
will
be
used.
At
that
apparent speed of
has
seconds to go across the face
scale
could
be,
for ex-
Model
of
four
follows:
adjusting SCAN -(MANUAL)
to
carefully trace presentation,
has
modes of horizontal
scale
is
determined
is
that
by
setting of TIME SCALE
x-Y
recorder work when
is
automatically swept
the
face of the tube
of
two
functions:
or
-
DENSITY (INTERNAL)
has
the effect of speed-
1)
2)
it
provides
set
to
INTER-
as
seen by the
TIME
it
this
no
by
by
is
part
is
that
at
pro-
i.e.,
-
a
the actual speed
oscilloscope face. However, turning
no effect on time scale in seconds per centimeter on
cathode-ray tube. On the MANUAL position of the
SCANNING switch, the SCAN (MANUAL)
(INTERNAL) control moves dot along presentation
explained in step
EXTERNAL positions of SCANNING, the SCAN
UAL) - DENSITY (INTERNAL) control
3-6.
REAR PANEL CONNECTORS.
3-7.
Figure
and gives a short description of their uses.
figure
the figure to the photograph and do not necessarily
indicate operational procedure. The following paragraphs give more detailed information about
connectors:
nector
sory instruments. The following voltages
able with reference to ground (pin
Channel
terminals for use in driving an
the Model
nector
0
(0.2
20,000
Channel
terminals. Output characteristics of CHAN B are
identical
scan voltage
available at this connector to drive horizontal
of X-Y recorder. The output from OUTPUT TIME
BASE SCAN
and
Source impedance
means of controlling scan with an external signal.
The scan voltages required
supplied by OUTPUT TIME BASE SCAN connectors,
i.e.,
about
HORIZONTAL POSITION centered). Input resistance
varies with setting
minimum
connects Model
voltage setting selected.
fuse
3-1,
a.
5401,
is
intended
1)
Pin b,
2)
Pin c,
3)
Pin d,
4)
Pin
e,
b.
OUTPUT CHAN
A
is
approximately
volt in center, and
volt/cm). Source impedance
ohms.
c.
OUTPUT CHAN
B
to
d. OUTPUT TIME BASE SCAN. The horizontal
+12
volts
e.
INPUT EXT SCAN. This connector provides
0
volt
+13
is
f.
115-230
(see
section VI).
to
produce one complete picture on
this
control
-
has
DENSITY
as
s(2).
On RECORD and RESET
3-2
shows
all
connectors on
the numbers in figure
is
3-2
inoperative.
rear
relate
OR
(MAN-
panel
As
in
text
in
these
regulated dc output connector. This con-
to
supply
dc
power tofuture acces-
are
avail-
a):
+12.6
volts
-12.6
volts
+250
volts
-100
volts
A.
The vertical output from
of the plug-in unit
187B
installed. The output from this con-
+1
-1
B.
of the plug-in unit
those of CHAN
that
has been selected
is
approximately 0 volt at sweep
at
sweep termination
is
approximately
is
available at these
X-Y
recorder with
volt at top of graticule,
volt at bottom of graticule
is
approximately
The vertical output from
is
available
A.
by
at
these
SCANNING
is
axis
start
(1.2
volts/cm).
20,000
ohms.
a
are
the same
to
position scan at left edge of crt, and
volts to position scan at right edge (with
of
time scale VERNIER
32.3K.
volt switch. This switch automatically
185B
power transformer for line
Be
sure
to use correct line
as
but
those
the
01255-1
3-3
Section
Paragraphs
III
3-8
to
3-15
Table
3-1.
Methods
of
Triggering
Model 185B
r
r
3
r
Waveform
Types
Trigger Freq
Over 10
Below
Below 10
Below 10
10
kc
kc
kc
kc
Possible
Trigger Ckt
Measurement?
to
under
no
Ext
Trig
Gen
4vailable
no
---
Delay
iequired
‘
(no,
test
delay
120 ns)
Line
no
Yes
no
yes
if
ckt
>
1
Xefer
Figure
3-
3
3-4
3-5
3-6
to
3-8.
THE
PRINCIPLE
OSCILLOSCOPE.
3-9.
The action of the sampling oscilloscope issim-
llar
to a strobe light that
nization with a turning device, resulting in an apparent slow motion--or the effect of taking moving pictures of a rapidly spinning wheel where
shutter speed
rotation of the wheel, causing the wheel to appear
be turning very slowly, or even turning backwards.
3-10. The sampling of the
oscilloscope
same way. The sampler plug-in in this case would
represent the camera shutter and lens. The sampling
circuit
input voltage
occurs over and over again,
sample taken on a succeeding repetition of input sig-
nal
reference point on input signal.
3-11. THE VIEWING ‘WINDOW”. There
maximum and minimum time during which sampling
oscilloscope may complete each observation. In the
Model 185B, maximum time
imum time approximately
is
and at a slightly later time with respect to same
is
nearly
is
accomplished in almost exactly the
opened for very short periods of time and
at
that
time
OF
THE
SAMPLING
is
slightly out of synchro-
the
the
same as the time for one
signal
is
carefully measured. This
is
0.3
with the sampling
with
each succeeding
is
acertain
100 pseconds and min-
nanoseconds. This time
camera
to
is
referred
120 nanoseconds for the “window” to open after trig-
ger
is
100
pseconds will require special triggering tech-
niques when
3-12. In
cuits must be delayed, or
used to permit viewing leading edge of this
signal. The following paragraphs will describe different methods available for solving some synchronization problems you may encounter in sampling oscilloscope technique.
3-13.
3-14. GENERAL.
3-15.
be
synchronized with a signal
signal received by vertical amplifiers. Furthermore,
the system must
because sampled signal never actually enters oscilloscope circuits, and therefore
ternally for synchronization (in the case
187B, signal
figures 3-3
and
measurement situations, and give recommended
instrument arrangement.
to
as
theviewing“window”. Since
received, signals with a period greater than
it
is
necessary to view rise time.
this
case, either
TRIGGERING THE MODEL
As
in any oscilloscope, the
be
is
sampled at probes). Table 3-1 and
to
3-6
list
the
signal
to
an
advanced trigger must
185B.
Model
that
is
time-related to
externally synchronized.
is
not available in-
common types of waveforms
it
sampling
type
185B must
This
of
Model
takes
cir-
be
of
is
3-4
01255-1
Model 185B
Section
III
Paragraphs 3-16 to 3-21
3-16. TRIGGER REPETITION RATE. Model 185B
accepts triggers
with
repetition
rates
between 50 cps
and 1000 mc. However, an internal hold-off circuit
limits
maximum sampling rate to about 100 kc. STABILITY provides limited control over hold-off circuit
to permit adjustment for maximum stability when
the
trigger rate exceeds 100 kc. For frequencies above
100 mc,
a
countdown circuit (adjusted with HIGH
FREQ. STABILITY) reduces frequency of trigger sig-
nal to approximately 10 mc
cuits
will
be
triggered reliably.
so
that synchronizing cir-
3-17. TRIGGERING METHODS.
is
3-18. When limited time-scale speed
lem, and
of synchronizing Model 185B
in a train, and
the screen. For
signal frequency must
than one pulse
jitter
is
not excessive, the simplest method
is
to
trigger on one pulse
to
view several succeeding pulses on
this
to
be
be
will
occur in the 100 Fsec viewing
possible, however, the
at
least
10
notaprob-
kc
so
that
more
“window” of oscilloscope. The following paragraphs
describe a few conditions that dictate
the
method
of
synchronization, and therefore the instrument setup
that
shculd
MODEL
be used.
1856
MODEL
A
IIOOA DELAY LINE
I1
L
1858
SYNC
185A-76A
DELAYED TRIGGER
TAKE-OFF
CIRCUIT
UNDER
TEST
J
LD-S-
574
Figure 3-4. Viewing Signals Below 10 kc by Using
Delayed Trigger to Drive Circuit under Test
to
fall
within 100 Fsec window.
In
order
to
see
the
leading edge, then, you must trigger oscilloscope
just ahead of an input pulse to allow the leading edge
to
fall within time window.
CIRCUIT
QQQ
185
B12IA SYNC PROBE
UNDER
TEST
L
LO-S-576
Figure 3-3. Viewing Signals Above 10 kc
3-19. SIGNAL REPETITION RATES ABOVE 10
Synchronizing on signals above 10
kc
presents few
problems since one or more pulses of a train may
KC.
be
viewed in oscilloscope “window”. The recommended
instrument setup for viewing signals above 10 kc
shown in figure 3-3, while operating procedures
described in figures 3-7 to 3-9. Remember
signals above 100 mc, TRIGGERING must
HIGH
FREQ., and STABILITY adjusted withbothHIGH
be
that
set
is
are
for
to
FREQUENCY STABILITY and STABILITY controls.
Note
If
it
is
necessary to examine very fast
time (e.g., 10 ns) signals between 10
approximately 100 kc,
it
will not
be
rise
kc
and
possible
using the above method, to magnify presen-
tation sufficiently to examine rise time. It
will
be
necessary in this
case
to resort to
one of the trigger methods given for signals
below 10 kc.
kc
LD
by
-8-173
LD-S-515
Using
SYNC PULSES
Figure
3-
on signals under
will
1100A-76A SYNC TAKE-OFF
DELAY LINE
3-5. Viewing Signals Below
SYNC PULSE OUT
to
Drive Circuit under Test
10
21. Several methods are available for synchronizing
10
kc. Generally the method used
depend on characteristics of circuit under
MODEL
11OOA DELAY LINE
LOAD
1858
test
3-20. SIGNAL REPETITION RATES BELOW
When signal repetition rate
is
below 10 kc,
10
signal
KC.
does not occur frequently enough to allow a full cycle
01255-1
Figure 3-6. Viewing Signals Below 10 kc
when the Circuit under Test Cannot
be
Driven by Synchronizing Pulses
3-5
Section
Paragraphs 3-22
III
to
3-38
Model 185B
and the associated equipment available. Table
and figures 3-4 to 3-6 give three additional basic instrument setups. Each of them
below:
3-22. METHOD ONE (figure 3-4). It
trigger Model 185B from a separate source and then
delay this same signal to trigger the circuit under
measurement
120 ns
An important consideration in this method
is
often possible to tolerate some deterioration of
the driving pulses by the delay line or delay circuit,
since these pulses serve only to trigger the measured
circuit.
with sufficiently fast rise time
triggering, there
presented to oscilloscope.
3-23. METHOD TWO (figure 3-5). You may use the
signal from SYNC PULSE OUTPUT connector to trigger circuit under test. This signal
that
is
scope sweep. See figure 3-12 for basic operating
procedures, If you cannot drive
frequency (100 kc), you can use a repetition rate generator such as
to trigger Model 185B at any rate between 50 cps and
100 kc to produce sync pulses. The sync pulse from
Model 185B
when connected to a 50-ohm load.
3-24. METHOD THREE. There are times when the
circuit under
10 kc and cannot be driven by any type of sync pulses.
This
is
erator
must be used
for instrument setups and figure 3-10 for operating
instructions.
is
(if
inherent delay of measured circuit
or
more, external delay will not
As
long as drive pulses arrive at the circuit
is
no loss of information on signals
suitably delayed from triggering of oscillo-
@
Model
211A
is
normally a +1.5 volt (minimum) pulse
test
operates at repetition rates under
typically the case when a mercury pulse gen-
used. In
this
to
trigger oscilloscope.
case, the signal to be viewed
is
described briefly
is
possible
be
to
provide reliable
is
a
fast
test
circuit at trigger
Square Wave Generator
See figure 3-6
3-1
to
is
required).
is
that
rise
pulse
it
signal fm and jitter become more severe as trigger
repetition rate increases.
maximum fm
trigger signal without affecting the display can
expressed as:
where
The formula indicates
mc trigger, 0.05% fm for a 100-mc trigger.
3-29. For trigger repetition
there
are
situation
should be reliable with up to
mc region and correspondingly
frequencies.
3-30.
3-31. OBSERVATION
3-32. Due
Model 185B screen represents true instantaneous
nal amplitude, useful information may
from signals
3-33. For instance,
occasionally missing,
normally on the crt except that
appear long the base line. The density of dots in
pulse relative to that in the
approximate percent
If
self
base
50% of the time, pulse and base line
same.
PULSE ANALYSIS.
the pulse
will
line will be continuous. If the pulse
or
jitter
Maximum fm
f
=
trigger repetition rate in mc
two
count-down circuits in series, and the
is
more complex. However, triggering
to
the
fact
that
are not 100% periodic.
if
of
is
missing most of the time, the pulse
be
represented
As
a
general guide, the
which can
a
maximum of 5% fm for
OF
INTERMITTENT PULSES.
that
one pulse in a pulse trainis
that
pulse will
be
present in the
=
5/f
a
rate8
above 100 mc,
0.05%
fm in the 100-200
less
fm at higher
each sample plotted on
be
a
be
series of dots
sig-
obtained
displayed
will
the
base
line
will
indicate the
time
that
the pulse
by
a
series of dots, while
is
will
missing.
is
appear the
the
present
be
1-
it-
3-25. The signal, decreased in amplitude by any attenuation in the resistive sync probe,
185A-76A sync take-off where
Half
the
signal
is
used directly to trigger Model 185B,
while the other
in unit. When using method three, remember that
the resultant vertical calibration
of SENSITIVITY setting times probe attenuation
times
2.
3-26. The Model
approximately
nals.
If
a
coil of 3/4 inch
mately 105 feet long to provide about 120-ns delay.
3-27.
3-28. When trigger repetition rate
each trigger actuates
jitter in trigger signal have no effect upon display
(provided there
being viewed).
100 kc and 100 mc, the internal hold-off circuit of
Model 185B comes into play, and
3-6
EFFECTS OF
half
llOOA
1
gc, which
still
wider bandwidth
is
no
For
is
delayed and
delay line
is
sufficient for most sig-
is
or
larger Styroflex cable approxi-
FM
AND JITTER.
a
sampling cycle, and fm and
jitter
between trigger andsignal
trigger repetition rates between
is
fed
to
Model
it
is
divided in
fed
to
vertical plug-
will
be the product
has
a
passband of
desired, you can use
is
below 100kc,
effects
half.
of trigger-
3-34. Conventional oscilloscopes indicate missing
pulses by allowing
brightness of wave compared
line indicates relative frequency of occurrences. You
can see that presentation in the form of relative num-